CELL-DYN 3000/3500 ONLY (For CELL-DYN 3700 see previous page) 1. Make sure the Analyzer is ON./ 2. Make sure Data Station power is OFF. 3. Insert Service Disc (CELL-DYN 3000 = 93168-01 Rev E) (CELL-DYN 3500 = 9381023 Rev A or higher). 4. Turn ON Data Station. 5. At the A> prompt, type "hsslcom", then ↵ENTER. 6. Press _ F1 key to download a program. 7. Type For CELL-DYN 3000 "C:\cd3000\cd3000". For CELL-DYN 3500 Rev F or lower "C:\cd3000\cd3000A" For CELL-DYN 3500 Rev G or higher "C:\cd3000\cd3500". ↵ENTER. 8. When downloading is complete, choose appropriate designation: "sample loader" or "closed system". 9. Download "sample loader" a. Type: "OKS (ZERO-UPPER CASE KS), then ↵ENTER. b. To verify programming type: OK (ZERO-UPPER CASE K), then ↵ENTER. c. The instrument will display "SKS". 10.Download "closed system" a. Type: "OKC (ZERO-UPPER CASE KC), then ↵ENTER. b. To verify programming type: OK (ZERO-UPPER CASE K), then ↵ENTER. c. The instrument will display "SKC", and a right facing dashed arrow symbol will appear. 11. Remove the Service Disc.
12.Turn Analyzer and Data Station OFF.
13.To return to normal operation, reboot according to normal startup procedures