Sie sind auf Seite 1von 19

Sample chapter from

BSAVA Manual of Canine and Feline


Anaesthesia and Analgesia
2nd edition
Edited by Chris Seymour and Tanya Duke-Novakovski
© BSAVA 2007

www.bsava.com
Chapter 7 Patient monitoring and monitoring equipment

7
Patient monitoring and
monitoring equipment
Yves Moens and Paul Coppens

Introduction
Anaesthetic machine
When performing general anaesthesia, a reversible Check gas source(s):
‘coma-like’ state is produced. Most drugs used in anaes- • Open cylinder(s)
thesia have an inherent degree of toxicity and often • Check quantity
produce cardiovascular and respiratory side effects. • Check output pressure
Check flowmeter(s)
These side effects jeopardize body homeostasis and
Check vaporizer(s):
make anaesthesia a potentially dangerous procedure. • Off position
In order to limit the risks of general anaesthesia, • Quantity
‘controllable’ problems must be prevented. Patient- Check oxygen bypass
related problems can be detected by the pre-anaes- Choose and connect a breathing system
thetic examination (see Chapter 2). Checking Check the breathing system; carry out a leak test:
equipment will help to prevent problems due to tech- 1. Close exhaust valve
2. Obstruct the patient extremity of the breathing system
nical errors (Figure 7.1; see also Chapters 4, 5 and
3. Fill system with oxygen until reservoir bag is under tension
6). Furthermore, basic prevention of problems involves (40 cmH2O)
maintaining the patient at an appropriate anaesthetic 4. Check if pressure is kept, otherwise find and fix leak
level throughout the procedure. However, even with 5. Open exhaust valve
adequate preparation, it is not possible to eliminate If ventilator available:
risk because every patient may exhibit individual and • Check function
unpredictable reactions to anaesthesia. Earliest de- • Carry out a leak test
• Preset the parameters
tection of any adverse reaction by the patient (by
closely observing physiological functions during an- Monitoring equipment
aesthetic administration) is vital. Therefore, attention • Power on
should be focused on the monitoring of vital systems, • Battery state
such as the cardiovascular and the respiratory sys- • Test
tems, which are responsible for correct delivery of • Calibration if necessary
oxygen to the organs. Additionally, other parameters Specific equipment
reflecting general homeostasis and the degree of anti- • Intravenous catheter
nociception and unconsciousness should be carefully • Intubation
observed. If deviations from normality outside defined • Fluids
limits occur, analysis of the situation is necessary and • Hypothermia prevention
corrective measures should be taken. The results of • Ophthalmic ointment
such action must also be evaluated. Resuscitation equipment availability
Monitoring is performed in various ways. Initially, Prepare the anaesthesia record
the veterinary surgeon must always make full use of
his/her own senses (vision, touch, smell, hearing) to The preventive approach: pre-anaesthetic
7.1
collect clinical information, on which decisions can be equipment check.
made as to whether intervention is necessary. Clinical
examination, although very relevant, does have limita- able to collect information on a continuous basis and
tions. The information is, in general, more qualitative can upgrade clinical observation with more sensitive
than quantitative. Moreover, when staff availability is information. The risk of anaesthesia in veterinary medi-
limited, or when the veterinary surgeon monitors more cine remains very high compared with that of human
than one patient (or performs some adjustments to anaesthesia. Healthy dogs and cats run a mortality risk
equipment), continuous collection of patient parameters of 0.054% and 0.112%, respectively (Brodbelt et al.,
will be difficult if not impossible. Therefore, monitoring 2005). Actual anaesthetic mortality in humans is re-
can be complemented by mechanical and electronic ported to be as low as 0.05 for every 10,000 patients
auxiliary equipment. During the last two decades, sev- (Arbous et al., 2001). One should bear in mind that in
eral new devices have come on to the market and more human medicine (unlike veterinary medicine), anaes-
affordable prices for these have led to their increasing thesia is performed by a medical professional specially
popularity in veterinary anaesthesia. These devices are trained in anaesthesia.

62
Chapter 7 Patient monitoring and monitoring equipment

Monitoring equipment occasionally provides observation of several parameters (integrated moni-


artefactual and wrong information. Hence, the func- toring) is necessary to build a more complete picture
tion and limits of these devices (as well the physio- of vital system function in the individual animal and
logical significance of the information they provide) the function of the anaesthetic machine. In this way,
must be very well understood if they are to be used monitoring is the central tool to ‘pilot’ anaesthesia with
to increase patient safety. Furthermore, this inform- enhanced safety.
ation must be used alongside clinical judgement
when evaluating a situation and deciding on possi-
ble therapeutic intervention.
It is erroneous to think that monitoring may be The main monitor: human senses
omitted for some ‘simple’ or ‘quick’ procedures, with
specific ‘safe’ drugs or during ‘stable’ episodes of Veterinary surgeons are used to performing a clini-
anaesthesia. Unexpected events often occur during cal examination, using their senses, on a daily
short and simple procedures on apparently healthy basis and should adapt this skill to conditions
patients. Both clinical and technical monitoring must during anaesthesia (Figure 7.2). Moreover, even
not be limited only to the patient but also include the today, the most appropriate method for assessing
anaesthetic machine. the depth of anaesthesia in veterinary practice
Monitoring must not concentrate either on one clini- remains the assessment of clinical signs (Figures 7.3
cal sign or one measured parameter. Simultaneous and 7.4).

Sense Observation Information


Sight Mucous membrane colour
Pale Hypoperfusion, anaemia
Pink Normal
Red Vasodilation, local congestion
Brick red Haemoconcentration
Blue Cyanosis
Surgical site
Tissue colour Cardiovascular and respiratory status
Bleeding: colour, intensity
Blood vessels: colour, turgescence
Thorax, bag of the breathing system, bellows of the ventilator
Rate Respiratory rate
Amplitude Tidal volume
Type Respiratory pattern
Eye
Position Depth of anaesthesia
Lacrimation
Pupil size
Degree of third eyelid protrusion
Movement
Spontaneous motor activity Depth of anaesthesia
Touch Pulse: femoralis, sublingual, metacarpal, metatarsal arteries
Tonus/amplitude Cardiovascular system: heart rate/rhythm, blood pressure
Rate/rhythm Autonomous response to noxious stimuli
Synchronicity with heart auscultation, ECG Depth of anaesthesia
pressure

Pulse amplitude:
Systolic–diastolic pressure difference

time
Tonus of the artery

Capillary refill time Cardiovascular status: peripheral perfusion


Palpebral/corneal reflex Depth of anaesthesia
Skin temperature Body temperature
Muscle relaxation
Relaxation of the jaw Muscle tone, depth of anaesthesia
Squeeze the bag after closing the exhaust valve
Mechanical integrity of the breathing system and respiratory tract Leak
Obstruction
Pulmonary compliance
Thoracic or extrathoracic resistance

7.2 Parameters that can be monitored during anaesthesia using the senses. (continues)

63
Chapter 7 Patient monitoring and monitoring equipment

Sense Observation Information


Hearing Abnormal noise
Mechanical integrity of the breathing system and respiratory tract Leak
Obstruction
Cardiac auscultation: stethoscope, oesophageal stethoscope
Rate Cardiac system: heart rate, rhythm, heart integrity
Rhythm Autonomic responses to noxious stimuli
Synchronicity with pulse, ECG Depth of anaesthesia
Murmur
Pulmonary auscultation: stethoscope, oesophageal
stethoscope
Rate Respiratory system: rate, rhythm, integrity
Rhythm Autonomic rsponse to noxious stimuli
Lung sounds Depth of anaesthesia
Smell Presence of abnormal odour
Mechanical integrity of the breathing system and respiratory tract Leak in the presence of halogenated anaesthetic agent

7.2 (continued) Parameters that can be monitored during anaesthesia using the senses.

Parameter Light anaesthesia Adequate anaesthesia Deep anaesthesia


Eye position Central Rotated Central

Palpebral reflex + – –
Jaw tonicity + – –
Movement Possible Absent Absent
Cornea Moist Moist Dry
Heart rate Usually increased Usually decreased
Respiratory rate Usually increased Usually decreased
Haemodynamic and/or respiratory Yes Usually no No
variations following surgical
stimulation

Clinical signs of depth of anaesthesia. Eye position, palpebral reflex and jaw tonicity are the most accurate
7.3
signs to evaluate depth of anaesthesia. Note that heart rate and respiratory rate are influenced by many other
factors. These signs are not valid for anaesthesia performed with a dissociative anaesthetic agent.

Parameter Light anaesthesia Adequate anaesthesia Deep anaesthesia


Eye position Central Central Central
Palpebral reflex + + ±
Jaw tonicity + + ±
Movement Possible Possible Possible/absent
Cornea Moist Moist Moist/dry
Heart rate Usually increased Usually decreased
Respiratory rate Usually increased Usually decreased
Haemodynamic and/or respiratory variations Yes Usually no No
following surgical stimulation

Clinical signs of depth of anaesthesia with a dissociative anaesthetic agent. Note that heart rate and
7.4
respiratory rate are influenced by many other factors.

64
Chapter 7 Patient monitoring and monitoring equipment

Initially, the person in charge of anaesthesia needs


to organize his/her workplace in such a way that the
collection of clinical information necessary to assess
depth of anaesthesia and the quality of the vital signs
is easy: the concept of ergonomics. The aim of this
concept is to be able to perform, from one location, a
quick and complete assessment of all monitored para-
meters and machine function (Figure 7.5). The layout
of the workplace will help concentration on the per-
ception and integration of measurements from the
patient, anaesthetic equipment and monitoring de-
vices. This includes:

• Clinical parameters, such as eye position, 7.6 The anaesthesia ‘screen’.


palpebral reflex, jaw tone, mucous membrane
colour, capillary refill time, sublingual pulse,
chest movement, skin temperature and, during 7.7
surgery, blood colour and bleeding intensity (a) Oesophageal
• Indirect clinical parameters, e.g. movement of stethoscope.
the reservoir bag (b) Tip of the
• Parameters collected from technical/electronic oesophageal
devices, e.g. wave forms and numeric displays stethoscope.
on the screen
• Monitoring of the anaesthetic machine: inlet
pressure (when available), flowmeters, vaporizer,
breathing system and function of any other
accessories like ventilators or syringe pumps.
(a)

Technical
monitors

Fluids

Patient

Anaesthetic (b)
machine
• Act as a source for retrospective analysis,
7.5 Ergonomics in the workplace. allowing further improvement of anaesthesia
strategies
• Represent a medicolegal document, attesting
The use of an anaesthesia ‘screen’ (Figure 7.6), so the presence of an active monitoring effort of the
the patient’s head can be seen, and the placement of patient during anaesthesia.
an oesophageal stethoscope (Figure 7.7), to make
cardiac and lung auscultation easier, are strongly rec-
ommended. However, the veterinary surgeon remains Monitoring of parameters that
the main monitor; he/she is the collector of parameters, reflect the integration of several
the central analyser and is responsible for all decisions physiological functions
(i.e. he/she is the ‘pilot’ on board). The limit of this ‘main
monitor’ is that its quality (and thus safety) is depend- Capnometry and capnography
ent on qualifications, knowledge and experience. Capnometry is the measurement and numerical dis-
A record of anaesthesia, even in a simple form, play of the carbon dioxide concentration in the respi-
must be completed. This will: ratory gas during the whole respiratory cycle
(inspiration and expiration). When a continuous
• Aid appreciation of trends in observed and graphical display (screen or paper) is available, the
measured parameters measurement produces typical curves (a capnogram)
• Act as a reference for future anaesthesia in an and the technique (instrument) is often referred to as
individual animal capnography (capnograph).

65
Chapter 7 Patient monitoring and monitoring equipment

The measuring principle relies on absorption of The capnogram


infrared light by carbon dioxide molecules. Capno- The production of the capnogram is based on the fact
meters are available as either sidestream or main- that carbon dioxide is produced in cells as a result of
stream infrared analysers. Sidestream capnometers metabolism and then carried by the circulation to the
require a sampling line connected to the airway and lungs, where it is removed by alveolar ventilation.
continuously sample the respiratory gases (Figure There are four distinct phases of the capnogram
7.8a). Samples are then analysed in the measuring (Figure 7.9a):
chamber of the instrument. With mainstream ana-
lysers, miniaturization allows placement of the meas- • Phase I is the inspiratory baseline. This phase
uring chamber directly in the airway and the represents the analysis of the gas mixture
measurement signal is generated here (Figure 7.8b). inspired by the patient for carbon dioxide. The

(b)
(a) Close-up of a connector for sampling gas
7.8
(sidestream capnograph). (b) Adaptor and sensor
(a) of a mainstream capnograph.

7.9
ICO2 ETCO2
(a) The normal capnogram.
• Anaesthetic machine • Metabolism Increased Decreased (b) Cardiogenic oscillations.
ETCO2 ↑ ETCO2 ↓ A–B: exhalation of carbon dioxide-
• Cardiovascular Cardiac output ↑ Cardiac output ↓
free gas contained in deadspace
ETCO2 ↑ ETCO2 ↓ at the beginning of expiration.
B–C: respiratory upstroke,
• Respiratory Hyperventilation Hypoventilation representing the emptying of
ETCO2 ↓ ETCO2 ↑ connecting airways and the
beginning of emptying of alveoli.
C–D: expiratory (alveolar) plateau,
representing emptying of alveoli.
D D: end-tidal carbon dioxide level –
C the best approximation of alveolar
carbon dioxide level.
D–E: inspiratory downstroke, as the
patient begins to inhale fresh gas.
E–A: continuing inspiration, where
carbon dioxide remains at zero.
ETCO2 = End-tidal carbon dioxide;
A B E ICO2 = Inspired carbon dioxide.

(a) I II III IV

D
C

A B E

(b)

66
Chapter 7 Patient monitoring and monitoring equipment

baseline should have a value of zero, otherwise where P bar is barometric pressure in mmHg.
the patient is rebreathing carbon dioxide Capnometers can often perform the conversion be-
• Phase II is the expiratory upstroke. This cause of a barometric pressure sensor in the device.
represents arrival at the sampling site of carbon End-tidal carbon dioxide concentrations between 5 and
dioxide from the alveoli and the mixing with gas 6% (35–45 mmHg; 4.6–6 kPa) are considered nor-
present in the conducting airways. It is usually mal in anaesthetized animals.
very steep
• Phase III is the expiratory (alveolar) plateau, Interpreting the capnogram
which represents pure alveolar gas (of The approach for proper interpretation of capno-
emptying alveoli). Due to uneven emptying of graphic information is initially to check if a normal
alveoli, the slope continues to rise gradually capnographic curve is present and then to note the
during the expiratory pause. The peak of this numerical value of end-tidal carbon dioxide. One
exhaled carbon dioxide is called the end-tidal must acknowledge the possible contribution of
carbon dioxide metabolism, circulation and ventilation to the pro-
• Phase IV is the inspiratory downstroke. This is duction of this number. When two functions are
the beginning of the inhalation and the carbon stable, capnometry monitors closely the third func-
dioxide curve falls steeply to zero. tion, e.g. if metabolism and circulation are stable,
the capnometer monitors ventilation. Conversely, if
During anaesthesia, with long expiratory times,
metabolism and ventilation are stable, the state of
phase III can show cardiogenic oscillations caused
the circulation will be more closely reflected. Addi-
by movement of the heart and this is considered
tionally, capnography will provide information
normal (Figure 7.9b).
about airway patency, technical faults and ade-
The capnometer usually displays respiratory rate,
quacy of gas flow in non-rebreathing systems.
the value of end-tidal carbon dioxide and sometimes
Capnography has become standard care for intra-
the value for inspiratory carbon dioxide. If a capno-
operative monitoring of ventilatory efficiency and
graph is used, the presence of a normal shape of the
thus decreases the need for invasive arterial blood
capnographic curve indicates that the number dis-
gas analysis.
played for end-tidal carbon dioxide is likely to repre-
sent a true end-expiratory sample. When grossly
normal ventilation/perfusion conditions in the lungs Increased ETCO2: Increases in end-tidal carbon
exist, the end-tidal carbon dioxide reflects the partial dioxide concentrations (Figure 7.11a) may be due to
pressure of carbon dioxide in arterial blood (PaCO2; impaired alveolar ventilation (anaesthetic-induced res-
Figure 7.10). Hence capnometry offers a continuous, piratory depression), increased metabolism (malignant
non-invasive way to reflect the partial pressure of car- hyperthermia or early sepsis), increased cardiac out-
bon dioxide in arterial blood, which is directly deter- put or the addition of carbon dioxide to the circulatory
mined by alveolar ventilation. system as a result of rebreathing carbon dioxide.
End-tidal carbon dioxide is displayed as a concen- In the latter case, increased inspired carbon dioxide
tration in volume percent (%) or as a partial pressure will be associated with increased end-tidal carbon
(in mmHg or kPa). Concentration can be converted dioxide (Figure 7.11b). Increased end-tidal carbon
into mmHg with the formula: dioxide is also observed during laparoscopy due to
absorption from the peritoneum of carbon dioxide used
PCO2 = (Pbar-47) x %CO2/100 to inflate the abdomen.

Capnography:
7.10
CO2 MONITOR the principle.
PACO2 = Alveolar carbon dioxide;
PaCO2 = Partial pressure of
carbon dioxide in arterial blood;
PvCO2 = Partial pressure of carbon
dioxide in venous blood.

DISPLAY UNIT

PACO2
PvCO2 PaCO2

67
Chapter 7 Patient monitoring and monitoring equipment

ICO2 ETCO2
• Anaesthetic machine • Metabolism Increased Decreased
ETCO2 ↑ ETCO2 ↓
• Cardiovascular Cardiac output ↑ Cardiac output ↓
ETCO2 ↑ ETCO2 ↓
• Respiratory Hyperventilation Hypoventilation
ETCO2 ↓ ETCO2 ↑

C D

A B E

(a)

ICO2 ETCO2
• Anaesthetic machine • Metabolism Increased Decreased
ETCO2 ↑ ETCO2 ↓
• Cardiovascular Cardiac output ↑ Cardiac output ↓
ETCO2 ↑ ETCO2 ↓
• Respiratory Hyperventilation Hypoventilation
ETCO2 ↓ ETCO2 ↑

C D

A B E

(b)
Capnographs showing (a) increased end-tidal carbon dioxide; (b) increased inspired carbon dioxide.
7.11
(continues)

68
Chapter 7 Patient monitoring and monitoring equipment

ICO2 ETCO2
• Anaesthetic machine • Metabolism Increased Decreased
ETCO2 ↑ ETCO2 ↓
• Cardiovascular Cardiac output ↑ Cardiac output ↓
ETCO2 ↑ ETCO2 ↓
• Respiratory Hyperventilation Hypoventilation
ETCO2 ↓ ETCO2 ↑

C D

A B E

(c)

7.11 (continued) Capnograph showing (c) decreased end-tidal carbon dioxide.

Decreased ETCO 2: Decreased end-tidal carbon Technical monitoring: Several technical aspects
dioxide concentrations (Figure 7.11c) may be due to of breathing systems, their function and connection
hyperventilation, low cardiac output (low blood vol- to the patient are also monitored by quantitative
ume delivered to the lungs) or pronounced hypother- (capnometer value) and qualitative analysis
mia. Rapidly falling end-tidal carbon dioxide, in the (capnogram morphology). The presence of inspired
presence of respiratory movements and absence of carbon dioxide (rebreathing) (see Figure 7.11b) can
hyperventilation, is a good indicator of failing circula- be due to soda lime exhaustion; an incompetent
tion and cardiac arrest. Absent end-tidal carbon diox- expiratory valve on a circle system allowing exhaled
ide can indicate respiratory arrest (no alveolar carbon dioxide to be re-inhaled (even with normal
ventilation), cardiac arrest (no circulation) or techni- function of soda lime); insufficient gas flow in a non-
cal problems (see below). Rapid diagnosis of cardiac rebreathing system. An abnormal capnogram may
arrest increases the chance of a successful outcome be due to:
for cardiopulmonary resuscitation. End-tidal carbon
dioxide values during resuscitation and cardiac mas- • A dislodged endotracheal tube
sage reflect efficiency of lung perfusion, and in hu- • A misplaced endotracheal tube (oesophageal
man medicine have been considered to have intubation)
prognostic value for successful restoration of sponta- • An obstructed endotracheal tube or airway
neous circulation. (endotracheal tube cuff hernia; Figure 7.12)

Capnograph showing
7.12
increased resistance at
D expiration.
C

A B E

69
Chapter 7 Patient monitoring and monitoring equipment

• A leak around the endotracheal tube cuff damaged, leading to erroneous end-tidal or inspired
(Figure 7.13) carbon dioxide values. Sidestream technology offers
• Disconnection of the endotracheal tube from the other (less important) possibilities like sampling
breathing system. directly from a nostril in non-intubated patients. Main-
stream technology can be used only in the presence
of an endotracheal tube, but copes better with rapid
Capnometry versus capnography: The drawback respiratory rates and very small tidal volumes.
of a capnometer compared with a capnograph is
that the former lacks the capnogram and, therefore, Pulse oximetry
cannot provide a qualitative analysis and precise The pulse oximeter is a device that allows non-
diagnosis from the morphological changes of exhaled invasive measurements of the saturation of haemo-
carbon dioxide. globin with oxygen (SpO2); in addition, it displays
When ventilation/perfusion mismatch in the lungs heart rate. The principle of measurement is as fol-
becomes important, the end-tidal carbon dioxide lows: a probe, which features a transmitter and re-
underestimates PaCO2 due to an increase in alveolar ceiver of red and infrared light, is positioned to
deadspace (e.g. low cardiac output, pulmonary transilluminate a pulsatile arteriolar bed. In small
thromboembolism). animal patients, it is usually placed on the tongue,
With sidestream devices, artefactually low end-tidal prepuce or other unpigmented area (Figure 7.14).
carbon dioxide is seen: Computer software analyses the total absorption
• When concomitant aspiration of ambient air of the light and detects the pulsating component of
occurs anywhere between sampling site and the absorption that originates from cyclic arrival of
measuring chamber arterial blood in the tissue. The absorption of infra-
• With rapid respiratory rates and very low tidal red and red light by the pulsatile component is meas-
volumes ured differentially. Because oxyhaemoglobin and
• When sampling at the level of fresh gas delivery reduced haemoglobin absorb more infrared and
in non-rebreathing systems. The sampling line red light, respectively (Figure 7.15), a ratio is calcu-
can become obstructed by condensation drops lated that corresponds to a percentage of haemo-
or aspiration of any fluids. globin saturated with oxygen (Figure 7.16). Because
maximal total light absorption corresponds to a
With mainstream capnographs, the measuring pulse, the pulse oximeter also provides a figure
chamber in the airway adaptor can become dirty or for pulse rate.

Capnograph showing
7.13
leakage at the level of
D endotracheal tube and/or breathing
C system.

A B E

(a) (b)

7.14 Sensor of a pulse oximeter. (a) On the tongue. (b) On a non-pigmented toe.

70
Chapter 7 Patient monitoring and monitoring equipment

Any general cause of poor peripheral perfusion,


such as vasoconstriction associated with shock,
Red Infrared hypothermia or use of alpha-2 agonists, causes erro-
660 nm 940 nm
Absorption (log scale)

neously low values or failure to measure. Local


hypoperfusion can also be induced by progressive
oxyHb
pressure from the clip holding the probe; changing
the position of the clip and the probe every so often
temporarily solves the situation.
reduced Hb Other factors can also interfere with pulse oximetry
readings:
• The presence of abnormal haemoglobin will
affect measurement. For instance, in the
Wavelength
presence of carboxyhaemoglobin SpO2 will be
Light absorption of reduced haemoglobin and overestimated
7.15
oxyhaemoglobin. • Interference from ambient light will reduce the
quality of the signal received
7.16 • Motion or shivering, use of electrosurgical
Relationship equipment and mucosal or skin pigmentation.
between the ratio
of red to infrared Some pulse oximeters display a photoplethysmo-
SpO2 (%)

light absorption graphic waveform. The presence of a regular and dis-


and the saturation tinct waveform similar to an arterial pressure waveform
of haemoglobin confirms the validity of the displayed SpO2 values (Fig-
with oxygen ure 7.18). Moreover, this can be supported by a dis-
(SpO2).
played heart rate equivalent to the true heart rate.
However, the amplitude of the waveform does not
Ratio red/infrared light absorption
necessarily reflect the quality of the signal because
the gain can be automatically adapted.
The threshold of hypoxaemia occurs at an arterial
oxygen tension of 60 mmHg (8 kPa). As shown on
the dissociation curve of haemoglobin (Figure 7.17),
this threshold corresponds to a saturation of 90%.
During anaesthesia, therefore, oxygen saturation must (a)
be kept above 90%. When oxygen is supplied (or even
in a healthy patient breathing room air), a saturation
closer to 100% should be expected. The usual preci-
sion of measurement of a pulse oximeter is ± 2%. It is
important to note that SpO2 only indicates that the
patient is receiving enough oxygen; SpO2 can be nor- (b)
mal in hypoventilating patients when the inspired oxy- (a) Pulse oximeter: display of SpO2, heart rate
gen fraction is increased (which normally should be 7.18
and waveform. (b) Display of capnography
the case during anaesthesia). In these circumstances, and pulse oximetry.
SpO2 gives no information about adequacy of ventila-
tion. Conversely, when a patient hypoventilates whilst Capnography and pulse oximetry
breathing room air, SpO2 will fall. Capnography and pulse oximetry (see Figure 7.18b)
provide complementary information. It has been
100 shown that the proper interpretation of information
received by capnography and pulse oximetry, used
80
together, have the potential to prevent 93% of the com-
plications during anaesthesia in human medicine
Hb saturation (%)

(Tinker, 1989).
60

Blood gas analysis


40
Blood gas analysers measure directly the pH and the
partial pressures of oxygen and carbon dioxide in a
blood sample. They also calculate derived variables,
20
such as plasma bicarbonate concentration (HCO3–),
base excess and saturation of haemoglobin with oxy-
0 gen (SaO2). In traditional analysers, the blood sample
0 20 40 60 80 100 600
is aspirated into a circuit bringing the blood in contact
PaO2 (mmHg)
with different electrodes to perform the measurements.
These machines are expensive and require careful
7.17 Haemoglobin–oxygen dissociation curve.
maintenance. Recently, portable and even handheld

71
Chapter 7 Patient monitoring and monitoring equipment

blood gas analysers have become available, allow- haematoma or bruise formation at the site of sampling
ing blood gas analysis to be performed in remote lo- can occur. In particular, pressure should be applied to
cations. These devices use disposable cartridges, the site for several minutes after sample collection to
which need only a minimal amount of blood to per- allow the vessel to seal. If it is likely that multiple sam-
form the measurement. They measure blood gas para- ples will be required, insertion of an arterial catheter
meters but offer the option to measure additional should be considered; this is often done in combina-
parameters, e.g. electrolytes, glucose and lactate. tion with invasive blood pressure measurement.
Blood gas analysis is the ‘gold standard’ method Blood gas analysis is useful in the management
for evaluation of gas exchange. It provides invaluable of critically ill patients and patients with ventilatory
information about the oxygenation, ventilation and problems. Specific treatment can be planned from the
acid–base status of the patient. Interpretation of the results of initial blood gas analysis, and the response
values for pH, PCO2 and HCO3–, coupled to the clinical to this treatment can be monitored by taking further
history of the patient, allows identification of respira- blood gas samples. The necessity of repeated blood
tory and metabolic acidosis and alkalosis, as well as sampling in anaesthesia is diminished when non-
ongoing compensatory mechanisms. The detailed invasive monitors of oxygenation and ventilation are used
interpretation of blood gas measurements can be simultaneously, e.g. pulse oximetry and capnometry.
complex as all the results are interrelated, and inter-
ested readers are advised to consult standard textbooks
on the subject. Figure 7.19 summarizes basic inter-
pretation of blood gas analysis. Cardiovascular monitoring
If the ability of the lungs to oxygenate blood is to
be assessed, sampling of arterial blood is mandatory. Heart rate indicator
The value of PaO2 (partial pressure of oxygen in arte- Heart rate indicators exist as separate devices but
rial blood) on its own is not meaningful to evaluate more often heart rate counting is a feature of an
oxygenation if the inspired oxygen fraction (FiO2) is electrocardiographic monitoring system. The heart
not known. Classical indices of oxygenation are: rate is deduced using detection of the QRS complex.
The number of complexes is counted over a certain
• The ratio PaO2/FiO2 (normal: >300) time period and calculated as beats per minute. Algo-
• The alveolar–arterial difference for PO2 rithms are based on human use and in veterinary
((A-a)PO2) where PAO2 = ((Pbar -47) x FiO2) – patients regularly cause the display of a double heart
1.2PaCO2 (normal: 5–10 mmHg if FiO2 = 0.21, rate when relatively high T waves are erroneously
100 mmHg if FiO2 = 1.0). detected as QRS complexes. The displayed heart rate
The ‘gold standard’ to evaluate the efficiency of must be regularly double-checked by clinical exami-
ventilation is PaCO2. The assessment of venous blood nation. There is no information about the pumping
samples does not provide straightforward information function of the heart (see Electrocardiography, below)
about oxygenation, but it does allow satisfactory moni- and consequently such heart rate indicators can give
toring of the ability to remove carbon dioxide by ven- a false feeling of security.
tilation and the acid–base status.
The classic blood gas machine uses samples of Blood pressure
heparinized blood. Failure to use the correct samples
causes clotting within the circuit. To obtain accurate Arterial blood pressure
results, blood samples should be withdrawn anaerobi- Arterial blood pressure is one of the most useful meas-
cally (no air bubbles) to avoid actual gas tensions be- ures of cardiovascular function available to the vet-
ing modified by gas exchange with room air. Analysis erinary surgeon. Two methods of measuring blood
is best done promptly, but storing blood samples for pressure can be used: invasive (direct) and non-inva-
less than 1 hour in iced water is acceptable. Taking an sive (indirect). Invasive blood pressure measurement
arterial blood sample is relatively easy in larger ani- is less straightforward to carry out in practice condi-
mals, but rather difficult in very small ones. Femoral tions but, on the other hand, non-invasive blood pres-
and dorsal metatarsal arteries are often used. Without sure measurement might not fulfil the expectations of
careful attention to technique, complications such as reliability and accuracy.

Parameter Normal values Decrease Increase


pH 7.35–7.45 Acidaemia Alkalaemia
PaCO2 35–45 mmHg Respiratory component: alkalosis Respiratory component: acidosis
HCO3– 22–26 mmol/l Metabolic component: acidosis Metabolic component: alkalosis
PaO 2 If oxygen level in inspired gas (FiO2) = Hypoxaemia: <60 mmHg (8kPa) Due to increased FiO2 or increased
21%: 80–100 mmHg (10.7–13.3 kPa) atmospheric pressure
SaO 2 95–100% Hypoxaemia: < 90 %

7.19 Basic interpretation of blood gas analysis.

72
Chapter 7 Patient monitoring and monitoring equipment

Invasive blood pressure measurement: Invasive the circumference of the limb. Wider cuffs will lead to
blood pressure measurement is considered the ‘gold underestimation of the true pressure, while narrow
standard’ technique. A catheter must be placed in a cuffs will tend to overestimate. Most cuffs have a mark
peripheral artery. In the dog, the dorsal metatarsal that should be placed directly over the artery.
artery is most commonly used, but the femoral Using the oscillometric method, returning blood
artery or the palmar artery can also be used. In the flow is detected by pulsatile pressure changes in the
cat, the femoral artery is the most common site. The cuff itself. The pressure at which pulses in the cuff
catheter is connected to a pressure transducer via appear is taken as systolic blood pressure. As the cuff
non-compliant tubing filled with heparinized saline. is further deflated, diastolic and mean blood pressure
Pure mechanical pressure transducers have been are measured. Oscillotonometric devices record mean
mostly abandoned for electronic ones connected blood pressure as the pressure at which the pulses in
to the monitoring device. The transducer must be the cuff are maximal and diastolic pressure as the
‘zeroed’ to ambient air at the level of the right atrium. pressure below which they disappear. The option of
The transducer–monitor combination gives a continu- automated function provides non-invasive blood pres-
ous reading of blood pressure and shows the pres- sure measurements at adjustable time intervals. Heart
sure waveform. Systolic and diastolic pressures are rate is also usually displayed.
taken as the cyclic maximum and minimum pres- Using the Doppler method, placement of an ultra-
sures, respectively. Mean pressure is automatically sound probe over an artery distal to the cuff is neces-
calculated. To allow trouble-free continuous moni- sary. The emitted ultrasound is reflected by the moving
toring (avoiding clotting in the arterial line), the set- blood and a frequency shift is converted to an audible
up is combined with a pressurized continuous flush signal. The pressure at which blood flow recom-
system that does not influence the accuracy of the mences, characterized by a typical ‘whoosh’ sound,
pressure reading (Figure 7.20). Repeated arterial is taken as systolic blood pressure. Diastolic pressure
blood sampling (blood gas analysis) is easy when is more difficult to determine. No automatic function
an arterial catheter is in place. is available but leaving the probe in place with a de-
flated cuff allows one to hear the blood flow continu-
Non-invasive blood pressure measurement: There ously, hence turning it into a continuous monitor.
are two methods generally used to measure blood Changes in the pitch of the tone suggest changing
pressure non-invasively: the oscillometric method and haemodynamics. The Doppler technique can be ap-
the Doppler method. Both methods are based on the plied to any size of animal, including cold-blooded
occlusion of blood flow to an extremity by the inflation animals, and is relatively inexpensive. It requires more
of a cuff and detection of reappearance of blood flow operator experience than automated oscillometric
during deflation. Any of the limbs or the tail can be measurement and the noise it produces can be dis-
used. The ideal cuff width is usually quoted as 40% of turbing to some unless headphones are used.

Set-up for
7.20
invasive blood
Monitor pressure measurement.

Pressurized
continuous flush
system

Electronic
pressure
transducer

Catheter in
peripheral
artery

73
Chapter 7 Patient monitoring and monitoring equipment

Significance of information: There is a wide range notch indicates vasodilation, steepness of upstroke
in blood pressure encountered in anaesthetized dogs represents strength of contraction.
and cats. Ranges commonly seen are between 90 Invasive blood pressure monitoring is less practi-
and 120 mmHg for systolic, between 55 and 90 mmHg cal than non-invasive techniques in the clinical set-
for diastolic, and between 60 and 120 mm Hg for mean ting. It requires more expensive equipment, skills in
blood pressure. placement of a catheter (especially in smaller patients)
Generally, arterial blood pressure gives an indica- and correct calibration. There is a risk of haematoma
tion of the adequacy of cardiovascular function. The formation, particularly after withdrawal of the catheter,
systolic blood pressure is determined by a combina- and a risk of infection.
tion of peripheral vascular resistance, stroke volume Individual blood pressure values obtained by non-
and intravascular volume, whereas diastolic blood pres- invasive methods must always be regarded cautiously.
sure primarily arises from peripheral vascular resist- Oscillometric blood pressure measurements tend to
ance. Measurement of systolic and diastolic blood be less accurate or the measuring devices regularly
pressures allows estimation of pulse pressure, which cease to function during hypotension and peripheral
is the difference between systolic and diastolic pres- vasoconstriction (especially in small patients <5 kg
sures (pulse quality). Mean blood pressure is an im- bodyweight). The Doppler technique, in principle, only
portant factor in relation to general tissue perfusion. indicates systolic pressure; in cats the Doppler tech-
Blood flow to the major organs of the body is nique underestimates true systolic pressure and tends
autoregulated across a range of mean blood pressures to be more closely related to mean blood pressure.
from about 60 mmHg to about 120 mmHg. When mean
blood pressure falls below this range, blood flow to the Central venous pressure
major organs is jeopardized. This results in inadequate Central venous pressure (CVP) represents the bal-
oxygen delivery and accumulation of lactic acid, lead- ance between the filling of the central venous reser-
ing to acidosis. On the other hand, it must be realized voir and the pumping ability of the heart. CVP is
that a normal mean blood pressure does not neces- measured in the vena cava within the thoracic cavity
sarily equate to good overall tissue perfusion. If a nor- via a long intravenous catheter. The classical ap-
mal blood pressure is generated by excessive proach is to position the catheter, via the jugular vein,
vasoconstriction, perfusion of tissue might be insuffi- in the thoracic portion of the cranial vena cava. Al-
cient. Acute hypertension can have deleterious effects ternatively, the femoral or saphenous approach can
on the central nervous system and the lungs. be used to position the catheter in the thoracic por-
Changes in blood pressure and heart rate can be tion of the caudal vena cava. CVP can be measured
seen with inadequate anaesthetic depth, anaesthetic using the technique described for invasive arterial
agent overdose, hypovolaemia and overhydration (see blood pressure. The zero reference point is the level
also Chapters 19 and 29). Analysis of the arterial of the right atrium. Landmarks are the manubrium
waveform with invasive blood pressure measurement (in lateral recumbency) and the scapulohumeral joint
allows semiqualitative estimation of ventricular (in patients in dorsal recumbency). Instead of elec-
contractility, hypovolaemia, cardiac output and the tronic transducers, a cheap saline manometer is of-
role of peripheral resistance, e.g. a low dichrotic ten used (Figure 7.21).

Set-up for CVP


7.21
measurement.
Note that the zero point on
the saline manometer is set
at the level of the
manubrium in the laterally
recumbent patient.
cm H2O

74
Chapter 7 Patient monitoring and monitoring equipment

A change of CVP over a short time period, in the improve electrical contact and is preferred to alcohol;
absence of a changed fluid load suggests changes in the latter tends to evaporate and the associated loss
cardiac function. A low or falling CVP suggests the of electrical contact can be the cause of artefacts.
presence of hypovolaemia. High or rising CVP can Placing adhesive electrode patches on the skin ne-
be due to fluid overload or to failing heart function. cessitates shaving, but this can be avoided when they
The administration of a fluid bolus, which is followed are attached to the pads of the paws. Patches can be
by a transient rise in CVP, suggests hypovolaemia reused when fixed with self-adhesive bandage.
but adequate cardiac function. Subsequent fluid A normal ECG trace is shown in Figure 7.22. The
therapy can be guided by CVP control. CVP can be a three-electrode system is effective for arrhythmia
valuable aid in differentiating low arterial blood pres- analysis, but in human medicine is considered insuf-
sure due to hypovolaemia from low arterial blood pres- ficient for detection of myocardial ischaemia (analysis
sure due to heart failure. The normal range is quite of the ST segment). When the surgical site does not
variable (3–8 cmH2O), so single values have limited allow normal electrode placement, a base–apex deri-
meaning. Conversely, repetitive measurements and vation can be used: one electrode is placed dorsal to
determination of trends provide valuable information. the base of the heart, another ventrally near the apex
CVP measurement is not a routine procedure dur- and the third peripherally. Transoesophageal ECG,
ing anaesthesia. In many cases, CVP measurement using a special oesophageal probe or stethoscope
may have already begun before anaesthesia. It is of equipped with ECG electrodes, is also a very con-
particular use in critical patients who need extensive venient method of monitoring cardiac rhythm.
and prolonged fluid therapy and/or are being treated The ECG:
for haemodynamic instability. During intermittent posi-
tive pressure ventilation (IPPV), CVP is influenced by • Provides information limited to electrical activity
the elevated intrathoracic pressure, especially when of the heart. It does not provide any information
positive end-expiratory pressure is used, and CVP about pumping function and haemodynamic
measurements have to be corrected accordingly. consequences. An extreme example is
electromechanical dissociation, where the ECG
Electrocardiography remains normal but the heart is not beating
The electrocardiogram (ECG) was the first monitor- • Allows quick diagnosis of the type of cardiac
ing to be made compulsory in human medicine and it arrest (asystole, ventricular fibrillation,
may be the most frequently used monitor in veteri- electromechanical dissociation) and is the only
nary anaesthesia. way to detect and analyse any other
In veterinary practice, the three electrode system arrhythmias.
is almost exclusively used: the electrical activity of the
heart is measured between two electrodes, with the Examples of common ECG abnormalities during
third electrode acting as the ground electrode. They anaesthesia are shown in Figure 7.23.
are normally located on left and right forelimbs and Common causes of artefacts in ECG monitoring
the left hindlimb. Perfect electrical contact with the are poor electrical contact of the electrodes (insuffi-
skin is essential to obtain reliable and prolonged ECG cient gel, dirty crocodile clips), detachment of elec-
tracings. This contact can be achieved with adhesive trodes, 50/60 Hz interference, electrical interference
electrode patches, crocodile clips or transcutaneous from electrosurgical devices and movement of the
needles. In general, electrode gel is recommended to patient.

7.22 Normal
R ECG.

P T

Q
S

75
Chapter 7 Patient monitoring and monitoring equipment

7.23
Abnormal ECGs,
with normal
ECG trace
superimposed
in grey.
(a) Ventricular
extrasystole.
(b) Atrioventricular
block, second
degree.

(a)

P P

(b)

Respiratory monitoring Airway pressure gauges


It is highly desirable to have an airway pressure gauge
Respiratory rate counters (Figure 7.24) in an anaesthetic breathing system. This
One type of respiratory rate monitor uses a connec- is a simple mechanical aneroid (or Bourdon-type)
tor attached to the end of the endotracheal tube. This manometer with a scale varying from –10 to +100
connector is equipped with a thermistor, which de- cmH2O. This instrument helps the performance of a
tects the movement of respiratory gases (expired gas pre-anaesthetic quantitative leak test of the anaes-
is warmer than inspired gas). These monitors can emit thetic machine and breathing system. The monitoring
an audible tone synchronous with the detection of of the typical pressure cycles of IPPV will serve as a
movement of gas. They can also display the time in guide during manual ventilation and for adjustment of
seconds since the last breath, or the respiratory fre- automatic ventilators. The absence of pressure swings
quency. They produce an alarm in case of apnoea indicates major leaks or disconnection of patient and
(‘apnoea detector’). However, the respiratory rate dis- anaesthetic apparatus. Airway pressure monitoring will
played should not be used as a safe indicator of the detect dangerously high airway pressure, e.g. due to
efficiency of ventilation, because no information is maladjustment of ventilator settings or when the ex-
provided about tidal volume. A ‘normal’ displayed res- haust valve remains inadvertently closed during spon-
piratory rate, and even tachypnoea, can be associ- taneous ventilation. If the delivered tidal volume is
ated with hypoventilation. In addition, if instrument known, the ratio of tidal volume to airway pressure is
sensitivity is not properly adjusted, it is possible for an indicator of compliance of the respiratory system
these monitors to produce artefactual acoustic sig- when IPPV is applied. Moreover, the level of positive
nals and display erroneous information. end-expiratory pressure can be checked.

76
Chapter 7 Patient monitoring and monitoring equipment

7.24 Oxygen analyser


The measurement of oxygen concentration in inspired
Airway gas (FiO2) is compulsory in human anaesthesia. When
pressure
gauge.
several gases are used (oxygen, nitrous oxide, air),
monitoring of the inspired oxygen fraction is a basic
method to prevent delivery of a hypoxic gas mixture
and hypoxaemia. The measurement of FiO2 also al-
lows safe use of low-flow anaesthesia techniques (see
Chapter 5). This measurement is often included in
respiratory gas analysis performed by multiparameter
anaesthetic monitors and different physical principles
are used. A popular simple stand-alone device is
based on a chemical reaction in a fuel cell. The lifespan
of this relatively costly cell is limited and depends on
Spirometers the amount and time of oxygen measurement. A mini-
A simple method to evaluate the efficiency of ventila- mum of 30% oxygen in the inspired gas should be
tion is the use of a mechanical respirometer (Figure provided during anaesthesia of healthy patients.
7.25). This instrument measures the volume of expired
gases and is often fixed in the expiratory limb of a
breathing system. Alternatively, it can be connected Temperature monitoring
directly to an endotracheal tube or facemask. A
respirometer indicates tidal volume and thus the minute The continuous measurement of body temperature is
volume must be calculated using a stopwatch. Some performed by a thermistor or a thermocouple mounted
respirometers have a built-in timer and automatically on an appropriate probe. The probe can be placed in
display the minute volume (minute volume (MV) is the the oesophagus, nasopharynx or rectum.
product of respiratory rate (RR) per minute and tidal Perioperative monitoring of body temperature is
volume (Vt) of the patient (MV = Vt x RR/minute)). important to detect hypothermia, which commonly
develops during anaesthesia. Among other reasons,
this is due to depression of hypothalamic thermo-
7.25
regulation and heat loss from surgical exposure of body
Spirometer. cavities. Hypothermia leads to reduced cellular meta-
bolism, decreased anaesthetic requirements (danger
of overdosage), delayed recovery, bradycardia and
increased morbidity. It is important to take preventive
measures, such as insulation of the surgical table, use
of a circulating warm water mattress, warm blankets
and heating pads. However, if not well controlled, there
is a potential risk of inducing iatrogenic hyperthermia.
Hyperthermia is also seen occasionally during low-flow
anaesthesia in a warm environment, and during
sepsis and malignant hyperthermia.
Electronic spirometers exist in different forms for
both research and clinical purposes and are based Inhalational anaesthetic agent
on different physical principles. However, they are monitoring
expensive and rarely found in veterinary practice.
Some devices do not provide any more information Monitors that continuously measure the concentra-
than the mechanical spirometer. tion of inhaled anaesthetic agents in the breathing
Spirometry is reliable and shows whether or not system are available. In many of these devices, infra-
ventilation is within acceptable limits (minute volume red analysis is used, and is often combined with the
100–300 ml/kg/minute, tidal volume 7–15 ml/kg). How- measurement of oxygen and carbon dioxide.
ever, acceptable values for ventilation are not neces- Measurement allows control of the dosage of
sarily associated with acceptable gas exchange; volatile agent administered to the patient and corre-
because the contribution of deadspace volume (about lation of this with its clinical effects. The principle is
one third in normal circumstances) remains unknown, that alveolar (end-tidal) concentration of an agent
the alveolar part of the minute volume (which deter- reflects its arterial and cerebral concentration; this
mines PaCO2) remains speculative. In this respect, value can be compared with the documented mini-
capnometry and blood gas analysis are most useful. mum alveolar concentration (MAC) value of the agent
Some spirometers allow advanced monitoring of (see Chapter 14). However, a very large number of
ventilation because they use physical principles that factors influence MAC value (e.g. individual varia-
permit measurement of respiratory flow and airway tion, sedative and analgesic drugs) and measured
pressure, calculate compliance of the respiratory values are indicative of the dosage of the agent but
system, and provide graphic presentation of the pres- not of the depth of anaesthesia. Clinical examina-
sure–flow and pressure–volume relationships in the tion remains the key factor in the assessment of an-
form of loops. aesthetic depth.

77
Chapter 7 Patient monitoring and monitoring equipment

If properly calibrated, these monitors can be used • The anaesthetic techniques used
to check the accuracy of the output of vaporizers. In • The need to monitor several cases
this case, it is imperative that sampling be done at the simultaneously
outlet of the vaporizer and not in the breathing system. • The need for post-anaesthetic monitoring.

The choice of which monitoring equipment to buy


Monitoring depth of anaesthesia depends on:

In clinical veterinary anaesthesia, the most accurate • The organization of the practice: if nobody is
method to assess the depth of anaesthesia is based available to monitor anaesthesia, more
on collection and interpretation of relevant clinical equipment will be necessary to alert the
parameters (see Figure 7.3). The use of neuro- surgeon, but this method has clear limits. It is far
muscular-blocking agents as part of a balanced an- preferable to have a trained person to monitor
aesthesia protocol eliminates many of these important anaesthesia, rather than rely exclusively on the
clinical parameters (see Chapter 15). In human medi- use of electronic monitors. It cannot be
cine, this has occasionally led to difficulties in estima- overemphasized that the purchase and use of
tion of anaesthetic depth, resulting in awareness monitors is of no use if:
during anaesthesia, or undesired recall afterwards. – Clinical monitoring is not mastered
For this reason, there is an increased use in human – Artefactual information cannot be identified
clinical practice of electronic devices to evaluate depth – The physiological meaning of the
of anaesthesia. These devices are based on compu- information provided is not understood and
terized analysis of the electroencephalogram (EEG) the right corrective steps cannot
(bispectral analysis) or EEG reaction to auditory stimu- consequently be undertaken. Adequate
lation (middle latency auditory evoked potentials). In training of staff is therefore of utmost
veterinary anaesthesia, this kind of monitoring is importance, and the practice must be
reserved for research purposes. organized in a way that allows someone
with adequate training to assess
anaesthetized patients on a regular basis
Monitoring of the neuromuscular • The budget: cost of the equipment is a limiting
junction factor. Also, to invest in expensive monitoring
equipment which is not properly understood (and
For information on monitoring of the neuromuscular
whose information cannot be correctly
junction, see Chapter 15.
interpreted) will be of no use in either improving
quality of anaesthetic management or reducing
the incidence of anaesthetic mishaps
Additional monitoring • Patient condition (ASA status): patients with
When a urinary catheter is in place, the urine output increased anaesthetic risk will require more
can easily be measured by collecting the urine in complete monitoring of their vital functions so
a bag (normal 1–2 ml/kg/h). A normal urine output that rapid detection of abnormalities is possible
reflects correct cardiovascular status and normal and correct assessment and therapeutic
renal perfusion. Renal perfusion can be reduced due decisions can be made quickly
to cardiovascular failure, hypovolaemia and shock • The chosen anaesthetic technique:
(see Chapter 23). Urine output is part of the fluid – Capnography in non-intubated patients is
balance evaluation comparing fluid input during an- difficult, whereas pulse oximetry can be used
aesthesia (intravenous infusions) and fluid losses – High doses of alpha-2 agonists in intubated
(blood, urine, transudates). patients may increase the likelihood of
In some circumstances, the perioperative moni- unreliable pulse oximeter function, but
toring of blood glucose concentration (e.g. anaes- capnography will still be reliable
thesia of diabetic patients and those with insulinoma) – When nitrous oxide is used, or when using
or electrolytes, haematocrit and total protein (unsta- low-flow techniques, there is a potential risk
ble critical patients with emergency surgery) may of administration of a hypoxic gas mixture. In
be necessary. this case, the use of an oxygen analyser in
the breathing system and/or a pulse oximeter
is recommended (SpO2)
Guidelines for monitoring strategy in • The case load and the types of surgery or
veterinary anaesthesia diagnostic examinations: it is not obligatory to
use all available monitoring equipment for very
When organizing and purchasing anaesthetic moni- short and simple procedures on healthy patients.
toring equipment, there are several factors to bear A pulse oximeter in non-intubated patients or a
in mind: capnograph in intubated patients can suffice.
Generally, the greater the disturbance in
• Staff numbers and qualifications homeostasis that is anticipated, the greater the
• Costs versus benefits number of physiological functions that must be
• Case load and nature of those cases monitored. Invasive surgery (e.g. thoracotomy),

78
Chapter 7 Patient monitoring and monitoring equipment

or surgery associated with specific Conclusion


pathophysiology or problems (e.g. bleeding) will
need appropriate monitoring. It is clear that monitoring equipment can never re-
place a suitably qualified person: it has not been
For practices performing invasive procedures or proven that monitoring equipment necessarily de-
dealing with emergency cases and postoperative care, creases anaesthetic risk. As complexity increases, the
capnography, pulse oximetry, ECG, non-invasive chances for errors to occur due to artefacts and mis-
blood pressure measurement (with the option of in- interpretations increase accordingly. The preparation
vasive blood pressure measurement) and temper- and set-up of monitoring equipment can be time-
ature monitoring are highly desirable. Nevertheless, consuming and the devices are usually expensive;
regular clinical assessment of depth of anaesthesia such considerations could lead one to question the
remains necessary because no technical device can true benefits.
replace this evaluation (Figure 7.26). One important aspect is that the whole concept of
monitoring provides continuous information about the
patient, and thus becomes a kind of teacher support-
ing the person in charge of anaesthesia. This teacher
demonstrates the physiological consequences of ad-
ministered drugs and dosages, and of the complex
interactions between anaesthesia, disease and sur-
gical manipulation. In this way one can become a
better veterinary surgeon, with more insight into what
one is actually ‘doing’ to the patient.
In conclusion, ‘when applied appropriately, oper-
ated properly, and interpreted correctly, however,
monitors afford the patient the best possible outcome’
(Lawrence, 2005).

Multiparametric monitor and anaesthesia


7.26
‘screen’. References and further reading
The use of normal anaesthetic and monitoring Arbous MD, Grobbee DE, van Kleef JW et al. (2001) Mortality associated
equipment inside a magnetic resonance unit is not with anaesthesia: a qualitative analysis to identify risk factors.
possible. The magnetic field attracts all ferrous metal Anaesthesia 56, 1141–1153
Bhavani-Shankar K http://www.capnography.com (a useful, free, website
and electronic functions are disturbed. Special moni- with animated capnographs)
toring equipment that can remain inside the room is Brodbelt D, Brearley J, Young L, Wood J and Pfeiffer D (2005)
Anaesthetic-related mortality risks in small animals in the UK. AVA
available, but is very expensive. For respiratory gas Spring Meeting, Rimini, Italy 20–23 April 2005
analysis, including capnography, a simple alternative Lawrence JP (2005) Advances and new insights in monitoring. Thoracic
solution is to sample via a very long line while the Surgery Clinics 15, 55–77
Tinker JH, Dull DL, Caplan RA, Ward RJ and Cheney FW (1989) Role
monitor is stationed outside the room. For ECG and of monitoring devices in prevention of anesthetic mishaps: a closed
pulse oximetry, special sensors are necessary. claims analysis. Anesthesiology 71, 541–546

79

Das könnte Ihnen auch gefallen