BRAWN LABORATORIES LIMITED

STANDARD OPERATING PROCEDURE


SOP No. RAW MATERIAL STORE CEPHALOSPORIN BLOCK Page No.
Title: Daily Verification of Weighing Balance in
BLL/STR/003/05 1 of 3
Cephalosporin Section
Effective Date 08-02-2011 Review Date 07-02-2013

1.0 OBJECTIVE :
To lay down the procedure for daily verification of all weighing balance in Cephalosporin
section store.
2.0 SCOPE :
This SOP is applicable for Store Cephalosporin for verification of all weighing
Balance, at Brawn Laboratories Ltd., Faridabad.

3.0 RESPONSIBILITY :
Stores Supervisors
4.0 DISTRIBUTION :
Master Copy : Quality Assurance Department
Control Copy : Store Department
Control Copy : Quality Assurance Department

5.0 PROCEDURE :
5.1 Select the standard calibrated weights for various balances based on the capacity and
operating range of the balance as mentioned in 5.7 and check the calibration status.
5.2 Before verification the balance, clean the weighing platforms and pans with dry mop.
Ensure that the balance rests evenly, which can be ensured with the help of sprit level.
5.3 Adjust the balance to zero.
5.4 Place the calibrated standard weight one by one on the weighing platform/pans at the *
marked places shown in the following diagram and check the readings. The different should not

Prepared by Reviewed by Approved by

Designation Officer-Store Department Head. Manager QA


Name Vibhor Kumar Pritam Singh K.S. Rao
Sign / Date
BLL/STR/003/F01-00
BRAWN LABORATORIES LIMITED
STANDARD OPERATING PROCEDURE
SOP No. RAW MATERIAL STORE CEPHALOSPORIN BLOCK Page No.
Title: Daily Verification of Weighing Balance in
BLL/STR/003/05 2 of 3
Cephalosporin Section
Effective Date 08-02-2011 Review Date 07-02-2013

be more then specified least count. If it is stable at all working points record the point E
observation in annexure.

A B
* *
E
*
* *
D C

5.5 If it conforms to acceptance criteria, affix the status label O.K with date.
5.6 Always use standard weights and balance, which are certified by Weights and Measures
department.
5.7 Selection of Standard Weights for various balances.
--------------------------------------------------------------------------------------------------------
Balance Operating
Capacity Range Standard Weights to be used Identification
No.
--------------------- -----------------------------------------------------------------------------------

5.8 If it does not conform to acceptance criteria, maintenance department shall be inform for
corrective measures and a status label under maintenance shall be affixed.

6.0 ABBREVIATIONS :

O.K. : All correct

Prepared by Reviewed by Approved by

Designation Officer-Store Department Head. Manager QA


Name Vibhor Kumar Pritam Singh K.S. Rao
Sign / Date
BLL/STR/003/F01-00
BRAWN LABORATORIES LIMITED
STANDARD OPERATING PROCEDURE
SOP No. RAW MATERIAL STORE CEPHALOSPORIN BLOCK Page No.
Title: Daily Verification of Weighing Balance in
BLL/STR/003/05 3 of 3
Cephalosporin Section
Effective Date 08-02-2011 Review Date 07-02-2013

7.0 REFERENCES:
NIL

8.0 ANNEXURE :
Annexure I Balance weigh check record

9.0 REVISION / REVIEW CARD:

Sr. Date Details of Reason for Issue Revision Approved


No. Revision/ Revision/ No. No. By/Date
Review Review

1. 08-02-2011 Change in format ------ 06 05

Prepared by Reviewed by Approved by

Designation Officer-Store Department Head. Manager QA


Name Vibhor Kumar Pritam Singh K.S. Rao
Sign / Date
BLL/STR/003/F01-00