Sie sind auf Seite 1von 13

Journal of Chromatography A, 1142 (2007) 56–68

Optimization of simulated moving bed and column chromatography for a


plasmid DNA purification step and for a chiral separation
Galatea Paredes, Marco Mazzotti∗
Institute of Process Engineering, ETH Zurich, Sonneggstrasse 3,
CH-8092 Zurich, Switzerland

Available online 22 December 2006

Abstract
This work analyzes the performance of the SMB and the column chromatography processes for two different case studies: the first stage of
the plasmid DNA (pDNA) polishing, and the Tröger’s base enantiomer separation, in which the adsorption isotherms are linear and non-linear,
respectively. Simulation tools are used together with an optimization routine (Non-Sorting Genetic Algorithm (NSGA)) in order to find the optimum
operating conditions leading to maximum productivity and minimum solvent consumption; the optimum solution for each of the processes is a
curve on the productivity-solvent consumption plane, the so-called Pareto set. The comparison between the column and the SMB processes is based
on the relative position of the two Pareto sets calculated at equal conditions and for the same final purity and recovery of the target species. The
results show that SMB is superior to column chromatography in the two case studies investigated, i.e. in the case of the linear isotherm (pDNA),
the productivity gain is up to a factor two for a given value of the solvent consumption. Furthermore, the flexibility of the SMB operation is larger,
since the Pareto sets are flatter and they prolong into regions of the productivity-solvent consumption plane that are not accessible with the column
chromatography process.
© 2006 Elsevier B.V. All rights reserved.
Keywords: Simulated moving bed; Column chromatography; Multiobjective optimization; Pareto set; Plasmid DNA; Tröger’s base enantiomers

1. Introduction matography constitutes another process option, which is less


common in practice, and will not be considered in this work
Chromatographic separations can be carried out either in [6–8].
the batch operating mode, using one single column (or sev- Column chromatography is extensively used in biosepa-
eral columns working in parallel) packed with a particulate rations due to its simplicity when compared to multicolumn
or monolithic stationary phase (alternatively fluidized beds are processes. The batch mode allows for cleaning in place after
sometimes used), or in the continuous mode in a multicolumn every cycle and often a few columns are used in parallel. Several
setup adopting the simulated moving bed (SMB) process or studies have reported procedures to find optimum operating con-
one of its modifications (Varicol, Powerfeed . . .). In the later ditions or minimum production costs for several modes of the
case, several fixed bed columns are used to separate continu- column chromatography process [9–13]. SMB has not been used
ously a binary feed mixture into two fractions; multicompo- so far in the field of bioseparations; however, and despite the dif-
nent mixtures can also be split in two or more fractions con- ferences with respect to the traditional SMB applications, there
taining species having similar retention properties, as described is no fundamental barrier that prevents the use of SMB in biosep-
in the literature [1–4]. SMB was first used in the petrochem- arations. It is established that for tasks such as chiral separations
ical industry, and then successfully downscaled to be used in SMB allows in most cases to increase productivity and reduce
the sugar industry and then for the separation of enantiomers the solvent consumption, when compared to the same process
of chiral molecules [5]. Closed loop recycling in column chro- carried out by column chromatography; furthermore, due to its
working principle, SMB may achieve total recovery in situations
∗ Corresponding author. Tel.: +41 44 6322456; fax: +41 44 6321141. where the column chromatography recovery is limited by low se-
E-mail address: marco.mazzotti@ipe.mavt.ethz.ch (M. Mazzotti). lectivities. The challenges in bioseparations are related (a) to the

0021-9673/$ – see front matter © 2006 Elsevier B.V. All rights reserved.
doi:10.1016/j.chroma.2006.12.009
G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68 57

nature of the molecules and often to the high density of the sam-
ples, (b) to the need for cleaning in place, (c) to the use of aqueous
mobile phases (usually buffer solutions) and of gradients ob-
tained using modifiers, and (d) to the use of compressible chro-
matographic media. As to cleaning in place (CIP), it has recently
been proved also experimentally that this can be carried out in an
SMB unit [14,15]; the CIP operation has to be designed accord-
ingly, i.e. the required number of columns both in the cleaning
and in the regeneration section have to be used with proper flow-
rates.
In order to introduce the SMB technique in the bioprocess
field, it is necessary to provide quantitative data about the SMB
performance and to establish a fair comparison with the corre-
sponding column chromatography alternatives. The aim of this
paper is to make such a comparison at isocratic conditions, and
for the cases of two model systems of interest requiring two
different types of stationary phases: a compressible media, typ-
ically used in bioseparations, and a rigid media used in chiral
separations.
For a given application, the comparison between the opti- Fig. 1. Scheme of a Pareto set in the plane of the performance indicators (I1 ,
mum SMB operation and the optimum column chromatogra- I2 ); the process aims at maximizing I1 and minimizing I2 .
phy operation is the only way to draw fair conclusions about
the relative performance of the two [16–18]. The performance
of any chromatographic process is described by two variables, better than q. However, comparing the performance of point q
or performance indicators, related by a trade-off: productivity with a point located on the Pareto set else than p, e.g. point s, it is
and solvent consumption. These are to be maximized and mini- readily seen that q is worse than s, hence it is excluded from the
mized, respectively, while taking into account the product qual- optimum solution. This finding can be generalized by saying that
ity requirements (e.g. purities) and the technical limitations of any point above the Pareto set has worse performance than the
the operation (e.g. pressure drop). The constrained optimization points constituting the Pareto set. Comparing now points p and
of productivity and solvent consumption for a given chromato- s, both located on the Pareto set, one point cannot be considered
graphic process constitutes a multiobjective optimization prob- better than the other since I1 is better in s whereas I2 is better
lem, namely a two objective optimization. The decision variables in p. This is the characteristic feature of any point on the Pareto
of the optimization problem are typically related to the size of the set and it is generally referred to as a non-dominance condition;
unit and to the operating conditions of the process. Looking for a when moving along the Pareto set, each of the points (corre-
global optimum for such a multiobjective optimization problem sponding to a specific set of decision variables) is non-dominant
would require making a choice about the relative importance with respect to the others. The points located below the Pareto
of the different performance indicators; such a choice might be set are not feasible because the product quality requirements are
based on contingency and therefore not general enough. A better not fulfilled (e.g. the purity of the target species is too low).
approach, that will be followed here, is to search for a set of opti- The Pareto set may therefore be considered as the boundary of
mal solutions, which are equally good with respect to all perfor- the domain of feasibility. The final choice of optimum operat-
mance indicators at the same time in a sense that will be clarified ing conditions, i.e the choice of a specific point on the Pareto
shortly. This is the so-called Pareto set, which in the case of a two set, depends upon additional considerations such as equipment
objective optimization results to be a curve (in the productivity, costs, equipment availability, stability of the sample, costs of the
solvent consumption plane). To make this clear let us consider precursors, value of the products, etc. As shown schematically
a general two objective optimization problem in which I1 is to in Fig. 1, a Pareto curve exists only in a finite domain of the (I1 ,
be maximized and I2 has to be minimized, subject to a num- I2 ) plane, and its extension is a measure of the versatility of the
ber of constraints such that maximum pressure drop, minimum operation. Once the Pareto sets for SMB and column chromatog-
product purity, etc. The solution of this constrained optimization raphy are obtained, their relative location on the productivity-
problem is the Pareto set shown in Fig. 1. Comparing points p solvent consumption plane allows to draw conclusions about
and r located on the Pareto curve and above it, respectively, it is the convenience of choosing one technique with respect to the
readily seen that the performance is certainly worse in r, since other.
I1 is smaller and I2 is larger; this implies that decision variables The paper is organized as follows. In Section 2 the model sys-
leading to point p constitute better operating conditions than the tems are presented; then, the chromatographic model is shown
ones corresponding to r. Let us now consider point q which is together with the optimization procedure in Section 3. In Sec-
located above the Pareto set and to the right of point p. Contrary tions 4 and 5 the optimization problems are solved and the results
to r, point q is better than p when I1 is considered, but it is worse for each model system are presented. Finally, Section 6 reports
when I2 is considered; in principle, point p could be considered our conclusions.
58 G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68

2. Background on the model systems Table 1


Physical properties of the pDNA system
In this work, two chromatographic operations are consid- Component γi DLi (cm2 /s) ki (s−1 )
ered. The first corresponds to the first step of the plasmid pDNA 0 0.015 u[cm/s] 10
DNA (pDNA) polishing stage described in previous publica- RNA 0.95 0.045 u[cm/s] 0.35
tions [14,15]. At this early stage of the purification cascade, n 0.95 0.045 u[cm/s] 0.85
high recovery of pDNA is required, at purity levels which are
less critical, since these can be improved during the next purifi- this is a rather accepted statement when it comes to enantiomer
cation stages. The second case considered in this work is the separations; this paper aims also at proving this for the TB enan-
separation of the Tröger’s base (TB) enantiomers [19]. tiomers, which are frequently used as model system in various
The choice of the first model system is motivated by the in- fields, including chiral chromatography [19,22,23].
terest in the high added-value pDNA product that has grown
over the last years, as the research on gene therapy and vac- 2.1. Plasmid DNA purification
cination with nucleic acids moves forward. Direct delivery of
pDNA is generally preferred over viral based vector adminis- In the step considered the pDNA is purified from the so-called
tration, since the former minimizes the risk of viral infection. cleared lysate, which is a rather complex mixture of many com-
However, targeting of pDNA vectors is less efficient and, as a ponents, namely RNA, chromosomal DNA fragments, proteins
consequence, larger amounts of pharmaceutical-grade plasmid and endotoxins. At the conditions selected during the experi-
are required for treatment (probably in the order of milligrams). ments reported in previous works [14,15], the pDNA (i.e. the
Plasmids are produced in Escherichia Coli by fermentation and ccc pDNA together with its isoforms) is completely excluded
then harvested after cell disruption. In order to satisfy the strict from the pores and elutes from the column as a single peak. The
relevant regulations, the plasmid must be highly purified and in other species can access the pores, and are thus retarded and
its supercoiled form (ccc pDNA). The main impurities are RNA, elute from the column later. In order to model the chromato-
proteins, chromosomal DNA fragments and endotoxins. A large graphic behavior of this mixture, a rather simple model system
scale pDNA manufacturing process needs to be established in has been selected, which is still able to capture the key features
order to satisfy the demand of pharmaceutical-grade pDNA for of the retention behavior of the cleared lysate. This is done by
clinical trials and eventually for commercialization. The down- considering a three component mixture (labelled pDNA, RNA
stream processing of pDNA aims at recovering the plasmid at and n) whose components exhibit linear retention behavior char-
the purity levels required by regulatory agencies, such as the acterized by the following isotherm [14,15]:
Food and Drug Administration (FDA) and the European Agency
ni = γi ci = [εpi + (1 − εpi )Ki ]ci (1)
for the Evaluation of Medical Products (EMEA). It is unlikely
that only one purification step serves to achieve the high purity where ci and ni are the fluid and the solid phase concentrations of
specifications required by the authorities. Indeed, most often a species i, and εpi , Ki and γi are the accessible intraparticle poros-
combination of purification steps is considered in the literature, ity, the adsorption equilibrium constant and the linear retention
and these are grouped in three stages: primary recovery, inter- parameter (pseudo Henry constant) of species i, respectively.
mediate recovery and final purification (or polishing). The chro- The first term of the second equation, εpi ci , accounts for the re-
matographic step investigated here corresponds to the first step tardation of species i due to its characteristic pore accessibility,
of the polishing stage, and it is based on a size exclusion (SE) whereas the second term, (1 − εpi )Ki ci , accounts for the adsorp-
mechanism, which implies a linear chromatographic behavior. tion on the stationary phase; this last term is zero, i.e. Ki = 0,
At the conditions selected, there exists a baseline separation be- for a purely size exclusion based mechanism. Table 1 reports the
tween pDNA and the main impurity, i.e. RNA [14,15]. model parameters for the three species considered here, namely
In order to extend the comparison results of column chro- the γi values used in Eq. (1), and the axial dispersion and mass
matography and SMB to a non-linear chromatographic problem, transfer coefficients, estimated from experiments. From Eq. (1),
the separation of the Tröger’s base enantiomers was chosen as a the total concentration of species i is given by:
second case study. Over the last decades it has been recognized
ciT = εci + (1 − ε)ni = [ε + (1 − ε)γi ]ci (2)
that the two enantiomers of a chiral drug may have different bi-
ological activity; as a consequence, producing enantiomerically where ε is the intraparticle bed void fraction. Using the last
pure substances has become a crucial issue [5]. The enantiose- equation with γi given by Eq. (1) allows using a standard force
lective synthesis is not always possible or even not profitable, model for the chromatographic column where the total porosity
and the authorities responsible for drug administration (such as ε∗ is given by ε. Fig. 2 shows a simulated chromatogram of the
FDA) impose high purity requirements [20]; the industry is thus cleared lysate using the model parameters reported in Table 1.
faced with the challenge of searching innovative separation pro- The feed concentrations used in this study and the characteristics
cesses with high yields. Chromatography is one of the key oper- of the columns used in the simulations are reported in Table 2.
ations in this field [21], and the simulated moving bed technique The CIP and regeneration steps are not considered in this
is generally preferred due to the larger productivities achieved work. In order to include them in the performance analysis,
and the reduced solvent consumption obtained in most of the equivalent cleaning protocols for the column and the SMB pro-
cases, when compared to column chromatography. Nowadays cesses have to be defined. This is a critical issue in chromato-
G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68 59

Fig. 2. Simulated overloaded chromatogram of cleared lysate. The species are Fig. 3. Simulated chromatogram of the Tröger’s base enantiomers in microcrys-
labelled pDNA (target), and RNA and n (impurities). ta and tb are the break- talline cellulose triacetate (CTA) (ethanol) at 50 ◦ C. The species are labelled A
through time of pDNA and the elution time of the injected volume, respectively; and B. ta and tb are the breakthrough time of B and the elution time of the in-
the cycle time, tc , is defined as tb − ta . Collecting the pDNA between times ta jected volume, respectively; the cycle time, tc , is defined as tb − ta . Collecting
and tY guarantees the minimum recovery (Ymin ). A and B between the times ta and tX1 and between tX2 and tb guarantees the
purities of the two fractions, XAmin and XBmin .
Table 2
Concentration of the feed mixture, column characteristics and constant process
parameters used during the optimization problems 3. Modelling and optimization
pDNA system TB system
3.1. Chromatographic model
A (cm2 ) 0.785 1
ε∗ 0.32 0.59
The column chromatography and SMB simulations were car-
m4 −0.05 Decision variables
ried out using a detailed one-dimensional model that accounts
configuration 1-2-2-1 1-2-2-1
Closed loop Closed loop for convection and axial dispersion in the fluid phase, and mass
F =6
transfer through a linear driving force model, assuming that solid
cF (mg/ml) cA cnF = 0.1
diffusion is the rate limiting step. The following model equations
cBF = 6 F
cRNA = 1.5 consist of material balances, and are integrated using the method
F
cpDNA = 0.4 of lines:

∂ 2 ci ∂ci ∂ni ∂ci


εDLi = ε∗ + (1 − ε∗ ) +u (4)
graphic bio-separations, whose clarification is beyond the scope ∂z2 ∂t ∂t ∂z
of this work.
∂ni
= ap ki (n∗i − ni ) (5)
2.2. Tröger’s base enantiomer separation ∂t

n∗i = fi (c) (6)


The competitive adsorption of the (+)-TB and the (−)-TB
enantiomers in ethanol on microcrystalline cellulose triacetate
here t and z are the time and space coordinates, respectively,
at 50 ◦ C can be rather well described by the following binary
ε the bed void fraction, ε∗ the total bed void fraction, ap the
Langmuir isotherm, at least for the purpose of this study [24]:
specific surface of the adsorbent particles, u the superficial ve-
γi ci locity, and ki and DLi are the mass transfer coefficient and the
ni = , i = A, B (3)
1 + K A cA + K B cB axial dispersion coefficient of species i, respectively. The func-
where A and B are the more (the (+) form) and the less (the tion fi in Eq. (6) is the adsorption isotherm of component i,
(−) form) retained species, respectively. The values of the pa- yielding its adsorbed phase equilibrium concentration, n∗i . The
rameters used in this study are shown in Table 3; Fig. 3 shows space discretization is based on finite differences; 30 points per
the corresponding simulated chromatogram of the TB racemic centimeter of column were used after verifying that this gives a
mixture. The feed concentrations and the characteristics of the good enough numerical accuracy. Integration in time is carried
columns are reported in Table 2. out using a commercial integrator for stiff equations (LSODA),
as appropriate when simulating the steep concentration profiles
occurring in high performance liquid chromatography. In the
Table 3 case of the SMB, the equations described above are coupled with
Physical properties of the TB system
mass balances at the nodes, and solved in time by accounting for
Component γi Ki (ml/mg) DLi (cm2 /s) ki (s−1 ) the switching mechanism. A detailed description of the model
A 6.45 0.39 0.01 u[cm/s] 0.09 equations has been reported earlier [25]. The code is written in
B 2.18 0.065 0.01 u[cm/s] 0.15 Fortran.
60 G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68

3.2. Optimization chromatographic media stability, or the maximum pressure drop


in the unit, or the minimum switch time, in the SMB case.
3.2.1. Decision variables and performance indicators In the column chromatography case, it is necessary to fix the
The performance of SMB and of column chromatography is times at which the fraction collection of a given species starts and
measured in terms of productivity and solvent consumption, i.e. ends in order to calculate the purity and the recovery; purities
good performance means large productivities and low solvent are based on the relative amounts collected during this period. In
consumptions. These are the so-called performance indicators, the case of the SMB, the purities are calculated considering the
Il , of the optimization problem. The operating variables consti- average concentrations of the species over one cycle at cyclic
tute the set of decision variables, i.e. the parameters to be varied steady state (a cycle is the time interval in which one column
in order to optimize the process; these depend upon the nature of goes through all positions in an SMB unit, i.e. nc t ∗ , where nc is
the problem considered, e.g. the injected volume belongs to the the total number of columns).
set of decision variables in the column chromatography case, but The constraint on the pressure drop plays a role whenever the
not in the case of the SMB operation, where the feed is continu- chromatographic media is rigid; indeed, overcoming the max-
ous. In the particular case of SMB, the so-called flow-rate ratios, imum allowable pressure drop would cause destruction of the
mj , are dimensionless quantities grouping all the important op- beads constituting the bed and would lead to unacceptable degra-
erating parameters, and may be used as decision variables: dation of performance. The pressure drop, p, is calculated ap-
Qj t ∗ − Vε∗ plying Darcy’s law
mj = (7)
V (1 − ε∗ ) p
= φQ (8)
t∗
where Qj is the flow-rate in section j, the switching time, L
and V is the volume of the column. In some cases, the space of where L is the length of the bed, Q the fluid-flow rate and φ is
decision variables may be reduced by fixing some parameters or a parameter depending on permeability of the bed and viscosity
by restricting the optimization to a given region of this space; of the mobile phase that can be easily measured. In the case of
this is used to simplify the optimization problem in the case SMB, since the flow-rates vary along the unit, Darcy’s law is
where many decision variables are involved. applied to each section (Q = Qj with j =1, . . . , 4) and then the
The productivity (P), is defined as mass of product recovered terms are summed up in order to get the total pressure drop in
per unit volume of stationary phase and time. Thus, it provides the unit.
information about the amount produced per unit time, with re- Most of the preparative chromatographic media available and
spect to the cost of the equipment required; indeed, the amount used in bioprocesses are compressible. For a given media, col-
of stationary phase is proportional to the size of the unit re- umn diameter, column length, solvent nature and temperature,
quired, which determines in turn the cost of the equipment. In there exists a critical velocity above which the properties of the
the context of this work, the solvent consumption (S) is defined bed change significantly, namely, the bed is compressed and
as the amount of solvent used per unit time and mass of product the pressure drop becomes infinite; this is the so-called critical
recovered; this quantity is multiplied by the feed concentration velocity [26]. It is evident that the packing velocity for a given
of the target species in order to get a dimensionless solvent con- column, which represents the maximum allowable operating ve-
sumption. The optimization requires I1 = P to be maximum locity, has to be chosen below the corresponding critical value,
and I2 = S to be minimum. The mathematical expressions of but not far from it since high velocities lead to larger produc-
the performance indicators depend upon the decision variables tivities and throughput. Overcoming the packing velocity, umax ,
corresponding to the specific problem studied, as it is explained leads to compression of the column and the subsequent degrada-
in detail below. tion in performance and increase in pressure drop. The critical
The concentration factor, α, achieved for a given set of op- velocity, uc , may be calculated through the following correlation
timum decision variables, is defined as the concentration of the [26]:
target species in the product fraction divided by the concentra-
Lo
tion of the target species in the feed mixture, i.e. ci /ciF . For a u c Lo = η +θ (9)
given pair (P, S) the value of the concentration factor is fixed, d
and as expected from the definition, this takes values smaller or where Lo is the gravity settled bed height, d the column diameter
equal to 1. Another important parameter is the throughput, T, and η and θ are the empirical constants that depend upon the
which is defined as the amount of product processed per unit of stationary phase, the mobile phase used, the temperature and the
time, which can also be calculated from the optimum solution. column section. In this work, umax = 0.8uc will be considered.
It is important to notice that the column section, being a scaling In the case of SMB, this condition must apply in section 1 of the
factor, is kept constant in all our computations. unit, where the flow is always larger than in the other sections,
i.e. u1 = umax .
3.2.2. Constraints It can be proved that in the case-studies analyzed in this work,
The objective functions are optimized subject to two kind the flow-rate in the unit at optimal operating conditions corre-
of constraints: constraints on the product quality, i.e. purity (X) sponds to the maximum flow-rate allowed. This is given by the
and recovery (Y) of the target compounds, and constraints on maximum pressure drop constraint in the case of rigid media,
the operation, i.e arising typically from considerations about the and by the critical flow-rate in the case of compressible media.
G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68 61

This rule holds only if the efficiency of the columns is high solved in this work is the following:
enough and the performance is only limited by the velocity at
max J1 (11)
which the process is run. In all the optimizations carried out in d1 ,d2 ,...
this work, the maximum flow-rate condition is used after veri- max J2 (12)
fying that this is indeed the one that applies for the two cases d1 ,d2 ,...
considered. The mathematical expression of the p or umax con- where d1 , d2 , . . . are the decision variables, which are defined
straint constitutes a relationship between the decision variables, below for each case.
which reduces the number of degrees of freedom in the system. There are unfortunately no general criteria to select the tun-
For SMB, there exists an additional operation constraint, able parameters in Eq. (10), i.e. μl and n. In this work, they have
which requires the switch time to be larger than a certain mini- been chosen through trial and error. It is however worth noting
mum value, typically of the order of 25–30 s. Below this value that wrong μl values lead to long convergence times or no con-
of the switch time the disturbances associated to port switch- vergence at all, and that the tuning of such parameters to find
ing would occur so frequently with respect to the switch time reasonably good values was typically rather short.
to hamper the possibility of achieving a stable and steady SMB The optimization routine used successfully in this work is
operation. a Non-Sorting Genetic Algorithm (NSGA) [27]. This method
has already been used in the frame of chromatographic process
3.2.3. Optimization algorithm, penalty functions and optimization and SMB technology [24,28]. The computational
objective functions time required to obtain a Pareto set depends on the complexity
The optimization consists of the following basic steps: for a of the process, on the nature of the separation and on the specific
given set of decision variables selected in the defined ranges or values of the constraints. In the cases analyzed here the compu-
windows (i.e. the subset of the space of the decision variables tational time was in the order of from a few hours to a few days
where optimum conditions are searched) the simulation of the (Intel Pentium Dual Core Processor, 3 GHz, 2 GB RAM).
chromatographic process is carried out, and the performance
indicators (Il ) and performance parameters (purities and/or re- 4. Plasmid DNA purification
coveries of the target species) are calculated. The values of the
constraints are taken into account during the optimization by 4.1. Definition of the column chromatography optimization
using a penalty function, as it is typically the case in constrained problem
optimization problems. Whenever the constraints are not met,
penalty functions are applied to the performance indicators, so Fig. 2 shows a simulated overloaded chromatogram of the
that this values are worse to an extent that is a function of the mixture with properties given in Table 1 in a 10 cm column at the
deviation from the constraints. The penalty function used here conditions described previously [15]. The first peak corresponds
takes the following form: to the pDNA which is completely excluded from the pores. The
times ta and tb are the breakthrough time of pDNA, and the
 c  ∗   total elution time of the injected sample, respectively. The first
 |Ci − Ci | n
Jl = Fl 1 − μl , l = 1, 2 (10) is defined as the time at which 0.25% of the injected mass of
Ci∗ pDNA is eluted, and the second as the time at which 0.25% of
i=1
the mass of species n injected is not yet eluted. The cycle time,
where c is the number of constraints that have to be tested; Ci tc , is defined as the minimum time interval required between two
the values of the constrained quantities and Fl depend on the consecutive injections, i.e the difference tb − ta .
performance indicators; Ci and Fl the calculated by the model; The search of the optimum operating conditions expressed
Ci∗ the relevant values of the upper or lower bounds given by by Eqs. (11) and (12) is carried out in the space of the decision
the constraints, μl the weighting factors for the deviation from variables, these being the length of the column, L, the flow-rate,
the constraints; n a tunable exponent (always equal to 2 in this Q, and the injected volume, VI (refer to Table 2 for other details).
work) and finally, Jl are the objective functions, i.e. the quantities The operation constraint is given by the maximum flow-rate,
to be optimized. The Jl values are used by the optimization since the media used for this application is compressible. As a
algorithm during the selection of a new decision variables set, consequence, the choice of the decision variable Q is subject to
i.e. the optimizer uses J1 and J2 to assign a performance index an upper constraint Qmax = 0.8Qc ; in this particular case, Qc
to the specific set of decision variables, and this performance is given by the expression:
index is utilized in the selection of the new decision variables.
1
The functions Fl (Il ) are generally equal to the performance Qc (ml/s) = 0.1055 + 0.6940 (13)
indicators; however, in the case studied here, the two functions L (cm)
F1 and F2 are selected such that F1 = I1 = P and F2 = 1/(1 + The other two constraints are related to the quality of the
I2 ) = 1/(1 + S). This is convenient because, by doing so, both product: these are the pDNA purity (X) and recovery (Y). The
objective functions have to be maximized. S may always be purity and the recovery constraints are fixed by the nature of the
back calculated and indeed, during the analysis of the results purification step. In this work, the choice of minimum recovery
presented below, the physically meaningful indicator, S, is used. is kept constant at Ymin = 0.98, and different minimum purities
The formal definition of the two objective optimization problem (Xmin ) are investigated.
62 G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68

The minimum recovery constraint, Ymin is used to determine above, the second product quality constraint, i.e. the recovery, is
the time at which the collection of the pDNA fraction ends; this used to determine the time at which the collection of the pDNA
is the time tY shown in Fig. 2, which is obtained by integrating fraction ends, and therefore it is always fulfilled by definition.
the pDNA concentration profile and by applying the following
equation: 4.2. Definition of the SMB optimization problem
 tY
Q
Ymin = 0.98 = F
cpDNA dt (14) The decision variables of the SMB optimization problem are
VI cpDNA ta the flow-rate ratios in the sections, i.e. m1 , m2 , m3 , the switch
F
where cpDNA is the concentration of pDNA in the feed mixture time, t ∗ , and the column length, L. In order to reduce the com-
and VI is the injected volume. Once tY is calculated using Eq. plexity of the problem, the flow-rate ratio in Section 4 is kept
(14), the recovery constraint is necessarily fulfilled and the purity constant at m4 = −0.05 (refer to Table 2 for other details).
of the pDNA fraction, X, is calculated as: The SMB is run in a closed loop mode and, as in the column
 tY chromatography case, the column section is kept constant (See
ta cpDNA dt Table 2). As explained in Section 3.2.2, the maximum flow-rate
X =  tY (15) constraint applies to section 1, hence Q1 = Qmax = 0.8Qc , thus
ta (cpDNA + cRNA + cn ) dt
implying that the columns are not compressed.
Whenever the selected operating conditions lead to baseline In the SMB case, pDNA is to be recovered in the raffinate
separation between the pDNA and the RNA peaks shown in stream whereas the impurities end up in the extract. The raffinate
Fig. 2, tY is set equal to the total elution time of pDNA and both, purity is defined as:
recovery (Y) and purity (X), are equal to 1. It follows that in our R
c̄pDNA
optimization procedure Y may be either 0.98 or 1. X= (18)
Productivity is calculated according to the definition given in R
c̄pDNA + c̄RNA
R + c̄nR
Section 3.2.1, which in the case studied here corresponds to the
following expression: where c̄iR is the concentration of the species i in the raffinate
averaged over one cycle:
F
YVI cpDNA  t+nc t ∗
P= (16) 1
tc AL c̄i =
R
ciR dt (19)
nc t ∗ t
where Y is the recovery of pDNA (0.98 or 1), tc the cycle time and
AL is the volume of the column. The productivity is therefore where the average can be calculated from any time t after reach-
expressed in terms of two of the three decision variables, i.e L ing cyclic steady state.
and VI , where L is a function of the maximum flow-rate through The recovery is given by the expression:
Eq. (13). R
Rc̄pDNA
According to the definition given in Section 3.2.1, S is written Y= F
(20)
in the following dimensionless form: FcpDNA

Qtc where F and R are the feed and the raffinate flow-rates.
S= (17) The performance indicators are defined as:
YVI
R
Rc̄pDNA
In order to write the penalty function, the corresponding pa-
rameters shown in the first column of Table 4 are used in the P= (21)
nc AL
general expression of Eq. (10). The two penalty functions, act-
D+F
ing on F1 = P and F2 = 1/(1 + S), respectively, have only one S= (22)
term that depends on the single constraint C1 = X; the penalty FY
functions are applied whenever C1 < C1∗ and the exponent of where nc is the total number of columns in the unit and D is the
these terms in Eq. (10) is n = 2 in both cases. As mentioned desorbent flow-rate.
Table 4
Penalty function parameters of Eq. (10) for the column chromatography and the SMB separation processes of the two case studies investigated in this work; the
parameter n is equal to 2 in all cases
Column pDNA SMB pDNA Column TB SMB TB
c =1, C1 X 0.85–0.99 X 0.90, 0.95 XA 0.97 XA 0.97
c =2, C2 – Y 0.98 XB 0.97 XB 0.97
c =3, C3 – – YA 0.97 –
c =4, C4 – – YB 0.97 –
l =1, F1 = P, J1 μ1 = 22, 000 if Ci < Ci∗ μ1 = 22, 000 if Ci < Ci∗ μ1 = 100, 000 if Ci < Ci∗ μ1 = 80, 000 if Ci < Ci∗
μ1 = 500 if Ci > Ci∗ μ1 = 200 if Ci > Ci∗
l = 2, F2 = 1
1+S , J2 μ2 = 1000 if Ci < Ci∗ μ2 =1000 if Ci < Ci∗ μ2 =50,000 if Ci < Ci∗ μ2 =80,000 if Ci < Ci∗
μ2 =10 if Ci > Ci∗ μ2 = 200 if Ci > Ci∗
G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68 63

Fig. 4. Column chromatography, pDNA case study. Pareto sets for X =


0.85–0.99 and Y = 0.98. On the X = 0.95 Pareto set the decision variables
and concentration factor corresponding to the larger and the smaller P, respec-
tively, are: L = 11.2 cm, VI = 2.6 ml, Q = 0.13 ml/s, α = 0.78 and L = 36 cm,
VI = 12 ml, Q = 0.10 ml/s, α = 0.90 (refer to Tables 1 and 2 for more details).

Contrary to the column chromatography case where the con-


straint on Y is always fulfilled since it is used to determine the
collection period for pDNA, in the SMB case the constraints on
X and Y may be fulfilled at the same time, or not. As a result both
constraints have to be tested and their fulfillment has to be ac-
counted for in the penalty functions. These are obtained by using
the parameters reported in the second column of Table 4 in the
general expression of Eq. (10). The two penalty functions, act-
ing on F1 = P and F2 = 1/(1 + S), may have two terms which
are dependent on the two constraints C1 = X and C2 = Y ; these
are applied whenever Ci < Ci∗ . As indicated in Table 4, a “soft”
penalty is also applied in the cases where Ci > Ci∗ since the opti-
mum is always expected at Ci = Ci∗ due to the trade off existing
between the fulfillment of the constraints and the optimization of
the objective functions; the exponents of these terms are n = 2
in both cases.

4.3. Results

4.3.1. Column chromatography


Fig. 4 shows the Pareto sets obtained for several purity val-
ues ranging from 0.85 to 0.99, and at a constant recovery of
0.98. As expected, the larger the purity, the larger the solvent re-
quirements for a given productivity. At X = 0.99, the optimum
operation cannot be run at P larger than about 1.7 mg/(h ml), and
a small increase in P affects strongly the solvent requirements, Fig. 5. Column chromatography, pDNA case study. Trends of the decision vari-
as indicated by the steepness of the 0.99 purity Pareto set in the ables along one Pareto set at X = 0.95 and Y = 0.98 (refer to Table 2 for more
region of maximum P. At X = 0.85 on the contrary, maximum details).
productivity is as large as about 0.43 mg/(h ml).
How the decision variables change along the Pareto set is tion (as expected from the trade off relating P and S): large flow-
illustrated in Fig. 5, where P and S are plot against the three rates, and smaller column lengths and injected volumes lead
decision variables for the case where X = 0.95 and Y = 0.98. to larger solvent consumptions. As expected, the concentration
In order to get large productivities the flow-rate must increase, factor, α, increases when going to smaller solvent requirements.
whereas the length and the injected volume have to be kept small. The fact that performance can be improved when lower purity
Similar trends are observed in the case of the solvent consump- values are specified is illustrated from a different point of view
64 G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68

Fig. 6. Column chromatography, pDNA case study. Trade-off between purity Fig. 8. SMB and column chromatography Pareto sets corresponding to the first
and productivity, and between purity and solvent consumption, for a column purification step of pDNA in Sepharose 6FF for X = 0.95, 0.90, Y = 0.98.
length of about 16.5 cm; X = 0.95, Y = 0.98 (refer to Table 2 for more details). On the X =0.95 Pareto set the operating conditions and concentration fac-
tor corresponding to the larger and the smaller P, respectively, are: t ∗ = 20 s,
in Fig. 6. In this Figure, the values of P and S corresponding L = 3.2 cm, Q1 = 0.22 ml/s, Q2 = 0.044 ml/s, Q3 = 0.036 ml/s, Q4 = 0.036
to the points on the Pareto sets of Figure where L =16.5 cm ml/s, α = 0.89 and t ∗ = 65 s, L = 7.2 cm, Q1 = 0.16 ml/s, Q2 = 0.027 ml/s,
are plotted against purity, thus showing that P decreases and S Q3 = 0.08 ml/s, Q4 = 0.025 ml/s, α = 0.94 (2M AS, 100 mM Tris buffer and
10 mM EDTA, refer to Table 2 for more details).
increases when X increases.
Fig. 7 shows the relation between V and VI for the points con-
SMB Pareto sets are far below those corresponding to column
stituting the 0.85 and 0.99 purity Pareto sets. The two variables
chromatography. For instance, in the case of X = 0.95 and Y =
are linearly related and these linear relationships may be used
0.98, and taking S = 4, the productivity is more than doubled
together with the pressure drop constraint in order to select the
when using SMB. Furthermore, in the column chromatography
optimum operating conditions for a given column.
case, it is not possible to access the region of high P even by
increasing S. This shows how the SMB is more versatile than
4.3.2. SMB
the column chromatography process.
Fig. 8 shows the SMB Pareto sets corresponding to Y = 0.98
In Fig. 8, the trends of L, t ∗ , α and Qj with P and S are also
and X = 0.95, and to Y = 0.98 and X = 0.90, together with the
indicated by arrows: larger productivities require smaller values
column chromatography Pareto sets (for the same values of X and
of L and t ∗ , and larger values of Qj , as expected. Again, the
Y). From this figure, it is clearly seen that the SMB operation
concentration factor increases at higher solvent requirements; it
is superior to the column chromatography operation since the
is worth noting that α is larger in the case of the SMB process
(the values are given in the figure captions).
In terms of throughput, with reference to the column cross
section considered here, i.e. S = 0.785 cm2 , and taking for in-
stance the case of X = 0.90, the values range from 80 to 96 mg/h,
while the maximum value achieved in column chromatography
(at this purity) is 36 mg/h.

4.3.3. Discussion
The optimization problem for column chromatography in-
volves three decision variables, namely column volume V, flow-
rate Q, and volume injected VI . The solution of the optimization
problem defines a Pareto set, on each point of which the three
decision variables are univocally defined and the separation per-
formances are given. Incidentally in the case considered here
the optimal flow-rate corresponds always to the maximum flow-
rate allowed by packing stability considerations. This is due to
the fact that column efficiency is high in this case and pack-
Fig. 7. Column chromatography, pDNA case study. Linear relationship be- ing stability constraints are controlling over column efficiency
tween the injected volume and the column volume. The linear regression pa-
limitations.
rameters in the equation VI (ml) = a + bV (ml) are the following: X = 0.85,
a = 0.6987, b = −2.376; X = 0.90, a = 0.6678, b = −2.6289; X = 0.95, With reference to model Eqs. (4)–(6) and considering a spe-
a = 0.628, b = −2.9819; X = 0.97, a = 0.6108, b = −3.3013; X = 0.99, a = cific system, i.e. assigned adsorption isotherms and packing
0.573, b = −3.7411 (refer to Table 2 for more details). characteristics (in terms of bed void fraction, mass transfer co-
G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68 65

efficients, and axial dispersions), it is quite clear that the three the maximum flow-rate is 60% of the critical one than when it
decision variables are not correlated. In other words, there is a is 80% of it.
one-to-one mapping between triplets of decision variables and
parameters in the model equations, thus making each triplet of 5. Tröger’s base enantiomer separation
decision variables associated in a unique way to specific sepa-
ration performances, in terms of productivity and solvent con- 5.1. Definition of the column chromatography optimization
sumption. problem
Previous literature reports different conclusions, namely that
only two independent variables exist for column chromatogra- Fig. 3 shows a simulated chromatogram of the Tröger’s
phy [16,7]. One of these is the number of theoretical stages, base enantiomers, (+)-TB and (−)-TB, which are referred to
which is defined as N = V/ h(Q), where the function h(Q) as species A and B, respectively. The times ta and tb are defined
expresses the dependence of the HETP on the flow-rate, e.g. as in the case of the pDNA purification, i.e the breakthrough
through the van Deemter equation. The other is the loading fac- time of the first eluting species (time for the elution of 0.25%
tor, Lf , which is by definition proportional to the ratio VI /V . of the mass of B injected), and the total elution time of the in-
The cited authors base their conclusions on the use of an equi- jected sample (time at which 99.75% of the mass of species A
librium dispersive model, consisting of only Eq. (4), where the injected is eluted), respectively. The cycle time, tc , is defined as
adsorbed phase concentration is directly given by the adsorption the difference tb − ta .
isotherm and the axial dispersion coefficient lumps the effect of The search of the optimum operating conditions is carried out
mass transfer resistance, axial mixing and axial molecular dif- in the same space of decision variables, as in the case of pDNA:
fusion into a single parameter. In the frame of this model and length of the column, L, flow-rate, Q, and injected volume, VI .
based on the definitions above, it is clear that there are infi- The column section is kept constant (refer to Table 2 for the
nite combinations of triplets of decision variables V, Q, and VI other model parameters).
yielding the same values of N and Lf , hence the same separation Since the microcrystalline cellulose triacetate is a rigid sta-
performances. tionary phase, the operation constraint is given by the maximum
Nevertheless, there is no contradiction between our analysis pressure drop. The parameter φ in Eq. (8) is 2.34 × 106 g/(s
and the previous literature that we have just mentioned [16,7]. cm5 ), and the maximum pressure drop is 40 bar, as to previous
In fact, it is well know that the equilibrium dispersive model is literature [24]. The product quality specifications are the puri-
only a good approximation of the detailed model used in this ties of the two enantiomers, XA and XB , and their recoveries,
work, in the limit of columns with a large number of stages [25]. YA and YB . In order to have data consistent with those reported
Simulations carried out with the detailed model of Eqs. (4)–(6) in the literature, the purity and recovery constraints investigated
(not reported here) confirm that choosing different triplets of the are XA = XB = YA = YB =0.97.
decision variables that correspond to the same values of N and For a given set of decision variables, the values of X, Y, P and
Lf lead to performances, which are indeed different, but only S are obtained by adopting the following procedure. The purity
slightly; the best performances are achieved for the triplet of of B, XB , is calculated and forced to be equal to the specific
decision variables in the calculated Pareto set. Thus, concluding, value Xmin , thus determining the time tX1 , at which the fraction
the correct number of decision variables depends on the model collection of species B ends, as shown in Fig. 3:
adopted; the larger the level of detail in the description of the  tX1
process, the more the decision variables. In the case of column t cB dt
XB = Xmin = 0.97 =  tX1 a (23)
ta (cA + cB ) dt
chromatography using efficient columns two decision variables,
i.e. N and Lf , are enough to capture most of the important effects
using a lumped description of mass transfer and axial dispersion. The purity of the A enantiomer is calculated for the fraction
Three decision variables are however required when the more collected between tX1 to tb :
detailed description of these mechanisms such as that provided  tb
t cA dt
by the model used in this work is preferred. XA =  tb X1 (24)
A related observation can be made with reference to the opti- tX1 (cA + cB ) dt
mization results obtained for the SMB process. Simulations not
If this purity is below the minimum purity required, i.e. XA <
reported here show that if the maximum flow-rate in section 1 is
Xmin , a shorter time interval has to be selected, i.e. the collection
constrained to be 60% of the critical flow-rate, instead of 80% as
of the A fraction has to be started later than tX1 . This is done by
in the results illustrated in Fig. 8, the Pareto set (calculated for a
integrating the concentration profiles from tb back to a certain
minimum purity of 0.90) is only slightly worse than that shown
time tX2 , with tX2 > tX1 , such that the purity requirement is
in Fig. 8. This result can be explained as follows: at a smaller
fulfilled:
flow-rate, the column length corresponding to optimal operat-  tb
ing conditions is down sized in such a way that the productivity, t cA dt
i.e. the amount of pDNA purified per unit time and unit volume XA = Xmin = 0.97 =  tb X2 (25)
tX2 (cA + cB ) dt
of the SMB unit, changes only to a small extent. Of course in
this case, if columns of the same cross-section are adopted the Using this procedure to find adequate collection time intervals
throughput of the SMB unit will be significantly smaller where for both fractions, the purity constraints are always fulfilled. The
66 G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68

recoveries of the two species can then be computed using the


following expressions:
 tb
Q
YA = F
cA dt (26)
VI cA tX2
 tX1
Q
YB = cB dt (27)
VI cBF ta

The values of YA and YB are used to compute the productivity,


P, as:
F + Y cF )
VI (YA cA B B
P= (28)
tc AL

The productivity is therefore expressed in terms of two of the Fig. 9. SMB and column chromatography Pareto sets corresponding to the
three decision variables, i.e L and VI , and these two are in this Tröger’s base enantiomer separation on microcrystalline cellulose triacetate
case related to the flow-rate Q by the maximum pressure drop (CTA) (ethanol) at 50 ◦ C for XA = XB = YA = YB = 0.97 (refer to Tables 1
given by Darcy’s law, Eq. (8). The dimensionless form of the and 2 for more details).
solvent consumption (S) is given by the following relationship:
following way
Qtc (cAF + cF ) E + RcR
EcA
S= B
(29) P= B
(30)
VI (YA cA + YB cBF )
F
nc AL

The penalty functions are obtained using the parameters (D + F )(cAF + cF )


S= B
F + Y cF )
(31)
shown in the third column of Table 4 (with n = 2) in the gen- F (YA cA B B
eral expression of Eq. (10). The two penalty functions, acting
where the concentrations are averaged over one cycle. The
on F1 = P and F2 = 1/(1 + S), may have four terms which are
penalty function parameters of Eq. (10) are shown in the fourth
functions of the four constraints: C1 = XA , C2 = XB , C3 = YA ,
column of Table 4 (with n = 2). The two penalty functions
C4 = YB ; these terms are considered whenever Ci < Ci∗ . As in-
may have two terms which are functions of the two constraints:
dicated in Table 4, a “soft” penalty is also applied in the cases
C1 = XA , C2 = XB and are applied when Ci < Ci∗ .
where Ci > Ci∗ .
The results of this optimization have been reported elsewhere
[24], and will be simply used in the following.
5.2. Definition of the SMB optimization problem
5.3. Results
In the case of the SMB process, the decision variables are the
flow-rate ratios in the sections, i.e. m1 , m2 , m3 , m4 , the switch Fig. 9 shows the SMB and the column chromatography Pareto
time, t ∗ , and the column length, L, which are the same ones sets for the TB enantiomer separation at 0.97 purity and recov-
considered for SMB optimization of the pDNA case, plus m4 ery of both enantiomers. Like in the linear pDNA separation
which is now included whereas it was kept constant in Section case, the SMB Pareto set is far below the column chromatog-
4.2. The characteristics of the columns are given in Table 2. raphy case. Furthermore, the last one is rather short, indicating
As in the column chromatography case studied before, the that only a narrow productivity range is accessible. As a result,
constraints are the purities and the recoveries of both enan- it can be concluded that the SMB process achieves better per-
tiomers, XA , XB , YA and YB . It can be easily proven through formance and is more flexible than the corresponding column
simple mass balances that for an equimolar feed mixture of A chromatography process.
and B and identical purity constraints on both the extract and
the raffinate fractions, the SMB operation leads to recovery val- 6. Concluding remarks
ues such that YA = YB = XA = XB . Therefore, once the purity
values are selected as XA = XB =0.97, the recovery constraints This work analyzes the performance of the SMB and the
are fulfilled and the basis for comparison between the SMB pro- column chromatography processes for two different separation
cess studied in this section and the column process studied in problems: the first stage of the polishing step within the down-
Section 5.1 are well established. stream processing of pDNA, and the Tröger’s base enantiomer
The purities are defined as usual, i.e. the average concentra- separation, in which the adsorption isotherms are linear and
tion of the target species (A in the extract, B in the raffinate) non-linear, respectively. Simulation tools are used together with
divided by the average total concentration. The expressions of an optimization routine in order to find the optimum operating
the recoveries are also straightforward and similar to Eq. (20). conditions leading to maximum productivity and minimum
The productivity and the solvent consumption are defined in the solvent consumption; the optimum solution for a given process
G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68 67

is a curve on the productivity-solvent consumption plane (P, DLi axial dispersion coefficient of component i (cm2 /s)
S), called Pareto set. The comparison between the column and F feed flow-rate (ml/s)
the SMB processes is based on the relative position of the two F1 , F2 functions of the performance indicators, i.e. Fl (Il )
Pareto sets in this plane. I1 , I2 performance indicators
The optimization problems have been formulated very care- J1 , J2 objective functions
fully so that the two processes are comparable and the con- ki mass transfer coefficient component i (1/s)
clusions meaningful, for each of the case studies. This proce- Ki Langmuir equilibrium constant of component i
dure involves the selection of the decision variables, the defi- (ml/mg)
nition of comparable objective functions and the consideration L column length (cm)
of equal operative and product quality constraints for the two Lf loading factor
processes. Lo gravity settled bed height (cm)
For the column chromatography pDNA purification process, mj flow-rate ratio in section j
several purities are considered leading to different Pareto sets; n penalty function parameter in Eq. (10)
this helps to confirm several effects, for instance the trade-off nc total number of columns
between purity and performance. In the SMB case, two purities j
ni adsorbed phase concentration of component i in the
are investigated. Comparing the position of two corresponding stream j (mol/l)
Pareto sets, the SMB process is found to outperform column N number of theoretical plates
chromatography as it may be seen in Fig. 8 for two different pu- p pressure (bar)
rities (X = 0.90, X = 0.95) and constant recovery (Y = 0.98), P productivity (mg/s ml)
i.e the productivity may be doubled for a given value of the sol- Qj volumetric fluid flow-rate in section j (ml/s)
vent consumption. Furthermore, the region of high P cannot be R raffinate flow-rate (ml/s)
achieved using the column process. S solvent consumption per unit mass of target species
The same is observed in the TB case, where, as expected con- produced (ml/(s mg))
sidering its non-linear adsorption behavior, the benefit provided t time (s)
by SMB is even larger than in the linear case. Indeed, one im- t∗ switch time (s)
portant feature of the SMB process is that high performances T throughput (mg/h)
can be achieved even at rather low values of selectivity and at u velocity (cm/s)
relatively low efficiencies. Fig. 9 also shows that the versatility uc critical velocity (cm/s)
of the SMB operation is significantly larger, as indicated by the V column volume (ml)
long and flat SMB Pareto set when compared to the short and VI injected volume (ml)
steep Pareto curve of column chromatography. x dimensionless space
The investigations presented in this work confirm the poten- Xi purity of component i
tial of the SMB process based on performance, throughput and yI dimensionless quantity defined by equation
versatility. These advantages have to be weighed up against the Yi recovery of component i
higher complexity of the SMB operation and the larger invest- z space coordinate (cm)
ment effort required, as compared to column chromatography.
This applies also to bio-separations, as it has been proved here for Greek letters
one case-study. It is worth noting that the case considered here α concentration factor
where the retention behavior was linear is a rather simple one, ε bed void fraction
but still SMB has proven to be better than column chromatogra- ε∗ overall bed void fraction
phy. We argue that in the case of more difficult bio-separations, εpi intraparticle void fraction related to species i
e.g. non-linear as the case of the enantiomers of the Tröger’s φ Darcy’s law parameter in Eq. (8) (g/(s cm5 ))
base, the improvement of performance using SMB would be γi Henry’s constant of component i
even larger and more attractive. η empirical constant in Eq. (9) (cm3 /s)
θ empirical constant in Eq. (9) (cm2 /s)
Nomenclature μ penalty function parameter in Eq. (10)
τ dimensionless time
ap specific surface of the adsorbent particles (cm−1 )
A cross-section (cm2 ) Subscripts and superscripts
c number of constraints A strong component label
j
ci concentration of component i in the stream j (mol/l, B weak component label
mg/ml) i component index (i = 1, 2, . . . , c)
C1 , C2 constraints j stream/section index (j = 1, 2, . . .)
C1∗ , C2∗ minimum value of the constraints max maximum
d column diameter (cm) min minimum
d1 , d2 , . . . decision variables n very retarded (and possibly retained) impurities
D desorbent flow-rate (ml/s) pDNA plasmid DNA
68 G. Paredes, M. Mazzotti / J. Chromatogr. A 1142 (2007) 56–68

R raffinate [10] A. Felinger, G. Guiochon, J. Chromatogr. 591 (1992) 31.


RNA retarded impurities (mainly RNA) [11] A. Felinger, G. Guiochon, AIChE J. 40 (1994) 594.
[12] A. Felinger, G. Guiochon, J. Chromatogr. A 752 (1996) 31.
[13] A. Felinger, G. Guiochon, J. Chromatogr. A 796 (1998) 59.
Acknowledgements [14] G. Paredes, M. Mazzotti, J. Stadler, S. Makart, M. Morbidelli, Adsorption
11 (2005) 841.
[15] G. Paredes, S. Makart, J. Stadler, M. Mazzotti, Chem. Eng. Technol. 28
The support of the Swiss Innovation Promotion Agency (KTI, (2005) 1335.
project number 6138.1 KTS) is gratefully acknowledged. The [16] A. Jupke, A. Epping, H. Schmidt-Traub, J. Chromatogr. A 944 (2002) 93.
authors are thankful to Joachim Stadler, Karol Lacki and Lars [17] A. Seidel-Morgenstern, Analusis 26 (1998) 46.
Hagel of GE Healthcare for useful discussions. [18] J. Strube, S. Haumreisser, H. Schmidt-Traub, M. Schulte, R. Ditz, Org.
Process Res. Dev. 2 (1998) 305.
[19] K. Mihlbachler, A. Seidel-Morgenstern, G. Guiochon, AIChE J. 50 (2004)
References 611.
[20] FDA, FDA Policy Statement for the Development of New Sterioisomeric
[1] S. Abel, M. Bäbler, C. Arpagaus, M. Mazzotti, J. Stadler, J. Chromatogr. Drugs. Chirality 4 (1992) 338.
A 1043 (2004) 201. [21] J.E. Rekoske, AIChE J. 47 (2001) 2.
[2] G. Paredes, S. Abel, M. Mazzotti, M. Morbidelli, J. Stadler, Ind. Eng. [22] M.P. Pedeferri, G. Zenoni, M. Mazzotti, M. Morbidelli, Chem. Eng. Sci.
Chem. Res. 43 (2004) 6157. 54 (1999) 3735.
[3] U. Voigt, J. Kinkel, R. Hempel, R.M. Nicoud, United States Patent No.: [23] J. Worlitschek, M. Bosco, M. Huber, V. Gramlich, M. Mazzotti, Helv.
US006306306B1, 2001. Chim. Acta 87 (2004) 279.
[4] R. Wooley, Z. Ma, N.H.L. Wang, Ind. Eng. Chem. Res. 37 (1998) 3699. [24] M. Amanullah, M. Mazzotti, J. Chromatogr. A 1107 (2006) 36.
[5] M. Juza, M. Mazzotti, M. Morbidelli, Trends Biotechnol. 18 (2000) 108. [25] C. Migliorini, A. Gentilini, M. Mazzotti, M. Morbidelli, Ind. Eng. Chem.
[6] A. Seidel-Morgenstern, G. Guiochon, AIChE J. 39 (1993) 809. Res. 38 (1999) 2400.
[7] F. Charton, M. Bailly, G. Guiochon, J. Chromatogr. A 687 (1994) 13. [26] J.J. Stickel, A. Fotopoulos, Biotechnol. Prog. 17 (2001) 744.
[8] I. Qui nones, C.M. Grill, L. Miller, G. Guiochon, J. Chromatogr. A 867 [27] V. Bhaskar, K. Gupta, A.K. Ray, Rev. Chem. Eng. 16 (2000) 1.
(2000) 1. [28] Z. Zhang, M. Mazzotti, M. Morbidelli, J. Chromatogr. A 989 (2003)
[9] A. Felinger, G. Guiochon, J. Chromatogr. 609 (1992) 35. 95.

Das könnte Ihnen auch gefallen