Avalon Cableless Monitoring

Avalon CL Base Station 866074


Avalon CL Toco+ MP Transducer 866075
Avalon CL Ultrasound Transducer 866076
Avalon CL ECG/IUP Transducer 866077
Technical Data Sheet

The Avalon CL solution offeres the choice of freedom eliminating cables for all possible
fetal and maternal measurements. With the Avalon CL you have the full performance and
features available as with the wired Smart Transducers: same ultrasound performance,
monitoring twins and triplets, Smart Pulse, and Transducer Finder LED. Maternal
measurement can be also monitored wireless using the IntelliVue CL Pods.

The Avalon CL combines convenience and versatility with ease of use; freeing the
mother from cables during all stages of care, especially in labor and delivery.
Features Assignment and Charging Slots
It offers: There are three CL transducer slots and one slot for the CL Pods. You
• Cableless, waterproof, light-weight transducers can assign up to four CL transducers and two CL Pods. Whenever the
• Constant strong performance of the CL ultrasound transducers Avalon base station is connected to a fetal monitor, the docked
throughout the operating time CL transducers and CL Pods are automatically assigned and recharged.
• Higher clinical confidence, by allowing expectant mothers
continuously be monitored, including during a bath or shower, or Transducers
while ambulating Each cableless transducer has an LED that lights up in multiple colors
• On demand audible patient call from the fetal monitor to indicate, for example, the battery charge status of the transducer, or
• Cableless monitoring for twins and triplets the transducer’s connectivity to the base station.
• Reliable fetal parameter signal quality with a line-of-sight operating For monitoring twins and triplets, additional transducers can be easily
range of up to 100 m/300 ft1 assigned by briefly docking them in the CL base station.
• Automatic out-of-range notification for the patient
• User-friendly intuitive controls at the display of the connected fetal Radio Frequency Transmission
monitor The Avalon CL Transducers transmit measurement data via radio to
• High capacity, rechargeable Lithium-Ion batteries, combining an the CL Base Station which provides it to the connected fetal monitor.
operating time of more than 8 hours per charge cycle with a short The transducers can be configured to available local radio frequency
recharge time of less than 3 hours channels operating within the country specific frequency range.
• Easy and quick assignment of transducers and CL Pods by simply Be sure to comply with any local regulations regarding approved
docking them at the base station. frequencies, and contact Philips for assistance with your specific
• The base station connects to any fetal sensor socket on the Philips requirements.
Avalon FM20, FM30, FM40 and FM50 fetal monitors, or to the rear
telemetry connectors on the FM40 and FM50. Frequency Ranges
• Eliminates the need of a power cable to the mains. Hardware RF
Country/Country Group Tuning Range
Description (MHz)
The Avalon CL solution consists of a base station and the following Japan 920.6 – 923.4
optional transducers: Most European Countries 433.05 – 434.79
• A cableless US transducer measuring fetal heart rate and fetal (covers Industrial, Scientific, and Medical (ISM)
movement profile band)
• A cableless Toco+MP transducer measuring uterine activity externally, USA, Canada, Australia 608 – 614
maternal pulse, and either DECG, MECG, or IUP (covers Wireless Medical Telemetry Service
• A cableless ECG/IUP transducer measuring either DECG, MECG, or (WMTS) band)
IUP
Alarms
• In addition the base station provides an interface to the IntelliVue CL
Alarms are available for the loss of the radio connection, low/out of
SpO2 Pod and IntelliVue CL NBP Pod, refer to the Technical Data
battery capacity, malfunctioning of the cableless transducers.
Sheet of the IntelliVue Cableless Measurements

Safety Features
Operation
• Optical alarm before the battery is empty
The Avalon base station is configured and operated via the touch
• Optical and acoustic alert if the patient moves out of the range of
screen of the connected fetal monitor.
monitoring
• Automatic resume after power failure

1
The effective operating range can vary depending on the geographical and constructional
characteristics of the building where the system is installed, and whether underwater
monitoring is performed.

2
Interface Capabilities • CAN/CSA C22.2#60601-1-08+TC2:2011
• JIS T 1303 2005
Avalon FM20/FM30/FM40/FM50 Fetal Monitors • AS/NZS 3200.1.0-1998
The Avalon CL interfaces to the Avalon family1 of fetal monitors via the The possibility of hazards arising from hardware and software errors
cable. The cable connects to any of the four red fetal sensor sockets on was minimized in compliance with ISO 14971:2012,
the monitor with the red connector. Alternatively for the FM40/FM50, IEC 60601-1:2005+A1:2012/EN 60601-1:2006+AC:2010.
an interface cable with a black connector connects to the black
telemetry sockets at the rear of the monitor. Whether the Avalon CL Electromagnetic Compatibility
base station is delivered with a red or black connector is an ordering • IEC 60601-1-2:2007/EN 60601-1-2:2007+AC:2010
option (K30/K40). • ICES-003:2012
866074 is classified as electromagnetic emissions Class A.
Related Products
For monitoring the maternal parameters you can use the IntelliVue Radio
CL SpO2 Pod and the IntelliVue CL NBP Pod with the Avalon CL base • ETSI EN 300 220-1:2012
station. Refer to the Technical Data Sheet for the IntelliVue Cableless • ETSI EN 300 220-2:2012
Measurements for details on their features. • ETSI EN 301 489-1:2011
• ETSI EN 301 489-3:2003
• FCC 47 CFR Part 95
• IC RSS-210 Issue 8
• ARIB STD-T108
• ETSI EN 300 328:2006
• ETSI EN 301 489-17:2009
• FCC 47 CFR Part 2 & 15
• AS/NSZ 4268
• AS/NSZ 4771:2000+A1:2003
The Avalon base station communicates vie a built-in short range radio
• ARIB STD-T66
and integrated antenna with the IntelliVue CL Pods (operating range:
in-room). The option to monitor with the IntelliVue CL Pods is not yet
Environmental Specifications
available for the FM20/30 #E25 Battery Option.
The base station and cableless transducers may not meet the given
performance specifications if stored and used outside the specified
Standards Compliance
temperature and humidity ranges.
Federal Law in the USA restricts this device to use by or on the order
of a physician. Environmental Specifications
The Avalon CL complies with the essential requirements of the Medical (Transducers and Base Station)
Devices Directive (93/42/EEC) and the Radio and Telecommunications Temperature Operating 0 – 45°C (32 – 113 °F)
Terminal Equipment Directive (1999/5/EC), and with the following Range
standards: Base Station
Transducers Operating 0 – 40°C (32 – 113 °F)
Safety, Performance Storage -20°C – +60°C (-4 – 140°F)
• IEC 60601-1:2005+A1:2012/EN 60601-1:2006+AC:2010 Humidity Range Operating <95% relative humidity @
• IEC 60601-1-6:2010/EN 60601-1-6:2010 45°C (113°F)
• IEC 60601-1-8:2006/EN 60601-1-8:2007+AC:2010 Storage <95% relative humidity @
• IEC 60601-2-27:2011+AC:2012 / EN 60601-2-27:2006+AC:2006* (* 60°C (140°F)
except clauses 201.6.2, 201.8.5.5, 201.12.1.101, 202.6.2.101)
• IEC 60601-2-37:2007 / EN 60601-2-37:2008
• IEC 60601-2-49:2011/EN 60601-2-49:2001
• ANSI/AAMI ES60601-1+C1:2009+A2:2010

1
Requires monitor software revision J.03.00 or higher.

3
Base Station Specifications Transducer Specifications
Battery Type Integrated rechargeable Li-ion
Base Station Specifications
battery with a battery gauge
Power Powered by connected
and a cycle counter
fetal monitor
Runtime (fully min. eight hours
Dimensions Size mm/(in): width x 349 x 183 x 74 mm
charged
and Weight depth x height (13.7 x 7.2 x 2.9 in)
battery)
Weight 0.985 kg (2.2 lbs)
Life > 500 charge/discharge cycles
without transducers
or older than 4 years
Type of Class II equipment
Transducer ≥ 1 year at 25°C (77°F)
Protection
Storage Time (battery should be charged to
Against
40% - 50% every six months)
Electrical
Shock Recharging 100% charged ≤ 3.0 h
Operating Receiving Range (line approximately 100 m Time
Range of sight) (300 ft.1) Recharging 100% charged >6.0 h
Water Ingress IP 31 Time with an
Protection FM20/30 #E25
Code Degree of For US, ECG/ Type CF
1 The effective operating range can vary depending on the geographical and
constructional characteristics of the building where the system is installed, and whether Protection IUP (DECG/
underwater monitoring is performed. Against MECG), and
Electrical Toco+MP
Transducer Specifications
Shock parameters.
Transducers
Transducer Specifications
are battery
Dimensions and Weight
operated.
Toco+MP Size (diameter/ 75.5 mm/36.6 mm (2.9/1.4 in)
Water Ingress IP 68 (1.0 m immersion for 5 hours)
Transducer height)
Protection
866075 Weight 123 g ±10% (4.3 oz ±10%)
Code
US Size (diameter/ 75.5 mm/36.6 mm (2.9/1.4 in)
Usability 1.0 m
Transducer height)
Underwater
866076
Weight 131 g ±10% (4.6 oz ±10%) Shock Withstands a 1.5 m drop to concrete surface with
ECG/IUP 75.5 mm/36.6 mm (2.9/1.4 in) Resistance possible cosmetic damage
Size (diameter/
Transducer height) Nominal RF 0.1 mW ERP (typical)
866077 Output Power
Weight 121 g ±10% (4.2 oz ±10%)

4
Performance Specifications
IUP
CL US
Measurement Passive Resistive Strain Gauge Elements
US Intensity Average output power P = (12.4 ± 0.4) mW
Method
Peak-negative acoustic p_ = (49.1 ± 5.2) kPa
Measurement Range -100 – +300 mmHg
pressure
Signal Range -99 – 127 mmHg or -13.2 – 16.9 kPa
Output beam intensity Isata =
Sensitivity 5 μV/V/mmHg
(Iob) (2.77 ± 0.56) mW/
Offset Compensation +100 – -200 mmHg
(= spatial average - cm2
Accuracy (not including sensor ± 0.5% per 100 mmHg
temporal average
accuracy)
intensity)
Auto Offset 3 seconds after connecting the transducer, the
Spatial-peak temporal Ispta =
Correction IUP value is set to 0 mmHg
average intensity (21.1 ± 5.1) mW/cm2
Effective radiating area A-12 dB =
@ -12 dB (4.47 ± 0.89) cm2 ECG
Thermal index (TI) and mechanical index (MI) are
Performance DECG Single Lead ECG (derived
always below 1.0.
Specifications Type from Fetal Scalp Electrode)
Beat-to-Beat change (max.) for Ultrasound 28 bpm
MECG Single Lead ECG (derived
US Frequency 1 MHz ± 100 Hz
from RA and LA electrodes)
US Signal range 3.5 μVpp to
Measurement Range 30 – 240 bpm
350 μVpp @ 200 Hz
Wave Speed (Global 6.25 mm/sec, 12.5 mm/sec, 25 mm/sec,
US Burst Repetition Rate 3.0 kHz
Speed) 50 mm/ sec
Duration ≤ 100 μs
Accuracy ± 1 bpm or 1%, whichever is greater
FMP Signal Range @ 33 Hz 200 μVpp to 40 μVpp
(non-averaging)
Beat-to-Beat change MECG: 28 bpm DECG: 28 bpm (with
Toco (max.) Artifact Suppression On)
Measurement Method Strain Gauge Sensor Differential Input Impedance > 15 MΩ
Element Electrode Offset Potential ± 400 mV
Sensitivity 1 unit = 2.5 g Tolerance
Measurement Range 400 units INOP Auxiliary Current (Leads Off < 100 μA
Signal Range 0 - 127 units Detection)
Maximum Offset Range -300 units Input Signal Range DECG 20 μVpp – 6 mVpp
Baseline Setting 20 units MECG 150 μVpp – 6 mVpp
Auto Offset 3 seconds after connecting the transducer, Dielectric Strength 1500 Vrms
Correction the Toco value is set to 20 units Defibrillator Protection None
Auto Zero Adjust Toco value is set to zero following a negative ESU Protection None
measurement value for 5 seconds Paced pulse detection None

Complies with IEC 60601-2-27:2011+AC:2012/EN 60601-2-


Maternal Pulse from Toco 27:2006+AC:2006 except clauses listed
Emitted Light Energy ≤ 15 mW below:
Wavelength Range 780 – 1100 nm • 201.6.2, 201.8.5.5
Range 40 – 240 bpm • 201.12.1.101
Accuracy ± 2% or 1 bpm, whichever is • 202.6.2.101
greater Sub clause 201.12.1.101.15 on heart rate range and accuracy is passed.
Update Rate every 4 seconds

5
Ordering Guide and Accessory Options Option
Description
You can order the Avalon CL Cableless Fetal Transducer System No.
under the product number 866074. To order, precede the required K03 3 Avalon CL Ultrasound Transducers
option number with 866074 (for example, you would order a base
station together with one Toco and one US transducer using 866074
K11 1 Avalon CL Toco+ MP Transducer
#A02). “K” options let you modify your order. Below you find tables K21 1 Avalon CL ECG/IUP Transducer
giving you a quick overview of the various options. Refer to “Standard
Accessories Included” on page 6 for a list of items shipped as part of
K22 2 Avalon CL ECG/IUP Transducers
the standard configuration. Order upgrade options by using the prefix
866074 AU. Standard Accessories Included

The following accessories are supplied with the Avalon CL:


System Hardware Components
System Accessories and Supplies
Product
System Hardware Components
Number Order
Accessory
Base Station 866074 Number
Cableless Toco Transducer 866075 Waterproof belts M1562B
Cableless Ultrasound Transducer 866076
Cableless ECG Transducer 866077 Disposable abdominal belts (case of 100) M2208A
Mounting brackets order directly Ultrasound 12 Bottles 40483A
from GCX gel
5 liter refill (with dispenser) 40483B
for 40483A
Shelf life: at least 6 months
866074 Modification Options
max.

DECG DECG reusable leg plate 9898 031 37651


866074 Modification Options
accessories adapter cable (with flushing
Option New Philips port)
Description DECG
No. DECG leg attachment 9898 031 39771
Solution
K30 Avalon CL (red connector), for connection to fetal electrode for the DECG leg
plate adapter cable
sensor sockets on Avalon FM20/FM30/FM40/FM50
fetal monitors (monitor requires software revision J.3 DECG fetal scalp electrode: 9898 031 37631
or higher) single spiral, worldwide
K40 Avalon CL (black connector), for connection to availability
Avalon FM40/FM50 fetal monitors (for rear DECG fetal scalp electrode: 9898 031 37641
connection (monitor requires software revision J.3 or double spiral, Europe only.
higher) Not for USA
K60 External Power Supply for the Avalon CL base station
MECG adapter cable M1363A
(only required if the base station is used as a charging
station; not yet available) MECG electrodes 40493E

Battery exchange kit 453564423251


Transducer Options

Option
Description
No.
K01 1 Avalon CL Ultrasound Transducer

K02 2 Avalon CL Ultrasound Transducers

6
Mounting Options

Upgrade Options

Product Option No. Description


M2740A C02 Avalon FM Cart

U01 Mounting Kit for Avalon FM Cart

Mounting Solutions
There are several ways to mount the Avalon CL:
• On a dedicated mounting shelf, available for Avalon FM Cart (see
“Mounting Options” on page 7)
• On top of Avalon FM40/FM50 fetal monitors using mounting brackets
• On top of carts or desks

Avalon FM Cart Avalon FM


with Avalon CL Cart with
mounting kit Avalon CL
mounting kit

Installation Considerations
For most countries the Avalon CL is delivered with a default
configuration and an automatic frequency allocation giving a potential
operating range of up to 100 m/300 ft. For availability, contact your
local Philips Medical sales representative. The method of installation
may vary according to country-specific requirements. Contact Philips
Support for assistance.

7
Philips Healthcare is part of
0366
Royal Philips
The 866074 Avalon CL Fetal Transducer
How to reach us System complies with the requirements of
www.philips.com/healthcare the Council Directives 93/42/EEC (Medical
healthcare@philips.com Device Directive) and 99/5/EC (Radio and
fax: +31 40 27 64 887 Telecommunications Terminal Equipment
Directive).
Asia
+852 2821 5888

Europe, Middle East, Africa


+49 7031 463 2254

Latin America
+55 11 2125 0744

North America
+1 425 487 7000
800 285 5585 (toll free, US only)

Please visit www.philips.com/

© 2014 Koninklijke Philips N.V.


All rights are reserved.

Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or
obligation and will not be liable for any consequences resulting from the use of this publication.

Printed in The Netherlands.


4522 962 94361 * Jan 2014