Sie sind auf Seite 1von 6

An experimental extrapolation technique using the Gafchromic EBT3

film for relative output factor measurements in small x-ray fields


Johnny E. Moralesa)
Department of Radiation Oncology, Chris OBrien Lifehouse, 119-143 Missenden Road,
Camperdown, NSW 2050, Australia and School of Chemistry, Physics, and Mechanical Engineering,
Queensland University of Technology, Level 4 O Block, Gardens Point, QLD 4001, Australia

Martin Butson
Department of Radiation Oncology, Chris OBrien Lifehouse, 119-143 Missenden Road,
Camperdown, NSW 2050, Australia and Institute of Medical Physics, University of Sydney,
NSW 2006, Australia

Scott B. Crowe
School of Chemistry, Physics, and Mechanical Engineering, Queensland University of Technology,
Level 4 O Block, Gardens Point, QLD 4001, Australia and Cancer Care Services, Royal Brisbane
and Womens Hospital, Butterfield Street, Herston, QLD 4029, Australia

Robin Hill
Department of Radiation Oncology, Chris OBrien Lifehouse, 119-143 Missenden Road,
Camperdown, NSW 2050, Australia and Institute of Medical Physics, University of Sydney,
NSW 2006, Australia

J. V. Trapp
School of Chemistry, Physics, and Mechanical Engineering, Queensland University of Technology,
Level 4 O Block, Gardens Point, QLD 4001, Australia
(Received 17 February 2016; revised 27 June 2016; accepted for publication 29 June 2016;
published 19 July 2016)
Purpose: An experimental extrapolation technique is presented, which can be used to determine the
relative output factors for very small x-ray fields using the Gafchromic EBT3 film.
Methods: Relative output factors were measured for the Brainlab SRS cones ranging in diameters
from 4 to 30 mm2 on a Novalis Trilogy linear accelerator with 6 MV SRS x-rays. The relative output
factor was determined from an experimental reducing circular region of interest (ROI) extrapolation
technique developed to remove the effects of volume averaging. This was achieved by scanning the
EBT3 film measurements with a high scanning resolution of 1200 dpi. From the high resolution
scans, the size of the circular regions of interest was varied to produce a plot of relative output factors
versus area of analysis. The plot was then extrapolated to zero to determine the relative output factor
corresponding to zero volume.
Results: Results have shown that for a 4 mm field size, the extrapolated relative output factor was
measured as a value of 0.651 0.018 as compared to 0.639 0.019 and 0.633 0.021 for 0.5 and
1.0 mm diameter of analysis values, respectively. This showed a change in the relative output factors
of 1.8% and 2.8% at these comparative regions of interest sizes. In comparison, the 25 mm cone
had negligible differences in the measured output factor between zero extrapolation, 0.5 and 1.0 mm
diameter ROIs, respectively.
Conclusions: This work shows that for very small fields such as 4.0 mm cone sizes, a measureable
difference can be seen in the relative output factor based on the circular ROI and the size of the
area of analysis using radiochromic film dosimetry. The authors recommend to scan the Gafchromic
EBT3 film at a resolution of 1200 dpi for cone sizes less than 7.5 mm and to utilize an extrapolation
technique for the output factor measurements of very small field dosimetry. C 2016 American
Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4958679]

Key words: small x-ray field dosimetry, Gafchromic EBT3 film, stereotactic radiosurgery, extrapola-
tion technique

1. INTRODUCTION nontissue equivalence of the detector active volume, and


overlapping penumbra.1,2 Therefore, the selection of a suitable
Accurate small x-ray field dosimetry is crucial in the area detector is crucial and should have minimal corrections
of stereotactic radiotherapy. There are a number of factors for each of these factors where possible. A number of
that can influence the accuracy of dosimetry measurements commercially available detectors introduce volume averaging
such as source occlusion, lack of charged particle equilibrium, effects in the field during measurements and a correction might

4687 Med. Phys. 43 (8), August 2016 0094-2405/2016/43(8)/4687/6/$30.00 2016 Am. Assoc. Phys. Med. 4687
4688 Morales et al.: An experimental extrapolation technique using Gafchromic EBT3 film 4688

be needed as shown by some studies.3,4 Volume averaging a PC desktop scanner and ImageJ (National Institutes of
effects are specifically due to the profile shape of the small Health, USA) software on a PC workstation at least 24 h
field beams and the influence of maximum dose deposition after irradiation to minimize effects from post-irradiation
that is averaged over a given area within the small field beams.3 coloration.21 The scanner was an Epson 10000XL dual lens
Some methods have been proposed to deal with volume system desktop scanner (Epson, NSW, Australia) using a
averaging effects in small field dosimetry including using scanning resolution of 1200 pixels/in. The images produced
film profiles to account for the shape of the small field3,4 or were 48 bit RGB color images and analyzed with the red
by performing full geometry calculations using Monte Carlo component of the signal making the final pixel density values
methods to obtain correction factors.5,6 However, Monte Carlo 16 bit information.22
simulations can be an arduous task requiring a full model of A control film was scanned with every experimental film
the treatment machine, a process not readily available to many in the same position for each measurement. The resulting
radiation oncology departments.510 scans were then corrected for any interscan variations based
Previous studies have been performed using radiochromic on the control films result compared to the average result in
film for small field dosimetry such as the work by Gonzalez- a technique similar to that used by Lewis and Devic.23
Lopez et al.1113 In this study, we introduce an experimental The dose delivered to each experimental film was calcu-
extrapolation technique using the Gafchromic EBT3 film lated by the creation of a calibration doseresponse curve for
(Ashland, NJ, USA), which can potentially eliminate the the Gafchromic EBT3 film by using standard x-ray fields of
volume effect in small field dosimetry. This technique consists 10 10 cm at given applied dose levels. This was performed
of varying the size of the region of interest (ROI) within the due to the known nonlinear relationship of net optical density
scanned area of the film. A method is presented which uses a to doseresponse of the Gafchromic EBT3 film when scanned
zero area extrapolation technique utilizing the high resolution using an Epson10000XL desktop scanner.
scanning in order to determine the final relative output factors
for very small fields.

2. MATERIALS AND METHODS


All relative output factor measurements were performed on
the 6 MV SRS x-ray beam as produced by a Novalis Trilogy
linear accelerator (Varian Medical Systems, Palo Alto, USA),
which has a thin flattening filter in order to produce a dose
rate of 1000 MU/min.1416 Beam collimation for the SRS x-ray
beams was achieved by using the Brainlab circular applicators
(Brainlab, Germany) with diameters between 4 and 30 mm as
defined at the isocenter. The X and Y collimator jaws were
set to 5 cm for all measurements. The relative output factor
is defined as the measurement for a given cone at a depth
of 1.5 cm and SSD of 98.5 cm divided by the measurement
for the reference field size at a depth of 1.40 cm and SSD
of 100 cm as per the manufacturers recommendations. The
reference field size was 10 10 cm2. Monitor units delivered
to each film ranged between 180 and 350 MU depending on
the cone size in an attempt to deliver absorbed doses close to
200 cGy to each film. Nine experimental films were irradiated
separately for each cone size evaluated from which the average
output factor and uncertainties were calculated.

2.A. Film dosimetry technique


Gafchromic EBT3 film (lot number 09151402) was used
for all relative output factor measurements in this work.
All pieces of film were used and handled in the process
as outlined in the AAPM TG-55 report17 and the Medical F. 1. (a) Measured and extrapolated output factor values for a 4 mm
Radiation dosimetry with radiochromic film report series.18 Brainlab cone using the Gafchromic EBT3 film with varying sizes of analysis
area. As the area of analysis decreases, an increase in measured output
It has been shown that the Gafchromic EBT3 film possesses
factor occurs due to the nonplateauing nature of the 4 mm cone profile. (b)
a minimal x-ray energy dependence19,20 and therefore should Measured net optical density profile for a 6 MV SRS x-ray beam, 4 mm cone
have a minimal impact on output factor assessment at small taken as the average of nine EBT3 film measurements. The inset in the figure
field sizes using 6 MV x-rays. All films were analyzed using includes details of the center 1 mm of the profile.

Medical Physics, Vol. 43, No. 8, August 2016


4689 Morales et al.: An experimental extrapolation technique using Gafchromic EBT3 film 4689

F. 3. Percentage difference in the relative output factor from the extrapo-


lated zero area value for the 4 and 25 mm diameter Brainlab cones. Variations
are seen using the 4 mm diameter cone, but negligible differences calculated
at larger cone sizes.

Relative output factor measurements determined with


a PTW 60019 microDiamond detector2428 (PTW, PTW-
Freiburg, Germany) have also been included for compar-
ison with the extrapolated Gafchromic EBT3 results. These
measurements were made under the same geometrical condi-
tions as the measurements for the Gafchromic EBT3 film.

3. RESULTS
F. 2. (a) Measured and extrapolated output factor values for a 25 mm diam-
eter Brainlab cone using the Gafchromic EBT3 film with varying diameters Figure 1(a) shows the measured values for the relative
of analysis area. As the area of analysis decreases, negligible differences in output factor for a 4 mm cone when different size circular
the measured output factor occur due to the plateauing nature of the 25 mm regions of interest ranging from 1.8 to 0.1 mm were used for
cone profile. (b) Measured net optical density profile for a 6 MV x-ray beam,
25 mm diameter Brainlab cone. Inset in the picture includes details of the
the analysis in ImageJ. The figure also shows the extrapolation
center 4 mm of the profile, for example, film showing the central plateau estimate of the relative output factor for a zero volume or
effect at this field size. area calculation, which was found to be 0.651 0.018 for the
4 mm Brainlab cone. The extrapolation was performed using
a second order polynomial line of best fit to provide the best
The net optical densities were calculated for each film piece estimate at zero volume output factor. The variation in the
using the circular regions of interest in ImageJ software. The measured output factor with different regions of interest areas
circular region of interest was centered on the cone produced is expected to change due to a number of reasons including the
radiation field. The relative output factors versus diameter size profile shape of the beam and the light scattering properties
was plotted in a curve. A best fit extrapolation technique was within the scanner and film. The uncertainty values quoted
applied to the results to determine the zero area output factor in the figures and text are the standard deviation in measured
utilizing a second order polynomial or linear function based results comparing the nine experimental films assessed for
on cone size and data requirements. each cone size measured. These values combine both type A

T I. Measured and extrapolated relative output factors for the Gafchromic EBT3 film with different analysis
diameters and for the PTW 60019 microDiamond detector.

PTW
60019 Gafchromic EBT3 filmAnalysis diameter % Difference
Cone size (mm) Measured 0 mm Extrapolated 0.5 mm Measured 1 mm Measured 00.5 mm 01 mm

4 0.661 0.651 0.639 0.633 1.8 2.8


7.5 0.821 0.810 0.804 0.803 0.7 0.9
10 0.879 0.870 0.869 0.866 0.1 0.5
12.5 0.912 0.903 0.902 0.901 0.1 0.2
15 0.932 0.932 0.932 0.931 0.0 0.1
17.5 0.946 0.947 0.948 0.947 0.1 0.0
20 0.956 0.960 0.959 0.959 0.1 0.1
25 0.964 0.971 0.970 0.970 0.1 0.1
30 0.975 0.981 0.980 0.981 0.0 0.1

Medical Physics, Vol. 43, No. 8, August 2016


4690 Morales et al.: An experimental extrapolation technique using Gafchromic EBT3 film 4690

and type B errors associated with the setup and dose delivery
uncertainty along with the experimental film analysis errors
combined. Results are quoted as 1 standard deviation of the
mean.
Figure 1(b) shows an example of net optical density profile
for one 4 mm cone as measured by the EBT3 film at a
resolution of 1200 dpi, which is the average of nine EBT3
film measurements. The central axis region of the profile is
also shown in more detail to highlight both the variation in
F. 4. Exaggerated example of a small field. The average value of pixels in
the net optical density and the uncertainty or noise level with
ROI A will be greater than that of ROI B.
the film scan for this typical measurement. In terms of very
small region of interest analysis, the user should make multiple
measurements using different films to minimize the impact of authors note that they are of similar size to those as shown
selecting a small region of interest for analysis around either in Figs. 1 and 2. Shown are the measured values at 0.5 and
a noise peak or trough, thus skewing the measured dose level 1 mm diameter circular ROI areas of analysis. Also shown are
either high or low by the magnitude of the noise which in our the percentage differences or percentage under prediction in
case was found to be between 0.5% and 1.5%. relative output factors which would occur if the measurements
Figures 2(a) and 2(b) show similar results but for a 25 mm were performed at either 0.5 or 1 mm effective detector
cone size produced by the same 6 MV SRS beam. The diameter.
extrapolated relative output factor for this cone was calculated These results highlight that even for a 0.5 mm effective
to be 0.971 0.017. The 25 mm diameter cone, however, detector size, the ROF for the 4 mm cone could be
has a negligible difference in the measured output factor underpredicted by approximately 2%. This effect is less than
when the diameter of the circular ROI analysis ranges from 1% for cone sizes of 7.5 mm or greater.
approximately 1.8 mm down to 0.1 mm. This is expected as An analysis of effective detector size compared to scan-
Fig. 2(a) shows that a percentage dose plateau does exist at ning resolution for the Epson 10000XL scanner and the
this field size. However, when the same comparison was made Gafchromic EBT3 film is shown in Table II.
for the very small 4 mm cone, the variation in the measured From these results, we can see that using a scanning
output factor was 2.8%. These results highlight the necessity resolution of 300 dpi and a 0.5 mm circular region of
of high spatial resolution scanning and analysis for very small interest, 27 pixels would be averaged to count toward net OD
(4 mm) cone field measurements. calculation. Whereas, using a scanning resolution of 1200 dpi,
Figure 3 shows the variation in the measured output factor 438 pixels would be used for the analysis. Similarly, for a
for these two circular cones as a percentage decrease when 0.1 mm effective detector size, the values would be 1 pixel
each measured point is compared to the extrapolated zero and 17 pixels, respectively, for 300 dpi and 1200 dpi.
area output factor value. As can be seen, over the range up to
2 mm diameter areas of analysis, an approximate percentage
difference for the 25 mm cone is negligible, whereas for the 4. DISCUSSION
4 mm cone the value rises to approximately 3%. These results For small fields which are almost entirely penumbral and
highlight the significance of the volume averaging effect and thus nonflat at the center of the field, the selection of ROI
effective detector size when dealing with very small fields like size in planar measurements can significantly impact the
a 4 mm cone. measurement result. This is illustrated in Fig. 4, where two
Table I shows the relative output factor determination different ROIs of the same field are selected; the average
when this extrapolation technique is applied to other cone value for net OD of pixels within ROI A will be greater
sizes available. The uncertainties in the film measurements than in ROI B due to the greater number of pixels in ROI
are not quoted to improve clarity in the table; however, the B that are of lesser value. Therefore, obtaining a series of

T II. Analysis of detector effective size versus pixels measured for various resolutions.

Pixels per effective


Scan resolution Size of pixel Equivalent circle detector size (circle)
Dots/in. Dots/mm Square (mm) Diameter (mm) 0.1 mm 0.5 mm

75 2.95 0.339 0.382 0.07 1.71


100 3.93 0.254 0.287 0.12 3.04
150 5.91 0.169 0.191 0.27 6.84
200 7.87 0.127 0.143 0.49 12.20
300 11.81 0.085 0.096 1.10 27.40
600 23.62 0.042 0.048 4.38 109.00
1200 47.24 0.021 0.024 17.53 438.00

Medical Physics, Vol. 43, No. 8, August 2016


4691 Morales et al.: An experimental extrapolation technique using Gafchromic EBT3 film 4691

T III. The DPI scanning resolution used across studies for small field dosimetry.

Study Film type Resolution DPI Year

Wilcox and Daskalov (Ref. 29) EBT 75 2007


Garca-Garduo et al. (Ref. 30) EBT 100 2010
Kairn et al. (Ref. 31) EBT2 72 2011
Aland et al. (Ref. 32) EBT2 75 2011
Huet et al. (Ref. 33) EBT2 150 2012
Fiandra et al. (Ref. 34) EBT2 and EBT3 72 2013
Huet et al. (Ref. 35) EBT3 150 2014
Moignier et al. (Ref. 36) EBT3 150 2014
Morales et al. (Ref. 13) EBT3 150 2014

ROIs of different diameters will enable extrapolation to the These resolutions are appropriate for cone sizes of 10 mm
zero volume to obtain the true output factor. and above; however, x-ray beams with dimensions less than
Table II highlights the standard scanning resolutions found 10 mm require greater resolution. Therefore, a high resolution
on the Epson 10000XL flatbed scanner. In normal clinical extrapolation technique is provided to best estimate the
operations, a scanning resolution of 75300 dots/in. would relative output factors for these very small cone sizes.
normally suffice. However, when dealing with very small fields, The PTW 60019 microDiamond detector has been exten-
the resolution plays a major role. In the table, the conversion to sively studied for small field dosimetry.2428,37,38 For the 4 mm
dots per millimetre is shown along with the pixel square side diameter cone used in this study, the ROF calculated using
size and equivalent circular diameter per pixel. When these the zero area extrapolation technique was in good agreement
values are applied to an area analysis using a circular region to the ROF measured with the microDiamond detector. As
of interest, the number of effective pixels which lie in the area such, we believe this agreement shows the suitability of
of analysis is shown for effective detector diameters of 0.1 and using a zero area technique for these smaller fields in film
0.5 mm. This means that if an equivalent detector size of 0.1 mm dosimetry.
is required, a scanning resolution of 300 dots/in. or below will
only give you one pixel or less information. At 1200 dpi reso-
lution, one would use approximately 1718 pixels for optical 5. CONCLUSIONS
density analysis. Using a high resolution provides substantially The technique presented in this study is useful for the
more data for analysis and subsequently should provide a more analysis of the Gafchromic EBT3 film results in order to elimi-
robust measurement to determine relative output factors and nate volume averaging effects,3 to measure very small relative
uncertainty levels. For this reason, we recommend a scanning output factors, and to correct for volume averaging effects,
resolution of at least 1200 dpi to perform the extrapolation of which can be caused by insufficient scanning resolution.
the relative output factor measurements for very small cones It provides enough detail and guidance to be able to
such as the 4 mm cone. reproduce the Gafchromic EBT3 film measurements with
The importance of dosimetric accuracy for a very small high resolution. It has been shown that very small field
field relative output factor lies in both planning data and relative output factor measurements are dependent on the
experimental dose verification. While planning computer effective detector size. When cones of size 4 mm are used, the
grid sizes which are often larger than the sub-millimeter Gafchromic EBT3 film can be used to measure the relative
measurement size in this work, they still require an accurate output factor. The measurement value is, however, affected
relative maximum dose factor for each cone as their defined by the average volume or effective diameter size of the area
value. For example, the Brainlab iPlan treatment system of analysis and up to 2.8% differences were measured when
utilizes an adaptive grid resolution down to 0.5 mm. By using effective detector sizes ranging from 0.0 (extrapolated)
utilizing the extrapolation technique, as shown in this work, to 1 mm effective diameter sizes. Field sizes ranging from
the peak relative output factor for the very small 4 mm cone 7.5 to 30 mm produced negligible differences for the same
can be accurately measured and given for dose calculation in effective detector diameters.
the planning system.
Radiochromic film has been used more extensively for the
determination of the small field size relative output factors CONFLICT OF INTEREST DISCLOSURE
and the data in Table III show the different DPI resolutions
used in various small field dosimetry studies.2936 None of The authors have no COI to report.
the studies shown in this table have been performed with
a resolution higher than 150 dpi and most of these studies a)Author to whom correspondence should be addressed. Electronic mail:
reported values based on a single scanning resolution. In johnny.morales@lh.org.au
addition, no extrapolation technique was reported to have been 1I. J. Das, G. X. Ding, and A. Ahnesjo, Small fields: Nonequilibrium

used in any of these studies. radiation dosimetry, Med. Phys. 35, 206215 (2008).

Medical Physics, Vol. 43, No. 8, August 2016


4692 Morales et al.: An experimental extrapolation technique using Gafchromic EBT3 film 4692

2S. Kumar, J. D. Fenwick, T. S. Underwood, D. D. Deshpande, A. J. Scott, 20T. A. Brown, K. R. Hogstrom, D. Alvarez, K. L. Matthews 2nd, K. Ham, and

and A. E. Nahum, Breakdown of Bragg-Gray behaviour for low-density J. P. Dugas, Doseresponse curve of EBT, EBT2, and EBT3 radiochromic
detectors under electronic disequilibrium conditions in small megavoltage films to synchrotron-produced monochromatic x-ray beams, Med. Phys.
photon fields, Phys. Med. Biol. 60, 81878212 (2015). 39, 74127417 (2012).
3G. Azangwe, P. Grochowska, D. Georg, J. Izewska, J. Hopfgartner, W. 21T. Cheung, M. J. Butson, and P. K. Yu, Post-irradiation colouration of

Lechner, C. E. Andersen, A. R. Beierholm, J. Helt-Hansen, H. Mizuno, A. Gafchromic EBT radiochromic film, Phys. Med. Biol. 50, N281N285
Fukumura, K. Yajima, C. Gouldstone, P. Sharpe, A. Meghzifene, and H. (2005).
Palmans, Detector to detector corrections: A comprehensive experimental 22M. J. Butson, T. Cheung, and P. K. Yu, Absorption spectra variations of

study of detector specific correction factors for beam output measurements EBT radiochromic film from radiation exposure, Phys. Med. Biol. 50,
for small radiotherapy beams, Med. Phys. 41, 072103 (16pp.) (2014). N135N140 (2005).
4C. Bassinet, C. Huet, S. Derreumaux, G. Brunet, M. Cha, M. Bau- 23D. Lewis and S. Devic, Correcting scan-to-scan response variability for

mann, T. Lacornerie, S. Gaudaire-Josset, F. Trompier, and P. Roch, Small a radiochromic film-based reference dosimetry system, Med. Phys. 42,
fields output factors measurements and correction factors determination for 56925701 (2015).
several detectors for a CyberKnife and linear accelerators equipped with 24A. Chalkley and G. Heyes, Evaluation of a synthetic single-crystal diamond

microMLC and circular cones, Med. Phys. 40, 071725 (13pp.) (2013). detector for relative dosimetry measurements on a CyberKnife, Br. J.
5P. Francescon, W. Kilby, and N. Satariano, Monte Carlo simulated correc- Radiol. 87, 20130768 (2014).
tion factors for output factor measurement with the CyberKnife system 25J. C. Barrett and C. Knill, Monte Carlo calculated correction factors for the

Results for new detectors and correction factor dependence on measurement PTW microDiamond detector in the Gamma Knife-Model C, Med. Phys.
distance and detector orientation, Phys. Med. Biol. 59, N11N17 (2014). 43, 10351044 (2016).
6H. Benmakhlouf, J. Sempau, and P. Andreo, Output correction factors 26P. Mancosu, G. Reggiori, A. Stravato, A. Gaudino, F. Lobefalo, V.

for nine small field detectors in 6 MV radiation therapy photon beams: A Palumbo, P. Navarria, A. Ascolese, P. Picozzi, M. Marinelli, G. Verona-
Monte Carlo study, Med. Phys. 41, 041711 (12pp.) (2014). Rinati, S. Tomatis, and M. Scorsetti, Evaluation of a synthetic single-
7S. Russo, L. Masi, P. Francescon, M. C. Frassanito, M. L. Fumagalli, M. crystal diamond detector for relative dosimetry on the Leksell Gamma
Marinelli, M. D. Falco, A. S. Martinotti, M. Pimpinella, G. Reggiori, G. Knife Perfexion radiosurgery system, Med. Phys. 42, 50355041
Verona Rinati, S. Vigorito, and P. Mancosu, Multicenter evaluation of a (2015).
synthetic single-crystal diamond detector for CyberKnife small field size 27J. M. Larraga-Guterrez, P. Ballesteros-Zebadua, M. Rodriguez-Ponce, O. A.

output factors, Phys. Med. 32, 575581 (2016). Garcia-Garduno, and O. O. G. de la Cruz, Properties of a commercial PTW-
8P. Francescon, W. Kilby, N. Satariano, and S. Cora, Monte Carlo simulated 60019 synthetic diamond detector for the dosimetry of small radiotherapy
correction factors for machine specific reference field dose calibration and beams, Phys. Med. Biol. 60, 905924 (2015).
output factor measurement using fixed and iris collimators on the Cy- 28J. E. Morales, S. B. Crowe, R. Hill, N. Freeman, and J. V. Trapp, Dosimetry

berKnife system, Phys. Med. Biol. 57, 37413758 (2012). of cone-defined stereotactic radiosurgery fields with a commercial synthetic
9G. Cranmer-Sargison, S. Weston, J. A. Evans, N. P. Sidhu, and D. I. diamond detector, Med. Phys. 41, 111702 (6pp.) (2014).
Thwaites, Implementing a newly proposed Monte Carlo based small field 29E. E. Wilcox and G. M. Daskalov, Evaluation of GAFCHROMIC

dosimetry formalism for a comprehensive set of diode detectors, Med. EBT film for CyberKnife dosimetry, Med. Phys. 34, 19671974
Phys. 38, 65926602 (2011). (2007).
10G. Cranmer-Sargison, S. Weston, J. A. Evans, N. P. Sidhu, and D. I. 30O. A. Garca-Garduo, J. M. Lrraga-Gutirrez, M. Rodrguez-Villafuerte,

Thwaites, Monte Carlo modelling of diode detectors for small field A. Martnez-Dvalos, and M. A. Celis, Small photon beam measurements
MV photon dosimetry: Detector model simplification and the sensitivity using radiochromic film and Monte Carlo simulations in a water phantom,
of correction factors to source parameterization, Phys. Med. Biol. 57, Radiother. Oncol. 96, 250253 (2010).
51415153 (2012). 31T. Kairn, S. Crowe, J. Kenny, and J. V. Trapp, Investigation of stereotactic
11A. Gonzalez-Lopez, J. A. Vera-Sanchez, and J. D. Lago-Martin, Small radiotherapy dose using dosimetry film and Monte Carlo simulations,
fields measurements with radiochromic films, J. Med. Phys. 40, 6167 Radiat. Meas. 46, 19851988 (2011).
(2015). 32T. Aland, T. Kairn, and J. Kenny, Evaluation of a Gafchromic EBT2 film
12S. Reinhardt, M. Hillbrand, J. J. Wilkens, and W. Assmann, Comparison of dosimetry system for radiotherapy quality assurance, Australas. Phys. Eng.
Gafchromic EBT2 and EBT3 films for clinical photon and proton beams, Sci. Med. 34, 251260 (2011).
Med. Phys. 39, 52575262 (2012). 33C. Huet, S. Dagois, S. Derreumaux, F. Trompier, C. Chenaf, and I. Robbes,
13J. Morales, R. Hill, S. Crowe, T. Kairn, and J. Trapp, A comparison of Characterization and optimization of EBT2 radiochromic films dosimetry
surface doses for very small field size x-ray beams: Monte Carlo calculations system for precise measurements of output factors in small fields used in
and radiochromic film measurements, Australas. Phys. Eng. Sci. Med. 37, radiotherapy, Radiat. Meas. 47, 4049 (2012).
303309 (2014). 34C. Fiandra, M. Fusella, F. R. Giglioli, A. R. Filippi, C. Mantovani, U.
14Z. Chang, Z. Wang, Q. J. Wu, H. Yan, J. Bowsher, J. Zhang, and F. F. Ricardi, and R. Ragona, Comparison of Gafchromic EBT2 and EBT3 for
Yin, Dosimetric characteristics of novalis Tx system with high definition patient-specific quality assurance: Cranial stereotactic radiosurgery using
multileaf collimator, Med. Phys. 35, 44604463 (2008). volumetric modulated arc therapy with multiple noncoplanar arcs, Med.
15A. Dhabaan, E. Elder, E. Schreibmann, I. Crocker, W. J. Curran, N. M. Phys. 40, 082105 (8pp.) (2013).
Oyesiku, H. K. Shu, and T. Fox, Dosimetric performance of the new high- 35C. Huet, C. Moignier, J. Fontaine, and I. Clairand, Characterization of the

definition multileaf collimator for intracranial stereotactic radiosurgery, J. Gafchromic EBT3 films for dose distribution measurements in stereotactic
Appl. Clin. Med. Phys. 11, 197211 (2010). radiotherapy, Radiat. Meas. 71, 364368 (2014).
16F.-F. Yin, J. Zhu, H. Yan, H. Gaun, R. Hammoud, S. Ryu, and J. H. Kim, 36C. Moignier, C. Huet, V. Barraux, C. Bassinet, M. Baumann, K. Sebe-

Dosimetric characteristics of Novalis shaped beam surgery unit, Med. Mercier, C. Loiseau, A. Batalla, and L. Makovicka, Determination of small
Phys. 29, 17291738 (2002). field output factors and correction factors using a Monte Carlo method for
17A. Niroomand-Rad, C. R. Blackwell, B. M. Coursey, K. P. Gall, J. M. a 1000 MU/min CyberKnife system equipped with fixed collimators,
Galvin, W. L. McLaughlin, A. S. Meigooni, R. Nath, J. E. Rodgers, and Radiat. Meas. 71, 287292 (2014).
C. G. Soares, Radiochromic film dosimetry: Recommendations of AAPM 37T. S. A. Underwood, B. C. Rowland, R. Ferrand, and L. Vieillevi-

Radiation Therapy Committee Task Group 55. American Association of gne, Application of the Exradin W1 scintillator to determine Ediode
Physicists in Medicine, Med. Phys. 25, 20932115 (1998). 60017 and microDiamond 60019 correction factors for relative dosim-
18M. J. Butson, P. K. N. Yu, T. Cheung, and P. Metcalfe, Radiochromic film etry within small MV and FFF fields, Phys. Med. Biol. 60, 66696683
for medical radiation dosimetry, Mater. Sci. Eng., R 41, 61120 (2003). (2015).
19R. Hill, B. Healy, L. Holloway, Z. Kuncic, D. Thwaites, and C. Baldock, 38A. Ralston, M. Tyler, P. Liu, D. McKenzie, and N. Suchowerska, Over-

Advances in kilovoltage x-ray beam dosimetry, Phys. Med. Biol. 59, response of synthetic microDiamond detectors in small radiation fields,
R183R231 (2014). Phys. Med. Biol. 59, 58735881 (2014).

Medical Physics, Vol. 43, No. 8, August 2016

Das könnte Ihnen auch gefallen