Sie sind auf Seite 1von 10

L A B O R A T O R Y

Preparing Your Laboratory for


PerkinElmer Atomic Absorption
Spectrometers

P R E P A R A T I O N
This document gives detailed 1. Suitable Working A means of electrically grounding
instructions regarding the space, Area the instrument and all accessories
accessories and utilities required The environment in which any must be available.
to operate PerkinElmer atomic instrument is housed is an important
absorption (AA) spectrometers, consideration. The room temperature Atomic Absorption System Layout.
THGA and HGA graphite should be between 15 and 35 °C In addition to the basic requirements
furnace systems, and other major (59-95 °F) with a maximum rate of for atomic spectroscopy systems,
AA accessories. change of 3 °C (5 °F) per hour. The preparation of your laboratory for
relative humidity should be between atomic absorption (AA) spectrometers
PerkinElmer atomic absorption 20 and 80%, noncondensing. equipped with graphite furnaces
instruments are complete systems requires that consideration be
with the exception of the A relatively dust-free environment given to the installation needs of
following items which must be is necessary. This is especially the furnace power supply when it
provided by the analyst: suitable important when working with is not built into the spectrometer.
working area; exhaust vent(s); ultra-trace techniques, such as For example, an AAnalyst™ 400
gases and regulators and light graphite furnace sampling. Other system with an HGA 900 graphite
sources. When a THGA or HGA important considerations are to furnace typically includes the
graphite furnace system is being locate the instrument in an area HGA 900 power supply, the HGA 900
used, additional items may be free of corrosive fumes and vibration furnace, an AS-800 autosampler,
required, which are described in and in an area that does not receive the AAnalyst 400 AA spectrometer
Section 8. General information on direct sunlight. and the system computer and printer.
each of the required items is given
in the appropriate section of this Atomic absorption spectrometer The HGA 900 furnace power supply
brochure as indicated below. systems are normally installed on mounts to the left side of the
laboratory work benches or tables. AAnalyst 400 spectrometer. The
TABLE OF CONTENTS The benchtop or area in which the system computer may be positioned
atomic absorption instrument is to on the bench next to the instrument
Section Subject
1 Suitable Working Area be installed should be large enough or on a separate computer table in
2 Exhaust Vent to accommodate the instrument and front of the instrument. However, it
3 Atomic Absorption Gases: all accessories. (Dimensions of should not be positioned immedi-
Compressed Air
those items are given in Section 10.) ately adjacent to the furnace power
Acetylene
Nitrous Oxide Make sure that there is space at the
Argon rear and sides of the system for air WARNING: The use of atomic
4 Gas Line Connections to circulate freely. Do not block the absorption instruments without
5 Handling of Gas Cylinders fan operating on the back or rear of adequate ventilation to outside
and Other Suggested the instrument. The back of the air may constitute a health hazard.
Safety Practices instrument should not be placed
6 Drain Vessels For example, the combustion of
permanently against a wall, as the halogenated hydrocarbons
7 Atomic Absorption
Source Lamps instrument must be accessible from produces toxic vapors. Extreme
8 Graphite Furnace the back for servicing purposes. An care should be taken that exhaust
Requirements accessible space of at least 60 cm gases are vented properly.
9 Training Courses (24 in.) should be available behind
10 Facilities Required the instrument. If this is not possible,
for PerkinElmer AA the table or bench on which the
Instruments and
Accessories
instrument is mounted should be on
wheels so that it can be easily moved.

w w w . p e r k i n e l m e r. c o m
supply, as the strong magnetic It is strongly recommended that the
Table I. U.S. Suppliers of
field generated during operation of instrument not be placed in a Flexible Metal Duct Tubing
the module might interfere with the chemical hood! If a chemical hood
computer’s CRT display. must be used, arrangements should be Flexaust Co. Potomac Rubber Co.
made to keep out corrosive vapors 11 Chestnut St. 9011 Hampton Overlook
Amesbury, MA Capital Heights, MD
FIAS systems and flame autosamplers and backdrafts from other hoods. 01913 20743
should be placed on a cart or table (508) 388-9700 (301) 336-7400
close to the AA spectrometer sample Sample preparation should not be
compartment to keep FIAS tubing carried out in the same hood where Triplex Inc. Fox Manufacturing Inc.
to a minimum length. The FIAS can the instrument is located. 1142 Kress St. P.O. Box 1047
also be placed to the side of the Houston, TX Clarkdale, AZ
77020 86324
instrument, however, maximum PerkinElmer offers an accessory
(713) 672-7521 (602) 634-5897
performance may be compromised Blower and Vent Kit (Part No. 0303-
due to the increased length of 0406) which will fulfill the exhaust Darcoid Co. of California
tubing required. requirements for all atomic 1742 Yosemite Ave.
absorption instruments (see Figure 1). San Francisco, CA 94124
Included in the kit are a rotary (415) 861-6984
2. Exhaust Vent blower with capacitor and hardware,
For the names of suppliers in other areas,
A venting system is required to a vent to be located above the
contact your PerkinElmer representative.
remove the combustion fumes and instrument, and an adapter to permit
vapors from the flame or graphite connection of the blower and vent
furnace for atomic absorption instru- with suitable metal tubing. The The capacity of the blower depends
ments. Exhaust venting is important adapter and vent are made of on the duct length and the number
for a number of reasons: stainless steel sheets. of elbows or bends used to install
the system. If an excessively long
• It will protect laboratory Notice: Local electrical codes do duct system or a system with many
personnel from toxic vapors not permit PerkinElmer Service bends is used, a stronger blower
which may be produced by Engineers to install the blower and may be necessary to provide
some samples. vent assembly. sufficient exhaust volume.
Alternatively, smooth stainless steel
• It will tend to remove the effects The metal tubing required to tubing may be used instead of
of room drafts and the laboratory connect the vent to the blower and flexible stainless steel tubing where
atmosphere on flame stability. to provide suitable exhaust from the flexibility is not required to reduce
blower is not included in the system friction loss or “drag.” A
• It will help to protect the accessory Blower and Vent Kit. length of smooth stainless steel
instrument from corrosive Flexible stainless steel tubing is ducting has 20-30% less friction
vapors which may originate recommended for this purpose and loss than a comparable length of
from the samples. can be obtained from the companies flexible ducting. When smooth
listed in Table I and from various stainless steel tubing is used, elbows
• It will remove dissipated heat other firms. In some instances this must be used to turn corners. These
which is produced by the flame type of flexible metal tubing is sold elbows should turn at a center line
or furnace. only in minimum lengths of 3 radius of 45 degrees to reduce
meters (10 feet). friction losses, and the number of
The venting system should provide a elbows should be minimized.
flow rate of approximately 7000-8500
liters/min (250-300 cubic feet/min). If a rigid tubing system is used, it is
strongly recommended that flexible
tubing be used from the vent hood
to the ceiling to facilitate hood
alignment and service access to
the instrument.

The dimensions for the various


parts of the Blower and Vent Kit are
shown in Figure 1. The vent i.d. is
slightly larger than the tubing o.d.
to allow for tubing tolerances. A
slight gap between the two units
is normal.

Figure 1. Blower and vent kit dimensions.

2 w w w . p e r k i n e l m e r. c o m
When installing such a venting • Design local exhaust ventilation WARNING: When using an air
system, all connections should be systems individually for each or air/acetylene accessory, NEVER
made with metal screws or rivets. specific atomic absorption exceed 1050 kPa (150 psig).
Solder must not be used. The blower instrument. Also, the opening of
should be located at least 4 meters the exhaust vent should be large
(12 feet) and not more than 6.5 enough to cover the graphite If there is any doubt regarding the
meters (20 feet) from the flame or the furnace or flame area completely. usability of a central air supply
graphite furnace and should exhaust (insufficient volume or pressure or
to the atmosphere or into a • Provide make-up air in the same excessive oil or water contamination),
considerably wider exhaust duct. quantity as is exhausted by the the quality of the supply should
Under these conditions, the system. An “airtight” lab will be checked prior to the arrival of the
following temperatures have been cause an efficiency loss in the instrument. A small, oil-less air
measured during operation of a exhaust system. compressor meeting the stated
nitrous oxide-acetylene flame: requirements is available from
310 °C at the vent intake; 160 °C at • Ensure that the system is drawing PerkinElmer (Part No. 0303-0313 for
2.4 meters (8 feet) from the vent properly by releasing smoke into 115 volts, 60 Hz or Part No. 0303-
intake; 105 °C at the blower intake; the mouth of the collector hood 0314 for 230 volts, 50 Hz service).
and 50 °C at the blower motor (vent). A synthetic “smoke” can
housing near the front bearing. be generated by placing open Air compressors are generally
bottles of hydrochloric acid and uncomfortably noisy to have in the
Instructions for installation (Part No. ammonium hydroxide in immediate vicinity of the instrument.
0990-9701) are provided with the proximity under the hood. Whenever possible, it is advisable to
Blower and Vent assembly. The locate them at some distance from
blower provided in the PerkinElmer • Equip the blower with a pilot laboratory workers in an area
accessory kit requires a line voltage light located near the instrument providing suitable ventilation.
of 115 V or 230 V. to indicate to the operator when
the blower is on. Cylinders of compressed air can also
Additional recommendations on the be used but are recommended only
venting system include: as an emergency or short-term
3. Atomic Absorption solution for the following reasons:
• Make sure the duct casing is Gases
installed using fireproof • A standard #1 size air cylinder
construction. Route ducts away contains about 6200 liters (220
NOTE: Standards for cylinder
from sprinkler heads. cubic feet) of air at standard
dimensions, regulator connections,
temperature and pressure (STP).
gas fittings, etc. vary from country to
• Locate the discharge outlet as A premix burner-nebulizer system
country. The information provided
close to the blower as possible. uses about 20 liters/min (0.7 cubic
here is for the U.S. Contact your
All joints on the discharge side feet/min), and, therefore, a
PerkinElmer representative for
should be airtight, especially if cylinder will last only about five
information on the specific
toxic vapors are being carried. hours. Unless an instrument is
requirements of your area.
used only a few hours per day,
• Equip the outlet end of the system changing cylinders becomes a
with a backdraft damper and take Compressed Air. For flame nuisance as well as being expensive.
the necessary precautions to keep operation, the air supply should
the exhaust outlet away from open provide a minimum of 28 liters/min • Occasionally, cylinder air may be
windows or inlet vents and to (1 cubic foot/min) at a minimum obtained from a liquefaction
extend it above the roof of the pressure of 350 kPa (50 psig). The process during which the oxygen-
building for proper dispersal of pneumatics of the AAnalyst 700 and to-nitrogen ratio can change.
the exhaust. AAnalyst 800 systems require an air Therefore, it is not uncommon to
pressure of 500 kPa (70-75 psig). find other than 20% oxygen in air
• Equip the exhaust end of the cylinders. This can cause erratic
system with an exhaust stack to It is desirable to have a water and burner operation and non-repro
improve the overall efficiency of oil trap or filter between the ducible analytical results and, in
the system. compressor and the instrument gas extreme cases, may provide a
control system. The use of an Air potential safety hazard. In general,
• Make sure the length of the duct Filter Accessory (Part No. 0047-0652) if cylinder air is to be used, it is
that enters into the blower is a or an Air/Acetylene Filter Accessory important to specify compressed
straight length at least ten times (Part No. N930-1398) is strongly air rather than breathing air
the duct diameter. An elbow recommended to remove entrained (i.e. medical grade) or an
entrance into the blower inlet water, oil, water aerosols and solid unspecified form.
causes a loss in efficiency. particles from compressed air lines.
WARNING: For safe operation,
oxygen must NEVER be used with
PerkinElmer premix burner systems.

3
The use of air cylinders requires The PerkinElmer Acetylene Since the acetylene is dissolved
the use of a suitable dual-stage Regulator Assembly includes an in acetone, the pressure drop is
regulator. A regulator for cylinders adapter so that the pressure regulator not linear with gas removal, and
with a CGA 590 connection is can be connected to cylinders a pressure of 600 kPa (85 psig)
available from PerkinElmer requiring either CGA 300 or CGA indicates that the cylinder is nearly
(Part No. 0303-0264). 510 fittings and a connector for empty, assuming the cylinder is
attaching the fuel hose assembly at room temperature.
Acetylene. For the overwhelming supplied with the instrument. The
majority of analyses, acetylene is the fuel hose assembly is constructed of Acetylene tanks should always be
preferred fuel gas with atomic red neoprene, reinforced with high stored and operated in a vertical
absorption spectrometers. Air- tensile strength rayon cord, and position, rather than horizontally, to
acetylene is the preferred flame for provides a rated working pressure of prevent liquid acetone from reaching
the determination of about 35 about 1700 kPa (250 psig). The the cylinder valve. New tanks should
elements by atomic absorption. connectors are permanently mounted be positioned vertically for at least 8
The temperature of the air-acetylene at each end of the hose assembly for hours prior to use. The practice of
flame is approximately 2300 °C. connection to the pressure regulator “cracking the valve” of an acetylene
For most air-acetylene flames, the and instrument gas controls, and use tank (that is, opening the valve
acetylene flow used is about 4 left-hand threads as per accepted slightly for a very short period prior
liters/min or 0.14 cubic feet/min. practice for fuel gas connections. to attaching the regulator) is not
Using a heat combustion value of (See Section 5 for more details.) recommended. While such an action
1,450 BTU per cubic foot, the heat will clear the valve opening of dust
given off would be approximately It may be desirable to have an or dirt particles and may remove
12,300 BTU per hour (3,600 W). An acetylene filter between the acetone from the cylinder valve, it is
air-acetylene flame can be used with acetylene tank and the instrument a potentially hazardous practice and
all PerkinElmer burner heads but is gas control system to remove one which should never be attempted
most commonly used with the particulates and acetone droplets in the presence of an open flame,
supplied 10-cm (4-inch) burner head from acetylene, protecting the gas sparks or other possible sources
(Part No. N040-0102 for AAnalyst controls and AA burner system from of ignition.
series instruments). contamination and corrosion.
An Acetylene Filter (Part No. N930-
CAUTION: Acetylene line
Suitable acetylene typically has a 1399) and an Air/Acetylene Filter
pressure from the cylinder to the
minimum purity specification of Accessory (Part No. N930-1398) are
instrument should never be allowed
99.6% with the actual assay being available from PerkinElmer. Some
to exceed 103 kPa (15 psig). At
about 99.8%. In general, ordinary countries also require the use of a
higher pressures, acetylene can
welding grade acetylene is adequate flashback arrestor such as PerkinElmer
spontaneously decompose or
for most atomic absorption analyses, Part No. N930-0068 in the acetylene
explode. PerkinElmer recommends
though sometimes a particular tank fuel line.
that a maximum acetylene line
may be contaminated. Special higher
pressure of 80-95 kPa (12-14 psig)
purity “atomic absorption” grade Acetylene is normally supplied
be used to provide a reasonable
acetylene is also available from some dissolved in acetone, and a small
margin of safety.
vendors, and its use is recommended amount of acetone carryover with
when the available welding grade the acetylene is normal. However, as
acetylene is not sufficiently pure. tank pressure falls, the relative Both fuel and oxidant gas lines
amount of acetone entering the gas should be relieved of pressure at the
A size 1A acetylene cylinder stream increases and can give erratic end of the working day or if the
contains about 8,500 liters (300 results, particularly for elements instrument is to be unused for an
cubic feet) of acetylene and usually such as calcium, tin, chromium, extended period. Cylinder valves
lasts about 30 hours of burning time molybdenum and others whose should be closed to avoid the
with an air-acetylene flame. The sensitivity is highly dependent on possibility of pressure regulators
cylinder requires an acetylene the fuel/oxidant ratio. For this failing and gas lines being subjected
pressure regulator, which can be reason, acetylene tanks should be to the full cylinder pressure.
obtained from the supplier of the replaced when the cylinder pressure
gas or from PerkinElmer drops to about 600 kPa (85 psig). Nitrous Oxide. The nitrous oxide-
(Part No. 0303-0106). acetylene flame has a maximum
IMPORTANT: Failure to change temperature of approximately 2800 °C
CAUTION: Acetylene may react the acetylene cylinder before the and is used for the determination of
with copper to form a potentially cylinder pressure drops below 600 elements that form refractory oxides.
explosive compound. Copper tubing kPa (85 psig) may cause damage It is also used to overcome chemical
or fittings for acetylene gas must to valves or tubing within the burner interferences that may occur in
be strictly avoided. gas control system due to high flames of lower temperatures. For
acetone carryover. Such damage the nitrous oxide-acetylene flame,
from acetone is not covered by the acetylene flow is about 14
instrument warranties. liters/min or 0.5 cubic feet per min.

4 w w w . p e r k i n e l m e r. c o m
Using a heat of combustion value of A nitrous oxide burner head (Part 4. Gas Line
1,450 BTU per cubic foot, the heat No. N040-0100 for AAnalyst series Connections
given off would be approximately instruments) must be used with PerkinElmer atomic absorption
43,000 BTU per hour (12,500 W). nitrous oxide-based flames. The instruments include the hoses
instructions provided with the necessary for connection to gas lines
The use of nitrous oxide requires a nitrous oxide burner head should (see Table III). It is the responsibility
number of accessories and precautions. be strictly followed. of the user to provide the appropri-
A size 1A cylinder of nitrous oxide ate gas lines, regulators, connectors
contains about 14,800 liters (520 There is generally little advantage to and valves to which the hoses
cubic feet) and will typically last for using nitrous oxide-acetylene to are connected.
10 to 12 hours of burning time. determine elements that are best
Cylinders of nitrous oxide (99.0% determined with air-acetylene. Air is
minimum purity) are available from less expensive than nitrous oxide; 5. Handling of Gas
local suppliers. A dual-stage regulator also, a number of elements will Cylinders and Other
is recommended (and is mandatory show poorer sensitivity and produce Safety Practices
in some countries.) ionization interferences in the
nitrous oxide flame.
Nitrous oxide is supplied in the NOTICE: The permanent
liquid state, initially at a pressure of Argon. Argon is required for external installation of gas supplies is the
about 5000 kPa (750 psig). Since the and internal gas streams through the responsibility of the user and
nitrous oxide is in liquid form, the THGA or HGA graphite furnace or as should conform to local safety
pressure gauge does not give a true a carrier gas with mercury/hydride and building codes.
indication of how much nitrous systems such as the FIAS or FIMS
oxide remains in the cylinder until flow injections systems or the MHS-15. • Fasten all gas cylinders securely
the pressure starts to fall rapidly as The quality criteria listed in Table II to an immovable bulkhead or a
the residual gas is drawn off. for argon are suitable for these permanent wall.
applications. Normally, for graphite
When nitrous oxide is rapidly furnace operation, gaseous argon is • When gas cylinders are stored in
removed from the cylinder, the used, although either liquid or confined areas, such as a room,
expanding gas causes cooling of the gaseous argon can be used. The ventilation should be adequate
cylinder pressure regulator and the choice of liquid argon or gaseous to prevent toxic or explosive
regulator diaphragm sometimes argon tanks is determined primarily accumulations. Move or store gas
freezes. This can create erratic flame by the availability of each and the cylinders only in a vertical position
conditions or, in the most extreme usage rate. Liquid argon is usually with the valve cap in place.
case, a flashback. It is therefore less expensive per unit volume to
advisable to heat the regulator using purchase, but cannot be stored for • Locate gas cylinders away from
either a built-in heater or an extended periods. If liquid argon is heat or ignition sources, including
externally supplied heat source, used, the tank should be fitted with heat lamps. Cylinders have a
such as an electrical resistance an over-pressure regulator which pressure-relief device that will
heating tape. will vent the tank as necessary in release the contents of the cylinder
order to keep the liquid argon cool if the temperature exceeds
CAUTION: All lines carrying enough to remain in the liquid state. 52 °C (125 °F).
nitrous oxide should be free of Gaseous argon tanks do not require
grease, oil or other organic material, venting and consequently can • When storing cylinders external
as it is possible for spontaneous be stored for extended periods to a building, the cylinders should
combustion to occur. Cylinders of without loss. be stored so that they are protected
nitrous oxide should be considered against temperature extremes
high-pressure cylinders and should A dual-stage cylinder regulator that (including the direct rays of the
be handled with care at all times. can be used with either gaseous sun) and should be stored above
argon or nitrogen is available from ground on a suitable floor.
A dual-stage heated nitrous oxide PerkinElmer (Part No. 0303-0284).
pressure regulator for use with gas The regulator has a CGA 580 fitting, • Mark gas cylinders clearly to
cylinders with a CGA 326 connection and includes a color-coded hose identify the contents and status
is available from PerkinElmer with 1/4-inch SWAGELOK® fittings (full, empty, etc.).
[Part No. 0303-0204 (115 volts) or to permit direct connection to the
0303-0349 (230 volts)]. These regulators regulator and to the instrument gas • Do not attempt to refill gas cylinders.
provide pressure control from controls (see Section 5).
350-520 kPa (50-75 psig) and contain • Use only approved regulators and
Table II.
an integral thermostatted heater to hose connectors. Left-hand thread
Quality Criteria for Argon
prevent freezing of the regulator fittings are used for fuel gas tank
diaphragm. A color-coded hose with Criterion Specification connections, whereas right-hand
suitable connectors at each end is Purity ................................ ≥ 99.996% fittings are used for oxidant and
supplied with the regulators to provide Oxygen ................................. ≤ 5 ppm support gas connections.
connection to the instrument gas Nitrogen ............................. ≤ 20 ppm
controls (see Section 5). Water .................................... ≤ 4 ppm
5
• Use galvanized iron tubing, steel, • Take suitable precautions when
WARNING: Contact between
wrought iron or other tubing that using volatile organic solvents. A
acetylene gas and copper or silver
will not react chemically with potentially flammable organic
(or high concentrations of silver
acetylene. Never use copper tubing vapor “cloud” can form around
salts), liquid mercury or gaseous
with acetylene. Joints may be the opening of the sample vessel.
chlorine can produce potentially
welded or made up of threaded Feeding the capillary tubing through
unstable acetylides. Always clean
or flanged fittings, typically a small hole in a covered sample
the burner thoroughly after
stainless steel, aluminum or brass container is one way of reducing
analyzing solutions with high silver
composed of less than 65% copper. the possibility for ignition.
or mercury concentrations, and
Rolled, forged or cast steel or
aspirate solution continuously
malleable iron fittings may also be • Never view the flame, hollow
during the analysis to prevent
used. Cast iron fittings cannot be cathode lamps, electrodeless
any residues from drying.
used safely for acetylene lines discharge lamps or deuterium
. background corrector lamps directly
• Arrange gas hoses where they will without protective eyewear.
not be damaged or stepped on and • When the equipment is turned Potentially hazardous ultraviolet
where things will not be dropped off (for example, at the end of the radiation may be emitted. Ordinary
on them. working day), close all gas safety glasses will in general provide
cylinder valves tightly at the sufficient protection, but additional
• Never run acetylene at a pressure tank. Bleed the remainder of the side shields will ensure a further
higher than 100 kPa (15 psig). At line to the atmosphere before the margin of safety. Safety glasses
pressures above this level, acetylene exhaust fan (vent) is turned off. will also provide mechanical
may spontaneously explode. protection for the eyes.
• When using premix burners with
• Perform periodic gas leak tests by cyanide solutions, check the pH • Never leave the flame unattended.
applying a soap solution to all of the liquid trap and drain vessel.
joints and seals. The pH of the liquid should be • Zeeman background-corrected
greater than 10. If the liquid is even AA instruments generate a strong
• Periodically check for the presence slightly acidic, highly toxic hydrogen magnetic field. People with cardiac
of acetylene in the laboratory cyanide gas may be released. pacemakers are advised not to
atmosphere, especially near operate or frequent the vicinity of
the ceiling. Zeeman-corrected instruments
while they are in operation.

Table III. Gas Line Connections (Note: Regulator, connector and fitting needs vary by country.
For information on what is required in your area, consult your local PerkinElmer Service Representative.)

Regulator
Regulator Regulator Connection to Hose Assembly Connections
Part to Cylinder Gas Hose Part Connection to Connection to
Gas Number CGA No. Assembly Number Color Regulator Instrument
Air 0303-0264 590 1/4" SWAGELOK 0057-0567 black 1/4" Swagelok 1/4" Swagelok
Argon 0303-0284 580 1/4" SWAGELOK 0057-0567 black 1/4" Swagelok 1/4" Swagelok
Nitrogen 0303-0284 580 1/4" SWAGELOK 0057-0567 black 1/4" Swagelok 1/4" Swagelok
N2O 0303-0204 326 1/4" N.P.T.* 0047-0258 blue 5/16" Swagelok 5/16" Swagelok
Acetylene 0303-0106 510 or 300 9/16" L.H.T.** 0057-0559 red 9/16" L.H.T. 3/8" L.H.T.

N.P.T. = Normal Pipe Thread, L.H.T. = Left-Hand Tread


* Supplied with 5/16" x 1/4" N.P.T. SWAGELOK Male connector Body (P/N 0990-3946).
** Supplied with Outlet Bushing (P/N 0990-3031), 1/4" N.P.T. to 9/16" L.H.T.

PerkinElmer Air Compressor (P/N 0303-0313, 0303-0314) provides a 1/4" SWAGELOK fitting.
PerkinElmer Air Filter Assembly (P/N N058-0531) provides 1/4" SWAGELOK inlet and outlet fittings.
PerkinElmer Air Dryer Filter Assembly (P/N 0047-0652) provides 1/4" SWAGELOK inlet and outlet fit-tings.
PerkinElmer Air/Acetylene Filter Assembly (P/N N930-1398) provides 1/4" SWAGELOK inlet and outlet fittings for air
and 3/8" LH (A size) inlet and 9/16" LH (B size) outlet fittings for acetylene.
PerkinElmer Acetylene Filter (P/N N930-1399) provides 3/8" LH (A size) inlet and 9/16" LH (B size) outlet fittings.

Also Available:
1. P/N 0990-3032 Connector for joining two P/N 0057-0559 fuel hose assemblies
2. P/N 0990-3898 Connector for joining two P/N 0057-0567 air/argon hose assemblies
3. P/N 0990-3196 Adapter, female 1/4" N.P.T. to male 1/4" SWAGELOK.
4. P/N 0992-0223 Connector for joining two 0047-0258 nitrous oxide hose assemblies.

6 w w w . p e r k i n e l m e r. c o m
6. Drain Vessels Hollow cathode lamps are excellent volts is recommended (with a
A specially-configured drain vessel for most elements; however, there minimum of 195 V under load
is supplied with all PerkinElmer are a number of “difficult” elements required) to enable the furnaces to
atomic spectroscopy instruments for which an improved light source reach maximum potential operating
with burner systems. That vessel is desirable. PerkinElmer System 2 temperatures and heating rate and,
must be used to gather the effluent Electrodeless Discharge Lamps (EDLs) for some systems with Zeeman
from the AA burner drain. provide improved performance in effect background correction, proper
most instances. EDLs are more magnetic field strength. For
The drain vessel should NOT be intense than their corresponding installations where the line voltage
stored in an enclosed storage area. hollow cathode lamps. Most also may drop below this level, the use
Rather, the drain vessel should be provide better lamp life and stability of a “buck boost”-type transformer is
stored in plain sight of the operator, and some also provide better recommended to maintain proper
usually on the floor in front of the sensitivity. EDLs do not require a analytical operating conditions.
instrument or on an open shelf separate power supply, except for
underneath the instrument table. the AAnalyst 200 without D2 An appropriately-rated female
background correction. electrical connector is required to
The drain system should be checked provide power for the graphite
regularly and replaced when necessary. A lamp mount or turret is supplied furnace. Please contact your local
Follow the directions in the instrument with all PerkinElmer AA instruments Service Engineer to determine
manuals regarding the proper place- and will accommodate all PerkinElmer the appropriate connector for
ment of the drain tube in the drain hollow cathode or electrodeless your laboratory.
vessel and the proper liquid level in discharge lamps. Users who may
the drain vessel. have lamps with 1.5-inch diameters Electrical supply circuitry, circuit
rather than the standard PerkinElmer breakers and wiring size for the
2-inch diameter can adapt those graphite furnaces should be selected
7. Atomic Absorption lamps for use in PerkinElmer lamp according to local regulations.
Source Lamps mounts with the Small Diameter Lamp Three types of 3-wire circuits used
Adapter Kit, (Part No. 0303-0870). in North America which provide
Atomic absorption spectrometers power adequate to run THGA or
require different source lamps, HGA systems are illustrated in
depending on the elements to be 8. Graphite Furnace Figure 2. The AAnalyst 700 AA
determined and the instrument to be Requirements requires a 230 volts (+5%, -10%),
used. Multielement lamps are available 50/60 Hz (±0.3 Hz ), 20- or 30-amp
for some elements, but most lamps are Location. The furnace power supply
is built into the SIMAA 6100 and line capable of delivering 3.5 kW of
constructed using a single element to peak power.
avoid potential spectral interferences the AAnalyst 600/700/800. With the
and reduced performance, especially AAnalyst 400, the HGA 900 must
be placed on the left side of The THGA furnace and Zeeman
when using a graphite furnace. magnet of the AAnalyst 600,
the spectrometer.
AAnalyst 800 and SIMAA 6100
PerkinElmer manufactures all of its systems operate from a single,
hollow cathode and electrodeless Services. Graphite furnaces require
electrical power, cooling water and a dedicated electrical supply of 230
discharge lamps. The Lumina™ volts (+5%, -10%), 30 amp, 50 or 60
series of hollow cathode lamps are supply of inert gas, normally argon.
A minimum input voltage of 208 Hz (± 0.3 Hz), single phase, capable
especially noted for spectral purity, of delivering 7.7 kW of peak power.
brightness, stability and long life.

Figure 2. HGA power circuit connections.

7
The SIMAA 6100, AAnalyst 600
and AAnalyst 800 systems are all CAUTION: Do not use
nitrogen as the furnace purge
provided with a 30-amp plug, and
gas, Its use may lead to
the HGA 900 and AAnalyst 700 AA
reduced sensitivity for some
is provided with a 20-amp plug. It is
elements, and it is also possible
recommended that 8-gauge (6 mm2)
for nitrogen to react with the
wire be used for the electrical supply
graphite tube at temperatures
for SIMAA 6100, AAnalyst 600, and
above 2300 °C to form
AAnalyst 800 systems, and that the
cyanogen, a toxic gas.
length of the wiring (circuit breaker
to instrument connection) not When operating the HGA
exceed 20 meters (65 feet). Although Graphite Furnace systems at
10-gauge (4 mm2) wire can be used high temperatures, do not
for the electrical supply of the look directly at the tube without
HGA 900 and AAnalyst 700 AA, suitable eye protection.
8-gauge (6 mm2) wire is recommended.

For all furnace systems, the Maximum gas consumption is


electrical supply should contain a 0.7 liters/min (0.19 cubic feet/min)
"slow blow" circuit breaker capable for THGA furnaces and 1.2 liters/min
of handling 300% of the rated (0.3 cubic feet/min) for the HGA
current for periods of 3 seconds. furnaces, both at 300-450 kPa
Also, the AA spectrometer, graphite (44-66 psig). Purity specifications
furnace, Zeeman magnet, computer for the gases to be used with HGA
and other accessories should all be Systems are given in Section 3.
connected to the same electrical Ventilation is required to remove
ground, and the power supply potentially toxic or corrosive
should be free of transients in excess gases which can be generated by
of 50 V over the nominal voltage. the samples.

Additional Furnace Requirements.


A water supply is required to cool 9. Training Courses
the furnace quickly to ambient
temperature after reaching high Training courses for users of
atomization temperatures. The water PerkinElmer atomic absorption
supply should be free of sediment, instruments are given at various
have a pH between 6.5 and 7.5, and locations. The courses cover basic
a hardness less than 14 °D (2.5 principles and applications of atomic
mMol/ liter). A maximum flow of absorption and detailed instruction
2.5 L/min (0.6 gal/min) is used for in the use of the instruments and
the THGA, 1.5 liters/min (0.4 major accessories.
gal/min) for the HGA at a
temperature between 20 °C and To gain the most benefit from the
40 °C. As both the flow rate and training course, it is strongly
water temperature affect the cooling recommended that the attendee
rate, it is desirable to be able to vary should have operated the instrument
the flow rate to compensate for for at least several weeks prior to
variations in cooling water the course.
temperature. A suitable recirculating
cooling system is included with a For additional information on
number of THGA systems. With an atomic spectroscopy training
HGA system, the use of an optional courses, please contact your local
Recirculating Cooling System is sales representative.
strongly recommended.
To prevent combustion of the
graphite tube at high temperatures,
the furnace is purged with argon.

8 w w w . p e r k i n e l m e r. c o m
10. Facilities Required
for PerkinElmer
AA Instruments
The following figure and tables
provide the dimensions and power
requirements for PerkinElmer
atomic absorption instruments and
major accessories. Dimensions are
defined in Figure 3. Required
services are shown in Table IV,
and product dimensions and
approximate power consumption
for AA spectrometers and major
accessories are given in Table V.
PerkinElmer instruments will
normally operate within a ±10%
range of the specified voltage and
within ±1 Hz of the specified
frequency, unless otherwise noted.
If the power line is unstable,
fluctuates in frequency or is subject
to surges or sags, additional control
of the incoming power may be
required. A means of electrically
grounding the instruments and A = length
accessories must be available. B = height to top of cover
C = depth
D = distance from center of atomizer compartment to right hand edge of instrument
E = protuberance of sample tray in front of instrument
F = height of flame shield above top of instrument cover
G = 10 cm (4 inches) in all cases
H = 30 cm (12 inches) in all cases
J = 15 cm (6 inches) in all cases

Figure 3. Spectrometer and accessories dimensions.

Table IV. Required Services


Gases Cooling Computer Req. No. Electrical
Instrument/Accessory Air N2O C2H2 Ar Water Controlled of Vents Power

AAnalyst 200 1 a
AAnalyst 300 1 a
AAnalyst 400 1 a
AAnalyst 600 d 1 b
AAnalyst 700 d 1 b
AAnalyst 800 1 b
SIMAA 6100 1 b
HGA 900 d e c b
FIAS 100/400 e c a
FIMS 100/400 c a
Amalgamation Attachment e a
MHS-15 c

a = 115 V, 50/60 Hz or 230 V, 50/60 Hz, single phase


b = 230 V (+5%, -10%), 50 or 60 Hz, single phase
c = uses the same vent as the spectrometer
d = optional (included with some versions of the AAnalyst 600)
e = controlled from the spectrometer computer/controller

9
Table V. Instrument and Accessory Specifications
(Refer to Figure 3)

Width Height Depth Power Weight


(cm) (cm) (cm) (cm) (cm) (cm)
A B C D E F VA Kg
Spectrometers:
AAnalyst 200 70 65 65 43 17 18 200 49
AAnalyst 400 70 65 65 * 43 17 18 200 49
AAnalyst 600 110 65 70 ** 43 30 6 5000 175
AAnalyst 700 110 64 70 ** 43 30 10 3600 147
AAnalyst 800 110 64 70 ** 43 30 10 5000 187
SIMAA 6100 110 65 70 42 30 7 5000 195

Graphite Furnace Components:


HGA 900 Controller 32 65 65 3600 43
AS-800 Autosampler 25 20 34 NA 6
HGA/THGA Cooling System 20 37.5 50 140 18

Flame Autosamplers:
AS-90A/90plus 41 56 *** 25 NA 3
AS-93plus 44 56 *** 34

Mercury/Hydride System:
MHS-15 17 38 16 NA 3

Flow Injection Systems:


FIAS 100/400 42 18 41 600 11
FIMS 100/400 42 26 41 450 12
Amalgamation Attachment 25 25 24 300 5

Dell® GX 270 Computers:


CPU 18 42 45 690 15
Monitor (17-inch) 41 42 6 210 15
Keyboard 46 5 18 NA 2

Accessories:
AutoPrep™ 50 Autodilutor 34 39 20 60 7.5
EDL System 2 Power Supply 18 16 50 170 9
Air Compressor 65 55 31 700 36
HP LaserJet® 4200 Printer 39 35 41 330 18
HP 5600 Series InkJet® Printer 44 20 37 250 6
Heated N2O Regulator 150 2

* With the AAnalyst 400, the furnace autosampler projects 30 cm from the left front of the instrument in
its storage position.
** With the AAnalyst 600, 700 or 800, the furnace autosampler projects 35 cm from the right front of the
instrument in its storage position.
*** Height at maximum sample arm upward travel.
NA Not applicable (powered by another device or included with another component)

PerkinElmer Life and


Analytical Sciences
710 Bridgeport Avenue
Shelton, CT 06484-4794 USA
Phone: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com

© 2003 PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark and HGA, Intensitron, AAnalyst, AutoPrep, Lumina, SIMAA and THGA are trademarks of PerkinElmer, Inc.
All trademarks depicted are the property of their respective holders or owners. PerkinElmer reserves the right to change this document at any time and disclaims liability for editorial, pictorial,
or typographical errors.

006660B 080300 Printed in USA

Das könnte Ihnen auch gefallen