SONOLINE G20

US

Replacement of Parts
System
Removing and Replacing FRUs

08647831
08647815

© Siemens AG 2004
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
Siemens Medical
Ultrasound
ejt DivisionSolutions
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: US06-102.841.01.02.02 English


Replaces: US06-102.841.01.01.02 Doc. Gen. Date: 10.05
n.a.

Part No.: n.a.


sd
2 Revision / Disclaimer
1Revision / Disclaimer

Copyright
Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy through-
out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-
grams are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Sub-
ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equip-
ment solely on the medium in which such program is delivered for the purpose of operat-
ing the equipment in accordance with the instructions set forth in the operator’s manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-
grams nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

SONOLINE G20 US06-102.841.01.02.02 Page 2 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Revision / Disclaimer 3

of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-
rect, consequential, incidental or other damages arising out of the sublicense of the com-
puter programs supplied with the equipment.

Siemens US06-102.841.01.02.02 Page 3 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
4 Table of Contents
1- 0Table of Contents

1 _______ Removing and Replacing Field Replacement Units ____________________ 6

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Before Starting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Verify the Operation of the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Remove the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Install the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Verify the Monitor Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Remove the Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Install the Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Trackball . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Transducer Holders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Remove the Transducer Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Install the Transducer Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
OEM AC Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Remove the OEM AC Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Install the OEM AC Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Left-Side Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Remove the Left-Side Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Install the Left-Side Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Right-Side Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Remove the Right-Side Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Install the Right-Side Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
OEM Shelf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Remove the OEM Shelf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Install the OEM Shelf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Caster Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Remove the Caster Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Install the Caster Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Caster . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
CD-RW Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Remove the CD-RW Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Install the CD-RW Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Verify the Operation of the CD-RW Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Front PCBs, Shielding Cover and Hard Disk Drive (HDD) . . . . . . . . . . . . . . . . . . . . . . . 35
Front Shielding Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
A34 AVIF2 PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
A33 ECPR PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Hard Disk Drive (HDD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
A33-2 SBC PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
A32 CFMS PCB (SLC Systems Only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

SONOLINE G20 US06-102.841.01.02.02 Page 4 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Table of Contents 5

A31 TRBF PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46


A30 TXRX PCB and A30-1 PBIF PCB Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Rear PCBs and Shielding Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Rear Shielding Cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
A36 RGLT PCB and Rear Plane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
A35 MTHR PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Switching Regulator (SWRG2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Remove the SWRG2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Install the SWRG2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
SWRG2 Fuse. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
SWRG2 Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Line Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Remove the Line Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Install the Line Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

Siemens US06-102.841.01.02.02 Page 5 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
6 Removing and Replacing Field Replacement Units
2-
Overview
1Removing and Replacing Field Replacement Units

This procedure describes the removal and installation of field replacement units (FRUs)
starting at the top of the system and then working down and into the system. Some proce-
dures overlap and duplicate the description of some tasks.
A Spare Part Catalog (SPC) has been established on the Internet at the following URL:
http://cs.med.siemens.de/
The following is a list of the FRU removal and installation procedures:
• Monitor
• Control Panel
- Trackball
• Transducer Holders
• OEM AC Outlet
• Right-Side Panel
- LC Transducer “Parking” Port
• Left-Side Panel
• OEM Shelf
• Caster Panel
- Caster
• CD-RW Drive
• Front PCBs and Shielding Cover
- Front Shielding Cover
- A34 AVIF2 PCB
- A33 ECPR PCB
- Hard Disk Drive (HDD)
- A33-2 SBC PCB
- A32 CFMS (SLC Only)
- A31 TRBF PCB
- A30 TXRX PCB and A30-1 PBIF PCB Assembly
- A37 MPXA
- A37-1 MPXT (SLC Only)
• HDD (part of the A33 ECPR assembly)
• Rear PCBs and Shielding Cover
- Rear Shielding Cover
- A36 RGLT PCB and Rear Plane
- A35 MTHR PCB
• Switching Regulator
• Fuse

SONOLINE G20 US06-102.841.01.02.02 Page 6 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 7

• Line Filter
- MAINS Circuit Breaker
- AC Power Outlet

Warnings and Cautions 0

When replacing spare parts on the SONOLINE G20 ultrasound system always observe
the safety cautions and warnings.

WARNING The G20 system contains hazardous voltages that can range from
90 VAC to 264 VAC and cause serious injury or death.
¹ Only authorized service personnel are permitted to
remove protective covers and access the energized com-
ponents of this system.

WARNING The weight of the G20 system represent lifting hazards.


The use of improper lifting methods or lifting in an awkward man-
ner can cause serious injury.
¹ Transport instructions for the G20 system are provided
on the shipping container. The G20 system is not
designed to be lifted once removed from its shipping
container. Transport the G20 system by rolling it on its
casters.
¹ The US National Institute for Occupational Safety and
Health publishes design guidelines indicating that items
weighing greater than 15.9 kg (35 lb.) must be lifted by
two people or mechanical lifting aids. Items weighing
greater than 34 kg (75 lb.) must be lifted only with
mechanical lifting equipment.

CAUTION When replacing fasteners, use fasteners of the correct length.


¹ Using the wrong length fasteners may cause damage to
underlying equipment.

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Siemens US06-102.841.01.02.02 Page 7 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
8 Removing and Replacing Field Replacement Units

Before Starting 0

Prior to performing any of the FRU removal or installation procedures, do the following:

1. Move the system to a level area with enough space to work around the system and set
the system brakes.
2. Power off the system, switch off the MAINS power circuit breaker, and leave the power
cord connected to the wall outlet.

Verify the Operation of the System 0

After performing any of the FRU installation procedures in this document verify the opera-
tion of the system by performing the following steps:

1. Plug the AC power cord into the wall outlet.


2. Connect a transducer to the transducer connector.
3. Switch the MAINS circuit breaker to on.
4. Press the Standby button to power on the system.
5. Confirm the boot sequence completes without error.
6. Perform a POST test and verify no errors are logged.
7. Verify a 2-D image is displayed.
8. Press the M button and verify an M-mode strip is displayed.
9. Press Standby to power off the system.

SONOLINE G20 US06-102.841.01.02.02 Page 8 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 9

Monitor 0

This section describes the removal and installation of the G20 monitor and then verifies
the functionality of the monitor after performing the installation. The installation of the
monitor requires approximately ten minutes.

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the Monitor 0

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Loosen (not remove) the screw that secures the monitor post cover to the system.

Fig. 1: Rear of Monitor


Pos. A Screw location
Pos. B Monitor post cover

3. Lift the monitor post cover up, removing it from the system.
4. Unscrew the monitor cable and then unplug it from the connector.

Siemens US06-102.841.01.02.02 Page 9 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
10 Removing and Replacing Field Replacement Units

NOTE The monitor cable connector may have one or two screws.

Fig. 2: Monitor Cable


5. Remove the two screws that secure the monitor to the post; locate the screws on the left
and right sides of the monitor.

Fig. 3: Monitor Securing Screw Locations


6. Lift the monitor from the post using proper lifting techniques.
7. Place the defective monitor in the packaging and prepare it for return.

Install the Monitor 0

WARNING The monitor is heavy. Improper installation of the threaded fasten-


ers and the mounting hardware that support the monitor can result
in serious injury.
¹ Ensure that the two cap screws are tightened securely.

1. Place the monitor on the post and align the two monitor mounting holes with the post
mounting holes.
2. Secure the two screws that secure the monitor to the post.

SONOLINE G20 US06-102.841.01.02.02 Page 10 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 11

3. Route the monitor cable through the base of the monitor mount, between the mounting
screws and the post; check that the cable moves freely and stays within the base with-
out binding.
- Rotate the monitor completely counterclockwise and then completely clockwise.
- Tilt the monitor completely forward and back.
- Secure the monitor cable to the connector
- Rotate the monitor completely counterclockwise and then completely clockwise.

Fig. 4: Monitor Cable Route in the Base


Pos. A Monitor cable
Pos. B Cable route between the mounting screws and the post
Pos. C Mounting screws
Pos. D Monitor post

4. Install the monitor cover.


- Place the cover over the cover mounting screw.
- Secure the cover mounting screw.

Fig. 5: Monitor Cover Mounting Screw

Siemens US06-102.841.01.02.02 Page 11 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
12 Removing and Replacing Field Replacement Units

Verify the Monitor Settings 0

1. Power on the system.


- Press Standby.
- Wait for the boot sequence to complete.
2. Adjust the monitor brightness and contrast as necessary.

NOTE The CONTRAST and BRIGHTNESS controls are located next to


each other on the lower left-side of the monitor.

3. Press F6 to access the Preset Main Menu screen.


4. Select Display from the System Configuration menu.
5. Adjust the Monitor settings for character brightness, flicker control and monitor contrast.

Fig. 6: Presets Menu - Display Screen


Pos. A Character Brightness settings
Pos. B Flicker control - on or off
Pos. C Monitor contrast settings - low or high

SONOLINE G20 US06-102.841.01.02.02 Page 12 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 13

Control Panel 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the Control Panel 0

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Loosen the screw that secures the monitor post cover to the system and remove the
monitor post cover. The control panel screws are accessible.

Fig. 7: Monitor Post Cover


3. Loosen the two screws that secure the control panel to the system.

Fig. 8: Control Panel Mounting Screws

Siemens US06-102.841.01.02.02 Page 13 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
14 Removing and Replacing Field Replacement Units

4. Remove the five screws from the bottom of the control panel.

Fig. 9: Bottom of the Control Panel (screw locations)


5. Tilt the monitor back and carefully lift the control panel off the system.

NOTE Be careful when lifting the control panel off the system, it is still at-
tached to a cable that is attached to the system.

6. Turn the control panel upside-down and disconnect the control panel cable, (Figure 10,
item A).

Fig. 10: Control Panel - (viewed from bottom)


Pos. A Control panel cable
Pos. B Trackball assembly
Pos. C Trackball assembly cable

SONOLINE G20 US06-102.841.01.02.02 Page 14 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 15

Install the Control Panel 0

1. Turn the control panel upside-down and connect the control panel cable (See Fig. 10,
Pos. A).
2. Turn the control panel over and place the control panel on the system.
3. Install the five screws in the bottom of the control panel.
4. Install the two screws that secure the control panel to the system.
5. Install the monitor post cover on the system and secure the screw that secures the
monitor post cover to the system.

Trackball 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the Trackball

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Rotate the trackball retaining ring counterclockwise, remove the trackball, and then re-
move the control panel (Remove the Control Panel / p. 13).
3. Turn the control panel over and disconnect P2 from J2.
Note the position of the trackball assembly when removing it from the control panel.

Fig. 11: Trackball Assembly


Pos. A P2/J2
Pos. B Screw locations

4. Remove the two screws that secure the trackball assembly to the control panel.

Siemens US06-102.841.01.02.02 Page 15 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
16 Removing and Replacing Field Replacement Units

Install the Trackball

1. Turn the trackball assembly over and install the two screws that secure the trackball as-
sembly to the control panel.
2. Connect P2 to J2.
3. Install the control panel (Install the Control Panel / p. 15).
4. Place the trackball into the trackball assembly and rotate the retaining ring clockwise to
secure the trackball.

SONOLINE G20 US06-102.841.01.02.02 Page 16 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 17

Transducer Holders 0

Remove the Transducer Holder 0

The short transducer holders and the Gel/Endo transducer holders are removed and
replaced on the system identically. The following steps refer to transducer holders, but the
directions apply to both transducer holders and gel bottle holders.

1. Squeeze and hold the retaining clip toward the holder.

Fig. 12: Transducer Holder


2. Firmly press down on the transducer holder until the holder is released from the system.

Install the Transducer Holder 0

1. Firmly press the transducer holder upward in the groove on the side of the control pan-
el.

Siemens US06-102.841.01.02.02 Page 17 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
18 Removing and Replacing Field Replacement Units

CAUTION Verify that the transducer holder locks into place on the side of the
control panel.
¹ Failure to do so may cause the transducer holder to
detach from the system, dropping the transducer to the
floor and possibly damaging the transducer.

Fig. 13: Transducer Holder

SONOLINE G20 US06-102.841.01.02.02 Page 18 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 19

OEM AC Outlet 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the OEM AC Outlet 0

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove all peripherals mounted on the OEM shelf.
See the Peripheral Configuration section for detailed steps to remove and install
peripherals.
3. Remove the three screws that secure the OEM AC outlet panel to the system.

Fig. 14: OEM AC Outlet Mounting Screw Locations


Pos. A Mounting Screws
Pos. B AC Outlet
Pos. C F1 fuse
Pos. D F2 fuse

4. Carefully pull the panel away from the system.

Siemens US06-102.841.01.02.02 Page 19 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
20 Removing and Replacing Field Replacement Units

5. Disconnect the cable connector.

Fig. 15: OEM Outlet Panel (viewed from the backside of the panel)
Pos. A OEM outlet connector

Install the OEM AC Outlet 0

NOTE When installing the panel, be careful not to pinch the cables be-
hind the OEM AC outlet panel.

1. Connect the OEM AC Outlet cable connector.


2. Align the OEM AC outlet panel with the mounting holes on the system and secure the
three screws.
3. Install the peripherals that were removed from the OEM shelf prior to removing the
OEM AC outlet panel.
See the Peripheral Configuration section for detailed steps to remove and install
peripherals.

SONOLINE G20 US06-102.841.01.02.02 Page 20 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 21

Left-Side Panel 0

This procedure assumes that right and left-side panels are attached to the system; for
systems that have one panel removed complete only steps 1 and 2.

Remove the Left-Side Panel 0

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Loosen the four captive screws that secure the left panel to the system.

Fig. 16: Left-Side Panel, Captive Screw Locations

Siemens US06-102.841.01.02.02 Page 21 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
22 Removing and Replacing Field Replacement Units

3. Detach the left-side panel from the system.


- Move to the front of the system.
- Place a coin in the lower groove that couples the right and left panels together and
turn the coin to pry the left-side panel loose from the system.

Fig. 17: Panel Grooves


Pos. A Lower groove location
Pos. B Upper groove location

4. Pull the panel away from the system.

Install the Left-Side Panel 0

1. Align the left-side panel four captive screws with their matching mounting holes and
tighten the captive screws.
2. Couple the left-side panel to the system.
- Move to the front of the system.
- Press the side panel into place. The panel should snap back into place.

SONOLINE G20 US06-102.841.01.02.02 Page 22 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 23

Right-Side Panel 0

There are two configurations for the right-side panel, one for the LC system and one for
the SLC system. The LC system is configured with a transducer “parking” port; the SLC
system is configured with a transducer sector port. The right-side panel is a spare part
that ships from the factory configured for the LC system. For SLC systems, the “parking”
port cover must be removed and replaced with the sector port cover. The sector port cover
components are shipped along with the right-side panel.
The right-side panel procedures assume that right and left-side panels are attached to the
system.

Remove the Right-Side Panel 0

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. If connected, disconnect the transducers from the system.
3. Loosen the four captive screws that secure the right panel to the system.

Fig. 18: LC Right-Side Panel, Captive Screw Locations

Siemens US06-102.841.01.02.02 Page 23 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
24 Removing and Replacing Field Replacement Units

Fig. 19: SLC Right-Side Panel, Captive Screw Locations


4. Detach the right-side panel from the system.
- Move to the front of the system.
- Place a coin in the lower groove that couples the left and right panels together and
turn the coin to pry the right-side panel loose from the system.

Fig. 20: Panel Grooves


Pos. A Lower groove location
Pos. B Upper groove location

SONOLINE G20 US06-102.841.01.02.02 Page 24 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 25

5. Pull the panel away from the system.

Install the Right-Side Panel 0

The installation of the right-side panel on a G20 LC system differs from that of an SLC
system, because the LC and the SLC use different port covers. For an LC system, install
the right-side panel as it is assembled from the factory. For an SLC system, the LC “park-
ing” port cover must be removed and replaced with the SLC sector port cover compo-
nents.

Install the Right-Side Panel (LC system)

1. Align the right-side panel four captive screws with their matching mounting holes and
secure the captive screws.
2. Couple the right-side panel to the system.
- Move to the front of the system.
- Press the side panel into place. The panel should snap back into place.

Install the Right-Side Panel (SLC system)

1. Remove the four screws that secure the LC “parking” port cover to the side panel.
Discard the metal plate.

Fig. 21: LC “Parking” Port Cover

Siemens US06-102.841.01.02.02 Page 25 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
26 Removing and Replacing Field Replacement Units

2. Mount the SLC sector port cover where the LC “parking” port cover was mounted.
- Align the SLC sector port cover to right-side panel.
- Tighten the four mounting screws that secure the SLC sector port cover to the side
panel.

Fig. 22: SLC Sector Port Cover


3. Align the right-side panel four captive screws with their matching mounting holes and
secure the captive screws.
4. Couple the right-side panel to the system.
- Move to the front of the system.
- Press the side panel into place. The panel should snap back into place.

SONOLINE G20 US06-102.841.01.02.02 Page 26 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 27

OEM Shelf 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the OEM Shelf 0

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove all peripherals from the OEM shelf.
3. Remove the OEM AC outlet panel (Remove the OEM AC Outlet / p. 19).
4. Remove the left-side panel.
5. Remove the right-side panel
6. Remove the two screws that secure the shelf to the right-side of the system.

Fig. 23: OEM Shelf Screw Location


7. Remove the two screws that secure the shelf to the left-side of the system.
8. Lift the back of the shelf up and then slide the shelf out from the groove and the system.

Siemens US06-102.841.01.02.02 Page 27 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
28 Removing and Replacing Field Replacement Units

Fig. 24: Remove the OEM Shelf


Pos. 1 Lift the back of the shelf
Pos. 2 Slide the shelf

Install the OEM Shelf 0

NOTE When installing the OEM shelf, do not to pinch the OEM AC outlet
cable.

1. Slide the shelf tab into the groove in the rear of the OEM bay.

Fig. 25: OEM Outlet Panel (viewed from the backside of the panel)
Pos. A OEM outlet connector

SONOLINE G20 US06-102.841.01.02.02 Page 28 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 29

2. Install the OEM AC outlet panel (Install the OEM AC Outlet / p. 20).
3. Install the two screws on the right-side of the system that secure the OEM shelf the sys-
tem.
4. Install the two screws on the left-side of the system that secure the OEM shelf the sys-
tem.

Siemens US06-102.841.01.02.02 Page 29 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
30 Removing and Replacing Field Replacement Units

Caster Panel 0

Remove the Caster Panel 0

1. Remove the side panel on the same side as the caster that will be removed.
2. Remove the two screws that secure the top of the caster panel to the system.

Fig. 26: Caster Panel Top Screw Locations


3. Loosen the five screws that secure sides of the caster panel to the system.

Fig. 27: Caster Panel Side Screw Locations


4. Lift the panel up off the system.

Install the Caster Panel 0

1. Place the panel on the caster frame; align the top two fastener holes in the cover with
the holes in the frame and secure the screws.
2. Install the five side screws that secure the caster panel to the system.
3. Install the side panel.

SONOLINE G20 US06-102.841.01.02.02 Page 30 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 31

Caster 0

Remove the Caster

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the side panel on the same side as the caster that will be removed.
3. Remove the caster panel.
4. Use a jack to raise the system one and a half inches off the floor; place blocks under the
caster frame next to each caster.

Fig. 28: Jack Placement and Caster Nut Locations

Pos. A Place the jack for lifting


Pos. B Caster nut

WARNING Be careful when lifting the system to change the caster.


¹ If the system is lifted too high it could become unstable
and tip over, resulting in injury to the CSE and/or damage
to the system.

5. Use an adjustable wrench to remove the retaining nut from the frame and remove the
caster.

Install the Caster

1. Apply Loctite # 262 to the thread of the caster and screw the caster into the frame; se-
cure the retaining nut.
2. Remove the blocks from under the caster frame and lower the system off of the jack.
3. Install the caster panel.
4. Install the side panel.

Siemens US06-102.841.01.02.02 Page 31 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
32 Removing and Replacing Field Replacement Units

CD-RW Drive 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the CD-RW Drive 0

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the OEM shelf (Remove the OEM Shelf / p. 27).
5. Remove the two screws that secure the front of the CD-RW drive mounting plate to the
system.

Fig. 29: CD-RW Front Mounting Screw Locations


6. Remove the two screws that secure the top of the CD-RW drive mounting plate to the
system.

Fig. 30: CD-RW Top Mounting Screw Locations

SONOLINE G20 US06-102.841.01.02.02 Page 32 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 33

7. Remove the rear shielding cover (Remove the Front Shielding Cover / p. 36); this al-
lows access to disconnect the CD-RW cable.

Fig. 31: CD-RW Assembly (rear shielding cover removed)


Pos. A CD-RW cable
Pos. B CD-RW drive

8. Disconnect the CD-RW drive cable and slide the CD-RW drive out from the system.

Install the CD-RW Drive 0

1. Slide the CD-RW drive assembly into the system and connect the CD-RW drive cable.
2. Install the rear shielding cover (Install the Front Shielding Cover / p. 37).
3. Install the two screws that secure the top of the CD-RW drive mounting plate to the sys-
tem.
4. Install the two screws that secure the front of the CD-RW drive mounting plate to the
system.
5. Install the OEM shelf (Install the OEM Shelf / p. 28).
6. Install the right-side panel (Install the Right-Side Panel / p. 25).
7. Install the left-side panel (Install the Left-Side Panel / p. 22).

Verify the Operation of the CD-RW Drive 0

1. Power on the system and allow the boot sequence to complete.


2. Create a new study, patient name, patient ID, and scan some images to use for the
testing of the CD-ROM function.

Siemens US06-102.841.01.02.02 Page 33 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
34 Removing and Replacing Field Replacement Units

3. Place a CD-R disk in the CD-RW drive and save the fake patient images to the CD-R
disk.
4. Eject the CD-R disk and re-insert the CD-R disk into the CD-RW drive.
5. Access the file on the CD-R disk. If the file cannot be accessed, or if it is damaged or in-
complete, do the following:
- Store the file to another CD-R disk.
- Check the CD-RW drive cable connections.
- Install the CD-RW drive.

SONOLINE G20 US06-102.841.01.02.02 Page 34 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 35

Front PCBs, Shielding Cover and Hard Disk Drive (HDD) 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

The front PCBs are located in the card cage behind the front shielding cover.
The G20 LC card cage includes the following PCBs:
• A34 AVIF2
• A33 ECPR
• A33-2 SBC
• A31 TRBF
• A30 TXRX/A30-1 PBIF Assembly
The G20 SLC card cage includes the following PCBs:
• A34 AVIF2
• A33 ECPR
• A33-2 SBC
• A32 CFMS
• A31 TRBF
• A30 TXRX/A30-1 PBIF Assembly
• A37 MPXA
• A37-1 MPXT

Siemens US06-102.841.01.02.02 Page 35 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
36 Removing and Replacing Field Replacement Units

Fig. 32: Front PCB Card Cage


Pos. A34 A34 AVIF 2
Pos. A33 A33 ECPR
Pos. A32 A32 CFMS (SCL Systems only)
Pos. A31 A31 TRBF
Pos. A30 A30TXRX/A30-1 PBIF Assembly
Pos. A37-1 A37-1 MPXT (SCL systems only)

Front Shielding Cover 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the Front Shielding Cover

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).

SONOLINE G20 US06-102.841.01.02.02 Page 36 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 37

4. Remove the six upper screws and loosen the two lower screws that secure the front
shielding cover to the system and remove the cover.

Fig. 33: Front Shielding Cover Screw Locations


Pos. A Remove
Pos. B Loosen

5. Loosen the three lower mounting screws.

Install the Front Shielding Cover

1. Install the six upper screws and tighten the two lower screws that secure the front
shielding cover to the system and remove the cover.
2. Install the right-side panel (Install the Right-Side Panel / p. 25).
3. Install the left-side panel (Install the Left-Side Panel / p. 22).

A34 AVIF2 PCB 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Siemens US06-102.841.01.02.02 Page 37 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
38 Removing and Replacing Field Replacement Units

Remove the A34 AVIF2 PCB

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Disconnect all connectors from the I/O panel.
3. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
4. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
5. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
6. Disconnect J20 and J21 from the I/O panel.
7. Remove the ten screws that secure the I/O panel to the system.

Fig. 34: I/O panel Screw Locations


8. Press the PCB release tab and carefully slide the A34 AVIF2 PCB approximately one
inch out from the card cage; disconnect the small ribbon cable and big ribbon cable
from the A34 AVIF2 PCB.

CAUTION The two ribbon cables and their corresponding connectors are
fragile; the cables are connected to both the A34 AVIF2 PCB and
the A33 ECPR PCB.
¹ Pulling the A34 AVIF2 PCB too far or too hard prior to
disconnecting the cable from the A34 AVIF2 PCB could
break the small ribbon cables.

9. Remove the A34 AVIF2 PCB from the card cage.

Install the A34 AVIF2 PCB

1. Slide the A34 AVIF2 PCB most of the way into the A34 slot of the card cage; connect
the small ribbon cable to the A33 ECPR PCB and big ribbon cable to the A34 AVIF2
PCB, and then seat the PCB into the card cage.

SONOLINE G20 US06-102.841.01.02.02 Page 38 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 39

CAUTION The two ribbon cables and connectors are fragile; the cable are
connected to both the A34 AVIF2 PCB and the A33 ECPR PCB.
¹ Pulling the A34 AVIF2 PCB too far or too hard prior to
disconnecting the cable from the A34 AVIF2 PCB could
break the small ribbon cables.

2. Install the ten screws that secure the I/O panel to the system.
3. Connect J20 and J21 to the I/O panel.
4. Install the front shielding cover (Install the Front Shielding Cover / p. 37).
5. Install the right-side panel (Install the Right-Side Panel / p. 25).
6. Install the left-side panel (Install the Left-Side Panel / p. 22).
7. Connect all connectors to the I/O panel.

A33 ECPR PCB 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the A33 ECPR PCB

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Locate and remove the A33ECPR PCB.
a) Press the PCB release tab.
b) Carefully disconnect the small ribbon cable from the A34 AVIF2 PCB.

CAUTION The two ribbon cables and connectors are fragile; the cables are
connected to both the A34 AVIF2 PCB and the A33 ECPR PCB.
¹ Pulling the A34 AVIF2 PCB too far or too hard prior to
disconnecting the cable from the A34 AVIF2 PCB could
break the small ribbon cable.

c) Remove the A33 ECPR PCB from the card cage and place it in an ESD work area.

Siemens US06-102.841.01.02.02 Page 39 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
40 Removing and Replacing Field Replacement Units

6. Remove the A33-3 HDD from the A33 ECPR PCB that was removed from the system.
Note the routing of all cables removed from the two PCBs.
a) From the removed A33 ECPR PCB, disconnect the power cable from the A33-3
HDD.
b) Disconnect the ribbon cable from the A33-3 HDD.
c) Remove the four screws that secure the A33-3 HDD to the removed A33 ECPR
PCB and remove the A33-3 HDD from the A33 ECPR.
d) Disconnect power cable P8 and ribbon cable P10 from the removed A33 ECPR
PCB.
e) Remove the four screws that secure the A33-2 SBC PCB to the removed A33 ECPR
PCB.
f) Disconnect the A33-2 SBC from the SBCR at J1 and remove the A33-2 SBC PCB
with the cables.

Install the A33 ECPR PCB

1. Install the components and cables removed in the previous step to the new A33 ECPR
PCB.
a) Place the new A33 ECPR PCB in an ESD work area.
b) Position the A33-2 SBC PCB on the new A33 ECPR PCB.
c) Route the cables on the A33 ECPR PCB as they were routed prior to their removal.
d) Connect the A33-2 SBC to the SBCR J1.
e) Install and tighten the four screws that secure the A33-2 SBC PCB to the new A33
ECPR PCB.
f) Connect power cable P8 and ribbon cable P10 to the new A33 ECPR PCB.
g) Place the A33-3 HDD on the new A33 ECPR PCB; install and tighten the four
screws that secure the HDD to the new A33 ECPR PCB.
h) Connect the ribbon cable to the A33-3 HDD.
i) Connect the power cable to the A33-3 HDD.
2. Slide the A33 ECPR PCB into the A33 slot of the card cage; connect the small ribbon
cable to the A33 ECPR PCB and seat the PCB into the card cage.

CAUTION The two ribbon cables and connectors are fragile; the cables are
connected to both the A34 AVIF2 PCB and the A33 ECPR PCB.
¹ Pulling the A34 AVIF2 PCB too far or too hard prior to
disconnecting the cable from the A34 AVIF2 PCB could
break the small ribbon cable.

3. Install the front shielding cover (Install the Front Shielding Cover / p. 37).
4. Install the right-side panel (Install the Right-Side Panel / p. 25).
5. Install the left-side panel (Install the Left-Side Panel / p. 22).

SONOLINE G20 US06-102.841.01.02.02 Page 40 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 41

Hard Disk Drive (HDD) 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the Hard Disk Drive

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Remove the A33 ECPR PCB (Remove the A33 ECPR PCB / p. 39).

CAUTION The two ribbon cables and connectors are fragile; the cables are
connected to both the A34 AVIF2 PCB and the A33 ECPR PCB.
¹ Pulling the A34 AVIF2 PCB too far or too hard prior to
disconnecting the cable from the A34 AVIF2 PCB could
break the small ribbon cable.

Siemens US06-102.841.01.02.02 Page 41 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
42 Removing and Replacing Field Replacement Units

6. Disconnect the power cable from the HDD.

Fig. 35: HDD Mounted on the A33 ECPR PCB


Pos. A Power cable
Pos. B Ribbon cable
Pos. C Mounting screws

7. Disconnect the ribbon cable from the HDD.


8. Remove the four screws that secure the HDD to its mounting bracket.

Install the Hard Disk Drive

1. Place the HDD in its mounting bracket and install the four screws that secure the HDD
to its mounting bracket.
2. Connect the ribbon cable from the HDD.
3. Connect the power cable from the HDD.
4. Install the A33 ECPR PCB (Install the A33 ECPR PCB / p. 40).

CAUTION The two ribbon cables and connectors are fragile; the cables are
connected to both the A34 AVIF2 PCB and the A33 ECPR PCB.
¹ Pulling the A34 AVIF2 PCB too far or too hard prior to
disconnecting the cable from the A34 AVIF2 PCB could
break the small ribbon cable.

5. Install the front shielding cover (Install the Front Shielding Cover / p. 37).
6. Install the right-side panel (Install the Right-Side Panel / p. 25).
7. Install the left-side panel (Install the Left-Side Panel / p. 22).

SONOLINE G20 US06-102.841.01.02.02 Page 42 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 43

A33-2 SBC PCB 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the A33-2 SBC PCB

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Remove the A33 ECPR PCB (Remove the A33 ECPR PCB / p. 39).

CAUTION The two ribbon cables and connectors are fragile; the cables are
connected to both the A34 AVIF2 PCB and the A33 ECPR PCB.
¹ Pulling the A34 AVIF2 PCB too far or too hard prior to
disconnecting the cable from the A34 AVIF2 PCB could
break the small ribbon cable.

Siemens US06-102.841.01.02.02 Page 43 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
44 Removing and Replacing Field Replacement Units

6. Disconnect the following cables from the A33-2 SBC PCB:

Fig. 36: A33-2 SBC PCB Cable Connections


Pos. A Connection from A33 ECPR - P10
Pos. B Connection from the A33 ECPR - HDD
Pos. C Connection from the A33 ECPR - Power
Pos. D Connection from the fan - Do not remove
Pos. E Connection from the A34 AVIF2 PCB
Pos. F Connection from the A34 AVIF2 PCB
Pos. G A33-1 SBC memory card connection - Do not remove

7. Remove the four screws that secure the A33-2 SBC PCB to the A33 ECPR PCB and re-
move the A33-2 SBC PCB.

Install the A33-2 SBC PCB

1. Position the A33-2 SBC PCB on A33 ECPR PCB and install the four screws that secure
the A33-2 SBC PCB to the A33 ECPR PCB.

SONOLINE G20 US06-102.841.01.02.02 Page 44 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 45

2. Connect the following cables to the A33-2 SBC PCB:

Fig. 37: A33-2 SBC PCB Cable Connections


Pos. A Connection from A33 ECPR - P10
Pos. B Connection from the A33 ECPR - HDD
Pos. C Connection from the A33 ECPR - Power
Pos. D Connection from the fan - Do not remove
Pos. E Connection from the A34 AVIF2 PCB
Pos. F Connection from the A34 AVIF2 PCB
Pos. G A33-1 SBC memory card connection - Do not remove

3. Install the A33 ECPR PCB (Install the A33 ECPR PCB / p. 40) and reconnect the rib-
bon cables to the A34 AVIF2 PCB.

CAUTION The two ribbon cables and connectors are fragile; the cables are
connected to both the A34 AVIF2 PCB and the A33 ECPR PCB.
¹ Pulling the A34 AVIF2 PCB too far or too hard prior to
disconnecting the cable from the A34 AVIF2 PCB could
break the small ribbon cable.

4. Install the front shielding cover (Install the Front Shielding Cover / p. 37).
5. Install the right-side panel (Install the Right-Side Panel / p. 25).
6. Install the left-side panel (Install the Left-Side Panel / p. 22).

Siemens US06-102.841.01.02.02 Page 45 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
46 Removing and Replacing Field Replacement Units

A32 CFMS PCB (SLC Systems Only) 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the A32 CFMS PCB

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Press the PCB release tab and remove the PCB from the card cage.

Install the A32 CFMS PCB

1. Slide the A32 CFMS PCB into the A32 slot of the card cage and seat the PCB into the
card cage.
2. Install the front shielding cover (Install the Front Shielding Cover / p. 37).
3. Install the right-side panel (Install the Right-Side Panel / p. 25).
4. Install the left-side panel (Install the Left-Side Panel / p. 22).

A31 TRBF PCB 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the A31 TRBF PCB

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).

SONOLINE G20 US06-102.841.01.02.02 Page 46 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 47

4. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Press the PCB release tab and remove the PCB from the card cage.

Install the A31 TRBF PCB

1. Slide the A31 TRBF PCB into the A31 slot of the card cage and seat the PCB into the
card cage.
2. Install the front shielding cover (Install the Front Shielding Cover / p. 37).
3. Install the right-side panel (Install the Right-Side Panel / p. 25).
4. Install the left-side panel (Install the Left-Side Panel / p. 22).

A30 TXRX PCB and A30-1 PBIF PCB Assembly 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the A30 TXRX PCB/A30-1 PBIF PCB Assembly

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Remove the eight screws that secure the transducer ports to the system.

Fig. 38: Transducer Port Screw Locations


6. Press the PCB release tab and remove the PCB assembly from the card cage.

Siemens US06-102.841.01.02.02 Page 47 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
48 Removing and Replacing Field Replacement Units

7. Place the PCB assembly in an ESD work area and remove the eight screws that secure
the A30-1 PBIF PCB to the A30 TXRX PCB.

Fig. 39: A30 TXRX PCB/A31 TRBF PCB Assembly Screw Locations

Install the A30 TXRX PCB/A30-1 PBIF PCB Assembly

1. Place the PCB assembly in an ESD work area and install the eight screws that secure
the A30-1 PBIF PCB to the A30 TXRX PCB.
2. Slide the A30 TXRX PCB/A30-1 PBIF PCB assembly into A30 TXRX PCB slot of the
card cage and seat the PCB into the card cage.
3. Install the eight screws that secure the transducer ports to the system.
4. Install the front shielding cover (Install the Front Shielding Cover / p. 37).
5. Install the right-side panel (Install the Right-Side Panel / p. 25).
6. Install the left-side panel (Install the Left-Side Panel / p. 22).

SONOLINE G20 US06-102.841.01.02.02 Page 48 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 49

A37 MPXA PCB (SLC Systems only)

Fig. 40: A37 MPXA PCB


Pos. A Front Side
Pos. B Back Side

Remove the A37 MPXA PCB

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Remove the four screws that secure the sector port to the system.
6. Disconnect the sector port PCB from the A37-1 MPXT PCB; gently pull the A37 MPXA
PCB until the board disconnects form the A37-1 MPXT PCB.
Set the sector port in an ESD-safe area.

Install the A37 MPXA PCB

1. Connect the A37 MPXA PCB to the A37-1 MPXT PCB; gently press the A37 MPXA
PCB until the connection is secure.
2. Align mounting holes and install the four screws that secure the sector port to the sys-
tem.
3. Install the front shielding cover (Install the Front Shielding Cover / p. 37).
4. Install the right-side panel (Install the Right-Side Panel / p. 25).
5. Install the left-side panel (Install the Left-Side Panel / p. 22).

Siemens US06-102.841.01.02.02 Page 49 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
50 Removing and Replacing Field Replacement Units

A37-1 MPXT PCB (SLC Systems only)

Fig. 41: A37-1 MPXT PCB


Pos. A A37 MPXA PCB Interface
Pos. B A35 Mother Board Interface

Remove the A37-1 MPXT PCB

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Remove the A37 MPXA PCB (Remove the A37 MPXA PCB / p. 49).
6. To create sufficient work space in the card cage, remove the A30 TXRX PCB/A30-1
PBIF PCB assembly (Remove the A30 TXRX PCB/A30-1 PBIF PCB Assembly / p. 47).
7. Remove the five screws that secure the A37-1 MPXT PCB to the system frame.
8. Gently pull the A37-1 MPXT PCB until the board disconnects from the A35 mother
board, and then set the sector port in an ESD-safe area.

Install the A37-1 MPXT PCB

1. Plug the A37-1 MPXT PCB into the A35 mother board.
2. Install and tighten the five screws that secure the A37-1 MPXT PCB to the system
frame.
3. Install the A30 TXRX PCB/A30-1 PBIF PCB assembly (Install the A30 TXRX
PCB/A30-1 PBIF PCB Assembly / p. 48).
4. Install the A37 MPXA PCB (Install the A37 MPXA PCB / p. 49).
5. Install the front shielding cover (Install the Front Shielding Cover / p. 37).

SONOLINE G20 US06-102.841.01.02.02 Page 50 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 51

6. Install the right-side panel (Install the Right-Side Panel / p. 25).


7. Install the left-side panel (Install the Left-Side Panel / p. 22).

Siemens US06-102.841.01.02.02 Page 51 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
52 Removing and Replacing Field Replacement Units

Rear PCBs and Shielding Cover 0

Rear Shielding Cover 0

Remove the Rear Shielding Cover

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the upper rear shielding cover screws and loosen the lower shielding cover
screws and remove the rear shielding cover.

Fig. 42: Rear Shielding Cover Mounting Screw Locations


Pos. A Rear shielding cover upper mounting screw locations (remove)
Pos. B A36 RGLT PCB lower mounting screw locations (loosen)

Install the Rear Shielding Cover

1. Install the rear shielding cover; install the four upper mounting screws and tighten the
two lower mounting screws.

CAUTION Do not pinch any of the cabling when securing the rear shielding
cover.
¹ The system may not function properly when a wire is
pinched.

2. Install the right-side panel (Install the Right-Side Panel / p. 25).

SONOLINE G20 US06-102.841.01.02.02 Page 52 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 53

3. Install the left-side panel (Install the Left-Side Panel / p. 22).

A36 RGLT PCB and Rear Plane 0

Remove the A36 RGLT PCB

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the rear shielding cover (Remove the Rear Shielding Cover / p. 52).

Fig. 43: System Rear - Fastener Hole Locations


Pos. A A36 RGLT PCB (five mounting screw locations)
Pos. B Back side connectors P362 and P363

5. Disconnect P1/J1 from the A36 RGLT PCB.


6. Remove the five screws that secure the A36 RGLT PCB to the A35 MTHR PCB and re-
move the board from the A35 MTHR PCB.

NOTE The A36 RGLT PCB has two connectors (P362 and P363) on the
back of the board that connect to the A35 MTHR PCB.

Siemens US06-102.841.01.02.02 Page 53 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
54 Removing and Replacing Field Replacement Units

7. Place the A36 RGLT PCB in the ESD packaging in which the replacement part was
shipped.

Install the A36 RGLT PCB

1. Connect P362 and P363 to the A35 MTHR PCB; install and secure the five screws that
secure the A36 RGLT PCB to the back pane.
2. Connect P1 to J1 on the A36 RGLT PCB.
3. Install the rear shielding cover (Install the Rear Shielding Cover / p. 52).
4. Install the right-side panel (Install the Right-Side Panel / p. 25).
5. Install the left-side panel (Install the Left-Side Panel / p. 22).

Verify the Operation of the A36 RGLT PCB


The A36 RGLT PCB generates +B VA power for Tx and supplies power to the analog
block. The DAC voltage check is mounted on this board.

1. Check the DAC voltages.


- Check the power at the A36 RGLT test points.
2. Perform a scan with a transducer.
3. If the voltages are not correct check the following:
- Check the A36 RGLT PCB cable connections.
- Install the A36 RGLT PCB.
4. If there is a problem with the scanned image check the following:
- Check that the transducer is properly connected.
- Check the A36 RGLT PCB cable connections.
- Install the A36 RGLT PCB.

A35 MTHR PCB 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the A35 MTHR PCB

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).

SONOLINE G20 US06-102.841.01.02.02 Page 54 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 55

3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).


4. Remove the front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Remove the rear shielding cover (Remove the Rear Shielding Cover / p. 52).
6. Remove the A36 RGLT PCB and place the board in an ESD work area.
7. Disconnect the CD-RW drive power cable from the A35 MTHR PCB.
8. Remove the front PCBs (Front PCBs, Shielding Cover and Hard Disk Drive
(HDD) / p. 35).
9. Disconnect P351/J351 from the A35 MTHR PCB.
10. Remove the sixteen screws that secure the A35 MTHR PCB cover to the A35 MTHR
PCB and remove the cover.

Fig. 44: A35 MTHR PCB Cover Screw Locations


11. Remove the A35 MTHR PCB.

Fig. 45: A35 MTHR PCB

Siemens US06-102.841.01.02.02 Page 55 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
56 Removing and Replacing Field Replacement Units

Install the A35 MTHR Board

1. Align the A35 MTHR PCB mounting holes with the mounting holes on the system and
secure the A35 MTHR PCB to the system.
2. Align the A35 MTHR PCB cover mounting holes with the A35 MTHR PCB mounting
holes and secure the sixteen screws that attach the A35 MTHR PCB cover to the A35
MTHR PCB.
3. Install the A36 RGLT PCB (Install the A36 RGLT PCB / p. 54).
4. Connect the CD-RW drive power cable.
5. Connect J351 to P351 of the A35 PCB.
6. Connect P1 to J1 of the A36 RGLT PCB.
7. Install the front PCBs and the front shield cover (Front PCBs, Shielding Cover and Hard
Disk Drive (HDD) / p. 35)
8. Install the rear shielding cover (Install the Rear Shielding Cover / p. 52).
9. Install the right-side panel (Install the Right-Side Panel / p. 25).
10. Install the left-side panel (Install the Left-Side Panel / p. 22).

SONOLINE G20 US06-102.841.01.02.02 Page 56 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 57

Switching Regulator (SWRG2) 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the SWRG2 0

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove front shielding cover(Remove the Front Shielding Cover / p. 36).
5. Remove the rear shielding cover (Remove the Rear Shielding Cover / p. 52).
6. Remove the two screws that secure the SWRG2 to the front of the system.

Fig. 46: Front SWRG2 Mounting Screw Location


7. Remove the two screws that secure the rear of the SWRG2 to the rear of the system.

Fig. 47: Rear SWRG2 Mounting Screw Location


8. Disconnect P1/J1 and P2/J2 on the rear of the SWRG2.

Siemens US06-102.841.01.02.02 Page 57 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
58 Removing and Replacing Field Replacement Units

9. Remove the SWRG2 from the rear of the system.


- Carefully slide the SWRG2 out until two of the three cable clamps are exposed.

CAUTION Do not allow the SWRG2 to contact the components mounted be-
low the SWRG2.
¹ If the SWRG2 contacts any of these components the
components could become damaged.

- Open the two cable clamps and free the cable bundle from the two clamps.
- Slide the SWRG2 until the third cable is exposed; open the clamp and free the cable
bundle.
- Disconnect the cable bundle from the front of the SWRG2 (P4/J4, P7/J7, P3/J3, and
P6/J6).

Install the SWRG2 0

1. Install the SWRG2 on the system.


- Connect the cable bundle to the front of the SWRG2 (P4/J4, P7/J7, P3/J3, and
P6/J6).
- Route the cable bundle through the cable clamp closest to the front of the SWRG2
and secure the cable at the front of the SWRG2.
- Carefully slide the SWRG2 approximately half way into the system; route the cable
through the remaining two clamps and secure the cable clamps.
- Slide the SWRG2 the all the way into the system.
2. Connect P1/J1 and P2/J2 at the rear of the SWRG2.
3. Install the two screws that secure the rear of the SWRG2 to the rear of the system
4. Install the two screws that secure the front of the SWRG2 to the front of the system.
5. Install the rear shielding cover (Install the Rear Shielding Cover / p. 52).
6. Install front shielding cover (Install the Front Shielding Cover / p. 37).
7. Install right-side panel (Install the Right-Side Panel / p. 25).
8. Install left-side panel (Install the Left-Side Panel / p. 22).

SWRG2 Fuse 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

SONOLINE G20 US06-102.841.01.02.02 Page 58 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 59

Remove the SWRG2 Fuse

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove the rear shielding cover (Remove the Rear Shielding Cover / p. 52).
5. Remove the fuse from the rear of the SWRG2.

Install the SWRG2 Fuse

1. Install the fuse in the fuse clip on the rear of the SWRG2.
2. Install the rear shielding cover (Install the Rear Shielding Cover / p. 52).
3. Install the right-side panel (Install the Right-Side Panel / p. 25).
4. Install the left-side panel (Install the Left-Side Panel / p. 22).
5. Connect the power cord to the rear of the system.

SWRG2 Fan 0

Remove the SWRG2 Fan

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the left-side panel (Remove the Left-Side Panel / p. 21).
3. Remove the right-side panel (Remove the Right-Side Panel / p. 23).
4. Remove front shielding cover (Remove the Front Shielding Cover / p. 36).
5. Remove the rear shielding cover (Remove the Rear Shielding Cover / p. 52).
6. Remove the SWRG2.
7. Disconnect the cable from the fan.
8. Remove the four screws that secure the fan to the top of the SWRG2.
9. Remove the fan.

Install the SWRG2 Fan

1. Install the fan.


2. Install the four screws that secure the fan to the top of the SWRG2.
3. Connect the cable to the fan.
4. Install the SWRG2.
5. Install the rear shielding cover (Install the Rear Shielding Cover / p. 52).

Siemens US06-102.841.01.02.02 Page 59 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
60 Removing and Replacing Field Replacement Units

6. Install front shielding cover (Install the Front Shielding Cover / p. 37).
7. Install the right-side panel (Install the Right-Side Panel / p. 25).
8. Install the left-side panel (Install the Left-Side Panel / p. 22).

SONOLINE G20 US06-102.841.01.02.02 Page 60 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.
Removing and Replacing Field Replacement Units 61

Line Filter 0

CAUTION The G20 system contains components that are sensitive to dam-
age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD pro-
tective equipment including an ESD wrist strap when the
protective covers are removed from the system.

Remove the Line Filter 0

1. Power off the system, switch off the MAINS power circuit breaker, and disconnect the
power cord from the wall outlet.
2. Remove the two screws that secure the line filter to the system.

Fig. 48: Line Filter Mounting Screw Location


3. Carefully pull the line filter away from the system; the line filter is still connected to the
system by its cables.
4. Disconnect the line filter cable (P1/J1) from the A07 PCB.

Install the Line Filter 0

1. Connect the line filter cable (P1/J1) to the A07 PCB.


2. Align the line filter mounting holes with the system mounting holes and install the two
screws that secure the line filter to the system.

Siemens US06-102.841.01.02.02 Page 61 of 62 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
62 Removing and Replacing Field Replacement Units

SONOLINE G20 US06-102.841.01.02.02 Page 62 of 62 Siemens


10.05 sd Medical Solutions USA, Inc.