Beruflich Dokumente
Kultur Dokumente
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Purpose of this article . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Dual Energy vs. Single Energy Dose Comparison Studies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Opportunities for Further Radiation Dose Reduction with DECT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Conclusion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
[1] Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA
[2] Institute of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, Medical Faculty Mannheim, Heidelberg University,
Germany
Introduction | 2
Introduction
visualization of lung perfusion and ventilation [1316],
and the automatic removal of bone from angiographic data
sets [1719].
Since then, various methods for acquiring DECT data have been Figure 1: Adenosine stress DECT coronary CT study of a 64-year-old
proposed for use with recent generations of advanced multi detec- man shows reversible ischemia of the left lateral wall and a papillary
tor-row CT systems: simultaneously applying two X-ray tubes and muscle (A and B). The diagnosis was confirmed by stress SPECT
two corresponding detectors at different kV and mA settings, as with (C and D).
dual-source CT; ultrafast kV switching based on single-source CT;
compartmentalization of detected X-ray photons into energy bins by
double layer detectors of a single-source CT scanner operating at
constant kV and mA settings; or the principle of a counting detector
system which allows dual-energy or spectral CT imaging. The former
methods all aim at acquiring multiple energy data of a given anatomic
area simultaneously, at the same time interval during the gantry rota-
tion. DSCT and double layer technology derive their information from
one set of data, acquired at identical time points. With ultrafast kV
switching there is a small delay of a few ms between readings; how-
ever, the data is fully interlaced as full images are based on several
hundred readings during which the system switched back and forth
multiple times. Another, conceptually different, approach revisits the Figure 2: Coronal DECT images of a 51 year-old man with lung can-
above described earlier concepts of generating dual energy data, cer. (A) represents 120 kV images, (B) shows a calculated virtual non
which involved scanning patients twice, by combining image data contrast image (VNC); (C) shows the color-coded iodine distribution
from two consecutive 270 rotations across the same anatomic re- within the tumor.
gion, one at low kV followed by one at high kV. This latter method
accordingly appears less suitable for imaging organs that change However, this exciting advancement has coincided with a growing
position during scan acquisition, foremost the heart, or patients with public awareness regarding the use of ionizing radiation for medical
limited compliance (trauma, pediatric patients). Also, with such an imaging, and increasing concerns over total radiation dose received
approach, quantitative dual-energy evaluation at applications involv- by the population from imaging tests. Accordingly, successful clini-
ing contrast administration is limited, since contrast concentration cal implementation of multiple-energy imaging will require convincing
changes between two acquisitions which are 270 apart. evidence that these techniques do not come at the price of higher
patient radiation exposure. Dual-source CT and dual-energy CT both
Over the last 5 years, the use of DECT has been evaluated for a va- carry the word dual in their name; likely on this semantic back-
riety of clinical applications. Studies, overwhelmingly based on dual- ground, concern has been raised, whether this technology involves
source CT technology (> 130 peer reviewed articles), have shown increased (as in double) radiation exposure as compared to single-
substantial clinical benefits e.g. in source CT, single-energy investigations.
son Studies
1st generation DSCT in single-energy mode, and 16-slice single
source CT. This investigation reports lower dose levels for DECT and
DSCT than for 16-slice CT [27]. In addition, the investigators con-
Until recently, most studies comparing radiation doses of DECT and clude that DECT delivers significantly lower dose than does regular
single energy CT were performed using DSCT scanners, with com- DSCT, but still maintains comparably good diagnostic image qual-
paratively sparse information available in the literature on the radia- ity. The authors report significantly higher CNR with DSCT in dual-
tion doses involved with the other above mentioned techniques for energy mode compared to both DSCT and 16-slice CT and attribute
acquiring DECT data. In general, considerable caution is in order this observation to the low kV data used in the DECT reconstruction
when interpreting comparative reports on radiation exposure with dif- algorithm. The superior CNR found with low kV protocols is a result
ferent techniques. A multitude of studies indiscriminately report radia- of the increase in the photoelectric effect at lower tube voltages, par-
tion doses with various routine techniques and conclude that one or ticularly in examinations of structures with a high anatomic number,
the other method results in higher or lower radiation, while no at- such as iodinated contrast material [28]. During evaluation, DECT
tention is paid to normalization of image quality, signal/noise ratio, or permits alteration of the balance between images acquired at low kV
DLP. In a recently published phantom study by Schenzle et al. such and highkV data for each study. For example, inclusion of more of
normalization was indeed performed and the authors reported the the low kV data set results in increased attenuation of contrast ma-
feasibility of DECT without increasing radiation dose [20]. Moreover, terial in the coronary arteries. Consequently, this ability to enhance
they observed no significant difference in image noise, but showed contrast attenuation could facilitate a reduction in the total amount
that the contrast to noise ratio (CNR) could be doubled with optimized of contrast medium required for this and other contrast medium en-
DECT reconstructions. Specifically, Schenzle et al. equipped an an- hanced CT applications without incurring higher image noise levels
thropomorphic Alderson phantom with thermoluminescent diodes involved with single-energy low kV imaging.
(TLDs) and scanned its chest with a 1st generation DSCT system in
dual energy mode at 140 and 80 kVp with 14 x 1.2mm collimation As mentioned above, very limited information is available in the lit-
as well as on a 2nd generation DSCT system at 140 kVp and 100 erature regarding radiation doses associated with multi-energy imag-
kVp with selective photon shielding at 128 x 0.6 mm collimation. For ing on platforms that are not based on dual-source CT technology.
dose comparison with single energy CT, the investigators obtained Ho et al. observed two to three times higher doses for DECT [25]
reference examinations at 120 kVp with 64 x 0.6 mm collimation at using a single-source system and rapid voltage switching (22.5 to
an equivalent CT dose index of 5.4 mGy*cm. In this phantom study, 36.4mSv for DECT vs. 9.4 to 13.8 mSv for single-source CT). This
the authors report effective dose measurements with TLDs, that were group did not perform normalization of image noise or dose; thus
equal for both DSCT systems in dual energy mode at 140/80 kV the lower energy spectrum was obtained with the same tube current
and 140/100 kV with selective photon shielding, when compared to time product as the single energy scan. In another study that did per-
the single energy 120 kV reference examination [2.61mSv (1st gen- form normalization to equivalent low-contrast detectability, Li et al.
eration DSCT in DECT mode), 2.69 mSv (2nd generation DSCT in observed an additional dose of 14% in the body and 22% in the head
DECT mode), and 2.70 mSv (2nd generation DSCT in single-energy using DECT with rapid kV switching in comparison to that of single
mode)]. Moreover, the image noise was reported similar for all three energy CT [29]. At this point, no study has systemically evaluated the
of the different imaging techniques. The selective photon shielding radiation dose of multiple energy image acquisitions using a single
evaluated in this study is based on filtering techniques, aimed at fur- source dual-layer detector CT system or combining image data from
ther spectral separation and thus higher discriminatory abilities for two consecutive 270 rotations. Furthermore, clinical studies specifi-
tissue characterization of dual-energy CT acquisitions. Similarly, in a cally comparing the dose efficiency of different DECT systems are
phantom study performed on a DSCT system, Yu et al. showed that lacking.
images blended from low and high tube-potential data yield similar or
even better iodine CNR than that of typical 120 kV images acquired
at the same radiation dose [21].
Conclusion
Of the various methods that have been proposed for acquiring DECT
data, image acquisition based on dual-source CT is the most in-
tensely evaluated approach in the literature. The available literature
suggests that there is indeed no radiation dose penalty with the use
of dual-source CT based DECT over single-energy protocols. Re-
ports on radiation dose with other approaches for DECT data acqui-
sition are scarce or non-existent, so that a conclusive evaluation of
Figure 3: 120 kV image in sagittal orientation demonstrates a small the radiation dose associated with these techniques remains elusive
isolated segmental pulmonary embolism of the right lower lobe (A). to date. Finally, judicious use of DECT techniques holds the potential
DECT iodine map demonstrates the corresponding wedge shaped of drastically reducing radiation exposure, e.g. via the elimination of
perfusion defect (B). unnecessary non-contrast CT scans.
Graser et al. published two studies using 1st and 2nd generation
DSCT systems to assess the performance of DECT in renal mass
characterization [6, 30]. In these studies, the replacement of a real
non-contrast CT scan with a virtual non-contrast scan resulted in a
35% (1st generation DSCT) and a 50% (2nd generation DSCT) re-
duction of the radiation dose. Similar findings of potential decreases
in radiation dose via virtual non-contrast DECT scans have been re-
ported by Leschka et al. and Brown et al. in phantom models [31,
32]. Furthermore, Zhang et al. compared virtual non-contrast scans
of the liver based on DECT with true non-contrast liver CT examina-
tions. Based on their results, they concluded that virtual non-con-
trast DECT could replace true non-contrast CT scans as part of a
multi-phase liver imaging protocol. In this study, the use of a bipha-
sic DECT examination with virtual non-contrast images instead of a
triphasic single energy CT study of the liver resulted in a 33% reduc-
tion of CTDIvol and a 34% reduction of DLP [33].
References 13. Chae EJ, Seo JB, Goo HW, et al. Xenon ventilation CT with
a dual-energy technique of dual-source CT: initial experience.
1. Graser A, Johnson TR, Bader M, et al. Dual energy CT char- Radiology. 2008;248:615624.
acterization of urinary calculi: initial in vitro and clinical experi-
ence. Invest Radiol. 2008;43:112119. 14. Thieme SF, Becker CR, Hacker M, et al. Dual energy CT for the
assessment of lung perfusion--correlation to scintigraphy. Eur J
2. Scheffel H, Stolzmann P, Frauenfelder T, et al. Dual-energy Radiol. 2008;68:369374.
contrast-enhanced computed tomography for the detection of
urinary stone disease. Invest Radiol. 2007;42:823829. 15. Henzler T, Meyer M, Reichert M, et al. Dual-energy CT angiog-
raphy of the lungs: Comparison of test bolus and bolus tracking
3. Thomas C, Heuschmid M, Schilling D, et al. Urinary calculi techniques for the determination of scan delay. Eur J Radiol.
composed of uric acid, cystine, and mineral salts: differentia- 2010.
tion with dual-energy CT at a radiation dose comparable to that
of intravenous pyelography. Radiology. 2010;257:402409. 16. Kerl JM, Bauer RW, Renker M, et al. Triphasic contrast injection
improves evaluation of dual energy lung perfusion in pulmo-
4. Thomas C, Krauss B, Ketelsen D, et al. Differentiation of uri- nary CT angiography. Eur J Radiol. 2010.
nary calculi with dual energy CT: effect of spectral shaping by
high energy tin filtration. Invest Radiol. 2010;45:393398. 17. Morhard D, Fink C, Graser A, et al. Cervical and cranial com-
puted tomographic angiography with automated bone removal:
5. Robinson E, Babb J, Chandarana H, et al. Dual source dual dual energy computed tomography versus standard computed
energy MDCT: comparison of 80 kVp and weighted average tomography. Invest Radiol. 2009;44:293297.
120 kVp data for conspicuity of hypo-vascular liver metastases.
Invest Radiol. 2010;45:413418. 18. Sommer WH, Johnson TR, Becker CR, et al. The value of dual-
energy bone removal in maximum intensity projections of lower
6. Graser A, Becker CR, Staehler M, et al. Single-phase dual-en- extremity computed tomography angiography. Invest Radiol.
ergy CT allows for characterization of renal masses as benign 2009;44:285292.
or malignant. Invest Radiol. 2010;45:399405.
19. Brockmann C, Jochum S, Sadick M, et al. Dual-energy CT an-
7. Chae EJ, Song JW, Seo JB, et al. Clinical utility of dual-energy giography in peripheral arterial occlusive disease. Cardiovasc
CT in the evaluation of solitary pulmonary nodules: initial expe- Intervent Radiol. 2009;32:630637.
rience. Radiology. 2008;249:671681.
20. Schenzle JC, Sommer WH, Neumaier K, et al. Dual energy CT
8. Schwarz F, Ruzsics B, Schoepf UJ, et al. Dual-energy CT of of the chest: how about the dose? Invest Radiol. 2010;45:347
the heart--principles and protocols. Eur J Radiol. 2008;68:423 353.
433.
21. Yu L, Primak AN, Liu X, et al. Image quality optimization and
9. Ruzsics B, Lee H, Zwerner PL, et al. Dual-energy CT of the evaluation of linearly mixed images in dual-source, dual-energy
heart for diagnosing coronary artery stenosis and myocardial CT. Med Phys. 2009;36:10191024.
ischemia-initial experience. Eur Radiol. 2008;18:24142424.
22. Deseive S, Bauer RW, Lehmann R, et al. Dual-Energy Com-
10. Weininger M, Schoepf UJ, Ramachandra A, et al. Adenosine- puted Tomography for the Detection of Late Enhancement in
stress dynamic real-time myocardial perfusion CT and adeno- Reperfused Chronic Infarction: A Comparison to Magnetic Res-
sine-stress first-pass dual-energy myocardial perfusion CT for onance Imaging and Histopathology in a Porcine Model. Invest
the assessment of acute chest pain: Initial results. Eur J Radiol. Radiol. 2011.
2010.
23. Kang DK, Schoepf UJ, Bastarrika G, et al. Dual-energy com-
11. Ruzsics B, Lee H, Powers ER, et al. Images in cardiovascu- puted tomography for integrative imaging of coronary artery
lar medicine. Myocardial ischemia diagnosed by dual-energy disease: principles and clinical applications. Semin Ultrasound
computed tomography: correlation with single-photon emission CT MR. 2010;31:276291.
computed tomography. Circulation. 2008;117:12441245.
24. Ruzsics B, Schwarz F, Schoepf UJ, et al. Comparison of dual-
12. Bauer RW, Kerl JM, Fischer N, et al. Dual-energy CT for the energy computed tomography of the heart with single photon
assessment of chronic myocardial infarction in patients with emission computed tomography for assessment of coronary
chronic coronary artery disease: comparison with 3-T MRI. AJR artery stenosis and of the myocardial blood supply. Am J Car-
Am J Roentgenol. 2010;195:639646. diol. 2009;104:318326.
25. Ruzsics B, Chiaramida SA, Schoepf UJ Images in cardiology: 38. Laks S, Macari M, Chandarana H Dual-energy computed
Dual-energy computed tomography imaging of myocardial in- tomography imaging of the aorta after endovascular repair
farction. Heart. 2009;95:180. of abdominal aortic aneurysm. Semin Ultrasound CT MR.
2010;31:292300.
26. Nance JW, Jr., Bastarrika G, Kang DK, et al. High-temporal res-
olution dual-energy computed tomography of the heart using a 39. Moscariello A, Takx RAP, Schoepf UJ, et al. Coronary CT An-
novel hybrid image reconstruction algorithm: initial experience. giography: Image Quality, Diagnostic Accuracy, and Potential
J Comput Assist Tomogr. 2011;35:119125. for Radiation Dose Reduction Using a Novel Iterative Image
Reconstruction Technique - Comparison with Traditional Fil-
27. Kerl JM, Bauer RW, Maurer TB, et al. Dose levels at coronary tered Back Projection. Eur Radiol. 2011;accepted manuscript
CT angiography-a comparison of Dual Energy-, Dual Source- in press.
and 16-slice CT. Eur Radiol. 2010.
40. Herzog C, Mulvihill DM, Nguyen SA, et al. Pediatric cardiovas-
28. Pflederer T, Rudofsky L, Ropers D, et al. Image quality in a low cular CT angiography: radiation dose reduction using automat-
radiation exposure protocol for retrospectively ECG-gated cor- ic anatomic tube current modulation. AJR Am J Roentgenol.
onary CT angiography. AJR Am J Roentgenol. 2009;192:1045 2008;190:12321240.
1050.
29. Li B Head and body CTDIw of dual energy x-ray CT with fast-
kVp switching. SPIE Medical Imaging. 2010:76227669.
36. Sommer WH, Graser A, Becker CR, et al. Image quality of virtu-
al noncontrast images derived from dual-energy CT angiogra-
phy after endovascular aneurysm repair. J Vasc Interv Radiol.
2010;21:315321.