Beruflich Dokumente
Kultur Dokumente
To cite this article: Svante Söderström, Anders Eklöf, Anders Brahme (1999) Aspects on the
Optimal Photon Beam Energy for Radiation Therapy, Acta Oncologica, 38:2, 179-187, DOI:
10.1080/028418699431591
Correspondence to: Dr Svante Söderström, Department of Medical Radiation Physics, Karolinska Institutet
and Stockholm University, P.O. Box 260, S-171 76 Stockholm, Sweden. Tel: +46 8 51 70 00 00.
Fax: +46 8 34 35 25. E-mail: svante@radfys.ks.se
The selection of optimal photon beam energy is investigated both for realistic clinical bremsstrahlung beams and for monoenergetic
photon beams. The photon energies covered in this investigation range from 60Co to bremsstrahlung and monoenergetic beams with
maximum energies up to 50 MeV. One head and neck tumor and an advanced cervix tumor are investigated and the influence of beam
direction is considered. It is shown that the use of optimized intensity modulated photon beams significantly reduces the need of beam
energy selection. The most suitable single accelerator potential will generally be in the range 6 – 15 MV for both superficially located and
deep-seated targets, provided intensity-modulated dose delivery is employed. It is also shown that a narrow penumbra region of a photon
beam ideally should contain low-energy photons ( 54 MV), whereas the gross tumor volume, particularly when deep-seated targets are
concerned, should be irradiated by high-energy photons. The regions where low photon energies are most beneficial are where organs at
risk are laterally close to the target volume. The situation is completely changed when uniform or wedged beams are used. The selection
of optimal beam energy then becomes a very important task in line with the experience from traditional treatment techniques. However,
even with a large number of uniform beam portals, the treatment outcome is substantially lower than with a few optimized
intensity-modulated beams.
Recei6ed 24 April 1998
Accepted 1 October 1998
During the past 30 years, optimization of external beam The selection and optimization of the photon beam
radiation therapy has been focused mainly on different energy for external beam radiation therapy have been
methods to improve the dose delivery by using multiple somewhat neglected during this period. However, during
uniform or wedge-shaped beams. The goal of the opti- recent years some attention has been devoted to the inter-
mization process has often been to deliver a dose to the esting and important subject of finding the ideal energy for
target volume as close as possible to the prescribed one radiation therapy (16 – 18). Most of these investigations
or, to achieve as small a deviation as possible from the used the fairly unimportant integral dose concept, or in
dose constraints for the tumor and the organs at risk. modern terminology the mean energy imparted, as a
For this purpose a large number of techniques and al- quantifier of the merits of treatment plans.
gorithms have been developed that optimize the beam In our study we use a more clinically relevant endpoint,
weights and/or the wedge angle with respect to a number namely the probability to control the tumor growth with-
of dose constraints or an objective function (1–3). More out inducing severe damage in irradiated normal tissues.
recently, arbitrary fluence profile optimizations have been This probability was calculated based on clinically estab-
studied by several authors (4–8). Another recent develop- lished dose-response relations for tumors and normal tis-
ment can be seen in stationary and dynamic multi-leaf sues (19 – 21). Today the beam energy is generally selected
collimation (9–13) where the goal is to deliver truly 3-D using more or less standardized protocols for each tumor
conformal irradiation by modulation of the photon site. Sometimes the standard energy selection is modified
fluence and/or electron fluence profiles. Dose optimiza- on a trial and error basis to increase the dose level and
tion with scanned photon and electron beams is a poten- uniformity in the target volume or to reduce the dose to
tially flexible and practical technique (12, 14). organs at risk. This can be useful when uniform or wedged
Furthermore, different methods for fan beam intensity beams are used but the role of beam energy in intensity-
modulation during arc therapy have recently been pro- modulated dose delivery has not yet been carefully investi-
posed by Mackie and co-workers (15). gated. In light of the new developments in the field of
external beam radiation therapy optimization and consid- side of Fig. 1. In this case the normal tissue stroma and the
ering the new possibilities to modulate the incoming beam spinal cord are the principal organs at risk. The gross
profiles, the best selection of beam energy and energy tumor and the local lymphatic spread are regarded as
distribution should also be carefully investigated. separate biological structures and are associated with dif-
One of the on-going developments of radiation therapy ferent biological responses (20). The gross tumor volume is
equipment to better treat deep-seated tumors is that to- assumed to consist of normal tissue infiltrated to 50% by
wards higher photon beam energies. The highest accelera- clonogenic tumor cells and the region of presumed
tor energy available today, the Racetrack microtron, lymphatic spread is assumed to contain a uniform clono-
covers the whole range from 2 MeV up to 50 MV (22). A genic tumor cell burden of 10%. When uniform, as well as
higher photon beam energy reduces the dose to shallow intensity-modulated, photon beams are used, this target is
organs but increases the exit dose and moves the maxi- treated with a three-field technique using one frontal field
mum dose deep into the patient (almost 7 cm at 50 MV) and two oblique lateral fields.
due to the increasing energy of the secondary electrons. At An advanced cervix cancer including locally involved
the same time a reduced difference in energy deposition in lymph nodes was taken as an example of a deep-seated
bone and soft tissue is seen at medium energies (4–25 MV) tumor. The geometry of the cervix target is shown on the
due to the domination of the Compton process. At higher left-hand side of Fig. 1. In this case the organs at risk are
energies the pair production process slowly raises the the bladder, the rectum, the hip joints, the small bowel,
energy deposition again. The value of using high-energy and the surrounding normal tissue stroma. The gross
Downloaded by [47.29.227.33] at 00:50 28 November 2017
&
pEi (r)= hEi (r, z− z%)mEie − mEi z% dz% [1]
A summary of the photon beam characteristics for the
presently used photon pencil beams is given in Table 2. It
is clear from previously published data (32) and Table 2
where the vector r =(r, z) and hEi (r, z−z%) is a monoen- that the build-up properties and the electron and photon
ergetic Monte Carlo calculated point energy deposition scatter are the main properties influencing the properties of
kernel with the energy Ei (31) and mEie − mEi z%is the depth the photon beams.
dependence of the TERMA of a monoenergetic and
monodirectional photon beam of energy Ei (32). The Optimization algorithm
bremsstrahlung spectra for different accelerator potentials The principal problem of forward radiation therapy plan-
have previously been calculated by semi-empirical methods ning can be formulated in the form of an integral equation
(33). These spectra have been used to generate polyen- that expresses the resultant dose distribution in the patient
182 S. Söderström et al. Acta Oncologica 38 (1999)
relative to the target volume and are thus in need of tissue involved with larger targets, the optimal energy rapidly
photon beams with a narrower penumbra. This is particu- increases beyond 2 – 2.5 cm sized tumors.
larly true for targets treated with conformal therapy but it
is also to some extent valid for few field treatment tech- niques. When a small number of fields are used, only a
small portion of the circumference of a spherical target is
irradiated tangentially, where a narrow penumbra is re-
quired. If the target, on the other hand, is more cube-like,
as much as four-sixths of the circumference may be irradi-
ated by one single field. The importance of a narrow
penumbra is thus exaggerated by the present somewhat
unclinical cubical target volume. In a more realistic clinical
situation a larger number of beam portals would reason-
ably be required to obtain a narrow penumbra at the
complete tumor – normal tissue interface. However, as the
number of beams increases, the need for a deep dose
maximum is relaxed and the relative importance of a
sharper penumbra increases if further improvements in the
dose distribution are required. This is a well-known princi-
ple utilized in, for instance, stereotactic radiosurgery with
the Gamma Knife® by Elekta.
A narrow penumbra enables a high dose delivery to the
target volume without causing severe injury to the sur-
rounding normal tissues. In addition, the treatment of
small tumors produces less normal tissue damage because
of the small beam portals and normal tissue volumes
irradiated. As the volume of the target increases, the ratio
of tumor to the normal tissue interface volume decreases
and the need to deliver large doses to bulky tumors
increases. To be able to deliver high doses to large vol-
umes, the dose maximum of the photon beam should
preferably be reached inside the target volume. Since only
one photon beam energy level is allowed during the
Fig. 6. The one-dimensional phase space of complication-free
1
present optimization process, the probability of achieving
tumor control, P + (V), for five different photon beam energies:
60
Co, 6 MV, 10 MV, 25 MV and 50 MV. (a) The phase space for
complication-free tumor control will decrease with increas-
the cervix target geometry. (b) The phase space for the head and ing target volume. The main reason for this is the increas-
neck target geometry. ing risk of normal tissue damage and thus an inability to
186 S. Söderström et al. Acta Oncologica 38 (1999)
modulated photon beams of a single fixed energy spectrum 8. Niemierko A. A random search algorithm (RONSC) for
are used, accelerator potentials higher than 25 MV are not optimization of radiation therapy with both physical and
generally required for treatment optimization. The mar- biological end points and constraints. Int J Radiat Oncol Biol
Phys 1992; 23: 89 – 98.
ginally decreasing probability of achieving a complication- 9. Källman P, Lind BK, Eklöf A, Brahme A. Shaping of arbi-
free cure above that energy is due to the larger lateral trary dose distributions by dynamic multileaf collimation.
spread of the intrinsic electron penumbra of the photon Phys Med Biol 1988; 33: 1291 – 300.
pencil beams. However, high photon energies are more 10. Convery DJ, Rosenbloom ME. The generation of intensity-
modulated fields for conformal radiotherapy by dynamic col-
desirable in the central part of deep-seated target volumes
limation. Phys Med Biol 1992; 37: 1359 – 74.
to improve the sparing of superficial tissues. Therefore, in 11. Bortfeld T, Kahler DL, Waldron TJ, Boyer AL. X-ray field
a strict optimization the beam should contain a central compensation with multileaf collimators. Int J Radiat Oncol
high-energy component with a low-energy component at Biol Phys 1994; 28: 723 – 30.
the periphery in order to sharpen the penumbra. 12. Gustafsson A, Lind BK, Svensson R, Brahme A. Simulta-
neous optimization of dynamic multileaf collimation and
When also considering the method of intensity-modu- scanning patterns or compensation filters using a general
lated dose delivery using scanned photon beams, it may be pencil beam algorithm. Med Phys 1995; 22: 1141 – 56.
desirable to use a very high photon energy with a thin 13. Svensson R, Källman P, Brahme A. An analytical solution
target so that the bremsstrahlung lobe becomes as narrow for the dynamic control of multileaf collimators. Phys Med
Biol 1994; 39: 37 – 61.
as possible (38). This interesting option for future dose
14. Lind BK, Brahme A. Optimization of radiation therapy dose
delivery optimization has been left largely untested in this distributions with scanned photon beams. Proc of the 9th
study. ICCR, The Use of Computers in Radiation Therapy. 1987;
Single-field techniques may require completely different 235 – 9.
beam energy selections due to the absence of cooperating 15. Mackie TR, Holmes TW, Swerdloff S, et al. Tomotherapy: A
new concept for the delivery of conformal radiotherapy. Med
beams in the target volume. In the periphery of deep-
Phys 1993; 20: 1709 – 19.
seated targets, lower photon energies are preferred in order 16. Hanks GE, Diamond JJ, Kramer S. The need for complex
to gain the steepest possible lateral dose fall-off to spare technology in radiation oncology, correlations of facility char-
normal tissues close to the target volume. For superficial acteristics and structure with outcome. Cancer 1985; 55:
targets, very low photon energies are the most advanta- 2198 – 201.
17. Laughlin JS, Mohan R, Kutcher GJ. Choice of optimum
geous, if intensity modulation is not available, suggesting megavoltage for accelerators for photon beam treatment. Int
that electron beams, possibly combined with photon J Radiat Oncol Biol Phys 1986; 12: 1551 – 7.
beams, might be preferred in many configurations. How- 18. Das IJ, Kase KR. Higher energy: is it necessary, is it worth
ever, the need to combine electron and photon beams the cost for radiation oncology? Med Phys 1992; 19: 917 –25.
19. Ågren-Cronqvist A-K, Källman P, Brahme A. Determination
could be avoided if the optimization and delivery of inten-
of the relative seriality of a tissue from its response to
sity modulated beams are at hand. non-uniform dose delivery. In: Baier K, Baltas D, eds. Mod-
elling in clinical radiobiology. Freiburg, Germany: Albert-
ACKNOWLEDGEMENTS Lundwigs-University, 1997: 127 – 41.
20. Ågren A-K, Brahme A, Turesson I. Optimization of uncom-
The guidance given by Anders Gustafsson in the handling of the plicated control for head and neck tumors. Int J Radiat Oncol
generalized pencil beam algorithm is gratefully acknowledged. Biol Phys 1990; 19: 1077 – 85.
Acta Oncologica 38 (1999) Optimal photon beam energy 187
21. Källman P, Lind BK, Brahme A. An algorithm for maximiz- 31. Ahnesjö A, Andreo P, Brahme A. Calculation and application
ing the probability of complication free tumor control in of point spread functions for treatment planning with high
radiation therapy. Phys Med Biol 1992; 37: 871–90. energy photon beams. Acta Oncol 1987; 26: 49 – 56.
22. Brahme A. Design principles and clinical possibilities with a 32. Eklöf A, Ahnesjö A, Brahme A. Photon beam energy deposi-
new generation of radiation therapy equipment. Acta Oncol tion kernels for inverse radiotherapy planning. Acta Oncol
1987; 26: 403 – 12. 1990; 29: 447 – 54.
23. Swanson WP. Estimate of the risk in radiation therapy due to 33. Andreo P, Brahme A. Stopping power data for high-energy
unwanted neutrons. Med Phys 1980; 7: 141–4. photon beams. Phys Med Biol 1986; 31: 839 – 58.
24. Nath R, Epp ER, Laughlin JS, Swanson WP, Bond VP. 34. Brahme A, Svensson H. Radiation beam characteristics of a
Neutrons from high-energy x-ray medical accelerators: An 22 MeV microtron. Acta Radiol Oncol 1979; 18: 244 – 72.
estimate of risk to the radiotherapy patient. Med Phys 1984; 35. Sorcini BB, Hyödynmaa S, Brahme A. Quantification of
11: 231 – 41.
mean energy and photon contamination for accurate dosime-
25. Dunscombe PB, McLellan J, Huda W, Gillies J. Patient
try of high-energy electron beams. Phys Med Biol 1997; 42:
activation with a 25 M therapeutic x-ray beam. Br J Radiol
1849 – 74.
1988; 61: 843 – 6.
36. Brahme A. Optimisation of radiation therapy for cancer
26. Followill D, Geis P, Boyer A. Estimates of whole-body dose
patients. Forum 1994; 4: 569 – 84.
equivalent produced by beam intensity modulated conformal
37. Källman P, Ågren A, Brahme A. Tumor and normal tissue
therapy. Int J Radiat Oncol Biol Phys 1997; 38: 667– 72.
27. Gudowska I, Brahme A. Neutron radiation from high-energy response to fractionated non-uniform dose delivery. Int J
x-ray medical accelerators. Nukleonika 1996; 41: 105– 18. Radiat Biol 1992; 62: 249 – 62.
28. Gustafsson A, Lind BK, Brahme A. A generalized pencil 38. Söderström S, Gustafsson A, Brahme A. Few field radiation
beam algorithm for optimization of radiation therapy. Med therapy optimization in the phase space of complication free
Downloaded by [47.29.227.33] at 00:50 28 November 2017
Phys 1994; 21: 343–56. tumor control. Int J Imag Syst Technol 1995; 6: 91 – 103.
29. Söderström S, Brahme A. Selection of suitable beam orienta- 39. Svensson R, Brahme A. Effective source size, yield and beam
tions in radiation therapy using entropy and Fourier trans- profile from multi-layered bremsstrahlung targets. Phys Med
form measures. Phys Med Biol 1992; 37: 911–24. Biol 1996; 41: 1353 – 79.
30. Söderström S, Brahme A. Which is the most suitable number 40. Söderström S, Brahme A. Optimization of the dose delivery in
of beam portals in coplanar radiation therapy? Int J Radiat a few field techniques using radiobiological objective func-
Oncol Biol Phys 1995; 32: 151–9. tions. Med Phys 1993; 20: 1201 – 10.