Calibration Certificate ID

gl0624-016-042617-ACC-USL
Accredited by the American Association
Mettler Toledo for Laboratory Accreditation (A2LA)
Service Business Unit Industrial
CALIBRATION CERT #1788.01
1900 Polaris Parkway
ISO 17025 Accredited
Columbus, OH 43240
ANSI/NCSL Z540-1 Accredited
1-800-523-5123

Accuracy Calibration Certificate


Customer

Company: Lucerne Foods Ltd


Address: 5115 57 St; Jalan Pekeling Tanjung 27
City: pipi Contact: N/A
Zip / Postal: T1G 1X1
State / Province: N/A

Weighing Device

Manufacturer: Mettler Toledo Instrument Type: Weighing Instrument


Model: PB1501-S Asset Number: LABY-7072
Serial No.: 1120263588 Terminal Model: N/A
Location: 1; 2; 3 Terminal Serial No.: N/A
Terminal Asset No.: N/A

Range Max. Capacity Readability (d)


1 1510 g 0.1 g

Procedure

Calibration Guideline: N/A


METTLER TOLEDO Work Instruction: N/A

This calibration certificate contains measurements for As Found and As Left calibrations.

Temperature Pressure conditions have been verified to ensure


Environmental Pressure
the accuracy of the
As Found Start: 20.0 °C End: 20.0 °C calibration.

As Left Start: 20.0 °C End: 20.0 °C

This certificate is issued in accordance with the conditions of accreditation granted by A2LA, which is based on ISO/IEC 17025. A2LA has
assessed the measurement capability of the laboratory and its traceability to recognized national standards.

As Found Calibration Date: 26-04-2017 Authorized A2LA


As Left Calibration Date: 26-04-2017 Signatory:
Chris Housden
Next Calibration Date: N/A

Software Version: 1.13.0.102 © METTLER TOLEDO Page 1 of 5


Report Version: 1 This is an original document and may not be partially reproduced without the
Form Number: written permission of the issuing calibration laboratory.
Calibration Certificate ID
gl0624-016-042617-ACC-USL

Measurement Results

Repeatability
Test Load: 700 g

As Found As Left As Found

1 700.0 g 700.0 g
◆ As Left
2 700.1 g 700.0 g
3 700.0 g 700.0 g
4 700.0 g 700.0 g
5 700.0 g 700.0 g
6 700.0 g 700.0 g

Standard
Deviation 0.04 g 0.00 g

The "d" in the graph represents the readability of the range/interval in which the
test was performed.

The results of this graph are based upon the absolute values of the differences
from the mean value.

Eccentricity
Test Load: 500 g

Position As Found As Left

1 500.0 g 500.0 g
2 500.0 g 500.0 g
3 500.1 g 500.0 g
4 500.0 g 500.0 g
5 500.0 g 500.0 g

Maximum
0.1 g 0.0 g
Deviation As Found As Left
The "d" in the graph represents the readability of the range/interval in which the
test was performed.

Error of Indication
As Found

Reference Value Indication Error of Indication Expanded Uncertainty k

1 0g 0.0 g 0.0 g 0.11 g 2.28


2 300.0 g 300.0 g 0.0 g 0.13 g 2.13
3 700.0 g 700.1 g 0.1 g 0.15 g 2.13
4 1100.0 g 1100.0 g 0.0 g 0.18 g 2.05
5 1500.0 g 1500.2 g 0.2 g 0.21 g 2.05

Software Version: 1.13.0.102 © METTLER TOLEDO Page 2 of 5


Report Version: 1 This is an original document and may not be partially reproduced without the
Form Number: written permission of the issuing calibration laboratory.
Calibration Certificate ID
gl0624-016-042617-ACC-USL

As Left

Reference Value Indication Error of Indication Expanded Uncertainty k

1 0g 0.0 g 0.0 g 0.06 g 2


2 300.0 g 300.0 g 0.0 g 0.08 g 2
3 700.0 g 700.0 g 0.0 g 0.08 g 2
4 1100.0 g 1100.0 g 0.0 g 0.08 g 2
5 1500.0 g 1500.0 g 0.0 g 0.08 g 2

As Found

◆ As Left

The uncertainty stated is the expanded uncertainty at calibration obtained by multiplying the standard combined uncertainty by the
coverage factor k – which can be larger than 2 according to EURAMET cg-18. The value of the measurand lies within the assigned range
of values with a probability of approximately 95%.

The user is responsible for maintaining environmental conditions and the settings of the weighing instrument when it was calibrated.

Software Version: 1.13.0.102 © METTLER TOLEDO Page 3 of 5


Report Version: 1 This is an original document and may not be partially reproduced without the
Form Number: written permission of the issuing calibration laboratory.
Calibration Certificate ID
gl0624-016-042617-ACC-USL

Test Equipment

All weights used for metrological testing are traceable to national or international standards. The weights were calibrated and certified by an
accredited calibration laboratory.

Weight Set 1: OIML F2

Weight Set No.: F2Test Date of Issue: 08-02-2016


Certificate Number: F2 Calibration Due Date: 11-08-2017

Weight Set 2: OIML E1

Weight Set No.: GL0101-Custom1 Date of Issue: 15-09-2016


Certificate Number: asfssss Calibration Due Date: 15-09-2017

Weight Set 3: OIML M2

Weight Set No.: M2 Date of Issue: 08-11-2016


Certificate Number: M2 Calibration Due Date: 08-11-2017

Weight Set 4: OIML M3

Weight Set No.: M3 Date of Issue: 08-11-2016


Certificate Number: M3 Calibration Due Date: 08-11-2017

Weight Set 5: OIML E2

Weight Set No.: Metric E2 Date of Issue: 27-04-2017


Certificate Number: Good Cert num Calibration Due Date: 27-04-2018

Remarks
Air drafts interfere with the balance
Balance integrated in other equipment

The user should re-verify the performance of the equipment on site with appropriate standards prior to use.

End of Accredited Section

The information below and any attachments to this calibration certificate are not part of the accredited calibration.

Software Version: 1.13.0.102 © METTLER TOLEDO Page 4 of 5


Report Version: 1 This is an original document and may not be partially reproduced without the
Form Number: written permission of the issuing calibration laboratory.
Calibration Certificate ID
gl0624-016-042617-ACC-USL

Measurement Uncertainty of the Weighing Instrument in Use


Stated is the expanded uncertainty in use with k = 2. The formula shall be used for the estimation of the uncertainty under consideration
of the errors of indication. The value R represents the net load indication in the unit of measure of the device.

Temperature coefficient for the evaluation of the measurement uncertainty in use: 10.0 · 10⁻⁶ / K

Temperature range on site for the evaluation of the measurement uncertainty in use: 10 K

Linearization of Uncertainty Equation


Range As Found As Left

1 0 g - 1510 g U₁ = 115 mg + 0.516 mg/g · R U₁ = 82 mg + 0.0353 mg/g · R

Examples for Various Net Indications


Net Indication As Found As Left

1.5 g 0.12 g 7.9% 0.082 g 5.3%


15.1 g 0.12 g 0.79% 0.083 g 0.53%
151.0 g 0.19 g 0.13% 0.087 g 0.060%
755.0 g 0.50 g 0.066% 0.11 g 0.015%
1510.0 g 0.89 g 0.059% 0.14 g 0.0093%

As Found As Left

Software Version: 1.13.0.102 © METTLER TOLEDO Page 5 of 5


Report Version: 1 This is an original document and may not be partially reproduced without the
Form Number: written permission of the issuing calibration laboratory.
Attachment to Calibration Certificate:
gl0624-016-042617-ACC-USL
Handbook 44 Tolerance Assessment

Handbook 44 Tolerance Assessment (Maintenance)


The measurements from the attached calibration certificate were assessed against the Tolerances defined by NIST Handbook 44.
The range of the measurements for both Eccentricity and Repeatability (if performed) tests is assessed against Maintenance Tolerances.

As Found As Left
Passed
Overall Failed

Weighing Device
Range Max. Capacity Readability (d) Verification Scale Interval (e) Class
1 1510 g 0.1 g 0.1 g II

Tolerances according to NIST Handbook 44

Test Load
Tolerance
From To
0.0 g 500.0 g 0.1 g
500.1 g 1510.0 g 0.2 g

As Found

◆ As Left
Tolerance

Eccentricity and Repeatability

As Found As Left
Test Test Load Tolerance
Maximum Error Result Maximum Error Result
Eccentricity 500 g 0.1 g 0.1 g 0.0 g
Repeatability 700 g 0.2 g 0.1 g 0.0 g

Error of Indication
As Found As Left
Reference Value Tolerance
Error of Indication Result Error of Indication Result
1 0g 0.1 g 0.0 g 0.0 g
2 300.0 g 0.1 g 0.0 g 0.0 g
3 700.0 g 0.2 g 0.1 g 0.0 g
4 1100.0 g 0.2 g 0.0 g 0.0 g
5 1500.0 g 0.2 g 0.2 g 0.0 g

Software Version: 1.13.0.102 © METTLER TOLEDO Page 1 of 1


Report Version: 1 This is an original document and may not be partially reproduced without the
Form Number: written permission of the issuing calibration laboratory.