Sie sind auf Seite 1von 9

Gas Chromatography

In gas chromatography (GC), the stationary phase is a high-boiling liquid and the mobile phase is an inert gas. In the organic
chemistry teaching labs at CU Boulder, GC is used as an analytical tool to find out how many components are in a mixture. It can also
be used to separate small amounts of material.

The GC Instrument

The process of gas chromatography is carried out in a specially designed instrument. A very small amount of liquid mixture is injected
into the instrument and is volatilized in a hot injection chamber. Then, it is swept by a stream of inert carrier gas through a heated
column which contains the stationary, high-boiling liquid. As the mixture travels through this column, its components go back and
forth at different rates between the gas phase and dissolution in the high-boiling liquid, and thus separate into pure components. Just
before each compound exits the instrument, it passes through a detector. When the detector “sees” a compound, it sends an electronic
message to the recorder, which responds by printing a peak on a piece of paper.

The two brands of GCs used in the organic chemistry teaching labs are shown below: Gow-Mac series 350 on the left, Varian
Aerograph Model 920 on the right. Click on each photo for a detailed enlargement.

The GC consists of an injection block, a column, and a detector. An inert gas flows through the system. The injection chamber is a
heated cavity which serves to volatilize the compounds. The sample is injected by syringe into this chamber through a port which is
covered by a rubber septum. Once inside, the sample becomes vaporized and is carried out of the chamber and onto the column by the
carrier gas.

The large photo below is of the inside of a GC, showing the column in the oven and the insulated chamber tht houses the detector.
Click on the thumbnails to see larger photos of the column and detector, as well as the inside of the injector port (showing the septum).
inside of the injector port

the septum

the column

the detector inside the


housing

On the Varian 920 and Gow Mac


350 chromatographs, detection of
the compounds is achieved with a
thermal conductivity (TC or hot
wire) detector.

The column (see the photo above) is an integral part of the GC system. On the outside, all you see is a long stainless steel tube, 1/8 to
1/4 inch in diameter and 4-5 feet long, which is coiled to fit inside the instrument. Inside the column is the important component: the
stationary phase composed of the high-boiling liquid. The liquid is usually impregnated on a high surface area solid support like
diatomaceous earth, crushed firebrick, or alumina. The liquid can be applied in various concentrations: the more liquid, the more sites
it has to interact with the compounds.

All of our GCs have columns which are five feet long and 1/8" or 1/4" in diameter and contain a methyl silicone polymer liquid phase
(OV-101, 1.5%) on a diatomaceous earth support (chromosorb G). Methyl silicone is a liquid phase of intermediate polarity, and non-
polar compounds such will separate according to their respective boiling points.

The carrier gas is an inert gas, helium. The flow rate of the gas influences how fast a compound will travel through the column; the
faster the flowrate, the lower the retention time. Generally, the flow rate is held constant throughout a run. (The GCs at CU Boulder
are set at a flow rate of 55 mL/min.)
This is where the carrier gas enters the Varian GCs and where the gas flow rate
can be adjusted. Click on the photo above for details.

In a professional laboratory, the GC conditions would be critical for another experimenter trying to duplicate your observations. All of
our GCs have the same columns (1.5% OV-101 on Chromasorb G) and the same flow rate (55 mL/minute) and detector bridge current
(150 mAmps). Each instrument will have a different setting for:

• column temperature
• injection port temperature
• detector temperature

It is a good practice to write down some of the settings on the instrument. The values for these temperaturs are viewed by turning the
knob on the instrument below the gauge -- click on the thumbnails below to see detailed photos of how to do this.

reading temperatures on the Gow-Mac

reading temperatures on the Varian

Recorders

Two devices are used to record the GC traces/areas under peaks:

• integrating recorders
• computer program

Each type of device records the messages sent to them by the detector as peaks, calculates the retention time, and calculates the area
under each peak; all of this information is included in the printout. For similar compounds, the area under a GC peak is roughly
proportional to the amount of compound injected. If a two-component mixture gives relative areas of 75:25, you may conclude that the
mixture contains approximately 75% of one component and 25% of the other.

An integrating recorder is pictured below. Click on the photo for a detailed picture and the location of the start button (press when
you inject), the stop button (press when you have seen your peaks, it tells the recorder to do the calculations and to print), and the
enter button (paper feed).
The screen of one of the computers is pictured below. Click on this image to link to screen shots of how to start, stop, calculate, and
print a GC trace using the computer program.

Retention Time (RT)

The retention time, RT, is the time it takes for a compound to travel from the injection port to the detector; it is reported in minutes on
our GCs. The retention time is measured by the recorder as the time between the moment you press start and the time the detector sees
a peak. If you do not press start at the same time you inject your sample, the RT values will not be consistent from run to run.

Factors which affect GC separations

Efficient separation of compounds in GC is dependent on the compounds traveling through the column at different rates. The rate at
which a compound travels through a particular GC system depends on the factors listed below:

• Volatility of compound: Low boiling (volatile) components will travel faster through the column than will high boiling
components
• Polarity of compounds: Polar compounds will move more slowly, especially if the column is polar.
• Column temperature: Raising the column temperature speeds up all the compounds in a mixture.
• Column packing polarity: Usually, all compounds will move slower on polar columns, but polar compounds will show a
larger effect.
• Flow rate of the gas through the column: Speeding up the carrier gas flow increases the speed with which all compounds
move through the column.
• Length of the column: The longer the column, the longer it will take all compounds to elute. Longer columns are
employed to obtain better separation.

Generally the number one factor to consider in separation of compounds on the GCs in the teaching labs is the boiling points of the
different components. Differences in polarity of the compounds is only important if you are separating a mixture of compounds
which have widely different polarities. Column temperature, the polarity of the column, flow rate, and length of a column are constant
in GC runs in the Organic Chemistry Teaching Labs. For each planned GC experiment, these factors have been optimized to separate
your compounds and the instrument set up by the staff.

Procedure for GC

Gas Chromatography: Procedure

(1) Add the sample to be injected to the syringe.

A 25µL glass Hamilton syringe is used to inject the GC samples. Only 2-4 µL of sample
is injected onto the column, which means that you fill only a small part of the barrel with
sample. Examine the syringe carefully before you fill it. The divisions are marked "5 - 10
- 15 - 20 - 25".

This is a 25 µL glass Hamilton syringe. You only


inject 2.5 µL, so it will NOT be filled to the top.

Click on the photo to see an enlargement.

Place the tip of the needle in the liquid. Slowly draw up a small amount of liquid by
raising the plunger, then press on the plunger to expel the liquid back into the liquid. This
serves to “rinse” the syringe with your sample, ensuring that what you will measure in the
GC run is the composition of your mixture. Repeat the rinse process one or two times.
Then, draw up the plunger slowly again while the needle is in the liquid and carefully fill
the syringe with liquid about halfway to the “5”.
It is often hard to see the liquid in the
syringe. If the syringe is clogged, the
plunger will be in the correct position but
the barrel of the syringe will be filled
with only air, as in the bottom syringe in
the photo to the left.

The best thing to do is to carefully


examing the syringe after you think that
you have filled it. Hold it up to the light
to get a better view.

Small air bubbles in the syringe will not


affect the GC run (middle syringe in the
photo to the left). As long as there is
enough liquid in the syringe, the GC run
will work fine. If you keep getting
bubbles, just pull the plunger up a bit
past the "halfway to the 5" mark to
compensate.

If you have a VERY large air bubble,


you will not have enough liquid to show
a reading on the GC (e.g., the bottom
syringe in the photo).

(2) Inject the sample into the injector port.

You are need to do two things sequentially and quickly, so make sure you know where
the injection port is and where the start button on the recorder is.

Push the needle of the syringe through the injection port and immediately press the
plunger to inject the sample, then immediately press the start button on the recorder.

You will feel a bit of resistance from the rubber septum in the injection port; this is to be
expected and you should be prepared to apply some pressure to the syringe as you force
the needle into the instrument all the way to the base of the needle.
(Click on each of the photos below for a larger view.)

Push the needle of the Remove the . . . and quickly press Here's a close-up of
filled syringe through syringe the start button on the the integrating
the injector (as far as immediately . . . integrating recorder recorder.
it will go) and quickly or the start recording
push the plunger. button on the
computer (ask your
TA which device is
connected to the GC
that you are using)
Here's a series of
pictures showing how
to run the computer
program.

(3) Sit back and wait.

Observe the recorder. Within several minutes, it should record several peaks.

(4) End the GC run.

When you have seen all of the peaks which you suspect are in the mixture, or when the
recorder has shown a flat baseline for a few minutes or so, press stop on the recorder.

When you press stop, the recorder will print out the peaks, the retention times, and the
areas under the peaks. When it is done printing, you can press “enter” a couple times to
advance the paper.

Carefully tear the paper off the recorder. The paper is not perforated, so do not try to pull
up and expect it to pop out of the recorder. Instead, pull it down to start a tear from one
edge, and then continue the tear until the paper is cut and free.
This may seem trivial -- showing you
how to tear the paper. But too many
times a student has tried to yank the
paper out instead of starting a tear and
tearing it neatly. Yanking the paper can
result in the paper being torn below the
plastic cutting surface on the recorder,
and the paper gets jammed down inside
the recorder.

If this happens, the entire recorder has


to be dis-assembled, a process which
takes about 15 minutes, thus putting the
entire GC out of service until it can be
fixed.

Das könnte Ihnen auch gefallen