Beruflich Dokumente
Kultur Dokumente
— Osteodensitometer —
CHALLENGER_Envision _Anglais_03
User Manual
— Osteodensitometer —
CHALLENGER_Envision _Anglais_03
CONTENTS
CHALLENGER_Envision _Anglais_03 1
Chapter 1 Introduction and Background Information
Challenger is an Osteodensitometer, using the technology known as DEXA (Dual Energy X-ray Absorptiometry) which itself
is controlled a standard Pentium IV 1.4 GHz (minimum), 512 RAM computer running Windows XP. It performs
measurements at various sites of the spine, the femur and the forearm using an x-ray tube and provides information on
projected area, BMD, and BMC at each measurement site.
1.2 Contraindications
Any female patients who are suspected of being pregnant should not undergo any examination, without first consulting their
personnel doctor.
1.3 Warnings
WARNING: Do not attempt to operate the equipment without first reading this manual / and receive the training from an
approved representative of DMS.
WARNING: This machine outputs a level of X-ray radiation. The applicable local and international regulations must apply
to the installation and operation of the machine.
WARNING: If the patient is or might be pregnant, contact the patient’s physician before performing an X-ray examination
WARNING: Do not look directly into the laser light
WARNING: Do not remove any paneling to obtain access to the interior of the equipment.
WARNING: Do not use portable cellular equipment (walkie-talkies, radiophones, and portable telephones) in the
proximity during operation, as this may cause an impact on the measurements.
WARNING: If the image of the patient becomes distorted, this should be considered as the machine not working correctly
and should be reacted to (refer to 16 Trouble-shooting p.219)
1.4 Cautions
CAUTION: Do not lie or sit on the machine for purposes other than scanning.
CAUTION: Remain in the room whilst performing the examination
CAUTION: Restrict the room to authorized personnel only
CAUTION: Use only accessories and peripheral devices conforming to IEC Standards.
CAUTION: Ensure that the correct voltages are set for the equipment to be connected to the mains power supply.
The quantity of examination per patient should be kept to a minimum, or in line with current European / International
Standards. No known adverse effects are known provided the limits are abided.
The public (patients) minimum recommended dosage is <1mGy/year, which is equivalent to approximately 100 examinations
in one year.
An operator must not be subjected to an average dose of more than 20mGy per year (averaged over 5 years). With an upper
limit of 50mGy in a year.
The physician / operator shall be responsible for the cleaning of CHALLENGER once a week using non abrasive cleaning
agent in order to ensure the cleanliness of the device.
CHALLENGER_Envision _Anglais_03 3
Chapter 1 Introduction and Background Information
The target population is any subject (patient) requiring an examination, with the exception of pregnant women or those
suspected of being pregnant.
CHALLENGER_Envision _Anglais_03 4
Chapter 2Using this manual
This handbook contains the latest information available at the time of writing.
Changes may be periodically made to the contents of this manual. Any change will be incorporated in form of updates.
Updates to the Manual, issued between new editions, shall contain additional and/or replacement pages to be either inserted in
or attached to the original by the user.
New editions may be released when a complete update of the manual is required.
D.M.S. reserves the right to modify at any time the product and (or) the software described in this user’s guide.
2.2 Copyright
2.2.2 Software
No part of the software may be reproduced either in whole or in part nor translated into another media or language without
prior written permission by Diagnostic Medical Systems.
The information in this document is subject to change without prior notice and does not represent a commitment on the part of
D.M.S.
D.M.S. acknowledges the importance of customer feedback and encourages users to contact D.M.S. for any suggestions or
questions about either the equipment or manual.
In case any mistake or misprint should have slipped through in this manual, please return a photocopy of the page involved to
the address below:
D.M.S.
Parc de la Méditerranée
District de Montpellier
34470 Pérols
FRANCE
E-mail dms@dms.com
Additional copies
Additional copies of this manual or other DMS literature may be obtained from the same address given above.
CHALLENGER_Envision _Anglais_03 5
Chapter 3 Device Description
3 Device Description
3.1 Description of the device
Challenger is an Osteodensitometers, using the technology known as DEXA (Dual Energy X-ray Absorptiometry) which itself
is controlled a standard Pentium IV 1.4 GHz (minimum), 512 RAM computer running Windows XP.
The different examination sites are the Spine, the Femur and the Forearm. The source and the detector are moved in the X and
Y directions to obtain the analysis zones of 150x200mm and 150x150mm, and 80x80mm respectively.
The X-ray source emits 3 intensities, 0.1mA, 0.4mA and 0.8mA. The detector assures a count of X-rays not absorbed by the
soft tissues and thus provides the density of the measured site.
These parameters are given for each area of analysis selected (either manually or automatically) on the image obtained from
the examination.
CHALLENGER_Envision _Anglais_03 6
Chapter 3 Device Description
Shown below in figure, here are the photographs of the both CHALLENGER:
CHALLENGER_Envision _Anglais_03 7
Chapter 4 Warranty Procedure
4 Warranty Procedure
4.1 DMS system warranty
DMS guarantees that each product sold is free from defects (either labor or material) and shall conform to its product
specifications as defined in the user's documentation.
If the product fails to function as specified during the warranty period, it shall be repaired or replaced free of charge.
The warranty period for products sold outside France is 12 months as from the date of installation and the date of shipment,
whichever comes first.
Misuse, accidental damage, modifications, unsuitable operating environment, unauthorized maintenance, or damage caused by
a product for which DMS cannot be held responsible may invalidate the warranty.
We do not guarantee:
i) Components and/or equipment which are not supplied by DMS but which constitute an integral part of the
product cannot be guaranteed by DMS.
ii) Non-DMS manufacturers or suppliers may provide their own guarantee to you.
CHALLENGER_Envision _Anglais_03 8
Chapter 5 Cleaning
5 Cleaning
5.1 Cleaning Challenger
Challenger should be cleaned once a week with a soft cloth and with clean tap water in order to avoid a build-up of dust and
dirt.
NB: Cleaning of the Mattress & Cushion may be carried out using a non-abrasive cleaning agent to ensure its cleanliness.
In order to reduce the cleaning requirements on the mattress, it is worthy to consider the use of Disposable Paper Sheets,
which can be changed following each patient examination.
The phantoms should be cleaned once a week with a soft cloth and with clean tap water in order to avoid a build-up of dust
and dirt.
CHALLENGER_Envision _Anglais_03 9
Chapter 6 Maintenance Requirements
6 Maintenance Requirements
There are no user-serviceable parts inside. Contact your authorized service representative for any servicing required.
A Technical Manual is available on request to those personnel technically qualified and authorized by DMS to carry out
technical servicing.
The unit should be considered as a system during the activities of preventative maintenance, such that the unit should be
inspected, including all connecting cables, the printer and the monitor. This should include the software in that the correct
software must be used for the machine.
This product has no particular requirements concerning the maintenance / servicing except those detailed in Chapters below.
This product and its components will perform reliably only when operated and maintained in accordance with the instructions
given in this manual.
The maintenance requirements by the user shall be limited to the tasks detailed below.
Do not attempt to repair or replace damaged parts. Only authorized representative of DMS may carry out any form of
maintenance or repair.
The User shall not be permitted to carry out any form of maintenance, other than the visual inspection that all cables are
securely positioned.
The User shall not be permitted to carry out any form of maintenance, other than the visual confirmation that the unit is not
damaged. Identifying any damage should be reported to the authorized representative of DMS.
This shall be carried out by the approved representative of DMS, and shall consist of testing the machine for its reliability /
performance, the verification of the verification and the safety of the machine. This should be carried out on a 12 monthly
basis.
CHALLENGER_Envision _Anglais_03 10
Chapter 7 Check-out
7 Check-out
7.1 Visual inspection
Check the unit regularly for exterior damage, including all cables, minimum of once per month.
Do not attempt to repair or replace damaged parts. Only qualified DMS representatives or technicians may carry out any form
of maintenance or repair.
CHALLENGER_Envision _Anglais_03 11
Chapter 8 Environmental conditions that affect use
Failure to comply within these limits could result in inaccurate results of the examination, or in inability to carry out the
examination.
CHALLENGER is not designed to be moved over a large distance (between buildings for example) or to be submitted to hard
blows.
The chair where the patient sits for a forearm examination must be positioned a certain way so as to permits accurate
measurements. A chair mounted on wheels should never be used.
CHALLENGER_Envision _Anglais_03 12
Chapter 9 Symbols & Labels Used
Earthing protection
CHALLENGER_Envision _Anglais_03 13
Chapter 9 Symbols & Labels Used
! Warning
9.3 Labels used
The following labels shall be mounted onto the exterior of the Challenger.
WARNING
This X-ray unit may be dangerous to patient and
operator unless the following factors are observed :
Estimated Operator Dose : 0.3m = 3.0 Gy/h
: 1.0m = NONE
Patient Dose <1Gy/h / examination
High Voltage = 86kV
CAUTION ELECTRIC SHOCK
Do not remove the protective paneling. Refer all
servicing to authorized personnel.
CHALLENGER_Envision _Anglais_03 14
Chapter 9 Symbols & Labels Used
Year of Manufacture
****
DMS
Par de la Méditerranée
District de Montpellier
34470 Pérols - FRANCE
Tel. 00 33 467 50 49 00
Fax. 00 33 467 50 49 09
CHALLENGER_Envision _Anglais_03 15
Chapter 9 Symbols & Labels Used
This label is to be found on the mains supply cable going from the Mains Voltage Socket (110V or 230V) to the Machine.
They are as follows:
CHALLENGER_Envision _Anglais_03 16
Chapter 10 Equipment supplied with CHALLENGER
STANDARD UNIT
- CHALLENGER machine
- Power Cable
- Serial cable
- Serial adapter
STANDARD ACCESSORIES
- Arm phantom
- Fuse kit
- Spine phantom (used also for the femur)
DOCUMENTATION
- User manual
- Form “Installation card”
- Software disks in English
- Packaging list (giving details of the items being delivered)
- Customer request for Assistance Card
- Quick reference to use CHALLENGER
CHALLENGER_Envision _Anglais_03 17
Chapter 10 Equipment supplied with CHALLENGER
The front view of the machine with the detail of the control panel fixed on the machine’s arm and the system display switched
on (ON) during X-ray emission (CAUTION! X-RAY) is detailed in chapter 10.6 Control panel p.21
Mobile version:
C H A L LE N G E R
C H A L LE N G E R M U L TISITE
M U LT IS I T E
D E X A S C A N N E R
D EX A S C A N N E R
P O W ER
O K
C A U T IO N
PO W E R
X-R A Y
O K
ON
C A U T IO N
X- R AY
O N
C H A L L E N G E R
Stationary version:
CHALLENGER_Envision _Anglais_03 18
Chapter 10 Equipment supplied with CHALLENGER
Rear-view with the detail of the electrical panel connector with 9 pins for the connection to the serial port RS232 of the
computer using a serial cable socket ~11O/230 Volt for the voltage supply of Challenger master switch can be seen in chapter
10.5 Electrical connection procedures p.21
CHALLENGER Mobile
M AI N S I N P U T IN P U T O U TP U T
V O LT A G E
S E LE C TO R
M A IN S O U T PU T 1 0 0V -1 1 0 V / 2 2 0 V - 2 4 0 V PC RE M O T E X - R A Y O N
LO C K V O LT A G E P = 1 50 V A m a xi C O M M U N I C A TI O N S I G N A L IS A T IO N L A M P
S E LE C TO R TO T A L O U T P U T -1 5 V / + 1 5 V m a x i
CHALLENGER stationary
MA I N S I NP U T IN PUT AT GE OU T P UT
S EV OL
LEC TO R
LM OC
AI NK S OU TT PA GE
S EV OL
LEC TO R
UT 1 0 0VP -1= 1 0V / 22 0m V -2ax4 0i V
T OTA15L 0 O VA
UT P U T
PC
-CO1 5 VM /M+ U 1N 5ICVA T ma
IO N xi S IGRE NMOA LI STE A TIX -RONA Y LAO MN P
CHALLENGER_Envision _Anglais_03 19
Chapter 10 Equipment supplied with CHALLENGER
The customer for future use in case of return should keep the shipping containers and all associated hardware in store.
To ensure correct working conditions, you have to supply ~110 / 220 Volt at 50-60 Hz. to CHALLENGER
The mains socket has to be earthen.
CHALLENGER_Envision _Anglais_03 20
Chapter 10 Equipment supplied with CHALLENGER
M A IN S IN P U T IN P U T O U TP U T
V O L TA G E
S E L E C TO R
O f O n
SC A N N E R
LO C K O U TP U T PC
V O L TA G E 1 0 0 V-1 1 0 V / 2 2 0 V-2 4 0 V R E M O TE X - R A Y O N
P = 1 5 0 V A m a xi C O M M U N IC A TIO N S IG N A L IS A TIO N L A M P
S E L E C TO R -1 5 V / + 1 5 V m a xi
TO TA L O U TP U T
i) MACHINE LOCK – cuts the power to the machine, thus the unit can be switched on, but cannot be used. In
both positions ON and OFF the key may be removed. It is the responsibility of the User to assure the safety
of the keys. DMS do not hold any supplementary copies of these keys.
ii) INPUT VOLTAGE can be set to either 110V or 230V is included with the Master Power Switch. This
switch is also to be used to control the power to the computer.
iii) OUTPUT VOLTAGE can be also set to 110V or 230V independently for that of the power supply input.
The output supply points shall be used for the Computer and the Printer.
iv) PC COMMUNICATION : is the connection cable between the PC and the X-ray machine
v) REMOTE X-RAY SIGNALISATION LAMP: is an option for the hospital to connect up an external
warning light. For example on the outside of the room. Refer to Chapter 10.9 Installation of Supplementary
X-ray Warning Light p.24
The control panel is positioned on top of the machine’s arm. It is composed of a number of buttons and an Emergency Switch.
The buttons are activated only in the Remote Control Mode.
i) Emergency Switch: This shall be used ONLY in an emergency; it is not to be used for switching off the machine on
an every day basis.
ii) Laser Button : Shall be used to activate and deactivated the laser light
iii) 4 x Position Control Buttons: shall be used to locate the arm of the X-ray machine into the correct position.
iv) Confirmation of Positioning of Arm: once the correct position has been obtained using the 4 x control buttons, the
position is confirmed using the OK button.
v) Release of the Arm: Is used to release the Electro-magnet to enable the Arm to fully swing away from the bed, to
enable the patient to lie easily onto the bed. The arm can then be swung back into place, and the Electro-magnet shall
be re-activated automatically.
vi) X-RAY LIGHT ON: shall light up to indicate that X-rays are being emitted.
CHALLENGER_Envision _Anglais_03 21
Chapter 10 Equipment supplied with CHALLENGER
vii) POWER SUPPLY ON: shall when lit indicate that the machine is switched on from the Master Switch – located at
the rear of the machine (See Figure 7 - Electrical Connections p.21)
C H A L LE N G E R
M U L TIS ITE
D E X A S C A N N E R P o w e r S u p p ly
E m e rg e n c y S w it c h L ig h t O N
PO W ER
C A U T IO N
X -R A Y
X-Ra y
O N
L ig h t O N
L a s e r B u tto n R e le a s e f o r A rm
4 x P o s it io n
C o n tro l B u tto n s C o n fi rm a tio n
o f P o s itio n in g o f A rm
W A R N IN G
Y O U W IL L D A M A G E TH E U N IT IF
Y O U D O N O T F O L L O W TH E
IN S TR U C TIO N S
A TTE N TIO N (1 ) Re m o ve s id e p a n e l
E N S U R E TR A N S P O R TA TIO N
SC R EW S A N D FO A M A R E (2 ) Re m o ve TR A N S P O R T X - R A Y F IX A TIO N B L O C K
R E M O V E D B E F O R E O P E R A TIO N
R E F E R TO U S E R M A N U A L (2 ) Re m o ve TR A N S P O R T X - R A Y F IX A TIO N B A R
A TT E N TIO N
E N S U R E TR A N S P O R TA TIO N
S C R EW S A N D F O A M A R E
R E M O V E D B E F O R E O P E R A T IO N
R E F E R TO U S ER M A N U A L
(4 ) Re m o ve th e f o a m
U n s c re w 2 s c re w s (A lle n k e y s iz e 3 ) (5 ) Re p la c e th e s id e p a n e l
PO W ER
O K
C A U TIO N
X-R AY
ON
R E M O TE X - R A Y O N
S IG N A L IS A TIO N LA M P
TR A N S P O R T X - R A Y F IX A T IO N B L O C K
TR A N S P O R T X - R A Y F IX A TIO N B A R
U n s c re w 2 s c re w s
(A lle n k e y s iz e 4 )
U n s c re w 2 s c re w s
(A lle n k e y s iz e 3 )
R e m o v e th e f o a m
CHALLENGER_Envision _Anglais_03 22
Chapter 10 Equipment supplied with CHALLENGER
Remove TRANSPORT X-
RAY FIXATION BAR
Unscrew 4 screws (Allen
key size 4)
CHALLENGER_Envision _Anglais_03 23
Chapter 10 Equipment supplied with CHALLENGER
Ensure that the screws and foam are stored safely, in case of need for future transportation.
Challenger is connected:
1. to an earthen socket with ~110 / 230 Volt
2. to the computer with a screened serial cable
3. ensure that the Isolating Key (found at the rear of the Machine) is switch into the on position
For the connections use ONLY the cables supplied with the device
The option is provided for an external light to be automatically activated the moment that the machine commences the use of
X-rays. The installation of such a system is at the discretion of the user, but must comply with the details shown below:
BNC Male
(Isolated) Power Source
Fuse 1.6 A
CHALLENGER_Envision _Anglais_03 24
Chapter 10 Equipment supplied with CHALLENGER
10.10Safety rules
For a correct use of CHALLENGER and to avoid damage to the device or to the operator please follow the instructions in this
chapter and the security rules listed below:
! Do not attempt to lower the bed, until the 2 bars have been fully extended
into position. Failure to do this could damage the hinge supports.
To close up the bed, is the reverse procedure as detailed above. Note that the slide bars when fully pushed inside the body of
the machine, automatically block the two folding ends, thus holding them in place whilst under transportation.
10.12Switching on Challenger
After having connected CHALLENGER to the computer and to the mains socket, and ensured that the voltages are correctly
selected for the equipment. The equipment is ready to be switched on using the Master Switch.
To use the device, please refer to the Operator manual and to its chapters including connection, scanning and analysis.
CHALLENGER_Envision _Anglais_03 25
Chapter 10 Equipment supplied with CHALLENGER
10.13Software Installation
4. Click on “Install” to accept the directory as default, then the program continues the installation.
CHALLENGER_Envision _Anglais_03 26
Chapter 10 Equipment supplied with CHALLENGER
5. After a few seconds, a window complete is displayed saying that the CHALLENGER software has been installed in the
computer as follows:
7. The installation of the software has been completed. Be ready to enter into the software!
CHALLENGER_Envision _Anglais_03 27
Chapter 11 System Safety/ Conformance to Standards
According to the Standard IEC - 601-1. The manufacturer may be held responsible for the effects on safety, reliability and
performance of the CHALLENGER only if:
Assembly operations, setting up, adjustments, extensions, modifications or repairs are carried out by authorized people.
The electrical system of the room in which CHALLENGER is installed complies with the requirements,
Challenger is used in accordance with the instructions for use.
This product has been tested and found to be conforming to EN 60601-1-2 (Electromagnetic Compatibility). Despite this
European standard this product may cause or suffer a certain level of interference. The user should consider the following if
levels of problems are encountered:
! Do not use portable cellular equipment (walkie-talkies, radiophones, and portable telephones) in the proximity of Explorer
during its operation, as this may cause an impact on the measurements.
If the problem persists please contact DIAGNOSTIC MEDICAL SYSTEMS for advice or the supplier of the other products
concerned.
In order to avoid electrical shocks, do not remove the cover from the unit.
Electrical Specifications:
Type of protection against electric shocks for a product powered by an exterior power supply : Class I
Degree of protection against electric shocks : Type B
Degree of penetration of liquids : IPX1 (product is protected against the vertical fall of drops of water)
Sterilization : not applicable
Product is not-adapted for use in the presence of a mix of inflammable anesthetic with air or with oxygen or with nitrous
oxide
Method of Operation : continuous service
CHALLENGER_Envision _Anglais_03 28
Chapter 11 System Safety/ Conformance to Standards
This system uses an X-ray source that emits ionizing radiation in the form of X-rays. It may be dangerous to both patient and
operator if handled incorrectly. To avoid any unsafe situations do not attempt to service this equipment yourself. Always refer
to an authorized representative of DMS.
Challenger complies with IEC 601-1-3 and IEC 601-2-7 (International Electromechanical Commission)
For details of the specification of the X-rays refer to chapter 11.5 Laser Light Radiation p.29.
It is the responsibility of the User to check the regulations concerning X-ray equipment. This may include notifying the
regulatory agency about your installation and / or operator training and certification. Please check the regulations in your
province or country.
This machine is equipped with a laser light to aid the operator to locate the correct position of the arm in relation to the
patient.
The label attached to the machine is detailed in chapter 9.1 Symbols used on the device p.13
The Challenger is equipped with an emergency stop push button. If this button is pushed the machine will immediately stop
and no more X-rays will be generated.
The Emergency Stop Button should be left on at all times. This button should not be used as a means to stop / start the power
on the machine.
If this button is used, then the following procedure must take place:
i) Switch off the Master Switch found behind the machine
ii) Reset the Emergency Button
iii) Switch on the master switch
iv) Carry out an Quality Control Verification Examination, before continuing with the patient scans
NOTE: If the operator suspects a problem then they should contact immediately, their authorized representative of
DMS.
This key is found at the rear of the machine, and enables the operator to close down completely the machine, cutting off the
power supply to both the machine and the computer. Thus ensuring there is no risk of any unauthorized use of the machine in
their absence.
The key should be placed in the key hole at all times, except when the machine is required to be isolated to prevent
unauthorized use.
CHALLENGER_Envision _Anglais_03 29
Chapter 11 System Safety/ Conformance to Standards
European Standards require that at the end of the useful life of a machine it should be disposed of in an appropriate manner.
DMS recommends disposing of this unit in accordance with the latest European Standards at that time.
If a doubt exists regarding the nature and type of components/materials used, please contact DMS prior to its disposal.
11.9 Registration
Government health departments often require medical facilities to register diagnostic x-ray equipment. Many municipal and
state health agencies require medical health facilities to employ certified radiologic technologists to operate diagnostic x-ray
devices. Contact the Department of Health office in your city, state, or country to ensure that proper guidelines for registration
and regulation compliance are being met.
Laser registration also varies from state to state. Contact the Department of Health in your city, state or country for
information about the laser registration.
11.10Facilities
Post a “Caution X-radiation” sign in the area or room where the system is operated. The system should not require additional
shielding in he walls, floor, or ceiling of the room where the system is installed. This is due to the low levels of leakage
radiation from the x-ray tube housing assembly. However consult with your state or local health and radiation safety
department regarding shielding requirements for your country.
No additional signs are required for the laser, but you may want to contact the Department of Health in your city, state or
country for possible regulations.
It is impossible that you will receive or excess the maximum permissible dose while working with a CHALLENGER bone
mineral analyzer.
However, some facilities may choose to use personnel monitors due to the surrounding environment, i.e. other x-ray machines
etc. Consult with your Radiation Safety Officer regarding your facility’s policy.
Film badges and thermal luminescent dosimeter (TLD) badges for personnel monitoring can be obtained from a supplier
accredited for personnel dosimetry processing.
CHALLENGER_Envision _Anglais_03 30
Chapter.14: Operator manual
12 DEXA Principle
12.1 General theory on the Dual Energy X-ray Absorptiometry
Dual energy X-ray Absorptiometry (DEXA) enables quantitative assessment of bone mineral in specific
regions of the body. The use of two photon energies minimizes errors that result from irregular soft body
masses and body contours.
In theory, in order to analyze a given number of substances, attenuation measurements at the same
number of discreet photon energies are required. Simultaneous equations obtained by applications of
Lambert’s law can be solved only if there are as many independent equations as unknowns. Since the
attenuation coefficient is correlated, the number of substances that can be determined is limited.
Two photon energies, therefore, allow discrimination of two substances in a given system. If only two
substances are present, the technique is capable of high accuracy. In the case where more than two
substances are present, the accuracy with which the two substances of interest can be measured depends
on the number of additional substances, their attenuation characteristics, and the fraction of total that
they represent.
Clinically, a two component system can be defined as consisting of bone mineral and soft tissue. The
equations describing the attenuation of two photons, as they pass through a bone tissue mass, are b:
Where:
IO = Unattenuated photon fluence rate (photon /sec)
I =Attenuated photon fluence rate (photon/sec)
M = Mass of given substance (g/cm²)
µ = Mass attenuation coefficient (cm²/g)
l = 40 keV
h = 65 keV
s, b = Subscripts for soft-tissue and bone
The photon energies are stated as 40 keV and 65 keV, since these are currently the most commonly used
energies for DEXA scanning. These low energy photons enable detection of very small changes in the
two component systems. The photon fluences can be measured directly and the attenuation coefficients
can be determined by various methods, the simplest being the use of published values. Only two
quantities, MS and MB, result in the following expression:
CHALLENGER_Envision _Anglais_03 31
Chapter.14: Operator manual
RB Ln(IH/IOH) – Ln(IL/IOL)
MS in g/cm² =
µSL- µSH RB
RS Ln(IH/IOH) – Ln(IL/IOL)
MB in g/cm² =
µBL - µBH RS
Where:
RS = µSL/µSH
RB = µBL/µBH
The R value is the ratio of soft tissue attenuation at 40keV to that at 65keV. A typical ratio is 1.35 with
higher values (1,40) in lean subjects and lower values (1,30) in obese subjects.
Approximately 60% of the weight of hydrated bone is due to bone mineral and 38% of the bone mineral
is calcium. The non-mineral components of bone include fat, collagen, and water of hydration. The
composition of the bone
Mineral is essentially invariable, while the non-mineral components of bone and the composition of
surrounding soft tissue vary considerably between individuals and over time. The data reduction
algorithm used by CHALLENGER makes corrections for the variability in the soft tissue composition.
The machine continuously samples the bone mineral content (BMC) as it moves in a rectilinear pattern
over the lumbar spine or other area of interest. The resulting series of transverse profiles provides the
data base fort quantitative analysis. The calculated BMC is calibrated against a set of bone simulating
standards of known bone mineral content in order to correct for systematic errors and simplify the data
processing.
The BMD is the BMC normalized for bone size and provides the most clinical information. BMC and
Area values are provided on analysis printouts, but they are intended only for research use, not for
clinical use.
Patient movement during the examination can compromise the precision of total and regional
measurements. Thus, the patient should be encouraged to lie quietly during the entire length of the
examination.
CHALLENGER_Envision _Anglais_03 32
Chapter.14: Operator manual
12.2.1 AP spine
In the lumbar spine region, the BMD is usually calculated from L2 through L4, although results can be
calculated for other regions-of-interest. Results other than L2 through L4 are useful in situations where
L2, L3 or L4 cannot be analyzed due to prostheses or fractures in those vertebral regions.
The precision of the machine on the spine may be adversely affected if fewer than three vertebrae are
used in longitudinal studies or if the exact same region is not analyzed – e.g., if L1-L3 is mistaken for
L2-L4. Factors that can affect results are the anatomical variability of L1 and the presence of ribs near
L1, the pelvis around L5, and the inconsistent use of only one or two vertebrae as the region-of-interest.
Inclusion of L14 in the region-of-interest (L1 though L4) will usually decrease the mean BMD of the
region; whereas, inclusion of L5 (L2 though L5) increases the mean density, because L1 typically has a
lower BMD and L5 a higher BMD than the L2-L4 sequence.
12.2.2 Femur
In the proximal femur, the region-of- interest is usually the area 1.5 cm wide across the entire femoral
neck. Additional regions are defined by the software in the lower density Ward’s triangle region, and in
the region of the greater Trochanter. Bone loss in the proximal femur begins in the Ward’s triangle
region and proceeds outward from there. This makes the region an early indicator of bone loss, but the
higher variance in measuring it, compared to the neck region, makes the latter zone a better
discriminator.
12.2.3 Forearm
In forearm scans there are three regions-of-interest: the Ultradistal site, the Mid-region site and the 1/3
site for Ulna and Radius. Repeat determinations of regions and subregions at different times can be best
compared by reviewing the print out from previous scans to verify that the regions are comparably
located (i.e., the cuts are located similarly). Small variations in delineation of these regions have a larger
effect on the BMC than on the BMD.
CHALLENGER_Envision _Anglais_03 33
Chapter.14: Operator manual
The T-score is defined as the difference between the patient’s value and the mean value in a population
of young healthy subjects, which are the same gender and race as the patient. “Young” means the age of
peak bone mass, typically between 20 and 40 years. This difference is expressed as a number of
standard deviations, which avoids any dependency on the parameter measured, its unit and its biological
variation. In a more mathematical expression:
During childhood, bone density regularly increases until approximately 20 years and then it stabilizes
for several years. After 40 years, it will start to decrease due to normal aging. There is therefore a
maximum value, the “peak bone mass”, obtained between 20 and 40, which is used as reference value.
It represents the typical BUA value in a healthy bone. The difference between the patient’s value and
this reference value is a logical index of bone health.
However, the simple difference cannot be directly exploited, because it depends on the scale of the
parameter used. It is therefore necessary to divide it by the standard deviation in the reference
population, which measures the dispersion of the values observed in the reference population. Let’s take
an example. The reference temperature in the human body is 37°C. If a patient’s temperature is 36.8°C,
we can say that this patient is 0.2°C below the normal, but should we worry about him? To obtain the
answer, we must know that the values measured in a healthy reference population can vary from –0.5°C
to +0.5°C compared to the mean value of 37°C. If we now compare our difference: 0.2°C, to the typical
variation: 0.5°C, we understand that our patient is healthy.
By definition of the standard deviation, 95% of the normal population has a value comprised between
the mean minus two standard deviations, and the mean plus two standard deviation, which we can note:
[mean-2SD, mean+2SD]. The World Health Organization defined that osteoporosis is established when
a BMD measurement gives a value below mean –2.5SD. It is equivalent to say that osteoporosis occurs
when the T-score is inferior to –2.5. Between –1, and –2.5, one says the patient has O steopenia (i.e.
“lack of bone”). In practice, there is of course no such clear limit, and the risk of fracture progressively
increases when the T-score decreases.
The Z-score is defined as the difference between the patient’s value and the mean value in a population
of healthy subjects of the same age, also divided by the standard deviation. In practice, a clinical study
allows one to determine a “reference curve”, representing the measurement mean value as a function of
age in a healthy population.
For young subjects, i.e. between 20 and 40, there is no significant difference between Z-score and T-
score.
For older patients however, the bone density decreases, and it is normal to observe a decreasing of T-
score. The Z-score is then useful to determine if the low bone density of an old patient is due to the
normal decreasing of bone density with age (in this case the Z-score is close to 0), or if there is a “high -
risk” of fracture (in this case the Z-score is negative).
CHALLENGER_Envision _Anglais_03 34
Chapter.14: Operator manual
The reference curve displays the Bone Mineral Density (BMD) in function of ages, for one or all region
of interest of the study.
.
The reference curve is provided for the spine’s L2, L3, L4 vertebrae, the Femur’s Neck, Greater
Trochanter, Intertrochanter and the Forearm’s Ultradistal, Midregion and 1/3 in order to generate T-score
and Z-score values as diagnostic values.
N o rm a l
O s t e o p e n ia - 1 .0
- 2 .5
O s t e o p o ro s is
Age
T- S c o re
- 1 .0
- 2 .5
Age
CHALLENGER_Envision _Anglais_03 35
Chapter.14: Operator manual
14 Operator Manual
1. DAILY SEASONING AND SOFTWARE PRESENTATION................................................41
1.1. DAILY SEASONING...............................................................................................................................................41
1.2. SOFTWARE PRESENTATION...................................................................................................................................44
2. START EXAMINATION AND ANALYSIS............................................................................47
2.1. HOW TO MAKE AN ACQUISITION.........................................................................................................................47
2.2. HOW TO READ THE EXAM WINDOW....................................................................................................................80
2.3. HOW TO MODIFY AN UNSATISFACTORY ANALYSIS..............................................................................................90
2.4. MEASURE: AREA , DISTANCE AND ANGLE...........................................................................................116
3. WORKING ON AN ANALYSIS INDEPENDENTLY OF AN EXAMINATION..............123
3.1. OPENING THE “VIEW SCAN” SECTION..............................................................................................................123
3.2. DISPLAYING AN ANALYSIS AFTER SELECTION WITH PATIENT’S LAST NAME...................................................124
3.3. DISPLAYING AN ANALYSIS AFTER RESEARCH WITH EXAM TYPE OR DATE........................................................125
3.4. DISPLAYING THE PATIENT FOLLOW-UP..............................................................................................................128
3.5. PRINTING THE MEDICAL REPORT.......................................................................................................................138
3.6. PRINTING MULTI SCAN REPORT.........................................................................................................................139
3.7. PRINTING ON A DICOM SYSTEM......................................................................................................................142
4. DATA MANAGEMENT............................................................................................................145
4.1. OPENING THE DATA MANAGEMENT FILE..........................................................................................................145
4.2. HOW TO PERFORM A RESEARCH........................................................................................................................146
4.3. HOW TO MANAGE A PATIENT FILE.....................................................................................................................152
5 DAILY QUALITY CONTROL..............................................................................................163
5.1 OPENING THE QUALITY CONTROL FILE.............................................................................................................163
5.2 HOW TO CARRY OUT THE QUALITY CONTROL...................................................................................................164
5.3 HOW TO VISUALIZE PREVIOUS CONTROLS........................................................................................................169
5.4 VIEWING THE QUALITY CONTROL RESULTS.....................................................................................................170
5.5 HOW TO EXPORT A FILE....................................................................................................................................176
5.6 PRINTING THE GRAPH AND THE RESULTS OUT..................................................................................................177
5.7 QUITTING THE “QUALITY CONTROL” FILE.......................................................................................................178
6 OPTIONS.................................................................................................................................181
6.1 DATABASE.........................................................................................................................................................182
6.2 ARCHIVAGE.......................................................................................................................................................183
6.3 SCAN.................................................................................................................................................................200
6.4 LANGUAGE........................................................................................................................................................202
6.5 CALCULATION...................................................................................................................................................204
6.6 PRINT................................................................................................................................................................205
6.7 INSTITUTE INFORMATION..................................................................................................................................206
6.8 WARMUP AND QUALITY CONTROL...................................................................................................................207
6.9 E-MAIL AND DICOM.......................................................................................................................................209
6.10 CONTACT...........................................................................................................................................................210
6.11 COMMENTS........................................................................................................................................................211
6.12 MAINTENANCE..................................................................................................................................................212
7 QUITTING THE CHALLENGER PROGRAM..................................................................213
CHALLENGER_Envision _Anglais_03 37
Chapter.14: Operator manual
This chapter will detail the different steps of the machine seasoning, before giving an overview of the
CHALLENGER Envision software content.
1. DAILY
CHALLENGER_Envision _Anglais_03 39
Chapter.14: Operator manual
2. If not, turn the machine on : set the main input buttons on “1”
4. Open Challenger software by double-clicking the icon on the desktop Windows screen. The following window
appears on the screen :
CHALLENGER_Envision _Anglais_03 41
Chapter.14: Operator manual
5. Click on “Perform Init”. The machine arm positions itself at the initialization place on the left hand side of the
device. The following window appears on the screen :
CHALLENGER_Envision _Anglais_03 42
Chapter.14: Operator manual
7. Monoblock warm up
! Make sure there is nothing on the bed during the X-Ray shoot time !
To perform the quality control, please refer to chapter 5 Daily Quality Control p.161.
CHALLENGER_Envision _Anglais_03 43
Chapter.14: Operator manual
The main menu introduces the user to the different options available in the software: the operator/the user uses them while
making an examination or working on a patient database or analysis independently of an examination.
CHALLENGER_Envision _Anglais_03 44
Chapter.14: Operator manual
This chapter will describe the stages the user must go through in order to perform an examination, from the
examination processing to the analysis of a scanned image.
It will show the user how to fill in or open a patient database previous to a scanning session.
It will also give the different stages of an examination processing from the positioning of the patient and the
Region of Interest adjustment to the scan processing and the analysis.
T his chapter will detail the stages the user must go through in order to carry out the analysis of a
scanned image, restart an unsatisfactory operation at any stage of the analysis process, either immediately after a
examination or independently of an examination session.
CHALLENGER_Envision _Anglais_03 46
Chapter.14: Operator manual
Click on “Start scan” on the main menu. The following window appears on the screen :
2. Click on “Start Scan” at the bottom left hand side of the screen. Then please follow instructions exposed in
chapter 2.1.4 How to start an examination p.50.
1. Click with the right mouse button on “Start Scan” button on the main menu and choose “Start Scan with patient…”
CHALLENGER_Envision _Anglais_03 47
Chapter.14: Operator manual
CHALLENGER_Envision _Anglais_03 48
Chapter.14: Operator manual
Type his / her surname, first name, birth date (month, day, year) and sex.
Click on the field and drag the mouse’s cursor down to the patient’s ethnic group.
Optionally, type his /her full address, telephone number and personal code.
3. Optionally, fill in the “Note” field with additional information about the patient (patient’s medical history, sports
practiced, etc.)
Refer to 2.2 “How to print the medical report” p.89to print the patient comment
CHALLENGER_Envision _Anglais_03 49
Chapter.14: Operator manual
1. After filling the patient’s private data part, click on “Start scan” at the bottom of the window. The following
window appears on the screen :
2. Patient height and weight allow the system to determine the mode (Normal or Large) for the acquisition; indicate
the patient’s height and weight and select the site to examine in the unrolling menu as follows. :
CHALLENGER_Envision _Anglais_03 50
Chapter.14: Operator manual
3. Do not forget to enter the forearm length (measure from the articulation wrist to the elbow) in the forearm case.
STEP 2: POSITIONING
Patient positioning
1. To let the patient lay on the table more easily, set the machine’s arm on the side of the table with the buttons of the
remote control:
!The patient must be correctly positioned on the imaging table, the cushion bearing his legs and arms along the body. In the
event of a femur examination, the limbs of the patient will be surrounded by two cushions filled with rice so as to prevent the
laser light from scanning air.
The doctor shall never position the patient’s arm upon the cushions.
CHALLENGER_Envision _Anglais_03 51
Chapter.14: Operator manual
CHALLENGER_Envision _Anglais_03 52
Chapter.14: Operator manual
2. Once the patient has been positioned on the examination table, click on “YES”.
Position the laser light at the center of the beginning of the scanned area using the arrow buttons on the remote control panel
of the machine ( )
CHALLENGER_Envision _Anglais_03 53
Chapter.14: Operator manual
Take the time to position the laser as shown on the figure! This will save examination time.
Figure 29 - Position of the laser for a spine examination Figure 30 - Position of the laser for a femur examination
5. Once the laser has precisely been positioned on the region to examination, click on “STOP” on the remote control
window on the screen or on “OK” on the machine‘s remote control.
CHALLENGER_Envision _Anglais_03 54
Chapter.14: Operator manual
STEP 3 : ACQUISITION
SPINE ACQUISITION
1. Select Spine area by clicking on the skeleton or by selection in the district menu.
CHALLENGER_Envision _Anglais_03 55
Chapter.14: Operator manual
3. The auto-verification starts up: the machine should emit X-rays for about 20 seconds.
CHALLENGER_Envision _Anglais_03 56
Chapter.14: Operator manual
4. If it is the first time you make a spine examination, the machine will ask you to confirm these values as references
values
The remote control window appears on screen to activate the arm remote control. To let the patient lay on the table more
easily, set the machine’s arm on the side of the table with the buttons of the remote control:
To position the patient on the table refer to Position the patient on the scanning table as follows :p.51.
CHALLENGER_Envision _Anglais_03 57
Chapter.14: Operator manual
5. Once the examination is finished the window as follow appears. You can open the arm, using the keyboard button,
and let the patient go down from the table
CHALLENGER_Envision _Anglais_03 58
Chapter.14: Operator manual
FEMUR ACQUISITION
This procedure will be applied as well as for right and left femur or both.
1. Select Femur area by clicking on the skeleton or by selection in the district menu.
CHALLENGER_Envision _Anglais_03 59
Chapter.14: Operator manual
3. The auto-verification starts up: the machine should emit X-rays for about 20 seconds.
CHALLENGER_Envision _Anglais_03 60
Chapter.14: Operator manual
4. If it is the first time you make a femur examination, the machine will ask you to confirm these values as references
values
The remote control window appears on screen to activate the arm remote control. To let the patient lay on the table more
easily, set the machine’s arm on the side of the table with the buttons of the remote control:
To position the patient on the table refer to Position the patient on the scanning table as follows :p.51.
CHALLENGER_Envision _Anglais_03 61
Chapter.14: Operator manual
5. Once the examination is finished the window as follow appears. You can open the arm, using the keyboard button,
and let the patient go down from the table
CHALLENGER_Envision _Anglais_03 62
Chapter.14: Operator manual
FOREARM ACQUISITION
This procedure will be applied as well as for right and left Forearm.
1. Select Forearm area by clicking on the skeleton or by selection in the district menu.
CHALLENGER_Envision _Anglais_03 63
Chapter.14: Operator manual
3. The auto-verification starts up: the machine should emit X-rays for about 20 seconds.
CHALLENGER_Envision _Anglais_03 64
Chapter.14: Operator manual
4. If it is the first time you make a Forearm examination, the machine will ask you to confirm these values as
references values
The remote control window appears on screen to activate the arm remote control. To let the patient lay on the table more
easily, set the machine’s arm on the side of the table with the buttons of the remote control:
To position the patient on the table refer to Position the patient on the scanning table as follows : p.51.
CHALLENGER_Envision _Anglais_03 65
Chapter.14: Operator manual
5. Once the examination is finished the window as follow appears. You can open the arm, using the keyboard button,
and let the patient go down from the table
CHALLENGER_Envision _Anglais_03 66
Chapter.14: Operator manual
1. Select Dual Femur area by clicking on the skeleton or by selection in the district menu.
CHALLENGER_Envision _Anglais_03 67
Chapter.14: Operator manual
3. The auto-verification starts up: the machine should emit X-rays for about 20 seconds.
CHALLENGER_Envision _Anglais_03 68
Chapter.14: Operator manual
4. If it is the first time you make a Femur examination, the machine will ask you to confirm these values as references
values
The remote control window appears on screen to activate the arm remote control. To let the patient lay on the table more
easily, set the machine’s arm on the side of the table with the buttons of the remote control:
To position the patient on the table refer to Position the patient on the scanning table as follows : p.51.
CHALLENGER_Envision _Anglais_03 69
Chapter.14: Operator manual
5. Once the examination is finished the window as follow appears. Click on OK to start the second verification.
!Before clicking on “OK”, make sure the patient is not on the cushion !
6. The auto-verification starts up to the second exam: the machine should emit X-rays for about 20 seconds.
CHALLENGER_Envision _Anglais_03 70
Chapter.14: Operator manual
The remote control window appears on screen to activate the arm remote control. To let the patient lay on the table more
easily, set the machine’s arm on the side of the table with the buttons of the remote control:
To position the patient on the table refer to Position the patient on the scanning table as follows : p.51.
7. Once the examination is finished the window as follow appears. You can open the arm, using the keyboard button,
and let the patient go down from the table
CHALLENGER_Envision _Anglais_03 71
Chapter.14: Operator manual
To modify the Left or Right Femur, click on the wanted button and refer to Chapter 2.3 How to modify
an unsatisfactory analysis p.99
CHALLENGER_Envision _Anglais_03 72
Chapter.14: Operator manual
1. Select CombiScan area by clicking on the skeleton or by selection in the district menu.
Figure 63 – CombiScan with Left Femur area selection by clicking on the skeleton
Figure 64 - CombiScan with Right Femur area selection by clicking on the skeleton
CHALLENGER_Envision _Anglais_03 73
Chapter.14: Operator manual
CHALLENGER_Envision _Anglais_03 74
Chapter.14: Operator manual
3. The auto-verification starts up: the machine should emit X-rays for about 20 seconds.
4. If it is the first time you make a Femur examination, the machine will ask you to confirm these values as references
values
CHALLENGER_Envision _Anglais_03 75
Chapter.14: Operator manual
The remote control window appears on screen to activate the arm remote control. To let the patient lay on the table more
easily, set the machine’s arm on the side of the table with the buttons of the remote control:
To position the patient on the table refer to Position the patient on the scanning table as follows : p.51.
Click on valid to start the examination.
5. Once the examination is finished the window as follow appears. Click on OK to start the second verification.
! Before clicking on “OK”, make sure the patient is not on the cushion !
6. The auto-verification starts up to the second exam: the machine should emit X-rays for about 20 seconds.
CHALLENGER_Envision _Anglais_03 76
Chapter.14: Operator manual
CHALLENGER_Envision _Anglais_03 77
Chapter.14: Operator manual
The remote control window appears on screen to activate the arm remote control. To let the patient lay on the table more
easily, set the machine’s arm on the side of the table with the buttons of the remote control:
To position the patient on the table refer to Position the patient on the scanning table as follows : p.51.
8. Once the examination is finished the window as follow appears. You can open the arm, using the keyboard button,
and let the patient go down from the table
CHALLENGER_Envision _Anglais_03 78
Chapter.14: Operator manual
To modify a Spine analysis, refer to Chapter 2.3 How to modify an unsatisfactory analysis p.90
To modify a Femur analysis, refer to Chapter 2.3 How to modify an unsatisfactory analysis
p.99
CHALLENGER_Envision _Anglais_03 79
Chapter.14: Operator manual
Functions available on the analysis window and for any site (Forearm, Femur, Spine):
CHALLENGER_Envision _Anglais_03 80
Chapter.14: Operator manual
In order to change the contrast, perform the image quality and allow a better reading, click on the moving command
with the mouse and move it as follows :
1. The doctor might want to remove the mask to obtain a quasi radiological image. Click on “Display” as follows,
CHALLENGER_Envision _Anglais_03 81
Chapter.14: Operator manual
2. A new window appears, click to “OUTLINE” to remove the mask, and click to “Standard ROI” to remove the
different ROI.
CHALLENGER_Envision _Anglais_03 82
Chapter.14: Operator manual
The user may use the “Zoom” function for better precision.
CHALLENGER_Envision _Anglais_03 83
Chapter.14: Operator manual
- Position the cursor in the blue frame to move the “Zoom” window at the wanted place :
It is possible for the user to consult the normality curve on the top right hand corner and the results board:
CHALLENGER_Envision _Anglais_03 84
1. Click on “Detailed results” in order to obtain the patient densitometry data. The following table appears on the
screen :
Refer to chapter 2.2 “How to print the medical report” p.89 In order to print comments on medical report.
How to display the patient follow-up
2. The following window appears on the screen, displaying results and graph per exam :
3. Click on a point on the graph, the corresponding exam image appears on the left hand side of the window.
If L2, L3, L4, L5 are displayed, it is necessary to modify the label of the first ROI. Press the “Modify ROI Label” button at the
bottom of the window:
2. The following window appears on the screen, displaying the printed report :
To print Patient /Exam comments, refer to chapter 6.6, Print p.205 and select this parameter.
2.3. How to modify an unsatisfactory analysis
The practitioner might be unsatisfied by the results and the analysis image; it is possible for him to work on the examination
just after or independently of the examination session:
He will redefine the outlines and the ROI positioning before BMD recalculation. Proceed as follows for any site examinations:
1. Immediately after the examination process, click on “Modify analysis” at the bottom left hand side of the analysis
window :
2. The analysis window appears on the screen. Please see below how to modify the analysis for each site :
After clicking on “Modify Analysis”, the doctor should obtain an image as follows :
3. The image appears automatically in “Mask” mode and the cursor is changed into a blue pastille.
The user may use the “Zoom” function if necessary at the bottom of the screen.
The cursor size does not change during the “Zoom” function using.
- To add some outlines, click on the left button of the mouse
Figure 103 - Before removing outlines Figure 104 - After removing outlines
o Cancel action
Click on “Cancel action” to erase last action(s) if necessary. Click in the unrolling field and select the number of actions to
erase.
o Modify the ROI
2. The user may work the ROI in the Normal or Mask mode : tick the box “Fill outline” in order to remove the mask
Click on yes. The new results will be displayed on the analysis window.
- To come back to the former results, click on “cancel analysis”. The following message appears on the screen :
The following window appears on the screen, displaying the regions of interest :
In a similar case and considering on the following drawing how the region of interest should be placed, it is highly
recommended to the doctor to proceed to a manual positioning.
The user may use the “Zoom” function if necessary at the bottom of the screen.
The cursor size does not change during the “Zoom” function using.
3. Modify the ROI with the help of the mouse by clicking in the blue frames and moving the mouse and following the
images exposed just below. First change the baseline positioning if wrong :
Click in the blue frame as indicated
opposite and place the baseline just
Figure 124 - Raising the baseline just below the small trochanter
below the small trochanter.
4. Modify the femoral neck ROI box, it should be perpendicular to the central axis of the femur neck
First modify the central axis of the femur neck until position it at equal distance from the femur neck edges:
- Then click in the blue frame to move the femoral neck ROI and take it down until it is tengential to the greater trochanter
ridge. It should include bone of the femoral neck and soft tissue at either end of the box. Please pay attention not to include
bone from the trochanter or ischium (it may occur if the ischio-femoral angle is closed).
Greater
trochanter ridge
Ischio-femoral
angle
The Ward’s triangle should be placed in the base of the femoral neck, in the area of lowest bone mass. Adjusting the
ward position to find the lowest BMD ; to do that consult the BMD value at the bottom of the window :
Click on Yes.
The following window appears on the screen, displaying the regions of interest :
In a similar case and considering on the following drawing how the region of interest should be placed, it is highly
recommended to the doctor to proceed to an analysis modification.
Proceed as follows:
1. After exam acquisition, click on “Outline” at the bottom left hand side of the analysis window :
The user may use the “Zoom” function if necessary at the bottom of the screen.
The cursor size does not change during the “Zoom” function using.
Cortical Bone
7. Next adjust the central and lateral axis as follows ; include soft tissues of the forearm in the ROI :
- To come back to the former results, click on “Cancel Analysis”. The following message appears on the screen :
Click on Yes.
You can have a measure of bone on any Exam Spine, Femur or Forearm
2. Click on “measurement”
This chapter will describe the functions of the software the doctor can use independently of an examination
session. It will show how to work on a patient analysis, and how to restart an unsatisfactory analysis,
independently of an examination.
2. The following window appears on the screen, displaying the existing patients and performed examinations lists
:
1. Unroll the patients list and select the patient name of whom the practitioner wants to display the analysis already
performed.
2. Then click on the wanted exam in the list at the bottom of the window; it is composed of the examination type,
date and time.
4. The analysis window appears on the screen, displaying the exam selected in the list :
3. The practitioner may perform a research either with exam type or with exam date :
1. Click in the “Exam Type” field at the top left hand corner of the window
The unrolling menu displays the different sites (All sites / left femur / Right femur / Right forearm / Left forearm / Spine /
paediatric Spine).
2. Select the wanted exam type with the help of the mouse. The list of all exams of this type is then displayed :
XXXX
XXXX
Figure 161 - Selection of the “Left femur” exam type for research
XXXX
XXXX
1. Fill in the fields “Scan date” at the top right hand side of the window :
XXXX
XXXX
3. Click on “View scan” at the bottom of the window or double-click on the wanted exam to display the analysis.
3.4. Displaying the patient follow-up
Parameters Selection
XXXXXXXX
2. Select the exam region wanted in the unrolling menu at the top left hand side of the window in order to perform the
patient follow-up :
XXXXXXXXX
Select the wanted scale, the corresponding co-ordinates will appear on the vertical axis of the graph :
5. Select the District we want to see the follow-up of in the spine case (L1 ; L2 ; L3 ; L4 ; Selection ; total)
The corresponding graph and results board are displayed on the screen :
L2 dictrict selected
See exam
1. Click on “See exam” at the bottom left hand corner of the window
3. Click on “Modify Analysis” at the bottom left hand side of the window.
1. Click on “Print trend” at the bottom left hand side of the window :
3. Select one this button “Curve” or “Pictures” or “Pictures and Curves” to print.
4. When you click on “Curve” displaying the graph and the results board according to the selected criteria :
3. Fill in the “Comment” part and click on “Update” at the bottom left hand side of the window.
3. Click on “Print” at the bottom right hand side of the window to launch the printing.
3.6. Printing Multi scan report.
The interest of this option is to print on the same page until 3 exams for the same patient.
1. Select a patient then click on “Print Multi scan” at the bottom of the “View Scan” window:
The different arrow on the right side permits to change the position of the examination site on the printing paper
5. Click on “Print” at the bottom right hand side of the window to launch the printing
3.7. Printing on a DICOM system
3. Click on “Print” at the bottom right hand side of the window to launch the printing.
4 Data management
This chapter will describe how performing exams or existing p patients researches, how reorganizing or
updating the patient database and the possible modifications of an unsatisfactory analysis independently of an
examination session.
4. DATA MANAGEMENT..............................................................................................................145
4.1. OPENING THE DATA MANAGEMENT FILE..........................................................................................................145
4.2. HOW TO PERFORM A RESEARCH........................................................................................................................146
4.2.1. Research data with the database..............................................................................................................147
4.2.2. Research data with the patient list...........................................................................................................147
4.2.3. Research data with the exams list............................................................................................................150
4.3. HOW TO MANAGE A PATIENT FILE.....................................................................................................................152
4.3.1. “Modify” : Modify an analysis or update patient data...........................................................................153
4.3.2. “Erase“ : Erasing a patient’s examination, analysis or file....................................................................157
4.3.3. “See”........................................................................................................................................................158
4.3.4. “Print”.....................................................................................................................................................158
4.3.5. “Add”........................................................................................................................................................159
4.Data Management
4.1. Opening the Data Management file
Once the database file has been opened (refer to chapter 2.1) the doctor can manage the patient files.
1. From the main menu, click on the “Data Management” button to open the file
2. The following window appears on the screen, displaying the patient list :
XXXXXXXX
XXXXXXXX
XXXXXXXX
XXXXXXXXX
XXXXXXXX
CHALLENGER_Envision_Anglais_03 145
4.2. How to perform a research
Please refer to the fields located on the left hand side of the window:
CHALLENGER_Envision_Anglais_03 146
4.2.1. Research data with the database
2. Select the database in which the user want s to find data, then launch the research by clicking on “Launch
search” at the bottom right hand side of the window :
1. Enter the patient’s name in the field “Last Name” of the search by patient part :
2. Click on “Launch search” at the bottom right hand side of the screen :
3. Data (Birth date, types, performed exams list and date…) of the wanted patient appear at the left hand side of
the window.
CHALLENGER_Envision_Anglais_03 147
Research data with the patient’s birth date
2. Click on “Launch search” at the bottom right hand side of the screen :
3. The data (Birth date, site, performed exam list and date…) of the patients born at the wanted date appear on
the left hand side of the screen.
2. Unroll the menu and choose to get the list of all existing patients, the list of the female patients only or the
list of the male patient only.
CHALLENGER_Envision_Anglais_03 148
3. Click on “Launch search” at the bottom right hand side of the window :
4. The patient data (Birth date, sites, exams performed list and date…) of male, female or all appear in the list
on the left hand side of the screen according to the request.
2. Unroll the menu and choose to get the database with patient from the wanted ethnic group.
3. Click on “Launch search” at the bottom right hand side of the window :
4. The data (Birth date, sites, performed exams list and date…) of the patient from the wanted ethnic group
appear in the list on the left hand side of the screen.
CHALLENGER_Envision_Anglais_03 149
4.2.3. Research data with the exams list
1. Click in the field “Site” in the part “Search by exam criteria” part :
2. Unroll the list and select the exam site of which you want to display the results.
3. Click on “Launch search” at the bottom right hand side of the window :
4. The data list of the patients who underwent those types of exam appears on the screen.
CHALLENGER_Envision_Anglais_03 150
Research data with the exam date
2. Click on “launch search” at the bottom right hand side of the window :
3. The list of the patients who underwent an exam at the selected date appears on the left hand side of the
screen.
2. Select the wanted operator name and click on “Launch search” at the bottom right hand side of the window :
3. The patient database of the selected operator appears on the left hand side of the window.
CHALLENGER_Envision_Anglais_03 151
4.3. How to manage a patient file
After performed the research, unroll the patients list, select the wanted patient or exam :
XXXXXXXX XXXXXXXX
XXXXXXXX XXXXXXXX
XXXXXXX XXXXXXX
XXXXXXXX XXXXXXXX
XXXXXXXX XXXXXXXX
Figure 212 - “Launch Search” button Figure 213 - “Launch Search” button
CHALLENGER_Envision_Anglais_03 152
4.3.1. “Modify” : Modify an analysis or update patient data
The operator will use this option in order to modify an exam or update data already stocked in a patient database (new
address, phone number, etc.)
Modify an analysis
Select the exam in the list of the “Data management” window (Please refer to the 4.1 chapter above). Click
on “Update” at the bottom of the window. The analysis window appears on the screen. Please follow
indications in the chapter 2.3 How to modify an unsatisfactory analysis p 59
CHALLENGER_Envision_Anglais_03 153
Modify patient data
1. Select the patient in the list of the “Database management” window (Please refer to 4.2 chapter above) :
XXXXXXXX
XXXXXXXX
XXXXXXXX
XXXXXXXXX
XXXXXXXX
2. Click on “Update” at the bottom left hand side of the window. The confidential patient data file appears on
the screen :
CHALLENGER_Envision_Anglais_03 154
XXXX
XXXX
XXXXX
XXXXX
CHALLENGER_Envision_Anglais_03 155
4. Click on the “Update” Button at the bottom left hand side of the window in order to record the new data.
XXXX
XXXX
CHALLENGER_Envision_Anglais_03 156
4.3.2. “Erase“ : Erasing a patient’s examination, analysis or file
The doctor will use this option to delete a patient’s examination, analysis or file when he won’t have anymore
examination.
Proceed in the same way for all the scans or analyses to erase.
First remove all the patients’ scans or analyses before erasing the pa tient’s file
1. Once all the patient’s scans have been erased, click on the “Erase” button
3. Click on OK
4. The database should definitely be cleared out of all the data previously stored in it
CHALLENGER_Envision_Anglais_03 157
4.3.3. “See”
The practitioner will use this option in order to view the patient’s analysis before performing updates.
4.3.4. “Print”
The doctor will use this option to print the medical report corresponding to the analysis chosen in the data list.
To print the results out, select the wanted analysis in the patient and exam list.
Click on Print again on the right hand side of the window, the results are automatically printed out.
To print all the patient data or patient follow up, select the patient wanted in the list and click on print.
The following message appears on the screen :
CHALLENGER_Envision_Anglais_03 158
4.3.5. “Add”
The doctor will use this option to add a new patient to the data list.
1. Click on “Add” at the bottom of the window. An empty patient file appears on the screen :
2. Fill in the fields and click on “Record”. The name of the new patient is added to the list.
CHALLENGER_Envision_Anglais_03 159
5 Quality control
This chapter will describe the steps to follow in order to verify the accuracy of the CHALLENGER
device and the reproducibility of the measurements already made.
An external phantom is supplied with the CHALLENGER. The values are known and stored in the
software. At the beginning of each day, the operator has to perform this test before beginning any
examination on patient. This chapter will detail how to make the acquisition of a phantom, then how to
perform an examination and how to analyze the image just scanned. It will show how to open the quality
control file, then how to fill in the data relating to the measurements to verify, and finally how to display the
graph relating to the measurements and the data themselves.
CHALLENGER_Envision_Anglais_03 161
5 Daily Quality Control
5.1 Opening the quality control file
Click on “Quality Control” in the main menu. This function displays the results of the tests on external
phantoms, with both a graphic and statistic display allowing the identification of any serious problem, as
well as the identification of any trend of gradual deviation from ideal results over a period of time.
Each point which appears on the graph corresponds with a quality control:
Tableau 15
CHALLENGER_Envision_Anglais_03 163
5.2 How to carry out the quality control
1. Click on “Start Control” at the bottom left hand corner of the window :
2. Remove all things laying on the bed and click on “Valid Site”.
Place the phantom correctly on the sticker
Click on start examination to perform the examination
CHALLENGER_Envision_Anglais_03 164
Place the phantom on the mask as follows:
CHALLENGER_Envision_Anglais_03 165
The following window appears on screen. Click on validate to start the acquisition.
CHALLENGER_Envision_Anglais_03 166
The phantom Examination is now finish
Click on “Validate” in order to accept the results and quit. Results are recorded when switching off the
window and will appear on the graph.
The following window appears on the screen, the last measured point appears on the graphic
CHALLENGER_Envision_Anglais_03 167
Click on “Close” in order to cancel and quit. The following message appears on the screen :
Rq: If any problem occurs during the verification, the following message appears on the screen and let choice for the
user to restart or not the quality control:
CHALLENGER_Envision_Anglais_03 168
5.3 How to visualize previous controls
1. On the graph of the quality control window, select the wanted control by clicking on the point ; the point
becomes red and the date appears as follows :
3. The Quality Control window corresponding with the selected exam is displayed :
CHALLENGER_Envision_Anglais_03 169
5.4 Viewing the Quality Control Results
CHALLENGER_Envision_Anglais_03 170
The “Control of” field selects the Region of Interest whose measurements are going to be analyzed.
Precise the dates between which the measurements will be checked, to allow the user to take a closer
look at a specific time period :
!The “Start Date” will always display by default the date of the first examination, and the “End Date” will always
correspond to the date of the most recent examination.
- In the “Start Data” part, enter the starting date of the measurements
- In the “End Data” part, enter the closing date of the measurements
The graph representing the measurements made between the dates selected is displayed.
CHALLENGER_Envision_Anglais_03 171
graph parameters :
2. The doctor can refine the graph characteristics by ticking the following parameters :
CURVES FUNCTIONS
“Display Average” Data average curve
“Display Variation margin” Variation margin curve at the control time
“Display Standard deviation” Curve of the positive and negative standard deviation at the
control time
Display markers To display the building points of the curve
Display lines To display lines that link building points together
Display limits To display allowed limits for BMD
Display time axis To display exams month per month or per exams number
Y Axis Auto range To adjust automatically the range of the Y axis
Tableau 17
CHALLENGER_Envision_Anglais_03 172
Markers
Average
Variation Margin
CHALLENGER_Envision_Anglais_03 173
Standard Deviation
Line
Limits
Time Axis
CHALLENGER_Envision_Anglais_03 174
Displaying the graph and the measurements selected
1. After validating the quality control and selecting eventually the graph parameters, the following window
appears on the screen, displaying the graph with the last control in addition (Compare with the Figure 225
- Quality Control Window p.163) :
The acceptable values are situated in the area delimited with the red dotted lines.
In the event of a technical problem being identified (i.e. several marks being outside of the area…, please contact
D.M.S. for assistance)
CHALLENGER_Envision_Anglais_03 175
5.5 How to export a file
It is possible for the user to export the quality control data in CSV (Text format with commas fields separator) or
DICOM format (In process).
2. Select CSV button to save data in a CSV file or cancel to quit. DICOM is available only if the DICOM
parameters are well configured.
4. Enter the file name and click on “save” to record the data.
CHALLENGER_Envision_Anglais_03 176
5.6 Printing the graph and the results out
1. To print the screen out, select “Print History” at the bottom of the window
3. The doctor might want to print the general report of the quality control. It is possible for him to print:
- a general report (displays the report of the current control with measurements and graph for each Region
Of Interest )
- a detailed report (Historic report of all performed controls)
- a GRIO report
CHALLENGER_Envision_Anglais_03 177
5. Click on “Detailed”. The following window is displayed :
Total
Click on “Close” in the bottom right hand corner of the quality control window.
CHALLENGER_Envision_Anglais_03 178
6 Options
The doctor will use this option to get more information on the composition of the software. Moreover this part will
enable him to consult, modify or update some functioning parameters settled by default.
6 OPTIONS....................................................................................................................................181
6.1 DATABASE..............................................................................................................................................182
6.2 ARCHIVAGE............................................................................................................................................183
6.3 Opening the “Archive scans” section...............................................................................................185
6.4 How to select an exam to archive.....................................................................................................186
6.5 How to archive an exam...................................................................................................................190
6.6 How to view or restore an archived exam........................................................................................197
6.7 SCAN......................................................................................................................................................200
6.8 LANGUAGE............................................................................................................................................202
6.9 CALCULATION........................................................................................................................................204
6.10 PRINT.....................................................................................................................................................205
6.11 INSTITUTE INFORMATION.......................................................................................................................206
6.12 WARMUP AND QUALITY CONTROL.......................................................................................................207
6.13 E-MAIL AND DICOM............................................................................................................................209
6.14 CONTACT...............................................................................................................................................210
6.15 COMMENTS............................................................................................................................................211
6.16 MAINTENANCE.......................................................................................................................................212
CHALLENGER_Envision_Anglais_03 179
6 Options
1. Click on “Software configuration” in the main menu. The following window appears on the screen :
CHALLENGER_Envision_Anglais_03 181
6.1 Database
1. Select “Database” in the list on the left of the screen in order to choose the database type :
Specify if the software must display only the Challenger envision exams on “Database Management” or
all exams made on different device.
Enable or disable the automatic exams display on “Database Management”.
CHALLENGER_Envision_Anglais_03 182
6.2 Archivage
CHALLENGER_Envision_Anglais_03 183
3. Tick off the wanted parameters to make archivage.
CHALLENGER_Envision_Anglais_03 184
In order to perform the archiving a CD/DVD Writer and NERO software (minimal version:
5.5.10) running with Windows 2000 or XP are necessary.
1. From the Archivage menu, click on the “Launch archivage” button to start archivage.
2. The following window appears on the screen, displaying the patients list and performed exams :
XXXXXXXX
XXXXXXXX
XXXXXXX
XXXXXXXX
XXXXXXXX
CHALLENGER_Envision_Anglais_03 185
6.2.2 How to select an exam to archive
1. The “Archive Scans” section lets the operator choose which patient’s exam you will copy. Click on the
“Selection” tab on the right hand side of the window :
2. The user has the choice to select the patient name or exam to archive among three different criteria :
-Database
-Patient Data
-Exam data
CHALLENGER_Envision_Anglais_03 186
Click in the unrolling menu and select the wanted database if several. Then click on “launch Search” at the
bottom of the window. The list corresponding to the selected database appears on the screen :
Click in the field, write and select the item characterizing the patient entirely or partially (last Name and/or
First Name and/or Birth date and/or sex and/or ethnic group). Click on “Launch Archive” at the bottom of the
window. The list corresponding to the selected criteria appears on the screen :
XXXXXXXXXX
CHALLENGER_Envision_Anglais_03 187
Click in the unrolling field and select the exam site and/or the exam date and/or the operator’s name and/or
the device of importation. Click on “launch Archive”, the list corresponding to the selected criteria appears
on the screen :
XXXXXXXXXX
XXXXXXXXXX
XXXXXXXXX
XXXXXXXXXX
XXXXXXXXXX
Figure 256 - Selection with a patient data Figure 257 - List corresponding to the criteria
Rq: in the event of an exam date selection, the operator may search an exam performed at a precised date, all exams
performed before or after one date, or all exams performed at a specific time period (Start date, Closing date):
CHALLENGER_Envision_Anglais_03 188
3. Then click in the list the patient or exam to archive as follows (Click on Ctrl to select separate exams, in
that case spine and left forearm exams are selected). Statistics part automatically appears on the right
hand side of the screen:
XXXXXXX
XXXXXXX
XXXXXX
XXXXXXX
XXXXXXX
4. Selected exams are referred in the statistics part displaying exams per site (in %) according to the
selection criteria (Query column), exams per site (in %) selected in the list on the left hand side of the
window and the place occupied on the hard disc (different regarding the media) :
CHALLENGER_Envision_Anglais_03 189
6.2.3 How to archive an exam
1. After selection (see chapter above), click on the “Launch Archive” button at the bottom of the screen.
The following message appears on the screen :
2. Then the following message appears on the screen, indicating the number of media needed to perform
the archiving (estimated according to the inserted media size) :
XXXXXXXX
XXXXXXXXX
XXXXXXXX
CHALLENGER_Envision_Anglais_03 190
3. Click on OK. If necessary change the description of the new media established by default :
4. Click on “validate”. The archive begins, the following messages are displayed :
XXXXXXXX
XXXXXXXX
XXXXXXXX
CHALLENGER_Envision_Anglais_03 191
XXXXXXXX
XXXXXXXX
XXXXXXXX
5. As mentioned on the screen, DO NOT REMOVE THE MEDIA : the software is going to perform a data
verification :
CHALLENGER_Envision_Anglais_03 192
XXXXXXXX
XXXXXXXX
XXXXXXXX
CHALLENGER_Envision_Anglais_03 193
6. The archives verification begins :
XXXXXXXX
XXXXXXXX
XXXXXXXX
7. At the end of the verification, the following message appears on the screen and the media is ejected :
CHALLENGER_Envision_Anglais_03 194
XXXXXXXX
XXXXXXXX
8. To finish archiving the selected exams, the operator should remove and label the media with the
reference corresponding to the archiving just performed; later it will give the operator the information
required to easily retrieve a patient’s exam for viewing :
XXXXXXXX
XXXXXXXX
XXXXXXXX
Make sure to carefully hand-label your media all the more reason if the operator uses several
database media!
A06PC300_3_1_26/01/2006
CHALLENGER_Envision_Anglais_03 195
9. Click on OK. The exam is archived and the following window appears on the screen:
XXXXXXXX
XXXXXXXX
XXXXXXXX
XXXXXXXXXX
XXXXXXXXXX
CHALLENGER_Envision_Anglais_03 196
6.2.4 How to view or restore an archived exam
Even though an exam has been archived, the list remains on the hard disk and you can view and reanalyze it if
necessary:
“Restore” lets you restore exams stored on archive media to the hard disk. As you restore the exams to the hard
disk, they are deleted from the archive media.
1. On the “Archive” screen select in the unrolling menu the media containing the exam to be restored to the
hard disk as follows :
2. The screen will display a list of all exams available to be restored to the hard disk. Select in the list the
exam(s) to be restored and click on “Launch Restore”. The following message appears on the screen :
CHALLENGER_Envision_Anglais_03 197
3. Click on yes. The following messages are displayed on the screen :
4. The
program prompts you to insert the archived media that contains the exam :
5. Insert the media, then click on OK. The following messages flash by on the screen :
CHALLENGER_Envision_Anglais_03 198
6. At the end of restoring, the updated list of the corresponding media is displayed on the screen.
1. Click on the “Archives” thumb on the top right hand-side of the window. Select the archive media to be
duplicated in the unrolling menu and click on “Duplicate a media”
CHALLENGER_Envision_Anglais_03 199
6.3 Scan
1. Select “Scan” in the list on the left of the screen in order to choose the default option scan
CHALLENGER_Envision_Anglais_03 200
4. Select the different ethnics to display by default
CHALLENGER_Envision_Anglais_03 201
6.4 Language
CHALLENGER_Envision_Anglais_03 202
4. Select the Measurement system – Metric or English – as follows :
CHALLENGER_Envision_Anglais_03 203
6.5 Calculation
2. Select the zone on which the Forearm calculation will be performed (Ulna, Radius or both) :
6. To memorise the modifications, click on “Validate” at the bottom left hand side of the window.
CHALLENGER_Envision_Anglais_03 204
6.6 Print
2. Tick off the wanted printing parameters. Click on “Validate”. The selected parameters will appear on the
printed report.
CHALLENGER_Envision_Anglais_03 205
6.7 Institute information
1. Click on “Institute” on the left of the screen. The following window is displayed :
3. Click on “Validate”. Those data are recorded and will appear on the printed reports.
CHALLENGER_Envision_Anglais_03 206
6.8 Warmup and Quality Control
1. Click on “Warm up and Quality Control” on the left of the screen. The following window appears on the
screen :
2. The practitioner may want to modify the interval between seasoning and quality control. Proceed as follows :
CHALLENGER_Envision_Anglais_03 207
Figure 287 - Changing Quality Control interval
CHALLENGER_Envision_Anglais_03 208
6.9 E-Mail and DICOM
1. Click on “E-Mail” on the left of the window : the following window appears on the screen :
3. Click on “Validate”.
“General”, “Push”, and “Print” are use to configured DICOM server network or DICOM system
CHALLENGER_Envision_Anglais_03 209
6.10 Contact
1. Click on “Contact” on the left of the screen. A window intended to provide DMS information
CHALLENGER_Envision_Anglais_03 210
6.11 Comments
Click on “Comments” on the left of the screen. A window intended to provide DMS information
CHALLENGER_Envision_Anglais_03 211
6.12 Maintenance
Click on “Maintenance” on the left of the window to open the file. The following window appears
asking the technician to enter his/her password :
CHALLENGER_Envision_Anglais_03 212
7 Quitting the Challenger program
7 Quitting the Challenger program
CHALLENGER_Envision_Anglais_03 213
Chapter 15 Patient counseling Information
Challenger is an Osteodensitometer, using the technology known as DEXA (Dual Energy X-ray Absorptiometry)
which itself is controlled a standard Pentium IV 1.4 GHz (minimum), 512 RAM computer running Windows XP.
The different possible examination sites are the spine, the femur and the forearm.
The X-ray source emits 3 intensities, 0.1mA, 0.4mA and 0.8mA. By assuring a count of X-rays that is not absorbed in
the same way in the air, the soft tissue and the bone, the software provides the density of the measured site.
The various technical approaches to bone mineral measurements using X-rays have the advantage to be understood in
terms of photon interactions with matter.
Dual X-ray Absorptiometry replaced the radionuclide source of Dual photon Absorptiometry previously used with an
X-ray tube, which has its origins in the 1970’s with the X-ray spectophometry.
This technique improved the performance of bone densitometers by combining higher photon flux with a smaller
diameter source. The availability of an intense, narrow beam of radiation shortened examination times, enhanced
image definition and improved precision. Another advantage of DXA is the relative stability of verification in clinical
use.
Clinicians requesting and technologists performing scans should be aware that any exposure to ionizing radiation
carries a risk.
However, studies of radiation exposure to patients from DXA scans have confirmed that dose levels are small
compared with many other radiological investigations involving ionizing radiation. With diagnostic procedures this is
always very small, and especially for bone densitometry studies where radiation levels are often so low that, for
example, the scatter dose to the operator from some pencil beam systems can be difficult to detect and quantify
accurately.
The risks involved are the induction of cancer by exposure to radiation and for women with child bearing potential,
genetic injury to future children. These risks are comparable to a risk of death in aeroplane accident, which means due
to chance, and increase in proportion to the exposure to it, that is to say to radiation.
However, it has not proved easy to quantify the risks of radiation because cancer has many other causes unrelated to
radiation.
Some risks in life comparable to having a DXA study are listed in table
CHALLENGER_Envision_Anglais_03 215
Chapter 15 Patient counseling Information
Some activities carrying a risk of death comparable to receiving an effective dose of 1µSv from a DXA
examination (data from Pochin).
Exposure to natural background radiation for 4 hours
Smoking one-tenth of a cigarette
Traveling 3 miles in an aero plane
Etc.
What is T-score?
The T-score represents the deviation between your value and the average value of a young healthy person of your sex
and ethnic. If your T-score is higher than –1, you are regarded as “normal”. If your T-score is comprised between –1
and –2.5, you have Osteopenia, i.e. low bone mass. This is not osteoporosis yet, but you should already consider a
treatment and/or changing your diet and physical activity. If your T-score is lower than –2.5, you have Osteoporosis.
What is Z-score?
The z-score represents the deviation between your value and the mean value in a population subject of the same age.
For young subjects, i.e between 20 and 40, there is no significant difference between Z-score and T-score.
For older patients however, the bone density decreases and it is normal to observe a decreasing of T-score.
Z-score is then useful to determine if the low density of an old patient is due to the normal decreasing of bone density
with age. (In this case the Z-score is close to 0), or if there is a high risk of fracture (in this case, Z-score is negative).
The key to the use of physical measurements in clinical trials is to ensure the consistency of the data for each patient.
Dual-energy X-ray Absorptiometry (DXA) measurements are the currently accepted standard for the measurement of
bone density. Indeed, it is known for its high precision, low radiation dose and long term stability of verification and
has been widely used in trials of new pharmaceuticals to prevent bone loss or treat Osteoporosis.
For DXA, the reliability of an instrument is assessed by daily measurements of phantoms, which permits to regularly
measure the coefficient of variations of the machine.
An important element of consistency is also to ensure that the same anatomical region is scanned in exactly the same
position each time the patient attends for a measurement. Following this the examination must be analyzed in the same
manner, selecting an identical region of interest.
Osteoporosis is a condition that thins and weakens your bones, making you more likely to break (fracture) bones, even
from minor knocks, bangs, or falls. People with osteoporosis most commonly experience fractures of the spine, hip or
wrist.
There are approximately 25 million people in the world affected by this « silent » disease, and most of them are not
even aware of the condition until a fracture occurs. Although osteoporosis is a significant health problem for many
people in the world, it is most common in middle-aged women. As many as one in three women, or one in five men
over the age of 50, will suffer a fracture related to osteoporosis during their lifetime.
In your teens and twenties, the “rebuilding” dominates, and your bones get stronger and stronger. In your thirties and
forties, the two activities more or less balance each other out, and your bones reach their peak strength. Peak bone
strength depends on many factors, including genetics, lifestyle (diet and exercise), medication and chronic illness.
After the age of forty, the “eroding” bone becomes dominant, and over time your bones gradually weaken.
CHALLENGER_Envision_Anglais_03 216
Chapter 15 Patient counseling Information
Normally, there is no cause for concern – remodelling is part of the natural life cycle for us all. However, if you
develop osteoporosis, “rebuilding” of the bone slows down earlier, or their “eroding” speeds up, or both. With
osteoporosis, the net result is that your bones become weaker than those of other people your own age and you have a
greater tendency to fracture do.
There are other risk factors that may lead to osteoporosis. These include a family history of osteoporosis, a small and
light body frame, smoking, and alcohol consumption, a lack of exercise, and long-term use of certain drugs such as
corticosteriods, and an early menopause.
In the 1970’s and 1980’s, special radiological tests were developed to help detect osteoporosis earlier. These tests are
called Bone Mineral Density (BMD) tests. BMD has been the most popular method for checking the healthiness of the
bone. Its measurement tells the doctor how much bone there is at a particular site on your body, referred to as your
bone density. The lower the density means the weaker your bone, and the more prone you are to having a bone
fracture. This type of measurement can be performed on various sites such as the spine, hip, arms and legs. BMD
measurements also involve x-ray radiation, although at lower levels than traditional x-rays.
In 1997, a new technology was approved by the U.S. Food and Drug Administration for testing bones. This new
method, called “Bone Sonometry” is based on ultrasound, similar to the one used by doctors to check the progress of
pregnant women and their unborn child.
BMD and Bone Sonometry are not always used on the same sites of the body, and the results of the tests are not
always the same. Each technology measures different parts of the body. Nevertheless, since osteoporosis is a
generalized disease, affecting your entire skeleton, it can be detected on various body sites. Your doctor understands
this, and knows how to interpret the results so that helped by him/her; you can make more informed decisions about
your health.
The examination length varies according to the speed selected in the patient database prior to an examination
performance.
In theory, the time attributed to an examination is:
Spine examination : 3-6min
Femur examination : 4-7 min
Forearm examination : 2-4 min
CHALLENGER_Envision_Anglais_03 217
Chapter 15 Patient counseling Information
All this information is then printed out onto a report showing the image of your scanned bone along with the results.
Your doctor knows how to interpret the numbers and the results, and can discuss with you what they mean.
Your doctor will advise you as to what best to do. If your test results indicate that your bones may be weak, he / she
will probably suggest preventive measures (like increasing his calcium or vitamin D to prevent the weakness of the
bone mass) and /or prescribe medication (like oestrogenic treatment). How long the doctor helps you to care for
yourself depends on many factors.
CHALLENGER_Envision_Anglais_03 218
Chapter 16 Trouble-shooting
16 Trouble-shooting
If any of the following problems keeps occurring, please contact your after-sales service.
1) The result of the previous verification and the result of the installation day reference verification are not close
from each other.
If the variation is only about 2 or 3 channels, and only in this case, you can accept the adjustment. But you
must be very careful that the red color is still on the second big peak. Otherwise do not accept.
2) While the auto verification, the patient lays his arm between the X-ray source and the detector
Ask the patient to put his arms under his head and do the verification again
If the image becomes deformed after an examination, this should be considered that there is a problem with the
machine, and the User should contact the authorized representative of DMS for advice.
Please note that if there is a slight displacement of the image, it is possible that the patient moved during the
examination. This should be confirmed by the patient. If the patient moved, the examination should be repeated as
many times as possible provided the recommended acceptable yearly doses are not acceded.
Ensure that the cable for power supply is connected to a socket with ~230 Volt and 50 Hz.
Confirm that the socket supplies the correct voltage required for the device.
If the machine is switched on but no x-rays are emitted while scanning confirm the connections of the serial cable
between computer and machine.
CHALLENGER_Envision_Anglais_03 219
Chapter 17 General Specifications
17 General Specifications
17.1 Declaration of Conformity to International Standards
IEC 60601-1-1
IEC 60601-1-2
IEC 60601-1-3
IEC 60601-1-4
On line with storage of measurements on a built in reference body every time an examination is performed. Statistical
control of these values.
17.4.2 Monitor
Recommendation of the following specifications of monitor:
15”or 17” SVGA (or larger)
Approved CE
1024x768 minimum
CHALLENGER_Envision_Anglais_03 220
Chapter 17 General Specifications
17.4.3 Printer
Recommendation of the following types:
Hewlett Packard DeskJet color
Epson Stylus Color
Dimensions:
During transportation (Folded) : 77 x 122 x 137 cm
During operate (Unfolded) : 197 x 122 x 137 cm
Weight:
ca. 200 kg
Dimensions:
During transportation (taken apart) : 77 x 122 x 137 cm
During operate (assembled) : 197 x 122 x 137 cm (79”x 45”x55”)
Weight:
ca. 300 kg
CHALLENGER_Envision_Anglais_03 221
Chapter 17 General Specifications
17.10Communication
CHALLENGER_Envision_Anglais_03 222
Chapter 17 General Specifications
manufactory Bosello
x ray tube model XRG 95.IT
work conditions tube 95 KVolt 1 mA max.
dimensions of x ray beam 2 mm at source spot
30mm a 40 cm on axis
17.12.2 Collimator
material : brass
height : 45 mm
hole diameter : 2 mm
patient skin collimator distance: 50 mm
patient skin focus x-ray tube distance: 145 mm
radiation exposure: <0.3Gy/h a 1mt.
3.0 Gy/h a 30 cm.
Exposure at patient skin: 700 Gy/h
kind of survey execution time (min.) Max. estimated Sv Approx. qty examinations / year
spine 5 33,3 30
femur 5 33,3 30
forearm 2 6.9 150
CHALLENGER_Envision_Anglais_03 223
Chapter 18 References / bibliography
18 References / bibliography
[1] Blake Glen M, Wahner Heinz W, Fogelman Ignac The evaluation of Osteoporosis: Dual Energy x-ray
Absorptiometry and Ultrasound in Clinical Practice - Second Edition -, Martin Dunitz LTD 1999
CHALLENGER_Envision_Anglais_03 224
Chapter 19 Glossary
19 Glossary
This chapter will explain the meaning of the elements on the analysis window and give some medical and technical
terms definitions.
L4 Vertebra
Soft tissues
Result board
(Refer to Tableau
21 p.205
CHALLENGER_Envision_Anglais_03 225
Chapter 19 Glossary
Normality curve:
situates the
patient’s results
among the
average
population for an
analysis (Refer to
13 T-score, Z-
score and
Reference Curve
p.34
CHALLENGER_Envision_Anglais_03 226
Chapter 19 Glossary
Gives information
where the cursor
is positioned as
well as on the
bone high and low
energy where the
cursor is
positioned.
(Refer to Tableau
23 p.207)
Analysis
Figure 297
results for L2,
Figure 298 – Validated spine analysis area L3 and L4
BUS: Electronic circuit used to establish communication between two other electronic devices: digital and analogical
data, voltages etc. can be transmitted.
Serial cable: Cable to connect the serial port of the computer with the machine.
Limit Switch: Microswitch, which reacts when a mechanical part of the machine reaches its initial or final position in
movement.
Fuse: Component, which has the purpose to protect electronic parts from excessive current.
CHALLENGER_Envision_Anglais_03 227
Chapter 19 Glossary
Earth: Electrical contact with the same electrical potential as earth. To prevent electrical hazard and electromagnetic
interference it is obligatory to earth the device.
Step by step motor: Electric motor able to perform precise movements controlled by the movement control.
Temporiser: Electronic device, which produces an electric contact after a prefixed time.
Monoblock: X-ray emitter composed of a control board and an x-ray tube positioned in a cooling oil bath. All parts
are contained in a housing of lead to screen the x-rays.
CHALLENGER_Envision_Anglais_03 228