Sie sind auf Seite 1von 67

Amit Manjure

Clinical & Scientific Excellence

Secondary
Method Validation
Secondary Method Validation

We run controls, sample, we collect data, we create graphs, charts, plots and
crunch numbers.

It is expensive and take time to it right.

Then we stuff it all in a folder and hand it to an inspector. It's called method
validation.

But what is it? How do we do it? Why do we do it?


Why Secondary MV?

§ Method validation studies verify the lab can achieve product performance
claims.

§ Provide useful information about what to expect with a new method & how
results may be different compared to the method being replaced.

§ Help Labs Detect Analytical Error


• Imprecision – Random Error
• Inaccuracy – Systematic Error
• Total Error = Combination

§ Assist in meeting regulatory requirements


Page 3
What studies are necessary?

CAP / CLIA Checklist recommends that laboratory must verify or establish


method performance specifications for each method.

Typical Studies needed for Quantitative Assay :


• Precision
• Analytical Measuring Range (AMR) Verification
• Method Comparison
• Verification of Reference Range
• Sensitivity – Functional Sensitivity
• Instrument Carry (for systems not using disposable tips)

Page 4
When is it Done

Analytical methods need to be validated

• before their introduction into routine use

& revalidated

• whenever the conditions change for which the method has been
validated (e.g., an instrument with different characteristics)

• whenever the method is changed and the change is outside the


original scope of the method

Page 5
Process of establishing a Routine Test

Select Diagnostics Test


Establishing
Routine Testing
Select - Method of
Analysis and Perform

Validate Method
Performance

Maintain Method
Implement Method
Prevent Problem

Check with Report


Acquire Sample Perform test
Statistical QC Results

A Routine Testing Process


Page 6
Agenda :

Precision
AMR
Method
Verification/
Comparison
Range

Overview & Reference


Cautions Interval

Page 7
Statistical tools

Common tools:
Mean, SD, CV, Confidence intervals, Chi Squared, Regression analysis

All calculations made with statistical tools are estimates, not absolute truth.
Estimates will vary between studies.

Points to keep in mind:

§ All statistical tools make assumptions about the data


§ All calculated estimates have limitations
§ Need to keep the assumptions & limitations in mind when
evaluating results of statistical tests

Page 8
Samples Used

Methods are designed to test fresh patient samples


Any other type of sample may perform differently

• Frozen / stored patient samples


• Accuracy validation samples
• Calibration Verification samples
• QC samples
• EQA samples

Page 9
Why is matrix an issue ?
Matrix: everything in the sample except what we are measuring
Sample matrix can be altered by handling / processing
• Matrix proteins in particular
• Proteins may denature

What can alter sample matrix?


• Refrigeration
• Freezing
• Pooling
• Diluting
• Spiking
• Manufacturing processes

Page 10
Reference Guidelines
Clinical & Laboratory Standards Institute (CLSI):
Accuracy / Range:
• EP15-A2 User Verification of Performance for Precision and Trueness (2005)
Precision:
• EP5-A2 Evaluation of Precision Performance of Quantitative Measurement
Methods (2004)
• EP15-A2 User Verification of Performance for Precision and Trueness (2005)
Method Comparison:
• EP9-A3 Measurement Procedure Comparison and Bias Estimation Using
Patient Samples (2013)
Reference Interval:
• C28-A3c Defining, Establishing, and Verifying Reference Intervals in the
Clinical Laboratory (2008)

Page 11
Agenda

Precision
AMR
Method
Verification
Comparison
/ Range

Overview & Reference


Cautions Interval

Page 12
AMR Verification / Assay Range

Analytical Range – aka: Calibration range, reportable range, analytical


measurement range, linear range

Study
Ideal: compare to true reference method using fresh patient samples
Next best: use special samples validated for the method
Study:
• Test accuracy samples in replicates, typically triplicate
• Compare mean of replicates to acceptable limits

Potential issues:
• Need to verify that assigned acceptable values are current to method and reagent lot number
• Comparing results to method in current use is NOT an accuracy study unless current method is a
true reference method

Page 13
TSH : Method Validation : Instrument name (S/N. 546722)
AMR/Linearity/Dilution Recovery using Linearity Panels Lot# 076859
Date: 12-Jan-16
Instrument: ADVIA CENTAUR
Reagent: Lot # 045 : Exp: 20 Mar 2016
Linearity Material: AUDIT
Calibrators Lot#: 076859 Exp.: 31 Jan 2014
Units of Measure: uIU/ml

Acceptability
Achieved Values LIMIT
Calibrators (Target) 1st Rep 2nd Rep 3rd Rep 4th Rep Mean bias % bias + / - 2 CV%
S0 0.40 0.35 0.32 0.40 0.38 0.36 -0.037 -9.1% 13.3% PASS
S1 2.50 2.30 2.40 2.35 2.33 2.345 -0.155 -6.2% 13.3% PASS
S2 5.50 5.60 5.67 5.45 5.89 5.652 0.152 2.8% 13.3% PASS
S3 10.00 11.00 10.40 10.54 10.50 10.61 0.61 6.1% 13.3% PASS
S4 25.50 24.50 24.70 25.50 23.50 24.55 -0.95 -3.7% 13.3% PASS
S5 45.50 43.78 44.01 43.68 43.89 43.84 -1.66 -3.6% 13.3% PASS
S6 75.00 72.08 71.09 73.00 73.23 72.35 -2.65 -3.5% 13.3% PASS
S7 105.00 99.00 99.32 100.00 98.50 99.21 -5.79 -5.5% 13.3% PASS

Maximum allowa ble CV (1/3 of Total Error Allowed of 20 %) = 6.7%

T SH Linearity

100 y = 0.9475x + 0.4558


R² = 0.9998

80
Achieved

60

40

20

0
-10 10 30 50 70 90 110

Target

0
25
Manufacturer's claimed linear range: 0.01 to 150 uIU/ml
Page 14
Conclusion: Linearity on Instrument name S/N. 546722 demonstrated from 0.0 to 150 uIU/ml.
Precision
AMR
Method
Verification
Comparison
/ Range

Overview & Reference


Cautions Interval

Page 15
Precision

Measure of the reproducibility of the results


Ability to reproduce a result to a given level of confidence
(typically 95% confidence)

Characterized using statistical tools


®®
• Primarily: SD & CV ®®
®
• All SD and CV values calculated from data are estimates ®

All estimates of precision can be expected to


cluster around the “true” precision

Page 16
Precision

Two Types of Estimates .

§ With in Run or Short term or Intra-assay Precision


§ Between Run or Long Term or Inter Assay Precision

Which Matrix to Use ?


• Calibrators
• Controls.
• Patient Sample Pools

Page 17
Precision - Recommendation
Atleast, Two or Three different control material that represent low and high
Medical decision points are used.

With in Run or Short term or Intra-assay Precision

• 20 replicates for each control material in one run.


• Calculate mean and SD and CV %
• Validate, determine if the short term precision is acceptable.

Between Run or Long Term or Inter assay Precision.

• 20/10/5 replicates of each control material one run for one day over 5
days
• Calculate mean, SD and CV %
• Validate, determine if the long term precision is acceptable.

Page 18
THCG ADVIA CENTUAR - Method Validation
Instrument Serial Number:
Date Performed:18 JAN 2008
Control Name: IMMUNOASSAY PLUS Manufactured by:
C LOT NO: 40191 Exp Date: 31 JULY 2009
Intra-Assay (Within Run) Precision Inter-Assay (Between Run) Precision
Conventional Units ( mIU/mL ) Conventional Units ( mIU/mL )
Level 1 Level 2 Level 3 Date Replicate Level 1 Level 2 Level 3
1 1
2 2
08/01/2012 Run
3 1 13:34PM
3
4 4
5 5
6 6
7 7
09/01/2012 Run
8 8
2 14:30 PM
9 9
10 10
11 11
12 12
10/01/2012 Run
13 3 15:30PM
13
14 14
15 15
16 16
17 17
11/01/2012 Run
18 4 16:50PM
18
19 19
20 20
21
n 0 0 0 22
12/01/2012 Run
Mean 5 18:45PM.
23
sd 24
CV(%) 25

n 0 0 0
Mean
sd

Page 19 CV(%)
What does the SD represent?

s=
å i
( x - x) 2
SD ?
(n -1)

Essentially the average size of the differences


between the mean and the individual values

Page 20
Precision: Assumptions & Limitations

n = 20 n = 10 n=6
93 93
96 Number of replicates affects calculated SD
100 100 100
93
89
94
89
94
§ Fewer replicates tend to make SD
87 87 appear larger
89
86 86 86 § Fewer replicates means less
88
86 86 confidence in how accurately the
91 91 calculated SD represents assay
92 92
91
performance
102 102 102
90
90 90
85 85
91 91
92
Mean 91.3 91.6 93.0
SD 4.39 5.52 7.04
%CV 4.8% 6.0% 7.6%

Page 21
How good are our estimates ?

87 Data Points
• Overall mean: 224
• Overall SD: 9.91
• Overall CV: 4.42%
How well do smaller data sets
estimate mean and SD ?

Page 22
Estimating Mean and SD

For a Mean:
Mean SD
Overall 224
8 – 10 replicates can give a useful estimate
Overall 9.90
Data Mean Δ
Data SD Δ
Groups of 10 results Groups of 10 results For the SD:
1-10 221.7 -1.1% 1-10 9.88 -0.2% t An SD calculated from a single study of 20
11-20 230.1 2.6% 11-20 11.76 18.8% replicates is an estimate of the “true” SD
21-30 224.3 0.0% 21-30 10.41 5.2% t This estimate can be from 30% smaller to
31-40 223.7 -0.2% 31-40 9.86 -0.4%
25% larger than the “true” SD and still be
equivalent to the “true” SD
41-50 224.7 0.2% 41-50 8.17 -17.5%
t To reliably estimate the SD to within ±10%
51-60 226.7 1.1% 51-60 7.72 -22.0%
of the “true” value requires 100+ values
61-70 226.1 0.8% 61-70 9.15 -7.6%
t To conclude that a problem exists anytime
71-80 220.2 -1.8% 71-80 11.42 15.4% an SD calculated from a single study
Groups of 20 results exceeds the reference SD (IFU) is
statistically incorrect and unrealistic
1-20 11.42 15.4%

21-40 9.87 -0.3%

41-60 7.80 -21.2%


61-80 10.52 6.3%

Page 23
Validating Precision

Is measured precision acceptable ?


If measured SD & CV are less than or equal to the reference precision
performance, method performance is validated
If measured precision exceeds reference precision …
u Performance may not meet requirements

u However, remember limitations on SD calculated from limited data

From CLSI EP5-A2 (Evaluation of Precision Performance of Quantitative


Measurement Methods):
“The user’s (precision) estimate can be larger than the SD claimed by the
manufacturer and still be acceptable. Since the user experiment is based on a limited
number of observations, there are expected sampling errors of the calculated (precision)
around the true values. … The chi-squared test is used to determine if your estimates are
significantly larger than those provided by the manufacturer.”

Page 24
Chi Squared

Chi squared (c2): tests whether a measured SD is statistically


comparable to a reference SD

Used to compare estimated precision to published performance

ü Calculated X2 is compared to the critical


s 2 * (n-1)
X2 = value of c2 for 95% confidence and degrees
σ2 of freedom (n-1)
ü If X2 is less than the critical value, precision
estimate is not different from reference
n = number of replicates
value
S = calculated SD
σ = reference SD

Page 25
Using Chi Squared
Replicate Sample1 Sample2 Sample3
1 93.00 93.00 100.00
2 96.00 100.00 94.00
3 100.00 89.00 86.00
Tests whether precision data collected on-
4 93.00 87.00 91.00
5 89.00 86.00 102.00
site is comparable to expected
6 94.00 86.00 85.00 performance
7 87.00 92.00
8 89.00 102.00 Provides objective test to support
9 86.00 90.00 comparability, even if study SD exceeds
10 88.00 91.00 reference SD
11 86.00
12 91.00
13 92.00
14 91.00
15 102.00
16 90.00
17 90.00
18 85.00
19 91.00
20 92.00
Mean 91.25 91.60 93.00
SD 4.39 5.52 7.04
CV 4.8% 6.0% 7.6%
N 20 10 6
IFU Mean 95 95 95
IFU claimed CV 5.2% 5.2% 5.2%
Chi-square 15.8 12.5 12.2
Crit. Chi-square 31.4 18.3 12.6
CV acceptable Yes Yes Yes

Page 26
Chi squared - Assumptions & Limitations

The calculation is based on SD, therefore it is essential that the SD’s


of the validation study and reference data be comparable

Since the SD may vary with concentration, we typically use a


reference SD at a concentration close to the mean of the precision
study data

2 * (n-1)
χ
2 = s
σ2

Page 27
Precision
AMR
Method
Verification
Comparison
/ Range

Overview & Reference


Cautions Interval

Page 28
What

Method comparison: testing the same group of patient


patientsamples
samplesby two
different methods to characterize the
relationship between the methods
Goal: to create a model of the relationship between the methods that
can be used to estimate average method to method difference

120

100

80

60

40

20

0
0 20 40 60 80 100 120

Page 29
Why

Commonly used in clinical laboratories to:


• Characterize the difference between a new method and the current one
• Understand how results will change with the new method
• Provide information to alert customers (clinicians) to a change
Validate reference interval for a new method

Page 30
Typical Scenario

§ Lab tests samples by current method


§ Identifies samples that will be used for method comparison
(typically 10 – 40 samples)
§ Writes initial results on tubes and stores the samples
§ Tests stored samples using new method
§ Uses regression analysis to evaluate results.

Page 31
Now there is an issue

§ Methods do not seem to agree


• Difference between results is more than 10% for some samples
• Regression slope is not 1.0, or within 0.9–1.1, or some other limit
• Correlation coefficient (r) is 0.998, but the results don’t agree

§ Regression slope is not the same as the method comparison published in the package
insert
§ Difference between methods not the same as between respective peer groups on EQA
reports or QC peer group reports
§ Most common conclusion:
There is something wrong with the new method or instrument

Most common actual root causes:


• Study done using less than optimal samples
• Data does not conform to assumptions of statistical tool(s) used
• Interpretation of regression statistics
Page 32
Considerations when selecting samples

Use samples compatible with both methods


• Avoid samples containing interferents known to affect either method
• lipemia, hemolysis, drugs, etc.
• anticoagulants
• Samples must meet requirements for BOTH methods
• Avoid using QC samples, proficiency testing samples, spiked samples, diluted samples,
linearity samples, etc.
Storage
• Avoid using stored specimens if possible
• Do not store samples in original collection tubes
• Must be stored per manufacturer recommendations (both methods)
• If stored samples are used, must be run by both methods after storage

DO NOT use the original results obtained with fresh samples by one method to
compare with results obtained later by another method using the same samples
after being stored, especially if stored frozen

Page 33
Considerations when testing samples

Range of concentration
• Cover as much of the shared analytical range as possible
• Focus on the range of clinical interest
• Sample concentrations should ideally be uniformly distributed over the range
covered
• Single samples at extreme limits of the range are generally not useful – can
strongly bias statistical analysis 80000

Good
Poor distribution
• Be very cautious about using proficiency testing, QC or other manufactured
70000

60000
samples to extend the range of concentration due to potential for matrix
50000

issues 40000

• Samples requiring dilution by either method should not be included


30000

20000

• If dilutions are important, keep the data for diluted samples separate
10000

0
0 10000 20000 30000 40000 50000 60000 70000 80000 90000

Page 34
Testing Protocol

§ Samples should be tested by both methods within a few hours or, at worst,
within the same day
§ If using stored samples, do not thaw or prepare samples until the day they will
be tested
§ Ideally, study should be done over several days, testing a different group of
fresh samples each day
§ Ideally, both methods should be recalibrated during course of study
§ If using pooled, diluted, spiked, or manufactured samples (EQA, QC, etc.), make
sure these samples are identified so they can be tracked throughout the study

Page 35
Reviewing the data

Step 1: Plot the data


• Scatter plot
• Difference plot
• Plots provide most of the necessary information
14

12
30

25
10
20

Difference (µg/L)
15
Method 2

8
10

5
6
0
0 100 200 300 400 500 600 700 800 900
-5
4
-10

2 -15

-20

0 Concentration (µg/L)
0 2 4 6 8 10 12 14
Method 1

Page 36
Scatter Plots

14

12
Always
the same X Y
10
scale Mean of Methodmethod
Comparative 2 7.72Test method

Prior method New method


Y 2

8 Mean of Method 1 8.19


Method

Reference method Candidate method


6 N 33
Current method Replacement method
Control method Evaluation method
4

0
0 2 4 6 8 10 12 14
X 1
Method

Page 37
Difference Plots

§ Plot can show…


• Graphically shows the difference
• Numerical difference
between the two methods
• % difference
• Useful for estimating average
§ Choice depends on method variance
difference and noting trends
• If the SD remains constant over the range – use
numerical value
• Useful when data is not optimal for
• If the CV remains constant over the range – use %
regression
Difference
analysis
• Correct choice improves ability to estimate
average difference

Page 38
Scatter Plots: Constant SD vs Constant
CV

Constant SD Constant CV

17 Scatter Plot 900 Scatter Plot

800
16

700
15

600

Test Method (µg/L)


Test Method (µg/L)

14
Identity
500

13
400 Identity

12
300

11 SD = +/- X 200 CV = +/- X%


10
At all concentrations 100
At all concentrations
0
9
0 100 200 300 400 500 600 700 800 900
9 10 11 12 13 14 15 16 17

Comparative Method (µg/L) Comparative Method (µg/L)

Typically chemistry methods Typically immunoassay methods

Page 39
Bland Altman Plot

60

Difference (New Method - Comparative Method)


§ Adaption of difference plot
40
§ Y data can be numeric or %
§ X data is average of both methods 20 +2 SD

§ Adds limits lines: 0

• Usually +/- 2 SD around the average


difference -20

• Can be any predetermined acceptable -40


limit
-2 SD
-60
50 150 250 350 450 550
Mean of Both

SD = standard deviation of the differences

Page 40
Reviewing the Data

Step 1: Plot the data


• Difference plot
• Scatter plot
• Plots provide most of the necessary information
Step 2: Regression analysis
• If appropriate – not always the best tool
• Powerful tool for estimating method to method bias

Page 41
Linear Regression

Linear regression analysis is a statistical tool

Used properly, provides an accurate, robust model of the relationship between


methods that can be very useful

Used carelessly, it can provide highly misleading misinformation

When using regression analysis, always keep two rules in mind …

Rule #1: Rule #2:


Always plot the data! Never use By eye, draw the best straight line you can
regression statistics without looking through the data evenly dividing the points on
at a graph both sides of the line.
If the calculated regression line is significantly
different from your line, do NOT trust the
calculated regression line

Page 42
Regression Analysis

Yields two parameter equation describing the relationship between


the methods: y = a*x + b
• Parameter 1: a = slope
• Proportional bias - Percentage difference
• If slope is 0.9 ð “y” results will be 90% of “x” results
• If both methods agree completely = 1.000
• Parameter 2: b = y intercept
• Constant bias - difference is same amount regardless of
concentration
• If both methods agree completely = 0
• Intercept estimated from equation: b = y – (a*x)
• using mean values for “X” and “Y” data

• Parameters are estimates


• Estimate is only as good as the data used
• More data gives a better estimate
• Different data sets can give somewhat different estimates
Page 43 • 95% confidence interval used to characterize estimate
Testing for adequate range

Correlation coefficient (r)


• Statistical estimate of how well the results of the two methods vary together
• If all data points fall exactly on the line: r = 1.000
• If data is randomly scattered with no relationship: r = 0.000
• Best used to confirm that range of values used was adequate to support use
of least squares linear regression
• To use least squares regression, ideally r ≥ 0.975
• Indicates that the range of data values used is adequate to compensate
for the actual scatter in the “X” data
• If r < 0.975 – use alternate approaches to estimate difference
• Use alternate regression models
• Estimate average difference from difference plot

Page 44
Misconceptions about correlation
coefficient

Common misconception about r….


is that it indicates agreement
Troponin I
Correlation coefficient tells us what 30

percentage of the differences between 25


methods is accounted for by the regression
line 20

Method 2
Good correlation means all the points fall on or 15

very near the line


10

It does not mean results agree numerically or


that the slope = 1 5

0
0 5 10 15 20 25 30

r ≠ Bias
Method 1

Method 2 = 0.37 (Method 1) + 0.11


r = 0.993

Page 45
Dealing with low r value

“Low” r value (less than 0.975)


• Generally indicates there is not enough data and /or it does not do not span enough range
for least squares to provide a good model of the relationship
• It also indicates scatter in the data which may point to assay specific matrix or interference
issues
• REMEMBER: the scatter is NOT always due to the new method –
current method may have poorer precision or interference / matrix issues

How to address:
• Obtain additional samples that span broader concentration range
• Often not practical – especially for some methods

• Instead of regression analysis, use difference plot and estimate average difference
• Use an alternate regression model (Weighted, Deming, Passing-Bablock)

Page 46
Case Studies
Dealing with data
Theory is great, but my data never looks like that !

Page 47
Case #1: Few data points

Method 1 Method 2 Bias % Bias


2.6 2.4 0.2 9.2%
2.8 3.0 -0.2 -8.2%
3.0 3.6 -0.6 -19.7%
3.6 4.1 -0.5 -13.6%
4.2 4.7 -0.5 -11.0%
4.9 5.2 -0.3 -6.1%
5.0 5.4 -0.4 -7.4%
6.1 6.7 -0.6 -9.3%
6.2 6.2 0.0 0.2%
7.2 9.0 -1.8 -25.4%
7.8 8.0 -0.2 -2.8%
7.9 7.8 0.1 1.8%
9.6 10.4 -0.8 -8.3%
10.9 11.3 -0.4 -3.7%
11.8 13.1 -1.3 -11.0%
11.9 12.2 -0.3 -2.5%
12.0 11.7 0.3 2.5%
12.4 11.7 0.7 5.6%
12.4 12.9 -0.5 -4.0%
13.8 14.2 -0.4 -2.9%

Average Bias: -0.4 -5.8%

Page 48
Difference plot

For regression analysis to be valid, need adequate data


Total T4
40 samples is recommended
4
20 – 40 may work, but may not 3.5 Average difference = 0.4
be reliable

Difference (M ethod 2 - M ethod 1)


3

Less than 20 … not worth doing 2.5

regression 2

1.5
How to address: 1
0.5
Test additional samples or …..
0
Estimate average difference -0.5

-1
Use difference plot
2 4 6 8 10 12 14 16
Mean of All

Page 49
Case #2: Troponin I

Number of Samples 30
14
Range of Observations 0.016 to 12.1
Correlation Coefficient (r) 0.999
12
Linear Slope 0.77

Linear Intercept 0.02


10

Most samples are “normal”


New Method
8

Only a very few are high 6

No samples are near upper assay limit


4
Limited number of high samples may
skew regression line
2

0
0 2 4 6 8 10 12 14

Comparative Method

Page 50
Troponin: Difference plot

Difference plot shows essentially no difference


between methods except for highest two 1

samples 0.5

D iffe re nc e (N e w M e thod - C om pa ra tiv e M e thod)


Regression slope, influenced by two high data 0
points, doesn’t show real relationship
-0.5

No difference noted at medical decision point


-1

Can get more samples if practical, but this shows -1.5


how methods relate at decision point
-2

-2.5

-3

-3.5
0.01 0.1 1 10 100
Comparative Method

Page 51
Troponin I: Excluding extreme samples

Number of Samples 26
0.11
Range of Observations 0.016 to 0.1
Correlation Coefficient (r) 0.962 0.1
Linear Slope 1.00
0.09
Linear Intercept 0.00

0.08

Allows focus on clinical decision 0.07

New Method
range 0.06

Shows method correlation at 0.05

decision point 0.04

0.03

0.02

0.01
0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 0.1 0.11

Comparative Method

Page 52
Case# 3

90,000
HCG
Number of Samples 18
80,000
Range of Observations 1.82 to 68,651

Correlation Coefficient (r) 1.000 70,000

Linear Slope 0.850


60,000
Linear Intercept 168
50,000
New

40,000

30,000

20,000

10,000

0
0 10,000 20,000 30,000 40,000 50,000 60,000 70,000 80,000 90,000

Comparison

Page 53
Dealing with diluted samples

With dilutions removed agreement HCG


900
much better

More accurate model of relationship 800

Do NOT include dilutions 700

All guidelines state dilutions should 600


not be used
500
• CLSI EP9
New

• Articles/books on method comparison 400

If dilutions are important – compare 300


separately Number of Samples 12
200 Range of Observations 1.82 to 848
Correlation Coefficient (r) 1.000
Linear Slope 0.990
100
Linear Intercept -2.19

0
0 100 200 300 400 500 600 700 800 900

Comparison

Page 54
Case #4

25 Absolute Neutrophil Count


Number of Samples 50
Range of Observations 0.09 to 23.05
Correlation Coefficient (r) 0.991 20

Linear Slope 0.893


Linear Intercept 0.399
15

New Method

10

0
0 5 10 15 20 25

Comparison Method

Page 55
Case #4 - suggestions

Several extreme results influencing line


Strategy: 25 Absolute Neutrophil Count

1. Collect more samples to fill in range


or..
20

2. Retest samples with discrepant


results to r/o testing error
15
3. Use difference plot

New Method
10

0
0 5 10 15 20 25

Comparison Method

Page 56
Case #4: Difference plot

Difference plot shows minimal bias


between methods Absolute Neutrophil Count
1.5

Overall method agreement is good

Difference (New Method - Com parison Method)


Only a couple of samples show larger 0.5

difference
• Is that difference significant ? -0.5

• Could it be an error ?
-1.5
Can collect more samples in higher
range, but if that is not practical this
-2.5
shows overall excellent agreement
-3.5
0 5 10 15 20 25
Comparison Method

Page 57
Dealing with Outliers, etc.

What’s an outlier ?
A result that does not represent overall relationship / performance due
to some sample specific characteristic
How do you know it’s an outlier ?
• EP9-A3:
• Samples with method to method differences more than 4x the
average difference can be considered outliers.
• Up to 2.5% of data points, identified as outliers, can be deleted
without re-assessing the study (1 point in 40)
• Statisticians recommend against removing outliers if no specific cause
can be found – suggest using statistics robust to outliers

Page 58
Outliers

Whenever possible find out reason for outlier


• Problem sample – interferences, fibrin, etc.
• Error on run – short sample
• Remember – “error” or interference may not be the new method

Exclude data with identified errors


• These results do NOT represent typical performance of the methods
• Including this data will make estimate of relationship inaccurate

Handle unexplained “outliers” as for extreme values


• Outliers near upper or lower limits of data may affect regression statistics
• Outliers near middle of the data range have less effect
• If using regression, use robust regression tool

Page 59
Case # 7

1800
LD
1600

1400

1200

New Method
1000

800

Nothing 600

Indicates real method 400

to method difference 200

0
0 200 400 600 800 1000 1200 1400 1600 1800
Comparison Method

Page 60
True Method to Method Difference

Not very common, but it happens 1800


LD
Validate by looking at other comparisons for 1600

same two methods 1400

Look for data from similar studies with same 1200


method – confirm this is what is expected

New Method
1000

Alternatively, use Truth Table/ Concordance


report to evaluate clinical equivalence 800

600
Will need to update reference interval and
notify clinicians of expected change to results 400

200

0
0 200 400 600 800 1000 1200 1400 1600 1800
Comparison Method

Page 61
Truth Table and Concordance

Used with:
• qualitative assays (infectious disease)
• large method to method differences to
show clinical equivalence

Page 62
Concordance and Quantitative Methods

§ Regression analysis has slope very


different from 1.0
§ Reference intervals are not similar
§ Concordance helps focus on clinical
interpretation rather than just looking at
numbers

Comparison Method

Reference Intervals: New Method


Comparison Method: Less than 30 U/ml
New Method: Less than 60 U/ml

Page 63
Precision
AMR
Method
Verification
Comparison
& Range

Overview & Reference


Cautions Interval

Page 64
Transferring Reference Intervals

Key: reference interval is being verified not established


Guidance:
C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory
• Establishing a reference interval requires at least 120 samples from reference individuals
• Verifying or transferring an established reference interval can be done with 20 samples

CLSI C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory (2008), p28

Page 65
Verifying a Reference Interval (CLSI C28-
A3)

1. Regression analysis
• If regression results are robust (enough samples, good distribution,
appropriate regression model, etc.) can use regression equation
• New Reference limit = slope x old reference limit + intercept
• If calculated new limits match proposed reference interval – verified
2. Using small study
• Select 20 individuals that match criteria used for proposed reference interval
and analyze with method
• If no more than 2 results exceed limits of proposed reference interval –
verified
3. Subjective judgment
• On careful review of all data and description of how proposed reference
interval was established, lab director may accept new reference interval
Page 66
Summary

• Method Validation studies are very useful, good lab practice, and
generally required by regulation

• Remember, all studies provide only estimates of precision, method


agreement, etc. – estimates will vary from study to study

• Keep the assumptions and limitations of all statistical tools used in mind
when reviewing studies

• There is no simple single answer for all data sets; each requires thoughtful
selection of the correct tool for the situation

Page 67

Das könnte Ihnen auch gefallen