Beruflich Dokumente
Kultur Dokumente
Application
Mark Stahl
Abstract
In terms of quality control and for all quantitative analysis peak purity
is a major task, which can be addressed in different ways. Often only
peak shapes and chromatograms are taken into consideration but a
very elegant possibility without the need to use a mass spectrometric
detector is the comparison of spectra recorded with a diode-array
detector during the registration of a chromatographic peak. Due to the
differences of these spectra within a single chromatographic peak, its
purity can be decided easily. The Agilent ChemStation software offers
an easy to use tool to perform these tasks routinely. In this note, the
theoretical background of peak purity determination is presented as
well as the practical use of the ChemStation software for these purposes.
Introduction ty detector, peak purity can only be In the elution dimension
judged from the peak profile of this Signals
An essential requisite of a separa- signal. Peak profile however is A valuable tool in peak purity
tion analysis is the ability to verify influenced by a variety of parame- analysis is the overlay of separa-
the purity of the separated species, ters and depends heavily on chro- tion signals at different wave-
that is, to ensure that no coeluting matographic or electrophoretic res- lengths to discover dissimilarities
or comigrating impurity contri- olution. Therefore peak purity of peak profiles. The availability of
butes to the peak’s response. The determination based on the peak spectral data in the three-dimen-
confirmation of peak purity should profile of a single signal is a very sional matrix generated by the
be performed before quantitative unreliable and insensitive method. diode-array detector enables sig-
information from a chromato- This is especially true in CE, where nals at any desired wavelength to
graphic or electrophoretic peak is mismatched sample and buffer be selected and reconstructed for
used for further calculations. zones always result in peak skew- peak purity determination after
Neglecting peak purity confirma- ing2. As a consequence, a second the analysis. A set of signals ca be
tion means, in quality control, an approach involving multiple wave- interpreted by the observer best
impurity hidden under a peak length detectors acquiring more when, before being displayed, it is
could falsify the results or, in than one signal in parallel has been normalized to maximum
research analysis, important infor- adopted. Such detectors enable absorbance or to equal areas. A
mation might be lost or scientific impurities to be uncovered by good overlap, where peak shape
observations rendered void should methods that involve overlaying sig- and retention or migration time
an impurity remain undiscovered. nals to compare peak profiles and match, indicate a pure peak while
Validated analytical methods usu- calculate signal ratios. These detec- a poor overlap indicates an
ally include the peak purity check tors have some disadvantages impure peak, as demonstrated in
as a major item in the list of their which will be discussed in the fol- figure 1.
method validation criteria (table 1). lowing section. They can be elimi-
nated easily by a third approach In addition to overlaying signals,
Method validation criteria based on diode-array technology1: their ratios ca be calculated and
• on-line acquisition of UV-Visible plotted. The resulting ratiograms
Selectivity (peak purity determination) ✔
Linearity ✔ spectra during the peak’s elution, are sensitive indicators of peak
Limits of detection and quantification ✔ • several signals at different wave- purity (figure 2). Any significant
Quality of data (accuracy and precision) ✔ lengths in parallel, distortion of the ideal rectangular
Ruggedness ✔
• signal extracts from a 3-D data form of the ratiogram indicates an
Table 1 matrix, containing spectral data impurity. However, there are some
Peak purity determination – a major criterion and separation signals, for peak limitations to be considered when
in method validation
purity analysis.. using signals for peak purity deter-
mination:
Techniques for peak purity Peak purity determination can be • The UV-Visible spectra of both
determination performed in different levels, tai- the main compound and the
lored to the complexity of the sepa- impurity must be well known in
Several techniques are currently ration and the user’s needs. order to select the most suitable
used for peak purity determination Detailed descriptions of the various wavelengths for the peak profile
in high performance liquid chro- peak purity routines for signals and comparison. This fact reduces
matography (HPLC) and in capil- spectra, such as normalization and the applicability of this type of
lary electrophoresis (CE)1. With a overlay, the mathematical opera- impurity detection to a routine
conventional single wavelength tions and the different display like search of known impurities
detector (or a monitor providing modes will be given in the follow- in known main peaks.
just one single output signal) such ing sections.
as a refractive index- or conductivi-
2
• The technique is not directly tinction between spectra and peak There are no hard and fast rules
applicable to research work or profile, higher resolution between as to which concentrations of
method development. The risk the main compound and the impu- impurities can be detected and
of overlooking an unknown rity, and with increasing absolute which not. In general, less than 0.5
impurity remains even when and relative concentration of the - 1 % can be determined if the
several wavelengths are selected impurity. spectra are distinct enough. If the
in parallel. spectra resemble each other very
3
closely, and the column or capil-
lary system does not resolve the
impurity and main component
well, then only 5 % impurity may
be feasible.
4
are linearly interpolated and
subtracted from the peak spec-
trum. The first integrated peak
start or end to the left of the peak 3
selected spectrum that is on the
baseline is taken as the time for Baseline Nearest
Peak segment baseline
left reference spectrum. For the spectrum
peak 5
right reference time the first inte-
1 b1 to b2 b1
grated peak start or end on the peak 4
2+3 b3 to b4 b3
peak 1
baseline to the right is taken (fig- peak 2 4 b3 to b4 b3
peak 6
ure 4 and 5). If a peak is not com- 5 b3 to b4 b4
pletely resolved from its neigh- 6 b5 to b6 b5
bour, automatic selection of refer-
ence spectra might lead to a refer- b1 b2 b3 b4 b 5 b6
ence selected from the valley
between the two peaks. Though
we already know that an unre-
solved peak cannot be pure, we Figure 5
Spectral Options, Reference Spectrum window of the ChemStation
may want to use the purity test to
look for other hidden components.
In that case, the manual reference
can be used to select a reference
spectrum from before and after
the group of peaks.
5
1. Maximum absorbance of the
spectrum (or a particular wave- (a) maximumóminimum (b) match normalization (c) area normalization
normalization
length range of the spectrum)
2. Area of the spectrum (or a maximum
wavelength range of the spec-
trum)
area
3. Best possible match of the
minimum
entire spectrum. This type of maximum
normalization is recommended difference spectrum
to display spectra for peak puri- area
ty evaluation because it tries to
minimum
make the difference of both
spectra as small as possible by
Both spectra have same Area of difference Both spectra have same
shifting and rescaling the spec- absorbance range spectrum as small area
tra. The ChemStation normal- as possible
izes spectra automatically using
the best possible match of the
entire spectrum.
6
in most practical cases this corre-
3 peak spectra
sponds to the three to five spectra apex
that will have been recorded. If 2 2
- width + 3 width
all spectra were acquired during the 3
7
parison of the spectra recorded
across the peak indicates the pres-
ence of an impurity. However, no
conclusion can be drawn concern-
ing the kind, number and level of
impurity. An additional aid in
interpreting spectral dissimilari-
ties is the display of difference
spectra, generated by subtracting
these normalized spectra from the
other peak spectra selected. The
profiles of the difference enable a
further conclusion to be drawn:
randomly distributed residual pro-
files result from spectral noise
which may be caused by the
instrument (figure 10, upper sec-
tion), whereas systematic trends
Figure 10
will be observed if real spectral Normalized spectra and randomly distributed residual spectra resulting from spectral noise
differences caused by a spectral
impurity occur (figure 11, upper
section).
8
At the extremes, a similarity factor
of 0 indicates no match and 1000
indicates indentical spectra. Gen-
erally, values very close to the 16 mAU
ideal similarity factor (greater
than 995) indicate that the spectra 260 nm
Absorbance
are very similar, values lower than spectrum 2 Similarity 999.968
990 but higher than 900 indicate Slope 1.06818
50
some similarity and underlying Intercept 0.04693
data should be observed more 40
carefully. Figure 13 shows exam- 30
ples of similarity factors for simi- 20
lar, different, noisy and spectra 10
with impurity. The slope of the 0
Absorbance
0 10 20 30 40 50
regression lines represents the spectrum 1
ratio of the concentration of the
two spectra.
18 mAU
260 nm
Figure 12
Similarity of absorbance at the individual wavelength plotted for a pair of spectra gives the
similarity factor
(a) very similar spectra (b) different spectra (c) spectra with impurity (d) spectra with noise
Spectral difference 0.6% Spectral difference 55% Spectral difference 4.8% Spectral difference 0.5%
Figure 13
Graphical display of similarity factor for different pairs of normalized spectra
9
3. Improving sensitivity and reli-
ability
The similarity curve and threshold
curve functions improve the sensi-
tivity and reliability of the peak
purity evaluation by using all the
spectra acquired during the elu-
tion or migration of a peak rather
than just three or four.
• Similarity curve
The mathematical fundamentals
used in the similarity curve cal-
culations are those used for the
purity factor, however, they are
displayed in another format. All
spectra from a peak are com- Figure 14
pared with one or more spectra Spectral Options, Advanced Peak Purity Options of the ChemStation. Here the spectrum
chosen for the calculation of the peak purity level can be selected and the display of the
selected by the operator (figure similarity curve can be customized.
14), an apex or an average spec-
trum for example. The degree of
match or spectral similarity is
plotted over time during elution. (a) peak without impurity (b) peak without impurity
An ideal profile of a pure peak is and noise but with noise
a flat line at 1000 as demonstrat-
ed in figure 15(a). At the begin-
ning and end of each peak,
where the signal-to-noise ratio
decreases, the constribution of
spectral background noise to the
peak’s spectra becomes impor-
tant. The contribution of noise 980 980
to the similarity curve as shown
similarity similarity
in figure 15(b). curve . curve
• Threshold curve
The influence of noise on weak- 1000 1000
ly absorbing spectra can be seen
in figure 16. The similarity factor
decreases with decreasing sig-
nal-to-noise ratio or constant
noise level with decreasing Peak spectrum, 20 mAU Noise, 0.1 mAU
absorbance range. A threshold
curve shows the effect of noise Figure 15
on a given similarity curve. The Similarity curves for a pure peak with and without noise, plotted in relation to the ideal
similarity factor (1000) and the user - defined threshold (980).
effect increases rapidly towards
both ends of the peak. In
essence, a threshold curve is a
10
similarity curve with back-
ground noise contribution. Compound Spectrum Minimal deviation
Figure 17(a) shows both the sim- caused by noise
Similarity curve
ilarity curve and the threshold without noise
Maximal deviation
curve for a pure peak with noise, caused by noise
figure 17(b) for an impure peak.
The determination of noise Similarity within
the noise
threshold is performed automat- 3 level
ically based on the standard
Similarity factor
Dissimilarity outside
deviation of pure noise spectra the noise level
at a specified time with a user 2
selectable number of spectra,
usually at 0 minutes with 14
spectra. In subsequent analyses 1
you can define the noise thresh-
1 2 3
old as a fixed value based on Spectra with
your experience. different S/N ratios Signal-to-Noise Ratio
The threshold curve, represent-
ed by the broken line, gives the Figure 16
range for which spectral impuri- Similarity factor as function of the noise level
ty lies within the noise limit.
Above this threshold, spectral
background noise and the simi-
larity curve intersects the
(a) peak without impurity but with noise (b) peak with impurity and noise
threshold curve indicating an
impurity providing the reference
and noise parameters have been
sensible chosen (figures 17 and 5% impurity
21).
980 980 similarity
threshold curve
The software offers four modes to curve
display the similarity and thresh- similarity threshold
old curves: curve curve
1000 1000
• Similarity & Threshold (with
out any transformation): The
similarity and threshold curves
are displayed as they are calcu- Noise, 0.1 mAU Impurity spectrum, 1 mAU
lated using values between 1
and 1000 (figure 18a).
•. Similarity & Threshold as the Figure 17
natural logarithm: Similarity Effect of impurity and noise on similarity and threshold curves
and threshold curves are dis-
played as the natural logarithm
of their calculated value. This
can give additional details in the
higher match factors (figure
18b).
11
•. Similarity / Threshold ratio:
The ratio of the similarity and
threshold curves is displayed as
a single curve.
ratio = 1000 – similarity
1000 – threshold
. The results for each spectrum
are compared to the expected
result based on the threshold
calculation. If the ratio is less
than 1 the test for that spectrum
passes, if it is greater than 1
then it fails (figure 18(c)).
•. Purity ratio: The purity value
of each single spectrum is dis-
played as the logarithm of the
difference from the threshold Figure 18
Threshold and similarity curves, (a) as calculated, (b) ln (threshold) and ln (similarity), (c) as a ratio
value. The chromatographic
and (d) as purity ratio
peak, similarity and threshold
curves are displayed in the
upper part of the display (figure
18(d) and 21). For a spectral
pure peak the ratio values are
below unity (the purity ratio is
in the green band) and for spec-
tral impure peaks the values are
above unity (the purity ratio is
in the red band). The advantage
of these modes are that only
one line is displayed, instead of
two which makes the interpre-
tation more simple. All the Figure 19
exact values for each single The peak purity information window of the ChemStation gives detailed information of the peak spec-
spectrum are not only graphi- tra recorded
cally displayed but can also be
reviewed in the Peak Purity Compare each individual spectrum where the impurity is, or needs to
information window (figure 19) ➝ to all others
improve the sensitivity of purity
of the ChemStation. ➝ to the apex spectrum evaluation. An example may help
➝ to the average spectrum (of all peak spectra) to show how this principle can be
4. Using specific target ➝ to the front spectrum applied. If the impurity is assumed
➝ to the tail spectrum
spectra ➝ to the front and the tail spectra to be located on the peak, select-
The ChemStation permits calcula- ing the tail or apex spectrum to be
tions of the purity factor and simi- Table 2 compared with all other spectra
Different reference spectra for
larity curves relative to different will provide the most significant
spectra comparison
target spectra, as shown in figure information. Figure 20 gives the
20 and table 2. As a general rule, impurities. The flexibility of being ratio curve for the front, apex, tail
the option Compare the average able to select a specific target and average spectrum of a peak
spectrum provides the most valu- spectra is valuable in those cases which contains an impurity after
able information for unknown where the analyst has to assume the response maximum (apex).
12
The front spectrum gives a small
spectral impurity at the end of the
peak. The deviation in this first
ratio curve is small since the front
spectrum absorbed little (giving a
rather high threshold curve). The Peak with
apex spectrum gives a low impuri- 2% impurity
ty in the front of the peak (the
apex spectrum contains only a
very small amount of the impuri-
ty) and high impurity at the tail.
The tail spectrum (with a high
amount of impurity) gives a spec-
tral impurity at the front of the
1
peak. The average spectrum (a
0
mean of the peak spectra selected,
in this case the upslope-, apex-, ratio of ratio of ratio of ratio of
and downslope spectra) indicates front spectrum apex spectrum tail spectrum average spectrum
spectral impurity in the total peak. to all other spectra to all other spectra to all other spectra to all other spectra
13
spectral data recorded with all
spectra set during acquisition. The
extraction is optimized to find as
much difference as possible in the
curvature of the signals. This gives
an extra conformation for peak
impurity and in many cases an
indication of the location of the
impurity (in some cases even
when more than one impurity is
present).
Examples
14
threshold curve limits and, the
peak purity ratio is clear within
the green band.
Impure peak
Figure 23 shows the peak purity
determination of an impure peak
containing an impurity with a
quite similar spectrum to that of
the main compound. The overlay
of the normalized spectra and the
extracted signals indicate the
presence of an impurity. The simi-
larity curve exceeds the threshold
curve between 9.5 and 9.6 min and
the peak purity ratio is clearly in
the red band thus leading to the
warning message. From these sig-
nals we can conclude that the
impurity is at the end of the peak. Figure 23
Figures 24 illustrates the Chem- Peak purity analysis of a peak containing an impurity.
Station’s capabilities to discover
very small impurities of 0.5 % and
less with almost indentical spectra
to the main analyte. Often the
spectra overlay, the residual spec-
tra and even the ratiogram do not
provide sensitive enough informa-
tion to uncover the impurity, how-
ever the similarity and threshold
curve and the peak purity ratio are
more sensitive and are able to
reveal the hidden impurity. 5%
15
step, unattended. Printed reports References
contain all the graphical informa-
tion from the screen plus 1.
additional numerical results. D. N. Heiger, “High performance
Despite the ChemStation’s capacity capillary electrophoresis,”
to run sample analyses automati- Agilent Primer 2000 publication
cally and unattended, we recom- number 5968-9936E.
mend that you examine the results
yourself first visually before 2.
passing any purity judgement. H.-J. P. Sievert and A. C. J. H.
Incorrect results most often arise Drouen, “Spectral matching and
when a column or capillary sys- peak purity” in Diode-Array
tem fails to resolve peak or the Detection in High-Performance
ChemStation integrates some Liquid Chromatography, 1993,
baseline event erroneously leading 51 – 125, Marcel Dekker, New York.
to a false reference spectrum
being selected.
Conclusion