Beruflich Dokumente
Kultur Dokumente
The Getinge Assured Test Swab provides a simple, reliable and rapid way to verify cleanliness and
measure microbial contamination on the surface of surgical instruments, flexible endoscopes, examination
rooms, operating rooms, restrooms, waiting rooms anywhere that can harbor possible contaminants.
When ATP is picked up by the tip of the Test Swab and brought into contact with the unique luciferase/
luciferin reagent in the Test Swab tube, energy is emitted in direct proportion to the amount of ATP
present. The Test Swab is then placed in the Getinge Assured SafeStep hand held unit where it measures that energy in the form of light
which is quantified by using Relative Light Units (RLUs), and displays the level of contamination present in just 15 seconds. The Getinge
Assured Quality Assurance System will detect 1/100,000th dilution of fresh blood, i.e. 0.2mg protein per swab.
The hand held unit can then be synced to a PC where the reading is downloaded to an easy-to-use database management software for
tracking the results. The software program can be used to run comparison charts and produce color trend analysis graphs and reports.
The Getinge Assured SafeStep Quality Assurance System has been independently lab validated. Testing data is available upon request.
Features
All-in-one sampling device Unique liquid-stable reagent
12-month shelf life
at refrigerated temperatures Swab tube
(36F to 47F) (2C to 8C)
4 week shelf life at room Swab with
Easy-to-activate
temperatures (70F) (21C) snap valve
premoistened
extractant
Unique liquid-stable reagent
Tolerant to temperature abuse
and sanitizers
page 1
Directions for Use - Test Swab
It is recommended that testing take place prior to high-level disinfection or sterilization.1
Follow the manufacturer's IFU for reprocessing of devices.
1. Take the Test Swab out of the tube and swab area of interest.
2. Place the Test Swab into the tube. Hold the swab tube firmly and use the thumb and forefinger to break the Snap Valve by bending
the bulb forward and backward. Squeeze the bulb twice, expelling all liquid down the swab shaft. Saturate the swab tip in liquid by
gently shaking for 3 seconds.
3. Insert the Test Swab tube into the Getinge Assured SafeStep Quality Assurance Hand Held Device, close and press the OK button.
In 15 seconds the hand held device will display the amount of ATP detected.
4. The Hand Held Device can then be synced to your computer and the reading downloaded to the software program provided. The
downloaded data can then be used to produce detailed reports that can be used as a part of the facility's QA system by providing
testing history on the effectiveness of cleaning procedures.
1 2 3 4
If the reading is greater than 100, the swabbed/tested area should be considered dirty and re-cleaning and re-testing is recommended.
Process should be repeated until swabbed/ tested area shows a pass.
If you are having repeated test failures on the same device please contact the device manufacturer for additional information on the best
cleaning practice and products available that may help to produce better outcomes.
Other Uses for the Getinge Assured SafeStep Quality Assurance System:
The Getinge Assured SafeStep Quality Assurance System, used in conjunction with Getinges Test Sponge, is the only product that can
provide an accurate numerical reading of the level of contaminants present inside the channel of scopes and cannulated instruments after
cleaning. Please contact your Getinge representative for more information.
Distributed by:
Getinge USA, Inc. Getinge Canada, Ltd. 1
Please note: Testing after HLD is an option for periodic
1777 E. Henrietta Road 90 Matheson Blvd. West Suite 300 testing of HLD processor maintenance. This would be
up to the facility to determine procedure and frequency.
Rochester, New York 14623 Mississauga, ON L5R 3R3
www.getinge.com
1015