How do I change the projection lamp on an HFA II-i?

ZEISS Quick Help: Humphrey® Field Analyzer II-i (HFA II-i)

This lamp is responsible for projecting the standard light stimulus. With the aid of color filters,
it is also used to create red and blue stimuli for color testing. If needed, you may order a new
lamp by contacting ZEISS and asking for a Projection Lamp (PN 2660021106082).

1. Lower the table and turn off the HFA II-i. To remove the top access panel, rotate it counterclockwise until you
align the lamp symbol on the cover with the raised dot that is molded into the top of the case (Figures 1 and 2).

Standing in front of the bowl opening, you will find that the projection lamp is located inside the open case
top, at the 12-o’clock position. Caution: Hot Lamp, allow lamp to cool completely (approximately five
minutes) before replacing lamp. Do not touch the two disk-shaped filters.

Figure 1 Figure 2

2. Remove the connector cable by pulling its connector straight upward (Figures 3 and 4).

Figure 3 Figure 4
How do I change the projection lamp on an HFA II-i | ZEISS Quick Help: Humphrey® Field Analyzer II-i

3. U
 se a screwdriver to slightly loosen (do not remove) the screw (Figure 5). Slide the wishbone-shaped plate
that holds the lamp assembly in place away from you (Figures 6 and 7).

Figure 5 Figure 6 Figure 7

4. R
 emove the old expended lamp assembly. Carefully open the new lamp. Do not touch the glass part of
the lamp with your fingers, as this will shorten the life expectancy of the lamp. If your finger touches
the glass portion of the lamp, wipe the lamp clean with a soft cloth. Insert the replacement lamp into
the housing. The notch in the base of the lamp housing lines up with the pin to the right of the assembly
(Figures 8 and 9).

Figure 8 Figure 9

5. Slide plate back into position (Figure 10), tighten screw and replace connector cable (Figures 11 and 12).

Figure 10 Figure 11 Figure 12

Page 2 of 3
How do I change the projection lamp on an HFA II-i | ZEISS Quick Help: Humphrey® Field Analyzer II-i

6. To replace the top access panel, insert the panel into the opening. As you do so, align the lamp
image on the lid with the raised dot on the underlying case. Rotate the panel clockwise until the lamp
symbol aligns with the open-circle symbol (Figure 13).

The contents of this reference guide may differ from the current status of the product in your country. Please contact your regional representative for more information. Subject to change in design and scope of delivery and as a result
of ongoing technical development. Humphrey and HFA are either trademarks or registered trademarks of Carl Zeiss Meditec, Inc. in the United States and/or other countries. © 2016 Carl Zeiss Meditec, Inc. All copyrights reserved.
Figure 13

Background Illumination Lamps


The lamps responsible for illuminating the bowl surface are a fluorescent type (not incandescent) and
have a long life expectancy. If you get a bowl illumination error, darken the room and restart the instrument
to see if this corrects the problem. If the bowl illumination error continues, do not attempt to replace the
lamp. These highly specialized lamps are to be replaced only by a Carl Zeiss Meditec Field Service Engineer.

Refer to the Humphrey Field Analyzer user manual Instructions for Use for safe and effective operation of the instrument.

SER.7972 Printed in the USA CZ-08/2016

0297

Carl Zeiss Meditec Inc. Carl Zeiss Meditec AG


5160 Hacienda Drive Goeschwitzer Str. 51-52
Dublin, CA 94568 07745 Jena
USA Germany
www.zeiss.com/med www.zeiss.com/med Page 3 of 3

Das könnte Ihnen auch gefallen