CALIBRATION STEPS
SETTING UP F90
STEP 17: In the EctAcqr93T, insert probe into cal. standard to pull probe past the graduals. Have probe pass
from largest gradual to smallest gradual. This will give a better signal for working with the Absolutes. Tip of
signal in F1-F4 ABS (XY) should reach bottom right hand corner on a 10 volt span.
STEP 18: Adjust all ABS so that tip of signal point toward bottom right hand corner. This is all done in the
rotation window (Ctrl-R), pulling probe back and forth thru the graduals.
STEP 19: Adjust the gain (Ctrl-G) for each ABS. In some cases you may also need to adjust the Driver Level at
this time (Ctrl-D), so tip of signal reaches the bottom right hand corner of XY plot. Watch out for saturation
with Driver Level and Gain adjustments. (Note: some higher Freq’s may not actually obtain a full response)
STEP 20: Balance probe in good portion of tube. (Ctrl-Z)
STEP 21: REPEAT STEPS 17-20, until tip of signal reaches bottom right hand corner.
TD-EC-006 (rev 11/22/2010)
SETTING UP THRU-HOLE
STEP 22: Hold calibration standard with hole up and away from you. Record as you pull probe past hole.
STEP 23: Hit F10 to view waveform for hole. Now look at voltage spans (lower left hand corner in XY Plot)
You want your F1 Diff to be in the range of .600 - .750 volts
STEP 24: Hit F10 to go back to EctAcqr93T to adjust the gain (Ctrl-G) for F1 Diff. At this time adjust F2, F3 &
F4 to match F1 voltage span.
STEP 25: After changing gain, balance (Ctrl-Z) probe in good portion of tube.
STEP 26: Repeat Steps 22-25, until you have matched the voltage spans and are in the range of .600 - .750
STEP 27: Hold calibration standard with hole up and away from you. Record as you pull probe past hole.
STEP 28: Look at the phase angle for F1, F2, F3 & F4 Diffs. You want to match F2, F3 & F4 to the same angle
as F1.
Ex: F1 @ 31.00, F2 @ 27.00, F3 @ 36.00 & F4 @ 31.00
STEP 29: Hit F10 to go back to EctAcqr93T to rotate the phase angles (Ctrl-R) for F2, F3 & F4.
Ex: F2 would rotate down by 4, F3 would rotate up by 5 and F4 would not need to be rotated.
STEP 30: Repeat Steps 27-29 until angles match.
TD-EC-006 (rev 11/22/2010)
STEP 31: Position probe in good portion of tube and baffle in same location on OD of tube.
STEP 32: In the EctAcqr93T window select Mix Setup and check mix cancellation.
STEP 33: Hold probe stationary. Hit space bar to record. Move baffle ring over probe coils from cable side to
tip. Hit space to stop recording.
STEP 34: Mix window will open. Select channels to be mixed. F90 over ½ F90 for best results. Click OK.
ALTERNATE STEP 31: Insert probe into one of the center tubes to be inspected until baffle signal is seen on F1
Diff strip.
ALTERNATE STEP 32: In the EctAcqr93T window select Mix Setup and check mix cancellation.
ALTERNATE STEP 33: Hit space bar to record as probe is pulled past baffle. Hit space bar to stop recording.
ALTERNATE STEP 34: Mix window will open. Select channels to be mixed. F90 over ½ F90 for best results.
Click OK.
STEP 35: Hold calibration standard with hole up and away from you. Record as you pull probe past hole.
STEP 36: Hit F10 to view waveform for hole. Now look at voltage span for MIX (XY Plot)
STEP 37: Hit F10 to go back to EctAcqr93T to adjust the gain select Mix Set Up (Mix Gain) for Mix 1
STEP 38: After changing gain, balance (Ctrl-Z) probe in good portion of tube.
STEP 39: Repeat Steps 35-38, until you have matched the voltage span to F1-F4 Diffs
STEP 40: Hold calibration standard with hole up and away from you. Record as you pull probe past hole.
STEP 41: Look at the phase angle for the Mix. Match the Mix angle to F1-F4 Diffs.
Ex: F1 @ 31.00, F2 @ 31.00, F3 @ 31.00, F4 @ 31.00 & Mix @ 34.00
STEP 42: Hit F10 to go back to EctAcqr93T to rotate the phase angles (Ctrl-R) for the Mix
Ex: The Mix would need to be rotated up by 3.
STEP 43: Repeat Steps 40-42 until angles match.
TD-EC-006 (rev 11/22/2010)