Collimator
XP

Replacement of Parts
Collimator

Replacing the focus-near Plates

Valid for collimators with 4-part focus-near plates


© Siemens, 2008

10092615
10092618
10847610
10569023
10092611
10092614
10092609
10092608
10092607
10092605
10092604
10092603
10092602
05662403
00468264
00468256
00468249
00468223
00468215

© Siemens, 2008
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.

Print No.: RX74-020.841.03.02.02 English


Replaces: RX74-020.841.03.01.02 Doc. Gen. Date: 03.12
CS PS AX R/F
n.a.

2008
2 Copyright / Version / Disclaimer

Copyright
1Copyright / Version / Disclaimer

“© Siemens, 2008“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA,
Siemens Healthcare Diagnostics Inc. - USA and/or Siemens Healthcare Diagnostics Prod-
ucts GmbH - Germany.

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.

Collimator RX74-020.841.03.02.02 Page 2 of 14 © Siemens, 2008


03.12 CS PS AX R/F
Replacing the focus-near Plates 3
2Replacing the focus-near Plates

General 0

Notes 0

Prior to performing the work steps, read the instructions thoroughly for each step.

Requirements 0

System knowledge is required for installation.

Required Materials 0

• Replacement Kit for the focus-near Plates


• Siemens standard tool kit
• If needed, system-specific Installation or Replacement Instructions.

General Safety Information 0

NOTE During execution of operations and tests the general safety


instructions for medical devices have to be regarded. See
Safety Information RAD-FLUORO XPRF-000.860.10.

Hazard Keys 0

The following Hazard Keys are part of this document


• hm_serv_Trained_Technician_only

WARNING System Operation


Loss of functionality: loss or deterioration of function,
assignment of image, data, objects, work flow and configura-
tion, inadequate user information.
Risk of injury!
Operational hazard due to lack of knowledge of the system,
e.g. safety measures have been ignored by the technician.
¹ Provide training to the technician.

• hm_serv_ProtectiveEarth

© Siemens, 2008 RX74-020.841.03.02.02 Page 3 of 14 Collimator


03.12 CS PS AX R/F
4 Replacing the focus-near Plates

WARNING System Operation


Danger from electric high voltage or line voltage, high volt-
age, line voltage, leakage current.
Risk of injury!
Problems with protective ground: protective ground is insuf-
ficient, disconnected, connecting screw loose.
¹ Setup Instructions, protective ground measurements
are described in the Startup Instructions and are per-
formed during system installation.

• hm_serv_DetOptLink_Eyeinjury

WARNING System Operation


Injuries from laser radiation
Risk of injury!
Technician looks into the laser of the optical link (Pix box/
docking station /detector).
¹ Service documentation, technician looks into the laser,
reason for CAUTION: Technician looks into the laser of
the optical link (Pix box/ docking station/ detector).
Possible consequences: Risk of eye injury. Remedy:
Do not look into the laser diodes of the optical link.

• hm_serv_CollimatorAdjust

WARNING FOV Position


Dose not as desired: ionizing radiation.
Risk of injury!
Incorrect collimator setting.
¹ Adjustment of the collimator plate: adjustment of the
collimator plate is designed so that the edge is visible
in the fluoro image.

• hm_serv_collimator_blades_checked

Collimator RX74-020.841.03.02.02 Page 4 of 14 © Siemens, 2008


03.12 CS PS AX R/F
Replacing the focus-near Plates 5

WARNING Dose control


Error in automatic change-over to middle measuring field.
Error in automatic change-over to middle measuring field
during collimation.
Generic Hazards: ionizing radiation, incorrect or inappropri-
ate output.
Unnecessary radiation, dose not as desired.
¹ Hint in service instructions: System has to be set up
that at least two collimator blades are visible in the
image.

• HM_F2_1G1U12M3_CS

WARNING Radiation release


Incorrect iris position respective incorrect collimation.
Patient exposed to radiation unnecessarily.
¹ Hint in service instructions: System has to be set up
that at least two collimator blades are visible in the
image.

• HM_F4_1G2U1M2_CS

WARNING Electrical safety


Working in energized environment.
Danger of electric shock or cardiac arrhythmia.
¹ Safety informations in service documents.

• HM_F5_6G2U1M2_CS

WARNING Mechanical safety


Assembly error
Injury of the patient during operation
¹ In the replacement of parts document the CSE will be
instructed about right assembly.

• hm_serv_replacement_parts

© Siemens, 2008 RX74-020.841.03.02.02 Page 5 of 14 Collimator


03.12 CS PS AX R/F
6 Replacing the focus-near Plates

WARNING System Operation


Loss of functionality. Generic Hazards: loss or deterioration
of function, assignment of image, data, objects, work flow
and configuration, inadequate user information.
Operational hazard due to lack of knowledge of the system,
e.g. safety measures have been ignored by the technician.
¹ In the "Replacement of Parts" document the CSE will
be instructed about the correct assembly.

Collimator RX74-020.841.03.02.02 Page 6 of 14 © Siemens, 2008


03.12 CS PS AX R/F
Replacing the focus-near Plates 7

Preparations 2.1

• System OFF.
• Remove the collimator from the system.

© Siemens, 2008 RX74-020.841.03.02.02 Page 7 of 14 Collimator


03.12 CS PS AX R/F
8 Replacing the focus-near Plates

Replacing the Plates 2.2

NOTE The plates and curved rings can be operated manually while
power OFF. This will make replacing the plates easier.

Removal 0

• Carefully turn the plates by the curved rings into the position shown in the illustration
(Fig. 1 / p. 8) (makes it easier to remove the plates).
• Remove the 4 mounting screws.
• Remove the defective plates. Make sure that the defective plates are complete after
removing them, and if applicable, remove broken parts from the collimator.

Fig. 1: Focus-near plates, mounting


Check the curved rings (visual check without removal). If they are okay, they can remain in
the collimator and can continue to be used. In this case, continue with the point “Inserting
the Plates“ (Inserting the Plates / p. 10).

Collimator RX74-020.841.03.02.02 Page 8 of 14 © Siemens, 2008


03.12 CS PS AX R/F
Replacing the focus-near Plates 9

Inserting the Curved Rings 0

• Position the drive bracket in the middle of the opening.

Fig. 2: Drive bracket


• Insert the curved rings into each another.
• Insert the studs for the curved rings into the slots of the drive bracket. The drive studs
for the inner ring must be inserted in the direction of the control panel or of the control
board.

Fig. 3: Curved rings

© Siemens, 2008 RX74-020.841.03.02.02 Page 9 of 14 Collimator


03.12 CS PS AX R/F
10 Replacing the focus-near Plates

Inserting the Plates 0

• Assemble the plate unit. All screws and spring must be inserted. The springs must be
pressed flush against the white support plate.

Fig. 4: Assembly of the plate unit


• Insert the plate unit into the collimator. The slot of the top pair of plates must be parallel
to the operating front or to the control board.

Fig. 5: Installation position of the plates


• Secure the plate unit with the 4 screws.

Collimator RX74-020.841.03.02.02 Page 10 of 14 © Siemens, 2008


03.12 CS PS AX R/F
Replacing the focus-near Plates 11

Adjusting the Plates 0

Check the position of the plates using the gauge with the collimator closed (mechanical
stop) and, if needed, readjust by carefully bending the metal plate part.
The adjustment must be performed for both pairs of plates.

Fig. 6: Adjusting the pairs of plates

© Siemens, 2008 RX74-020.841.03.02.02 Page 11 of 14 Collimator


03.12 CS PS AX R/F
12 Replacing the focus-near Plates

Final Work Steps 2.3

• Install the collimator in the system.


• Switch the system ON.
• Generate several test exposures at different collimator settings (minimum/maximum
collimator opening, as well as with the height plate opened to the minimum and with the
width plates opened to the maximum, and vice versa).
• If additional settings are required on the collimator, use the current version of the docu-
ment “Notes on Replacing, Repairing and Adjusting a Collimator”, RX74-020.842.01

Collimator RX74-020.841.03.02.02 Page 12 of 14 © Siemens, 2008


03.12 CS PS AX R/F
Replacing the focus-near Plates 13

Changes to Previous Version 2.4

Chapter Change
General Remarks “General Safety Information“ and “Hazard
Keys“ added.
Miscellaneous Editorial revisions.

© Siemens, 2008 RX74-020.841.03.02.02 Page 13 of 14 Collimator


03.12 CS PS AX R/F
14 Replacing the focus-near Plates

List of Hazard IDs 2.5

There are no Hazard IDs in this document.

Collimator RX74-020.841.03.02.02 Page 14 of 14 © Siemens, 2008