Sie sind auf Seite 1von 2

Instruction Manual Dear Customer, Specifications Display Code Guide

Thank you for choosing a Hanna product. This manual will


HI 93701 provide you with the necessary information for the correct
This indicates that the meter is in a
operation of the meter. Please read it carefully before using
Free Chlorine ISM the meter. If you need additional technical information, do ready state and zeroing can be per-
not hesitate to e-mail us at tech@hannainst.com. formed.
This instrument is in compliance with directives. Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
Preliminary Examination This indicates that the meter is in a
zeroed state and measurement can be
Remove the instrument from the packing material and performed.
examine it carefully to make sure that no damage has A zero reading was not taken. Insert a
occurred during shipment. If there is any damage, notify sample before adding reagent and press
your Dealer. ZERO.
Each Ion Specific Meter is supplied complete with Under range. A blinking "0.00" indicates
9V Battery SPECIFICATIONS that the sample absorbs less light than
Two Sample Cuvets and Caps Range 0.00 to 2.50 mg/L the zero reference. Check the procedure
One Transport Cap Resolution 0.01 mg/L and make sure you use the same cuvet
Note: Conserve all packing material until the instrument Accuracy 0.03 mg/L 3% of reading for reference (zero) and measurement.
has been observed to function correctly. Any defective Typical EMC Over range. A flashing value higher than
This Instrument is in item must be returned in its original packing. Deviation 0.01 mg/L the maximum readable concentration (see
www.hannainst.com Compliance with the CE Directives
Light Source Light Emitting Diode @ 555 nm specifications) indicates that the sample
Method Adaptation of the EPA recommended absorbs too much light, meaning that
DPD method 330.5. The reaction between
General Description free chlorine and the DPD reagent causes
the concentration is too high. Dilute the
sample.
Warranty The HI 93701 portable microprocessor meter measures the
a pink tint in the sample
Light over range. The cuvet is not in-
Light Detector Silicon Photocell
HI 93701 is warranted for two years against defects in workman- free chlorine (Cl2) content in water and wastewater in the serted correctly and an excess ambient
Environment 0 to 50C (32 to 122F);
ship and materials when used for its intended purpose and maintained 0.00 to 2.50 mg/L (ppm) range. light is reaching the detector. If the cover
max 95% RH non-condensing
according to instructions. The meter uses an exclusive positive-locking system to ensure is properly installed, then contact your
This warranty is limited to repair or replacement free of charge.
Battery Type/Life 1 x 9 volt/40 hours
that the cuvet is in the same place every time it is placed Auto-Shut off After 10' of non-use dealer or the nearest Hanna Customer
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
into the measurement cell. Dimensions 80 x 83 x 46 mm (7.1 x 3.3 x 1.8") Service Center.
If service is required, contact the dealer from whom you purchased the The reagents are in liquid or powder form and are supplied Weight 290 g (10 oz.) Light under range. The zero sample is
instrument. If under warranty, report the model number, date of pur- in bottles or in packets. The amount of reagent is precisely too dark for proper zeroing. If this is not
chase, serial number and the nature of the failure. If the repair is not dosed to ensure maximum repeatability. the case, contact your dealer or the
covered by the warranty, you will be notified of the charges incurred. If REQUIRED REAGENTS nearest Hanna Customer Service Center.
the instrument is to be returned to Hanna Instruments, first obtain a Display codes aid the user in routine operations.
Returned Goods Authorization Number from the Customer Service de- The meters have an auto-shut off feature that will turn the POWDER: The "V" indicates that the battery volt-
partment and then send it with shipment costs prepaid. When shipping Code Description Quantity age is getting low and the battery needs
any instrument, make sure it is properly packaged for complete protec-
instrument off after 10 minutes of non-use.
HI 93701-0 DPD 1 packet to be replaced.
tion.
To validate your warranty, fill out and return the enclosed warranty card LIQUID: This indicates that the battery is dead
within 14 days from the date of purchase. Code Description Quantity and must be replaced.
All rights are reserved. Reproduction in whole or in part is prohibited HI 93701A-F DPD1 Indicator 3 drops Note: once this indication is displayed,
without the written consent of the copyright owner, Hanna Instru- HI 93701B-F DPD1 Buffer 3 drops the meter will lockup. Change the bat-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
tery to restart.
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.

Downloaded from www.Manualslib.com manuals search engine


Operational Guide The instrument directly displays concentration in mg/L of Tips for an Accurate Measurement Accessories
free chlorine on the Liquid Crystal Display.
MEASUREMENT PROCEDURE The instruction listed below should be carefully followed
ON REAGENT SETS
Turn the meter on by pressing Liquid reagents procedure during testing to ensure best accuracy.
ON/OFF.
OFF HI 93701-F Reagents for 300 tests (liquid)
To an empty cuvet add Do not touch the cuvet walls with hands.
HI 93701-01 Reagents for 100 tests (powder)
3 drops of HI 93701A-F In order to maintain the same conditions during the
When the LCD displays - - -, it zeroing and the measuring phases, it is necessary to HI 93701-03 Reagents for 300 tests (powder)
DPD1 indicator, 3 drops
is ready. of HI 93701B-F DPD1 close the cuvet to prevent any contamination. OTHER ACCESSORIES
buffer. Swirl gently to Do not let the test sample stand too long after reagent HI 710009 Blue rubber boot
Fill the cuvet with 10 mL of 10 mL mix and immediately add is added or accuracy will be lost. HI 710010 Orange rubber boot
unreacted sample, up to the 10 mL of unreacted Whenever the cuvet is placed into the measurement cell, HI 721310 9V battery (10 pcs)
mark, and replace the cap. sample. it must be completely free of fingerprints, oil or dirt.
HI 731318 Tissue for wiping cuvets (4 pcs)
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion. HI 731321 Glass cuvets (4 pcs)
Place the cuvet into the holder
It is important that the sample does not contain any HI 731325 Caps for cuvets (4 pcs)
and ensure that the notch on
the cap is positioned securely into debris. This would corrupt the readings. HI 93703-50 Cuvets cleaning solution (230 mL)
Replace the cap and
the groove. shake gently again. It is possible to take multiple readings in a row, but it
Press ZERO and SIP will appear on the display. is recommended that a zero reading be taken for each CE Declaration of Conformity
sample and that the same cuvet is used for zeroing and
measurement.
ZERO Reinsert the cuvet into
It is important to discard the sample immediately after
the instrument.
the reading is taken because the glass might become
permanently stained.
Wait for a few seconds and the Wait for 1 minute and then press READ DIRECT. The Shaking the cuvet can generate bubbles in the sample,
display will show -0.0-. Now display will show SIP during measurements. causing higher readings. To obtain accurate measure-
the meter is zeroed and ready for ments, remove such bubbles by swirling or by gently
measurement. READ tapping the vial.
DIRECT
Remove the cuvet. All the reaction times reported in this manual are
referred to 20C (68F). As a general rule of thumb,
Powder reagent procedure they should be doubled at 10C (50F) and halved at
The instrument directly displays concentration in mg/L of 30C (86F).
Add the content of one packet of free chlorine on the Liquid Crystal Display.
HI 93701 DPD reagent. Replace
the cap and shake gently for 20 INTERFERENCES
seconds (or 2 minutes in case of Interference may be caused by:
Battery Replacement
seawater analysis). Bromine Iodine
Replace the cuvet into the holder Battery replacement must only take place in a non-hazard-
Chlorine dioxide Ozone Recommendations for Users
and ensure that the notch on ous area using a 9V alkaline battery.
Hydrogen peroxide Before using these products, make sure that they are entirely suitable for
the cap is positioned securely Oxidized Chromium (Cr6+) and Manganese (Mn4+, Mn7+). Simply slide off the battery cover on the back of the meter. the environment in which they are used.
into the groove. Alkalinity above 250 mg/L or acidity above 150 mg/L will Detach the battery from the terminals and attach a fresh Operation of these instruments in residential area could cause unaccept-
Wait for 1 minute and press READ DIRECT. The display not reliably develop the full amount of color or it may 9V battery while paying attention to the correct polarity. able interferences to radio and TV equipments, requiring the operator to
will show SIP during measurements. rapidly fade. To resolve this, neutralize the sample with Replace the battery and the cover. take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may
diluted HCl or NaOH. degrade the instruments' EMC performance.
READ
DIRECT In case of water with hardness greater than 500 mg/L 9V To avoid damages or burns, do not perform any measurement in micro-
CaCO3, shake the sample for approximately 1 minute after SLIDE
wave ovens.
adding the powder reagent. OFF
IST93701R4 09/03

Downloaded from www.Manualslib.com manuals search engine