Key facts
J Frequency range from 9 kHz to 3.6 GHz or 8 GHz
J Highsensitivity(<–141dBm(1Hz),withpreamplifier
<–161 dBm (1 Hz ))
J Low measurement uncertainty (<1 dB)
Easy-to-replace battery J Measurement functions for all important measurement
tasks related to the startup and maintenance of trans-
mitter systems
J Internal tracking generator and VSWR bridge with built-in
DC voltage supply (bias)
J Two-port network analyzer
J Easy-to-replace Li-ion battery for up to 4.5 h of operation
J Rugged,splash-proofhousingforroughworkinthefield
J Easy handling due to low weight (3 kg with battery) and
easy-to-reach function keys
J Saving of measurement results on SD card
J LAN and USB interface for remote control and transfer of
measurement data
J R&S®FSH4View software for simple documentation of
measurement results
Easy operation
J Quick function selection via keypad and rotary knob
J Optimal reading of measurement results in any situation
J Setting of frequency via channel tables
J Easy-to-access, well-protected connectors
▷ page 10
Distance-to-fault measurements
The distance-to-fault, caused by a pinched cable or by
loose or corroded cable connections, is determined quickly
and precisely. The built-in threshold function ensures that
only true cable faults, i.e. faults that exceed a tolerance
limit, are displayed in a list. This considerably simplifies the
evaluation of the measurement.
1)
CDMA2000® is a registered trademark of the Telecommunications Industry
Association (TIA - USA).
of electromagnetic
J
using an isotropic antenna
J Determinationofelectricfieldstrengthinatransmission
channelwithdefinedbandwidth(channelpowermea-
fields surement)
applications in the J
J
Frequency and level measurements
Power measurements up to 18 GHz with the accuracy of
a power meter
lab or in service J
J
Measurementsonamplifiers,filters,etc.,usingvector
network analysis
Automated creation of test sequences by remote control
via LAN or USB
The fold-out stand turns the R&S®FSH into a desktop
analyzer for work in the lab or in service. Location of EMC problems
The R&S®HZ-15 near-field probes are used as diagnostic
tools for locating EMC problems, e.g. on circuit boards,
integrated circuits, cables and shieldings. The R&S®HZ-15
near-field probe set is adequate for emission measure-
ments from 30 MHz to 3 GHz. The R&S®HZ-16 preampli-
fier improves measurement sensitivity up to 3 GHz, with
approx. 20 dB gain and a noise figure of 4.5 dB. In combi-
nation with the R&S®FSH, the preamplifier and near-field
probe set are a cost-effective means of analyzing and
locating disturbance sources during development.
R&S®FSH4View software
10
Additional connectors (e.g. for LAN and USB) are protected by caps
Front view
Cursor keys
Standard functions
Application/ TDMA power Channel power Field strength Occupied Frequency Scalar Scalar
product measurements measurements measurements/ bandwidth settings via transmission reflection
measurements easurements
m channel table measurements easurements
m
with isotropic
antenna
R&S®FSH4/8,
√ √ √ √ √ – –
model .04/.08
R&S®FSH4/8,
√ √ √ √ √ √ –
model .14/.18
R&S®FSH4/8,
√ √ √ √ √ √ √
model .24/.28
Options
Application/ Distance-to-fault Vector reflection One-port Power measure- Directional power Remote control
product (DTF) measure- and transmission cable loss ments up to measurements up via LAN or USB
ments measurements easurements
m 8 GHz/18 GHz to 1 GHz/4 GHz
R&S®FSH4/8,
– – – R&S®FSH-Z1/-Z18 R&S®FSH-Z14/-Z44 R&S®FSH-K40
model .04/.08
R&S®FSH4/8,
– – – R&S®FSH-Z1/-Z18 R&S®FSH-Z14/-Z44 R&S®FSH-K40
model .14/.18
R&S®FSH4/8,
R&S®FSH-K41 R&S®FSH-K42 R&S®FSH-K42 R&S®FSH-Z1/-Z18 R&S®FSH-Z14/-Z44 R&S®FSH-K40
model .24/.28
Models
R&S®FSH model Frequency range Preamplifier Tracking generator Built-in VSWR bridge
12
1)
With carrying handle.
14
Regional contact
Europe, Africa, Middle East
+49 1805 12 42 42* or +49 89 4129 137 74
customersupport@rohde-schwarz.com
North America
1-888-TEST-RSA (1-888-837-8772)
customer.support@rsa.rohde-schwarz.com
Latin America
+1-410-910-7988
customersupport.la@rohde-schwarz.com
Asia/Pacific
+65 65 13 04 88
customersupport.asia@rohde-schwarz.com
ISO 9001
DQS REG. NO 1954 QM
ISO 14001
DQS REG. NO 1954 UM
Rohde & Schwarz GmbH & Co. KG
Mühldorfstraße 15 | 81671 München
Phone +49 89 41 290 | Fax +49 89 41 29 121 64
www.rohde-schwarz.com
*0.14 €/min within German wireline network; rates may vary in other
networks (wireline and mobile) and countries.