To run the installation open the Upgrade Utility executable file, when prompted click on
Setup.
After clicking Next on the first window you will have the option to install the program to
the default location (C:\Program Files\Olympus NDT\Upgrade2010\) or to select your own.
Once you have made this choice click Next and then click Next again on the following
screen.
Towards the end of the installation the below window will pop up make sure you click
Continue Anyway.
Olympus NDT
48 Woerd Ave, Waltham, MA 02453-3826 USA
Tel: 781-419-3900 Fax: 781-419-3980
www.olympus-ims.com
After this it will say that that Upgrade2010 has been successfully installed click on Close
Click on Open Device button. Once the program connects to the unit a message
will appear The gage 38DL PLUS is found.
Olympus NDT
48 Woerd Ave, Waltham, MA 02453-3826 USA
Tel: 781-419-3900 Fax: 781-419-3980
www.olympus-ims.com
The next step is to load New Bin Files this is done by clicking on New Bin Files
button and then browsing to the folder (probably on the desktop) that the upgrade Bin
files were saved. A dialog box will come up and say 4 File(s) Copied click OK.
Make sure the Monitor, Application and FPGA check boxes are selected then click
on the Upgrade Grade button and the software will begin upgrading the gage. Once
it is done the program will ask you to turn the 38DL PLUS off then back on and click
[OK] button. This will allow upgrade2010 to upgrade the language file.
After completion click on the EXIT button. Power the gage on and off. On startup
the unit will display the latest version of 38DL Plus Software.