Thank you for purchasing MRC Microbial air sampler, this manual will
guide you through the proper use of this equipment, please keep it for
future reference.
Directory
x. Accessories --------------------------------------------9
1
I. Equipment introduction
Thank you for purchasing and using AIS-SERIES type microbial air sampler,
it represents the latest air floating microbial detection concept.
AIS-SERIES type microbial air sampler is a high effective multi-nozzle inhaling
style plankton sampler. It is based on isokinetic sampling design,
sampling is accurate. Sampling head wind speed is consistent with that of the
clean room, so it more accurately reflects the concentration of
microorganisms in the clean room. When collecting samples, the air
carrying particles with a certain impact speed into the surface of the medium,
the principle of this process is equivalent to level five of Anderson theory, to
ensure all particles larger than 1 micron are collected. Particles hit the surface
of the medium uniformly in the Petri dish, and the medium will maintain the
activity of the microbial by creating the best environment, which ensures
accurate and reliable measurement results. After automatic sampling is
completed, place the Petri dish in the incubator and culture, living
microorganisms in appropriate circumstances will obtain nutrients from the
medium and grow in high speed, finally the result comes out.
II. Application areas
1. Pharmaceutical industry:
Monitor microorganism number of laminar flow hood
Clean area cleanliness trend analysis
Evaluate the effectiveness of the sterilization
2. Food and beverage industry:
Detect microbial contamination in clean area, it is required in HACCP.
3. Hospitals
Detect air pathogenic microorganisms in the operating room, pharmacy,
disposal room and special ward room
4. Electronics industry
Measure the particle and microbiological contamination difference in the clean
room
V. Technical parameters
1. Sampling head impaction nozzle number: 300;
2. High-capacity rechargeable lithium polymer battery: 6400 mA
3. Sampling flow rate: 100L / min;
4. Sampling head port flow rate: 0.38m/s, basically same with clean room wind
speed (isokinetic sampling);
5. Automatic control sample volume: 10 ~ 6000L, minimum unit:10L;
6. Parameter settings and sampling records storage: 99 sampling parameters
setting record; 2000 sample data records.
7. Applicable Petri dish: Φ 70 ~ 90 × 15 mm;
8. Charging power: AC 110 ~ 220V
9. Dimensions: Φ 120 × 280 mm
10. Weight: 2 .5 kg
3
VII. Operation instructions
Main interface
Start panel)
Enter default setting (User not need to set Sampling location setting
Date setting Time setting Back light delay time Delete sampling setting
3. Quick settings:
Switch on the device, it enters the main interface. Press the "Menu Button" to enter
the parameter setting mode:
5
⑴ Sample volume setting: Press “Select” button to adjust the cursor position, press
the “up” and “down” button to adjust to the desired sample volume, press "Exit" to
exit the settings, press "Menu" to enter the next parameter setting.
⑵ Sampling delay time setting: Press “up” and “down” button to adjust to the desired
sampling delay time, press "Exit" to exit the settings, press the "Menu" to enter the
next parameter setting.
⑶ Sampling location setting: Press “up” and “down” button to set the sampling point
code, press the "Exit" button to exit the setting, press the "Menu" to enter the next
parameter setting.
⑷ Save sampling setting: Press “up” and “down” button to choose "Yes" and press
"Select" to save the current sampling parameters. Press "Exit" to return to the main
menu, setup is complete.
⑸ Main menu will show the latest settings of the sampling parameters.
8
IX. Troubleshooting
9
Factory test report
Temperature: Humidity: Series No.:
1. Appearance:
2. Function tests:
4. Parametric tests:
Test point
Item & Criteria a b c d e Average
Sampling
0.38m / s
head wind
Sampling
± 0.5S
time error
Measuring
Φ75 ± 0.1
tube size
5. Conclusion:
Inspector:
Reviewer:
Inspection date:
10