Beruflich Dokumente
Kultur Dokumente
Daily Guide
Ref: RAB275CEN
Daily Guide
Daily Guide
Ref: RAB275CEN
Contents
1. Revisions..................................................................................................1
3. Safety Precautions...............................................................................3
3.1. Warning and Precautions..............................................................................3
3.2. Environment................................................................................................3
3.3. Labels and Connections...............................................................................4
3.4. Power Requirements....................................................................................6
3.5. Electronic and Moving Parts..........................................................................6
3.6. Biological Risk.............................................................................................7
4. Workflow..................................................................................................8
4.1. Sample Materials.........................................................................................8
4.2. Start of Day.................................................................................................8
4.3. Reagent Cassette Management...................................................................10
4.4. Calibration and Control...............................................................................12
4.5. Patient Samples.........................................................................................17
4.6. End of Day................................................................................................22
4.7. To Stop in Emergency During Analysis.........................................................26
5. Flags........................................................................................................27
5.1. Error Flags.................................................................................................27
5.2. Error Code for ISE......................................................................................29
Daily Guide i
Ref: RAB275CEN
Daily Guide ii
Ref: RAB275CEN
Revisions
1. Revisions
This document applies to the latest software version listed and higher versions.
When a subsequent software version changes the information in this document, a new
electronic issue (CD-ROM and/or online help) is released and supplied by
HORIBA Medical.
To update a paper document, please contact your local HORIBA Medical representative.
If a problem occurs during the instrument operation, refer to the user manual
instructions for recommendations. If the problem persists, please contact your
local HORIBA Medical representative.
Daily Guide 1
Ref: RAB275CEN
Software Icons Description
Start: to run patient's sample, calibration or control. Shutdown: to open Shutdown window.
STAT: to run an emergency sample. Print screen: to print data / to send data to the host.
MSDS: to open the reagent online help. Help: to open the Pentra C200 online help.
Audible Alarm: to stop the sound alarm in progress. Keyboard: to display Virtual Keyboard screen.
Instrument status
Daily Guide 2
Ref: RAB275CEN
Safety Precautions
Warning and Precautions
3. Safety Precautions
3.2. Environment
3.1. Warning and Precautions
The operation of the Pentra C200 should be restricted to indoor location use only!
Instrument is operational at an altitude of maximum 2000 m (6562 ft).
Work safety reliability and general characteristics are guaranteed by HORIBA Medical
under the following conditions: The instrument is designed for safety from voltages surges according to INSTALLATION
CATEGORY II and POLLUTION DEGREE 2 (IEC 61010-1).
■ User manual must be entirely read, and personnel trained by a HORIBA Medical
representative before attempting to operate the instrument. Please contact your local representative for information regarding operation locations,
■ The user always operates with full knowledge and appreciation of instrument when it does not comply with the recommended specifications.
warnings and alarms.
■ Always refer to labelling and HORIBA Medical instructions in order to avoid to
compromise system integrity.
This instrument must be operated as instructed in the user manual. Any other use might
compromise system integrity and might be hazardous for the operator.
This instrument complies with Standards and Directives named in the Declaration of
Conformity. The latest version of the Declaration of Conformity for this instrument is
available online at www.horiba.com.
Daily Guide 3
Ref: RAB275CEN
Safety Precautions
Labels and Connections
The Power switch and Power supply connection should always be accessible!
When positioning the system for operational use, leave the required amount of
space for easy accessibility to these items.
A color code is used for connections, tubing and tank identification. Blue color is used
for distilled water whereas red is used for waste.
Daily Guide 4
Ref: RAB275CEN
Safety Precautions
Labels and Connections
■ PFA tube for water supply (2) needs to be tightened by hand without any
tool.
■ Water tank needs to be kept at the instrument level and silicon tube (3) is
cut to be as short as possible between connector and tank.
■ Waste tank needs to be located below the instrument and loops must be
avoided between connector and tank. Cut the silicon waste water tube to
avoid loops. Whenever it is possible, leave USB port n°4 available for technical operations.
If it is not possible, use the USB port n°4 for mouse (option) or keyboard (option).
ISE concentrated waste can be separated from other non concentrated waste
for special disposal of contaminated liquid.
3.3.5. Supplied USB Key
The USB key is preloaded with the relevant data pertinent to the system.
Dispose of waste according to your local/national guidelines for biohazard
waste disposal.
Daily Guide 5
Ref: RAB275CEN
Safety Precautions
Power Requirements
3.3.6. Warnings, Cautions and Biological Risks Labels Labels Warning Location
Risk of injury.
Turn the power OFF before
Halogen lamp cover, pump Fuses characteristics:
unit panel, ISE module panel
opening this panel.
Type Quantity Characteristics
Warning, hot surface.
Ceramic tube fuse 2 6.3 A / 250 V
Risk of personal injury due to hot
surfaces within the instrument. Detector unit
Wait for the lamp to cool down
before proceeding.
General warning, caution, risk of
danger.
Risk of personal injury during the
reagent and sample container 3.5. Electronic and Moving Parts
Sample and reagent
unit cover operation (hands/
container unit, cuvette
fingers). Ensure that the unit
loading unit, reaction unit
cover closure is done cautiously.
Risk of personal injury during
cuvette loading unit operation The following parts must not be handled or checked by the user:
(hands/ fingers). ■ Electrical Power supply
■ Electronic circuit boards
Daily Guide 6
Ref: RAB275CEN
Safety Precautions
Biological Risk
Troubleshooting > Maintenance > Other Procedures > To Decontaminate the Instrument
Operator injury may occur from an electric shock. Electronic components can chapter of the user manual to perform the instrument cleaning and decontamination
shock and injure the user. Do not dismantle the instrument nor remove any procedure.
components (covers, doors, panels and so on) unless otherwise instructed
within this document.
Danger of explosion if battery is not replaced correctly! When replacing the
battery, always use the same and/or equivalent type recommended by the
manufacturer. Dispose of used batteries according to the manufacturer's
specific instructions.
The manufacturer uses disinfectant product for instrument decontamination and highly
recommends it to decontaminate your instrument. See Maintenance and
Daily Guide 7
Ref: RAB275CEN
Workflow
Sample Materials
4. Workflow
At the beginning of a day of work, the instrument wakes from sleeping mode.
4.1. Sample Materials When the instrument is in sleeping mode, the green LED on the right of the screen blinks
and the screen is black. During automatic wake up the instrument performs initialization.
When the preparation process is finished the status changes from "START UP" to
"READY" and the pop-up disappears.
Samples such as serum, plasma, urine and homogeneous liquid are to be used on the
Pentra C200.
Ensure that a sufficient sample volume without air bubble or foam is placed on 4.2.1. To Log on the Instrument
the instrument whatever the sample type (calibrator, control, patient sample).
A pop-up with several choices (Sleep, Power Off, Logout, Restart, Cancel) is
displayed.
2. Press Logout.
A pop-up is displayed to confirm system logout.
Daily Guide 8
Ref: RAB275CEN
Workflow
Start of Day
3. Press OK to confirm system logout or Cancel to cancel the ordered logout. 7. Press Login.
A pop-up is displayed for user confirmation.
8. Press OK.
9. Press the Back button in the generic toolbar to enter the main screen.
Daily Guide 9
Ref: RAB275CEN
Workflow
Reagent Cassette Management
4.3. Reagent Cassette Management 3. Check reagents for the tests you intend to run on the instrument.
Total displays the total number of tests that can be performed on the instrument with
the reagents registered on board.
The Inventory tab allows you to check reagent availability and volumes. Volumes are 4. Check the Inner (R2) tab.
deduced from the last use of reagent cassettes. The number of tests available is in
accordance with the method configured with the reagent. For new cassettes, theoretical
values are displayed and actualized after the first pipetting.
Daily Guide 10
Ref: RAB275CEN
Workflow
Reagent Cassette Management
6. Put back the reagent and sample trays in the SRCU by inserting the two guide pins Access: Inventory > Reagent Tray (tab)
in the two bigger holes of the reagent tray.
7. Remove the caps from the cassettes.
8. Press on the center of the trays to make sure the trays are correctly placed.
9. If necessary, remove bubbles with a pasteur pipette.
10. Put back the SRCU cover in correct position.
11. Check that the probe has no obstacle on its way.
12. Close the main cover.
Do not forget to register the cassettes after their loading (see Workflow >
Reagent Cassette Management > Cassette Registration > To Register a
Cassette chapter).
To get information about a cassette, press the concerned circle on the tray
representation and information about the cassette compartment is displayed in the
cassette information area.
Daily Guide 11
Ref: RAB275CEN
Workflow
Calibration and Control
4.3.3.2. To Register a Cassette 2. Check the identification of the selected cassette on the cassette identification area.
3. Press Delete.
Access: Inventory > Reagent Tray (tab)
4. Press Save to validate the deletion.
A pop-up is displayed and allows you to confirm or to cancel the deletion.
■ To proceed to cassette registration, reagents must already be registered on the
instrument ( Main menu > Services > System configuration > Reagent (tab) ),
All the cassettes registered on the tray can be deleted by pressing Delete
■ Reagents provided by HORIBA Medical are already registered on the All button.
instrument,
■ Only reagents for open channel must be registered by the user,
■ Only level 2 users (L2) are allowed to register new reagents.
■ Open cassettes must be registered on the instrument (see User Manual: Workflow >
Reagent Cassette Management > To Register an Open Cassette chapter for open
cassette registration),
■ Reagent cassettes must be loaded on the instrument.
4.4. Calibration and Control
1. Press Barcode Scan to register new cassettes.
The instrument checks the cassettes available on the instrument.
2. Press Save to save cassette registration. Before running any sample, the user needs to check daily calibration and
control validity for the concerned tests.
If the instrument encounters a barcode scan problem for a cassette, follow the
procedure described in the troubleshooting chapter of the user manual. See
Maintenance and Troubleshooting > Troubleshooting > Undetected Cassette
chapter. 4.4.1. To Calibrate ISE
1. Select the cassette position by presssing the corresponding circle on the tray 2. Remove the Sample and Reagent Container Unit (SRCU) cover.
representation. 3. Place Standard 2 solution (2 mL) for ISE calibration in a tube.
Daily Guide 12
Ref: RAB275CEN
Workflow
Calibration and Control
4. Load the tube at the first position on the sample tray. 4.4.2. Check of Calibration and Control Status
5. Close the SRCU cover.
6. Close the main cover.
7. In Main menu > Order > Selection (tab) , register ISE calibration: 4.4.2.1. Test Color Code
a. Select position 1.
b. Select ISE Calibration from the Type dropdown list. Access: Main menu > Order > Selection (tab)
c. Press Save.
All tests are displayed on the Selection screen. According to the sample category
selected, they are associated with a color code corresponding to the calibration and
8. In Main menu > Services > Customer Services > Sequence (tab) , press ISE control status.
Calibration button.
Daily Guide 13
Ref: RAB275CEN
Workflow
Calibration and Control
Daily Guide 14
Ref: RAB275CEN
Workflow
Calibration and Control
Daily Guide 15
Ref: RAB275CEN
Workflow
Calibration and Control
The instrument displays the start time and the estimated finish time for
If not enough cuvettes on board, the message "Shortage of cuvette or measurements in the main screen.
disposal space for all orders. Continue the measurement?" is displayed.
The following pop-up message may be displayed: "Conduct DTR automatic 4.4.4. To Check Calibration and Control Results
measurement?".
During measurements and when run is complete, the Run screen displays calibration
The autogain measurement of the spectrophotometer is only performed for and control results.
the first measurement of the day.
1. Check for calibrations validity in the Calibration tab.
■ If an error is displayed for a calibration, the calibration and the controls
A pop-up is displayed: "Preparing for measurement start". associated must be rerun.
■ If no error is displayed, check for the controls validity.
Daily Guide 16
Ref: RAB275CEN
Workflow
Patient Samples
If necessary, register patient data in the Patient screen before sample registration ( Main
menu > Order > Patient (tab) ).
1. Select an available position in the worklist ( Main menu > Order > Selection (tab) ).
4.5.1. To Order Patient Samples
2. Select Normal in the Type dropdown list.
3. Select the cup type (Normal or Biocup) from the Cup dropdown list.
Access: Main menu > Order > Selection (tab) When a sample is registered, the instrument automatically gives a SID corresponding
to the sample position on the tray.
1. Open the main cover.
2. Open the Sample and Reagent Container Unit (SRCU) cover. 4. If using a SID, type the SID in the SID field (up to 12 digits).
Daily Guide 17
Ref: RAB275CEN
Workflow
Patient Samples
■ Ensure that a sufficient sample volume without air bubble or foam is ■ Samples registered by this method stay in memory from run to run as long
placed on the instrument (refer to Workflow > Sample Materials chapter). as they are not run.
■ Caps must be removed from the sample cups as they may hit the ■ When you start the run, the instrument scans the tubes for barcode
Sample and Reagent Container Unit (SRCU) cover. identification and assigns the corresponding positions on the sample tray
to each tube.
If necessary, register patient data in the Patient screen before sample registration ( Main
menu > Order > Patient (tab) ).
4.5.1.3. To Register Sample with Host Connection (L.I.S)
1. Select the empty position in the Start Pos field (top of the dropdown list).
2. Select Normal in the Type dropdown list.
Access: Main menu > Order > Selection (tab)
3. Select the cup type Normal from the Cup dropdown list.
4. Type the barcode in the SID field. The host communication mode is set in the Host Connection screen ( Main menu >
5. If using a PID, select the corresponding patient in the PID field. Services > System Configuration > Host Connection (tab) ). Three communication
options are available:
6. Select the sample category (Common, Serum, Urine, Plasma).
■ Off Line
7. Select patient range.
■ On Line Batch
8. Choose tests to run for the selected sample by checking the corresponding tests. ■ On Line Query
9. Save sample registration.
The sample registered appears in the worklist below the STAT position 16 (E) at the 1. If the host communication mode is set to On Line Batch:
bottom of the list. a. If starting a day of work, press Acquire Orders from the Selection screen. The
next orders during the day are automatically acquired by the system.
The worklist registered on the laboratory information system is loaded or updated
10. Repeat the procedure for each sample to register.
on the instrument.
11. Load the samples on the sample tray.
b. Load the samples on the sample tray.
■ Ensure that a sufficient sample volume without air bubble or foam is
placed on the instrument (refer to Workflow > Sample Materials chapter). 2. If the host communication mode is set to On Line Query, load the samples on the
■ Caps must be removed from the sample cups as they may hit the sample tray.
Sample and Reagent Container Unit (SRCU) cover.
Daily Guide 18
Ref: RAB275CEN
Workflow
Patient Samples
When a sample tube is detected on board, the instrument automatically checks for A pop-up is displayed: "Preparing for measurement start".
orders registered on the laboratory information system and loads orders concerning
the tube on the instrument.
■ If a cassette has been removed or if a new reagent lot number has been
loaded since the last barcode scan, the following pop-up message is
displayed: "The reagents placed on the SRCU do not match with the
■ Ensure that a sufficient sample volume without air bubble or foam is registered information. Start measurement?".
placed on the instrument (refer to Workflow > Sample Materials chapter).
■ Caps must be removed from the sample cups as they may hit the Sample ■ If a cassette has been removed, the pop-up message displays
and Reagent Container Unit (SRCU) cover. "Cassette removed" and indicates the cassette position and the
reagent name. Press Cancel, put back the cassette on the reagent tray
and press Start button for more than two seconds.
■ If a new reagent lot number has been loaded, the pop-up message
displays "Lot number different for" and indicates the cassette position
4.5.1.4. To Start Sample Measurement and the reagent name. Press Cancel, delete the worklist, run a new
calibration and controls for the test before ordering the worklist again.
■ If not enough reagent to perform the ordered tests, the instrument displays
Before you start measurements, the following conditions must be fulfilled: the following pop-up message: "Reagent inventory may be insufficient to
■ Enough new cuvettes must be loaded on the instrument. process all test ordered, continue measurement?".
■ Dust pod must allow to store cuvettes used during measurements.
■ Reagents, diluents and wash solutions are on board and enough for ordered
measurements. The instrument starts measurement and the status "ANALYSING" is displayed.
■ Check that the water tank is full, if not fill it up with distilled water. The instrument displays the start time and the estimated finish time for measurements in
■ Check that the water tubing end stands in the bottom of the water tank. the main screen.
■ Check that the waste tank is empty, if not empty it.
■ The cover of the Sample and Reagent Container Unit (SRCU) must be closed.
■ The instrument main cover must be closed.
■ The STAT unit must be closed.
4.5.2. To Run Emergency Samples
Press Start button for more than two seconds.
STAT position on the left-hand side of the instrument is dedicated to emergency samples.
Daily Guide 19
Ref: RAB275CEN
Workflow
Patient Samples
For the STAT position, sample height within the tube must be higher than 12.6 As soon as the STAT button is available, you are able to run a new
mm. emergency sample.
Before replacing the emergency sample, check the STAT sample status
from the main screen.
This action enables you to open the STAT unit. Open the STAT unit when the status
changes from "ANALYSING" to "WAITING". ■ If this one is available (green), you can replace it immediately and
without condition.
2. Open the STAT unit by pushing slightly on the STAT cover on the left-hand side of ■ If this one is still in progress (orange), you can replace it now but if an
the instrument. automatic rerun is needed, it will not be performed!
3. Insert the emergency sample (sample tube or sample cup with adaptor) and push the
STAT unit back into the instrument.
4. Select position 16 from the Start Pos dropdown list and Normal from the Type a. Press the STAT button from the generic toolbar.
dropdown list in the Selection screen.
5. Select the cup type (Normal or Biocup) from the Cup dropdown list.
6. If using a SID, type the SID in the SID field (up to 12 digits).
7. If using a patient ID, select the corresponding patient in the PID field.
This action enables you to open the STAT unit.
8. Select the sample category (Common, Serum, Urine, Plasma).
9. If necessary, select the patient range. b. Select the Replace check box in the Selection screen.
10. Choose tests to run for the selected sample by checking the corresponding tests. c. Repeat steps from 2 to 12 to replace and to run the emergency sample.
11. Press Save to save emergency sample registration.
12. Press Start for more than two seconds to restart measurements.
The emergency sample requested is run in priority before any remaining samples
from the worklist.
Daily Guide 20
Ref: RAB275CEN
Workflow
Patient Samples
4.5.3. To Check Results 4.5.3.2. To Check or To Print Results from the Results Screen
Results are displayed in the Run screen as soon as the measurement is done.
1. Check the results from the run in the Patient tab from the Run screen.
Only on this screen, the instrument adds to the SID of each sample a prefix
corresponding to the sample type and the cup type (N for normal cups, n
for biocups, E for emergency samples and R for replicates).
2. If results are flagged, please refer to Workflow > Patient Sample > To Check Results
> To Manage Flagged Results chapter.
■ Results from the run are automatically transferred to the Results screen.
■ When a new run is started, results from the previous run are deleted from
1 = SID/PID dropdown list
the Run screen.
2 = SID/PID fields
5 = Search button
Daily Guide 21
Ref: RAB275CEN
Workflow
End of Day
If more than 2000 items are available for the specified search, a pop-up is
displayed: "Too many results meet search condition. Narrow the condition
and try again." Press OK and narrow the search specifications.
To print standard report, first select Normal from the Sample Type
dropdown list, then select Standard report from the Result Output 4.6.1. To Empty the Worklist
dropdown list.
8. If results are flagged, please refer to Workflow > Patient Sample > To Check Results After a day of work, before performing the instrument cleaning, the worklist needs to be
> To Manage Flagged Results chapter. deleted and the samples removed from the Sample and Reagent Container Unit (SRCU).
Daily Guide 22
Ref: RAB275CEN
Workflow
End of Day
Daily Guide 23
Ref: RAB275CEN
Workflow
End of Day
According to your geographical location, the SRCU may have condensation and may
4.6.3. To Check System Solutions need to be wiped after every day of work.
To follow the SRCU cleaning procedure, please refer to the maintenance chapter of the
user manual. See Maintenance and Troubleshooting > Maintenance > Daily Procedures >
To Clean Sample and Reagent Container Unit chapter.
1. Check that the water tank is full, if not fill it up with distilled water.
2. Check that the water tubing end stands in the bottom of the water tank.
3. Check that the waste tank is empty, if not empty it.
Daily Guide 24
Ref: RAB275CEN
Workflow
End of Day
4.6.6. Sleeping Mode 2. Press the Shutdown button in the generic toolbar at the top of the screen.
Daily Guide 25
Ref: RAB275CEN
Workflow
To Stop in Emergency During Analysis
4. Press OK to confirm wake up from the pop-up: "Conduct the cancel of sleep ■ A pop-up with several choices (Power Off, Restart, Cancel) is displayed.
operation?".
The instrument performs an initialization.
2. Press Cancel.
The door is unlocked and can be opened by the user.
1 = Green LED
2 = Red LED
Daily Guide 26
Ref: RAB275CEN
Flags
Error Flags
5. Flags
Daily Guide 27
Ref: RAB275CEN
Flags
Error Flags
Error flag Description Error Cause Error flag Description Error Cause
Volume Zero in R2 wash The remaining volume of R2 wash solution for IRU temperature extremely
R2WS TE2 IRU temperature > 37+2(°C).
solution method-to-method wash has reached zero. high
DB No dilution bottle The dilution bottle has not been registered. RCU temperature extremely
TE3 RCU temperature > 15°C.
The remaining volume of the dilution bottle has high
DS Volume Zero in dilution bottle
reached zero. RCU temperature extremely
TE4 RCU temperature < 2°C.
Wash solution bottle for PT wash has not been low
WB No wash solution bottle
registered.
WS
Volume Zero in wash solution The remaining volume of the wash solution bottle Error rank 4
bottle for PT wash has reached zero.
The reagent lot number used is not as previous lot Error flag Description Error Cause
LOT Lot inconsistency
recorded. EXP The reagent is no longer valid. An expired reagent was used in the measurement.
PT wash has been failed. Check the alarm to Invalid reagent stability was used with the
SPW Failure of PT wash (SPT) STB Reagent stability expired
determine the reason for its fail. measurement.
Method-to-method washing has failed prior to a The reason is that the measurement was taken by
R1W Failure of PT wash (RPT1) R1 operation.. Check the alarm to determine the CTO Calibration expired a method of which calibration has not been
reason for its fail. implemented within a setting term.
Method-to-method washing has failed, prior to a The reason is that the measurement was taken by
R2W Failure of PT wash (RPT2) R2 operation.. Check the alarm to determine the CXP Control expired a method of which QC measurement has not been
reason for its fail. implemented within a setting term.
DWPB No DET.W bottle for PT DET.W bottle is not placed beside PT unit. Q Control error Executed the measurement with control error.
Volume Zero in DET.W The remaining volume of DET.W for PT has been An error of concentrations of No calibration curve, or ISE calibration ever has
DWPS CA?
solution for PT zero. conversions. been executed after the startup.
DWMB No DET.W bottle for MIX DET.W bottle is not placed beside MIX unit.
The remaining count of MIX washing by DET.W Error rank 5
EUT Effective count Zero of DET.W
has been zero.
Cuvettes in the cuvette rack are insufficient for Error flag Description Error Cause
CS Cuvette shortage
measurement.
An error out of the range of The value is not between Calibrator’s minimum
DF Full cuvette in DP DP full up with used cuvettes. OVR
calibrator. and maximum.
EST Anomalous measurement Error sampling stop. Check the alarm for the reason.
Error rank 6
Error rank 3
Error flag Description Error Cause
Error flag Description Error Cause LIN Linearity Limit Error Linearity Limit Error.
TE1 IRU temperature extremely low IRU temperature < 37-2(°C). PRO Prozone Limit Error Prozone Limit Error.
Daily Guide 28
Ref: RAB275CEN
Flags
Error Code for ISE
Error flag Description Error Cause 5.2. Error Code for ISE
The flag "AB1" is displayed when all measuring
points exceed the absorbance limit except the first
AB1 Absorbance Limit1 Error
one. No result is displayed (*****) and a rerun with a
post-dilution is performed.
When an error occurs during ISE calibration, an error is displayed in the Result of ISE
The flag "AB2" is displayed when more than 1 Calibration screen. The table below displays the error codes with their description.
point and less than 5 points are within the
AB2 Absorbance Limit2 Error absorbance limit. The first result is then calculated The error code for ISE module is a four-digit code:
with those points and a rerun with post-dilution is
performed. ■ 0XXX: electrode errors
■ 1XXX: ISE module errors
Error rank 7
Electrode errors
Error flag Description Error Cause ■ The first digit is 0 (zero) for electrode errors.
Used reagent lot number in the specimen or ■ The second digit is the error code for Na electrode.
Using the different lot number ■ The third digit is the error code for K electrode.
CLT control measurement has been different from the
of R1 or R2
reagent in the calibration measurement. ■ The fourth digit is the error code for Cl electrode.
The measurement was taken by sample type of
each method that does not match the sample type. Error code Description
STM Sample type Inconsistent The concentration value was converted with a 1 2-point calibration: response alert.
calibration curve that does not match the sample
type. 2 2-point calibration: stability alert.
3 2-point calibration: reference potential alert.
Daily Guide 29
Ref: RAB275CEN
Flags
Error Code for ISE
Daily Guide 30
Ref: RAB275CEN