Filtration Basics
EN 1822 Short Overview
Solution for EN 1822 conform characterization
of filter media
2
2007, TSI Incorporated
Filter Testing Introduction
Downstream Concentration
Penetration, P =
Upstream Concentration
Efficiency, E = 1 - P
Interception Particles travel along streamlines and are caught because of their
size. Ratio of particle to fiber size is important.
Inertial Particles deviate from the streamlines due to inertia and impact on
the fibers. Large particles, high density, high velocity and small fibers
Impaction is most efficient.
Deposition
Inertial Impaction Electrostatic
Attraction
(-)
(+)
Filter
Fiber
Brownian Diffusion
Interception
4
2007, TSI Incorporated
Filter Efficiency Curve
100%
80%
Efficiency
60% INTERCEPTION
DIFFUSION AND
INERTIAL
IMPACTION
40% DIFFUSION
AND
INTERCEPTION
20%
0%
0.01 0.1 1
Particle Diameter [m]
5
2007, TSI Incorporated
Filter Penetration
100
0.01
0.001
0.01 0.1 1 10
Effect of
Particle Deposition Mechanisms
and Flow Rate (Face Velocity) MPPS
on Penetration and
Diffusion Interception
PENETRATION
Impaction
Most Penetrating Particle Size MPPS
lower velocity
Leads to
Increase of MPPS
PARTICLE SIZE - log scale Decrease of Penetration
7
2007, TSI Incorporated
Filter Testing Standard EN 1822
8
2007, TSI Incorporated
Filter Testing Standard EN 1822
9
2007, TSI Incorporated
Classification by En 1822 1
HEPA & ULPA Filters
High Efficiency
Particle Air Filter
Ultra Low
Penetration Air Filter
10
2007, TSI Incorporated
Example for Decrease of Number
Concentration
1.0E+07
1.0E7 Particle/ccm on raw gas side
1.0E+04
1.0E+03
1.0E+02
1.0E+01
1.0E+00
1.0E-01
1.0E-02
1.0E-03
H10 H11 H12 H13 H14 U15 U16 U17
Filter Type 11
2007, TSI Incorporated
Filter Testing Standard EN 1822
12
2007, TSI Incorporated
1822-3 Testing flat sheet filter media
13
2007, TSI Incorporated
1822-3 Testing flat sheet filter media
14
2007, TSI Incorporated
1822-3 Testing flat sheet filter media
15
2007, TSI Incorporated
50 nm
Atomizer with
0.03% DEHS in Isopropanol
1.6L/min
Neutralizer 3012
16
2007, TSI Incorporated
100 nm
Atomizer with
0.03% DEHS in Isopropanol
1.6L/min
Neutralizer 3012
17
2007, TSI Incorporated
150 nm
Atomizer with
0.03% DEHS in Isopropanol
1.6L/min
Neutralizer 3012
18
2007, TSI Incorporated
Creating Penetration Curve
with Monodisperse Aerosol
Penetration
dn/dlogdp
Particle Diameter
Penetration
logdp
Particle Diameter
19
2007, TSI Incorporated
1822-3 Testing flat sheet filter media
20
2007, TSI Incorporated
Disadvantages of OPCs for
Determination of MMPS
Comparatively poor size resolution (high
resolution needed)
No channels below 200 nm
Some filter media (e.g. electrostatic) can have
MMPS of below 50nm
21
2007, TSI Incorporated
Monodisperse test aerosol and CPC(s)
for determination of MPPS
Biggest advantage:
Excellent coverage of the possible size
range for the MPPS.
Disadvantage: More time needed to carry
out test
22
2007, TSI Incorporated
Instrument Solution
Model 3160 Fractional Efficiency Filter Tester
Oil Atomizers
Bleed Flow
System Downstream
CPC
24
2007, TSI Incorporated
Instrument Solution (2)
25
2007, TSI Incorporated
3160 Printouts
26
2007, TSI Incorporated
Applications
27
2007, TSI Incorporated
3160 Operation and Setup Tips
28
2007, TSI Incorporated
3160 Calibration
Instrument uses electrical mobility which is the
standard method for checking submicron particle
sizes and CPCs which are the concentration
standard
Flows and pressure drops can be independently
verified
Field Installation is provided with new instruments
to verify performance
We provide Field Service and Service Contracts
when requested by customers
29
2007, TSI Incorporated
3160 Features and Benefits
Feature Benefit
Automated Fractional Efficiency test Simpler and easier than developing this system
from components
Includes computer and Windows software Familiar Windows look and feel minimizes operation
for current operating systems and training time
Flows and size controls are automatic or Doesnt require additional equipment or lots of time
simple to use to set up tests
New expanded size range up to 0.8 Provides information that previously required a
microns different tester
Fast Test Mode Adds ability to use an instrument designed for
fractional efficiency for fast quality control tests
Automatically adjusts concentration and Eliminates the need to monitor the test for
test time to get the best results within the concentrations that are higher or lower than desired
parameters you set
30