Scope: This test method covers the measurement of refractive indexes, accurate to six units in the fifth decimal place,
and refractive dispersions, accurate to twelve units in the fifth decimal place, of transparent and light-colored
hydrocarbon liquids that have refractive indexes in the range from 1.33 to 1.50, and at temperatures from
+20 to +30 °C.
The test method is not applicable within the accuracy stated to liquids having colors darker than No. 4 ASTM
Color as determined by Test Method D 1500, to liquids having bubble points so near the test temperature that
a reading cannot be obtained before substantial weathering takes place, to liquids having a refractive index
above 1.50, or to measurements made at temperatures above +30 °C.
Refractometry has been a well-proven method to determine The typical accuracy for Abbe-type refractometers is in the third
the concentration of liquid mixtures or their purity or quality decimal place with the fourth decimal place estimated.
characteristics. This can either be identification or quality control.
Page: 1
28.07.2010
13-0285.pdf
data & photo not mandatory
Main Unit
13-0285 High Precision Modified Abbe Refractometer
ASTM D 1218 - ASTM D 1747 - DIN 51 423-2 -
ISO 5661 - JIS C2101 (ASTM D 2159)
Modified Abbe-type refractometer for refractive index measurements nD = 1.3 to 1.74 with evenly
divided scale and micrometer screw.
The refractive index is found by use of a conversion table. Average accuracy throughout the range is
0.00004.
Consisting of:
1 conversion table for refractive indexes related to sodium-D light (589.6 nm). Prisms with hose
nozzles for a thermostatic circulator, built-in temperature probe and digital temperature display, scale
illumination by LED, prism illumination by spectral light.
Supplied with:
1 scale lamp outfit and 1 sodium spectral lamp, traceable test plate (certificate, however, subject to
particular order), bottle of contact liquid (monobromonaphthalene) and software for conversion of the
scale readings to refractive index or Brix.
Consisting of:
Cooler with stainless steel bath.Thermostat with digital display, splash-proof keypad, integrated heater,
pump, RS232-interface, over/under-temperature & low-level protection, adjustable pump-pressure,
cooling control.
Technical Data:
Working range: -28 to +200 °C (at +5 to +40 °C ambient temp.)
Display / Resolution: LED / 0.1 °C
Control: PID
Heater capacity: 2000 W
Cooling capacity: 260 W at +20 °C
Pump-Pressure capacity: 11-16 l/min, 450 mbar
Bath opening: 12 x 14 x 14 cm (WxDxH)
Safety class / Bath fluid: 3 / FL (Flammable Liquids possible)
Volume: 4.5 l
Sockets for: hose (Ø 8 / 10 mm), solenoid valve
Dimensions / Weight: 23 x 42 x 61 cm (WxDxH), 31 kg
Power supply: 230 V, 50/60 Hz, EU-plug
Page: 2
28.07.2010
13-0285.pdf
data & photo not mandatory
13-1505 PC-Software - ABBE (Windows®)
to convert refractometer readings to wavelengths
other than 589.6 nm
Note!
For free submission or download from "MyPetrotest"
please contact: service@petrotest.com
Spare Parts
13-1523 Contact Fluid for test plates and solid samples
with R.I. up to 1.65: Monobromonaphthalene, 6 ml bottle
Order Guideline
Minimum equipment: 1x 13-0285
Spares (approx. 1 year): 13-1523, 13-0287
Additional requirements: