Sie sind auf Seite 1von 8

Analytica Chimica Acta 383 (1999) 283±290

Determination of paracetamol in intact tablets by use


of near infrared transmittance spectroscopy
A. Eustaquioa, M. Blancoa,*, R.D. Jeeb, A.C. Moffatb
a
Departamento de QuõÂmica, Unitad de QuõÂmica AnalõÂtica, Universidad AutoÂnoma de Barcelona, E-08193 Bellaterra, Barcelona, Spain
b
Centre of Pharmaceutical Analysis, The School of Pharmacy, University of London, 29/39 Brunswick Square, London WC1N 1AX, UK

Received 10 July 1998; received in revised form 16 November 1998; accepted 16 November 1998

Abstract

Near infrared transmittance spectroscopy was used to determine the analgesic paracetamol in a pharmaceutical preparation
commercially available as tablets. Spectra were recorded on a dedicated instrument that measures the transmission of intact
tablets over the wavelength range 600±1900 nm. Spectral data were processed by using two multivariate calibration methods,
viz. stepwise multiple linear regression (SMLR) and partial least-squares regression (PLSR). The analgesic contents provided
by the two calibration methods were comparable and differed by less than 1% from the reference (UV spectrophotometric)
value. The calibration graphs constructed to determine the mean paracetamol content were used to analyse content uniformity
in the tablets. # 1999 Published by Elsevier Science B.V. All rights reserved.

Keywords: Near infrared transmittance spectroscopy; Intact tablet analysis; Stepwise multiple linear regression; Partial least-squares
regression; Content uniformity

1. Introduction The NIRS technique has proved a highly useful tool


for controlling powdered pharmaceuticals (intact
The high industrial interest aroused by near infrared powdered raw materials, intermediates and dosage
spectroscopy (NIRS) in recent years is a direct result forms), which it can readily identify and quantify
of dramatic advances in this technique including the [4]. The ability to analyse intact dosage forms (parti-
increasing availability of fast-scan instruments that cularly tablets, which are the most widely used form)
enable measurements with little or no sample manip- has dramatically facilitated monitoring of every single
ulation and of software for implementing powerful step in the production process as well as of the ®nal
chemometric techniques. Also the increasing aware- form, all with minimal sample manipulation. On the
ness and acceptance of NIRS has facilitated the expe- other hand, direct tablet analysis by NIRS has devel-
ditious analysis of many complex samples with oped to a much lesser extent owing to technical
no need for reagents or solvents in a variety of ®elds constraints [5]. The earliest applications of NIRS to
[1±3]. intact tablets were reported in 1988 [6]; the whole
surface of a double-re¯ecting aluminium sample con-
*Corresponding author. Fax: +34-93-581-24-77; tainer was illuminated for the qualitative analysis
e-mail: iqan8@blues.uab.es of tablets. Later applications have been primarily

0003-2670/99/$ ± see front matter # 1999 Published by Elsevier Science B.V. All rights reserved.
PII: S0003-2670(98)00815-0
284 A. Eustaquio et al. / Analytica Chimica Acta 383 (1999) 283±290

concerned with qualitative analyses [7]. Most reported In this work, near infrared transmittance spectro-
direct NIRS analyses of tablets are subject to two scopy was tested for the analysis of tablets about 4 mm
hindrances, namely: (a) the lack of appropriate devices thick. This type of sample is representative of many
for making reproducible sample measurements and (b) commercially available tablets and consequently, if
the fact that measurements are made by re¯ectance, so thick tablets can be satisfactorily analysed, the NIR
the information acquired corresponds to the sample technique should be applicable to virtually any other
surface, which may lead to spurious results with type of tablet.
coated tablets.
These shortcomings are circumvented by recently 2. Experimental
developed NIR instruments of improved design. Their
measuring modules provide more reproducible and 2.1. Apparatus and software
reliable measurements of individual tablets. Also, the
new instruments use transmission measurements (i.e. Near infrared spectra were recorded on an NIRSys-
the NIR beam passes through the whole sample thick- tem 6500 spectrophotometer (FOSS, NIRSystems)
ness). Using these instruments provides a number of equipped with an Intact Tablet Analyser that was in
advantages, namely, turn furnished with an indium±gallium±arsenide
(InGaAs) detector which provides highly linear,
1. the ability to draw information for single tablets low-noise measurements of sample spectra and allows
from their individual spectra; processing of bands up to 3 absorbance units above the
2. the ability to analyse coated tablets, the coating baseline [8]. The spectrophotometric system was opti-
thickness posing no special problem since the mised to measure tablet absorbances over the visible
spectrum is due to the whole tablet; (Vis) and near infrared (NIR) spectral ranges.
3. increased reproducibility in the measurements ± The equipment was controlled via the NSAS v. 3.52
those of individual tablets included ± by virtue of software package (FOSS, NIRSystems), which
the response being produced by a sizeable amount enables automated acquisition and processing of
of sample ± its entire thickness; NIR spectra. The same software was used to perform
4. simple calibration models relative to those typi- SMLR computations.
cally used with reflectance measurements ± the PLS calibration was done by using the multivariate
recorded signal (absorbance) can be assumed to calibration program Unscrambler v. 6.1, from CAMO
obey the Lambert±Beer law since the improved AS (Trondheim, Norway), which affords direct import
instrument design results in minimal scatter. of NIR spectra.
The reference spectroscopic method was implemen-
On the other hand, the most severe constraint on ted on a Perkin-Elmer Lambda 15 UV/Vis spectrometer.
transmission measurements is posed by the thickness
of typical tablets. Because the light beam must travel 2.2. Samples
through a large amount of product, only a small
fraction of incident light reaches the detector. The The specimens studied were circular cores about
high absorbances involved in this type of measure- 4 mm thick consisting of paracetamol as active prin-
ment ± most often in excess of 4 absorbance units ± ciple (nominal content 84% m/m) and six different
have compelled instrument manufacturers to improve excipients.
existing commercial equipment by minimising losses Twenty-®ve production batches containing 99±
of incident light on its way to the detector (the 102% of the nominal paracetamol value, and 10
instrument design should also minimise stray light batches of tablets made at a pilot plant were studied.
reaching the detector, e.g. by passing around the sides The latter allowed the content range to be extended to
of the tablet sample) and using highly sensitive detec- 90±110% of the nominal value.
tors affording readings of much higher absorbances The UV spectrophotometric procedure used for
than those usually recorded in other spectroscopic reference was that of®cially recommended by the
techniques. British Pharmacopoeia [9].
A. Eustaquio et al. / Analytica Chimica Acta 383 (1999) 283±290 285

2.3. Recording of NIR spectra Tablets were always placed in the holder the ¯at
side up (this was facilitated by the tablets having a
The instrument used permits the recording of different mark on each side) and the holder inserted
transmission spectra of intact tablets. The measuring into the instrument. The spectra (average of 32 scans)
system was optimised by constructing a holder to for each of 20 tablets from each batch were recorded
closely ®t a tablet in order to ensure maximal from 600 to 1900 nm, using air as reference. First and
reproducibility in the sample position and minimal second derivative absorbance spectra were calculated
stray light. The holder, which held a single tablet using the NSAS software with zero gap size and 10
of the product being tested, was made as per the and 20 segment sizes, respectively.
instructions given in the instrument's user guide Fig. 1 shows the NIR spectra obtained in different
[10]. modes for the samples containing the extreme para-

Fig. 1. Absorbance, first-derivative (absorbance) and second-derivative (absorbance) near infrared transmittance spectra for three different
samples: (A) and (C) sample with a paracetamol content equal to 90% and 110%, respectively, of the nominal value; (B) production sample.
286 A. Eustaquio et al. / Analytica Chimica Acta 383 (1999) 283±290

cetamol contents examined as well as that for a normal The results provided by the different models tested
production sample. There are signi®cant differences were compared via the relative standard error of
on the spectra around 1400±1500 nm; these differ- prediction (RSEP), given by [13]
ences have been only observed in overdosed pilot sP
n
plant samples and are not correlated to the paraceta- iˆ1 P …CLABi ÿ CNIRi †2
RSEP …%† ˆ n 2
 100; (3)
mol content in the samples. Its source is unknown. iˆ1 CLABi

2.4. Data processing where CLAB and CNIR are the paracetamol content deter-
mined by the reference and NIR methods, respectively.
The mean paracetamol content was determined
from the mean spectrum for each batch (i.e. the mean 3. Results and discussion
of the 20 individual spectra recorded). Paracetamol
was quanti®ed by using two different multivariate Available samples were split into two different sets
calibration methods, viz. inverse stepwise linear mul- for calibration and validation, respectively. The cali-
tiple regression (SMLR) and partial least-squares bration set consisted of 20 batches with paracetamol
regression (PLSR). contents that evenly spanned the range from 90% to
With the SMLR method, each model was optimised 110% of the nominal value; it included production and
in terms of the number of wavelengths used in order to prepared at the pilot plant samples. The remaining 15
avoid over®tting; an F statistical criterion was applied batches were used as the validation set and spanned the
to the sum of the squares of the residuals for the range from 95% to 105% of the nominal paracetamol
calibration samples in the models [11]. The different content; it also included both types of samples.
models thus constructed were compared by estimating
the goodness of ®t from the multiple correlation 3.1. Determination by stepwise multiple linear
coef®cient regression
s
Pn 
…C NIRi ÿ 
C LABi † 2 The wavelengths used in constructing the SMLR
R ˆ Pniˆ1 ; (1) model were chosen in the ascending mode. The start-
…CLABi ÿ C  LABi †2
iˆ1 ing wavelength range was 600±1350 nm ± the region
where CNIR is the paracetamol content calculated by from 1350 to 1900 nm was excluded owing to the high
using the NIR method and C LAB the mean paracetamol absorbances and noise associated with it.
content in the calibration samples provided by the Results were compared on the basis of two different
reference method. For accurate comparison, this para- criteria for wavelength selection, namely:
meter must be corrected for the number of degrees of A. Automatic selection. The wavelength resulting
freedom: in the highest possible correlation between
  absorbance and concentration was chosen and
2 mÿ1
Raj ˆ 1 ÿ …1 ÿ R2 †; (2) the equation was successively expanded with
mÿn
new terms by incorporating those previously
where n is the number of coef®cients in the equation unused wavelengths that met the same criterion.
and m the number of calibration samples used. While The number of wavelengths used in each model
R2 increases and approaches unity as the number of was optimised in terms of the above-mentioned
terms is increased, R2aj tends to level off after the F criterion.
optimum model has been reached. B. Imposing a first wavelength based on a
PLS models were constructed by cross-validation, chemical criterion. The first wavelength used
using as many segments as samples were included in was that where differences between the spectra
the calibration matrix. The optimum number of PLS for the samples containing the extreme con-
components was chosen according to the criterion of centrations of paracetamol (Fig. 1) were max-
Haaland and Thomas [12]. The calibration procedure imal, the software was allowed to automatically
was applied to centred, unscaled data. select successive wavelengths as in A.
A. Eustaquio et al. / Analytica Chimica Acta 383 (1999) 283±290 287

Table 1
Relative standard error of prediction for the calibration set (RSEP(C)), validation set (RSEP(V)) and R2aj as obtained by using different SMLR
methods

Mode R2aj RSEP(C) (%) RSEP(V) (%) Wavelengths (nm)

Absorbance A 0.989 0.6 0.8 1206, 1186


B 0.989 0.6 0.8 1206, 1186

First derivative A 0.989 0.6 0.7 900


B 0.991 0.6 0.8 1190, 886

Second derivative A 0.989 0.6 1.0 1212, 850


B 0.990 0.6 0.9 1202, 820
A: automatic wavelength selection; B: first wavelength based on chemical criterion (see text for more details).

Calibration equations were derived from spectra than ttab (2.14, Pˆ0.05, dfˆ14), the proposed proce-
subjected to various treatments (absorbance, ®rst deri- dure exhibits no statistically signi®cant differences
vative, second derivative). from the reference method.
Table 1 gives the wavelengths used for the models
and their corresponding errors of prediction; in the 3.2. Determination by partial least-squares
absorbance spectral mode, criterion B was equivalent regression
to criterion A. It can be seen that R2aj was similar for all
the models examined, and therefore that with the Although PLSR is a whole-spectrum technique,
smallest RSEP(V) was chosen. This was one of the quantitations based on it can bene®t from careful
simplest models, using ®rst derivative (absorbance) selection of spectral wavelengths. We examined two
spectra and a single wavelength. different ranges, viz. 600±1350 and 600±1900 nm.
Table 2 gives the results obtained for the validation Although the high end of the spectrum exhibited high
samples using the optimum model. The results were absorbances and considerable noise, we assayed using
compared to the reference values using a paired PLSR to model the system in order to determine
samples Students' t test. As texp (0.354) was smaller whether the broader range would lead to improved
results.
Table 2
Table 3 shows the errors of prediction for the
Paracetamol contents (% m/m) in the validation samples provided models tested. As can be seen, the predictive capacity
by the optimum SMLR and PLSR models was not affected by the spectral treatment used; how-
ever, the models constructed from whole spectra were
Sample [Paracetamol]LAB [Paracetamol]SMLR [Paracetamol]PLSR
generally more complex (they used more PLS com-
3 80.3 79.2 79.3 ponents); this can be ascribed to small baseline shifts
6 83.9 84.6 85.0
among spectra, shifts that were minimised by deriva-
9 84.1 84.2 83.8
11 84.2 85.2 84.6 tion. Models constructed by applying Savitzky±Golay
13 84.2 84.5 84.7 smoothing (9 data points, third-order polynomial) to
15 84.3 84.5 84.6 the spectra were also tested; however, their results are
17 84.4 85.0 85.1 not reported because they did not improve on those of
19 84.4 84.8 84.7
the other models in any way.
21 84.7 83.4 83.5
23 84.9 84.7 84.8 As the prediction capability is comparable in all
25 85.1 84.5 84.4 models assayed, the best PLS model was selected on
27 85.1 84.6 84.5 the basis of its simplicity. In this case, the simplest
28 85.5 85.0 85.0 model needed only one PLS-component and was
30 85.7 85.5 84.7
calculated using the shorter spectral range and ®rst-
33 88.5 88.5 88.5
derivative spectra. The results for the samples in the
288 A. Eustaquio et al. / Analytica Chimica Acta 383 (1999) 283±290

Table 3
Relative standard error of prediction calculated for the calibration set (RSEP(C)) and validation set (RSEP(V)) as obtained by using different
PLSR models

Range (nm) No. PLS-comp. RSEP(C) (%) RSEP(V) (%)

Absorbance 600±1350 3 0.6 0.8


600±1900 5 0.6 0.8

First derivative 600±1350 1 0.7 0.8


600±1900 3 0.6 0.8

Second derivative 600±1350 2 0.6 0.8


600±1900 2 0.6 0.8

validation set obtained by this model are given in Table 4


Table 2. Again, a paired samples Students' t test Paracetamol contents (% m/m) in validation samples obtained by
using a calibration set consisting of pilot plant samples only
revealed that the proposed procedure is not signi®-
cantly different from the reference method (texpˆ Sample [Paracetamol]LAB [Paracetamol]PLSR
0.136, ttabˆ2.14, Pˆ0.05, dfˆ14). 5 83.5 85.0
The simplicity of the optimum model, which uses a 6 83.9 85.3
single PLS component, suggested that the calibration 7 84.0 85.6
set could be constructed from only samples made at 8 84.1 85.1
10 84.1 84.6
the pilot plant permitting a decreased number of 12 84.2 85.1
samples for calibration. To verify this hypothesis, 13 84.2 85.1
we constructed a new PLSR model using the previous 14 84.3 85.2
optimum parameters (viz. the 600±1350 range and 15 84.3 85.0
®rst-derivative spectral mode) and the 10 spectra for 16 84.3 85.0
17 84.4 85.4
samples made at the pilot plant. Table 4 shows the
18 84.4 84.8
paracetamol contents in the 25 production batches 19 84.4 85.0
obtained with this new calibration model; the RSEP 20 84.5 85.1
was 0.9%, indicating that the predictive capacity is 21 84.7 83.8
somewhat lower than the model which included pro- 22 84.8 85.1
23 84.9 85.1
duction samples in the calibration set. Although the 24 85.0 84.9
pilot plant samples are very similar to the production 25 85.1 84.7
samples, the predictive capacity of the model is 26 85.1 85.0
improved if variability in the production samples is 27 85.1 84.8
introduced in the calibration. This agrees with pre- 28 85.5 85.3
29 85.6 85.2
vious studies using calibrations including synthetic 30 85.7 85.0
[14] or overdosed samples [15,16]. 31 85.8 85.3
RSEP (%) 0.9

3.3. Content uniformity


The European Pharmacopoeia [17] test for unifor-
A measuring system capable of recording spectra mity of content requires the speci®ed active principle
for individual tablets provides an additional advan- in 10 randomly chosen tablets to be measured.
tage, viz. the ability to simultaneously determine, Although the test is strictly only required for tablets
using the same calibration, the content in individual in which the active principle is less than 2 mg or less
tablets. Fig. 2 shows the spectra of 20 individual than 2% (m/m), we have applied the ``spirit'' of the
tablets from a single production batch. test to the paracetamol tablets. Using the ®rst 10
A. Eustaquio et al. / Analytica Chimica Acta 383 (1999) 283±290 289

Fig. 2. Near infrared transmittance spectra for 20 individual paracetamol tablets from a production batch.

Table 5 models provided the same coef®cient of variation


Study of content uniformity in a production batch
(CV). The relative paracetamol contents obtained in
SMLR PLSR all the determinations were all within the accepted
Tablet limits of the standard (viz. 85±115% of the nominal
[Parac.]SMLR Nominal [Parac.]PLSR Nominal
(%) value (%) (%) value (%) value).
1 84.6 100.5 84.6 100.5
2 84.3 100.1 84.3 100.1
3 84.6 100.5 84.8 100.7 4. Conclusions
4 84.5 100.4 84.7 100.6
5 85.0 101.0 85.0 101.0 The results obtained in this work reveal that the
6 84.5 100.4 84.7 100.6 direct analysis of intact tablets by near infrared trans-
7 84.4 100.3 84.8 100.7
mittance spectroscopy is an effective way of quantify-
8 84.3 100.1 84.3 100.1
9 84.5 100.4 84.7 100.6 ing this type of end product. The ensuing methodology
10 84.3 100.1 84.4 100.3 is accurate enough to ascertain whether the content in
CV (%) 0.3 0.3 the active principle assayed is within the speci®ed
range and can thus be used to control the process. In
addition, a single calibration allows the simultaneous
individual spectra previously recorded we evaluated determination of the mean content and content uni-
them by using the optimum SMLR and PLSR models. formity in a production batch.
Table 5 shows the per cent contents obtained relative The two calibration techniques tested, SMLR and
to the nominal value. As can be seen, both calibration PLSR, provide quite satisfactory results and exhibit a
290 A. Eustaquio et al. / Analytica Chimica Acta 383 (1999) 283±290

comparable predictive ability. The simplicity of the [3] K.A. Bunding Lee, Appl. Spectrosc. Rev. 28 (1993) 231.
SMLR and PLSR calibration models used suggests [4] M. Blanco, J. Coello, H. Iturriaga, S. Maspoch, C. de la
Pezuela, Analyst, 123 (1998) 135 R.
that transmittance measurements are highly correlated [5] J.D. Kirsch, J.K. Drennen, Appl. Spectrosc. Rev. 30 (1995)
with the analyte content since the typical scatter in 139.
diffuse re¯ectance measurements is negligible here. [6] R.A. Lodder, G.M. Hieftje, Appl. Spectrosc. 42 (1987)
556.
[7] M.A. Dempster, J.A. Jones, I.R. Last, B.F. MacDonald, K.A.
Acknowledgements Prebble, J. Pharm. Biomed. Anal. 11 (1993) 1087.
[8] Intact1 Tablet Analyzer Manual, FOSS (NIRSystems), 1996,
The authors are grateful to Spain's DGICyT (Pro- p. 1.
ject PB96-1180) for funding this work. A. Eustaquio [9] British Pharmacopoeia 1993, vol. II, London, HMSO, 1993,
p. 1043.
also acknowledges additional funding from Spain's [10] Intact1 Tablet Analyzer Manual, tablet construction design,
Ministry of Education and Science in the form of an FOSS (NIRSystems), 1996.
FPI grant. The authors also wish to thank Sano® for [11] M. Blanco, J. Coello, H. Iturriaga, S. Maspoch, E. Rovira, J.
kindly supplying the production and pilot plant sam- Pharm. Biomed. Anal. 16 (1997) 255.
ples, and A.D. Trafford for providing the reference [12] D.M. Haaland, E.V. Thomas, Anal. Chem. 60 (1988)
1193.
values for all the samples. [13] M. Otto, W. Wegscheider, Anal. Chem. 57 (1985) 63.
[14] M. Blanco, J. Coello, H. Iturriaga, S. Maspoch, C. de la
References Pezuela, Anal. Chim. Acta 333 (1996) 147.
[15] M. Blanco, J. Coello, H. Iturriaga, S. Maspoch, C. de la
Pezuela, Analyst 112 (1997) 761.
[1] B.G. Osborne, T. Fearn, Near Infrared Spectroscopy in Food [16] E. Dreassi, G. Ceramelli, P. Corti, P.L. Perruccio, S. Lonardi,
Analysis, Longman, Harlow, 1986. Analyst 121 (1996) 219.
[2] M. Blanco, J. Coello, H. Iturriaga, S. Maspoch, E. Bertran, [17] European Pharmacopoeia 1997, 3rd ed., Strasbourg, Council
Analyst 119 (1994) 1779. of Europe, 1997, p. 1753.

Das könnte Ihnen auch gefallen