Sie sind auf Seite 1von 8

TML Pam E-3001A

TML PORTABLE DATA LOGGER

G7F $(#

Strain Gauges Strain Gauge- DC voltage Thermocouple Pt RTD


transducers FSW-10
Channel unit
Max 50CH

002-0170.indd 1 10.2.22 4:08:45 PM


Field use, battery drive
This portable data logger consists of a control unit (TDS-150) and exclusive decade channel units (FSW-
10) and measures strain gauges, strain-gauge-based transducers, DC voltages, thermocouples and Pt
RTD. The decade channel unit (FSW-10) can be cascaded up to 5 units to total 50 channels. In addition,
by using a switching box/TML-NET driving board (option), up to 100 channels can be extended. TDS-
150 operates on not only AC mains but alkaline D-cells or battery and has data memory and sleep inter-
val timer functions for long term automatic measurement. It is possible to store measurement data and
setup condition on compact flash memory card. Interfaces are USB and RS-232C, and reading of
various setting conditions and measurement data can be conducted from a PC.
Functions Decade channel unit connector

AC adaptor connector LR20 alkaline battery storage


DC external power input
(9~18V dc)
Battery Unit

Compact flash card


USB interface and card slot

RS-232C interface
Power switch
Switching box/TML-NET
driving board
(Factory-installed option)

Ten-key
LAN board
(Factory-installed option)

LCD display with backlight ENT/START


Cursor
Function keys F1 F2 F3 ESC/Key lock

FEATURES
Connectable Five decade channel units (FSW-10) Control unit TDS-150
for 50 channels max. with Battery unit

Possible long-term automatic measurement using sleep


interval timer.
Low power consumption
Measurement of strain, DC voltage and temperature
using thermocouples and Pt RTD.
Large capacity data memory available.
1-gauge 4-wire method available.
TEDS compatible
Strain complete compensation method available
Channel unit FSW-10 (option)
Connectable network modules expandable upto 50 channels
(Factory installed option)
LR20 battery drive with
low power consumption TML-NET driving board
(Factory-installed option)
connectable for digital net-
work measurement

002-0170.indd 1 10.2.22 4:08:52 PM


System block diagram

Compact Flash
EDU-11
CONTROL UNIT card
External Display Unit G7F $(# Capacity:
2GBytes max.
RS-232C cable Data storage
Data cable CR-5360 USB cable
CR-3830 CR-6187
RPC-05A
Remote Power Controller USB port

RS-232C port

DPU-H245AS-A03A
External Printer
Printer cable
CR-5370

Switching box/TML-NET driving board LAN board (Factory


(Factory installed option) installed option)
TML-NET
Digital network

CHANNEL UNIT
9FJ $# (Option)

Switching Box FSW-21C Sensor input


Strain gauge-based
Strain gauges
transducers

DC voltage Thermocouples

Pt RTD 1-Gauge 4-Wire strain


gauges with modular plug

INTERFACE DATA MEMORY


USB port The maximum 80,000 data can be recorded. The data
Using the USB cable CR-6187 (option), control of TDS- memory is one area only and the data stored in the area
150 from a computer and data read of online measure- in order of measurement. One data are composed of
ment are possible. The USB driver is contained in TML channel, measurement time, measured data and physi-
measurement software Visual LOG Light (option). cal unit.

RS-232C port
80000
By connecting the RS-232C cable CR-5360 (option),
control of TDS-150 from a computer and data read of Data file
online measurement can be done. Also, connection with Data memory Channel
external devices using the external cable is possible. 40000 Measurement time
Measured data
Monitoring on TML External Display EDU-11 Unit
The use of EDU-11 enables monitoring at a place away
from TDS-150. 0
Measurement with Remote Power Controller RPC-05A a). The number of recordable data is 80,000
By setting up RPC-05A between TDS-150 and a computer maximum.
or modems, power on/off, control for solar power charge, b). When the number of data reaches 80000,
etc. in long-term measurement are possible. M is indicated on the Sub-LCD and no
more data recorded.
Printout of data c). Even if the channel is changed the storing
The measured data and stored data are printed on the destination of the data is not changed.
external printer DPU-H245AS-A03A (option). d). The data after storing in a PC should be
sorted out by channel.

002-0170.indd 2 10.2.22 4:08:58 PM


Pop-up operation guide
3 Interface USB/RS-232C/LAN
CF card Monitor comparator
Simple MEASURE Date and Time
Data Memory
Key lock
Scanning counts (remaining)
Battery charge

[D]: Direct Sensor mode


[M]: Measure Strain, DC voltage, Load, Displacement
Pressure, Temperature, etc.
[m]: COMET
[J]: Jump Measuring mode
Measured data/Monitored value
Channel number with engineering unit

[Proc Init] [Ch. Setting] [To Menu]


Initial value processing Monitor channels set

Initial-in functions to take initial unbalanced Selectable display for numerical value
value of the strain gauge or transducer from Y-T waveform, or OFF.
measurement value.

Applicable sensor

Strain
gauges Thermo-
couples

Strain gauge Pt RTD


transducers

DC voltage
TEDS
complatible
TEDS sensor
TML-NET Digital network module

Various checks

Resistance check of leadwire on the 3-wire Bridge output voltage check on the 3-wire
quarter bridge mode in COMET function. quarter bridge

Data record Interface

File and data managements RS-232C : Baud rate/Data bit/Parity/Stop bit, etc.
Reading/Deletion of data memory LAN : Factory installed option
File dump, Card format, etc. of CF card Data output through the RS-232C port to a printer, display unit or PC
File copy of specified file name Data format : Setting of data format with TDS/CSV and display of
Record of measured data in data memory, CF, and both header, and time

002-0170.indd 3 10.2.22 4:09:05 PM


Various applications
1-Gauge 4-Wire strain measurement method Patented

Our developed method makes strain


measurement possible by plugging
strain gauges with lead wires in
4-wire system and modular plug in
the input receptacles of the FSW-10.

TML-NET for 2-wire digital network measurement


Strain gauge-based transducers : Load cells, Displacement transducers
Civil engineering transducers, etc.
and DC voltage, Thermocouples

Network module
NSW

Switching box/TML-NET driving board


(Factory-installed option)

Free combination

Compact design enabling to store in any storage box

002-0170.indd 4 10.2.22 4:09:10 PM


TDS-150 SPECIFICATIONS

Number of channels Thermocouple temperature measurement


Maximum 100 channels In conjunction with external switching box Thermo- Measuring range Resolution Accuracy (%rdg+ C) (23 C5 C)
NB: Switching box/TML-NET driving board couple ( C) ( C) External RJC Internal RJC
(Factory-installed option) is required.
250 ~ 200 0.1 0.38 + 0.6 0.38 + 3.9
50 channels In conjunction with 5 units of FSW-10
T 200 ~ 100 0.1 0.15 + 0.2 0.15 + 1.4
Applicable sensors (Sensor mode setting) 100 ~ + 400 0.1 0.10 + 0.2 0.10 + 0.8
210 ~ 160 0.1 0.19 + 0.3 0.19 + 1.6
1-gauge 4-wire 120G
160 ~ 0 0.1 0.12 + 0.2 0.12 + 1.0
method 240G K
0 ~ + 960 0.1 0.08 + 0.1 0.08 + 0.5
350G
+ 960 ~ +1370 0.1 0.10 + 0.9 0.10 + 1.4
3-wire quarter 120G
Bridge excitation voltage 200 ~ 160 0.1 0.16 + 0.2 0.16 + 1.2
Strain bridge 240G
DC1V 48ms (50Hz) J 160 ~ 0 0.1 0.12 + 0.1 0.12 + 0.8
350G
0 ~ + 700 0.1 0.08 + 0.1 0.08 + 0.5
Half bridge 120~1000G
+ 700 ~ +1200 0.1 0.08 + 0.6 0.08 + 0.9
Full bridge 120~1000G + 200 ~ + 280 0.04 + 4.0 0.04 + 4.0
0.5~0.4
Full bridge con- 350G B + 280 ~ + 800 0.3~0.1 0.04 + 1.2 0.04 + 1.2
stant current + 800 ~ +1760 0.1 0.05 + 0.4 0.05 + 0.4
Bridge excitation voltage 10 ~ + 200 0.1 0.09 + 0.6 0.09 + 1.2
Full bridge 0-2V 120~1000G DC2V 24ms (50Hz) S
+ 200 ~ +1760 0.1 0.07 + 0.4 0.07 + 0.7
Thermocouple T 10 ~ + 150 0.1 0.09 + 0.7 0.09 + 1.2
Thermocouple K R + 150 ~ +1760 0.1 0.07 + 0.4 0.07 + 0.7
Thermo- Thermocouple J
Thermocouple B Linearization: 210 ~ + 550 0.1 0.17 + 0.2 0.17 + 1.4
couple E
Thermocouple S Digital operation + 550 ~ +1000 0.1 0.09 + 0.4 0.09 + 0.8
Thermocouple R 200 ~ 0 0.1 0.18 + 0.4 0.18 + 1.6
Thermocouple E N 0 ~ +1090 0.1 0.08 + 0.2 0.08 + 0.6
Thermocouple N +1090 ~ +1300 0.1 0.08 + 0.9 0.08 + 1.2
Input impedance The accuracy of thermocouples is not included. Thermocouple B does not use RJC.
DC voltage Voltage V 1/1 300mV V 1/1 more than 500MG RJC: Reference junction compensation
Voltage V 1/100 30V V 1/100 more than 1MG
Linearization:
Pt RTD Pt RTD 3-wire Display unit LCD with backlight
Digital operation
Display Resolution 255x160 dot
Operating NSW series Data reading from Network
TML-NET [Option] sensors Contents Measuring data, Setting list, Y-T monitor
Setting Year, Month, Day, Hour, Min. and Sec.
Clock
Measuring Range Accuracy 1 sec./day (23 C5 C)
USB/RS-232C, LAN (Option)
Item Range Measuring range Initial memory Sampling speed Interface
Function Control from PC and Data transfer
x1 30000 x106 strain 160000 Measurement INITIAL, DIRECT & MEASURE for each channel
Strain
x10 300000 x106 strain x106 strain mode (DIRECT only for temperature)
V 1/1 V 1/1 Scanning Automatically from First to Last channel
x1 30.000mV 160.000mV Channel (Jump available)
x10 300.000mV 80ms switching
DC voltage Monitor Repetition of monitor channel
V 1/100 V 1/100 (50Hz area)
Time-independent graphic monitor
x1 3.0000 V 16.0000V 67ms
x10 30.0000 V (60Hz area) Measurement Start key switch, Interval timer, Monitor comparator
start USB, RS-232C and LAN (Option)
T: - 250 ~ + 400 C
K: - 210 ~ +1370 C Capable of setting for each channel
J: - 200 ~ +1200 C Coefficient (0.0001 to 99999)
Thermo- B: +200 ~ +1760 C Program Unit 40 kinds such as , mV, C, kN and mm
couple S: - 10 ~ +1760 C Decimal point Any 0 ~ 6 decimal places
R: - 10 ~ +1760 C Initial value Writing for every channel
E: - 210 ~ +1000 C Sesor mode Setting for every sensor
N: - 200 ~ +1300 C Coefficient 1.0000
Pt RTD - 200 ~ +850 C SIMPLE Unit As per sensor mode
Note : Measuring range in Full bridge 0-2V mode for TML LVDT is measure
Decimal point As per sensor mode
15000 x106 strain (x1) and 150000 x106 strain (x10). Insulation resistance, Dispersion, Sensitivity, Thermocouple
Self-diagnosis wire burnout, Bridge output and coefficient setting
Measuring accuracy
Tempera- Aging TEDS Standard IEEE1451.4 Class 2
Sensor Accuracy Function Readout of TEDS sensor parameter
Range Resolution ture effect effect
mode (23 C5 C)
(%rdg/ C) (%rdg/year) Function Automatic start according to the set time
1 1x10-6 (0.08%rdg+1digit) 0.002 0.02 interval and time
Strain
10 10x10-6 (0.08%rdg+1digit) 0.002 0.02 Interval Hour, min. and sec. up to 99h 59m 59s for
1 0.001mV (0.08%rdg+3digit) 0.0024 0.02 each step
DC voltage
V1/1 10 0.010mV (0.08%rdg+3digit) 0.0024 0.02 No. of starts Programmable 99 times at max. or infinite
DC voltage 1 0.0001V (0.08%rdg+2digit) 0.002 0.02 Interval timer per step
V 1/100 10 0.0010V (0.08%rdg+2digit) 0.002 0.02 No. of steps Programmable 10 steps at max.
Real time start Sets a start time (day: hour: minute: second)
Pt RTD 0.1C (0.08%rdg+3C) 0.0020 0.05 for each step
Pt100 3W
GOTO step Looping previous step
Range : in auto-ranging
Sleep ON/OFF Switches on 5 sec. before measurement
Leadwire resistance correction start and turns off automatically after meas-
urement finish
Comet B Leadwire resistance
Gauge resistance correction range Function Automatic measurement based on a change
(3-wire quarter bridge) Monitor com-
120G Less than 100G amount set by monitor channel (1point)
parator
240G Less than 200G Comparative Amount settable every step (999999 at max.)
350G Less than 300G amount

002-0170.indd 5 10.2.22 4:09:11 PM


SPECIFICATIONS TDS-150

Comparative Available either amount of change or absolute Operational environ- 10 ~ +50 C <85%RH without condensation
ment
Monitor com- method
parator Cycles of start Max. 99 times/step or infinite Storing temperature 20 ~ +60 C
Cycles of step Max. 10 steps programmable LR20 Alkaline cell 4 pieces
GOTO step Programmable loop to previous step Power requirement Exclusive AC adaptor (CR-1861)
External battery 9 ~ 18Vdc
GOTO interval Move to step 1 of interval
TDS-150 Control unit including battery unit
Function Storing and reading of measurement data 280(W) x 60(H) x 162(D) mm
Contents Measure mode, channel number, measure- Dimensions FSW-10 Unit channel (Option)
ment data, time data and data number 280(W) x 60(H) x 100(D) mm
Data memory
Capacity Maximum 80000 data excluging projecting parts
or 16,000 scans per 10 channels TDS-150 Control unit : 1.0 kg
Storage period About 20 days (with full charge) Weight Battery unit : 0.6 kgs (No battery installed)
Standard Compact Flash card FSW-10 Unit channel (Option) : 1.5 kgs.
Memory card
Capacity 32MB ~ 2GB (FAT 16)
Automatically turns off when neither receiving any key opera-
Auto-power Standard accessory
tion nor interface commands for any set time. Switchable LR20 Alkaline cell 4 pieces
OFF
On/ Off. Philips driver 1 piece
Using 4 LR20 alkaline battery Operation manual 1 copy
Operational Full bridge 120G about 40 hours at 23 C5 C Carrying belt 1 piece
time in con- 1 hour sleep interval about 8 months for 10 channels
tinuous use
scanning, Sleep ON at 23 C5 C

Outer View and Dimensional Diagram

302 66
TDS-150 280 60
Control unit
(Battery unit included)
DC9-18V
0.5A MAX

PUSH
62
CARD POWER

USB
179

RS-232C MEMORY RUN-MONI EXT-DC


To FSW FULL
7 8 9

Switching box/TML-NET
100

4 5 6
driving extension board To TML-NET

(Factory installed option) LAN


F1 F2 F3
1 2 3
279

0 ESC ENT

KEY LOCK START


PORTABLE
DATA LOGGER
| TDS-150
TML Tokyo Sokki Kenkyujo Co.,Ltd.

LAN board
(Factory installed option)

SWITCHING BOX

FSW-10
FSW-10
Channel unit CH.NO.
101

(OPTION)
100

A A A A A A A A A A
117

B B B B B B B B B B
(H) (H) (H) (H) (H) (H) (H) (H) (H) (H)

C C C C C C C C C C

D D D D D D D D D D
(L) (L) (L) (L) (L) (L) (L) (L) (L) (L)

E E E E E E E E E E

CH. 0 CH. 1 CH. 2 CH. 3 CH. 4 CH. 5 CH. 6 CH. 7 CH. 8 CH. 9 TML TYPE B-2950

Unit : mm

002-0170.indd 6 10.2.22 4:09:13 PM


Related products
Channel Unit FSW-10 External Printer DPU-H245AS-A03A
Combination with the exclusive switching box makes The measurement data of TDS-150 is printed out..
maximum 50 channels automatic and interval measu-
rement possible.

Printer cable CR-5370 supplied


Dsub9P-10P(mini) thru 0.5m
Exclusive cable

Remote Power Controller RPC-05A RS-232C cable CR-5360


Dsub9P-9P Cross
In combination with RPC-05A and an external battery, Exclusive cable for connection with personal computer
long-term measurement with TDS-150 using sleeping
function becomes possible. USB cable CR-6187
Mini B-A with ferrite core 1.5m
Exclusive cable for connection with personal computer
AC adaptor CR-1861
Compact Flash card
Capacity : 32MB~2GB

External Display Unit EDU-11 TEDS compatible sensor


The monitor value of TDS-150 can be displayed at a To use TEDS functioin of the
remote place. Features high visibility with high- TDS-150, TEDS compatible
brightneess LED. sensor is required to recog-
nize its own parameters such
as measuring capacity, rated
Data cable CR-3830
output, etc. registered in the
BNC output cable
built-in IC chip.
CR-31 belonging to
EDU-11
TEDS compatible load cell
TCLZ with the built-in IC chip

1-Gauge 4-Wire method Strain Gauges

modular plug attached

Specifications subject to change without prior notice

Approval Certificate ISO9001 www.tml.jp/e


Design and manufacture of 8-2, Minami-Ohi 6-Chome, Shinagawa-Ku, TOKYO 140-8560, JAPAN
strain gauges, strain measuring TEL: TOKYO 03-3763-5611 FAX: TOKYO 03-3763-5713
equipment and transducers -mail address: