CMAC 5208 AC/DC Current Clamp CMSS 6187-CE AC/DC Current Clamp
The AC/DC Current Clamps (Models CMAC The AC/DC Current Clamps allows a user to
5208 and CMSS 6187-CE) are battery- easily and safely incorporate the capability for
powered clamp-on current probes that is used current monitoring within Microlog routes.
on primary circuits with the SKF Microlog® The jaws clamp around a conductor allowing
Portable Data Collector/FFT Analyzer to current measurements without breaking the
measure AC or DC currents up to 600 amps circuit. A Hall Sensor mounted in a gap in
peak AC, 600 amps DC. It is one of a range of the core senses the magnetic field produced
accessories offered by SKF Reliability Systems by the current being measured. An amplifier
to enhance the predictive maintenance efforts circuit generates an output signal of 1 mV per
of the proactive machinery analyst. amp (600 amp range) or 10 mV per amp (60
amp range). Demagnetizing (degaussing) is
accomplished by means of a push-button.
Features
Lightweight and easy to carry, the AC/DC
• Solid state hall effect transducer
Current Clamp may be used throughout the
• AC (1 Hz to 5 kHz) and DC (0 to 1 Hz) plant or in control room environments.
measurement
For more information on utilizing the AC/DC
• Ranges – 60 and 600 amps Current Clamp along with the Microlog and
SKF Machine Suite software, contact your
• Degaussing adjustment for zero noise local SKF Reliability Systems Representative.
offset
AC/DC Current Clamp
CMAC 5208 and CMSS 6187-CE
www.skf.com/cm
2
Specifications
Model CMAC 5208 Model CMSS 6187-CE
ELECTRICAL
Current Range: 40 A (AC) / 60 A (DC) Range: 0.2 to 40 A (AC), (60 A Peak) 0.4 to 60 A (DC)
400 A (AC) / 600 A (DC) Range: 0.5 to 400 A (AC), (600 A Peak) 0.5 to 600 A (DC)
400 A (AC) / 600 A (DC) Range: 1.5% of Reading ± 1.0 A (0.5 to 100 A)
2.0% of Reading (100 to 400 A)
2.5% of Reading (400 to 600 A DC)
400 A (AC) / 600 A (DC) Range: ≤ 3.0° (10 to 100 A) ≤ 2.0° (10 to 100 A)
≤ 1.5° (100 to 400 A)
* Reference Conditions: +64 °F to +84 °F (+18 °C to +28 °C), 20% to 75% relative humidity, external magnetic field < 40 A/m, no DC component, no external current carrying
conductor, test sample centered, 1 MΩ ≤ 100 pF load, zero adjustment prior to measurement (DC only) DC to 65 Hz, Battery voltage = 9 V ± 0.1 V.
ENVIRONMENTAL
Case Protection: IP30 per IEC 529
Operating Relative Humidity: +10 °C to +35 °C, 90% ± 5% Relative Humidity (Without condensation)
+40 °C to +55 °C, 70% ± 5% Relative Humidity (Without condensation)
MECHANICAL
Zero Adjustment: Automatic zero
Battery: 9 V Alkaline
Battery Life: Approximately 120 hours with alkaline battery Approximately 50 hours with alkaline battery
Overload Indication: Red LED indicates input greater than the selected range
MATERIAL
Output Lead: Double/reinforced insulated 5 foot (1.5m)
CM2016 (Revised 8-06) • Copyright © 2006 by SKF Condition Monitoring, Inc. ALL RIGHTS RESERVED.