Service Manual for Dameca MRI 508 Dameca
II Oct. 08
Contents 1
User Responsibility 2
Safety Precautions 3
Introduction 4
Description 5
Troubleshooting 6
Maintenance 7
Calibration 8
Spare Parts 9
Schematics 10
Technical Data 11
Earlier Revisions 12
Dameca Service Manual for Dameca MRI 508
Contents
1. Contents (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Oct. 08 1-1
Service Manual for Dameca MRI 508 Dameca
Contents (continued)
7. Maintenance (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.1 Maintenance, General (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.2 Tools and Test Equipment (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7.3 Service Kits (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
7.4 Annual Maintenance (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
7.5 Post Maintenance Check (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
7.6 Four Year Maintenance (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
7.7 Service Menu (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
7.8 Post Maintenance/Repair Check List (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
7.9 Check Scheme for Post Maintenance / Repair (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . 7-37
1-2 Jan. 10
Dameca Service Manual for Dameca MRI 508
List of Figures
Contents Page
Fig. 5-1 Pneumatic diagram, Dameca MRI 508 (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Fig. 5-2 Mains Power Diagram, Dameca MRI 508 (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Fig. 7-1 Service kit Dameca MRI 508, 3 rotameters (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Fig. 7-2 Service kit Dameca MRI 508, 5 rotameters (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Fig. 7-3 Service kit Dameca MRI 508, 4 years (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Fig. 7-4 Inlet manifolds (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Fig. 7-5 Pin index yoke for spare cylinder (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Jan. 10 1-3
Service Manual for Dameca MRI 508 Dameca
Fig. 10-1 Wiring diagram for ventilator, Old version (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . 10-2
Fig. 10-2 Wiring diagram for ventilator, New version (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . 10-3
Fig. 10-3 Wiring diagram for ventilator (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Fig. 10-4 Wiring diagram for ventilator (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Fig. 10-5 Wiring diagram for ventilator (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
Fig. 10-6 Pneumatic diagram Dameca MRI 508 (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
1-4 Oct. 08
Dameca Service Manual for Dameca MRI 508
2. User Responsibility
This anaesthesia machine has been built to instructions issued by DAMECA a/s only, and
conform to the specifications and operating must not be altered or modified in any way
procedures stated in this manual and/or without the written approval of DAMECA a/s.
accompanying documentation, labels and
notices when checked, assembled, operated, The user of this machine has sole responsibility
maintained, and serviced in accordance with for any malfunction which results from impro-
the instructions provided. per use, maintenance, repair, damage or altera-
tion by anyone else than DAMECA a/s or their
To ensure the safety of this machine it must be appointed agents.
checked and serviced to at least the minimum
standards in this manual.
Note:
A defective or suspected defective product Always give as much of the following
must under no circumstance be used. information as possible.
Worn, broken, distorted, contaminated or
missing components must be replaced immedia-
tely. Should such repair become necessary, it is O Type of equipment
recommended that a request for service advice
is made to the nearest DAMECA a/s service O Serial number
centre or agent.
O Date of purchase
This machine and any of its constituent parts
must be repaired in accordance with written O Apparent fault
Oct. 08 2-1
Service Manual for Dameca MRI 508 Dameca
2-2 Oct. 08
Dameca Service Manual for Dameca MRI 508
3. Safety Precautions
Read this manual carefully and familiarize O Do not use this machine in an environment
yourself with the machine before you start exceeding the EMC levels in EN 60601-1-2.
to operate the machine.
O External patient breathing systems used
Always check the machine for proper function with this machine must conform to EN 740,
before use. EN ISO 8835-2 or similar standard.
A WARNING indicates a situation in which O Bacterial filters used with this machine must
the user or the patient may be in danger of conform to EN 13328-1 or similar standard.
injury.
O An anaesthetic vapour delivery device
A CAUTION indicates a situation that may connected to the machine shall comply
lead to equipment damage or malfunction. with ISO 8835-4 or similar standard.
A NOTE indicates a point of particular interest O Heat and moisture exchangers used with this
for more efficient and convenient operation. machine must conform to EN ISO 9360 or
similar standard.
Oct. 08 3-1
Service Manual for Dameca MRI 508 Dameca
3-2 Oct. 08
Dameca Service Manual for Dameca MRI 508
Oct. 08 3-3
Service Manual for Dameca MRI 508 Dameca
O Use only original DAMECA hoses and O Do not leave spare cylinder valves open
accessories; other hoses and accessories when using the machine on central gas out-
may cause malfunctions. lets. Pressure from cylinders may be higher
than the pressure from central outlets, and
O Use only original DAMECA O-rings; the cylinder supply could be depleted.
other O-rings may cause problems if they
are not of the same size or material. O Disconnect the machine from mains and
medical gas supply, when not in use, to
prevent pollution.
Note:
Tubings used with this product must fulfill the EN12342 standard, and
respiration bags must fulfill the EN1820 standard.
3.3 Symbols
Protective earth.
Consult accompanying documents.
Equipotentiality.
Attention, consult accompanying
documents.
ON/OFF
3-4 Oct. 08
Dameca Service Manual for Dameca MRI 508
4. Introduction
This manual is related to ventilator O An APL (Adjustable Pressure Limiting)
software version 2.1. valve is integrated into the IBS, giving the
user a choice between manual ventilation
Dameca anaesthesia machine model Dameca or spontaneous breathing.
MRI 508 is applicable with most types of
inhalation anaesthesia. O2 and N2O, or O2 O The ventilator is microprocessor controlled,
and AIR can be administered, and anaesthetic with software installed.
vaporizer(s) is connected to a back-bar.
The gas supply can be obtained from a central The machine can be equipped with various
gas supply system. additional equipment such as:
O Active scavenging system
O Transfusion stand
The machine is equipped with: O Arm for patient monitor
O A rotameter unit (minoxyguard) giving O Pin-index reduction valves for 4 L. gas
user selectable gas mixes (O2, AIR and cylinders.
N2O) in a range of 0 - 10 L/min for each
gas. If N2O is selected as a mix to O2, the The design of the machine is very flexible
N2O percentage is limited to max.75% towards mounting of extra equipment, and
(minimum 25% O2). additional rails, clamps etc. can easily be
mounted on the machine.
O A microprocessor controlled anaesthesia
ventilator with user selectable ventilation
modes: 4.1 Intended Use
- VCV Volume Controlled Ventilation The Dameca MRI 508 anaesthesia machine
- PCV Pressure Controlled Ventilation is intended for composition of anaesthetic
- SIMV Synchronised Intermittent breathing gas and manual or automatic
Mandatory Ventilation ventilation of patients undergoing anaesthesia,
- PSV Pressure Supported Ventilation in MRI scanning rooms.
- PRVT Pressure Regulated Volume
Target (option) The Dameca MRI 508 anaesthesia machine is
intended for adult and paediatric patients
A real time clock is also included in the Qualified anaesthesia personnel should operate
ventilator. the Dameca MRI 508 anaesthesia machine only.
Before operating the machine mains electrical
O An integrated breathing system (IBS) power and pressurised gas must be connected
where Bag in Bottle and patient re-breathing to the machine.
system (incl. absorber) are integrated into
one block. Thereby there are only 2 tubes
(inspiratory and expiratory) from the IBS The machine is MRI compatible, for use in
system to the Y-piece. magnetic fields up to 1000 Gauss. *
Oct. 08 * This applies for machines marked with a yellow Gauss level label. 4-1
Service Manual for Dameca MRI 508 Dameca
4.2 Description
The rotameter unit (minoxyguard) of the two valves automatically close the connections
Dameca MRI 508 controls the freshgasflow to the absorber, in order not to loose gases.
given to the patient breathing system.
The flow is adjusted by the user according
to the measured flow of the rotameter tubes. Note:
The parts of the Dameca MRI 508
The ventilator of the Dameca MRI 508 uses
do not contain any Latex.
compressed AIR or O2 to ventilate the patient
through the Bag in Bottle system, placed in
the integrated breathing system (IBS).
The ventilator continuously measures airway
pressure and inspired/expired volume (optional). Note:
The ventilator can also measure inspired O2 by The Dameca MRI 508 contains
means of a fuel cell sensor (optional) placed on batteries that must be exchanged
the inspiratory port of the patient breathing system. by a technician at certain intervals,
as prescribed in the service manual.
The integrated breathing system (IBS) of the
Dameca MRI 508 is a semi-closed patient
breathing system with one-way valves in both
the inspiratory and expiratory limb, and an 4.3 Performance
absorber for eliminating CO2. The system is
suitable for high flow (approx. the patient minute The rotameter unit (minoxyguard) of the
volume), medium flow (approx. 1-1.5 L/min) Dameca MRI 508 is designed to give a
and low flow (approx. 0.5-1 L/min) of freshgas freshgas mixture selected by the user of:
for patient ventilation.
O 0 to 10 L/min for O2, AIR and N2O.
The IBS is a combination of a Bag in Bottle unit
and a patient re-breathing circuit. The IBS has This flow is led through the vaporizer rail,
a standing bellows in order for the user to see where anaesthetic agent can be added by
the controlled ventilation of the patient by the means of a user selectable vaporizer.
movement of the bellows. Besides, in case of By activating an oxygen by-pass valve,
a leakage, room air will not be drawn into the approx. 45 L/min of O2 will be added to
system, and the leakage can be seen immediately the freshgas flow, after the vaporizer rail.
by the position of the bellows.
The Dameca MRI 508 has an integrated gas
Integrated into the IBS is an APL valve which scavenging system which removes excess
is used for manual ventilation only. gases from an open reservoir. This system
must be connected to an AGSS power unit
The IBS uses a CO2 absorber i-SORB which with a flow of 30-40 L/min. If an external
contains 900 gr. Sodalime. The absorber is AGSS power unit is not available, the
available as a re-usable or a disposable version. Dameca MRI 508 can be equipped with an
If the absorber is disconnected from the IBS, integrated AGSS power unit (optional).
4-2 Oct. 08
Dameca Service Manual for Dameca MRI 508
Full spirometry
If the volume monitor is configured to »full
spirometry«, the volume sensor must be placed
at the Y-piece of the patient breathing system.
qusjob8.absorberkurver.tif
Oct. 08 4-3
Service Manual for Dameca MRI 508 Dameca
All these settings can be changed in the service- An MPL (Maximum Pressure Limiting) valve
menu, by a super-user or a technician. opens if the pressure of the freshgas flow to the
auxiliary freshgas outlet exceeds 90 (x100Pa).
4-4 Oct. 08
Dameca Service Manual for Dameca MRI 508
5. Description
Note:
In this manual pneumatic symbols are used in pneumatic diagrams and
some figures, in order to simplify service and troubleshooting.
CONNECTION POINT
ON/OFF VALVE
FILTER
PRESSURE GAUGE
CHECK VALVE
PROPORTIONAL VALVE
PRESSURE REDUCTION
VALVE
Oct. 08 5-1
Service Manual for Dameca MRI 508 Dameca
Oxygen
Oxygen is supplied from either the central If the user selects AIR the valve (29) is open
inlet (1). and AIR will flow to the regulator knob (30)
for AIR freshgas flow, controlled by the user.
Via filters (7) it is let to the pressure gauge (8) After that the flow enters a filter (31) and then
and then to a check valve (9). the AIR rotameter (32).
O2 enters the O2 tank (10) which has a separate The flow is let to the freshgas flow sensor (18)
check valve (11) and a regulator (12) that opens of the ventilator and then to the back bar (19),
for O2 flow to a whistle (13), if the O2 pressure and the fresh gas selector switch (20). The user
drops to app. 2.5 (x100Pa), thereby giving can select either the AUX fresh gas outlet (21)
an oxygen failure alarm. or the integrated breathing system (22).
Additionally, AIR is let to the auxiliary AIR
The O2 freshgas flow is controlled by a outlet (33).
regulator knob (14), by the user.
5-2 Oct. 08
Dameca Service Manual for Dameca MRI 508
Driving gas is used for the ventilator, patient The ventilator (62) is pneumatically powered
suction and active AGSS. The driving gas is by the driving gas through the inlet filter (4).
normally AIR, but O2 can also be used as A pressure switch (63) monitors the driving gas
driving gas. pressure, giving an alarm if the pressure is
below 2.5 bar.
AGSS Driving gas is let to the pressure regulator (64)
Driving gas is reduced to 4 bar by a pressure reducing the pressure to 3 bar for the Insp. flow
regulator (44), and is supplied to the ON/OFF control valve (65).
switch (45). When ON, the ejector (46) will
create a flow from the spill valve in the inte- This valve is controlled by the ventilator
grated breathing system (22) through the flow electronics, selecting a correct inspiratory flow
marker (47) to the AGSS outlet (48) on the according to the ventilator settings. A differen-
back of the machine. tial pressure transducer connected to the
pressure ports P1 and P2 measures the flow
For hospitals with active scavenging systems, and during inspiration the signal to the insp.
the machine can be delivered without the flow valve is continously adjusted, in order
ON/OFF switch (45) and the power unit (46). for the ventilator to give correct inspiratory
flow to the bag in bottle.
If an external patient system is used, it can be
connected to the Aux. AGSS outlet (49). A pressure transducer connected to the pressure
port P3 measures the driving gas pressure
An open reservoir (50) prevents vacuum to (Bag-in-Bottle chamber pressure).
be built-up in the AGS system.
During inspiration the Expiratory valve (66),
and the PEEP valve (67) are closed, and the
Suction unit gas powered driving gas is let to the chamber (68) of the
Driving gas is reduced to 4 bar by a pressure Bag-in-bottle system. The exhalation valve (69)
regulator (44), and is supplied to the jet suction is closed during inspiration, and open during
unit (51) through the ON/OFF switch (52). The expiration when the bellows (70) is in top and
vacuum is controlled by the regulator (53), and the pressure in the bellows is approx. 3 hPa
monitored by the vacuum gauge (54). A suction higher than the pressure from the ventilator.
jar can be connected to the suction connector (55). Excess gases are evacuated through the AGS
system.
Suction unit VAC powered During expiration the expiration valve (66) is
VAC is supplied from the central inlet (56). open, and the pressure in the chamber (68) is
It is led to the auxiliary VAC outlet (57), and released to the AGS system.
then to the ON/OFF switch (58). The vacuum
is controlled by the regulator (59) and monitored If a PEEP pressure (4 - 20 hPa) is selected, the
by the vacuum gauge (60). A suction jar can be insp. flow controller is set to a driving gas flow
connected to the suction connector (61). during expiration.
Oct. 08 5-3
Service Manual for Dameca MRI 508 Dameca
The Expiratory valve (66) and the PEEP valve expiration valve (76), the Bag/vent valve (72)
(67) are fully open, at the beginning of expira- to the bag (73).
tion. When the pressure in the patient breathing
system reaches the set PEEP level, the Expira- The inspiratory pressure is limited in the APL
tory valve (66) is closed and the PEEP valve valve (77), set by the user.
(67) now controls the PEEP pressure. Excess gases are evacuated through the AGS
system.
The ventilator airway pressure sensor P8
measures the PEEP pressure. This measured If the absorber is changed during use, the two
pressure controls the PWM signal to the PEEP absorbervalves (78) and (79) will bypass the
valve (67) keeping the PEEP pressure constant absorber connection, so that gases are not lost.
to the selected value.
The negative pressure relief valve (80) opens if
The driving gas is reduced to 3 bar in the the patient breathes spontaneous and the patient
pressure regulator (64), and led to the pilot breathing circuit is emptied for fresh gas (relief
valve (71). The pilot valve is controlled by the pressure is - 6 hPa).
ventilator electronics. During STBY the valve
is closed, and the Bag/vent valve (72) in the
integrated breathing system (22) is set to manual
Automatic Ventilation
ventilation. During automatic ventilation the Bag/vent
valve (72), controlled by the ventilator, is set
When the ventilator is set to VENT, the pilot to automatic ventilation.
valve (71) is opened and the Bag/vent valve
(72) in the integrated breathing system (22) The patient breathing system is filled with
is set to automatic ventilation. fresh gas from the fresh gas selector switch
(20), through the inspiratory limb, the
expiratory limb, and to the bellows (70).
Integrated Breathing System
Manual Ventilation When the bellows (70) is forced down during
During manual ventilation the Bag/vent valve inspiration, the gas will flow through the Bag/vent
(72), controlled by the ventilator, is set to valve (72), the absorber (74), the inspiratory
manual ventilation. valve (75), and the inspiratory limb to the patient.
The patient breathing system is filled with When the pressure in the chamber (68) is
fresh gas from the fresh gas selector switch released, the expiration gases flow through
(20), through the inspiratory limb, the the expiratory limb, expiration valve (76),
expiratory limb, and to the bag (73). the Bag/vent valve (72) to the bellows (70).
When the bag (73) is squeezed during inspiration, The inspiratory pressure is limited by the
the gas will flow through the Bag/vent valve ventilator high pressure alarm, set by the user.
(72), the absorber (74), the inspiratory valve
(75), and the inspiratory limb to the patient. Excess gases are evacuated through the
When the bag (73) is released, the expiration exhalation valve (69) to the AGS system,
gases flow through the expiratory limb, through the ventilator.
5-4 Oct. 08
Dameca Service Manual for Dameca MRI 508
Inspiratory/expiratory volumes can be measured During start-up and every 10 min. the two zero-
by a flow sensor (option), which can be placed valves (83 and 84) are activated by the ventila-
at the Y-piece (81) or the expiratory port (82). tor for app. 3 sec. in order to do zero calibration
of the P9 pressure sensor.
The flowsensor gives a pressure signal to the P9
pressure sensor of the ventilator. The ventilator The inspired O2 concentration can be measured
then calculates the actual patient flow over time by an external O2 fuel cell sensor (85), (option).
which gives a volume reading.
Oct. 08 5-5
Service Manual for Dameca MRI 508 Dameca
5-6 Oct. 08
Dameca Service Manual for Dameca MRI 508
48
61
22 46 45
58
56 59 68
60 70 49
57 67
47
55 69
54 66 77
52
44
53 50
51
72 73
25 71
26 65 76 82
24 64
63
74 78
23
8 62
4
1
7 9 79 81
13
11 80 75
12 20
11 85
28 17 32 40
10 43 21 83 84
3
7 9 16 31 39 18
33 15 38
34
37
5 14 30 36
7 9
19
35 29
qusjob42.10650-whispa-mri-pne-manual.png
Fig. 5-1 Pneumatic diagram, Dameca MRI 508
Oct. 08 5-7
Service Manual for Dameca MRI 508 Dameca
5-8 Oct. 08
Dameca Service Manual for Dameca MRI 508
Mains filter
Mains F1 Switch mode
230V or
100/115V F2 Power supply
Battery
charger
Display F5 / F6
Battery
F4
E-box
Ventilator
Oct. 08 5-9
Service Manual for Dameca MRI 508 Dameca
5.4 Ventilator
The main components of the Whispa ventilator are:
qusjob42.10659m-1000.pdf
Fig. 5-3
5-10 Oct. 08
Dameca Service Manual for Dameca MRI 508
Mains power enters the power supply box, The pneumatic manifold controls the drive gas
which contains a battery charger and a flow to and from the bag in bottle.
switch-mode power supply.
Oct. 08 5-11
Service Manual for Dameca MRI 508 Dameca
5-12 Oct. 08
Dameca Service Manual for Dameca MRI 508
6. Troubleshooting
6.1 General 6.2 Maximum Allowable Leaks
(according to ISO 5358)
Due to the construction and selection of
materials, problems are held at a minimum High pressure section 25 ml/min. (each gas)
in this high quality machine. Dose section 50 ml/min. (each gas)
Normally a carefully performed, planned Patient system 100 ml/min.
preventive maintenance is sufficient to keep
the machine in perfect working order for
many years.
Note:
If a problem should occur, this section of the
These values are absolute max.
manual can be a useful help to solve the problem.
values, normally the actual leaks
will be much less than these values,
The most common problems in anaesthesia
or no leaks at all.
machines are leaks. In order to locate a leak
there are some guidelines to be followed.
Oct. 08 6-1
Service Manual for Dameca MRI 508 Dameca
Table 6-1 Troubleshooting the O2/AIR supply for leaks in the high pres. sect.
Table 6-2 Troubleshooting the N2O supply for leaks in the high pressure section
6-2 Oct. 08
Dameca Service Manual for Dameca MRI 508
Oct. 08 6-3
Service Manual for Dameca MRI 508 Dameca
If there is a problem with one of the units, it must be exchanged. External components such
as proportional valves, flow sensors, and other external parts are available as spare parts.
Main fuses
The two main fuses are placed on the power supply box, and can be changed without
opening the box. The fuses are (Fuse 1 and 2):
Inside the ventilator box (inside the machine) is placed a black plastic box in which
the fuses for the ventilator components are placed. Open the box by lifting the lid in one end.
Caution:
Use only fuses of same type and rating.
6-4 Oct. 08
Dameca Service Manual for Dameca MRI 508
2. Change main fuses on the power supply box. 110/115V: T4A (P/N 40055-4AT)
230V: T2A (P/N 40055-2AT)
Carry out a functional check of the machine as described in the User Manual.
Oct. 08 6-5
Service Manual for Dameca MRI 508 Dameca
qusjob42.10659-40 manual-2.png
6-6 Oct. 08
Dameca Service Manual for Dameca MRI 508
Enter super user password (26-23-8) or technician password (91-34-80) in order to get access
to the service menu.
Oct. 08 6-7
Service Manual for Dameca MRI 508 Dameca
SIMV
Select »SIMV« <ETR>.
Select »Yes« <ETR> if the machine should be used with SIMV.
Select »No« <ETR> if the machine should NOT be used with SIMV.
PSV
Select »PSV« <ETR>.
Select »Yes« <ETR> if the machine should be used with PSV.
Select »No« <ETR> if the machine should NOT be used with PSV.
Note: PSV can only be used if volume sensor is set to »full spirometry«.
N2O
Select »N2O« <ETR>.
Select »Yes« <ETR> if N2O can be present in the freshgas flow from the machine.
Select »No« <ETR> if N2O is NEVER present in the freshgas flow from the machine.
Volume
Select »Set volume« <ETR> in order to change the volume setting of the ventilator.
Select »Volume sensor« <ETR> in order to change the volume measurement of the ventilator.
Set volume
Select »Set volume« <ETR>.
Select »TV« <ETR> if the ventilator volume set by the user should be »tidal volume«.
Select »MV« <ETR> if the ventilator volume set by the user should be »minute volume«.
Volume sensor
Select »Volume sensor« <ETR>.
Select »Full spirometry« <ETR> if the volume measurement should include both inspired a
nd expired volumes (volume sensor placed at the Y-piece).
Select »Only exp. volume« <ETR> if the volume measurement should only include expirered
volumes (volume sensor placed at the exp. cone of the IBS).
Select »No volume sensor« <ETR> if the machine should not have any volume measurement.
Language
Select »Language« <ETR> in order to change the language of the user interface
(not the service menu).
6-8 Oct. 08
Dameca Service Manual for Dameca MRI 508
Enter the technician password (91-34-80) in order to get access to the full service menu.
Oct. 08 6-9
Service Manual for Dameca MRI 508 Dameca
Analog calib
Select »Analog calib.« <ETR> in order to get access to the calibration menu of the ventilator.
See section 8 regarding calibration of the ventilator.
Regulator param
The regulator parameters are used for controlling the electronic valves of the ventilator.
These parameters should not be changed, but should be as shown in the menu figure.
Version
Select »Version« <ETR> in order to see the software version of the ventilator display unit
(called »MMI«) and the ventilator electronic box (called »CPU«).
An »*« is shown after »Whispa«, in the top of the menu. This setting must NOT be changed.
6-10 Oct. 08
Dameca Service Manual for Dameca MRI 508
7. Maintenance
7.1 Maintenance, General
In order to assure maximum safety and reliability
of all functions of this machine, planned preven-
tive maintenance must be carried out once a
Warning:
year as a minimum by an authorized technician. Use of standard oil or grease in this
machine could cause an explosion.
Use special grease P/N 36825 only.
Before any maintenance or repair is started the
machine must be disconnected from the mains,
and all gas pressures must be released. Be sure
to release the N2O pressure before the O2 pressure.
Caution:
Do not open the Electronic box of the
Due to safety and performance reasons, it is ventilator. Opening the box void the
highly recommended not to repair or disassemble warranty of the ventilator.
the proportional regulator. The regulator can be
repaired by DAMECA a/s only.
Important:
Always check the function of the parts being After maintenance and/or repair a
serviced. full function check should be
carried out according to the post
maintenance/repair check list.
Protect the table plate with a piece of carpet
or similar.
Oct. 08 7-1
Service Manual for Dameca MRI 508 Dameca
7-2 Oct. 08
Dameca Service Manual for Dameca MRI 508
Oct. 08 7-3
Service Manual for Dameca MRI 508 Dameca
7-4 Jan. 10
Dameca Service Manual for Dameca MRI 508
1,67
3DWLHQWEUHDWKLQJV\VWHP
)LOWHU
,%6ERG\
<RNH
%DFNEDU
$*6
5RWDPHWHU 9DOYH$ $3/YDOYH
9DOYH%
9DOYH&
,%6EDVH
qusjob42.36459-1.pdf
Jan. 10 7-5
Service Manual for Dameca MRI 508 Dameca
7-6 Jan. 10
Dameca Service Manual for Dameca MRI 508
1,67
3DWLHQWEUHDWKLQJV\VWHP
)LOWHU
,%6ERG\
<RNH
%DFNEDU
$*6
5RWDPHWHU
9DOYH$ $3/YDOYH
9DOYH%
9DOYH&
,%6EDVH
qusjob42.36459-15.pdf
Jan. 10 7-7
Service Manual for Dameca MRI 508 Dameca
5
5 4
3
4 2
3
2 O2 inlet manifold
Gas selector block
1
Inlet manifolds
9
5 7
4
3
2 6
O2 alarm
Rotameterbox
5
4 Ventilator
3
2
8
qusjob42.36459M-4 man.pdf
7-8 Jan. 10
Dameca Service Manual for Dameca MRI 508
Caution:
Before any maintenance or repair is started the machine must be disconnected
from the mains, and all gas pressures must be released. Be sure to release the
N2O pressure before the O2 pressure.
Note:
Special grease P/N 36825, 36810 or 36811 should be applied to all
new O-rings, before they are used.
2
3 Pin Pin
22
2
O2
Pin
3 Pin 2
AIR
22 Pin
N2 O
Oct. 08 7-9
Service Manual for Dameca MRI 508 Dameca
qusjob42.940fi7-6.pcx
Change gaskets: 1
O O2 gasket P/N 36610 (red).
O N2O gasket P/N 36609 (black).
7.4.4 Back-bar
qusjob42.13680-36457-1.pdf
7-10 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.11013manual.pdf
7.4.7 Ventilator
O Change filter for lower fan (6) (on the back of the machine) P/N 36715-01.
Oct. 08 7-11
Service Manual for Dameca MRI 508 Dameca
7-12 Oct. 08
Dameca Service Manual for Dameca MRI 508
1. Remove the rotameter tubes and change O-rings, filters, seals etc. (pos. 27, 28, 29, 30 and 31).
2. If the machine is installed with rotameterbox with 5 pcs. rotametertubes, additional O-rings and
seals (pos. 32, 33, 34, 35 and 36) must be changed.
Oct. 08 7-13
Service Manual for Dameca MRI 508 Dameca
N2O AIR
9 9
6 6
qusjob42.31350A-36457-4.pdf qusjob42.31350C-36457-4.pdf
O Change Filter (6), P/N 36701 and O-ring (9), P/N 36269.
7-14 Oct. 08
Dameca Service Manual for Dameca MRI 508
2
9
3
6
4
O Change Membrane (2), P/N 33005-71. O Change Membrane (2), P/N 33005-71.
O Change Valve seat (3), P/N 33005-72. O Change Valve seat (3), P/N 33005-72.
O Change Valve (4), P/N 33005-73. O Change Valve (4), P/N 33005-73.
O Change Friction Ring (5), 33005-75. O Change Friction Ring (5), P/N 33005-75.
O Change filter (6), P/N 36701 and O-ring
(9), P/N 36269.
Oct. 08 7-15
Service Manual for Dameca MRI 508 Dameca
7.6.4 O2 Alarm
qusjob42.10654-30 36457-4.png
7-16 Oct. 08
Dameca Service Manual for Dameca MRI 508
7.6.6 Ventilator
Pin 8
4
3
2
5
5-36457-4.SLDDRW
qusjob42.10987-25-36457-4.pdf
Oct. 08 7-17
Service Manual for Dameca MRI 508 Dameca
7-18 Oct. 08
Dameca Service Manual for Dameca MRI 508
The password for a super user is: The password for a technician is:
26 91
23 34
8 80
Oct. 08 7-19
Service Manual for Dameca MRI 508 Dameca
7-20 Oct. 08
Dameca Service Manual for Dameca MRI 508
Oct. 08 7-21
Service Manual for Dameca MRI 508 Dameca
7-22 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.72231.pdf
Special tool for pressure regulator (P/N 72231)
If necessary re-adjust the reduction valves and check that the pressure is stable after adjustment,
and does not rise slowly.
Oct. 08 7-23
Service Manual for Dameca MRI 508 Dameca
3UHVVXUHJDXJH
'
3UHVVXUHJDXJH
GLVFRQQHFWHGH[LVWLQJKRVH
3UHVVKHUHWR
FKHFNSUHVVXUH
&
& &
7-24 Oct. 08
Dameca Service Manual for Dameca MRI 508
Then set Change-over valve to AIR, close O2 freshgas flow and connect a flowmeter
(0-250 ml/min) between the central gas supply outlet and the O2 and AIR gas supply
hoses to the machine.
Alternatively, connect the O2 and AIR gas supply hoses to a Dameca test suitcase.
Check that the leakage for both gases is less than 25 ml/min.
Leakage:_________ml/min
Connect all gas supply hoses of the machine to the central gas outlets.
Open AIR freshgas flow until the AIR flow goes to zero.
Then set Change-over valve to N2O and connect a flowmeter (0-250 ml/min) between the
central gas supply outlet and the N2O gas supply hose to the machine. Alternatively, connect
the N2O gas supply hose to a Dameca test suitcase.
Leakage:_________ml/min.
Connect all gas supply hoses of the machine to the central gas outlets.
Open the flowmeter and adjust until the pressure is stable at 30 cm H2O.
Check that the flowmeter indicates a flow (leakage) of less than 25 ml/min.
If the machine is not equipped with an auxiliary freshgas outlet, this test should not be performed.
Oct. 08 7-25
Service Manual for Dameca MRI 508 Dameca
7.8.6 Selftest
If the ventilator says that the freshgas flow is too high or too low when the freshgas flow is
set to 5 L/min AIR or O2, the freshgas flowsensor must be re-calibrated (see section 8.4).
Set the freshgas flow to 10 L/min and connect reference flow measuring instrument to BAG
connection on the IBS and check that measured flow is 0.00 L/min.
Connect reference flow measuring instrument to the auxiliary freshgas outlet and check that
measured flow is 0.00 L/min.
If this test is not OK, the freshgas selector valve could be defective.
Set AIR freshgas flow to zero, set change-over valve to N2O and set a freshgas flow
of 5 L/min for O2 and N2O.
Set change-over valve to AIR and check that the AIR flow on the rotameter goes to zero.
Set change-over valve to N2O and check that the N2O flow again shows 5 L/min.
7-26 Oct. 08
Dameca Service Manual for Dameca MRI 508
If the machine is equipped with change-over valve for AIR and N2O, the following test must
be carried out
Set N2O freshgas flow to zero, set change-over valve to AIR and set a freshgas flow of 5 L/min
for O2 and AIR.
Set change-over valve to N2O and check that the N2O flow on the rotameter goes to zero.
Set change-over valve to AIR and check that the AIR flow again shows 5 L/min.
7.8.11 O2 alarm
Set O2 freshgas flow to 5 L/min and disconnect the O2 gas supply to the machine.
Check that the »O2 alarm (whistle)« alarm starts when the pressure on the O2 pressure
gauge on front of the machine is app. 2.5 (x100 kPa).
PS: O2 supply hose must be equipped with a male-connector with one-way valve, in order for
this test to be carried out.
Disconnect the drivegas (AIR or O2) supply hose to the machine and check that the »drivegas low«
alarm starts when the pressure of the drivegas is 2.0 - 2.8 (x100 kPa), by reading the pressure gauge
on front of the machine.
Oct. 08 7-27
Service Manual for Dameca MRI 508 Dameca
Connect the drivegas (AIR or O2) supply hose to the machine to the central gas outlet and check
that the ventilator starts to function again, and that all alarms disappear.
If this test indicates a malfunction, the drivegas alarm must be re-calibrated (see section 8.7).
7.8.13 Fans
Check that lower fan is running.
7.8.14 AGSS
Activate the AGSS flow and check that the yellow ball in the flowmarker is visble.
7.8.15 O2 flush
Activate the auxiliary freshgas outlet and connect a reference flow measuring instrument to
the outlet.
If the machine is not equipped with an auxiliary freshgas outlet, connect the reference flow
measuring instrument to the insp. cone on the IBS, set APL valve to max. and block the cone
for the respiration bag on the IBS.
Connect an oxygen analyzer and check that the flow is 100 % O2.
Activate the auxiliary freshgas outlet and connect a reference pressure measuring instrument
(0-100 cm H2O) to the outlet.
Set O2 freshgas flow to 2 L/min and check that the pressure is 90-100 cm H2O when the outlet is
blocked.
If not, the MPL valve placed on the auxiliary freshgas outlet block (facing the inside of
the machine) must be adjusted.
Set the freshgas switch to »circle system« and close the freshgas flow.
7-28 Oct. 08
Dameca Service Manual for Dameca MRI 508
7.8.17 Minoxyguard
Activate the auxiliary freshgas outlet and connect a reference instrument for measurement of
O2 % to the outlet.
If the machine is not equipped with an auxiliary freshgas outlet, connect the reference instrument
to the insp. cone on the IBS, set APL valve to max. and block the cone for the respiration bag
on the IBS.
Set change-over valve (if present) to N2O and set N2O freshgas flow regulator knob to max.
Set O2 freshgas flow to 0.5 L/min and wait for stable O2 % reading on the reference instrument.
The O2 % must be 23 - 29. If not re-adjust the minoxyguard as described in section 8.8.
Set O2 freshgas flow to 2.0 L/min and wait for stable O2 % reading on the reference instrument.
The O2 % must be 22 - 28. If not re-adjust the minoxyguard as described in section 8.8.
Set O2 freshgas flow to 4.0 L/min and wait for stable O2 % reading on the reference instrument.
The O2 % must be 24 - 30. If not re-adjust the minoxyguard as described in section 8.8.
7.8.18 Rotameters
Check that all freshgas flows for O2, AIR and N2O can be adjusted from 0 to 10 L/min, and
that the rotors of the rotameters rotate freely, from 0 to 10 L/min.
If not, and the patient suction is powered by AIR or O2, the suction unit placed under the stand
of the machine could be defective.
Oct. 08 7-29
Service Manual for Dameca MRI 508 Dameca
If the machine is equipped with an auxiliary O2 flowmeter, the following test must be carried out.
Connect a reference flow measuring instrument to the outlet of the auxiliary O2 flowmeter.
Set the flowmeter to 5 L/min O2 (reading is middle of the ball) and check that the measured
flow is 4.3 to 5.7 L/min.
Connect an oxygen analyzer and check that the flow is 100 % O2.
7.8.21 IBS
Place a testlung (f.ex. a respiration bag) at the Y-piece.
Start the ventilator with default settings, fill bellows by pressing the O2 flush and set
freshgas flow to 0.3 L/min.
Disconnect the CO2 absorber from the IBS and check that both absorber valves close.
Check that the bellows still goes to the top of the bag in bottle chamber, during expiration.
Connect the CO2 absorber to the IBS.
Connect a testlung (for example a respiration bag) to the Y-piece and start the ventilator with
default settings, incl. tidal volume 500 ml.
Set high alarm to 80 and fill the bellows by pressing the O2 flush.
Set respiration rate to 12, wait 3 respirations and check that the measured TV exp shown on the
display is the same as before (within +/- 30 ml).
TV exp after:_________ml
If not, the zero valves for the volume measurement might be leaking.
7-30 Oct. 08
Dameca Service Manual for Dameca MRI 508
Connect a reference volume measuring instrument to the INSP cone on the IBS.
If the drivegas for the ventilator is AIR, the volume measuring instrument must be set to AIR.
If the drivegas for the ventilator is O2, the volume measuring instrument must be set to O2.
If a Mallinckrodt PTS 2000 is used, the trigger flow for volume measurement should be set to 5 L/min.
Remove the bellows in the bag in bottle and mount the chamber.
Start the ventilator with default settings and set freshgas flow to zero.
Set ventilator to STBY and mount the bellows in the bag in bottle.
Place a testlung (NOT a respiration bag) at the Y-piece. Place a PAED. volume sensor in the patient
system. Remember to place a filter between the testlung and the volume sensor.
Go to main menu and to sub-menu »spirometry« and set sensor to PAED. <ETR>.
Start the ventilator with default values and respiration rate set to 10.
Oct. 08 7-31
Service Manual for Dameca MRI 508 Dameca
Fill the bellows by pressing the O2 flush and set freshgas flow to 10 L/min O2.
Connect an O2 sensor to exp. cone on IBS and wait until O2% is higher than 95.
Close freshgas flow, wait 3 respirations and check that O2% is still higher than 95.
Set tidal volume to 200 ml, wait 3 respirations and read the measured expired tidal volume
on the ventilator:_________ml
on the instrument:_________ml
If not, the Paed. volume sensor must be re-calibrated (see section 8.3.2).
Go to main menu and to sub-menu "spirometry" and set sensor to ADULT <ETR>.
Start the ventilator with default values and respiration rate set to 10.
Set high alarm to 80.
Fill the bellows by pressing the O2 flush and set freshgas flow to zero.
Check that O2% at exp. cone on IBS is still higher than 95. If not, open O2 freshgas flow
and after a while, close the flow and check that O2% is still higher than 95.
Set tidal volume to 500 ml, wait 3 respirations and read the measured expired tidal volume;
on the ventilator:_________ml
on the instrument:_________ml
These volumes should be identical, within +/- 50 ml.
If not, the Adult volume sensor must be re-calibrated (see section ).
Set tidal volume to 700 ml, wait 3 respirations and read the measured expired tidal volume;
on the ventilator:_________ml
on the instrument:_________ml
These volumes should be identical, within +/- 70 ml.
If not, the Adult volume sensor must be re-calibrated (see section 8.3.1).
Set ventilator to STBY and disconnect the reference volume measuring instrument.
7-32 Oct. 08
Dameca Service Manual for Dameca MRI 508
7.8.26 PEEP
Place a testlung (NOT a respiration bag) at the Y-piece. Connect a reference pressure measuring
instrument (0-100 cm H2O) to the patient system.
Start the ventilator with default settings, set respiration rate to 6 and set high alarm to 80.
Set PEEP to 5, 10 and 20 and at each setting check that the pressure measured by the reference
pressure instrument, at the end of the expiration phase, matches the set PEEP within +/- 2 cm H2O.
If not, the Pat. and BiB pressure sensors must be re-calibrated (see section 8.2).
Also, check that the airway pressure graph shown on the display indicates a stable pressure during
the expiratory phase.
Set ventilator to STBY, place the O2 sensor on the INSP cone on the IBS and set freshgas flow
to 10 L/min O2. Check that the O2 % shown on the display is 98-102.
If one of these two tests indicates an inaccuracy, the O2 sensor must be re-calibrated, as described
in the user manual (via the main menu).
7.8.28 PCV
Place a testlung (NOT a respiration bag) at the Y-piece.
Start the ventilator with default settings, set respiration rate to 10 and set ventilation mode to PCV.
Fill the bellows by pressing the O2 flush and set freshgas flow to app. 1.0 L/min.
Set Insp. Pressure to 17, wait 3 respirations and check that the measured PEAK pressure shown on
the display is 18-22 hPa.
Also, check that the airway pressure graph shown on the display indicates a stable pressure during
the inspiratory phase.
Oct. 08 7-33
Service Manual for Dameca MRI 508 Dameca
Set Insp. Pressure to 27, wait 3 respirations and check that the measured PEAK pressure shown on
the display is 28-32 hPa.
Also, check that the airway pressure graph shown on the display indicates a stable pressure during
the inspiratory phase.
Set Insp. Pressure to 47, wait 3 respirations and check that the measured PEAK pressure shown on
the display is 48-52 hPa.
Also, check that the airway pressure graph shown on the display indicates a stable pressure during
the inspiratory phase.
If this test indicates an unstable airway pressure during inspiratory phase, the PEEP valve must be
re-calibrated (see section 8.5)
Remove the testlung and check that »system pressure high« alarm start when the bellows goes to
the bottom of the chamber.
Connect the testlung, activate O2 flush until bellows is in top position, and check that both alarms
stop after a couple of respirations.
If this test is not OK, the Pat. and BiB pressure sensors must be re-calibrated (see section 8.2).
Check also, that the PEAK pressure shown on the display is 14-16 hPa.
Set high alarm to 80 and check that the alarm stops after one respiration cycle.
7-34 Oct. 08
Dameca Service Manual for Dameca MRI 508
Disconnect the mains cable to the machine from the mains outlet.
Check that »mains failure« alarm starts and that the machine functions as before.
If a »battery low« alarm starts during these 10 minutes, charge the batteries for at least 8 hours.
Oct. 08 7-35
Service Manual for Dameca MRI 508 Dameca
7-36 Oct. 08
Dameca Service Manual for Dameca MRI 508
Technician Date Machine S/N IBS S/N APL valve S/N Signature
Oct. 08 7-37
Service Manual for Dameca MRI 508 Dameca
7-38 Oct. 08
Dameca Service Manual for Dameca MRI 508
Oct. 08 7-39
Service Manual for Dameca MRI 508 Dameca
7-40 Oct. 08
Dameca Service Manual for Dameca MRI 508
Oct. 08 7-41
Service Manual for Dameca MRI 508 Dameca
7-42 Oct. 08
Dameca Service Manual for Dameca MRI 508
8. Calibration
Before calibration can be carried out, the tests If the primary drivegas to the ventilator is AIR,
described in section 7.8.1-7.8.9 must be carried the external adult patient flow sensor used by
out. the integrated volume monitor can be calibrated
at the same time.
The calibration requires a reference measuring
instrument for pressure (0-80 hPa) and flow If the primary drivegas to the ventilator is O2,
(0-80 L/min.) the external adult patient flowsensor must
NOT be calibrated at the same time.
Activate MAIN menu, select service menu and
enter the service technician password (91-34-80). Calibrate only inspiratory flow sensor:
Select Analog calib. <ETR>. In the installation menu (see section 6.7),
set »Volume sensor« to »No vol. sensor«.
When calibration of the inspiratory flow sensor
In the calibration menu you get is completed, set »Volume sensor« as desired
access to: (see section 6.7).
Insp. Flow & Adult pat. Flow Remove bellows in Bag in Bottle, and connect
BiB & Pat. Pressure inspiration port of the integrated patient system
Pat. Flow to flow measuring reference instrument.
Freshgas Flow
PEEP valve Press enter to confirm <ETR>.
Temperature calib.
Calibrate both inspiratory flow sensor and
adult patient flow sensor:
Note: Remove bellows in Bag in Bottle, and connect
inspiratory port of the integrated patient
The accuracy of the ventilator depends
breathing system to external adult patient
on the accuracy of the calibration.
flow sensor (MUST be mounted with patient
Therefore the ventilator must have
port facing flow instrument) and then to flow
room temperature, and must be switched
measuring reference instrument.Press enter
on for at least 30 min., before
to confirm <ETR>.
calibration is carried out.
Note:
When calibrating the patient flow
8.1 Insp. Flow & Adult Pat. Flow sensor, it must be placed with the
port labeled »PATIENT« facing the
Select »Insp. Flow & Adult pat. Flow« to cali- flow measuring reference instrument.
brate the sensor that measures the inspiratory
flow from the ventilator to the Bag in Bottle.
Oct. 08 8-1
Service Manual for Dameca MRI 508 Dameca
Step 1
Check that the insp. flow pressure regulator
on the pneumatic block inside the ventilator $GMXVWVFUHZRQ
LQVSLUDWRU\IORZYDOYH
is adjusted to 2.8-3.2 (x100kPa).
Note:
During calibration, the inspiratory flow qusjob42.10987-25 manual 1.pdf
8-2 Oct. 08
Dameca Service Manual for Dameca MRI 508
Step 6 Step 12
Adjust inspiratory flow from ventilator, until Adjust inspiratory flow from ventilator, until
flow reading on flow measuring instrument flow reading on flow measuring instrument
is 4.95-5.05 L/min. is 39.9-40.1 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 5.0 L/min <ETR>. Press enter to calibrate at 40.0 L/min <ETR>.
Step 7 Step 13
Adjust inspiratory flow from ventilator, until Adjust inspiratory flow from ventilator, until
flow reading on flow measuring instrument flow reading on flow measuring instrument
is 9.9-10.1 L/min. is 49.9-50.1 .0 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 10.0 L/min <ETR>. Press enter to calibrate at 50.0 L/min <ETR>.
Step 8 Step 14
Adjust inspiratory flow from ventilator, until Adjust inspiratory flow from ventilator, until
flow reading on flow measuring instrument flow reading on flow measuring instrument
is 14.9-15.1 L/min. is 59.9-60.1 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 15.0 L/min <ETR>. Press enter to calibrate at 60.0 L/min <ETR>.
Step 9 Step 15
Adjust inspiratory flow from ventilator, until Adjust inspiratory flow from ventilator, until
flow reading on flow measuring instrument flow reading on flow measuring instrument
is 19.9-20.1 L/min. is 79.8-80.2 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 20.0 L/min <ETR>. Press enter to calibrate at 80.0 L/min <ETR>.
Step 10 Step 16
Adjust inspiratory flow from ventilator, until Press enter to save calibration <ETR>.
flow reading on flow measuring instrument Press »Normal screen« to cancel the calibration.
is 24.9-25.1 L/min.
Adjust by turning the control wheel.
Press enter to calibrate at 25.0 L/min <ETR>. Note:
If the ventilator does not accept the
Step 11 calibration, the filter inside the venti-
Adjust inspiratory flow from ventilator, until lator pneumatic manifold could be
flow reading on flow measuring instrument defective.Change filter (P/N36709),
is 29.9-30.1 L/min. gaskets (2 pcs. P/N36498 and 1 pcs.
Adjust by turning the control wheel. P/N36499), O-ring (P/N36423) and
Press enter to calibrate at 30.0 L/min <ETR>. redo the calibration.
Oct. 08 8-3
Service Manual for Dameca MRI 508 Dameca
Assemble the integrated patient breathing Press »Normal screen« to cancel the calibration.
system, incl. hoses and Y-piece.
Step 3 Note:
Remove the Bag in Bottle chamber. When calibrating the patient flow
Press enter to re-calibrate at 0 hPa <ETR>. sensor, it must be placed with the
port labeled »PATIENT« facing the
Step 4 flow measuring reference instrument.
Re-mount only the Bag in Bottle chamber and
open freshgas flow until stable reading on the
8-4 Oct. 08
Dameca Service Manual for Dameca MRI 508
Step 1 Step 7
Check that reading on flow measuring Adjust flow from ventilator, until flow reading
instrument is 0.0 L/min. on flow measuring instrument is 19.9-20.1 L/min.
Press enter to calibrate at 0.0 L/min <ETR>. Adjust by turning the control wheel.
Press enter to calibrate at 20.0 L/min <ETR>.
Note: Step 8
During calibration, the patient flow Adjust flow from ventilator, until flow reading
measured by the ventilator according on flow measuring instrument is 24.9-25.1 L/min
to data from last calibration, is shown .
on the display. Adjust by turning the control wheel.
Press enter to calibrate at 25.0 L/min <ETR>.
Step 9
Step 2 Adjust flow from ventilator, until flow reading
Adjust flow from ventilator, until flow reading on flow measuring instrument is 29.9-30.1 L/min.
on flow measuring instrument is 0.95-1.05 L/min. Adjust by turning the control wheel.
Adjust by turning the control wheel. Press enter to calibrate at 30.0 L/min <ETR>.
Press enter to calibrate at 1.0 L/min <ETR>.
Step 10
Step 3 Adjust flow from ventilator, until flow reading
Adjust flow from ventilator, until flow reading on flow measuring instrument is 39.9-40.1 L/min
on flow measuring instrument is 1.95-2.05 L/min. .
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 2.0 L/min <ETR>. Press enter to calibrate at 40.0 L/min <ETR>.
Step 4 Step 11
Adjust flow from ventilator, until flow reading Adjust flow from ventilator, until flow reading
on flow measuring instrument is 4.95-5.05 L/min. on flow measuring instrument is 49.9-50.1 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 5.0 L/min <ETR>. Press enter to calibrate at 50.0 L/min <ETR>.
Step 5 Step 12
Adjust flow from ventilator, until flow reading Adjust flow from ventilator, until flow reading
on flow measuring instrument is 9.9-10.1 L/min. on flow measuring instrument is 59.9-60.1 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 10.0 L/min <ETR>. Press enter to calibrate at 60.0 L/min <ETR>.
Step 6 Step 13
Adjust flow from ventilator, until flow reading Adjust flow from ventilator, until flow reading
on flow measuring instrument is 14.9-15.1 L/min. on flow measuring instrument is 79.8-80.2 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 15.0 L/min <ETR>. Press enter to calibrate at 80.0 L/min <ETR>.
Oct. 08 8-5
Service Manual for Dameca MRI 508 Dameca
Step 14 Step 2
Press enter to save calibration <ETR>. Adjust flow from ventilator, until flow reading
Press »Normal screen« to cancel the calibration on flow measuring instrument is 0.95-1.05 L/min.
Adjust by turning the control wheel.
8.3.2 Calibrate paediatric patient Press enter to calibrate at 1.0 L/min <ETR>.
flow sensor
Select »Paed. sensor« to calibrate the Step 3
external paediatric patient flow sensor. Adjust flow from ventilator, until flow reading
on flow measuring instrument is 1.95-2.05 L/min.
Remove bellows in Bag in Bottle, and connect Adjust by turning the control wheel.
inspiratory port of the integrated patient
breathing system to the external paediatric Press enter to calibrate at 2.0 L/min <ETR>.
patient flow sensor (should be mounted with
patient port facing flow instrument) and then Step 4
to flow measuring reference instrument. Adjust flow from ventilator, until flow reading
on flow measuring instrument is 3.95-4.05 L/min.
Press enter to confirm <ETR>. Adjust by turning the control wheel.
Step 1 Step 6
Check that reading on flow measuring Adjust flow from ventilator, until flow reading
instrument is 0.0 L/min. on flow measuring instrument is 9.9-10.1 L/min.
Adjust by turning the control wheel.
Press enter to calibrate at 0.0 L/min <ETR>.
Press enter to calibrate at 10.0 L/min <ETR>.
Note: Step 7
During calibration, the patient flow Adjust flow from ventilator, until flow reading
measured by the ventilator according on flow measuring instrument is 14.9-15.1 L/min.
to data from last calibration, is shown
on the display. Adjust by turning the control wheel.
8-6 Oct. 08
Dameca Service Manual for Dameca MRI 508
Step 10 Step 3
Adjust flow from ventilator, until flow reading Close the freshgas flow and check that the refe-
on flow measuring instrument is 29.9-30.1 L/min. rence instrument reads 0.0 L/min. Press <ETR>.
Adjust by turning the control wheel.
Step 4
Press enter to calibrate at 30.0 L/min <ETR>. Adjust AIR freshgas flow until reference instru-
ment reads 10.0 L/min. AIR flow. Press <ETR>.
Step 11
Adjust flow from ventilator, until flow reading Step 5
on flow measuring instrument is 34.9-35.1 L/min. Press enter to save calibration <ETR>.
Adjust by turning the control wheel. Press »Normal screen« to cancel calibration.
Oct. 08 8-7
Service Manual for Dameca MRI 508 Dameca
During this period, the temperature of the PEEP If enter is not pressed within 20 min., the
valve coil must be measured, using an infrared calibration is automatically cancelled.
temperature sensor. If the temperature of the
coil reaches 35-40OC before the 4 min., press Step 4
enter to start next step of the calibration <ETR>. Now the ventilator checks the characteristics
of the PEEP valve.
Note:
The PEEP valve calibration can be
qusjob42.10987-25 manual 2.pdf cancelled at all times, by pressing
»Normal screen«.
Step 2
The PEEP valve temperature is now stabilised
during a 5 min. period.
8.6 Temperature Calibration
When the 5 min. period is ended, the next step
of the calibration is activated automatically. The«Temperature calib.« is for definition of
the temperature characteristics of the pressure
Step 3 sensors used in the ventilator.
The PEEP valve is now warmed up slowly, and This definition can NOT be changed in the field,
the temperature of the PEEP valve coil must be as it requires specific control of room tempera-
measured, using an infrared temperature sensor. ture for the electronic box of the ventilator.
8-8 Oct. 08
Dameca Service Manual for Dameca MRI 508
6FUHZIRUDGMVXWHPHQW
RIGULYHJDVORZ
DODUP
Oct. 08 8-9
Service Manual for Dameca MRI 508 Dameca
8.8 Minoxyguard
Set ventilator to STBY, close O2 and AIR Step 4
freshgas flows, set change-over valve to Adjust O2 freshgas flow regulator knob until the
N2O and set N2O freshgas flow regulator O2 flow is 2 L/min.
knob to max.
Step 5
Step 1 Turn the adjustment screw on the rear side of
Slowly open O2 freshgas flow regulator knob the lower minoxyguard manifold (positioned
until the O2 flow is 250 ml/min. behind the N2O freshgas flow regulator) until
the N2O flow is 6 L/min.
Step 2
Turn the adjustment screw on the rear side of Step 6
the lower minoxyguard manifold (positioned Close the O2 freshgas flow and repeat step 3.
behind the O2 freshgas flow regulator) until If a re-adjustment is required, step 4 and 5
the N2O float (rotor) just begins to move. must also be performed again.
Step 3
Close O2 freshgas flow regulator knob and
slowly open it again. Check that the N2O flow
starts when the O2 flow is 250 - 300 ml/min. If
not, re-adjust as described in section 1 and 2.
8-10 Oct. 08
Dameca Service Manual for Dameca MRI 508
9. Spare Parts
Contents Page
Dameca MRI 508 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Stand with Integrated Suction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Gas- and Electrical Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Gasbox for Dameca MRI 508 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Gasbox Assy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10
Togggle Switch Valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
Turn Switch Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
Rotameterbox with 3-tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-14
Minoxyguard with 3 tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-16
Rotameterbox with 5-tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-18
Drive Gas Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-22
Backbar Selectatec . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-24
Backbar Dräger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-26
Inlet Manifold O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-28
Inlet Manifold AIR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-30
Inlet Manifold N2O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-32
Inlet Manifold VAC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-34
Inlet Manifold AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-35
Whispa Ventilator Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-36
E-Box for Whispa ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-38
Power Supply Box, Old version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-40
Valve Block Assy. for Whispa . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-42
Rear Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-48
Flowmarker Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-49
Suction Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-50
Mains Switch, standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-52
Vacuum Suction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-54
AGSS Ejector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-56
O2 Flush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-58
IBS System Complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-59
IBS Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-60
IBS System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-62
Spare parts for DAMECA products should be ordered through our local agents.
When ordering spare parts, identify the part, and use the part number stated in this manual.
Oct. 08 9-1
Service Manual for Dameca MRI 508 Dameca
9-2 Jan. 10
Dameca Service Manual for Dameca MRI 508
6
13
4
3
8
7
15
4
2
21
5 25
20
19
22 24
12
18
10
23
14
1 11
16
17
qusjob42.10651M-00 manual-1.pdf
Jan. 10 9-3
Service Manual for Dameca MRI 508 Dameca
qusjob42.10651M-01 manual.png
9-4 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.10657MS manual.pdf
Oct. 08 9-5
Service Manual for Dameca MRI 508 Dameca
9-6 Oct. 08
Dameca Service Manual for Dameca MRI 508
Oct. 08 9-7
Service Manual for Dameca MRI 508 Dameca
Gasbox Assy.
Item Qty Description Part No.
1 1 Fordampergasbox Dameca MRI 508, complete 10653M-20109
2 1 Back Bar selectatec compatible (optional) 11080-650
3 1 Cone 30 mm female 13952
4 1 Nipple VAC 31094b
5 1 Mini Coupling O2, white 31081
6 1 Pneumatic control panel for Siesta i Whispa MRI 10653M-204
7 18 Spacer bushing. M4, l: 30mm 40098-77
8 1 Equipotential connector 40055-01
9 1 Potential equalising label (green/yellow) 11360
10 4 Nut DIN 6923 A2 70424
11 2 Screw RHR M4 X 8MM A2 70046-0408
12 2 Screw DIN 912 M4X12 70009
13 2 Nylon washer 70467
14 1 Negative pressure valve 10653-150
15 2 Ball chain end 36871-02
16 1 Ball chain 36871-01
17 2 Screw RHR M3 X 8MM A2 70046-0308
18 1 Plug for AGSS 13951-11
19 1 Code ring O2 (white) 11365
20 1 Code ring (blue) 11365A
21 1 Code ring (black/white) 11365C
22 1 Strain relief 40061-63
23 2 Nut M3 DIN 934 70447
24 4 Screw M5x12 70414-0512
25 1 Reservoir assembly for Siesta i 10653M-60
26 2 Spacer bush 10653-63
27 2 Nut M5 DIN 934 70441
28 1 Snap bushing 40058-26
29 1 Fresh Gas Switch Valve 10653-220
30 1 Vacuum Switch Valve 10653-230
31 1 Rear cover for Dameca MRI 508 10653M-021
32 3 Quick connector angled, ø6/M5 36690-22
33 3 Mounting ring for ø40 gauges 10653-52
34 3 Gauge 0-10 bar, plastic housing 34108
35 3 Mounting plate (SS) for ø40 gauge 10653M-53
36 6 Screw DIN 912 M5X10 70013
37 1 Pin-index pressure regulator N2O (optional) 33052-50
38 1 Pin-index pressure regulator O2 (optional) 33053
9-8 May 09
Dameca Service Manual for Dameca MRI 508
qusjob42.10653MS manual.pdf
May 09 9-9
Service Manual for Dameca MRI 508 Dameca
Control Panel
Item Qty Description Part No.
1 1 Front plate for control panel, SS 10653M-2041
2 1 O2 flow meter 0-10 L/min (optional) 32022
3 1 Gauge 1-0 bar, plastic housing 34109
4 1 Regulator forsuction unit (optional) 10995-155
5 1 Foil for flow meter (optional) 10653-06
6 1 Foil for suction unit (optional) 10653-107
7 1 Blind foil for short model 10653-09
8 1 Foil for outlet selector valve (optional) 10653-11
9 1 Acrylic plate for vacuum gauge 10653-12
10 2 Screw DIN 912 M5X8 70017-0508
11 2 Screw DIN 912 M4X8 70007
12 1 Clamp for flowmeter 10653M-05
13 2 Support for flowmeter, MRI 10653M-055
14 16 Nut DIN 6923 A2 70424
15 1 Knob for O2 flow meter (optional) 10653-044
16 1 Allen screw 70107
17 1 Quick connector angled, ø6/1/8" 36690-20
18 1 Label O2 White 11336
19 1 Toggle switch valve 11951
20 1 Turn switch valve 11982
21 1 Quick connector angled, ø6/M5 36690-22
22 1 T-connector ø6/1/8"RG 36690-60
23 1 Lynkobling, T-stykke, Ø6 36690-18
24 1 Silencer 10653-208
25 2 Spacer bushing. l: 28mm 10653-78
26 1 Mounting plate (SS) for ø40 gauge 10653M-53
27 1 Spacer for vacuum gauge, MRI 10653M-043
9-10 Oct. 08
Dameca Service Manual for Dameca MRI 508
*
qusjob42.10653M-204S manual.pdf
Oct. 08 9-11
Service Manual for Dameca MRI 508 Dameca
qusjob42.11951.png
9-12 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.11982.png
Oct. 08 9-13
Service Manual for Dameca MRI 508 Dameca
9-14 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob4210654-80 manual.png
Oct. 08 9-15
Service Manual for Dameca MRI 508 Dameca
9-16 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob4210654-75 manual.pdf
Oct. 08 9-17
Service Manual for Dameca MRI 508 Dameca
9-18 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.10774-750 manual.pdf
Oct. 08 9-19
Service Manual for Dameca MRI 508 Dameca
9-20 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.10774-750 manual.pdf
Oct. 08 9-21
Service Manual for Dameca MRI 508 Dameca
9-22 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.10659-40 manual.png
Oct. 08 9-23
Service Manual for Dameca MRI 508 Dameca
Backbar Selectatec
Item Qty Description Part No.
1 1 Selectatec back bar block 11080-651
2 4 Valve 11080-70
3 2 Cover plate 11080-66
4 2 Code pin 11080-53
5 2 Spring 11080-59
6 4 Philips screw M2.5x6mm 70273-256
7 2 Allen point screw 70135-06
8 1 Label 11338
9 2 Quick connector angled 36690-20
10 1 Quick connector T-piece 36690-18
11 1 Tube connector 36690-60
12 4 Philips screw M3x8mm 70276
13 4 Screw 70056
9-24 Oct. 08
Dameca Service Manual for Dameca MRI 508
12
4 7
2
8
13
10
qusjob42.11080-650s manual.pdf
Oct. 08 9-25
Service Manual for Dameca MRI 508 Dameca
Backbar Dräger
Item Qty Description Part No.
1 1 Back bar block Dräger 11060-651
2 4 Screw 70056
3 2 Quick connector angled 36690-20
4 2 Screw DIN 912 M5X30 70017
5 2 Allen point screw 70135-06
6 4 Valve 11080-70
7 2 Lock 11066-62
8 4 Screw "70273-02,505"
9 1 Tower for back bar 11060-653
10 2 Pin 11060-654
11 1 Arm 11060-652
12 2 Screw DIN 912 M4X12 70009
13 2 Cover cap 36653-09
14 1 Spring 11060-659
15 1 Ball ø4 mm 30122-08
16 1 Lock pin 11060-658
17 1 Handle 11060-655
18 1 Label 11338
19 1 Quick connector T-piece 36690-18
20 1 Tube connector 36690-60
9-26 Oct. 08
Dameca Service Manual for Dameca MRI 508
13
12
16
11
17
14
9
10 15
18
2
5
4
3
Use for 10651-00 (Rotameter) only
20
Use forl 10650-00 (TS) only
19
qusjob42.11060-650s manual.pdf
Oct. 08 9-27
Service Manual for Dameca MRI 508 Dameca
Inlet Manifold O2
Item Qty Description Part No.
1 1 Inlet manifold O2 31350-05
2 1 Filter nut 31350-06
3 1 Nut for check valve 31350-07
4 1 Filter housing 11213
5 1 Valve housing w/filter w/non-return valve 11215
6 1 Check valve 11214
7 1 Zeeger circlip ring 36735-09
8 3 Quick connector 31405
9 5 Quick connector angled 36690-20
10 1 Label O2 White 11336
11 1 Retainer nut 31381-03
12 1 Hose connector NIST O2 31381-02
13 1 Spring 33005-74
14 1 Valve seat 33005-72
15 1 Membrane, black 33005-71
16 1 Friction ring 33005-75
17 1 Spring 33005-76
18 1 Nut 10967-29
19 1 Adjustment screw 10967-291
20 1 Nut 36083
21 1 Valve 33005-73
22 1 O-ring Ø16,1x1,6 36256
23 3 Allen point screw 70135-05
24 1 O-ring 36305
25 1 O-ring Ø9,1x1,6 36271
26 1 Filter bronze, Ø8x2mm 36701
27 1 O-ring Ø5,1x1,6 36269
28 1 Tube connector 36690-60
29 1 Quick connector T-piece 36690-18
9-28 Oct. 08
Dameca Service Manual for Dameca MRI 508
19
20
18
8
11
12
17
Loctite 270
24 16
7 15
29
4 14
8
28 23
21
Filter facing inwards
13 Loctite 270
Loctite 270
27
9 Loctite 270
26
8
10
6
22
2
1 5
25
9 3
qusjob42.31350 manual.pdf
Oct. 08 9-29
Service Manual for Dameca MRI 508 Dameca
9-30 Oct. 08
Dameca Service Manual for Dameca MRI 508
10
13
Loctite 270
4 1 6
12
2
Loctite 270
6
3
5
11
15 8
14
16
17
qusjob42.31350C manual.pdf
Oct. 08 9-31
Service Manual for Dameca MRI 508 Dameca
9-32 Oct. 08
Dameca Service Manual for Dameca MRI 508
10
11 Loctite 270 18
19
13
17
5
16
2
8
15
14
7
1 4
12
9
3
qusjob42.31350A manual.pdf
Oct. 08 9-33
Service Manual for Dameca MRI 508 Dameca
1 2
qusjob42.31350B manual.pdf
9-34 Oct. 08
Dameca Service Manual for Dameca MRI 508
5
3
qusjob42.31350F manual.pdf
Oct. 08 9-35
Service Manual for Dameca MRI 508 Dameca
9-36 Oct. 08
Dameca Service Manual for Dameca MRI 508
0RXQWLQJSODWH 0
6SDFHU3ODWH
qusjob42.10659M-100 manual.pdf
Fig. 9-20 Whispa ventilator
Oct. 08 9-37
Service Manual for Dameca MRI 508 Dameca
9-38 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.10987M-50S manual.pdf
Fig. 9-21 E-Box for whispa Ventilator
Oct. 08 9-39
Service Manual for Dameca MRI 508 Dameca
9-40 Jan. 10
Dameca Service Manual for Dameca MRI 508
7
6
19 2
8 23 20 10 3
5
24 22
13
25 15
18 14 17
21
9
11
1 26
12
qusjob42.40098-30 manual.pdf
Jan. 10 9-41
Service Manual for Dameca MRI 508 Dameca
9-42 Oct. 08
Dameca Service Manual for Dameca MRI 508
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
7RUTXH1P/RFWLWH
qusjob42.10987M-25S
Fig. 9-23 Valve block assy. for Whispa
Oct. 08 9-43
Service Manual for Dameca MRI 508 Dameca
9-44 Oct. 08
Dameca Service Manual for Dameca MRI 508
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
7RUTXH1P/RFWLWH
qusjob42.10987M-25S
Oct. 08 9-45
Service Manual for Dameca MRI 508 Dameca
9-46 Oct. 08
Dameca Service Manual for Dameca MRI 508
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH /RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
/RFWLWH
7RUTXH1P
qusjob42.10987M-25 .pdf
Oct. 08 9-47
Service Manual for Dameca MRI 508 Dameca
Rear Cover
Item Qty Description Part No.
1 1 Rear cover for Dameca MRI 508 10651M-13
2 1 Fan filter complete 36715
3 4 Screw 70058
4 4 Toothed lock washer ø4 70476
5 4 Nut M4 DIN 934 70440
qusjob42.10651M-12 manual.pdf
9-48 Oct. 08
Dameca Service Manual for Dameca MRI 508
Flowmarker Assembly
Item Qty Description Part No.
1 1 Housing for flow marker 13680-01
2 1 Holder for flow marker 13680-02
3 1 Flowmarker 25 L/min 13676
4 1 Piston for valve, green 30325-10
5 1 Piston for valve, red 11014-11
6 2 Quick connector 36690-12
7 2 Nipple for hose 10894-62
8 3 O-ring ø7,1x1,6 36268
9 1 O-ring ø12,29x3,53 36247
10 1 O-ring ø10,69x3,53 36246
11 2 O-ring ø17,17x1,78 36248
Loctite 242
5 8 6 7 1 8 8 4
Loctite 242
3
6
Loctite 242
10
11
11
qusjob42.13680 manual.pdf
Oct. 08 9-49
Service Manual for Dameca MRI 508 Dameca
Suction Unit
Item Qty Description Part No.
1 1 Body for integrated suction 32525-03
2 1 Diaphrame house 32525-02
3 1 Diaphrame for suction, red 32515-13
4 1 O-ring 36288
5 1 House for valve 32515-12
6 1 Bottum nozzle 32515-21
7 1 O-ring Ø13,1x1,6 36272
8 1 O-ring 36281
9 1 Top nozzle 32515-17
10 1 O-ring Ø9,1x1,6 36271
11 1 Middle nozzle 32515-19
12 1 Bushing for silencer 32515-22
13 1 House for silencer 32525-04
14 2 Silencer 32010-22
15 2 Screw DIN 912 M5X8 70017-0508
16 1 O-ring Ø14,1x1,6 36307
17 1 Diaphram, yellow 32515-14
18 2 Quick connector angled, ø6/M5 36690-22
19 1 Quick connector angled, ø6/1/8 36690-20
20 2 Washer 70468
9-50 Oct. 08
Dameca Service Manual for Dameca MRI 508
19
2
18
17
16
1
20
15
8
9
10
11
7
6
12
14
14
13
qusjob42.32530-100 manual.pdf
Oct. 08 9-51
Service Manual for Dameca MRI 508 Dameca
9-52 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.40090-70 manual.png
Oct. 08 9-53
Service Manual for Dameca MRI 508 Dameca
Vacuum Suction
Item Qty Description Part No.
1 1 House for valve, front part 10653-41
2 1 Housing for vacuum suction 10653-72
3 1 Screw for regulation 10653-43
4 1 Vacuum diaphragm 32550-08
5 1 Cone for diaphragm 32550-07
6 1 Shaft for valve 10653-73
7 1 O-ring 36294
8 1 Zeeger ring 36738
9 1 Spring 32550-15
10 1 Nylon gasket 32550-10
11 1 Button for vacuum 10653-48
12 1 Shaft for valve 10653-44
13 1 Allen screw 70101
14 2 Washer 70467
15 1 Nut M3 DIN 934 70439
16 1 Label VAC, red 11336B
17 1 Label VAC, yellow 11336B-S
18 1 Label Yellow 11361-04
19 1 Nut 10653-47
20 1 Quick connect Ø8-1/8" Viton 36690-14
21 1 Quick connector angled, ø6/M5 36690-22
22 1 Quick connector 90 deg. angel, Ø8x1/8" 36690-21
9-54 Oct. 08
Dameca Service Manual for Dameca MRI 508
qusjob42.10653-70s.png
Oct. 08 9-55
Service Manual for Dameca MRI 508 Dameca
AGSS Ejector
Item Qty Description Part No.
1 1 Ejector body 11014-15
2 1 Nozzle 0.6 mm 11014-60
3 1 Filter bronze 36132
4 1 Manifold for ejector 11013-01
5 2 Screw DIN 912 M3X20 70001-0320
6 1 Hose nipple w/1/2-26 10894-63
7 1 O-ring ø7,65x1,78 36309
8 1 Nipple w/M10x1 11014-16
9 1 Spring 5,0/3,4/8,0 10887-27
10 1 O-ring teflon Ø6,07x1,78 36311
11 1 Spindle short M6x0.5 10887-26
12 1 Spindle guide M6x0.5 10887-24
13 1 Retainer w/slot 11004-18
9-56 Oct. 08
Dameca Service Manual for Dameca MRI 508
13 Loctite 270
1
12 8
2
11
3
10 7
Loctite 270
qusjob42.11013 manual-2.pdf
Oct. 08 9-57
Service Manual for Dameca MRI 508 Dameca
O2 Flush
Item Qty Description Part No.
1 1 House for O2 flush 10658-31
2 1 Valve shaft for O2 flush 10658-32
3 1 Pushbutton for O2 flush 10658-33
4 1 Label O2+, white 11356
5 1 Spring for O2 flush 10784-07
6 1 O-ring ø4,48x1,78 36281
7 1 O-ring ø2,57x1,78 36279
8 1 Quick connector straight 36690-10
9 1 Label QC 11385-11
10 1 Label: BATCH LABLE YY-WW 11385-01
11 1 Quick connector angled, ø6/M5 36690-22
qusjob42.10658-30 manual.png
Fig. 9-31 O2 flush
9-58 Oct. 08
Dameca Service Manual for Dameca MRI 508
'HIDXOW
47<
qusjob42.11960 manual.pdf
Fig. 9-32 IBS system complete
Oct. 08 9-59
Service Manual for Dameca MRI 508 Dameca
IBS Console
Item Qty Description Part No.
1 1 Absorber lock 11960-60
2 1 IBS base 11960-81
3 3 Friktion washer 11960-82
4 1 Freshgas connector 11960-84
5 1 Connector M18x1 11960-85
6 1 Hinge for absorber 11960-13
7 1 Mount block for switch 11960-93
8 1 Quick connector straight, ø6/1/8" 36690-10
9 1 Piston for changeover valve 11960-87
10 1 Cylinder for changeover valve 11960-88
11 1 Connector Ø22 12064-02
12 2 Dowel pin for base 11960-105
13 2 Lock bolt for base 11960-102
14 1 Mount plate for base 11960-101
15 4 Dowel pin Ø3x10 10991-256
16 1 Banjo bolt M5x14 36681-05
17 2 Alu. washer ø8/ø5 x1 36596
18 1 Freshgas connector 11960-841
19 3 Screw DIN 7991 M4x10 70056
20 2 Screw DIN 912 M4X25 70012
21 2 O-ring, silicone red, ø9,19x2,62 36343
22 1 O-ring, silicone red, ø15,08x2,62 36344
23 2 Screw DIN 912 M3X8 70001-0308
24 2 Screw DIN 912 M4X16 70010
25 2 O-ring Ø27,1x1,6 EPDM 36295
26 2 O-ring Ø24x1,6 EPDM 36242
27 2 Screw DIN 912 M8X25 70026
28 4 Screw DIN 912 M5X10 70013
29 1 Banjo with hose connector 36692-7405
30 1 Quick connect Ø8-1/8" Viton 36690-14
31 1 Lock ring w/handle 11960-140
32 2 Dowel pin 3mm 36753-20
33 2 Spring 5,0/3,4/8,0 10887-27
34 1 lock nut DIN 985 M6 70436
35 1 Pin M6 11960-104
36 4 Screw DIN 912 M2X10 70001-0210
37 1 Mount block for switch 11960-92
38 1 Switch for IBS 40090-330
39 1 Switch for absorber 40090-325
9-60 Oct. 08
Dameca Service Manual for Dameca MRI 508
/XEULFDWHZLWK0RO\FRWH';
/RFWLWH
/RFWLWH
qusjob42.11960-80 manual.pdf
Oct. 08 9-61
Service Manual for Dameca MRI 508 Dameca
IBS System
Item Qty Description Part No.
1 1 IBS body 11960-01
2 1 ISO 22/15 Cone 11960-02
3 5 Plug M18x1 11960-03
4 1 Plug 7/8"-26wg 11960-13
5 1 Bellows 11960-05
6 1 Chamber 11960-09
7 1 Valve pin exp. 11960-11
8 1 Valve pin INSP. 11960-12
9 1 Inspiratory valve for absorber 11960-230
10 1 Expiratory valve for absorber 11960-460
11 1 Changeover valve for IBS 11960-550
12 1 Bag connector 11960-70
13 1 Lid for overflow membrane 51525-10
14 1 APL valve 11960-330
15 5 O-ring Ø15,1x1,6 EPDM 36240
16 2 O-ring Ø19,1x1,6 EPDM 36244
17 8 Screw BN 9995, (ISO 14585C), 2,9x16 70128-2,9x16
18 4 Screw BN 9995, (ISO 14585C), 2,9x9,5 70128-2,9x9,5
19 2 One-way valve 11890-31
20 1 O-ring ø49,6x2,40 36261-4924
21 3 Dowel pin Ø2 x 10 36757-10
22 1 Release button for APL valve 11960-77
23 6 O-ring Ø18,1x1,6 EPDM 36267
24 1 Membrane 51025-08
25 1 Membrane weight 51025-07
26 2 Spring 4,3/3,3/15mm 11080-62
27 2 Valve dome with o-ring 11890-42
28 1 ISO 22/15 cone 11960-10
29 1 O-ring Ø27,1x1,6 EPDM 36295
30 1 O-ring, silicone blue, ø129,77x3,53 36330
9-62 Jan. 10
Dameca Service Manual for Dameca MRI 508
qusjob42.11960-100 manual.pdf
Oct. 08 9-63
Service Manual for Dameca MRI 508 Dameca
9-64 Oct. 08
Dameca Service Manual for Dameca MRI 508
10. Schematics
Contents Page
Oct. 08 10-1
10-2
Wall Power Supply Unit External cable Internal cable Ventilator
5m
20
24V DC B1
-B1
PIN 5 F5
A (pin 5) 13 16 21
+24V DC +24V DC B2
2.3 2.4
2AT 4AT
Service Manual for Dameca MRI 508
Battery F6
Charger 4AT
MAINS 2AT
24V DC 18
+
24V Battery Pack
PIN 4 B (pin 4) 15 17
0V DC 0V DC
-B1
PIN 3 C (pin 3) F4
4 5 26
+24V DC +24V DC
22
4AT 84
PSU
24V DC
PIN 2 D (pin 2) 85
23 23
0V DC 0V DC 3
24
0V DC 4
25
0V DC 2
Oct. 08
Dameca
Oct. 08
Dameca
+
24V Battery Pack
PIN 3 PIN 3 15 17
0V DC 0V DC
-B1
PIN 2 PIN 2 F4
4 5 26
+24V DC +24V DC
22
4AT 84
PSU
24V DC
PIN 5 PIN 5 85
23 23
0V DC 0V DC 3
24
0V DC 4
25
0V DC 2
CPC CPC
10-3
qusjob3.siesta i whispa mri wiring a3 new.pdf
Service Manual for Dameca MRI 508
Service Manual for Dameca MRI 508 Dameca
10-4 Oct. 08
Dameca Service Manual for Dameca MRI 508
Oct. 08 10-5
Service Manual for Dameca MRI 508 Dameca
10-6 Oct. 08
Dameca Service Manual for Dameca MRI 508
48
61
22 46 45
58
56 59 68
60 70 49
57 67
47
55 69
54 66 77
52
44
53 50
51
72 73
25 71
26 65 76 82
24 64
63
74 78
23
8 62
4
1
7 9 79 81
13
11 80 75
12 20
11 85
28 17 32 40
10 43 21 83 84
3
7 9 16 31 39 18
33 15 38
34
37
5 14 30 36
7 9
19
35 29
qusjob42.10650-whispa-mri-pne-manual.png
Fig. 10-6 Pneumatic diagram, Dameca MRI 508
Oct. 08 10-7
Service Manual for Dameca MRI 508 Dameca
Dameca Service Manual for Dameca MRI 508
Note:
All flow and volume measurements are related to STPD conditions
(Standard Temperature Pressure Dry).
11.1 General
Dimensions:
Height : 1510 mm
Width : 760 mm (including IBS)
Depth : 670 mm
Weight: : Approx. 110 kg (depending on configuration)
MRI Compatibility:
Max. magnetic field : 1000 Gauss
Electrical:
Class :I
Type :B
Mains voltage : 100/115/230 volt 50/60 Hz.
Power consumption (typical) : 80 VA
Battery capacity : 2.2 Ah (ventilator)
Battery operation time : App. 60 minutes (only by mains failure)
Battery type : Pb rechargeable
Charge time : Approx. 12 hour
Note:
There are no changes in performance when the ventilator is battery powered.
11.2 Gases
Note:
All gases and anaesthetic agents must conform to the European Pharmacopeia.
Oct. 08 11-1
Service Manual for Dameca MRI 508 Dameca
Pressure gauges:
Central gas supply : 0-10 (x100kPa) for O2, AIR and N2O
Gas cylinder (optional) : 0-315 (x100kPa) for O2
: 0-100 (x100kPa) for N2O
AGSS:
Scavenging flow : 30 to 40 L/min
Gas Alarm:
Alarm start-pressure for O2 : 2.5 (x100kPa)
11.3 Rotameterbox
Rotameters:
Measuring range : 0.1-10 L/min for O2, AIR and N2O
optional: 0.05-1 L/min for O2 and N2O
1-10 L/min for O2 and N2O
11-2 Oct. 08
Dameca Service Manual for Dameca MRI 508
Minoxyguard:
Min. O2 concentration in N2O/O2 mix : 25% O2
Min. O2-flow in order for N2O-flow to open : app. 0.3 L/min O2
O2 flush flow:
Fixed flow : App. 45 L/min
11.4 Ventilator
Driving gas (AIR or O2):
Pressure : Min. 3.0 (x100kPa) at 80 L/min.
Max. consumption (peak flow) : 120 L/min
Average consumption : Max. 80 L/min at 2.8 (x100kPa)
Pressure range:
Pressure limitation (Plim max) : 90 (x100Pa) (mechanical pressure limitation valve)
Max. adjustable working pressure : 80 (x 100Pa)
High pressure alarm : 10 to 80 (x100Pa)
Min. expiration pressure (Plim min) : +1 (x100Pa)
Cycling pressure : 1 to 80 (x100Pa)
Note:
Negative airway pressures are possible if the patient is breathing spontaneously only.
Set parameters:
Tidal Volume : 20 to 1500 ml.
Accuracy : 20 to 250 ml: +/-10%, min. 10 ml
: 250 to 1500 ml: +/-5%, min. 25 ml
Minute volume : 0.2 to 60 L/min.
Frequency : 4 to 80 resp./min.
Accuracy : +/- 5 %
I:E Ratio : 3:1 and up to 1:9,9
PEEP : OFF, 4 to 20 (x100Pa)
Inspiratory pressure : 10 to 70 (x100Pa)
Inspiratory pause : 0 to 70 %
Ventilation mode (controlled) : VCV, SIMV (option), PCV, PSV (option), PRVT (option)
Oct. 08 11-3
Service Manual for Dameca MRI 508 Dameca
SIMV settings:
Trigger point : -0.5 to -10.0 (x100Pa)
PSV settings:
Support pressure : 4 to 50 (x100Pa)
Inspiratory trigger : 1 to 10 L/min.
Expiratory trigger : 10 to 80 %
PSV backup : 10 to 40 sec.
Sigh settings:
Interval : 50 to 150 respirations
Added tidal volume : 30 to 80 %
Note:
The sigh function is active in VCV and SIMV mode only.
11-4 Oct. 08
Dameca Service Manual for Dameca MRI 508
11.6 Environment
Storage temperature : -20 to + 50OC (optional O2-sensor: 0OC to +50OC)
Working temperature : 15 to 35OC
Relative humidity : 10 to 90% RH (non condensing)
Storage pressure : 630 to 1060 (x 100Pa)
Working pressure : 960 to 1060 (x 100 Pa)
11.7 Standards
Standards fulfilled:
Software : EN 60601-1-4
EMC : EN 55011; Group 1; Class B
Electrical safety : EN 60601-1
Function safety : EN ISO 10524-1, EN ISO 5359, EN 740,
EN 837-1, EN 980, EN ISO 5356-1
Oxygen monitor (optional) : EN ISO 19054, EN ISO 21647, EN ISO 15002
Oct. 08 11-5
Service Manual for Dameca MRI 508 Dameca
11-6 Oct. 08
Dameca Service Manual for Dameca MRI 508