Beruflich Dokumente
Kultur Dokumente
Chapter 5
Copyright 1996 Soil Science Society of America and American Society of Agronomy, 677 S.
Segoe Rd., Madison, WI 53711, USA. Methods of Soil Analysis. Part 3. Chemical Methods-SSSA
Book Series no. 5.
91
92 SOLTANPOUR ET AL.
Bradford & Bakhtar, 1991); use of computer programs such as orthogonal poly-
nomials (Hassan & Loux, 1989), simplex optimization (Belchamber et aI., 1986),
and that recommended by Taylor and Schutyser, 1986, for optimization of spec-
trometer operating conditions and automatic correction for spectral interferences;
and compilation of ICP emission lines (McLaren & Berman, 1985; Boumans,
1984; Parsons et aI., 1980).
The ICP-MS technique has been developed over the last 20 yr, particular-
ly by Houk et ai. (1981) who showed suprathermal ionization in an ICP Ar plas-
ma. Within about the last 10 yr it has become possible to apply the method to
routine analytical concentration determinations. Several review articles docu-
ment the ICP-MS developmental milestones (Beauchemin, 1989; Hieftje &
Vickers, 1989; Douglas, 1989; Houk & Thompson, 1988; Houk, 1986; Gray,
1985; Douglas & Houk, 1985). Between 1986 and 1988, ICP-MS enjoyed a
surge of popularity. According to Cresser et ai. (1988), the late AR. Date attrib-
uted the success of ICP-MS to spectral simplicity, very high sensitivity, and iso-
tope ratio capability. Cresser attributes the following quote to Date (1986), "ICP-
MS is the greatest thing to happen to atomic spectroscopy since chopped light."
Each year since 1986, papers dealing with ICP-AES and ICP-MS analysis of
environmental samples including soil samples have been reviewed by Malcolm
S. Cresser, and coworkers (Cresser et aI., 1986, 1988, 1989, 1990, 1991, 1992;
~bdon et aI., 1987). Another source of information for the ICP-MS analysis of
geological and inorganic materials is the biennial review publication appearing
in Analytical Chemistry (Jackson et aI., 1989, 1991). The ICP Information
Newsletter publishers an ICP bibliography each January (Barnes, 1992) and-
like the Cresser review-abstracts papers on ICP-MS presented at national and
international conferences.
An inorganic mass spectrometry, ICP-MS, and x-ray fluorescence spec-
trometry review has been published yearly since 1988 (Ure et aI., 1988; Bacon
et aI., 1989, 1990, 1991). A text on applications of ICP-MS was edited by Date
and Gray (1989), and a compilation of 21 selected papers from the Second
International Conference on Plasma Source Mass Spectrometry held at Durham
University, 24 to 28 Sept. 1990, was edited by Holland and Eaton and published
as a bound volume (1991). A comprehensive book on ICP-MS and its applica-
tions has been written by Jarvis (1992).
Isotopes of 71 naturally occurring elements can be monitored using con-
ventional positive ion, solution nebulization ICP-MS. Accuracies of theconcen-
trations estimated using these measurements at the Division of Agriculture and
Natural Resources (DANR) Analytical Lab at the University of California,
Davis, using internal standard corrections, are typically within 2.5% of the true
concentrations in favorable cases. For about 70% of these elements, more than
one stable isotope occurs in nature. Thus, analyses can be done for them using
isotope ratios and/or isotope dilution. Isotope ratios show precision of 0.1 to
0.3% (Gregoire, 1989). Concentrations calculated using the method of isotope
dilution (Fasset & Paulsen, 1989) are generally within 1% of their true concen-
trations; a higher accuracy and precision than ICP-MS analyses done without the
use of stable isotope addition (Viczian et aI., 1990; Van Heuzen et aI., 1989,
Garbarino & Taylor, 1987; McLaren et aI., 1987; Dolan et aI., 1990).
ICP EMISSION SPECTROMETRY 93
Concentrations for 13 other nuclei that are not naturally occurring also can be
estimated using the ICP-MS (Brown et aI., 1988; Igarashi et aI., 1990; Kim et aI.,
1989a,b, 1991).
While elemental coverage and detection limits under relatively ideal con-
ditions are excellent, there are some problem areas in ICP-MS that must be
investigated before deciding whether or not the ICP-MS technique "will work for
you (Hieftje, 1992)." While most of the following problems have been overcome
or circumvented to meet analytical needs in selected instances, the statements
that follow are generally valid for a generic, normal resolution (i.e., peak widths
between 0.5 and 1.0 dalton), normal aqueous aerosol generation ICP-MS:
1. Although the ICP-MS has been found to be satisfactory for soils work,
compared to ICP-AES the accuracy and precision of ICP-MS data are approxi-
mately three times worse. However, for concentrations determined from isotope
dilution/ratio measurements, precision and accuracy is somewhat better
(Gregoire, 1989; Dolan et aI., 1990; Taylor et aI., 1991).
2. Isobaric overlaps (spectral interferences) occur with some regularity for
elements between approximately 28 to 80 daltons, and do occur throughout the
mass range. They are a result of a common unit mass shared by more than one
element, doubly charged ions overlapping an isotope with half the unit mass of
the doubly charged species (Vaughan & Horlick, 1986), elemental oxide, ele-
mental hydride, and/or elemental hydroxide ions overlapping isotopes of other
elements (Vaughan & Horlick, 1986; Munro et aI., 1986; Date et aI., 1987; Gray,
1986), and background spectral problems (Vaughan & Horlick, 1986; Gray,
1986; Tan & Horlick, 1986). The isobaric interferences involving O2 can be
eliminated using techniques such as electrothermal vaporization (ETV) atomiz-
ers or Laser ablation sample aerosol production (Gregoire, 1989).
3. Ion response is significantly suppressed by matrix elements. The thresh-
old matrix element values are low compared to emission suppressions noted for
ICP-AES. Nonspectroscopic interferences result from excessive dissolved solids
in the test solutions. For a number of reasons, the analyte ion arrival rate at the
detector (i.e., analyte response) is suppressed under these circumstances
(Beauchemin et aI., 1987; Olivares & Houk, 1986; Douglas & Kerr, 1988;
Gregoire, 1987a,b; Hieftje, 1992). Although the onset of suppression is usually
observed in the neighborhood of 100 to 500 mg L-l at the DANR Analytical Lab,
Gregoire (1989) indicated somewhat higher levels using the same instrument
model/manufacturer (i.e., Perkin-Elmer SCIEX 250).
4. The ICP, generated in Ar with normal aqueous solution nebulization,
may be unable to produce measurable amounts of positive ions for some analytes
that could be of interest, e.g., F, CI, and/or S. However, the halogens can be deter-
mined in the negative ion mode (Hieftje et aI., 1988; Chisum, 1992). Sulfur can
be detected if the water is removed from the sample prior to injection of dry
aerosol into the ICP and/or a high resolution ICP-MS is available (Douhitt,
1994a,b). Some water vapor can be removed from the sample aerosol using a
cooled spray chamber (Hutton & Eaton, 1987). Water can be completely sepa-
rated from the S using an electrothermal atomizer (Gregoire, 1989) or partially
removed using nebulization-desolvation equipment (Veillon & Margoshes,
1968).
94 SOLTANPOUR ET AI...
Instrumentation
Inductively Coupled Plasma Generation
The ICP is produced by passing initially ionized Ar gas through a quartz
torch located inside a Cu coil connected to a radio frequency (RF) generator. The
RF generator provides up to 3 kW forward power (in most commercial units) at
a frequency of 27.1 MHZ. "The high-frequency currents flowing in the Cu coil
generate oscillating magnetic fields whose lines of force are axially oriented
inside the quartz tube and follow elliptical closed paths outside the coil," Fassel,
1977. The Cu coil-torch assembly is shown schematically in Fig. 5-1 (Fassel &
ICP EMISSION SPECTROMETRY 95
E
E
...J
5 TEMPERATURE (K)
u ( IO%)
a
<{
25 6000
0
...J 20 6200
w
>
0
15 6500
CD 6800
<{ 8000 (ESTIMATE)
..... 10000
:J:
(!)
iii
:J:
SAMPLE AEROSOL
Fig. 5-2. Temperatures in the plasma as measured by the spectroscopic slope method (reproduced
from Fassel, 1977, in Pure and Applied Chemistry, with permission obtained from Pergamon Press
Ltd., Oxford).
Kniseley, 1974). Electrons and ions passing through the oscillating electromag-
netic field flow at high acceleration rates in closed annular paths varying in their
direction and strength, resulting in electron acceleration on each half cycle.
Collisions between accelerated electrons and ions, and ensuing unionized AI gas
cause further ionization. The collisions cause ohmic heating and, when measured
spectroscopically, give thermal temperatures ranging from 6000 to 10 ooooK
(Fig. 5-2; Fassel, 1977).
The quartz torch has three concentric channels. The outer channel conducts
AI gas at about 15 to 17 L min-I to the plasma to sustain the plasma and to iso-
late the quartz tube from high temperatures. The innermost channel is for intro-
duction of sample into the plasma. The middle channel conducts the auxiliary AI
gas at about 1 L min-I and is used in ICP-AES only when starting the plasma or
for organic samples, and is routinely used for all types of samples for ICP-MS
(Fig. 5-3; Fassel & Knisely, 1974). The ICP has an annular, or donut, shape
when it is viewed from above. The hole has a lower temperature than the donut
and, once the flow of vaporized sample carried in a flow of nebulizer argon is
established, offers less resistance to the sample injection. The sample is injected
into the plasma by using Ar carrier gas at a rate of about 1 L min- 1 for ICP-AES
work. For ICP-MS work the aerosol flow is approximately 1.5 L min-I.
begins in the sample aerosol channel inside the load coil for ICP-AES. The IRZ
extends upward to 1 or 2 mm above the load coil, taking on the appearance of an
amber "bullet" during nebulization of soil, plant and water samples. The CaO
molecules on the surface of the "bullet" emit an amber colored radiation which
changes to a deep blue- or purple-colored radiation further downstream as emis-
sion from Ca atoms and ions dominate. The blue/purple region is termed the nor-
mal analytical zone (NAZ), and is the region in which the analyte emission is
observed by the spectrometer. Color photographs illustrating the appearance of
the IRZ and NAZ while nebulizing an elevated concentration of Y into an ICP
have recently been published for ICP-AES (Winge et ai., 1988), which clearly
define these critical regions. The NAZ is 15 to 20 mm above the coil, or about
14 to 19 mm above the tip of the IRZ, in an environment relatively low in back-
ground emission.
The background emission consists of Ar lines and some weak band emis-
sion from OH, NO, and CN molecules present in the plasma (Ward, 1978a). By
the time the decomposition products of the sample reach the NAZ, they have had
a residence time of about 2 ms at spectroscopically measured temperatures rang-
ing from about 8000 to 5000 K (Fassel, 1977). The residence time and temper-
0
ature experienced by samples introduced into the plasmas are about twice as
large as those in the hottest flames, e.g., N20-C2H2. The high temperature and
residence time combination, at the sample aerosol flow rates typical for the ICP-
AES, lead to complete sample vaporization and atomization in contrast to flames
that require releasing agents for refractory compounds (Larson et ai., 1975).
Once the free compounds, atoms, and ions are formed in ICP-AES, they are in a
chemically inert environment in contrast to highly reactive combustion flame
environments. Ionization interferences are generally negligible in ICP-AES.
Self-absorption (a phenomenon responsible for the flattening of the standard
curve at high analyte concentrations) is practically absent, which leads to a wide
linear dynamic analytical range of three to five decades. No sampling or skim-
mer cones are used in the ICP-AES, and therefore contamination from them is
absent.
ICP EMISSION SPECTROMETRY 97
(Lim et aI., 1989). Although the position of the sampler with respect to the
extended IRZ of the ICP is adjusted before each analytical run to achieve maxi-
mum ion arrival rate at the detector, it also is sampling aerosol that has under-
gone solute vaporization and atomization reactions outside the hottest regions of
the ICP. This is thought to contribute to the appearance of more molecular ions
in the mass spectra and higher susceptibility to nonspectroscopic matrix effects
than if the aerosol flow rate and/or auxiliary Ar flow rate could be slowed down
enough to put the IRZ back to within 1 or 2 mm of the downstream side of the
load coil. However, this is not possible because of the arcing that occurs between
the load coil and the metallic sampling cone in instances in which the cone is
placed too close spatially to the load coil.
In summary, the NAZ is much closer to the tip of the IRZ in ICP-MS (~1O
mm) than the NAZ is to the tip of the IRZ in ICP-AES (14-19 mm). The closer
proximity used for the ICP-MS measurements increases the concentration of ions
to a level at which they are analytically useful; but, to prevent arcing between the
load coil and the metal sampler cone, high flow rates for the nebulizer and/or
auxiliary gases must be used to push the IRZ tip away from the load coil. Thus,
analyte solute vaporization and atomization occur in a region of the plasma with
98 SOLTANPOUR ET AL.
Load Coil
Fig. 5-4. Spatial nomenclature for ICP (adapted from Koirtyohann, 1980), where IRZ denotes the
Initial Radiation Zone. In agronomic work, it likely will appear red to pink in color due to emis-
sion of CaO and CaOH molecules. NAZ denotes Normal Analytical Zone, the location of the NAZ
depending on whether analytical measurements are being perfonned for AES (A) or MS (B).
tions into the spray chamber. The Babington requires a pump to deliver the solu-
tion to a pinhole orifice from which Ar gas is emerging at high velocity. The
USN also requires a pump to deliver the solution, this time to a vibrating plate.
There are two common types of pneumatic nebulizers: cross flow and con-
centric. For the cross flow, as the solution emerges from the rigid capillary tube
carrying the sample solution, another tube positioned at a right angle blasts Ar
past it to shear off fine aerosol particles. The cross-flow nebulizers are often
made of highly corrosion-resistant capillary metal tubes, e.g., Pt-Ir alloy. One
capillary carries Ar at approximately 1 L min- 1 and the other, sample solution.
The orientation of the tips is fixed by the manufacturer, and may include a sap-
phire edge at the tip of the solution tube to produce a fine, uniform mist out of
approximately 10% of the solution drawn in. The cross-flow systems in the
authors' laboratories have held up to the most demanding applications for 2 or 3
yr with no sign of degradation.
The concentric flow (Meinhardt type) glass nebulizers are routinely used
at the DANR Analytical Lab for both ICP-AES and ICP-MS work. These are
made entirely of glass in a "T" type configuration. The main barrel of the nebu-
lizer consists of a fine glass tube tapered to capillary size. The capillary portion,
measuring 2.5 cm. in length, carries the sample solution and is through a tube
joined in a "T" shape to this barrel. The Ar pressure can be anywhere from 124
kPa to 345 kPa (18-50 psi), resulting in flow rates of between approximately
0.75 to 1.5 L min- 1. The open ends of the argon tube and the capillary tube meet
at a taper, and a fine mist is produced as the Ar (flowing concentrically around
the capillary) shears off small fragments of water droplets at the capillary tip.
These nebulizers are very steady and, like the cross-flow type, produce aerosol
from about 10% of the solution going through the tip.
The cross-flow and concentric nebulizers easily clog with high salt solu-
tions. Soltanpour et a1. (1979a) treated 1 M NH4HC0 3-O.005 M DTPA (diethyl-
enetriaminepentaacetic acid) soil extracts with 0.5 M HN0 3 to overcome clog-
ging. However, the Colorado State University Soil Testing Laboratory
(CSUSTL) presently uses a Legere 1 teflon nebulizer (Legere & Burgener, 1985)
attached to a peristaltic pump; eliminating the need for acid pre-treatment.
Wolcott and Butler (1979) designed a pneumatic nebulizer that could aspirate
solutions containing up to 36% suspended solids. To overcome differences in
surface tension, density, and viscosity, the analyst can use a peristaltic pump to
introduce sample solutions into the nebulizer (Beasecker & Williams, 1978). For
concentric nebulizers, care must be taken to eliminate small insoluble particles
from test solutions that would otherwise clog the capillary. If a particle becomes
lodged in the capillary or between the capillary and the tapered tip, then great
care must be exercised while removing the blockage to avoid breaking the frag-
ile glass tubing. One method is to carefully remove the nebulizer from the Ar and
sample delivery tubes and squirt acetone from the nebulizer tip into the barrel
while tapping with a finger, and then force Ar through the tip into the barrel
while continuing to tap2.
1 Distributed by Burtec Instrument Corporation, Post Office Box 235, Delmar, NY 12054.
2 Petrie, K., Precision Glassblowing of Colorado, Englewood, CO.
100 SOLTANPOUR ET AL.
and the portion of the sample to be ablated is located within the ocular of a light
microscope. The specimen is focused using the X-Y-Z movement of the sample
stage. The computer is notified of impending analysis, and the LASER fired.
Preablation times, LASER repetition rates and LASER power per pulse are some
of the more important variables.
The detection limits of metals in the solid are usually less than 1 mg kg-1.
The dry sample aerosol produced by the laser bursts is free of many of the recom-
bination polyatomic ions that would ordinarily accompany a major element (M)
in a nebulized sample (e.g., MO+, MW, MOH+, see Date et al., 1987). Argide
polyatomic ion species, e.g., MAr+, may persist however. Sample analysis rate
can be rapid, but depends on the analytical objectives and the variability between
samples. The accuracy of the analyses are highly dependent on the availability of
certified materials of composition similar to the sample. At ultralow concentra-
tions, memory effects must be taken into consideration. For example, assume a
Au nugget is to be ablated to determine approximate elemental composition. On
the next sample an elemental assay is requested on a metallic inclusion in a piece
of quartz for Au content. To reduce the Au background between the two samples,
the entire ICP-MS system should be shut down to permit thorough cleaning of
the sample and skimmer cones, the ICP torch, and aerosol carrier line from the
LASER stand, and the interior of the glass sample stage cover. Cleaning the glass
sample stage cover is probably the most critical because the interior of the
LASER ablation window becomes coated with a metallic film of elemental com-
position generally representative of the ablated sample, and reablation of the film
can occur during ablation on subsequent samples. Thus, for the analysis problem
at hand, the total analysis time can be a few minutes or a few hours depending
on whether the quartz piece can be run ahead of the Au nugget, and, more gen-
erally, what the detection limit and accuracy requirements are.
Direct Injection Nebulizers. Direct Injection Nebulizers (DIN) provide
for the direct injection of microvolume aqueous liquid samples into the base of
torch plasma using fused silica capillary tube and a high pressure HPLC type
pump. It can be utilized on either ICP-AES or ICP-MS instrumentation. Direct
injection nebulizers is uniquely suited for the determination of nebulizer memo-
ry-prone mass isotopes of B, Hg, I, C, S, and Br or where sample volume is lim-
ited. Studies of Smith et al. (1991) have shown it capable of detecting as little as
1 nglg of B in biological materials using ICP-MS. Powell and Boomer (1995)
have shown that the technique under optimal conditions accurately capable of
detecting Cr at the 30 ngIL concentration range for multiple Cr(lll) and Cr(VI)
species. In addition, the technique provides the fast sample wash-out and high
sample through-put.
Spectrometers
Atomic Emission Spectrometry. Atoms of elements in a sample when
excited emit light of characteristic wavelengths with an intensity directly pro-
portional to the element concentration. The light is focused on the entrance slit
of the spectrometer to illuminate the diffraction grating. The diffraction grating
separates light into its component wavelengths of lines (spectrum). The spectral
line of an analyte passes through the aperture of an exit slit and strikes a photo-
ICP EMISSION SPECTROMETRY 103
Analytical Capabilities
Selection of Wavelength
The number and wavelengths of spectral lines generated by de-excitation
of atoms depends on their electron configuration. The theory and explanation of
atomic emission make for exciting reading but are beyond the scope of this dis-
cussion. Those wishing to explore spectral theory more thoroughly can read
Boumans' (1966) book on the subject.
ICP EMISSION SPECTROMETRY 105
For the analyst using the spectrometric technique, line selection becomes a
matter of finding the most useful line, sufficiently intense to be easily detected
with a minimum of spectral interference from other spectral lines and back-
ground. Line selection can be a difficult process requiring careful examination of
the spectrum. In some instances, the most useful lines may lie outside the spec-
tral range of the spectrometer or fall in areas of high background. If an analyst
wants to measure the concentration of an element such as S that emits the most
suitable radiation for analytical purposes in the ultraviolet (UV), then that ana-
lyst should have a vacuum or Ar-purged spectrometer available to him because
much of the radiation in the UV spectrum, including UV emission lines of S, is
efficiently absorbed by 02.
For some elements, only one or two useful lines are available, whereas
other elements offer several useful lines.
Winge et at. (1979) determined the relative intensities of atomic and ionic
lines of elements excited in ICP. This information is partially reproduced in Table
5-1.
Table 5-1. Prominent lines of the elements emitted by the ICP. The list is arranged alphabetically
with the lines of each element listed in order of decreasing I..JIb ratio (adapted from Winge et aI.,
1979).
Estimated
*
detection
Element S of It Wavelength I..JIb Concentration Iimi~ Comments#
nm l1g/mL
Ag I 328.068 38.0 10.0 0.007
I 338.289 23.0 10.0 0.01
II 243.779 2.5 10.0 0.1
II 224.641 2.3 10.0 0.1
II 241.318 1.5 10.0 0.2
II 211.383 0.9 10.0 0.3
II 232.505 0.7 10.0 0.4
II 224.874 0.6 10.0 0.5
II 233.137 0.5 10.0 0.6
AI 309.271 13.0 10.0 0.02 OH band, NRtt
309.284 13.0 10.0 0.02 OHband, NR
396.152 10.5 10.0 0.03
237.335 10.0 10.0 0.03 NR
237.312 10.0 10.0 0.03 NR
226.922 9.0 10.0 0.03 NR
226.910 9.0 10.0 0.03 NR
308.215 6.6 10.0 0.04 OHband
394.401 6.3 10.0 0.05
236.705 5.8 10.0 0.05
226.346 5.0 10.0 0.06
221.006 4.8 10.0 0.06
257.510 4.0 10.0 0.08
As 193.696 56.0 100.0 0.05
197.197 39.0 100.0 0.08
228.812 36.0 100.0 0.08
200.334 25.0 100.0 0.1
189.042 22.0 100.0 0.1
(continued on next page)
106 SOLTANPOUR ET AL.
nm llg!mL
234.984 21.0 100.0 0.1
198.970 16.0 100.0 0.2
200.919 6.1 100.0 0.5
278.022 5.7 100.0 0.5
199.048 5.5 100.0 0.5
B 249.773 63.0 10.0 0.005
249.678 53.0 10.0 0.006
208.959 30.0 10.0 0.01
208.893 25.0 10.0 0.01
Ba II 455.403 230.0 10.0 0.001
II 493.408 130.0 10.0 0.002
II 233.527 75.0 10.0 0.004
II 230.424 73.0 10.0 0.004
II 413.066 9.1 10.0 0.03
II 234.758 7.8 10.0 0.04
II 389.178 5.2 10.0 0.06 H 388.905
II 489.997 3.7 10.0 0.08
II 225.473 2.0 10.0 0.2
II 452.493 1.9 10.0 0.2
Be II 313.042 110.0 1.0 0.0003 OHband'
I 234.861 96.0 1.0 0.0003
II 313.107 41.0 1.0 0.0007 OHband
I 249.473 8.0 1.0 0.004 Group NR**
I 265.045 6.4 1.0 0.005 Group NR
I 217.510 2.5 1.0 0.01 NR
I 217.499 2.5 1.0 0.01 NR
I 332.134 1.4 1.0 0.02 Group NR
I 205.590 0.7 1.0 0.04 NR
I 205.601 0.7 1.0 0.04 NR
Bi I 223.061 87.0 100.0 0.03
I 306.772 40.0 100.0 0.08 OHband
I 222.825 36.0 100.0 0.09
I 206.170 35.0 100.0 0.09
I 195.389 14.0 100.0 0.2
I 227.658 12.0 100.0 0.3
II 190.241 10.0 100.0 0.3
I 213.363 10.0 100.0 0.3
I 289.798 9.0 100.0 0.3
I 211.026 7.8 100.0 0.4
Ca II 393.366 89.0 0.5 0.0002
II 396.847 30.0 0.5 0.0005 H 397.007
II 317.933 1.5 0.5 0.01 OHband
I 422.673 1.5 0.5 0.01
Cd II 214.438 120.0 10.0 0.003
I 228.802 110.0 10.0 0.003
II 226.502 89.0 10.0 0.003
I 361.051 1.3 10.0 0.2
I 326.106 0.9 10.0 0.3
I 346.620 0.7 10.0 0.4
I 231.284 0.5 10.0 0.6
nm JlglmL
II 263.105 19.0 10.0 0.02 NR
II 263.132 19.0 10.0 0.02 NR
II 274.932 19.0 10.0 0.02
II 275.574 16.0 10.0 0.02
II 233.280 15.0 10.0 0.02
II 273.955 15.0 10.0 0.02
Ga I 294.364 64.0 100.0 0.05
I 417.206 45.0 100.0 0.07
I 287.424 38.0 100.0 0.08
I 403.298 27.0 100.0 0.01
I 250.017 16.0 100.0 0.2
II 209.134 11.0 100.0 0.3
I 245.007 10.0 100.0 0.3
I 294.418 9.4 100.0 0.3
I 271.965 5.7 100.0 0.5
I 233.828 3.9 100.0 0.8
I 265.987 3.6 100.0 0.8
Hg II 194.227 120.0 100.0 0.03
I 253.652 49.0 100.0 0.06
I 296.728 1.7 100.0 1.8
I 435.835 1.1 100.0 2.7
I 265.204 0.7 100.0 4.3
I 302.150 0.6 100.0 5.0
I 365.483 0.3 100.0 10.0
In II 230.606 47.0 100.0 0.06
I 325.609 25.0 100.0 0.1
I 303.936 20.0 100.0 0.2
I 451.131 16.0 100.0 0.2
I 410.176 6.4 100.0 0.5 H 410.174
I 271.026 5.4 100.0 0.6
I 325.856 5.0 100.0 0.6
II 207.926 4.2 100.0 0.7
I 256.015 4.2 100.0 0.7
I 293.263 2.0 100.0 1.5
II 197.745 1.7 100.0 1.8
I 175.388 1.6 100.0 1.9
K 404.721 0.7 1000.0 42.9
404.414 NM 1000.0 NM AR404.442
Li 460.286 3.5 100.0 0.9
323.263 2.8 100.0 1.1 OHband
274.118 1.9 100.0 1.6
497.170 1.4 100.0 2.1
256.231 0.7 100.0 4.3
413.262 0.4 100.0 7.5 NR
I 413.256 0.4 100.0 7.5 NR
Mg II 279.553 195.0 1.0 0.0002
II 280.270 100.0 1.0 0.0003
I 285.213 19.0 1.0 0.002
II 279.806 2.0 1.0 0.02
I 202.582 1.3 1.0 0.02
II 279.079 1.0 1.0 0.03
(continued on next page)
ICP EMISSION SPECTROMETRY 109
*
detection
Element S of It Wavelength IJlb Concentration Iimit~ Comments#
nm llg/mL
I 383.826 0.9 1.0 0.03
I 383.231 0.7 1.0 0.04
I 277.983 0.6 1.0 0.05
II 293.654 0.5 1.0 0.06
Mn II 257.610 220.0 10.0 0.001
II 259.373 190.0 10.0 0.002
II 260.569 145.0 10.0 0.002
II 294.920 39.0 10.0 0.008
II 293.930 29.0 10.0 0.01
I 279.482 24.0 10.0 0.01
II 293.306 22.0 10.0 0.01
I 279.827 18.0 10.0 0.02
I 280.106 14.0 10.0 0.02
I 403.076 6.8 10.0 0.04
II 344.199 6.6 10.0 0.05
I 403.307 6.3 10.0 0.05
II 191.510 5.8 10.0 0.05
Mo II 202.030 38.0 10.0 0.008
II 203.844 24.0 10.0 0.01
II 204.598 24.0 10.0 0.01
II 281.615 21.0 10.0 0.01
II 201.511 16.0 10.0 0.02
II 284.823 15.0 10.0 0.02
II 277.540 12.0 10.0 0.03
II 287.151 11.0 10.0 0.03
II 268.414 10.0 10.0 0.03
II 263.876 8.0 10.0 0.04
II 292.339 8.0 10.0 0.04
Na I 588.995 101.0 100.0 0.03 AR588.859
I 589.592 43.0 100.0 0.Q7
I 330.237 1.6 100.0 1.9
I 330.298 0.7 100.0 4.3
I 285.301 1.1 1000.0 27.3 NR
I 285.281 1.1 1000.0 27.3
II 288.114 0.6 1000.0 50.0
Nb II 309.418 83.0 100.0 0.04 OHband
II 316.340 75.0 100.0 0.04 OHband
II 313.079 60.0 100.0 0.05 OHband
II 269.706 43.0 100.0 0.Q7
II 322.548 42.0 100.0 0.Q7 OHband
II 319.498 41.0 100.0 0.Q7 OHband
II 295.088 40.0 100.0 0.08
II 292.781 40.0 100.0 0.08
II 271.662 34.0 100.0 0.09
II 288.318 31.0 100.0 0.1
II 210.942 31.0 100.0 0.1
II 272.198 30.0 100.0 0.1
II 287.539 28.0 100.0 0.1
Ni II 221.647 29.0 10.0 0.01
I 232.003 20.0 10.0 0.02
II 231.604 19.0 10.0 0.02
(continued on next page)
110 SOLTANPOUR ET AL.
*
detection
Element S of It Wavelength IJlb Concentration limit'!! Comments#
nm llglmL
II 216.556 17.0 10.0 0.02
II 217.467 13.0 10.0 0.02
II 230.300 13.0 10.0 0.02
II 227.021 12.0 10.0 0.03
II 225.386 12.0 10.0 0.03
I 234.554 9.5 10.0 0.03
II 239.452 7.8 10.0 0.04
I 352.454 6.6 10.0 0.05
I 341.476 6.2 10.0 0.05
P I 213.618 39.0 100.0 0.08
I 214.914 39.0 100.0 0.08
I 253.565 11.0 100.0 0.3
I 213.547 8.5 100.0 0.4
I 203.349 7.4 100.0 0.4
I 215.408 7.2 100.0 0.4
I 255.328 5.2 100.0 0.6
I 202.347 3.8 100.0 0.8
I 215.294 3.4 100.0 0.9
I 253.401 3.0 100.0 1.0
Pb II 220.353 70.0 100.0 0.04
I 216.999 33.0 100.0 0.09
I 261.418 23.0 100.0 0.1
I 283.306 21.0 100.0 0.1
I 280.199 19.0 100.0 0.2
405.783 11.0 100.0 0.3
224.688 3.0 100.0 0.3
368.348 8.6 100.0 0.3
266.316 7.7 100.0 0.4
239.379 6.3 100.0 0.5
363.958 5.2 100.0 0.6
247.638 5.1 100.0 0.6
S 180.669 30.0 100.0 0.1 Vac.line
181.979 30.0 100.0 0.1 Vac. line
Sb 206.833 91.0 100.0 0.03
217.581 68.0 100.0 0.04
231.147 49.0 100.0 0.06
252.852 28.0 100.0 0.1
259.805 28.0 100.0 0.1 NR
259.809 28.0 100.0 0.1 NR
217.919 19.0 100.0 0.2
195.039 18.0 100.0 0.2
213.969 16.0 100.0 0.2
204.957 15.0 100.0 0.2
214.486 12.0 100.0 0.3
209.841 8.7 100.0 0.3
203.977 6.6 100.0 0.5
220.845 6.5 100.0 0.5
287.792 4.7 100.0 0.6
Se 196.026 40.0 100.0 0.08
203.985 26.0 100.0 0.1
206.279 10.0 100.0 0.3
(continued on next page)
ICP EMISSION SPECTROMETRY 111
*
detection
Element S of It Wavelength IJlb Concentration Iimit~ CommentS#
nm I!glmL
207.479 1.9 100.0 1.6
199.511 0.6 100.0 5.0
Si 251.611 250.0 100.0 0.01
212.412 180.0 100.0 0.02
288.158 110.0 100.0 0.03
250.690 100.0 100.0 0.03
252.851 95.0 100.0 0.03
251.432 79.0 100.0 0.04
252.411 75.0 100.0 0.04
221.667 72.0 100.0 0.04
251.920 61.0 100.0 0.05
198.899 50.0 100.0 0.06
221.089 47.0 100.0 0.06
243.515 36.0 100.0 0.08
190.134 23.0 100.0 0.1
220.798 23.0 100.0 0.1
205.813 23.0 100.0 0.1
Sn II 189.989 120.0 100.0 0.03
I 235.484 31.0 100.0 0.1
I 242.949 31.0 100.0 0.1
I 283.999 27.0 100.0 0.1
I 226.891 25.0 100.0 0.1
I 224.605 25.0 100.0 0.1
I 242.170 19.0 100.0 0.2
I 270.651 18.0 100.0 0.2
I 220.965 16.0 100.0 0.2
I 286.333 14.0 100.0 0.2
I 317.505 14.0 100.0 0.2 OH band
Sr II 407.771 72.0 1.0 0.0004
II 421.552 39.0 1.0 0.0008
II 216.596 36.0 10.0 0.008
II 215.284 29.0 10.0 0.01
II 346.446 13.0 10.0 0.02
II 338.071 8.8 10.0 0.03
II 430.545 4.8 10.0 0.06
I 460.733 4.4 10.0 0.07
II 232.235 2.9 10.0 0.1
II 416.180 2.4 10.0 0.1
Te 214.281 73.0 100.0 0.04
225.902 17.0 100.0 0.2
238.578 17.0 100.0 0.2
I 214.725 14.0 100.0 0.2
I 200.202 12.0 100.0 0.3
I 238.326 11.0 100.0 0.3
I 208.116 11.0 100.0 0.3
I 199.418 6.3 100.0 0.5
I 225.548 2.7 100.0 1.1
I 226.555 2.7 100.0 1.2
Ti II 334.941 79.0 10.0 0.0004
II 336.121 57.0 10.0 0.0005
II 323.452 56.0 10.0 0.005 OHband
(continued on next page)
112 SOLTANPOUR ET AL.
nm ~g!mL
t S of I = state of ionization. Symbols I, II, and III indicate that the spectral lines originate from the
neutral atom, singly ionized, and doubly ionized states, respectively.
+ InlIb =ratio of net analyte intensity to background intensity.
Concentration = concentration of the single element analyte solution used for the wavelength
scans from which the prominent lines were determined.
11 Detection limits estimated from the IJIb ratios using the formula: DL =O.03C/(IJIb) where C is
concentration of analyte.
# Includes interference information when a component of the background spectrum overlaps an
analyte line (e.g., the Ba 389.178-nm line is located on the H 338.905-nm line) or when an ana-
Iyte line is located in complex molecular band system (e.g., the OH 306.36-nm system) where
band components may cause spectral interferences. The notation of molecular bands does not pre-
clude the use of analyte wavelengths within the band region.
tt NR = not resolved. This description indicates components of an unresolved pair of lines.
+:!: Group NR indicates components of an unresolved group (three or more lines). Only the wave-
length of the strongest line is listed.
Vacuum lines for S require either a vacuum or an Ar purged spectrometer, see "Selection of
Wavelength."
Selection of Isotope
Ideally, the most abundant isotope is selected for an element because it
usually will produce the strongest signal to background ratio. Thus, the analyst
obtains the lowest detection limit, the best accuracy, and the best sensitivity.
However, the analytical isotope selection process can be complicated by the
presence of isobaric interference from atomic and molecular ions originating
from the plasma, the solvent, and the reagents used to prepare the test solutions
as well as from nonanalyte sample elemental constituents (Vaughan & Horlick,
1986; Munro et ai., 1986; Date et ai., 1987; Gray, 1986; Tan & Horlick, 1986).
In the isotope selection process the following are considered: analytical isotope
abundance, background isobaric species, and isobaric species resulting from
sample solution composition. Relative abundance of the isotopes of all natural
elements are given by Date and Gray (1989). Tables by Holden and Walker
(1972) and Weast and Astle (1979) contain not only the naturally occurring iso-
114 SOLTANPOUR ET AL
topes but those produced as a result of man's activities as well. These are useful
when used with the relative abundances in accurately calculating atomic masses
of the elements, and by themselves for those considering high resolution, double-
focusing ICP-MS.
Table 5-2. Detection limits in an ideal solution (pure water), 10% HCl, and in a simulated arid soil
digest.
ICP
Element Ideal solutiont 10% HCl:j: Simulated soil digest
l1g!mL
Ag 0.004
AI 0.0002 0.03
As 0.04 0.02 0.7
Au 0.04
B 0.0007 0.002 0.03
Ba 0.00002 0.001 0.001
Be 0.0004
Bi 0.05
Ca 0.00002 0.002
Cd 0.002 0.001 0.006
Ce 0.0007
Co 0.003 0.01 0.02
Cr 0.0003 0.003 0.01
Cu 0.0001 0.003 0.01
Fe 0.0003 0.003
Ga 0.0006
Hf 0.01
Hg 0.001 0.02 0.2
In 0.03
K 0.1
La 0.00005
Li 0.0003
Mg 0.00001 0.0004
Mn 0.00006 0.001
Mo 0.0002 0.004 0.02
Na 0.0002 0.01
Nb 0.00007
Ni 0.0004 0.005 0.009
p 0.02 0.03
Pb 0.002 0.03 0.07
Pt 0.08
Sb 0.2
Se 0.02 0.2
Si 0.01
Sn 0.03
Sr 0.00002 0.0006 0.007
Ta 0.002
Ti 0.00007 0.002
U 0.03
V 0.0002
W 0.0007
Zn 0.002 0.002 0.008
Zr 0.0004
t Data from Robin (1979).
:j: Data from CSUSTL.
Data from CSUSTL. Simulated soil digest contained 10% HCl and the following elements at the
given concentrations in micrograms per milliliter: AI = 1500, Fe = 500, K = 400, Ca = 200, Na =
200, Mg = 100, Ti = 60, Mn = 20, P = 15.
116 SOLTANPOUR ET AI...
Table 5-2 reveals the deterioration of the detection limits due to interele-
ment spectral interferences. For example, in the case of As, a detection limit of
0.7 ppm is shown. This detection limit will preclude determination of As in total
soil digests using direct nebulization. In this case, As is separated from major soil
constituents by the hydride generation technique (see "Determination of
Ultratrace Levels of Arsenic, Selenium, and Mercury by Use of Hydride-
Mercury Vapor Generator"). When interelemental interference is not as severe as
in the case of As, other correction techniques can be used (see "Correction for
Interferences (Inductively Coupled Plasma-Atomic Emission Spectrometry),,).
value of 3.66. In this case, the spectrometer was not restandardized before the
high background Pb solution was read. This may explain the reason for obtain-
ing 0.84 value instead of 1.0 for Pb. However, when a solution containing all the
above element concentrations except Pb was aspirated, it gave an apparent Pb
concentreation of 2.80, which is almost identical to the sum of the apparent Pb
concentrations in AI, Fe, and Ti solutions. The above example shows the validi-
ty of interelemental interference correction method discussed above.
Some precautions to be observed in interelemental spectral interference
correction are discussed in the remainder of this section. Care must be taken to
ensure that an adequate rinse between the introduction of each interferential solu-
tion is performed. Pure chemicals such as SpecPure (Curtis Matheson Scientific,
Inc. Chern Pure Brand, Houston, TX) reagents should be used for determination
of interelemental interferences. If the chemicals are not pure, an impurity of ana-
lyte in these chemicals will create rather large errors in the results. For soil HF-
HCI0 4 digests, the interference of major soil constituents on other elements
should be determined. Soil extracts also should be examined for possible
interelemental interferences. These interferences are specific for a given instru-
ment depending on the wavelength used for each element and effective use of
baffles and black interiors to reduce stray light interferences.
In the instrument (Jarrell-Ash Model 975 ICP AtomComp) used at the
CSUSTL, it was found tha~ the following elements significantly interfere with
some trace elements in HF-HCI0 4 soil digests: AI, Fe, Mg, Mn, and Ti. However,
one should be on guard against interference from other elements that may be
found in large quantities in contaminated soils. The interelemental spectral inter-
ferences found in the ICP-AES system mentioned above are shown in Table 5-3.
In the Colorado State University instrument, Ca interference with other elements
is very low in contrast with published reports on another instrument (U.S. Dep.
Commerce Natl. Tech. Inform. Serv., 1977).
The order of elemental correction is important (Marciello & Ward, 1978).
As a general rule, interelement correction should be programmed into the com-
puter in the following order:
1. Major matrix components as interfering elements.
2. In order of magnitude of interference effect for minor matrix compo-
nents.
In correcting for interelemental interference, one should be aware of the
fact that inference per unit concentration of the interfering element may not be
linear. This has been shown to be the case in the particular instrument used for
the Ca interference on As, Se, Pb, and Sn (U.S. Dep. Commerce Natl. Tech.
Inform. Serv., 1977). In this event, curves should be plotted showing the appar-
ent concentration of the affected element as a function of the concentration of the
affecting elements. These curves should then be used to correct for interelemen-
tal interferences. Computer programs and computers interfaced with the direct-
readers capable of performing these tasks wit make the interelemental interfer-
ence corrections easier and much faster.
Another scheme for interference correction, called background correction
by some, is by the use of a device called a spectrum shifter. The spectrum shifter
N
...<=
Table 5-3. Examples of some interelemental spectral interferences observed in an ICP spectrometer at the CSUSTL.
Wavelength (nm)
Affecting 324.7 206.2 x 3 213/6 x 2 202.0 214.4 228.6 267.7 407.7 249.7 455.4 220.3 253.6 193.6 196.0
element Concentration Cu Zn Ni Mo Cd Co Cr Sr B Ba Pb Hg As Se
llg/mL Apparent concentration of affected elements llg/mL
K 400t
Ca 50 0.003
200t 0.002 0.D15 0.024
500 0.004 0.039 0.079
Mg 50 0.007 0.009 0.005 0.004 0.006 0.024 0.054
lOOt 0.012 0.012 0.005 0.008 0.010 0.039 0.043 0.087
200 0.024 0.038 0.008 0..013 0.004 0.066 0.083 0.168
Na 200t
P 15t 0.030
Ti 60t 0.026 0.003 0.111 0.008 0.028 0.038 0.197
Mn 20t 0.002 0.003 0.010 0.D18
Fe 100 0.005 0.006 0.011 0.010 0.008 0.016 0.004 0.285 0.064 1.36 0.053 0.083
500t 0.014 0.021 0.048 0.044 0.046 0.072 0.024 1.32 0.164 6.96 0.250 0.564
1000 0.025 0.040 0.093 0.082 0.088 0.140 0.046 2.51 0.004 0.302 13.1 0.494 1.07
AI 500 0.088 0.002 0.253 0.080 0.008 0.053 0.006 0.006 0.866 0.019 5.96 1.46
rJl
1000 0.002 0.176 0.013 0.517 0.164 0.008 0.016 0.106 0.020 0.012 1.76 0.030 12.2 2.90 0
1500t 0.002 0.261 0.016 0.760 0.242 0.018 0.024 0.156 0.020 0.D18 2.62 0.054 18.0 4.28 ~
High background solutiont 0.046 0.295 0.080 0.807 0.288 0.200 0.074 0.172 1.26 0.023 2.80 6.84 18.2 4.98
~
t Indicates high background solution was made from the elements and concentrations marked by t. All solutions were 10% in HCI. -=0
~
:=
~
>
r
ICP EMISSION SPECTROMETRY 121
measures the background radiation close to the analyte exit slit. This radiation is
assumed to originate from nonanalyte sources and to increase the radiation at the
analyte wavelength. The average background radiation signal is subtracted from
the signal observed at the analyte wavelength to correct for the interference. This
correction method may be used when the analyst is not aware of all the con-
stituents of the sample or deals with samples of varying matrixes. This correc-
tion method is obviously not valid when the background radiation received at the
analyte wavelength is grossly different from the one measured by the spectrum
shifter.
discussed below are: (i) Solute vaporization, (ii) Ionization suppression, and (iii)
Space-charge.
Solute vaporization interference occurs in instances in which the solute
does not have sufficient time and/or the source does not have sufficient energy to
dissociate the solute before the analytical species moves into and through the
region of observation. Typical manifestations of the interference are suppression
and/or increased variability of the analyte signal as a function of increased con-
comitant concentration. For example, Johnson et al. (1979a) showed that AI sup-
presses the Ca signal in a direct current plasma. Winge et al. (1991) published
high-speed photographs providing evidence that species generally associated
with low spectroscopic temperatures can persist through the central channel of
the ICP to enter the sampling cone of the ICP-MS, and discussed similar reports
of unvaporized solvent doing the same. Ionization interferences can be noted for
ICP-MS. Partially ionized elements in an ICP, e.g., Au and B, are susceptible to
ion suppression from fully ionized interferants (smaller effective analyte ion
molar ratio/interferant ion) and less effective in causing ion suppression for fully
ionized analytes (larger analyte ion molar ratio/interferent ion) (Gregoire, 1989).
In ICP-MS measurements, ionization interferences cause suppression of analyti-
cal response (Houk & Thompson, 1988; Tan & Horlick, 1986).
Easily ionized matrix element interference has been recognized in ICP-MS
data. For a given matrix element ion interference, the analytical response for the
lighter atomic mass isotopes are suppressed more than the heavier ones. For a
given analyte, lighter matrix elements suppress the response less than the heav-
ier ones. These trends are consistent with what would be expected from a space-
charge effect (Gregoire, 1989; Hieftje, 1992). Tracing the course of the ions from
the point of ion production in the plasma, the ions that move through the sampler
and skimmer, through the ion optics, and through the quadrupole, are reflected
away from the deflector and accelerated onto the detector. As the particle beam
electrons diffuse out of the beam more quickly than the positive ions-a process
termed ambipolar diffusion (Ahearn, 1972). As a consequence, coulombic repul-
sion spreads the ion beam. The larger ions stay on course better than the lighter
ones. Equivalently, the trajectories of the lighter ions are affected more by the
ions of the heavier mass isotopes than the trajectories of the heavier ions are
affected by the lighter mass isotopes.
In practice, it is difficult to separate the nonspectroscopic interferences.
These can be compensated for by matrix matching, but that seriously limits the
range of matrix element concentration variability between test solutions within
the run. Another approach that has been used successfully by many is the inter-
nal standard calibration method, discussed in "Methods of Correction for
Interferences and Standardization Procedures (Inductively Coupled Plasma-
Mass Spectrometry)."
Mass Discrimination. Gregoire (1989) defines mass discrimination as
bias in ion transmission to the detector the magnitude of which is dependent
upon the mass of the analytical isotope. Further, the effects can be divided into
two categories depending upon the origin of the mass bias. The first category is
referred to as instrumental mass discrimination effects which are the result of
mass discrimination occurring at the interface (sample and skimmer cones), ion
ICP EMISSION SPECTROMETRY 123
lenses, quadrupole mass filter and detector. Correction factors for instrumental
mass discrimination are normally found by comparing measured isotope ratios to
the known isotope ratio for a substance of known or certified isotopic composi-
tion, and applying the correction to the samples run during the same time.
Instrumental mass discrimination can range from 50% per dalton for light ele-
ments to 2% per dalton for heavy elements.
The other type of mass discrimination results from the presence of matrix
elements, has been reported only twice, and effects only Li and B. Briefly, the
magnitude of the effect is dependent upon five factors:
1. Absolute mass of the analyte.
2. Degree of ionization of the analyte.
3. Difference in mass between the two isotopes.
4. Mass of the matrix elements.
5. Degree of ionization of the matrix elements.
The above dependent factors involved with this effect are very similar to the
space-charge interpretation of analyte response suppression, so for all practical
purposes matrix dependent mass discrimination can be considered a special case
of the more general space-charge effect. The interested reader is referred to
Gregoire (1989) for more details.
Unwanted Ions (Inductively Coupled Plasma-Mass Spectroscopy).
Unwanted ions refer to that part of the ion beam other than the analyte ion part
reaching the detector. In a mass spectrometer system, particularly one operated
with peak widths in the neighborhood of one dalton, the analyte signal may be
accompanied by unwanted ions. These unwanted ions occur as a consequence of
several factors (Gray, 1985; Date & Gray, 1989; Gregoire, 1989; Vaughan &
Horlick, 1986; Munro et aI., 1986; Date et aI., 1987; Gray, 1989; Tan & Horlick,
1986), and include the following:
1. Elemental ions (NM+) of the same unit mass as the analyte isotope
(NN).
2. Elemental hydride (N{N-IMIHV) molecular ions that are of the same
unit mass as the analyte isotope (NA +).
3. Element oxide (N{N-16M160+) molecular ions that are of the same unit
mass as the analyte isotope (NA +).
4. Element hydroxide (N{N-17M1601HV) molecular ions that are of the
same unit mass as the analyte isotope (NA +).
5. Elemental (2NM2+) ions that are doubly charged and twice the unit mass
of the analyte isotope (NA +). Also, for unit resolution spectrometers,
elemental 2N-l)M2+) and (2N+IlM2+) ions must be included as ions that
will overlap, albeit partially with the analytical mass (NA +).
6. Elemental Argide (N{N-36M36Ar}+' N{N-38M38Ar}+' N{N-40M 40 ArV)
molecular ions that are of the same unit mass as the analyte isotope
(NN).
7. Elemental Hydrogen Argide (N{N-37MIW6Ar}+' N{N-39MIH38Ar}+'
N{N-41MIH40ArV) ions that are of the same unit mass as the analyte iso-
tope (NA+).
124 SOLTANPOUR ET AL.
same with sequential vs. simultaneous, but the simultaneous is easier to apply.
Trends for improvement in analytical sensitivities (Skogerboe & Grant, 1970)
were about the same as noted for accuracies. While the isobaric overlap correc-
tion procedure is still being evaluated, it appears that the calibration - subtrac-
tion approach, supplemented by assumptions concerning the natural distribution
of the interfering isotopes where convenient and/or necessary, can yield correc-
tions within 2.5% of the true correction. The accuracy of the final results then
depends on the severity of the specific interference (Johnson et aI., 1979b).
Practical Applications
Grinding Soil Samples
Soil samples should be air dried as close to the time of sampling as possi-
ble (drying and grinding of soil samples are not recommended for Mn and Cr).
Soil samples may contain clods or large aggregates. Many laboratories use auto-
matic grinders to crush soil. Recent studies have shown that the amount of
extractable micronutrients from soils is affected largely by the degree of grind-
ing (Soltanpour et aI., 1976, 1979a); therefore, grinding variables such as force
and time should be standardized. When soils are ground for extractable microele-
ments, care must be taken to avoid excessive grinding. It is important to use
grinders that do not contaminate the soil. At the CSUSTL, a high-density alu-
minum oxide auger made by the Coors Porcelain Co., Golden, Colorado, is
attached to a Nasco-AspIin automatic grinder. The grinder is equipped with a 2-
mm stainless steel sieve. This grinder minimizes the degree of soil contamina-
tion with trace elements. In soil analysis, it is a standard procedure to pass the
soil through a 2-mm sieve after mild grinding. Then all analytical results are
based on a 2-mm soil. For total elemental analysis, the 2-mm soil may be further
ground so that all of it passes through a O.16-mm (lOO-mesh) polyvinyl chloride
(PVC) sieve.
ously for Ca, Mg, Na, and K and then calculating the Na absorption ratio. Plant
digests also are analyzed by ICP-AES.
The CSUSTL and many environmental laboratories in the western U.S.
analyze mine overburden and mine spoil materials by use of AB-DTPA and ICP-
AES for P, Zn, Cu, Mn, B, Cd, As, Se, Mo, Pb, Ni, and other elements for screen-
ing these materials for their potential toxicity to plants and their consumers
(Soltanpour, 1991). The AB-DTPA extract is low in Ca, Mg, AI, Fe, Mn and Ti,
which cause interelemental interference, and therefore is well adapted to ICP-
AES analysis. Obviously water extracts are ideal for ICP-AES determinations,
but some elements in water extracts are found in very low concentrations that are
below the ICP-AES detection limits.
Jones (1977) determined Ca, K, Mg, and P simultaneously in the double-
acid extracts of Georgia soils.
The type of vessel and shaker and the speed of the shaker may affect the
amount of some extractable elements, but their effect is small compared with the
grinding variables (Soltanpour et aI., 1976).
For obtaining saturation extracts, refer to Chapter 14 (Rhoades, 1996). For
obtaining double-acid extractions, refer to Chapter 19 (Helmke & Sparks, 1996).
For obtaining Mehlich 3 extracts (a modification of double-acid extract for mul-
tielement extraction) refer to Chapter 19 (Helmke & Sparks, 1996).
Special Equipment
1. Reciprocal shaker.
Reagents
1. Ammonium bicarbonate (NH4HC03).
2. Diethylenetriaminepentaacetic acid (DTPA).
3. Ammonium hydroxide (N~OH).
5. Ammonium bicarbonate-diethylenetriaminepentaacetic acid solution
(AB-DTPA).
To make aIM NH4HCOr -O.OO5 M DTPA solution, add 1.87 g of DTPA
to 800 mL of distilled-deionized water (DDW). Add approximately 2 mL of 1:1
NH40H to facilitate dissolution and to prevent effervescence. Shake until most
of the DTPA is dissolved. Then add 79.06 g of NH4HC0 3, and stir gently until
dissolved. Adjust the pH to 7.6 with NH40H. Dilute the solution to 1.0 L with
DDW. This solution is unstable with regard to pH. If the solution is stored under
about 3 cm of mineral oil, the pH remains stable. However, it is preferable to use
a fresh solution.
Procedure. Weigh 10 g of a 2-mm soil into a 125-mL conical flask. Add
20 ml of AB-DTPA solution. Shake on the reciprocal shaker for 15 min at 180
cycles/min with flasks kept open. Filter the extracts through Whatman no. 42 fil-
ter paper or its equivalent. Take a 2-mL aliquot of the extract, and add 0.25 mL
of concentrated HN0 3 Shake for 10 min to get rid of carbonate-bicarbonate
matrix to prevent clogging of the capillary tip in cross-flow or concentric nebu-
lizers. This solution is now ready for simultaneous multielement determination.
With high-salt nebulizers (Babington type) the acid pretreatment is not required.
ICP EMISSION SPECTROMETRY 127
Special Equipment
1. Stainless steel perchloric hood.
2. Hot plate.
3. Teflon beaker, 100 mL.
4. Teflon watch cover.
Reagents
1. Perchloric acid (HCI04).
2. Nitric acid (HN03).
3. Hydrofluoric acid (HF).
4. Hydrochloric acid (HCl).
Procedure. Weigh 1.0 g of the lOO-mesh (0.15-mm) soil into a 100-mL
Teflon beaker. Add 10 mL of HN0 3 and 10 mL of HCI04 Cover with a Teflon
watch cover, and heat at 200C for I h. Remove the cover and continue heating
until the volume is 2 to 3 mL. Cool the sample, add 5 mL of HCI04 and 10 mL
of HF, cover with a Teflon watch cover, and heat overnight at 200C. Overnight
digestion is for convenience, but one may terminate digestion as soon as all
siliceous material has been dissolved. Remove the cover, and continue heating
until the volume is 2 to 3 mL. Cool the digest, add 10 mL of 50% HCI, cover,
and heat at 100C for 30 min. Remove from the hot plate, and allow it to cool.
Transfer the solution quantitatively into a 50-mL volumetric flask, and bring to
volume. The solution is then ready for ICP determination.
Comments. The reason a 100-mesh (0.15-mm) soil sample instead of a 2-
mm sample is used in digestion is to speed up the breakdown of silicates. At the
CSUSTL, 2-mm soil samples have been digested with no difficulty when soils
were digested overnight in the presence of HF-HCI04
The HCI04 hood should be washed periodically to remove perchlorates
and to avoid danger of explosion. Do not let HCl04 solution get dry, anhydrous
HCl04 is explosive.
128 SOLTANPOUR ET AL.
Table 5-4. Primary standard solution preparation (adapted from Ward, 1978).t
Element Compound Weight Solvent
g
AI AI 1.0000 6MHCI
AlCI3 6HzO 8.9481 1MHCI
Sb Sb 1.0000 Aqua regia
SbCI 3 1.8736 1MHCI
As As 1.0000 4MHN03
Asz0 3 1.3203 4MHCI
Ba BaClz* 1.1516 Water
BaC03* 1.4369 0.05MHN0 3
BaN03 1.9029 Water
Be Be 1.0000 0.5MHCI
BeO* 2.7753 0.5MHCI
Bi Bi 1.0000 4MHN0 3
Bi z0 3 1.1149 4MHN03
Bi (N03)3 5HzO 2.3211 1MHN03
B B 1.0000 4MHN03
H3B03 5.7195 Water
Cd Cd 1.0000 4MHN0 3
CdO 1.1423 4MHN03
Ca CaC03 2.4972 0.5MHN0 3
Ca (N03)z 4HzO* 5.8920 Water
Cr Cr 1.0000 4MHCI
CrCI 3 (6H20) 5.1244 Water
Co Co 1.000 4MHCI
CoCl z 6 HzO 4.0373 Water
Cu Cu 1.0000 4MHN0 3
CuO 1.2518 4MHN0 3
In In 1.0000 Aqua regia
Fe Fe 1.0000 4MHCI
Fez03 1.4297 4MHCI
Pb Pb 1.0000 4MHN03
PbO 1.0772 4MHN03
Pb (N0 3)z 2.6758 Water
Li Li zC0 3 5.8241 1MHCI
LiCI 6.1092 Water
Mg MgO 1.6581 0.5MHCI
MgCLz 6HzO* 8.3625 Water
Mn Mn 1.0000 4MHN0 3
MnOz 1.5825 4MHN0 3
Hg HgClz 1.3535 Water + 1 g (NH4)zSzOg
Mo Mo 1.0000 Aqua regia
Mo0 3 1.5003 Aqua regia
Ni Ni 1.0000 4MHCI
NiO 1.2725 4MHCI
NiClzo 6HzO 4.0489 Water
Nb NbzOs 1.4305 Minimum quantity of HF, add 1 M HCI
P NaHzP04 3.8735 Water
NH4HzP04 3.7137 Water
K KCI 1.9067 Water
KZC03 1.7673 1MHCI
Se SeOz 1.4053 Water
Si Na zSi0 3 9HzO* 10.1190 Water
Ag Ag l.OOOO 4MHN0 3
AgzO 1.0742 4MHN03
NaBH4
Solution
To
plasma
Sample
60ml coarse
fritted disc
Buchner funnel
Drain
Argon
Pumps l2Imin
SUMMARY
newsletters available from the manufacturer of their unit and other sources so
that they can be kept abreast of new developments in ICP-AES and ICP-MS.
Journals references and other significant sources of information are given in the
reference section.
REFERENCES
Abell, J.D. 1991. Performance benefits of optimisation of laser ablation sampling for ICP-MS. p.
209-217. In G. Holland and AN. Eaton (ed.) Applications of plasma source mass spectrom-
etry. R. Soc. Chern. Cambridge, England.
Ahearn, A.J. (ed.). 1972. Trace analysis by spark source mass spectrometry. Acad. Press, New York.
Bacon, J.R., A.T. Ellis, and J.G. Williams. 1989. Atomic spectrometry update-inorganic mass spec-
trometry and x-ray fluorescence spectrometry. J. Anal. At. Spectrom. 4:199R.
Bacon, J.R., AT. Ellis, and J.G. Williams. 1990. Atomic spectrometry update-inorganic mass spec-
trometry and x-ray fluorescence spectrometry. J. Anal. At. Spectrom. 5:243R.
Bacon, J.R., A.T. Ellis, and J.G. Williams. 1991. Atomic spectrometry update-inorganic mass spec-
trometry and x-ray fluorescence spectrometry. J. Anal. At. Spectrom. 6:229R.
Bajo, S. 1978. Volatilization of arsenic (III, V) antimony (III, V) and selenium (IV, VI) from mix-
tures of hydrogen fluoride and perchlonc acid solution: Application to silicate analysis. Anal.
Chern. 50:649-651.
Barnes, R.M. (ed.). 1992. ICP inform. newsletter. Univ. Massachusetts, Amherst, MA.
Beasecker, D.R., and LL Williams. 1978. An improved sample delivery system for ICAP analysis.
Jarrell-Ash Plasma Newslett. 1(3):5-9.
Beauchemin, D. 1989. Early experience with inductively coupled plasma mass spectrometry. J. Anal.
At. Spectrom. 4:553.
Beauchemin, D., J.w. Mclaren, and S.S. Berman. 1987. Study of the effects of concomitant elements
in inductively coupled plasma mass spectrometry. Spectrochim. Acta 42B:467.
Belchamber, R.M., D. Betteridge, AP. Wade, AJ. Cruickshank, and P. Davison. 1986. Removal of a
matrix effect in ICP-AES multi-element analysis by simplex optimisation. Spectrochim. Acta
41:503-505.
Bernas, B. 1968. A new method for decomposition and comprehensive analysis of silicates by atom-
ic absorption spectrometry. Anal. Chern. 40:1682-1686.
Boumans, P'W.J.M. 1966. Theory of spectrochemical excitation. Plenum Press, New York.
Boumans, P.W.J.M. 1984. Line coincidence tables for inductively coupled plasma atomic emission
spectrometry. 2nd ed. Pergamon, New York.
Bradford, G.R., and D. Bakhtar. 1991. Determination of trace metals in saline irrigation drainage
waters with ICP-OES after preconcentration by chelatiOn/solvent extraction. Environ. Sci.
Technol. 25:1704--1708.
Bradshaw, N., E.F.H. Hall, and N.E. Sanderson. 1989. Inductively coupled plasma as an ion source
for high-resolution mass spectrometry. J. Anal. At. Spect. 4:801.
Braverman, D.S. 1992. Determination of rare earth elements by liquid chromatography separation
using inductively coupled plasma mass spectrometric detection. J. Anal. At. Spect. 7:43.
Brown, P.G., T.M. Davidson, and J.A Caruso. 1988. Application of He microwave induced plasma
mass spectrometry to the detection of high ionization potential gas phase species. J. Anal. At.
Spect. 3:763.
Chisum, M.E. 1992. Applications of negative ion analyses on the ELAN 250 ICP/MS. At. Spectrosc.
12:155.
Cresser, M.S., LC. Ebdon, C.W. McLeod, and J.C. Burridge. 1986. Atomic spectrometry update-
environmental analysis. J. Anal. At. Spectrom. 1:1R.
Cresser, M.S., LC. Ebdon, and J.R. Dean. 1988.Atomic spectrometry update-environmental analy-
sis. J. Anal. At. Spectrom. 3:1R.
Cresser, M.S., LC. Ebdon, and J.R. Dean. 1989. Atomic spectrometry update-environmental analy-
sis. J. Anal. At. Spectrom. 4:1R.
Cresser, M.S., LC. Ebdon, J. Armstrong, J.R. Dean, M.H. Ramsey, and M. Cave. 1990. Atomic spec-
trometry update_nvironmental analysis. J. Anal. At. Spectrom. 5:1R.
Cresser, M.S., J. Armstrong, J.R. Dean, P. Watkins, and M. Cave. 1991. Atomic spectrometry
update-environmental analysis. J. Anal. At. Spectrom. 6:1R.
Cresser, M.S., J. Armstrong, J.R. Dean, P. Watkins, and M. Cave. 1992. Atomic spectrometry
update-environmental analysis. J. Anal. At. Spectrom. 7:1R.
ICP EMISSION SPECTROMETRY 135
Dahlquist, R.L., and J.w. Knoll. 1978. Inductively coupled plasma-atomic emission spectrometry.
Analysis of biological materials and soils for major, trace, and ultra-trace elements. Appl.
Spectrosc. 32:1-29.
Date, AR. 1986. ICP-MS: The best thing in analytical chemistry since chopped light? p. 20-26. In
Biennial Natl. Atom. Spectrosc. Symp., Bristol, England. July 1986.
Date, AR., Y.Y. Cheung and M.E. Stuart. 1987. The influence of polyatomic ion interferences in
analysis by inductively coupled plasma source mass spectrometry (ICP-MS). Spectrochim.
Acta 42B:3.
Date, AR, and AL. Gray. (ed.). 1989. Applications of inductively coupled plasma mass spectrom-
etry. Blackie and Sons, London.
Dean, J.R., L.c. Ebdon, H.M. Crews, and RC. Massey. 1988. Characteristics of flow injection induc-
tively coupled plasma mass spectrometry for trace metal analysis. J. Anal. At. Spect. 3:349.
Denoyer, E.R. 1991. Analysis of powdered samples by laser sampling ICP-MS. p. 199-208. In G.
Holland and AN. Eaton (ed.) Applications of plasma source mass spectrometry. R Soc.
Chern., Cambridge, England.
Denoyer, E., L. Qinghong, A Stroh, U. Volkopf, and R. Thomas. 1991a. Extending ICP-MS capa-
bilities with flow injection analysis. In Elem. Mass Spectrom. Workshop. Am. Soc. Mass
Spectrom. San Francisco, CA November 1991.
Denoyer, E.R, KJ. Fredeen, and J.W. Hager. 1991b. Laser solid sampling for inductively coupled
plasma mass spectrometry. Anal. Chern. 63:445A
Denoyer, E.R., and A Stroh. 1992. Expanding ICP-MS capabilities using flow injection. Am. Lab.
February 74.
Doherty, W. 1989. An internal standardization procedure for the determination of Yttrium and the
rare earth elements in geological materials by inductively coupled plasma-mass spectrometry.
Spectrochim. Acta 44B:263.
Dolan, R., J. Van Loon, D. Templeton, and A Paudyn. 1990. Assessment of ICP-MS for routine mul-
tielement analysis of soil samples in environmental trace element studies. Fres. J. Anal.
Chern. 336:99.
Douglas, DJ., and R.S. Houk. 1985. Inductively-coupled plasma mass spectrometry (ICP-MS). Prog.
Analyt. At. Spectrosc. 8:1.
Douglas, DJ., and L.A Kerr. 1988. Study of solids deposition on inductively coupled plasma mass
spectrometry samplers and skimmers. J. Anal. At. Spect. 3:749.
Douglas, DJ. 1989. Some current perspectives on ICP-MS. Can. J. Spectrosc. 34:38.
Douthitt, c.B. 1994a. A new technology in high performance ICP-MS. p. 16. In Proc. 36th Rocky
Mountain Conf. Anal. Chern., Finnigan MAT, Pap. 2, Denver, CO. 1 August.
Douthitt, C.B. 1994b. High resolution ICP-MS for ultra trace analysis in difficult matrices. p. 16. In
Proc. Rocky Mountain Conf. Anal. Chern .. Finnigan MAT, Pap 6. Denver, CO. 1 August.
Ebdon, L.c., M.S. Cresser, and C.W. Mcleod. 1987. Atomic spectrometry update-environmental
analysis. 1. Anal. At. Spectrom. 2:1R
Ek, P.G., S.G. Hulden, E. Johansson, and T. Liljefors. 1991. A continuous hydride generation system
for ICP-MS using separately nebulized internal standard solutions. p. 178-198. In G. Holland
and AN. Eaton (ed.) Application of plasma source mass spectrometry. R. Soc. Chern.
Evans, E.H., and L.C. Ebdon. 1989. Simple approach to reducing polyatomic ion interferences on
arsenic and selenium in inductively coupled plasma mass spectrometry, J. Anal. At. Spect.
4:299.
Fassel, V.A 1977. Current and potential applications of inductively coupled plasma (ICP)-atomic
emission spectroscopy (AES) in the exploration, mining, and processing of materials. Pure
Appl. Chern. 49:1533-1545.
Fassel, V.A., and RN. Kniseley. 1974. Inductively coupled plasmas. Anal. Chern. 46:1155A-1164A
Fasset. J.D., and PJ. Paulsen. 1989. Isotope dilution mass spectrometry for accurate elemental analy-
sis. Anal. Chern. 6L643A.
Garbarino, J.R., and H.E. Taylor. 1987. Stable isotope dilution analysis of hydrologic samples by
inductively coupled plasma mass spectrometry. Anal. Chern. 59:1568.
Gervais, L.S., and E.D. Salin. 1991. Inserted injector tubes for inductively coupled plasma spec-
trometry. J. Anal. At. Spect. 6:493.
Gray, AL. 1985. The ICP as an ion source-origins, achievements and prospects. Spectrochim. Acta
40B:1525.
Gray, A.L. 1986. Mass spectrometry with an inductively coupled plasma as an ion source: The influ-
ence on ultratrace analysis of background and matrix response. Spectrochim. Acta 41B:151.
Gray, AL. 1989. Visual observation of shock waves in an inductively coupled plasma mass spec-
trometry expansion stage. J. Anal. At. Spect. 4:371.
136 SOLTANPOUR ET AL.
Greenfield, S. 1983. Inductively coupled plasma-atomic emission spectroscopy (ICP-AES) with flow
injection analysis (PIA). Spectrochim. Acta 38:93-105.
Gregoire, D.C. 1987a. The effect of easily ionizable concomitant elements on non-spectroscopic
interferences in inductively coupled plasma-mass spectrometry. Spectrochim. Acta 42B:895.
Gregoire, D.C. 1987b. Influence of instrument parameters on nonspectroscopic interferences in
inductively coupled plasma-mass spectrometry. Appl. Spectrosc. 41:897.
Gregoire, D.C. 1989. Application of isotope ratios determined by ICP-MS to earth science studies.
Prog. Anal. Spectrosc. 12:433.
Hager, J.W 1989. Relative elemental responses for laser ablation-ICP-MS. Anal. Chern. 61:1243.
Hassan, S.M., and N.T. Loux. 1989. Elimination of spectral interferences in inductively coupled plas-
ma-atomic emission spectroscopy using orthogonal polynomials. Spectrochim. Acta
45:719-729.
Helmke, P.A., and D.L. Sparks. 1996. Lithium, sodium, potassium, rubidium, and cesium. p.
551-574. In D.L. Sparks et al. (ed.) Methods of soil analysis. Part 3. Chemical methods.
SSSA Book Ser. 5. SSSA and ASA, Madison, WI.
Hieftje, G.M. 1992. Toward the next generation of plasma-source mass spectrometers.p. 16. In Proc.
1992 Winter Conf. Plasma Spectrochem., San Diego, CA. 11 January.
Hieftje, G.M., and G.H. Vickers. 1989. Developments in plasma source/mass spectrometry. Anal.
Chim. Acta 216:1.
Hieftje, G.M., G.H. Vickers, and D.A. Wilson. 1988. Detection of Negative Ions by ICP-MS. Anal.
Chern. 60:1808-1812.
Holden, N.E., and EW Walker. 1972. Chart of the nuclides. Wall chart. 11th ed. Knolls Atomic
Power Lab. General Electric Co. and Naval Reactors, U.S. At. Energy Comm. Educ.
Relations, G.E. Company, Schenectady, NY.
Holland, G., and A.N. Eaton. 1991. Applications of plasma source mass spectrometry. R Soc. Chern.
Hossner, L.R 19%. Dissolution for total elemental analysis. p. 000-000. In D.L. Sparks et al. (ed.)
Methods of soil analysis: Chemical methods. Part 3. SSSA Book Ser. 5. ASA, CSSA, and
SSSA, Madison, WI.
Houk. R.S. 1986. Mass spectrometry of inductively coupled plasmas. Anal. Chern. 58:97A.
Houk, RS., Hl. Svec, and V.H. Fassel. 1981. Mass spectrometric evidence for suprathermal ioniza-
tion in an inductively coupled argon plasma. Appl. Spectrosc. 35:380.
Houk, R.S., and J.J. Thompson. 1988. Inductively coupled plasma mass spectrometry. Mass
Spectrom. Rev. 7:425.
Huang, Y.Q., anc C.M. Wai. 1986. Extraction of arsenic from soil digests with dithiocarbanates for
ICP-AES analysis. Commun. Soil Sci. Plant Anal. 17:125-133.
Hutton, RC., and A.N. Eaton. 1987. Role of aerosol water vapour loading in inductively coupled
plasma mass spectrometry. J. Anal. At. Spect. 2:595.
Igarashi, Y., C.K. Kim, Y. Takaku, K. Shiraishi, M. Yamamoto, and N. Ikeda. 1990. Application of
inductively coupled plasma mass spectrometry to the measurement of long-lived radionu-
c1ides in environmental samples. A review. Anal. Sci. 6:157.
Isaac, RA., and WC. Johnson. 1982. High speed analysis of agricultural samples using inductively
coupled plasma-atomic emission spectroscopy. Spectrochim. Acta 38:277-282.
Jackson, L.L., D.M. McKown, J.E. Taggart, Jr., Pl. Lamothe, and EE. Lichte. 1989. Geological and
inorganic materials. Anal. Chern. 61:109R
Jackson, L.L., T.L. Fries, J.N. Grossman, B.S.W King, Pl. Lamothe. 1991. Geological and inorgan-
ic materials. Anal. Chern. 63(biennial Applications Reviews issue):33R-48R
Jarvis, KE. 1992. Handbook of inductively coupled plasma mass spectrometry. Chapman and Hall,
New York.
Johnson, G.W. 1979. Trace analysis with a direct current plasma echelle spectrometer system. Ph.D.
diss. Colorado State University, Fort Collins (Diss. Abstr. 40104:1673-B).
Johnson, G.W., H.E. Taylor, and RK Skogerboe. 1979a. Evaluation of solute vaporization interfer-
ence effects in a direct current plasma. Anal. Chern. 51:2403.
Johnson, G.W., H.E. Taylor, and R.K Skogerboe. 1979b. Evaluation of spectral interferences asso-
ciated with a direct current plasma-multielement atomic emission spectrometer (DCP-MAES)
system. Appl. Spectrosc. 33:451.
Johnson, G.W., RD. Miller, and P.H. Brown. 1992a. ICP-MS for plant analysis. p. 212. In Proc. 1992
Winter Conf. on Plasma Spectrochem., San Diego, CA. 10 January.
Johnson, G.W., E.S. Littlefield, and R.O. Miller. 1992b. Botanical tissue analysis using the USN ICP-
MS technique. p. 49. In Proc. Pacific Conf. Chern. and Spectrosc., Foster City, CA. 21
October.
ICP EMISSION SPECTROMETRY 137
Jones, lB., Jr. 1977. Elemental analysis of soil extracts and plant tissue ash by plasma emission spec-
troscopy. Commun. Soil Sci. Plant Anal. 8:349-365.
Kim, C.K., Y. Oura, Y. Takaku, H. Nitta, Y. Igarashi, and N.Ikeda. 1989a. Measurements of
240pu: 239pu ratio by fission track method and inductively coupled plasma mass spectrometry.
1 Radioanal. Nucl. Chern. 136(5):353.
Kim, C.K, Y. Takaku, M. Yamamoto, H. Kawamura, K Siraishi, Y. Igarashi, I. Igarashi, H.
Takayama, and N. Ikeda. 1989b. Determination of Neptunium - 237 in soil samples using
inductively coupled plasma mass spectrometry. Radioisotopes 38:153.
Kim, C.K, R. Seki, S. Morita, S.1. Yamasaki, A. Tsumura, Y. Takaku, Y. Igarashi, and M. Yamamoto.
1991. Application of a high resolution inductively coupled plasma mass spectrometer to mea-
surement of long-lived radionuclides. J. Anal. At. Spect. 6:205.
K1esta, E.J., Jr., and J.K Bartz. 1996. Quality assurance and quality control. p. 19-48. In D.L. Sparks
et al. (ed.) Methods of soil analysis. Part 3. Chemical methods. SSSA Book Ser. 5. SSSA and
ASA, Madison, WI.
Koirtyohann, S.R., IS. Jones, and D.A. Yates. 1980. Nomenclature system for low-power inductive-
ly coupled plasma. Anal. Chern. 52:1965.
LaFemiere, KE., G.w. Rice, and V.A. Fassel. 1985. Flow injection analysis with inductively coupled
plasma-atomic emission spectroscopy: Critical comparison of conventional pneumatic, ultra-
sonic and direct injection nebulization. Spectrochimica Acta 40:1495-1504.
Laird, D.A., R.H. Dowdy, and R.e. Munter. 1991. Suspension nebulization analysis of clays by
inductively coupled plasma-atomic emission spectroscopy. Soil Sci. Soc. Am. J. 55:274-278.
Lam, J.W., and lW. Mclaren. 1990. Use of aerosol processing and nitrogen-argon plasmas for
reduction of oxide interference in inductively coupled plasma mass spectrometry. J. Anal. At.
Spect. 5:419.
Langmyhr, F.J., and P.E. Paus. 1968. The analysis of inorganic siliceous materials by atomic absorp-
tion spectrophotometry and the hydrofluoric acid decomposition tehcnique. Anal. Chim. Acta
43:506-507.
Larson, G.F., W.Y. Fassel, R.H. Scott, and R.N. Kniseley. 1975. Inductively coupled plasma-optical
emission analytical spectrometry. A study of some interelement effects. Anal. Chern. 47:238.
Larson, G.F., V.A. Fassel, F.K Winge, and R.N. Kniseley. 1976. Ultratrace analyses by optical emis-
sion spectroscopy: The stray light problem. Appl. Spectrosc. 30:384-391.
Lechler, P.J., and R.K Leininger. 1979. Analysis of black shale by ICAP spectroscopy. Jarrell-Ash
Plasma Newslett. 2(1):8-10.
Legere, G., and P. Burgener. 1985. Elimination of high-salt interference effects caused by lithium
metaborate using a new teflon high-salt nebulizer. ICP Inform. Newslett. 11:447-456.
Lim, H.B., R.S. Houk, M.e. Edelson, and KP. Carney. 1989. Some fundamental characteristics of a
reduced-pressure plasma extracted from an inductively coupled plasma. J. Anal. At. Spect.
4:365.
Lindsay, W.L., and W.A. Norvell. 1978. Development of a DTPA soil test for zinc, iron, manganese,
and copper. Soil Sci. Soc. Am. J. 42:421-428.
Marciello, I., and A.F. Ward. 1978. Interelement corrections for spectral line interferences. Jarrell-
Ash Plasma Newslett. 1(2):12-13.
Mclaren, J.W., and S.S. Berman. 1985. Wavelength selection for trace analysis by ICP-AES.
Spectrochimica Acta 40:217-225.
Mclaren, J.W., D. Beauchemin, and S.S. Berman. 1987. Application of inductively coupled plasma
mass spectrometry to analysis of marine sediments. Anasl. Chern. 59:610.
McQuaker, N.R., P.D. K1uckner, and G.N. Chang. 1979. Calibration of an inductively coupled plas-
ma-atomic emission spectrometer for the analysis of environmental materials. Anal. Chern.
51:888-895.
Munro, S., L. Ebdon, and D.J. McWeeny. 1986. Application of inductively coupled plasma mass
spectrometry (ICP-MS) for trace metal determination in foods. 1 Anal. At. Spec!. 1:211.
Odegard, M. 1979. Determination of major elements in geological materials by ICAP spectroscopy.
Jarrell-Ash Plasma Newslet!. 2(1):4-7.
Olivares, lA., and R.S. Houk. 1986. Suppression of analyte signal by various concomitant salts in
inductively coupled plasma mass spectrometry. Anal. Chern. 58:20--25.
Olson, KW., W.J. Haas, Jr., and Y.A. Fassel. 1977. Multielement detection limits and sample nebu-
lization efficiencies of an improved ultrasonic nebulizer and a conventional pneumatic nebu-
lizer in inductively coupled plasma-atomic emission spectrometry. Anal. Chern. 49:632...Q37.
Parsons, M.L., A. Forster, and D. Anderson. 1980. An atlas of spectral interferences in ICP spec-
troscopy. Plenum, New York.
Pearce, N.J.G., W.T. Perkins, I. Abell, G.A.T. Duller, and R. Fuge. 1992. Mineral microanalysis by
LASER ablation inductively coupled plasma mass spectrometry. 1. Anal. At. Spect. 7:53.
138 SOLTANPOUR ET AL.
Plantz, M.R., I.S. Fritz, F.G. Smith, and R.S. Houk. 1989. Separation of trace metal copmlexes for
analysis of samples of high salt content by inductively coupled plasma mass spectrometry.
Anal. Chern. 61:149.
Powell, M.J., and D.W. Boomer. 1995. Determination of chromium species in environmental sam-
ples using high-pressure liquid chromatography direct injection nebulization and inductively
coupled plasma mass spectrometry. Anal. Chern. 67:2474-2478.
Robin, I. 1979. Emission spectrometry with the aid of an inductive plasma generator, ICP Inform.
Newslett. 4(11):495-509.
Roychowdhury, S.B., and I.A Koropchak. 1990. Thermospray enhanced inductively coupled plasma
atomic emission spectroscopy detection for liquid chromatography. Anal. Chern. 62:484-489.
Sah, RN., and RO. Miller. 1992. Spontaneous reaction for acid dissolution of biological tissues in
closed vessels. Anal. Chern. 64:230.
Serfas, RE., I.1. Thompson, and RS. Houk. 1986. Isotope ratio determinations by inductively cou-
pled plasma/mass spectrometry for Zinc bioavailability studies. Anal. Chim. Acta 188:73.
Skogerboe, RK., and S.R Koirtyohann. 1976. Accuracy assurance in the analysis of environmental
samples. Natl. Bur. Stand., Stand. Publ. 422. u.S. Gov. Print. office, Washington, DC.
Skogerboe, RK., and c.L. Grant. 1970. Comments on the definitions of the terms sensitivity and
detection limit. Spectrosc. Leu. 3:215.
Slavin, M. 1971. Emission spectrochemical analysis. Wiley-Intersci., New York. p. 53-80.
Smith, F.G., D.R. Wiedrin and R.S. Houk. 1991. Measurement of boron concentrations and isotope
ratios in biological samples by inductively coupled plasma mass spectrometry with direct
injected nebulization. Anal. Chim. Acta. 248:229-234.
Soltanpour, P.N., A Khan, and W.L. Lindsay. 1976. Factors affecting DTPA-extractable Zn, Fe, Mn,
and Cu from the soils. Commun. Soil Sci. Plan Anal. 7:797-821.
Soltanpour, P.N., and AP. Schwab. 1977. A new soil test for simultaneous extraction of macro- and
micro-nutrients in alkaline soils. Commun. Soil Sci. Plant Anal. 8:195-207.
Soltanpour, P.N., and S. Workman. 1979. Modification of the N~HCOrDTPA soil test to omit car-
bon black. Commun. Soil Sci. Plant Anal. 10:1411-1420.
Soltanpour, P.N., A. Khan, and A.P. Schwab. 1979a. Effect of grinding variables on the NH4 HCOr
DTPA soil test values for Fe, Zn, Mn, Cu, P, and K. Commun. Soil Sci. Plant Anal.
10:903-909.
Soitanpour, P.N., S. Workman, and A.P. Schwab. 1979b. Use of inductively - coupled plasma spec-
trometry for the simultaneous determination of macro- and micronutrients in N~HCOr
DTPA extracts of soils. Soil Sci. Soc. Am. J. 43:75-78.
Soltanpour, P.N., and S. Workman. 1981. Soil testing methods used at Colorado State University Soil
Testing Laboratory for the evaluation of fertility, salinity, sodicity, and trace element toxicity.
Colorado State Univ. Exp. Stn. Tech. Bull. 142.
Soltanpour, P.N., J.B. Jones, Jr., and S.M. Workman. 1982. Optical emission spectrometry. p. 29-65.
In AL. Page (ed.) Methods of soil analysis. Part 2. 2nd ed. Agron. Monogr. 9. ASA and
SSSA, Madison, WI.
Soltanpour, P.N. 1991. Determination of nutrient availability and elemental toxicity by AB-DTPA
soil test and ICPS. Adv. Soil Sci. 16:165-190.
Suddendorf, R.F., and K.w. Boyer. 1978. Nebulizer for analysis of high salt content samples with
inductively coupled plasma emission spectrometry. Anal Chern. 50: 1769-1771.
Tan, S.H., and G. Horlick. 1986. Background spectral features in ICP-MS. Appl. Spectrosc. 40:445.
Taylor, H.E., J.R. Garbarino, and S.R Koirtyohann. 1991. Flame ionization mass spectrometry:
Isotope ratio determinations for potassium. Appl. Spectrosc. 45 :886.
Taylor, P., and P. Schutyser. 1986. Description of a computer program for quantitative spectral analy-
sis of ICP-AES spectra generated with a high resolution computer-controlled monocrhoma-
tor. 1985. Spectrochim. Acta 41:81-103.
Thompson, M., B. Pahlavanpour, and S.J. Walton. 1978a. Simultaneous determination of trace con-
centrations of arsenic, antimony, bismuth, selenium, and tellurium in aqueous solution by
introduction of the gaseous hydrides into an inductively coupled plasma source for emission
spectrometry. Part I. Preliminary studies. Analyst (London) 103:568-579.
Thompson, M., B. Pahlavanpour, and S.J. Walton. 1978b. Simultaneous determination of trace con-
ncentrations of arsenic, antimony, bismuth, selenium and tellurium in aqueous solution by
introduction of the gaseous hydrides into an inductively coupled plasma source for emission
spectrometry. Part 2. Interference studies. Analyst (London) 103:705-713.
Thompson, M., and J.N. Walsh. 1983. A handbook of inductively coupled plasma spectrometry.
Blackie, London.
ICP EMISSION SPECTROMETRY 139
Thompson, J.J., and R.S. Houk. 1986. Inductively coupled plasma mass spectrometric detection for
multielement flow injection analysis and elemental speciation by reversed- phase liquid chro-
matography. Anal. Chern. 58:2541.
Thompson, J.J., and R.S. Houk. 1987. Study of internal standardization in inductively coupled plas-
ma-mass spectrometry. Appl. Spectrosc. 41:801.
Tsumura, A., and S. Yamasaki. 1991. Determination of ultra-trace levels of rare earth elements in ter-
restrial water by high resolution ICP-MS with an ultrasonic nebulizer. p. 119-129. In G.
Holland and AN. Eaton (ed.) Applications of plasma source mass spectrometry. R. Soc.
Chern., Cambridge, England.
Ure, AM., AT. Ellis, and J.G. Williams. 1988. Atomic spectrometry update-inorganic mass spec-
trometry and x-ray fluorescence spectrometry. J. Anal. At. Spectrom. 3:174R.
U.S. Department of Commerce National Technical Information Service. 1977. Evaluation of an
inductively coupled plasma, multichannel spectrometric analysis system. Ames Lab., Ames,
Iowa, and Environ. Res. Lab., Athens, GA Anal. Chern. Branch PB-270-918.
Van Heuzen, AA., T. Hoekstra, and B. van Wingerden. 1989. Precision and accuracy attainable with
isotope dilution analysis applied to inductively coupled plasma mass spectrometry: Theory
and experiments. J. Anal. At. Spectrom. 4:483.
Vaughan, M.A., and G. Horlick. 1986. Oxides, hydroxide and doubly-charged analyte species in
inductively coupled plasma/mass spectrometry. Appl. Spectrosc. 40:434.
Veillon, C., and M. Margoshes. 1968. A pneumatic solution nebulization system producing dry
aerosol for spectroscopy. Spectrochim. Acta 23B:553.
Viczian, M., A Lasztity, X. Wang and R.M. Barnes. 1990. On-line isotope dilution and sample dilu-
tion by flow injection and inductively coupled plasma mass spectrometry. J. Anal. At.
Spectrom.5:125.
Ward, A.F. 1978a. Inductively coupled argon plasma spectroscopy. Development, technique, and
applications. Am. Lab. 10:79-87.
Ward, AF. 1978b. Stock standard preparation. Jarrell-Ash Plasma Newslett. 1(2):14-15.
Ward, AF., and R.B. Myers. 1979. The effect of concomitant species upon the analyte signal
observed from a modem ICAP direct reading spectrometer. ICP Inform. Newslett. 4(9):403.
Weast, R.C., and MJ. Astle. (ed.). 1979. CRC handbook of chemistry and physics. 59th ed. CRC
Press, Boca Raton, FL.
Winge, R.K., V.I. Peterson, and V.A. Fassel. 1979. Inductively coupled plasma-atomic emission
spectroscopy. Prominent lines. Appl. Spectrosc. 33:206-219.
Winge, RK, D.E. Eckels, E.L. DeKalb and V.A. Fassel. 1988. Spatiotemporal characteristics of the
inductively coupled plasma. J. Anal. At. Spect. 3:849.
Winge, R.K., J.S. Crain, and R.S. Houk. 1991. High speed photographic study of plasma fluctuations
and intact aerosol particles or droplets in inductively coupled plasma mass spectrometry. J.
Anal. At. Spect. 6:6Ol.
Wolcott, J.F., and C.S. Butler. 1979. A simple nebulizer for the inductively coupled plasma source.
ICP Inform. Newslett. 4(10):461.
Workman, S.M., and P.N. Soltanpour. 1980. Importance of pre-reducing selenium (VI) to selenium
(IV) and decomposing organic matter in soil extracts prior to determination of selenium using
hydride generation. Soil Sci. Soc. Am. J. 44:1331-1333.