gl0624-016-042617-ACC-USL
Accredited by the American Association
Mettler Toledo for Laboratory Accreditation (A2LA)
Service Business Unit Industrial
CALIBRATION CERT #1788.01
1900 Polaris Parkway
ISO 17025 Accredited
Columbus, OH 43240
ANSI/NCSL Z540-1 Accredited
1-800-523-5123
Weighing Device
Procedure
This calibration certificate contains measurements for As Found and As Left calibrations.
This certificate is issued in accordance with the conditions of accreditation granted by A2LA, which is based on ISO/IEC 17025. A2LA has
assessed the measurement capability of the laboratory and its traceability to recognized national standards.
Measurement Results
Repeatability
Test Load: 700 g
1 700.0 g 700.0 g
◆ As Left
2 700.1 g 700.0 g
3 700.0 g 700.0 g
4 700.0 g 700.0 g
5 700.0 g 700.0 g
6 700.0 g 700.0 g
Standard
Deviation 0.04 g 0.00 g
The "d" in the graph represents the readability of the range/interval in which the
test was performed.
The results of this graph are based upon the absolute values of the differences
from the mean value.
Eccentricity
Test Load: 500 g
1 500.0 g 500.0 g
2 500.0 g 500.0 g
3 500.1 g 500.0 g
4 500.0 g 500.0 g
5 500.0 g 500.0 g
Maximum
0.1 g 0.0 g
Deviation As Found As Left
The "d" in the graph represents the readability of the range/interval in which the
test was performed.
Error of Indication
As Found
As Left
As Found
◆ As Left
The uncertainty stated is the expanded uncertainty at calibration obtained by multiplying the standard combined uncertainty by the
coverage factor k – which can be larger than 2 according to EURAMET cg-18. The value of the measurand lies within the assigned range
of values with a probability of approximately 95%.
The user is responsible for maintaining environmental conditions and the settings of the weighing instrument when it was calibrated.
Test Equipment
All weights used for metrological testing are traceable to national or international standards. The weights were calibrated and certified by an
accredited calibration laboratory.
Remarks
Air drafts interfere with the balance
Balance integrated in other equipment
The user should re-verify the performance of the equipment on site with appropriate standards prior to use.
The information below and any attachments to this calibration certificate are not part of the accredited calibration.
Temperature coefficient for the evaluation of the measurement uncertainty in use: 10.0 · 10⁻⁶ / K
Temperature range on site for the evaluation of the measurement uncertainty in use: 10 K
As Found As Left
As Found As Left
Passed
Overall Failed
Weighing Device
Range Max. Capacity Readability (d) Verification Scale Interval (e) Class
1 1510 g 0.1 g 0.1 g II
Test Load
Tolerance
From To
0.0 g 500.0 g 0.1 g
500.1 g 1510.0 g 0.2 g
As Found
◆ As Left
Tolerance
As Found As Left
Test Test Load Tolerance
Maximum Error Result Maximum Error Result
Eccentricity 500 g 0.1 g 0.1 g 0.0 g
Repeatability 700 g 0.2 g 0.1 g 0.0 g
Error of Indication
As Found As Left
Reference Value Tolerance
Error of Indication Result Error of Indication Result
1 0g 0.1 g 0.0 g 0.0 g
2 300.0 g 0.1 g 0.0 g 0.0 g
3 700.0 g 0.2 g 0.1 g 0.0 g
4 1100.0 g 0.2 g 0.0 g 0.0 g
5 1500.0 g 0.2 g 0.2 g 0.0 g