USERS MANUAL
Version 2.31.01
CONTENTS
CHAPTER 1 MEDEX PRODUCT SUPPORT......................................................................................... 7
2
MedEx DIGITAL EEG USER MANUAL
FILE MENU............................................................................................................................24
RECORD MENU....................................................................................................................27
BOOKMARK MENU...............................................................................................................28
FILTER TOOLBAR.................................................................................................................31
REVIEW TOOLBAR...............................................................................................................32
3
MedEx DIGITAL EEG USER MANUAL
REVIEW IN EVENTS.............................................................................................................57
4
MedEx DIGITAL EEG USER MANUAL
SUBSECTION 3 CSA....................................................................................................................77
5
MedEx DIGITAL EEG USER MANUAL
6
MedEx DIGITAL EEG USER MANUAL
You can get technical support, check on the status of problems, or report new problems by
China
Tel: +86-10-62974855
Fax: +86-10-62978861
E-mail: marketing@medextech.com
info@medextech.com
This file provides information on installing and operating MND-3000 Digital EEG Workstation.
As it is assumed that you are familiar with Windows 2000 above OS, the Microsoft software used
7
MedEx DIGITAL EEG USER MANUAL
The EEG Acquisition Box is certified by CE and meets the requirements of the 93/42/EEC. It is
and complies with the guidelines issued by German Association of Medical Insurance Companies.
The functionality of the EEG Acquisition Box may be disturbed by heavy electromagnetic radiation.
Try to avoid standing right next to mobile phones, microwaves and radio broadcast masts if
possible.
Up to now, when using the device correctly, no side effects and risks are known to the
manufacturer.
SECTION 2 CLASSIFICATION
CE-Certification
Type BF
8
MedEx DIGITAL EEG USER MANUAL
Notice
2. Dont turn them upside down or place them on side during transport and storage.
SECTION 5 STANDARDS
MDD: class II a
9
MedEx DIGITAL EEG USER MANUAL
2. 256MB RAM
2. MND3000 software
6. Computer
7. Printer
Because the EEG acquisition box gets power by USB port, please ensure that the power
10
MedEx DIGITAL EEG USER MANUAL
The MND-3000 EEG system should install the software first and then connect the hardware. It is
not suggested to connect the hardware before installing the software successfully.
The system consists of several software and hardware components, which must be installed in
Step Comment
6 Restart the PC
It is also important to install the software components into defined directories on the hard disk:
Component Directory
Note:
1. All directory statements made in this manual refer to the table above.
2. The installation directory and the data directory can be modified manually during the
installation course.
11
MedEx DIGITAL EEG USER MANUAL
1. Before installing MND-3000 8/16/19 channel EEG system, please make sure that the
2. Place the EEG workstation CD-ROM into the CD-ROM drive of the computer.
directory of the CD-ROM drive, and pop up the following figure: (The file name is different with
different channels.)
4. Click the Next button and the system prepares to install the software. In about some seconds,
12
MedEx DIGITAL EEG USER MANUAL
7. Then there is an install driver window pops up as the following figure. Click the Install button
8. Then the install driver was successfully installed window appears as the following figure.
13
MedEx DIGITAL EEG USER MANUAL
9. Then the following window pops up to suggest MedEx System Support installation. Click the
10. Then the following window pops up. Choose Yes, I want to restart my computer now, and
14
MedEx DIGITAL EEG USER MANUAL
After installing the software, then an icon appears on the desktop. Shut down the PC.
Note: After installing the software, you have to shut down the PC.
The Security Key is used to provide a unique system ID. This system ID is used to generate the
license key. License files now protect the software. These files are stored on EEG workstation
only. When EEG software starts, the license file will be read.
Also, the security key you use will decide what functions your EEG program can provide to you.
That means, if you purchase the software with different functions, the security key you have may
vary.
Note: The Security key must be connected after installing the MND3000 software and before
2. Connect the USB cable of the Acquisition Box to the USB port of the PC.
3. Connect one end of the short grounded cable to the Acquisition Boxs ground cord with the;
15
MedEx DIGITAL EEG USER MANUAL
connect the other end to the PC (which connects to the standard ground later).
Note:
1. The length of USB cable cannot extend 1.5 meter. It is not suggested to connect to the USB
2. Before running MND-3000 system, please ensure that the security key and the EEG Acquisition
Box have been connected to the PC and the Acquisition Boxs power has been switched on.
Photic Stimulator is an option of the system. If it is not ordered, this section can be ignored.
1. Fix the photic stimulator on the stimulator holder. Please note that the stimulator holder is the
3. Connect one end of the USB data cable to the photic stimulator and connect the other end of
Note:
16
MedEx DIGITAL EEG USER MANUAL
2. Before running MND-3000 system, please ensure that the photic stimulator has been
connected to the PC and the photic stimulators power has been switched on.
Video Module is an option of the system. If it is not ordered, this section can be ignored.
acceptable.
After installing the video driver, connect the camera with the PC. Then run the MND3000
software.
Right click on the window, a dialog pops up. Click connect, then the video is connected with the
17
MedEx DIGITAL EEG USER MANUAL
Right click on the video model, the following menu pops up.
Users can adjust preview rate. The higher the video preview rate is, the smoother the video
Users can adjust video compress to select compress programs and compression quality as the
following figure.
18
MedEx DIGITAL EEG USER MANUAL
If users do not want to use Video, right click on the window and click disconnect to close the
video.
When you use the system to do patients EEG tests, please connect the grounded cable.
Connect one end of the long grounded cable to the PC as the following figure and connect the
19
MedEx DIGITAL EEG USER MANUAL
Fp1-F3 is an operation of the value of Fp1 electrode minus the value of F3. Fp1+A1 is
Besides mono-polar and Bi-polar, you can also configure the montages according to your needs.
20
MedEx DIGITAL EEG USER MANUAL
The system uses international 10-20system electrode placing methods. That is the distance
Each electrode has its name. Those locating on the left are with odd numbers and those on the
right are even. Those closing the middle have small numbers and those further have big numbers.
The electrodes name also includes the first letter of the head areas. The following figures show
the electrodes placing and the relative EEG Acquisition Boxs picture.
After making sure the electrodes placing, use acetone to clean the related head place and then
use paste or EEG cap to fix each electrode. Connect each electrode to the relative socket on the
Note:
1. The panel of EEG Acquisition Box has many sockets. Plug the patient leads into the sockets.
Hold the position of the rubber when plug in or out, otherwise it is easy to broke leads.
2. On the panel of the EEG Acquisition Box, there are 2 sockets labeled ISO GND. Just use one
of them and put its electrode on the middle of the forehead, in the area between Fp1 and Fp2.
4. Look carefully at the electrode map, and place the electrode properly as showed in the map.
For example, the electrode plugging in Fp1 should be placed on the left prefrontal, the O2
on right occipital.
5. Any two of the electrodes should be dispatched with each other to avoid mutual interference.
21
MedEx DIGITAL EEG USER MANUAL
After the PC enters the OS, double-click the icon on the desktop or run
pops up. It includes software name, company information and so on, shown as the following.
In a moment, the startup window disappears and the MND-3000 main interface pops up.
Note: Before running the MND-3000 system, please ensure that the security key and the EEG
Acquisition Box has been connected to the PC. Also the EEG Acquisition Boxs power has been
switched on.
22
MedEx DIGITAL EEG USER MANUAL
The main interfaces display may vary with what user selectable toolbars are selected, and where
these toolbars are placed on the screen and what the user ordered. When the user orders the
The boundary splitters can also be dragged, which separate the Waveform window, Video
window and Analysis window to resize these windows, even to hide any of them.
There are 3 main windows on the main interface. They are the Waveform window, Video window
and Analysis window. Each windows display size can been adjusted by dragging the boundary
splitters which separate each window with the others. These windows can even be hidden.
On the waveform window, it displays the waveform capture and review. On the analysis window, it
23
MedEx DIGITAL EEG USER MANUAL
The title bar is located on the uppermost of the main interface. It displays the patients basic
information and the systems model number. When the current case is the new generated case, it
displays the patient ID number and MND3000; when the current case is the previously saved
case, it displays the patient ID number, the patient name, sex, age and MND3000. In this way, it is
The menu bar is located on the second row of the main interface. All functions in the system can
be realized by clicking the corresponding items. Here each item in the menu bar is introduced
briefly.
FILE MENU
In the file menu, you can realize to create a new case, open a case, save, print waveform, print
topography, print setup, exit functions and so on. Here, some functions are introduced.
: Save the current case to a directory, in the pop-up routine window, input the
storage directory or review to get the routine. Then click OK button. If the input routine does not
exist, the system can create a new directory according to the input directory.
24
MedEx DIGITAL EEG USER MANUAL
is a patient list. It can help to look for the cases with the searching terms within the
: When you order ambulatory EEG, this function can be used to open
EDIT MENU
In the Edit Menu, you can realize to review cases in events, add comments, add events, create
new montages, edit the current using montages, select\save a section of waveform, input\edit
case information.
: review all events, add or edit the events, review the waveform in Comments, Events,
: After selecting a section of waveform, this function can be used to reverse the
selection.
25
MedEx DIGITAL EEG USER MANUAL
VIEW MENU
In the View Menu, you can realize to show frequently-used functions and operations shortcut
buttons, such as standard toolbar, review toolbar, sample toolbar, filter toolbar, bookmark toolbar,
status bar and pre-defined event dialog. You can also choose different background grid color,
window layouts; edit color for texts, events. Set whether operation labels, video window and
analysis window are visible. If pitching on these windows, they are visible, otherwise, they are
invisible.
: This is used for the SpO2 function, which cannot be used in the system.
: This is used to set different formats for the EEG tracing grid. It can be set to
26
MedEx DIGITAL EEG USER MANUAL
: This is used to select between horizontal window layout, vertical window layout.
RECORD MENU
In the Record Menu, you can realize to waveform monitor, record, pause and complete functions.
REVIEW MENU
In the Review Menu, you can realize to play EEG waveform in different modes and also jump to
the certain time waveform by second/minute or drag to review certain time waveform. You can
27
MedEx DIGITAL EEG USER MANUAL
also set Automatic Montage Shifting to realize reviewing EEG waveform in the montages as
recording.
There are shortcut keys for different playing modes. For example, Number 6 is one second
forward. Number 4 is one second backward. Number 3 is one page forward. Number 9 is
one page backward. Number 7 is to jump to the Home page. Number 1 is to jump to the end
page.
automatically. Spacebar is the shortcut key for pause playing. : This is used to
In the Data Process Menu, you can realize to do power spectrum analyses, Topography Mapping
analyses to the selected data. In the parameter settings, it can realize the frequency subsection
setting, whether the Topography Mapping, power spectrum shows the real time analyses results,
BOOKMARK MENU
28
MedEx DIGITAL EEG USER MANUAL
In the Bookmark Menu, you can add new bookmarks during recording or reviewing. Here, you can
also review the data by bookmarks or delete all bookmarks during reviewing.
TOOLS MENU
In the Tools Menu, you can do system settings and photic stimulation settings, make reports, use
: convert the current data to the EDF format (European Data Format).
: This is used for the network EEG, which cannot be used in the system.
HELP MENU
In the Help Menu, you can check the software information, such as the version and release date
and so on.
The tool bar is located under the menu bar. It displays the shortcut buttons which are selected
from the Menu bar. The user can set the toolbar according to his requirement. Common used
functions and operations can be set in the toolbar. The user can drag the toolbar conveniently.
STANDARD TOOLBAR
29
MedEx DIGITAL EEG USER MANUAL
: Check it to display exact 10-second waveform in a page, ignore the paper speed which
users select. And the printer setting changes to 10 Divisions mode and print exact 10-second
: Change the mouse to standard point style. Mouse in this style shows mouse pointer shape
as an arrow. Double-click left mouse button or click right mouse button, Waveform Measuring
: Increase or decrease sensitivity of the current used montage. Press the upward arrow to
enhance the sensitivity. Press the downward arrow to lower the sensitivity. The results can be
seen on the left by the operation names. Move the mouse cursor to any operation name, its
current sensitivity, high cut and low cut can be shown as the figure.
: Combo box for selecting paper speed. Select the value from the pull down list. Paper
30
MedEx DIGITAL EEG USER MANUAL
: Combo box for selecting current montage. Select the montage from the pull down
montage list. All montages of the current file are listed in the list box.
: Create a new montage and select it as the current one. The New Montage dialog pops up.
FILTER TOOLBAR
: Combo box for selecting low cut frequency of the filter. Select the value from the pull
: Combo box for selecting high cut frequency of the filter. Select the value from the pull
: Industrial frequency (50Hz) cut button. Check it to turn on the industrial frequency cut
RECORD TOOLBAR
RecordMonitor.
: Begin recording or switch from monitoring to recording as the menu item RecordBegin.
: Start or stop photic stimulation. Press the unchecked button to perform stimulation and
press it again to stop stimulation. If the checked photic stimulation is active, montage can be
31
MedEx DIGITAL EEG USER MANUAL
REVIEW TOOLBAR
: Stop reviewing.
: Combo box for selecting jump step (waveform length for one jump while auto playing).
Select jump step from the pull down list. Page means one page for each jump when automatically
playing. 1.0sec is one second for each jump when automatically playing. The value of jump step is
: Combo box for selecting frequency (jump frequency) of the timer for auto playing.
Select frequency from the pull down list. The jump frequency value is to be set in System Settings.
: Combo box for the play speed of the auto playing. The value of the play speed= jump
step * jump frequency. The value here will change if you change either value of jump step and
jump frequency.
: Combo box for selecting an operation for the sound playing. Select one operation
from the pull-down list and when automatically playing, the certain operations sound can be
played.
32
MedEx DIGITAL EEG USER MANUAL
BOOKMARK TOOLBAR
: Jump to the previous bookmark. Its grayed if theres no previous bookmark available. :
a b c d e f
(a) Information related to the prompt button in the toolbar when you put your mouse on it,
(b) Paging indicator: show current page number and total waveform page number. (One page
(c) Clock Time: When reviewing, it indicates the recording time of the waveform at the leftmost
of the waveform window. When recording, it shows the system time. It takes format as
year-month-day hour-minute-second.
(d) Eclipse Time: Indicate the Eclipsed Time of the leftmost waveform of the waveform window.
(e) Playing/Recording status indicator: Indicate recording or playing status. It includes Ready,
Monitoring, Monitoring Paused, Recording, Recording Paused, New Case (Saved), Open
an old case, Playing Forward, Playing Forward Paused, Playing Backward, and Playing
Backward Paused.
(f) A/D card indicator: Indicate connection status of A/D card, 1A/D Card or 0 A/D Card. When it
displays 0 A/D card, it means the software has not recognized the EEG Acquisition Box.
33
MedEx DIGITAL EEG USER MANUAL
Before doing EEG examination, normally the users create the new montages according to their
habits.
There are two types of montages in the MND-3000 system. They are system montages and case
montages. The system montage exists in the system and will not disappear. Users can create
system montages by themselves. But the case montage exists in each case, it is user defined.
1. The system montages can not be used directly, during recording and reviewing, the used
2. The case montage exits in the case, but the system montage exists in the system library, so
the case montage can disappear with the changes of the cases, but the system montage can only
3. The system montage can be transferred to the case montage by taking out from the montage
library. But the case montage can not be transferred to the system montage.
The case montage is located in the pull-down list. The case montage can be
edited by selecting Edit Montage from the Edit Menu. If deleting the case montage, delete it in the
1. In the EEG main interface, Select Tools in the Menu, and then click System settings, the
34
MedEx DIGITAL EEG USER MANUAL
2. Then select the tab Montage to bring forth the montage page which is shown as the
following. Two list boxes are presented in this page. One is the Montages and the other is
the Operations.
3. Press right mouse button at the blank place of Montages list box and then the following
35
MedEx DIGITAL EEG USER MANUAL
4. Choose the New item and then the following figure appears
5. Choose the EEG amplifier type which you are using from the lists and then click OK. Here you
should choose MedEx, Standard Config-24 Channels (which is for the MEA-06 amplifier) as
shown in the last figure. Then a new montage named NoName is added. To rename it,
36
MedEx DIGITAL EEG USER MANUAL
choose it and press left button on it. Then enter the new name for the montage. For example,
6. After renaming the new montage, create its new operations in the following ways. Move the
mouse cursor to a lead on the right picture and click the left mouse button on it. The lead turns
into pink, such as Fp1 in the above figure. Loose the mouse button and move the cursor to
another lead, then the mouse drags a blue line as the following figure.
7. When the cursor arrives at another lead, the lead turns pink. Click the left mouse button. The
37
MedEx DIGITAL EEG USER MANUAL
line with an arrowhead appears between 2 leads and one operation between the two leads
also appears on the left Operations list box as the following figure.
8. Add other operations in the above method. Then the following figure can be appeared.
9. If you want to modify any one operation, choose the operation from the operations list box and
click right mouse button then choose the Delete from the pop-up menu. Then redraw the
38
MedEx DIGITAL EEG USER MANUAL
10. After adding all operations for the new montage, click Ok button to finish creating a new
montage.
If the new montage is not often used, it is suggested to create as a case montage. The new
created case montage is saved in the current case. If the montages which are used have no
changes, the new created case montages can exist and be used. If the used montages have
1. In the EEG main interface, click button or select New Montage from Edit Menu. The
2. Input the montage name on the left top corner, then link two leads together as described
above. On the left column, the related operation appears. If the operation is linked wrong,
select the corresponded operation and right click the mouse button and then delete it from the
list.
39
MedEx DIGITAL EEG USER MANUAL
3. On the left operation list, double click on the operation, or right click and then select the
property item, the following figure pops up. Here you can set the operation definition and
parameters.
4. After adding all operations for the new montage, click Ok button to finish creating a new case
montage.
Besides linking each operation one by one, you can also choose All Channels after creating the
40
MedEx DIGITAL EEG USER MANUAL
new montage name, and then edit each operation by editing its property.
When using the system montage, it should be transferred to the case montage by taking it out
1. Click the button on the main interface, the following figure pops up.
2. Click From Library button and then the following figure pops up
41
MedEx DIGITAL EEG USER MANUAL
3. Choose the new montage; here is montage24Bipolar and then click Select button.
4. The system returns to the preview interface and then click OK button. Then the new montage
can be added to montage list. You can find it from down list.
After using the new montage for the first time, users can choose the montage from the
down list.
42
MedEx DIGITAL EEG USER MANUAL
When running the MND-3000 EEG system to enter the main interface, the system automatically
creates a new case. Its ID number is generated automatically as the current system date plus a
3-digits and order generated natural number. Or click the button on the main interface to
Note: The MND-3000 system is designed to record the EEG tracing first and then input the
patients information. If the user wants to input the patients information before recording the EEG
tracing, click the Edit/Case item from the menu bar, and then the Case Property window pops up
43
MedEx DIGITAL EEG USER MANUAL
Select Information label and then fill in patient information. This page involves patient information
and case history information. The patient name could not be empty and the defaulted is
NoName. Main ID, Case ID and Patient ID are inputted by user or can be left blank. Examine
Date is the recording date and not editable. And Department and Physician either are inputted
by typing or are selected from the lists (history list) of the combo boxes. The new value is
automatically added to the history list and can be selected later after inputting. After inputting the
After creating a new case and applying a full set of disk scalp electrodes using the International
10/20 System of Electrode Placement, the user can start monitoring and recording the EEG
tracings.
The Default Montage appears on the screen. The default filters, sensitivity and paper speed are
shown on the screen as well. Users can select the other values from the corresponding pull-down
lists.
Click to monitor the EEG waveforms. You can adjust operations display room, edit the
montages, and change each operations properties before recording the data.
When one operation or some operations waveform is too high or too small, you can adjust their
display room on the screen. And the printing room for the leads is automatically adjusted too as
the displaying room. In the main interface, left click on the operation name you want to adjust,
drag it up or down to the desired place. Then the room for displaying the operations waveform is
increased or decreased. Also that operations room can be increased or decreased on the printing
reports as well.
When recording, right click on the operation name as the following figure to select one operation
whose name turns into red and its function menu pops up.
44
MedEx DIGITAL EEG USER MANUAL
Users can edit the current operation and edit the current montage as well.
Users can choose property to enter its property window. Or double click on the operation name
and its property window pops up. Users can edit the operations property here.
The channel definition is two physical leads relation (addition or subtraction).The operation name
is default defined by the selected two leads and their relation. Users can define it by themselves.
The operations sensitivity, high cut, low cut and voltage range values can be selected from their
combo boxes down lists.The values in the combo boxes are pre-set in the system settings.
If pressing on Use Filter, the high cut and low cut values cannot be selected here, they follow the
Users can set each operations waveform color by choosing different color.
Note: If checking any items left box, its value is used for all operations in the montage.
45
MedEx DIGITAL EEG USER MANUAL
Right click on the operation name and pop up its function menu, users can add a new operation or
: This is to add a new operation after the current operation. If without selecting the
: This is to add a new operation before the current operation. If without selecting the
: This is to move the current operation before the previous operation. If the current
: This is to move the current operation after the next operation. If the current operation is
EDIT MONTAGE
Right click on the operation name and pop up its function menu, select item from
the function menu, the Edit Montage dialog pops up. Users can edit the properties of the current
46
MedEx DIGITAL EEG USER MANUAL
During recording and reviewing, you can change filters or sensitivity or paper speed.
The height of the label form is related to the sensitivity and the voltage range of the
operation (height = range / sensitivity). For example, if you select 50uV as the voltage range and
10uV/mm as the sensitivity for an operation, the height of the corresponding label form will be
5mm (50/10). If the label form is high enough, the voltage range of the operation is also displayed,
shown as .
If you change the paper speed by selecting the different value in the combo box of paper speed,
Change sensitivity: Click on the down arrow in the Parameter Toolbar. Select the new value
from the down list. After adjusting, move the mouse cursor to any one operation name and then its
current used sensitivity values are shown on the window as the following figure.
47
MedEx DIGITAL EEG USER MANUAL
Change Paper Speed: Click on the combo box down list in the Parameter Toolbar.
Select the new value from the down list. After adjusting, the background grid is changed as well.
Change filters: Click on the filter down arrow. Select the new value from the down list. Users can
Click the check box of power filter to filter the power disturbance. There are 50Hz and 60Hz
according to different countries power. It is set by both the hardware and the software in the
factory.
Double click the mouse on an operations name. The Channel Properties window will be displayed.
Or, right click on the operation name and select the Properties from the context menu. Then
48
MedEx DIGITAL EEG USER MANUAL
When the waveform is steady and the right montage and each operations filter, sensitivity and
paper speed are right set, click to begin recoding EEG with the default montage.
The users can also change the settings during or after recording. For example, users can edit the
montage; add events, change filters/ sensitivity/ paper speed and so on.
During recording, users can change a montage, add an event or a comment and use Clock
user montages push-down list and then click on it. The desired montage is the
If the desired montage are set in the system montage and has not been selected from the library,
1. Click the button on the main interface, the following figure pops up.
2. Click From Library button and then the following figure pops up
49
MedEx DIGITAL EEG USER MANUAL
3. Choose the new montage; here is montage24Bipolar and then click Select button.
4. The system returns to the preview interface and then click OK button. Then the new montage
can be added to montage list. You can find it from down list.
After using the new montage for the first time, users can choose the montage from the
down list.
During recording and reviewing, you can add a comment and event.
Select Pre-defined Event Dialog from the View pull-down menu. Then the pre-defined dialog
50
MedEx DIGITAL EEG USER MANUAL
Click on the appropriate event button then click left mouse button on the EEG where you want to
place the event. Then the event log is marked on the waveforms. Or you can use shortcuts of
keyboard for quicker operations: The number keys on the keyboard from 0 to 9 serve as the
shortcut for the 10 anterior preset events on the Pre-defined Event Dialog. Press key 1 to add
the first preset comment, press key 2 to add the secondary one, and so on. The key 0 is for the
If you do not locate where the event is, it is locked in place when it scrolls off the screen.
Click on the button in the Standard Toolbar or select Add Comment from the Edit
pull-down menu then click left mouse button on the EEG where you want to place the comment.
Type your comment in Event Property dialog and press OK. Color and Priority can be set here.
51
MedEx DIGITAL EEG USER MANUAL
If you do not locate where the event is, it is locked in place when it scrolls off the screen.
When recording, press the button on the Record Toolbar or select Timer from the Tools
Enter the second value (initialized with history value) in the middle edit box.
Check the button to count from zero to the value of the middle edit box. And check the button
Note: When recording, in the beginning and end of counting, events are marked automatically.
And for each one minute, one event is produced also. But if Timer dialog is closed in the middle
of counting, the clock stops and the ending event is not marked.
When finishing recording, click to stop, and Case Property window pops up. Select
Information button to fill in patient information as the followings. Click OK, the case is saved
automatically.
52
MedEx DIGITAL EEG USER MANUAL
This page involves patient information and case history information. The patient name could not
be empty and the defaulted is NoName. Main ID, Case ID and Patient ID are inputted by user or
can be left blank. Examine Date is the recording date and not editable. And Department and
Physician either are inputted by typing or are selected from the lists (history list) of the combo
boxes. The new value is automatically added to the history list and can be selected later after
inputting.
Press the button on Standard Toolbar or select Open from the File pull-down menu. Pick
Or quick search the case according to the exam date and the patients sex within the appointed
Run in the File Menu and the cases in the default saving directory are listed as the
following figure.
53
MedEx DIGITAL EEG USER MANUAL
In the left columns, there are three directories, all patients, male patients and female patients.
Select anyone directory, the cases which meet with the items are shown in the middle lists within
the working routine. If select all patients, all patients cases are listed; if select male patients, all
male patients cases are listed; if select female patients, all female patients cases are listed.
Or search the case by examination date period. Input the start and the end examination date, and
then click Refresh button, the cases which meet with the items are listed in the middle.
Click Recent button on the right column, the recently opened cases are listed.
Select one case in the middle list, and then click Open button, the case is opened in the main
interface.
Click the Edit/Case item from the menu bar, and then the Case Property window pops up as the
following figure.
54
MedEx DIGITAL EEG USER MANUAL
Re-input the information in the desired modifying items and then click OK button to save the
The MND-3000 system has many review modes. During, users can also change montages,
change filters, sensitivity, paper speed and other parameters, add comments or edit comments,
During reviewing EEG waveform, users can set the current display as 10 Divisions. Check
button on the toolbar to display exact 10-second waveform in a page, ignore the paper speed
which users select. And the printer setting changes to 10 Divisions mode and print exact
55
MedEx DIGITAL EEG USER MANUAL
Click the button on the toolbar, the system plays the EEG waveforms forward automatically.
Click the button on the toolbar, the system plays the EEG waveforms backward
automatically. Click button, the system pauses/continues auto playing. Click button to
stop auto playing.
Select the different step values from the down list. (Step value is the waveform length
of each step when auto displaying.) Among the value lists, Page is one page for each step;
1.0sec is one second for each step. The step values can be set in the system settings.
Select the different frequency values from the down list. (Frequency value is the
waveform frequency of each step when auto displaying.) The frequency values can be set in the
system settings.
Select the different play speed values from the down list. When auto playing, play
speed=step value X frequency value. The speed values vary with the step values and frequency
values.
Select one operation from down list. When auto-playing, the system plays the
Click button on the toolbar. Move the mouse cursor to the different positions on the screen,
the system reviews the EEG waveform in Page or second. As the following figure, when the
mouse cursor is moved to the left or right, the cursor changes into dual-arrowhead, now the
system plays the waveform in page. Click the left mouse button on the right side of the screen, 1
page advances forward; click the left mouse button on the left side of the screen, 1 page moves
56
MedEx DIGITAL EEG USER MANUAL
When the mouse cursor is moved to the middle leaning left or leaning right, the cursor changes
into one-arrowhead and now the system plays the waveform in second. To move in one-second
increments, use the button on the Review Toolbar. Click to review the waveform of last
REVIEW IN EVENTS
Using the EEG scroll bar is located at the bottom of the screen to move to the next event. Press
left mouse button on the color vertical line. The waveform goes to that event.
Event markers are displayed as vertical lines color-coded by priority, shown as following figure.
Use the Events button to display a list of all events/comments entered in the record. You can
57
MedEx DIGITAL EEG USER MANUAL
REVIEW IN BOOKMARKS
Seek position by virtue of bookmarks. Press button and button on the Bookmark
Toolbar to jump.
Drag the thumb of the scrollbar on the bottom of the waveform window. The waveform window
goes forward or backward following the mouse pointer. When the mouse button is down, the real
time and the eclipsed time are visible at the left-bottom corner of the waveform window.
58
MedEx DIGITAL EEG USER MANUAL
the following figure shows, then the EEG waveforms will be displayed with the montages same as
Here users cannot do changes on montages. Montage lists are in grey. The used montage name
during recording is shown on the grey montage name in the toolbar as shown in the following
figure.
59
MedEx DIGITAL EEG USER MANUAL
Right click on the operation name as the following figure to select one operation whose name
Users can edit the current operation and edit the current montage as well.
Users can choose property to enter its property window. Or double click on the operation name
and its property window pops up. Users can edit the operations property here.
60
MedEx DIGITAL EEG USER MANUAL
The channel definition is two physical leads relation (addition or subtraction).The operation name
is default defined by the selected two leads and their relation. Users can define it by themselves.
The operations sensitivity, high cut, low cut and voltage range values can be selected from their
combo boxes down lists.The values in the combo boxes are pre-set in the system settings.
If pressing on Use Filter, the high cut and low cut values cannot be selected here, they follow the
Users can set each operations waveform color by choosing different color.
Note: If checking any items left box, its value is used for all operations in the montage.
Right click on the operation name and pop up its function menu, users can add a new operation or
: This is to add a new operation after the current operation. If without selecting the
: This is to add a new operation before the current operation. If without selecting the
: This is to move the current operation before the previous operation. If the current
61
MedEx DIGITAL EEG USER MANUAL
: This is to move the current operation after the next operation. If the current operation is
EDIT MONTAGE
Right click on the operation name and pop up its function menu, select item from
the function menu, the Edit Montage dialog pops up. Users can edit the properties of the current
If the desired montage has been selected from library, select the montage from the
user montages push-down list and then click on it. The desired montage is the
If the desired montage are set in the system montage and has not been selected from the library,
1. Click the button on the main interface, the following figure pops up.
62
MedEx DIGITAL EEG USER MANUAL
2. Click From Library button and then the following figure pops up
3. Choose the new montage; here is montage24Bipolar and then click Select button.
4. The system returns to the preview interface and then click OK button. Then the new montage
can be added to montage list. You can find it from down list.
After using the new montage for the first time, users can choose the montage from the
down list.
63
MedEx DIGITAL EEG USER MANUAL
If the comment or event you want to add is in the Pre-defined Event Dialog, select Pre-defined
Event Dialog from the View pull-down menu. Then the pre-defined dialog pops up as the
following figure.
Click on the appropriate event button then click left mouse button on the EEG where you want to
place the event. Then the event log is marked on the waveforms. Or you can use shortcuts of
keyboard for quicker operations: The number keys on the keyboard from 0 to 9 serve as the
shortcut for the 10 anterior preset events on the Pre-defined Event Dialog. Press key 1 to add
the first preset comment, press key 2 to add the secondary one, and so on. The key 0 is for the
If you do not locate where the event is, it is locked in place when it scrolls off the screen.
64
MedEx DIGITAL EEG USER MANUAL
If the comment or event you want to add is not in the Pre-defined Event Dialog, click on the
button in the Standard Toolbar or select Add Comment from the Edit pull-down menu then click
left mouse button on the EEG where you want to place the comment. Type your comment in
Event Property dialog and press OK. Color and Priority can be set here.
If you do not locate where the event is, it is locked in place when it scrolls off the screen.
EDIT AN EVENT
Press button on the Standard Toolbar or select Event Dialog from the Edit pull-down
menu. The Event dialog pops up. Select the event you want to edit from the left list box, and all
events in that category are listed in the right list box. Select the event that you want to edit or
delete from the right list box and press the right mouse button. Following context menu appears.
65
MedEx DIGITAL EEG USER MANUAL
Select Event Property from the menu. The Event Property dialog pops up.
Enter the new text of the event and/or change the priority. Press Ok to confirm the changes or
Press button on the Standard Toolbar or select Event Dialog from the Edit pull-down menu.
The Event dialog pops up. Select Comments from the left list box and all comments are listed in
the right list box. Select the comment that you intend to delete from the right list box and press the
66
MedEx DIGITAL EEG USER MANUAL
The height of the label form is related to the sensitivity and the voltage range of the
operation (height = range / sensitivity). For example, if you select 50uV as the voltage range and
10uV/mm as the sensitivity for an operation, the height of the corresponding label form will be
5mm (50/10). If the label form is high enough, the voltage range of the operation is also displayed,
shown as .
If you change the paper speed by selecting the different value in the combo box of paper speed,
Change sensitivity: Click on the down arrow in the Parameter Toolbar. Select the new value
from the down list. After adjusting, move the mouse cursor to any one operation name and then its
current used sensitivity values are shown on the window as the following figure.
67
MedEx DIGITAL EEG USER MANUAL
Change Paper Speed: Click on the combo box down list in the Parameter Toolbar.
Select the new value from the down list. After adjusting, the background grid is changed as well.
Change filters: Click on the filter down arrow. Select the new value from the down list. Users can
Click the check box of power filter to filter the power disturbance. There are 50Hz and 60Hz
according to different countries power. It is set by both the hardware and the software in the
factory.
Double click the mouse on an operations name. The Channel Properties window will be displayed.
Or, right click on the operation name and select the Properties from the context menu. Then
68
MedEx DIGITAL EEG USER MANUAL
When one operation or some operations waveform is too high or too small, you can adjust their
display room on the screen. And the printing room for the leads is automatically adjusted too as
the displaying room. In the main interface, left click on the operation name you want to adjust,
drag it up or down to the desired place. Then the room for displaying the operations waveform is
increased or decreased. Also that operations room can be increased or decreased on the printing
reports as well.
Click the button on the Standard Toolbar. Double-click the left mouse button on the
Move three vertical lines to the target position by pressing left mouse button on any vertical line
and when it is moved to suitable place, press the left button again. The line is changed to the new
position. Two blue vertical lines are the start and the end; the red vertical line is the middle.
Measuring results are showed under the waveform window. Time(T) refers to the time interval of
the two vertical blue-colored lines. Frequency (1/T) is the reciprocal of Time(T). Voltage(G) is the
69
MedEx DIGITAL EEG USER MANUAL
voltage difference between two ends of the green line. Voltage(R) is the voltage difference
between two ends of the red line. Whether to check Show Background is to select showing grid
or not on waveforms.
Click to select a segment of waveform. The mouse pointer shapes like . And click left
mouse button on the left of segment to set the left edge; then the mouse pointer shapes like ,
click left mouse button on the right of segment to set the right edge. The selected waveform is
70
MedEx DIGITAL EEG USER MANUAL
To select multiple segments of waveforms, keep the key Shift down and also perform as the
After selecting the waveforms, press right mouse button on the selected waveform segment, a
Select Delete item and then the selected waveforms are deleted. (The selected waveforms are
changed to zero). Use this function to delete the waveforms with high disturbance.
After selecting the waveforms, press right mouse button on the selected waveform segment and
Click Power chart and the power chart window pops up. (Note: The selected waveforms length
should be among 3 seconds to 60 seconds.) In the power chart window, users can view any one
operations Power chart. Select one montage and then select one operation on the left columns.
71
MedEx DIGITAL EEG USER MANUAL
Click topography mapping and the updated topography window with the calculation result is
shown on the right data process window. (Note: The selected waveforms length should be among
2 seconds to 68 seconds.)
Press right mouse button on the selected waveforms and click print or print preview in the context
After selecting a segment of waveform as the above description, and then click Selection Save
Then the Workpath status dialog is shown and user can select the saving path.
It is saved in another directory that is located in the same directory of its record. Add a suffix to
that of the record and the new directorys name will be generated. For example, if the directory of
72
MedEx DIGITAL EEG USER MANUAL
the current file is 20060518002, the new directory is named 20060518002-2. If the directory of
In the Data Process Menu, it can realize to do power spectrum analyses, Topography Mapping
analyses and CSA analyses. The processing window includes Mapping, Power Spectrum and
A B C
Click view on the main menu of the main interface and then select from
the pop-up list. Then the processing window shows on the right of the screen.
You can select different processing windows to see between Mapping, Trendgraph and CSA.
These three above graphs are synchronized with the waveform on same Eclipsed Time.
Topography is the calculating result of waveforms that start at the current Eclipsed Time or the left
A small blue slider block located below the topographys time axis indicators the current Eclipsed
Time.
For CSA, the time of the most front frame has the current Eclipsed Time.
The most important parameters for data processing are the frequency bands. 6 frequency bands
are provided. The start and the end of these bands can be adjusted in Parameter Settings from
73
MedEx DIGITAL EEG USER MANUAL
the pull-down menu Data Process. The name of a frequency band is different with different
frequency setting.
Six colorful head portraits are displayed, with a color ruler located at the right of them and the
The numbers at the right of the color ruler are the power values that are shown by the
corresponding color (parallel with the underline or the head-line of the number). The numbers and
There are two ways to show topography: Absolute and Relative. Press the mid combo box,
In Absolute model, six color head mappings use the same rulers. The minimal value is the lowest
of all six frequency bands and the maximal value is the greatest of them.
In Relative model, six color head mappings use the different rulers. The minimal value and the
maximal value are all based on each frequency band (each head portrait).
Change view aspect: press the left combo box and select top, left or right from the pull down list.
Change size of the head portraits: press the right combo box and select large, medium or small
Change Eclipsed Time: Scroll the waveform window; eclipsed time shown on the bottom
changes and topography is refreshing (the topography is the calculating result of waveforms that
begin at the left edge of the window and last for 6 seconds).
74
MedEx DIGITAL EEG USER MANUAL
SUBSECTION 2 TRENDGRAPH
Two trend graph windows are listed in gray-colored, respectively for the operation Fp1-A1 and
The power value of each frequency band is displayed in form of percentages. The vertical axis
stands the percentage. The horizontal axis on the bottom of the color graph stands the Eclipsed
Time of selected waveform. The unit of it is minute. Double click a certain time on the horizontal
At the left of trend graph, frequency bands are displayed according to the checkboxes. Uncheck
any of them to diminish the corresponding color band. For example, uncheck the band for
Press right mouse button in the area of the container window that is not covered by any trend
75
MedEx DIGITAL EEG USER MANUAL
Cascade Windows: Cascade all trend graph windows in the container window.
Tile Windows: Tile all trend graph windows in the container window.
Recalculate All: Recalculate all trend graph and update them with the new result.
Press right mouse button in the area of any trend graph window when the window is not
A B
Cascade Windows: Cascade all trend graph windows in the container window.
Tile Windows: Tile all trend graph windows in the container window.
Recalculate: Recalculate the trend graph that the mouse button is pressed on, and update it
Recalculate All: Recalculate for all trend graph and update them with the new result.
Press left mouse button in area of trend graph, and press Ctrl button in the keyboard at the
same time, the horizontal and vertical axes of the graph, shown as the following.
The left and right numbers are zoom-in or out of horizontal and vertical axes
76
MedEx DIGITAL EEG USER MANUAL
Press Ctrl and the left mouse button, and move the mouse up/down and the vertical axis zoom
in/out; move the mouse right/left and the horizontal axis zoom in/out. Then release the key and
mouse button, the trend graph zoom in/ out in different rates of horizontal and vertical axes.
POWER SPECTRUM
Double-click left mouse button on the graph area and bring forth the Power Spectrum dialog
shown as the following. It shows the power histogram for the point where the mouse is pointing at.
Double-click left mouse button below the time axis, the time indicating slider block will jump to the
point where mouse button is pressed, and the current Eclipsed Time changes. As a result, the
waveform window jumps to the position with the same Eclipsed Time.
If the slider block is currently visible, it could be moved by another way. Single-click left mouse
button on the slider block to pick it up. The slider block follows the mouse pointer. Press left button
Click the button at up-right corner of trend graph window to maximize it. Click the button
SUBSECTION 3 CSA
77
MedEx DIGITAL EEG USER MANUAL
In the above figure, two CSA windows are listed in a gray-colored container window. They are
related for the operation Fp1-A1 and Fp2-A2. The operation list can be set in Parameter
The vertical glide adjusts CSAs Y-scale. (Four scale grades 4: 1, 2: 1, 1: 1, 0.5: 1, 0.25: 1 from
the up to down).
The horizontal glide is the time axis. Move (by dragging the slider block or clicking on the glide or
the sideward arrows) the slider block to change the current Eclipsed Time and then go through the
whole file. Meanwhile other CSA graphs, the waveform window is refreshed synchronically.
Double-click left mouse button on CSA and show the Power Spectrum dialog as follows. It
shows the histogram and Power-Frequency chart for the point where the mouse button is
pressed.
78
MedEx DIGITAL EEG USER MANUAL
Press right mouse button on the area of the container window that is not covered by any CSA
Recalculate All: Recalculate all CSA and update them with the new result.
If press right mouse button in the area of any CSA window when the window is not maximized,
A B
Check the checkbox on the side of the color ruler to draw CSA colorfully, different
power value with different color.
79
MedEx DIGITAL EEG USER MANUAL
Click the button at up-right corner of CSA window to maximize it. Click the button to
During reviewing the waveform, choose better waveforms shown on the screen. Click the
button on the toolbar or select Reporting from the Tools pull-down menu, the Reporting dialog
Users can input report diagnosis and conclusions here and also edit the patients information.
Change the report titles and case information if necessary. These changes are only for temporary
purpose, they do not affect the system settings and the case information of the current patient.
The checkbox Waveform and Topography are initially checked for printing. Uncheck the
corresponding checkbox if not print. Select orientation (Portrait or Landscape) for papers. Press
the button Preview to preview the report. Press the button Print to print the report. The output
80
MedEx DIGITAL EEG USER MANUAL
The patients information, EEG waveforms, Topography mapping, patients video (if users choose
video option), what to see and diagnosis are all listed in the report. The printed waveforms are the
waveforms which start from the left edge of the shown waveform window and last for 6 seconds.
There are 3 title lines which can be used. For example, the second line can be used to set case ID
or leave as empty. In the report, only the first line with the Hospital name is shown.
Select the montage that you want to print of current one, and adjust some parameters of the
montage if necessary.
If you prefer to print the current shown screen waveforms, click the button on the Standard
Toolbar.
If you prefer to print exactly 10-second waveform for one page, check the button on
Standard Toolbar.
Ensure that the waveform window has the focus. If you are not sure, click mouse button (left or
right) on the waveform window. Select Print Setup from the File pull-down menu, set parameters
of the printer, especially paper size and papers orientation. Press the button on the
81
MedEx DIGITAL EEG USER MANUAL
Standard Toolbar or select Print Waveform from File the pull-down menu, following dialog is
presented.
Enter start time and end time of waveform to print. Check the From the left of the screen to set
the left edge of the waveform window as the start point. Checkbox From the beginning of file
means printing from the very beginning of the file. To the right of the screen means the
waveform block to be printed ends at the right edge of the waveform window. The checkbox To
Select the from the File menu, the cases in the default saving directory are listed
82
MedEx DIGITAL EEG USER MANUAL
In the left columns, there are three directories, all patients, male patients and female patients.
Select anyone directory, the cases which meet with the items are shown in the middle lists within
the working routine. If select all patients, all patients cases are listed; if select male patients, all
male patients cases are listed; if select female patients, all female patients cases are listed.
Or search the case by examination date period. Input the start and the end examination date, and
then click Refresh button, the cases which meet with the items are listed in the middle.
Click Recent button on the right column, the recently opened cases are listed.
Select one case in the middle list, and then click Delete button, the reminding dialog pops up as
the following figure, click Yes button to delete the selected case.
83
MedEx DIGITAL EEG USER MANUAL
dialog pops up. Save file as EDF format. (*.EDF) Select montage that you want to export from the
list box at bottom of the dialog. Press the button Save to begin converting.
EEG impedance check is used to check whether the electrode is well connected with skin when
we examine the EEG. With this function, we can check whether electrodes are off work and make
sure the EEG waveform is more accurate. There are two modes of display impedance check:
Usage: In the process of monitoring or gathering the waveform, press green button on the left
side of the Acquisition Box. Then a impedance check window pops up as below:
When this window pops up, it is in the impedance measurement mode, and top screen displays as
Initialing. It will display the result of impedance check after about 10 seconds. Every electrode
is marked as different color, and the screen will display as Check Status as picture below:
84
MedEx DIGITAL EEG USER MANUAL
: According to the signal given by Acquisition Box, calculate the status of the
: If green signal lamp and white signal lamp wink alternately, it means
85
MedEx DIGITAL EEG USER MANUAL
: In this mode, the color will change from green to yellow and then to red, that shows
the approximate range of the impedance. Minimum color is green and maximum color is red. The
impedance value changes from minimum to maximum when the color changes from green to red.
: Shows upper limit of EEG impedance range, this value and the value higher
than this will all be marked as red.
: Shows lower limit of EEG impedance range, this value and the value lower
than this will all be marked as green.
: The upper limit and the lower limit will be changed when this button pressed down.
: The impedance color comparison bar change from green to red gradually,
shows that the impedance value changes from minimum to maximum.
impedance value is higher than this parameter, and when the value is lower than it, the color will
be changed to green.
86
MedEx DIGITAL EEG USER MANUAL
: The impedance color comparison bar shows that if the impedance value is higher
than 10K the color will be red and lower than 10K color will be green.
87
MedEx DIGITAL EEG USER MANUAL
Select the Case from the Edit menu, the case window pops up. In the case window, there are 4
pages.
In the Information page, it involves patient information. Examine Date is the recording date and
not editable.
88
MedEx DIGITAL EEG USER MANUAL
In the storage label, users can set working path. Input the path or click the button to find a
path. If it is a previously saved file, users cannot change the working path. If the file has video,
open the saving file, the video channel and the file name can be shown on the left list. Double
click on the file name to rename it. Patient File is to save the patients basic information;
Waveform File is to save the waveforms data; Event File is to save the added events; Montage
File is to save the montages data. Users can check the data storage files and can rename them.
But for the previously saved files, users cannot rename the File name.
89
MedEx DIGITAL EEG USER MANUAL
Set the A/D Record Rate: pull down the list of the combo box, and select one between 100Hz,
Change target gain for a channel: select one channel on the left table and then press left button
on it again. The target gains combo box is shown and then users select one from the pull-down
Users can save the current scheme of each channels gain as the default scheme by pressing the
Note: normally users do not need to change them and they are pre-set by the manufacturer.
Video settings:
AVI File Size Limit is used to set the maximum size of one video file. When the video file reaches
Video Record Rate is used to set the recording frequency of the video. Choose one value from
the pull-down list. For the previously saved files, users cannot set the value.
In the Recording Time Control, users can select manually stop recording or automatically
90
MedEx DIGITAL EEG USER MANUAL
In the montage label, users can check, edit, add and delete the case montages and their
operations. The detailed operation about it please refer to section create a case montage.
This pre-defined event dialog is the container for all pre-defined events. For each pre-defined
event, users can modify them by themselves. Users can add or delete event buttons and also edit
Select Pre-defined Event Dialog from the View pull-down menu. The Pre-defined Event Dialog
pops up.
91
MedEx DIGITAL EEG USER MANUAL
Press right mouse button on the event button of the Pre-defined Event Dialog you want to
operate with, select menu item Property to show and/or edit property of the corresponding
event. Select Delete to delete this event button, and Delete All to delete all event buttons.
Press right mouse button on the empty area where there are no any event buttons, another menu
Select New to add a new event button to the tail of the button list.
Note: If the events type is Events, its priority cannot be changed. If the events type is Comments,
92
MedEx DIGITAL EEG USER MANUAL
Select System Settings from the pull-down menu Tools, and input the password medex. The
System Settings window pops up, which includes five pages of Review, Storage,
Montage, Report and Other. They are respectively designed for system parameters of
reviewing, storage scheme, system montages, report format, and other miscellaneous system
parameters. Set the system parameters by selecting different tabs. Finally press the button OK
to confirm the changes or press the button Cancel to exit without saving the changes. The
system parameters changes will not take effect until you restart the program MND3000 again.
Select the tab Review to bring forth the reviewing parameters page shown as follows.
The high cuts, low cuts, play speeds, sensitivities, paper speeds, voltage ranges is listed
respectively in the list box High Cut, Low Cut, Play Speed, Sensitivity, Paper Speed
and Voltage Range. Add new items to the list boxes or delete items from them by pressing right
93
MedEx DIGITAL EEG USER MANUAL
mouse button in the list boxes and select New from the sequent context menu. To edit an item of
a list box, just single-click left button on the item to make it the current one and then press left
mouse button on it again, or press right mouse button and select Property from the above context
menu. Then input the new value for it. To delete an item, select the Delete from the above context
menu.
Select industrial frequency of your country region from the combo box. The selected value is
Press left mouse button on the upward or downward arrow at right of the edit box
Press left mouse button on the upward or downward arrow at right of the edit box
Check the radio button to set the default Time Axis Orientation as horizontal or
The Play Speed column is for the parameter settings for the auto playing. You can set different
values for the jump frequency and jump step, and then the play speed will be auto generated.
Their relationship is: play speed= jump frequency X jump step. After you set here, you will see
the values here to be the option items in the combo box in the Review Toolbar.
Select the tab Storage to bring forth the storage parameters page shown as following.
94
MedEx DIGITAL EEG USER MANUAL
Pre-defined Event Files, System Montage Files and External Event Files are all stored in disk files.
Enter the paths and names for these files in the corresponding edit box word by word. If the
Pre-defined Event File or the System Montage File you selected dose not exist, a new file will be
created and. when you exit from MND3000, the Pre-defined Event File or System Montage File is
saved in the new file. Press the button to show and/or modify the content of the external
events file.
The work path and the temp path are respectively the path where recorded files (data) and temp
data (files) are to be saved. Enter the pathnames in the edit boxes. Remember that these paths
must already exist; otherwise you will have some troubles when begin recording.
The default file names (Waveform File, Video File, Montage File and Event File) are used as
names for the new files. But you do not need to specify them. Leave these edit boxes blank. Some
rules are used to construct the file names if you do not specify the default values. For details
95
MedEx DIGITAL EEG USER MANUAL
extension part (they are respectively . dat, .avi, .vnt and .mtg).
The ID is a text string identifying the local system (MND3000). It distinguishes the local MND3000
system from MND3000 systems working on other computer systems within one department. Set it
The yyyy is a four-digit string presenting the year part of the date. The mm is a two-digit string
presenting the month part of the date. The dd is a two-digit string presenting the day part of the
date. The iii is a three-digit string presenting the index value of the file within the day.
Select the tab Montage to bring forth the montage page which is shown as following.
Two list boxes are presented in this page. One is the Montages and the other is the
Operations.
To add a new system montage, press right mouse button in the Montages list box and select
New from the context menu. A new montage named NoName is added. (The other menu item
Delete is designed for deleting the current list box item and its corresponding montage).
96
MedEx DIGITAL EEG USER MANUAL
To rename a montage, press left mouse button on the corresponding list box item to make it to be
the current one and press left button on it again. Then enter the new name for the montage.
Select a montage in the Montages list box, and all operations of it is listed in the Operations list
box.
Press right mouse button in the Operations list box, and select New from the context menu. A
new operation with default properties is inserted into the montage and the corresponding list box
item is added.
Double-click or press right mouse button on the operations name, and the Channel Property
dialog is displayed to set the properties for the operation. Also you can select Delete to omit the
Press the button on the Parameter Toolbar. New Montage dialog pops up.
97
MedEx DIGITAL EEG USER MANUAL
Enter name for the new montage in the top-most edit box (e.g. Montage2).
Press right mouse button in the operation list box (below the edit box), select New from the
sequent context menu. A new operation with default properties is created and the corresponding
text item is added to the list box. Press left mouse button on the newly added item to make it the
current one. Then press right mouse button on it and select Property from the context menu or
simply double-click left mouse button on it to bring forth the Channel Property dialog. Set the
properties for the new operation now. To add more operations, repeat the above steps. If want to
omit an operation, press right mouse button on the corresponding list box item and select Delete
All channels (Press the button All Channels to form a Single-polar montage with 16 operations)
or system montage (Press the button From Library, and select a system montage from the
pop-up dialog. The system montages were set as system parameters previously. Then do some
adjustments with the operations properties. Press the button OK. The montage is finally added
98
MedEx DIGITAL EEG USER MANUAL
Select the tab Report to bring forth the report format page which is shown as following.
Enter texts for three title lines in the corresponding edit box. These three title lines are printed on
the head of the report papers. If you leave any edit box blank, no paper space reserves for the
corresponding title line. Select the right fonts size and font formats for each line by clicking the
button.
Check the button , or to respectively align the lines right, left or middle.
99
MedEx DIGITAL EEG USER MANUAL
The ExternalLeading is the space between the intended title line and the next line. Click on the
upward arrow or the downward arrow to increase or decrease the space value.
Press the button in the group box Text to set the font for report text.
Here you can the report format which is mainly composed of 4 parts as shown on the left column
in the last figure, EEG Waveform, Topograph, Video and Text. Tick or untick on certain item, the
Users can choose the paper layout between Portrait and Landscape and can also adjust each
100
MedEx DIGITAL EEG USER MANUAL
Select the tab Other to bring forth the page for other system parameters for Ambulatory EEG,
Enter the hospital name in the first edit box. The second edit box is designed for the Local
Identifier of this copy of MND3000 system. This Local Identifier distinguishes this system and
MND3000 systems running in other computer systems, and is used to construct default file names
101
MedEx DIGITAL EEG USER MANUAL
Select Parameter Settings from the pull-down menu of Data Process, users can do frequency
bands settings, set whether the processing window shows real time analyses results, and choose
There are 6 frequency bands provided. Click upward arrow or downward arrow to increase or
Click on the button to set all the low-points and high-points to default values. The
six default frequency bands are Delta (0.8-3.8 Hz), Theta (4.0-7.8 Hz), Alpha1 (8.0-9.8 Hz),
Alpha2 (9.8-12.8 Hz), Beta1 (13.0-20.0 Hz) and Beta2 (20.0-30.0 Hz).
Note: 4 basic system frequency bands are defined. They are Delta (0.8-3.8 Hz), Theta (4.0-7.8
Hz), Alpha (8.0-12.8 Hz) and Beta (13.0-30 Hz). Set six frequency bands for the application.
These frequency bands are named based on the above four basic system frequency bands. If the
average of the high frequency and the low frequency is within the range of any basic band, this
102
MedEx DIGITAL EEG USER MANUAL
frequency band is named as the basic one. For example, the average (2.45) of the frequency
band (0.9-4.0 Hz) is within the range from 0.8 to 3.8, so the band is named as Delta. If there are
more than one application frequencies bands are within the same basic system frequency band,
indexes are appended after the basic name. For example, you may get Deta1, Deta2 and Deta3
and so on, if there are three frequency bands within the range of the basic frequency band Deta.
If you want to perform topography calculating during recording, check the checkbox
Select the tab Trend Graph, the Trend Graph page pops up.
103
MedEx DIGITAL EEG USER MANUAL
If you want to perform trend graph calculating during recording, check the checkbox
Check the checkbox to execute EEG waveform filtering before EEG waveforms are
The Calculation Interval is the interval (in second) of Eclipsed Time between two adjacent
calculation steps. The Data Length is the time length (in second) of EEG data, which is fetched to
complete one step of calculation. The value of Data Length should be more than Calculation
Interval. So there are some data overlaps for several adjacent calculation steps, and the result
trend graph may seem smoother. Click on upward arrow or downward arrow to change the
In the following group box Channel Selection, Select operations to be calculated. There is a
Trend graph window in the Trend graph container for each operation you selected here.
Two list boxes are provided. The left one contains the operations that have not select and the right
one contains the operations that have selected. To select an operation, first select the operation in
the left list box, and then press the button to move it from the left to the right. To unselect an
operation, select the operation in the right list box and press the button to move it to the left.
Press the button to prepare all operations in the left list box for selecting. And
reversal, pressing the button will cancel the selecting preparing status of all the
104
MedEx DIGITAL EEG USER MANUAL
Check the radio box , then the operations of the Single-polar montage (simply
composed by the 32 channels) except the selected operations, are listed in the left list box. Check
the radio box , and then select operations from the montages of the current record. All
the montages are listed in the list of the combo box . Pull down the montage list
, and select the montage. All operations, except the operations have selected, are
listed in the left list box. Select the intended operations.
Operating with CSA page is the same as operating with the page of Trend Graph.
105
MedEx DIGITAL EEG USER MANUAL
If the photic stimulator is ordered, users can pre-set photic stimulation programs.
Select Photic Stimulation Settings from the Tools pull-down menu to show the following dialog.
There are five scheme programs from Config1 to Config 5. Select one from the Active
Configuration list. For example, you can select the Config 1 as the above figure. There are
twelve sections of stimulation you can perform for one program. The frequency or the duration of
Serial Port is the port used for the photic simulator. For example, COM1 is that the stimulator
106