RBE
IWIS container

Integral solution
for Infectious Waste Integral
System (IWIS) Technical features
MATACHANA offers a global solution for the infectious
waste treatment in a container, equipped with all the Model IWIS 1006 IWIS 1008 IWIS 1010
necessary elements to inactivate the potentially Overall dimensions 6058 x 2591 x 2438 mm 6058 x 2591 x 2438 mm 6058 x 2591 x 2438 mm
contaminated materials with the best guaranties. (width x height x depth)
We provide a complete solution for the waste storage, Sterilizer 1006 E-1 1008 E-1 1010 E-1
sterilization and post shredding, optimizing the space.
This container can be installed in an external place, near to Liters in chamber 445 565 774
the healthcare centre with no workmanship required.
This means a clear advantage over traditional waste Load capacity/approx. cycle 45 - 75 65 - 105 80 - 135
treatment plants. (kg/cycle) (*)
Total power (kW) 38.5 60.5 72.5

This system is composed by: Maximum total weight (kg)(**) 5230 5390 5820
• REFRIGERATION CHAMBER – To store the infected waste waiting to be processed in the sterilizer. With (*) Production can vary depending on the loading system and receptacles used
two 240 l. DIN container storage capacity (included).
(**) With the autoclave full of water for the hydraulic test
• STERILIZER – To process with saturated steam the infectious waste. Different models under request,
depending on the bed numbers of the centre (this determines the volume of the waste to be pro-
cessed). With built-in electric steam generator. With air and condensates treatment system. “Plug and play” design & installation.
• SHREDDER – Shredder for the sterile waste. With space to allocate under it a 240 l. DIN container. Specially developed so that when it
has reached its final installation place,
• WASHBASIN – Made of stainless steel. Pedal-operated.
only electricity, water and drainage
• ADDITIONAL EQUIPMENT – It is fitted with additional equipment to cover the container needs (water facilities (to an existing connection)
softener, air compressor). are required.
• ACCESS RAMP AND SEPARATION BETWEEN ZONES – With an access ramp for the easy movement It can be installed in an annex
location to the site, occupying the
of the 240 l. containers with wheels, inside and outside the IWIS working area.
space of only two car places.
A variant of this container is available
The IWIS is designed with two totally separated areas with different access points: a working area on request, prepared to be installed
where the waste is processed and a technical area for the maintenance of the equipment (with an in very cold countries or places.
access door placed in the lateral side of the container).

67
www.matachana.com

RBE
Sterilizers general features

Infectious Medical Waste Treatment


• Technology:
New control system more powerful and modern. TFT touch
screen for loading area with integrated PLC. Frontal made of
Matachana RBE an innovative material that prevents dirt and reflections. Easy
to clean.
Steam sterilization • Ergonomics:
SC500 RBE More accessible displays with a modern and versatile inferface.
The MATACHANA SC500, S1000 and S2000 RBE sterilizers can cope with the
needs of any hospital. The only difference between all the series is the volume • Sustainability:
of waste they can treat. These series of sterilizers use a “clean” treatment system. Energy and water consumption reduced thanks to incorporated
The air extracted and the liquids generated during the process are treated economizer systems.
using a special technique and monitored by a microprocessor control unit to
ensure total asepsis without any environmental contamination. The autoclave • Efficiency:
technique by means of fractionated vacuum is the most efficient alternative for Redesigning of vacuum system by Venturi.
the treatment of special biocontaminated waste. It is reliable, easy to operate,
S1000 RBE
has low operating costs and it is a clean system, as it does not produce any • Comfort:
contaminants whatsoever. New range of more functional accessories, adapted to the
loading requirements.

• Accessibility:
Full circuits and components redistribution that facilitates
the accessibility of the Technical Support Service for easy
maintenance procedures.

• Remote diagnosis:
RBE programs Remote connection via Ethernet port to know the sterilizer’s
condition and enable the Technical Support Service to diagnose
The MATACHANA sterilizers, for the treatment of special healthcare waste, incidences/failures remotely for faster solution.
operate with fractionated vacuum and sterilization programs to 134 ºC tem-
perature. Standard type sterilizers and specifically designed sterilizers to treat S2000 RBE
special contaminated waste must be differentiated. To serve this purpose,
MATACHANA has developed its special line of RBE sterilization, with the fol-
lowing distinctive features:
PRESSURE

Technical features
RBE Model Chamber dimensions Overall dimensions Usable chamber
The L3 sterilization programs are progressing with the same stages as the Series (width x height x depth) (width x height x depth) volume
standard solid programs, except for the particularities stated below that SC500 SC501 VR/ER-1 RBE 1 door 500 x 500 x 1020 mm 900 x 1824 x 1342 mm 252 l
prevent the risk of contamination towards the outside.
SC501 VR/ER-2 RBE 2 doors 500 x 500 x 1020 mm 900 x 1824 x 1355 mm 252 l

• The condensates generated during the phases previous to sterilization re- SC502 VR/ER-1 RBE 1 door 500 x 500 x 1300 mm 900 x 1824 x 1622 mm 321 l
Start
TIME
End of cycle
main inside the chamber and are sterilized during the sterilization phase. SC502 VR/ER-2 RBE 2 doors 500 x 500 x 1300 mm 900 x 1824 x 1635 mm 321 l
S1000 1006 VR/ER-1 RBE 1 door 670 x 670 x 998 mm 996 x 1954 x 1314 mm 445 l
• The potentially contaminated air leaving the chamber during the pro- 1006 VR/ER-2 RBE 2 doors 670 x 670 x 996 mm 996 x 1954 x 1336 mm 445 l
Cycle profile of a sterilization program cess passes through a 0.02 microns HEPA absolute filter, ensuring
non-contamination towards the outside. 1008 VR/ER-1 RBE 1 door 670 x 670 x 1296 mm 996 x 1954 x 1614 mm 565 l
1008 VR/ER-2 RBE 2 doors 670 x 670 x 1265 mm 996 x 1954 x 1636 mm 578 l
• Treatment of fluid evacuated from the chamber at the beginning of the
1010 VR/ER-1 RBE 1 door 670 x 670 x 1735 mm 996 x 1954 x 2054 mm 774 l
cycle, when the load has not been sterilized yet. These fluids may be liquid
(condensate) or gaseous (air) and may contain microorganisms susceptible to 1010 VR/ER-2 RBE 2 doors 670 x 670 x 1733 mm 996 x 1954 x 2076 mm 773 l
be pollutants. They are therefore treated by a special system that guarantees 1012 VR/ER-1 RBE 1 door 670 x 670 x 2000 mm 996 x 1954 x 2316 mm 893 l
sterility before being discharged into the general drainage and environment.
1012 VR/ER-2 RBE 2 doors 670 x 670 x 1998 mm 996 x 1954 x 2338 mm 892 l
S2000 2.1830 LVR-1 RBE 1 door 960 x 1440 x 1520 mm 2390 x 1950 x 1880 mm 2100 l
• The sterilizer is provided with two types of processes with fixed times
and temperatures. They are used according to the type and weight of the 2.1830 LVR-2 RBE 2 doors 960 x 1440 x 1520 mm 2390 x 1950 x 1980 mm 2100 l
container in which the contaminated waste is housed. Thus, it is important 2.3730 LVR-1 RBE 1 door 960 x 1440 x 2290 mm 2390 x 1950 x 3330 mm 4037 l
to distinguish the different densities and containers’ type for the load to be 2.3730 LVR-2 RBE 2 doors 960 x 1440 x 2290 mm 2390 x 1950 x 3430 mm 4037 l
processed, such as autoclavable bags, semi-rigid containers (cartons with
inner polyamide bag) and rigid containers. 2.6200 LVR-1 RBE 1 door 960 x 1440 x 4300 mm 2390 x 1950 x 4670 mm 5920 l
2.6200 LVR-2 RBE 2 doors 960 x 1440 x 4300 mm 2390 x 1950 x 4770 mm 5920 l

89
www.matachana.com

RBE

Infectious Healthcare Waste Treatment


Infectious Healthcare Waste Treatment
Consumables
Post-treatment
With the aim to reduce the waste volume and morphology, shredding and Autoclavable bags
compacting equipment may be included after the sterilization process in
accordance with the user’s specific requirements. Made in red polypropylene, labelled with infectious materials
identification (international symbol).
Shredder with lifting and tilting Include sterilization control indicator.
device for large production plants We are totally in favour of minimal manipulation in regards to management Supplied in boxes of 200 u.
and maintenance. Therefore, we firmly believe that the shredding/compacting Each box comes with a bag of 200 flanges for closing the
process to reduce volume should be performed at a later stage to sterilization waste bag.
to minimize contamination risks.
Bag 50 liters 100 liters
The containers used for the treatment of healthcare waste inside the sterilizer Dimensions 64 x 90 cm 80 x 100 cm
chamber, are designed and manufactured in their shape, size and Thickness 50 microns 50 microns
anchoring system according to DIN 30700. This makes them compatible with
the lifting and tilting mechanism used by the urban waste transport.

After sterilization, should you wish to transfer the waste to a rented industrial
compactor from the center; this way avoiding any manipulation, you need to
place a lifting and tilting system.

Shredder for treatment Monitoring systems for


plants
Other accessories steam sterilization
Several ways are available to monitor the steam sterilization
Depending on the required production/project, the following equipment
cycles and ensure that the processed material has been
shall be considered in order to complete and facilitate the workflow in a waste
properly sterilized.
treatment plant:
- Containers weighing scale Bowie & Dick Test
Chemical controls
- Radioactivity detector
- Waste cooling before sterilization There are different types of indicators which are sensitive
- Waste containers and bags for collection and treatment to process variables, providing important information about
- Water treatment the cycle. For example: adhesive tapes, Bowie & Dick test,
multidual indicators, integrator indicators or emulator
- Steam generator chemical indicators.
- Refrigeration and recycling of sterilizer’s clean condensates All these chemical indicators must comply with the International
Multipack 650 Chemical controls
Standard EN ISO 11140.
waste compactor
Biological controls
They monitor the efficiency of the sterilization process by using
real living organisms. The spores and medium will activate
themselves in order to enter in contact after completion of
the cycle and thus verify the effectiveness of process in real
conditions. As a rule, a 24 hours incubation is sufficient to
ensure that the sterilization process is successful.
There are other forms of biological indicators such as dry
strip, metal plate, discs, etc., with different incubation times
according to the systems and formats.
Biological indicators must comply with the International
Biological controls Standard EN ISO 11138.

Water treatment Refrigerated chamber for big-sized