Dameca Service Manual for Dameca MRI 508

Service Manual for Anaesthesia Machine


Model Dameca MRI 508

Important !
We are constantly developing our products, and
consequently reserve the right to make alterations
in design and equipment without prior notice.

DAMECA a/s issue a guarantee against manu-


facturing and material defects for a period of
one year from the date of delivery. Excepted
from this guarantee are rubber articles, anaest-
hesia valves with plastic parts and batteries.

DAMECA a/s will repair or replace all defective


components during the period of guarantee,
provided (1) that the apparatus has been used
correctly in accordance with the instructions for
use and (2) that the instructions for maintenance
have been observed carefully.

Defective products or components must be


returned to DAMECA a/s with a detailed
description of the defect in question.

This manual is valid from the date of issue. If repairs are carried out by anyone else than
Questions concerning updates can be DAMECA a/s or by a DAMECA a/s appointed
obtained at the manufacturer. agent, the guarantee no longer applies.

Nobody but DAMECA a/s can issue any other


form for guarantee, either in writing or orally.

Preventive maintenance of this apparatus should


be carried out at least once a year either by
DAMECA a/s or an authorised technician.

Conformity according to the


Council Directive 93/42/EEC Document No. 10651MRI-91
concerning Medical Devices. October, 2008

Oct. 08 I
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

II Oct. 08
Contents 1

User Responsibility 2

Safety Precautions 3

Introduction 4

Description 5

Troubleshooting 6

Maintenance 7

Calibration 8

Spare Parts 9

Schematics 10

Technical Data 11

Earlier Revisions 12
Dameca Service Manual for Dameca MRI 508

Contents
1. Contents (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

2. User Responsibility (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

3. Safety Precautions (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1


3.1 Warnings (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 Cautions (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.3 Symbols (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4

4. Introduction (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1


4.1 Intended Use (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 Description (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.3 Performance (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.4 User Setup (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.5 Responsibility of the Manufacturer (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.6 Safety Features of the Dameca MRI 508 Anaesthesia Machine (Oct. 08) . . . . . . . . . . . 4-4
4.7 Equipment Classification (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4

5. Description (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1


5.1 Pneumatic Symbols (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.2 Description Pneumatics (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.3 Description Electronics (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
5.4 Ventilator (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10

6. Troubleshooting (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1


6.1 General (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.2 Maximum Allowable Leaks (according to ISO 5358) (Oct. 08) . . . . . . . . . . . . . . . . . . . 6-1
6.3 Troubleshooting Pneumatics (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.4 Troubleshooting Electronics (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
6.5 Changing Mains Power Voltage (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.6 Changing Drivegas (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
6.7 Changing the Setup (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7

Oct. 08 1-1
Service Manual for Dameca MRI 508 Dameca

Contents (continued)
7. Maintenance (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.1 Maintenance, General (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.2 Tools and Test Equipment (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7.3 Service Kits (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
7.4 Annual Maintenance (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
7.5 Post Maintenance Check (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
7.6 Four Year Maintenance (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
7.7 Service Menu (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
7.8 Post Maintenance/Repair Check List (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
7.9 Check Scheme for Post Maintenance / Repair (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . 7-37

8. Calibration (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1


8.1 Insp. Flow & Adult Pat. Flow (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8.2 Bag in Bottle and Patient Pressure (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
8.3 Patient Flow (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
8.4 Freshgas Flow (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
8.5 PEEP Valve (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
8.6 Temperature Calibration (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
8.7 Drivegas Adjustments (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
8.8 Minoxyguard (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10

9. Spare Parts (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1

10. Schematics (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1

11. Technical Data (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1


11.1 General (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
11.2 Gases (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
11.3 Rotameterbox (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-2
11.4 Ventilator (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
11.5 Integrated Breathing System (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
11.6 Environment (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
11.7 Standards (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5

12. Earlier Revisions (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1

1-2 Jan. 10
Dameca Service Manual for Dameca MRI 508

List of Figures
Contents Page

Fig. 5-1 Pneumatic diagram, Dameca MRI 508 (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Fig. 5-2 Mains Power Diagram, Dameca MRI 508 (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9

Fig. 6-1 Drivegas block (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6

Fig. 7-1 Service kit Dameca MRI 508, 3 rotameters (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Fig. 7-2 Service kit Dameca MRI 508, 5 rotameters (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Fig. 7-3 Service kit Dameca MRI 508, 4 years (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Fig. 7-4 Inlet manifolds (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Fig. 7-5 Pin index yoke for spare cylinder (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10

Fig. 9-1 Dameca MRI 508 (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3


Fig. 9-2 Stand with integrated suction (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Fig. 9-3 Gas- and electrical panel (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Fig. 9-4 Gasbox for Dameca MRI 508 (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Fig. 9-5 Gasbox assy. (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
Fig. 9-6 Control panel (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
Fig. 9-7 Toggle switch valve (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
Fig. 9-8 Turn switch valve (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
Fig. 9-9 Rotameterbox with 3 tubes (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-15

Fig. 9-10 Minoxyguard with 3 tubes (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-17


Fig. 9-11 Rotameterbox with 5 tubes (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-19
Fig. 9-12 Drive gas block (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-23
Fig. 9-13 Backbar Selectatec (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-25
Fig. 9-14 Backbar Dräger (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-27
Fig. 9-15 Inlet manifold O2 (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-29
Fig. 9-16 Inlet manifold AIR (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-31
Fig. 9-17 Inlet manifold N2O (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-33
Fig. 9-18 Inlet manifold VAC (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-34

Fig. 9-19 Inlet manifold AGSS (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-35


Fig. 9-20 Whispa ventilator (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-37
Fig. 9-21 E-Box for Whispa ventilator (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-39
Fig. 9-22 Power supply box (Jan. 10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-41
Fig. 9-23 Valve block assy. for Whispa (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-43
Fig. 9-24 Valve block assy for Whispa (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-47
Fig. 9-25 Rear cover (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-48

Jan. 10 1-3
Service Manual for Dameca MRI 508 Dameca

List of Figures (continued)


Contents Page

Fig. 9-26 Flowmarker assembly (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-49


Fig. 9-27 Suction unit (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-51
Fig. 9-28 Mains switch, standby (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-53
Fig. 9-29 Vacuum suction, New version (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-55
Fig. 9-30 AGSS ejector (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-57
Fig. 9-31 O2 Flush (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-58
Fig. 9-32 IBS system complete (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-59
Fig. 9-33 IBS console (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-61
Fig. 9-34 IBS system (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-63

Fig. 10-1 Wiring diagram for ventilator, Old version (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . 10-2
Fig. 10-2 Wiring diagram for ventilator, New version (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . 10-3
Fig. 10-3 Wiring diagram for ventilator (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Fig. 10-4 Wiring diagram for ventilator (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Fig. 10-5 Wiring diagram for ventilator (Oct. 08). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
Fig. 10-6 Pneumatic diagram Dameca MRI 508 (Oct. 08) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7

1-4 Oct. 08
Dameca Service Manual for Dameca MRI 508

2. User Responsibility
This anaesthesia machine has been built to instructions issued by DAMECA a/s only, and
conform to the specifications and operating must not be altered or modified in any way
procedures stated in this manual and/or without the written approval of DAMECA a/s.
accompanying documentation, labels and
notices when checked, assembled, operated, The user of this machine has sole responsibility
maintained, and serviced in accordance with for any malfunction which results from impro-
the instructions provided. per use, maintenance, repair, damage or altera-
tion by anyone else than DAMECA a/s or their
To ensure the safety of this machine it must be appointed agents.
checked and serviced to at least the minimum
standards in this manual.
Note:
A defective or suspected defective product Always give as much of the following
must under no circumstance be used. information as possible.
Worn, broken, distorted, contaminated or
missing components must be replaced immedia-
tely. Should such repair become necessary, it is O Type of equipment
recommended that a request for service advice
is made to the nearest DAMECA a/s service O Serial number
centre or agent.
O Date of purchase
This machine and any of its constituent parts
must be repaired in accordance with written O Apparent fault

Agent Name and Address:

Oct. 08 2-1
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

2-2 Oct. 08
Dameca Service Manual for Dameca MRI 508

3. Safety Precautions
Read this manual carefully and familiarize O Do not use this machine in an environment
yourself with the machine before you start exceeding the EMC levels in EN 60601-1-2.
to operate the machine.
O External patient breathing systems used
Always check the machine for proper function with this machine must conform to EN 740,
before use. EN ISO 8835-2 or similar standard.

A WARNING indicates a situation in which O Bacterial filters used with this machine must
the user or the patient may be in danger of conform to EN 13328-1 or similar standard.
injury.
O An anaesthetic vapour delivery device
A CAUTION indicates a situation that may connected to the machine shall comply
lead to equipment damage or malfunction. with ISO 8835-4 or similar standard.

A NOTE indicates a point of particular interest O Heat and moisture exchangers used with this
for more efficient and convenient operation. machine must conform to EN ISO 9360 or
similar standard.

3.1 Warnings O Humidifiers used with this machine must


conform to EN ISO 8185 or similar standard.
O Do not open service hatches or covers. No
user-serviceable parts inside the machine. O Suction units used with this machine must
A qualified service technician should conform to EN ISO 10079-1,
perform service. EN ISO 10079-2 and EN ISO 10079-3
or similar standard.
O The machine must be connected to correct
grounded hospital grade electrical outlets O Medical gases and anaesthetic agents used
only. with this machine must conform to the
European pharmacopoeia.
O Always unplug the electric power cord
before cleaning or maintenance of the O Central gas supply used with this machine
machine. Let it dry completely before must comply with EN ISO 9170-1 and
reconnecting it to the electrical outlet. ISO 7396 or similar standard.

O Spillage of liquids may cause electrical


shock. Note:
If the central gas supply fails, the
O Check that all cables are in good working auxiliary gas outlet (optional) of
condition before connecting the machine this machine will also fail.
to mains or patient.

Oct. 08 3-1
Service Manual for Dameca MRI 508 Dameca

O A patient monitor connected to the patient Patient safety


breathing system, may cause a leakage, O Do not use the machine for a clinical
even if it is in stand-by mode. procedure without first ensuring that the
O The AGSS must be turned on whenever the complete system is operating correctly.
machine is in use. O The machine should always be used with a
O Serious personal injury and property damage separate monitor for O2, anaesthetic agent
has resulted from the use of steel gas cylinders and CO2 concentrations. This monitor must
in close proximity to MRI devices. comply with the relevant standards under
the EU Medical Device Directive.
O Do not add, remove or modify any part of
the machine, unless authorised by Dameca a/s. O If the volume monitor of the Dameca MRI
508 is not installed, the machine should
O Do not place magnetic items on the machine, always be used with a separate monitor for
as they will pose a danger in an MRI expired volume. This monitor must comply
environment ! with the relevant standards or the EU
O Compressed gas cylinders, cylinder valves, Medical Device Directive
cylinder regulators, other assemblies inclu- O Ensure that all connections are secure and
ded with cylinders or any other equipment
check for proper operation of the machine
used to supply compressed gases shall not
before use. In particular the patient breathing
be present in the same room as an MRI
system and O-rings should be checked to
device, unless proven non magnetic and/or
ensure that connections are not loose, and
tested and deemed suitable for use in an
the patient breathing circuit is not leaking.
MRI environment.
O The use of antistatic or electrically condu-
Explosion hazard
ctive breathing tubes when utilizing high
O No oil or grease should be used in this frequency surgery equipment may cause
machine. This also applies to silicone burns and is therefore not recommended
based lubricants and inflammable oil or in any application of this machine.
grease. Only PTFE based oxygen
compatible lubricants must be used, O Constant attention by a qualified individual
(e.g. DAMECA special grease P/N 36825). is required whenever a patient is under
O
anaesthesia or connected to a ventilator.
The machine is not intended for use in the
Some malfunction of equipment requires
presence of flammable anaesthetic agents
immediate action.
such as ether and cyclopropane. Only
anaesthetic agents complying with EN 740 O When using the ventilator, alternative means
or similar standards must be used. for ventilating the patient must be present.
O Combustible materials that could be found O When operating the machine, an alternative
near patients, should be avoided in the area oxygen supply must always be present.
where oxygen is being administered. This
includes: hair oils, oil-based lubricants, skin O Use only the anaesthetic agent for which
lotions, facial tissues, clothing and linen, the vaporizer is calibrated. Using the
rubber and plastic articles, suction tubing, wrong agent in the vaporizer can be fatal
alcohols and acetone. for patients.

3-2 Oct. 08
Dameca Service Manual for Dameca MRI 508

O If a re-usable CO2 absorber is used, the Occupational hazard


following must be taken into account during O Do not use this machine in insufficiently
change of soda-lime: Soda-lime dust is ventilated areas. Use scavenging system to
caustic and can produce burns in the respira- avoid increased concentrations of anaest-
tory tract if inhaled. Breathing protection hetic gases in operating rooms. Observe
must be used to ensure that no soda-lime that the ball of the flowmarker is visible.
dust is inhaled by personnel or patient.
O Do not start any test procedure of the
ventilator with a patient connected to the
3.2 Cautions
machine. The test procedures are designed O Before connecting the power cord to the
to ensure correct and safe function of the
electrical power supply outlet check that
machine. Do not bypass the test procedures
the local voltage and frequency are as
unless patient safety requires an immediate
stated on the TYPE label on the rear
starting up procedure.
of the machine.
O The O2 percentage in the patient circuit may O Maximum load on side rails at table is 20 kg.
differ significantly from the O2 percentage
Too much heavy equipment attached at one
of the freshgas.
side of the machine may cause the machine
O When using soda lime in absorbers, follow to tilt.
the manufacturers instructions. O Maximum total load on table plate
O If the machine has been used with different (including side rails) is 40 kg.
anaesthetic agents, residues of these agents O Maximum load on top plate is 40 kg.
(PPM) may still remain in the machine
(vaporizer rail etc.). By setting a high O Maximum torque on siderails is 20Nm.
freshgas flow during a long period of O When closing the drawers of the machine,
time these residues can be minimized.
observe that nothing gets squeezed or
O Means to avoid cross infection between damaged, with special attention to fingers.
patients must be used. See section 8 O Vaporizers used for Fluothane, Halothane or
regarding maintenance and cleaning.
other anaesthetic agents containing Tymol or
other stabilisers must be emptied every other
Note: week, and the agent that is drained must be
The auxiliary O2 flowmeter (optional) destroyed.
is by-passing the MPL valve. The pres- O Check calibration of vaporizers regularly
sure from the flowmeter may exceed using a proper gas indicator (e.g.refracto-
the pressure from the freshgas outlet. meter or similar).
O Authorised service personnel should service
Failure of operation this machine at least once a year.
O If the machine fails to respond as described, O Vaporizers should be serviced according
do not use until qualified service personnel to the manufacturer’s instructions by an
have corrected the situation. authorised service centre.

Oct. 08 3-3
Service Manual for Dameca MRI 508 Dameca

O Use only original DAMECA hoses and O Do not leave spare cylinder valves open
accessories; other hoses and accessories when using the machine on central gas out-
may cause malfunctions. lets. Pressure from cylinders may be higher
than the pressure from central outlets, and
O Use only original DAMECA O-rings; the cylinder supply could be depleted.
other O-rings may cause problems if they
are not of the same size or material. O Disconnect the machine from mains and
medical gas supply, when not in use, to
prevent pollution.

Note:
Tubings used with this product must fulfill the EN12342 standard, and
respiration bags must fulfill the EN1820 standard.

3.3 Symbols

Type B (EN60601-1) protection ON/OFF switch for lamp.


against electrical shock.

Protective earth.
Consult accompanying documents.

Equipotentiality.
Attention, consult accompanying
documents.
ON/OFF

Adjusment O2-, AIR- or N2O-


freshgas flow and for for auxiliary WEEE Symbol, meaning that the
O2 flow or patient suction: electronics must not be regarded
Turn counter-clockwise to increase. as normal waste.
Turn clockwise to decrease.

Activate brakes on the front


wheels after the machine has
Gauss level. been placed correctly in the
MR-scanner room.

3-4 Oct. 08
Dameca Service Manual for Dameca MRI 508

4. Introduction
This manual is related to ventilator O An APL (Adjustable Pressure Limiting)
software version 2.1. valve is integrated into the IBS, giving the
user a choice between manual ventilation
Dameca anaesthesia machine model Dameca or spontaneous breathing.
MRI 508 is applicable with most types of
inhalation anaesthesia. O2 and N2O, or O2 O The ventilator is microprocessor controlled,
and AIR can be administered, and anaesthetic with software installed.
vaporizer(s) is connected to a back-bar.
The gas supply can be obtained from a central The machine can be equipped with various
gas supply system. additional equipment such as:
O Active scavenging system
O Transfusion stand
The machine is equipped with: O Arm for patient monitor
O A rotameter unit (minoxyguard) giving O Pin-index reduction valves for 4 L. gas
user selectable gas mixes (O2, AIR and cylinders.
N2O) in a range of 0 - 10 L/min for each
gas. If N2O is selected as a mix to O2, the The design of the machine is very flexible
N2O percentage is limited to max.75% towards mounting of extra equipment, and
(minimum 25% O2). additional rails, clamps etc. can easily be
mounted on the machine.
O A microprocessor controlled anaesthesia
ventilator with user selectable ventilation
modes: 4.1 Intended Use
- VCV Volume Controlled Ventilation The Dameca MRI 508 anaesthesia machine
- PCV Pressure Controlled Ventilation is intended for composition of anaesthetic
- SIMV Synchronised Intermittent breathing gas and manual or automatic
Mandatory Ventilation ventilation of patients undergoing anaesthesia,
- PSV Pressure Supported Ventilation in MRI scanning rooms.
- PRVT Pressure Regulated Volume
Target (option) The Dameca MRI 508 anaesthesia machine is
intended for adult and paediatric patients
A real time clock is also included in the Qualified anaesthesia personnel should operate
ventilator. the Dameca MRI 508 anaesthesia machine only.
Before operating the machine mains electrical
O An integrated breathing system (IBS) power and pressurised gas must be connected
where Bag in Bottle and patient re-breathing to the machine.
system (incl. absorber) are integrated into
one block. Thereby there are only 2 tubes
(inspiratory and expiratory) from the IBS The machine is MRI compatible, for use in
system to the Y-piece. magnetic fields up to 1000 Gauss. *

Oct. 08 * This applies for machines marked with a yellow Gauss level label. 4-1
Service Manual for Dameca MRI 508 Dameca

4.2 Description
The rotameter unit (minoxyguard) of the two valves automatically close the connections
Dameca MRI 508 controls the freshgasflow to the absorber, in order not to loose gases.
given to the patient breathing system.
The flow is adjusted by the user according
to the measured flow of the rotameter tubes. Note:
The parts of the Dameca MRI 508
The ventilator of the Dameca MRI 508 uses
do not contain any Latex.
compressed AIR or O2 to ventilate the patient
through the Bag in Bottle system, placed in
the integrated breathing system (IBS).
The ventilator continuously measures airway
pressure and inspired/expired volume (optional). Note:
The ventilator can also measure inspired O2 by The Dameca MRI 508 contains
means of a fuel cell sensor (optional) placed on batteries that must be exchanged
the inspiratory port of the patient breathing system. by a technician at certain intervals,
as prescribed in the service manual.
The integrated breathing system (IBS) of the
Dameca MRI 508 is a semi-closed patient
breathing system with one-way valves in both
the inspiratory and expiratory limb, and an 4.3 Performance
absorber for eliminating CO2. The system is
suitable for high flow (approx. the patient minute The rotameter unit (minoxyguard) of the
volume), medium flow (approx. 1-1.5 L/min) Dameca MRI 508 is designed to give a
and low flow (approx. 0.5-1 L/min) of freshgas freshgas mixture selected by the user of:
for patient ventilation.
O 0 to 10 L/min for O2, AIR and N2O.
The IBS is a combination of a Bag in Bottle unit
and a patient re-breathing circuit. The IBS has This flow is led through the vaporizer rail,
a standing bellows in order for the user to see where anaesthetic agent can be added by
the controlled ventilation of the patient by the means of a user selectable vaporizer.
movement of the bellows. Besides, in case of By activating an oxygen by-pass valve,
a leakage, room air will not be drawn into the approx. 45 L/min of O2 will be added to
system, and the leakage can be seen immediately the freshgas flow, after the vaporizer rail.
by the position of the bellows.
The Dameca MRI 508 has an integrated gas
Integrated into the IBS is an APL valve which scavenging system which removes excess
is used for manual ventilation only. gases from an open reservoir. This system
must be connected to an AGSS power unit
The IBS uses a CO2 absorber i-SORB which with a flow of 30-40 L/min. If an external
contains 900 gr. Sodalime. The absorber is AGSS power unit is not available, the
available as a re-usable or a disposable version. Dameca MRI 508 can be equipped with an
If the absorber is disconnected from the IBS, integrated AGSS power unit (optional).

4-2 Oct. 08
Dameca Service Manual for Dameca MRI 508

The ventilator of the Dameca MRI 508 is 4.4 User Setup


designed to give an inspiratory flow of:
The Dameca MRI 508 can be set up in order to
O Continous flow: 2 to 80 L/min match user preferences.
O Max. Peak flow: 120 L/min.
The following features can be installed when
In VCV and SIMV mode the Tidal volume, the machine is manufactured:
Resp. rate, I:E ratio and Insp. pause user settings
are continuously controlled by the ventilator, O Integrated patient suction.
in order to stay within the 2 to 80 L/min O Auxiliary freshgas outlet and integrated
continuous flow limits. auxiliary O2 flowmeter
O Extra O2 , AIR or VAC gas outlet
Inspiratory flow = Tidal volume * Resp. rate * O Power unit for AGSS
((I+E)/I) * (100/(100-Insp. pause)) O PSV accessible or not
O PRVT accessible or not
The integrated breathing system (IBS) of
the Dameca MRI 508 has an expiratory The following parameters can be changed
and inspiratory resistance of: after the machine has been manufactured:

60 L/min 30 L/min 5 L/min O Ventilator volume setting based on


Expiratory 4.5 2.0 0.3 Tidal volume or Minute volume.
resistance (x100Pa) (x100Pa) (x100Pa) O SIMV accessible or not.
Inspiratory 4.5 2.0 0.3 O PSV accessible or not.
resistance (x100Pa) (x100Pa) (x100Pa) O Oxygen monitor based on external fuel cell
sensor can be activated or de-activated.
Compliance = 4 ml/hPa with adult tubings O Volume monitor can be activated or
(22mm diameter). de-activated.

The CO2 absorber i-SORB of the integrated


breathing system IBS is designed to optimize If volume monitor is activated, the volume
the flow through the absorber, so that all of measurement can be configured in two ways:
the sodalime is being used.
Only expired volume
If the volume monitor is configured to »only
exp. volume«, the volume sensor must be placed
at the expiratory port of the patient breathing
system. The volume monitor will then measure
the expired tidal- and minute-volume only.

Full spirometry
If the volume monitor is configured to »full
spirometry«, the volume sensor must be placed
at the Y-piece of the patient breathing system.
qusjob8.absorberkurver.tif

Oct. 08 4-3
Service Manual for Dameca MRI 508 Dameca

The volume monitor will then measure expired Minoxyguard


tidal- and minute volume, inspired tidal volume N2O cut-off if O2 supply fails.
(during manual ventilation) and a pressure/
volume graph (spirometry loop) can be selected O2/N2O ratio control (secures a minimum
by the user. of 25% O2 in an O2/N2O mix).

All these settings can be changed in the service- An MPL (Maximum Pressure Limiting) valve
menu, by a super-user or a technician. opens if the pressure of the freshgas flow to the
auxiliary freshgas outlet exceeds 90 (x100Pa).

4.5 Responsibility of the Vaporizer rail


Manufacturer Back-bar for one or two vaporizers. The back-
bar for two vaporizers is fitted with interlock
Dameca a/s is responsible for the effects on system to assure that only one vaporizer can
safety, reliability and performance of the be activated at a time.
equipment only if:
Ventilator
O Personnel authorized by Dameca a/s A high-pressure alarm will automatically force
carry out assembly, operations, extensions, the ventilator into its expiratory state, avoiding
re-adjustments, modifications, periodical dangerous overpressure situations. In addition
service or repairs. the ventilator has a built-in pressure relief valve
which opens at 90 (x100Pa).
O The pneumatic installation of the relevant
rooms complies with appropriate
requirements. 4.7 Equipment Classification
O The equipment is used in accordance with Classification according to EN60601-1
this user manual.
Class: I According to the type
of protection against
4.6 Safety Features of the Dameca electrical shock.
MRI 508 anaesthesia Machine
Type: B According to the degree
Gas supply of protection against
O2, supply failure alarm. Built-in check valves electrical shock.
to prevent reverse flow of gases.
Continuous operation: According to the mode
The ventilator, patient suction (optional) and of operation.
AGSS power unit (optional) is powered by the
driving gas (AIR or O2 , normally AIR).
Classification according to EU
Medical Device Directive
The optional patient suction can also be The Dameca Dameca MRI 508 anaesthesia
powered by VAC (option). machine is classified as IIb.

4-4 Oct. 08
Dameca Service Manual for Dameca MRI 508

5. Description

Note:
In this manual pneumatic symbols are used in pneumatic diagrams and
some figures, in order to simplify service and troubleshooting.

5.1 Pneumatic Symbols

CONNECTION POINT
ON/OFF VALVE

FILTER

PRESSURE GAUGE

CHECK VALVE

PROPORTIONAL VALVE
PRESSURE REDUCTION
VALVE

O2 FLUSH VALVE PILOT VALVE

Oct. 08 5-1
Service Manual for Dameca MRI 508 Dameca

5.2 Description Pneumatics

Oxygen
Oxygen is supplied from either the central If the user selects AIR the valve (29) is open
inlet (1). and AIR will flow to the regulator knob (30)
for AIR freshgas flow, controlled by the user.
Via filters (7) it is let to the pressure gauge (8) After that the flow enters a filter (31) and then
and then to a check valve (9). the AIR rotameter (32).

O2 enters the O2 tank (10) which has a separate The flow is let to the freshgas flow sensor (18)
check valve (11) and a regulator (12) that opens of the ventilator and then to the back bar (19),
for O2 flow to a whistle (13), if the O2 pressure and the fresh gas selector switch (20). The user
drops to app. 2.5 (x100Pa), thereby giving can select either the AUX fresh gas outlet (21)
an oxygen failure alarm. or the integrated breathing system (22).
Additionally, AIR is let to the auxiliary AIR
The O2 freshgas flow is controlled by a outlet (33).
regulator knob (14), by the user.

After that the flow enters a regulator (15) for


N2O
minoxyguard characteristics, a filter (16) and N2O is supplied from either the central inlet (5).
then the O2 rotameter (17). The flow is let to the Via filters (7) it is let to the pressure gauge (34)
freshgas flow sensor (18) of the ventilator and and then to a check valve (9).
then to the back bar (19), and the fresh gas
selector switch (20). The user can select either If the user selects N2O the valve (35) is open
the AUX fresh gas outlet (21) or the integrated and N2O will flow to the flow to the regulator
breathing system (22). Additionally, oxygen is knob (36) controlled by the user.
let to the auxiliary oxygen outlet (23).
After that the flow enters the proportional
The pressure regulator (24) reduces the incoming regulator (37) securing min. 25% O2 in an
pressure to 4 bar for the Aux. O2 flow meter O2/N2O freshgas flow, a regulator (38) for
(25) giving 0 - 10 l/min O2. The reduced minoxyguard characteristics, a filter (39)
pressure is also let to the O2 flush valve (26) and then the N2O rotameter (40).
giving an O2 flow of 40 (+/- 5) l/min. to the
fresh gas selector switch (20), by-passing the The flow is let to the freshgas flow sensor (18)
vaporisers connected to the back-bar (19). of the ventilator and to the back bar (19), and
the fresh gas selector switch (20). The user can
select either the AUX fresh gas outlet (21)
AIR or the integrated breathing system (22).
AIR is supplied from the central inlet (3).
Via a filter (7) it is let to the pressure gauge The pressure of the AUX freshgas outlet (21), is
(28), and then to a check valve (9). limited to 90 (x100Pa) by the MPL valve (43).

5-2 Oct. 08
Dameca Service Manual for Dameca MRI 508

Drive Gas Ventilator

Driving gas is used for the ventilator, patient The ventilator (62) is pneumatically powered
suction and active AGSS. The driving gas is by the driving gas through the inlet filter (4).
normally AIR, but O2 can also be used as A pressure switch (63) monitors the driving gas
driving gas. pressure, giving an alarm if the pressure is
below 2.5 bar.
AGSS Driving gas is let to the pressure regulator (64)
Driving gas is reduced to 4 bar by a pressure reducing the pressure to 3 bar for the Insp. flow
regulator (44), and is supplied to the ON/OFF control valve (65).
switch (45). When ON, the ejector (46) will
create a flow from the spill valve in the inte- This valve is controlled by the ventilator
grated breathing system (22) through the flow electronics, selecting a correct inspiratory flow
marker (47) to the AGSS outlet (48) on the according to the ventilator settings. A differen-
back of the machine. tial pressure transducer connected to the
pressure ports P1 and P2 measures the flow
For hospitals with active scavenging systems, and during inspiration the signal to the insp.
the machine can be delivered without the flow valve is continously adjusted, in order
ON/OFF switch (45) and the power unit (46). for the ventilator to give correct inspiratory
flow to the bag in bottle.
If an external patient system is used, it can be
connected to the Aux. AGSS outlet (49). A pressure transducer connected to the pressure
port P3 measures the driving gas pressure
An open reservoir (50) prevents vacuum to (Bag-in-Bottle chamber pressure).
be built-up in the AGS system.
During inspiration the Expiratory valve (66),
and the PEEP valve (67) are closed, and the
Suction unit gas powered driving gas is let to the chamber (68) of the
Driving gas is reduced to 4 bar by a pressure Bag-in-bottle system. The exhalation valve (69)
regulator (44), and is supplied to the jet suction is closed during inspiration, and open during
unit (51) through the ON/OFF switch (52). The expiration when the bellows (70) is in top and
vacuum is controlled by the regulator (53), and the pressure in the bellows is approx. 3 hPa
monitored by the vacuum gauge (54). A suction higher than the pressure from the ventilator.
jar can be connected to the suction connector (55). Excess gases are evacuated through the AGS
system.
Suction unit VAC powered During expiration the expiration valve (66) is
VAC is supplied from the central inlet (56). open, and the pressure in the chamber (68) is
It is led to the auxiliary VAC outlet (57), and released to the AGS system.
then to the ON/OFF switch (58). The vacuum
is controlled by the regulator (59) and monitored If a PEEP pressure (4 - 20 hPa) is selected, the
by the vacuum gauge (60). A suction jar can be insp. flow controller is set to a driving gas flow
connected to the suction connector (61). during expiration.

Oct. 08 5-3
Service Manual for Dameca MRI 508 Dameca

The Expiratory valve (66) and the PEEP valve expiration valve (76), the Bag/vent valve (72)
(67) are fully open, at the beginning of expira- to the bag (73).
tion. When the pressure in the patient breathing
system reaches the set PEEP level, the Expira- The inspiratory pressure is limited in the APL
tory valve (66) is closed and the PEEP valve valve (77), set by the user.
(67) now controls the PEEP pressure. Excess gases are evacuated through the AGS
system.
The ventilator airway pressure sensor P8
measures the PEEP pressure. This measured If the absorber is changed during use, the two
pressure controls the PWM signal to the PEEP absorbervalves (78) and (79) will bypass the
valve (67) keeping the PEEP pressure constant absorber connection, so that gases are not lost.
to the selected value.
The negative pressure relief valve (80) opens if
The driving gas is reduced to 3 bar in the the patient breathes spontaneous and the patient
pressure regulator (64), and led to the pilot breathing circuit is emptied for fresh gas (relief
valve (71). The pilot valve is controlled by the pressure is - 6 hPa).
ventilator electronics. During STBY the valve
is closed, and the Bag/vent valve (72) in the
integrated breathing system (22) is set to manual
Automatic Ventilation
ventilation. During automatic ventilation the Bag/vent
valve (72), controlled by the ventilator, is set
When the ventilator is set to VENT, the pilot to automatic ventilation.
valve (71) is opened and the Bag/vent valve
(72) in the integrated breathing system (22) The patient breathing system is filled with
is set to automatic ventilation. fresh gas from the fresh gas selector switch
(20), through the inspiratory limb, the
expiratory limb, and to the bellows (70).
Integrated Breathing System
Manual Ventilation When the bellows (70) is forced down during
During manual ventilation the Bag/vent valve inspiration, the gas will flow through the Bag/vent
(72), controlled by the ventilator, is set to valve (72), the absorber (74), the inspiratory
manual ventilation. valve (75), and the inspiratory limb to the patient.

The patient breathing system is filled with When the pressure in the chamber (68) is
fresh gas from the fresh gas selector switch released, the expiration gases flow through
(20), through the inspiratory limb, the the expiratory limb, expiration valve (76),
expiratory limb, and to the bag (73). the Bag/vent valve (72) to the bellows (70).

When the bag (73) is squeezed during inspiration, The inspiratory pressure is limited by the
the gas will flow through the Bag/vent valve ventilator high pressure alarm, set by the user.
(72), the absorber (74), the inspiratory valve
(75), and the inspiratory limb to the patient. Excess gases are evacuated through the
When the bag (73) is released, the expiration exhalation valve (69) to the AGS system,
gases flow through the expiratory limb, through the ventilator.

5-4 Oct. 08
Dameca Service Manual for Dameca MRI 508

Inspiratory/expiratory volumes can be measured During start-up and every 10 min. the two zero-
by a flow sensor (option), which can be placed valves (83 and 84) are activated by the ventila-
at the Y-piece (81) or the expiratory port (82). tor for app. 3 sec. in order to do zero calibration
of the P9 pressure sensor.
The flowsensor gives a pressure signal to the P9
pressure sensor of the ventilator. The ventilator The inspired O2 concentration can be measured
then calculates the actual patient flow over time by an external O2 fuel cell sensor (85), (option).
which gives a volume reading.

Oct. 08 5-5
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

5-6 Oct. 08
Dameca Service Manual for Dameca MRI 508

48
61
22 46 45
58
56 59 68
60 70 49
57 67

47

55 69
54 66 77
52
44
53 50
51
72 73

25 71
26 65 76 82

24 64
63
74 78
23
8 62
4
1
7 9 79 81

13
11 80 75
12 20
11 85
28 17 32 40
10 43 21 83 84
3
7 9 16 31 39 18
33 15 38
34
37

5 14 30 36
7 9
19

35 29

qusjob42.10650-whispa-mri-pne-manual.png
Fig. 5-1 Pneumatic diagram, Dameca MRI 508

Oct. 08 5-7
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

5-8 Oct. 08
Dameca Service Manual for Dameca MRI 508

5.3 Description Electronics


Power Supply Fuses
Mains power is supplied via fuses F11 &F12 to F1 T2A/T4A (230/115V)
the power supply box, with the switch mode F2 T2A/T4A (230/115V)
power supply and the Battery Charger. F4 T4A
F5 T4A
See section 10 for electrical diagrams. F6 T4A

Mains filter
Mains F1 Switch mode
230V or
100/115V F2 Power supply

Battery
charger

Power Supply Box

Display F5 / F6

Battery

F4
E-box

Ventilator

F1: T2A/T4A (230/115V)


F2: T2A/T4A (230/115V)
F4: T4A
F5: T4A
F6: T4A
qusjob3.siesta i whispa mri drawing.pdf

Fig. 5-2 Mains Power Diagram, Dameca MRI 508

Oct. 08 5-9
Service Manual for Dameca MRI 508 Dameca

5.4 Ventilator
The main components of the Whispa ventilator are:

O Power supply box (25) O Display unit (24) O Batteries (12)

O Electronics box (4) O Pneumatic manifold (5)








 
  
 
  









qusjob42.10659m-1000.pdf

Fig. 5-3

5-10 Oct. 08
Dameca Service Manual for Dameca MRI 508

Power Supply Box Pneumatic manifold

Mains power enters the power supply box, The pneumatic manifold controls the drive gas
which contains a battery charger and a flow to and from the bag in bottle.
switch-mode power supply.

A battery charger is continously controlling the


Hoses to pressure sensors
condition of the ventilator batteries. S1 and S2 : To ventilator manifold for
measurement of inspiratory
The switch-mode power supply controls the flow.
24V DC supply for the electronic box.
S3 : To ventilator manifold and then
to large connector on sidepanel,
Display unit for measurement of patient
The display unit is placed on the front panel volume.
as user interface, and is connected to the
electronics box with 3 wires, S4 : To ventilator manifold and then
to small connector on sidepa-
nel, for measurement of patient
Batteries volume.
A 24V DC battery is placed in the ventilator,
for back-up during a mains power failure. S5 : To IBS for measurement of
airway pressure.
Electronics box
The electronics box contains all pressure sensor
used by the ventilator, and the electronics
printboard controlling the ventilator.

Oct. 08 5-11
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

5-12 Oct. 08
Dameca Service Manual for Dameca MRI 508

6. Troubleshooting
6.1 General 6.2 Maximum Allowable Leaks
(according to ISO 5358)
Due to the construction and selection of
materials, problems are held at a minimum High pressure section 25 ml/min. (each gas)
in this high quality machine. Dose section 50 ml/min. (each gas)
Normally a carefully performed, planned Patient system 100 ml/min.
preventive maintenance is sufficient to keep
the machine in perfect working order for
many years.
Note:
If a problem should occur, this section of the
These values are absolute max.
manual can be a useful help to solve the problem.
values, normally the actual leaks
will be much less than these values,
The most common problems in anaesthesia
or no leaks at all.
machines are leaks. In order to locate a leak
there are some guidelines to be followed.

For tools and test equipment see section


1. Locate the gas that is leaking (normally 7. Maintenance.
only one gas is leaking in the high
pressure and dose section).
Warning:
2. Locate the leaking section.
Do not lubricate any part of the pneu-
matic system with any type of combu-
stible lubricant as this would lead to a
3. Use the corresponding tables in this section
danger of explosion. Use only special
of the manual for most common problems.
grease i.e. DAMECA P/N 36825.

A very useful aid in locating leaks is leak-finder 6.3 Troubleshooting Pneumatics


spray, or a DAMECA test suitcase.
Table 6-1 AIR leaks in the high pressure section.
Table 6-2 N2O leaks in the high pressure section.
Table 6-3 Troubleshooting the high pressure
Warning: section.
Table 6-4 Troubleshooting the dose section.
Avoid spraying leak-finder spray
Table 6-5 Troubleshooting the AGS system.
on electrical or electronic wires
Table 6-6 Troubleshooting the Bag-in-Bottle
and components.
system.
Sect. 6.4 Troubleshooting electronics.

Oct. 08 6-1
Service Manual for Dameca MRI 508 Dameca

Table 6-1 Troubleshooting the O2/AIR supply for leaks in the high pres. sect.

Item Possible Fault Remedy


Quick connectors Valve seat leaking Change connector
O-ring leaking
AGSS ON/OFF valve Leaking Change valve
O2 flush valve Leaking Change valve
O2 flowmeter Leaking/Other problems Change flowmeter and
flowmeter front foil

Table 6-2 Troubleshooting the N2O supply for leaks in the high pressure section

Item Possible Fault Remedy


Quick connectors Valve seat leaking Change connector
O-ring leaking

Table 6-3 Troubleshooting the high pressure section


Item Possible Fault Remedy
Pressure gauges Zero or wrong pressure displayed, Change pressure gauge
pointer does not »0« at no pressure
Check valves Pressure drops to zero when Change check valve
disconnecting inlet hose from
wall outlet
Pressure regulators To low pressure Increase/Decrease inlet pressure
To high pressure Check reduced pressure,
use P/N 72231, and adjust
Reduced pressure increases Change Valve P/N 33005-73
(internal leakages) Valveseat P/N 33005-72
Membrane P/N 33005-71

6-2 Oct. 08
Dameca Service Manual for Dameca MRI 508

Table 6-4 Troubleshooting the dose section


Table
Item 6-4 Troubleshooting
Possiblethe dose section
Fault Remedy
Back bar valves Leaking gaskets or O-rings Change gaskets and O-rings
(grease slightly)
Aux. freshgas outlet Leaking Change O-rings (grease slightly)

Table 6-5 Troubleshooting the the AGS system


Item Possible fault Remedy
Ejector Too little flow Change ejector filter
Change nozzle
Check pressure to ejector
(4-4.5 (x100kPa))
Too big resistance at the AGS
outlet connector, due to too
long or too narrow tubing
from outlet to wall/column
connection.
Flowmarker Ball invisible Change flowmarker
Reservoir Reservoir blocked Change reservoir tubing and
clean inlet opening.

Table 6-6 Troubleshooting the Integrated Breathing System


Item Possible fault Remedy
IBS Leakage between IBS and base Change O-rings and check the
surface of the base and IBS.
Bellows Bellows penetrated or deformed Change the bellows
Spill valve Leaking membrane Change membrane
Defective membrane seat Change seat
O-ring on base Leakage Change O-ring
Absorber valves Leakage when absorber mounted Change seal between valve
and absorber
Leakage when absorber not mounted Change valve

Oct. 08 6-3
Service Manual for Dameca MRI 508 Dameca

6.4 Troubleshooting Electronics


The ventilator cannot be repaired on component level except for changing fuses.

If there is a problem with one of the units, it must be exchanged. External components such
as proportional valves, flow sensors, and other external parts are available as spare parts.

6.4.1 Troubleshooting Ventilator

Main fuses
The two main fuses are placed on the power supply box, and can be changed without
opening the box. The fuses are (Fuse 1 and 2):

O 115V version : T4A (P/N 40055-4AT).


O 230V version : T2A (P/N 40055-2AT).

Inside the ventilator box (inside the machine) is placed a black plastic box in which
the fuses for the ventilator components are placed. Open the box by lifting the lid in one end.

The fuses are:

Fuse 4: Protection of power supply : T4A (P/N 40055-4AT).

Fuse 5: Protection of battery charger : T4A (P/N 40055-4AT).

Fuse 6: Protection of battery : T4A (P/N 40055-4AT).

Caution:
Use only fuses of same type and rating.

6-4 Oct. 08
Dameca Service Manual for Dameca MRI 508

6.5 Changing Mains Power Voltage


If the mains power to the machine is changed from 230V to 110/115V, or from 100/115V to 230V,
the machine must be modified by the following procedure:

1. Disconnect the machine from mains power and driving gas.

2. Change main fuses on the power supply box. 110/115V: T4A (P/N 40055-4AT)
230V: T2A (P/N 40055-2AT)

Labels Mains cable:


Mount new labels at the power supply box. 110/115V: 11395-110
230V: 11395-230

Reconnect the machine to Mains and driving gas.

Carry out a functional check of the machine as described in the User Manual.

Oct. 08 6-5
Service Manual for Dameca MRI 508 Dameca

6.6 Changing Drivegas


If the drivegas for the ventilator, patient suction 3. The hose going from the inlet manifold con-
(option) and active AGSS (option) should be nection called »Drivegas« to the drivegas
changed from AIR to O2 or from O2 to AIR, block connection called »Inlet« is the drivegas.
the machine must be modified by the Connect O2 hose (clear) or AIR hose (black)
following procedure: as desired. For the hose which is not the
drivegas, the »Drivegas« connection on the
1. Locate the drivegas manifold placed in the inlet manifold must be blocked.
top section of the machine, see below:
4. If the drivegas has been changed from AIR
2. Disconnect O2- and AIR-supply to the to O2, or from O2 to AIR, the »ventilator
machine and release the pressure inside insp. flow« must be re-calibrated, as
the machine. described in section 8.

qusjob42.10659-40 manual-2.png

Fig. 6-1 Drivegas block

6-6 Oct. 08
Dameca Service Manual for Dameca MRI 508

6.7 Changing the Setup


If the setup of the Dameca MRI 508 should be changed, these changes can be made
in the service menu.

Enter super user password (26-23-8) or technician password (91-34-80) in order to get access
to the service menu.

Oct. 08 6-7
Service Manual for Dameca MRI 508 Dameca

The following settings can be changed in the service menu:


O2 sensor
Select »O2 sensor« <ETR>.
Select »Yes« <ETR> if the machine should measure insp. O2 concentration using an external
O2 fuel cell sensor (placed on a T-piece at the insp. cone of the IBS).
Select »No« <ETR> if the machine should not have any measurements of O2 concentrations.

SIMV
Select »SIMV« <ETR>.
Select »Yes« <ETR> if the machine should be used with SIMV.
Select »No« <ETR> if the machine should NOT be used with SIMV.

PSV
Select »PSV« <ETR>.
Select »Yes« <ETR> if the machine should be used with PSV.
Select »No« <ETR> if the machine should NOT be used with PSV.
Note: PSV can only be used if volume sensor is set to »full spirometry«.

N2O
Select »N2O« <ETR>.
Select »Yes« <ETR> if N2O can be present in the freshgas flow from the machine.
Select »No« <ETR> if N2O is NEVER present in the freshgas flow from the machine.

Volume
Select »Set volume« <ETR> in order to change the volume setting of the ventilator.
Select »Volume sensor« <ETR> in order to change the volume measurement of the ventilator.

Set volume
Select »Set volume« <ETR>.
Select »TV« <ETR> if the ventilator volume set by the user should be »tidal volume«.
Select »MV« <ETR> if the ventilator volume set by the user should be »minute volume«.

Volume sensor
Select »Volume sensor« <ETR>.
Select »Full spirometry« <ETR> if the volume measurement should include both inspired a
nd expired volumes (volume sensor placed at the Y-piece).
Select »Only exp. volume« <ETR> if the volume measurement should only include expirered
volumes (volume sensor placed at the exp. cone of the IBS).
Select »No volume sensor« <ETR> if the machine should not have any volume measurement.

Language
Select »Language« <ETR> in order to change the language of the user interface
(not the service menu).

Languages will be listed here, as they become available.

6-8 Oct. 08
Dameca Service Manual for Dameca MRI 508

Enter the technician password (91-34-80) in order to get access to the full service menu.

Oct. 08 6-9
Service Manual for Dameca MRI 508 Dameca

Analog calib
Select »Analog calib.« <ETR> in order to get access to the calibration menu of the ventilator.
See section 8 regarding calibration of the ventilator.

Regulator param
The regulator parameters are used for controlling the electronic valves of the ventilator.
These parameters should not be changed, but should be as shown in the menu figure.

Version
Select »Version« <ETR> in order to see the software version of the ventilator display unit
(called »MMI«) and the ventilator electronic box (called »CPU«).

An »*« is shown after »Whispa«, in the top of the menu. This setting must NOT be changed.

6-10 Oct. 08
Dameca Service Manual for Dameca MRI 508

7. Maintenance
7.1 Maintenance, General
In order to assure maximum safety and reliability
of all functions of this machine, planned preven-
tive maintenance must be carried out once a
Warning:
year as a minimum by an authorized technician. Use of standard oil or grease in this
machine could cause an explosion.
Use special grease P/N 36825 only.
Before any maintenance or repair is started the
machine must be disconnected from the mains,
and all gas pressures must be released. Be sure
to release the N2O pressure before the O2 pressure.
Caution:
Do not open the Electronic box of the
Due to safety and performance reasons, it is ventilator. Opening the box void the
highly recommended not to repair or disassemble warranty of the ventilator.
the proportional regulator. The regulator can be
repaired by DAMECA a/s only.

After any maintenance or repair, a performance Caution:


check should be carried out, and a post mainte- Vaporizers should be serviced at least
nance/repair check list filled in. once a year by an authorized service
centre.

Always use the correct tools to avoid damage


to screws, nuts, and other parts.

Important:
Always check the function of the parts being After maintenance and/or repair a
serviced. full function check should be
carried out according to the post
maintenance/repair check list.
Protect the table plate with a piece of carpet
or similar.

Oct. 08 7-1
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

7-2 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.2 Tools and Test Equipment

Standard tools Part No.


1 set of Allen keys 1.5 - 10 mm
1 set of angled Allen keys 1.5 - 10 mm
1 set of Philips screwdrivers No. 0 - 1 - 2
1 set of screwdrivers 3.5 - 4 - 5 mm
1 set of spanners 6 - 24 mm

Special tools Part No.


Soft claw spanner 72203
Special tool for plastic tubing ø5/3 72205
Special tool for plastic tubing ø6/4 72206
Pressure gauge 72231
Special tool for pressure regulator 72230
Mount tool for guidance ring 72235-01
Mount tool for guidance ring 72235-02
Special tool for change over valve 72236
Special tool for plugs (IBS) 72236-10
Special tool for BAG connection piece 72236-12
Special tool for 11960-30 72237-10
Special tool for 11960-30 72237-20

Special tools Part No.


Test suitcase 72250
Mallinckrodt PTS-2000, TSI Certifier FA system w/ model 4070
Timeter or similar flow, volume, pressure and O2% measuring device
Digital voltmeter
Test instrument for leakage current

Oct. 08 7-3
Service Manual for Dameca MRI 508 Dameca

7.3 Service Kits


7.3.1.1 Annual Service Kit for 3 rotameters P/N 36459-1

Item Description Qty. Part no.

1. Flowmarker 25 L/min 1 13676


2. Filter 3 11213
3. Valve housing w/filter w/non-return valve 2 11215
4. O-ring ø13,95x2,62 4 36300
6. Filter white for fan 1 36715-01
7. O-ring Ø18,1x1,6 EPDM 5 36267
11. O-ring, silicone red, ø15,08x2,62 1 36344
12. O-ring Ø25x2.5, EPDM 1 36252
13. O-ring, silicone red, ø9,19x2,62 2 36343
14. O-ring ø32.99x2.62, EPDM 1 36346
15. O-ring, silicone blue, ø129,77x3,53 1 36330
16. O-ring, EPDM, ø49,6x2,40 1 36261-4924
19. Membrane 1 51025-08
20. One-way valve 2 11890-31
21. Valve dome kit 1 11890-40
22. O-ring Ø9,1x1,6 EPDM 2 36271
23. O-ring Ø7,65x1,78 EPDM 1 36309
24. Filter bronze 1 36132
25. Transparent tube, 1200mm 1 23101
27. Filter bronze Ø9,8xØ3x21mm 3 36704
28. O-ring ø6,07x1,78 teflon 3 36311
29. Silicone gasket Ø15,5xØ11x2mm 6 36505
30. Silicone gasket Ø16,1xØ7x2mm 6 36509
31. Gasket Ø21xØ16x1mm EPDM 3 36513
37. Changeover valve for IBS 1 11960-550
38. Expiratory valve for absorber 1 11960-460
39. Inspiratory valve for absorber 1 11960-230

7-4 Jan. 10
Dameca Service Manual for Dameca MRI 508

1,67

 3DWLHQWEUHDWKLQJV\VWHP
 
 
 )LOWHU 

 

,%6ERG\

<RNH
%DFNEDU
 
  

$*6 
 


5RWDPHWHU 9DOYH$  $3/YDOYH

9DOYH%


9DOYH& 
,%6EDVH

qusjob42.36459-1.pdf

Fig. 7-1 Service kit Dameca MRI 508, 3 rotameters

Jan. 10 7-5
Service Manual for Dameca MRI 508 Dameca

7.3.1.2 Annual Service Kit for 5 rotameters P/N 36459-15

Item Description Qty. Part no.

1. Flowmarker 25 L/min 1 13676


2. Filter 3 11213
3. Valve housing w/filter w/non-return valve 2 11215
4. O-ring ø13,95x2,62 4 36300
6. Filter white for fan 1 36715-01
7. O-ring Ø18,1x1,6 EPDM 5 36267
11. O-ring, silicone red, ø15,08x2,62 1 36344
12. O-ring Ø25x2.5, EPDM 1 36252
13. O-ring, silicone red, ø9,19x2,62 2 36343
14. O-ring ø32.99x2.62, EPDM 1 36346
15. O-ring, silicone blue, ø129,77x3,53 1 36330
16. O-ring, EPDM, ø49,6x2,40 1 36261-4924
19. Membrane 1 51025-08
20. One-way valve 2 11890-31
21. Valve dome kit 1 11890-40
22. O-ring Ø9,1x1,6 EPDM 2 36271
23. O-ring Ø7,65x1,78 EPDM 1 36309
24. Filter bronze 1 36132
25. Transparent tube, 1200mm 1 23101
27. Filter bronze Ø9,8xØ3x21mm 3 36704
28. O-ring ø6,07x1,78 teflon 3 36311
29. Silicone gasket Ø15,5xØ11x2mm 1 36505
30. Silicone gasket Ø16,1xØ7x2mm 3 36509
31. Silicone gasket Ø10,9xØ5x2mm 4 36503
32. Silicone gasket Ø16,1xØ12,7x2mm 5 36517
33. Silicone gasket Ø16,9xØ7x2mm 3 36508
34. O-ring Ø8,1x1,6 EPDM 4 36304
35. O-ring Ø14,1x1,6 EPDM 6 36307
36. Silicone gasket Ø15,5xØ11x2mm 4 36502
37. Changeover valve for IBS 1 11960-550
38. Expiratory valve for absorber 1 11960-460
39. Inspiratory valve for absorber 1 11960-230

7-6 Jan. 10
Dameca Service Manual for Dameca MRI 508

1,67
 3DWLHQWEUHDWKLQJV\VWHP




 )LOWHU 

 

,%6ERG\

<RNH
%DFNEDU 

  


$*6 
 
 
 
5RWDPHWHU
9DOYH$ $3/YDOYH
 


9DOYH% 

 9DOYH&
,%6EDVH

qusjob42.36459-15.pdf

Fig. 7-2 Service kit Dameca MRI 508, 5 rotameters

Jan. 10 7-7
Service Manual for Dameca MRI 508 Dameca

7.3.2 4-Year Service Kit P/N 36459M-4

Item Description Qty. Part no.

1. Proportional regulator with 1 11780-50


plug- in connection
2. Membrane, black 4 33005-71
3. Valve seat 4 33005-72
4. Valve 4 33005-73
5. Friction ring 4 33005-75
6. Filter bronze, Ø8x2mm 3 36701
7. Battery 12V 2,2Ah lead-acid 2 40099-62
8. Filter housing 1 11213
9. O-ring Ø5,1x1,6 3 36269

5
5 4
3
4 2
3
2 O2 inlet manifold
Gas selector block
1

Inlet manifolds

9
5 7
4
3
2 6
O2 alarm
Rotameterbox

5
4 Ventilator
3
2
8

qusjob42.36459M-4 man.pdf

Fig. 7-3 Service kit Dameca MRI 508, 4 years

7-8 Jan. 10
Dameca Service Manual for Dameca MRI 508

7.4 Annual Maintenance

Caution:
Before any maintenance or repair is started the machine must be disconnected
from the mains, and all gas pressures must be released. Be sure to release the
N2O pressure before the O2 pressure.

Note:
Special grease P/N 36825, 36810 or 36811 should be applied to all
new O-rings, before they are used.

7.4.1 Central Gas Inlets


7.4.2 Inlet manifolds

2
3 Pin Pin
22

2
O2
Pin

3 Pin 2
AIR
22 Pin

N2 O

qusjob42.31350, 31350A, 31350C manual.pdf


Fig. 7-4 Inlet manifolds
1. Loosen the union nut. 6. Change the Check valve with filter (3)
2. Remove the zeeger circlip ring. P/N 11215 and O-ring (22) P/N 36271,
3. Remove the filter by pulling the pin. (only if the machine is equipped with yokes
4. Change the filter without check valve (2) for spare gas cylinders). Place new check
P/N 11213. Place new filter with pin facing valve with pin facing inside the manifold.
the circlip ring. 7. Reassemble the manifold, and carry out a
5. Mount the zeeger circlip ring. high-pressure leak test.

Oct. 08 7-9
Service Manual for Dameca MRI 508 Dameca

7.4.3 Pin index yokes 7.4.5 Evacuation (AGSS)

qusjob42.940fi7-6.pcx

Fig. 7-5 Pin index yoke for spare


cylinder

Change gaskets: 1
O O2 gasket P/N 36610 (red).
O N2O gasket P/N 36609 (black).

These gaskets are not included in the servicekit.

7.4.4 Back-bar

qusjob42.13680-36457-1.pdf

O Change flowmarker (1), P/N 13676.


O Change reservoir tubing (25), P/N 10653-60
placed above the ventilator.
O Activate the AGSS flow and check that the
yellow ball is visible.

If the machine is equipped with active scaven-


ging system, using an ejector to generate AGSS
flow, the following parts must be changed in the
qusjob42.11080-650-36457-1.pdf
ejector unit, placed in the middle section of the
Change O-rings on top of valves (4), P/N 36300. machine, to the right (seen from behind).

7-10 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.4.6 Ejector unit





qusjob42.11013manual.pdf

O Loosen the top of the ejector unit.


O Remove the filter and the O-ring.
O Change the filter (24) P/N 36132 and the O-ring (23) P/N 36309.
O Mount the top of the ejector unit.
O Activate AGSS flow and check that the yellow ball is visible.

7.4.7 Ventilator
O Change filter for lower fan (6) (on the back of the machine) P/N 36715-01.

Oct. 08 7-11
Service Manual for Dameca MRI 508 Dameca

7.4.8 Integrated Breathing System


O Disassemble the IBS unit as shown on the
figure.

O Change spill-valve membrane (19),


P/N 51025-08. Weight on top of old spill
valve membrane should be used on new
membrane.

O Change O-ring on cover for membrane



(16), P/N 36261-4924.
 
O Change blue O-ring for Bag in Bottle
chamber (15), P/N 36330. Old O-ring can be
  removed by using a screwdriver. Be sure
that the new O-ring is positioned correctly
in the groove of the IBS body.

 O Change red O-ring (12), P/N 36252 and red


 O-ring (14), P/N 36346 on the APL valve.

O Change 5 pcs. O-ring (7), P/N 36267; on


 insp. cone, on exp. cone, on cone for respi-
  ration bag, on connection for insp. valve
and on connection for exp. valve.

 O Change insp. and exp. valve (20),


 P/N 11890-31.

 O
 Change dome for insp. and exp. valve (21)
kit, P/N 11890-40.

O On IBS base, change O-ring (11),


P/N 36344 and 2 pcs. red O-ring (13),
P/N 36343.

O Change valve A, B and C, (37, 38 and 39).


qusjob42.11960-36457-1.pdf

7-12 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.4.9 Rotameter Box

1. Remove the rotameter tubes and change O-rings, filters, seals etc. (pos. 27, 28, 29, 30 and 31).

2. If the machine is installed with rotameterbox with 5 pcs. rotametertubes, additional O-rings and
seals (pos. 32, 33, 34, 35 and 36) must be changed.

3. Clean the rotameter tubes and float (rotor).








 



 


 



 

 

 

 







qusjob42.10654-75 36459-1 pdf qusjob42.10774-750 36459-15.pdf

Oct. 08 7-13
Service Manual for Dameca MRI 508 Dameca

7.5 Post Maintenance Check


Perform post maintenance/repair check as described in section 7.8

7.6 Four Year Maintenance


Additional to annual maintenance, perform the following:

7.6.1 Inlet Manifolds

N2O AIR

9 9

6 6

qusjob42.31350A-36457-4.pdf qusjob42.31350C-36457-4.pdf

O Change Filter (6), P/N 36701 and O-ring (9), P/N 36269.

7-14 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.6.2 O2 Inlet Manifold 7.6.3 Drivegas Block


2

9
3

6
4

qusjob42.31350-36457-4.pdf qusjob42.10659-40 36457-4.pdf

O Change Membrane (2), P/N 33005-71. O Change Membrane (2), P/N 33005-71.
O Change Valve seat (3), P/N 33005-72. O Change Valve seat (3), P/N 33005-72.
O Change Valve (4), P/N 33005-73. O Change Valve (4), P/N 33005-73.
O Change Friction Ring (5), 33005-75. O Change Friction Ring (5), P/N 33005-75.
O Change filter (6), P/N 36701 and O-ring
(9), P/N 36269.

Oct. 08 7-15
Service Manual for Dameca MRI 508 Dameca

7.6.4 O2 Alarm

qusjob42.10654-30 36457-4.png

O Change Membrane (2), P/N 33005-71. 7.6.5 Rotameterbox


O Change Valve seat (3), P/N 33005-72.
O Change Valve (4), P/N 33005-73. O Change proportional regulator (1),
O Change Friction Ring (5), P/N 11785-50. Return the old proportional
P/N 33005-75. regulator to Dameca.

7-16 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.6.6 Ventilator

Pin 8

4
3
2
5

5-36457-4.SLDDRW

qusjob42.10987-25-36457-4.pdf

O Change Membrane (2), P/N 33005-71. 7.6.7 Integrated Breathing System


O Change Valve seat (3), P/N 33005-72.
O Change Valve (4), P/N 33005-73. O Change the APL valve (10), P/N 11960-30.
O Change Friction Ring (5), P/N 33005-75.
O Change Battery (7) P/N 40099-67, placed
on the ventilator base plate.
O Change filter without check valve (8),
P/N 11213. The pin of the filter must
face away from the manifold.

Oct. 08 7-17
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

7-18 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.7 Service Menu


Select MAIN menu in order to get access to the service menu.

In the MAIN menu, select SERVICE.

The password for a super user is: The password for a technician is:
26 91
23 34
8 80

Oct. 08 7-19
Service Manual for Dameca MRI 508 Dameca

7-20 Oct. 08
Dameca Service Manual for Dameca MRI 508

Oct. 08 7-21
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

7-22 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.8 Post Maintenance/Repair Check List


Connect the machine to central gas outlets (O2, AIR and N2O), VAC and AGSS.

7.8.1 Internal reduction valves


Check the adjustment of all the internal reduction valves, by using test instrument P/N 72231.

qusjob42.72231.pdf
Special tool for pressure regulator (P/N 72231)

The valves should be adjusted as:

A) 1 pcs. at O2 inlet manifold: 4 - 4.5 (x100 kPa)


B) 1 pcs. at drivegas selector block: 4 - 4.5 (x100 kPa)
C) 1 pcs. at ventilator manifold: 2.8 - 3.2 (x100 kPa)

If necessary re-adjust the reduction valves and check that the pressure is stable after adjustment,
and does not rise slowly.

7.8.2 O2 alarm adjustment


Open O2 fresgas flow and disconnect O2 supply to the machine.
Close O2 freshgas flow when the O2 pressure gauge on front of the machine reads 2.5 (x100kPa).
Adjust the pressure regulator (D) for the O2 alarm until the O2 alarm (whistle) sounds.
Connect O2 supply to the machine, open O2 freshgas flow and check that the O2 alarm starts
at app. 2.5 (x100kPa). If not, re-adjust.

Oct. 08 7-23
Service Manual for Dameca MRI 508 Dameca

3UHVVXUHJDXJH

'

3UHVVXUHJDXJH
GLVFRQQHFWHGH[LVWLQJKRVH

3UHVVKHUHWR
FKHFNSUHVVXUH

&
& &

qusjob42.10651MRI-00 manuel 02.pdf


Dameca MRI 508 reduction valves

7-24 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.8.3 High pressure leakage O2 and AIR

Turn ON the machine and disconnect N2O supply to the machine.


Bypass the selftest, open O2 and N2O freshgas flow until the N2O flow goes to zero.

Then set Change-over valve to AIR, close O2 freshgas flow and connect a flowmeter
(0-250 ml/min) between the central gas supply outlet and the O2 and AIR gas supply
hoses to the machine.
Alternatively, connect the O2 and AIR gas supply hoses to a Dameca test suitcase.

Check that the leakage for both gases is less than 25 ml/min.

Leakage:_________ml/min

Connect all gas supply hoses of the machine to the central gas outlets.

7.8.4 High pressure leakage N2O


Disconnect AIR supply to the machine.

Open AIR freshgas flow until the AIR flow goes to zero.

Then set Change-over valve to N2O and connect a flowmeter (0-250 ml/min) between the
central gas supply outlet and the N2O gas supply hose to the machine. Alternatively, connect
the N2O gas supply hose to a Dameca test suitcase.

Check that the leakage is less than 25 ml/min

Leakage:_________ml/min.

Connect all gas supply hoses of the machine to the central gas outlets.

7.8.5 Dose system leakage


Activate the auxiliary freshgas outlet, open vaporizer (if present) and connect a flowmeter
(0-250 ml/min) and a pressure gauge (0-100 cm H2O) to the auxiliary freshgas outlet.

Open the flowmeter and adjust until the pressure is stable at 30 cm H2O.

Check that the flowmeter indicates a flow (leakage) of less than 25 ml/min.

If the machine is not equipped with an auxiliary freshgas outlet, this test should not be performed.

Set the freshgas switch to »circle system«.

Oct. 08 7-25
Service Manual for Dameca MRI 508 Dameca

7.8.6 Selftest

Turn the machine OFF, and turn it ON again.


Perform the selftest according to the instructions on the display.
Check that no error messages appear during the selftest.

Leakage in VENT system:_________ml/min

If the ventilator says that the freshgas flow is too high or too low when the freshgas flow is
set to 5 L/min AIR or O2, the freshgas flowsensor must be re-calibrated (see section 8.4).

7.8.7 Auxiliary freshgas outlet


If the machine is equipped with an auxiliary freshgas outlet, set the »freshgas switch« to
»auxiliary outlet«.

Open the Y-piece of the patient system to ambient air.

Set the freshgas flow to 10 L/min and connect reference flow measuring instrument to BAG
connection on the IBS and check that measured flow is 0.00 L/min.

Set the »freshgas switch« to »circle system«.

Connect reference flow measuring instrument to the auxiliary freshgas outlet and check that
measured flow is 0.00 L/min.

If this test is not OK, the freshgas selector valve could be defective.

Close the freshgas flow.

7.8.8 Change-over valve AIR


If the machine is equipped with change-over valve for AIR and N2O, the following test
must be carried out.

Set AIR freshgas flow to zero, set change-over valve to N2O and set a freshgas flow
of 5 L/min for O2 and N2O.

Set change-over valve to AIR and check that the AIR flow on the rotameter goes to zero.

Set change-over valve to N2O and check that the N2O flow again shows 5 L/min.

7-26 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.8.9 Change-over valve N2O

If the machine is equipped with change-over valve for AIR and N2O, the following test must
be carried out

Set N2O freshgas flow to zero, set change-over valve to AIR and set a freshgas flow of 5 L/min
for O2 and AIR.

Set change-over valve to N2O and check that the N2O flow on the rotameter goes to zero.

Set change-over valve to AIR and check that the AIR flow again shows 5 L/min.

Close the freshgas flow.

7.8.10 N2O slave function


Set freshgas flow to 1 L/min. for O2 and N2O.
Disconnect O2 supply to the machine and check that N2O flow on the rotameter goes to zero,
when the O2 flow goes to zero.

7.8.11 O2 alarm
Set O2 freshgas flow to 5 L/min and disconnect the O2 gas supply to the machine.
Check that the »O2 alarm (whistle)« alarm starts when the pressure on the O2 pressure
gauge on front of the machine is app. 2.5 (x100 kPa).

If not, the O2 alarm must be re-adjusted (see section 7.8.2).

Connect the O2 gas supply to the machine.


Close the freshgas flow.

PS: O2 supply hose must be equipped with a male-connector with one-way valve, in order for
this test to be carried out.

7.8.12 Drivegas alarm


Connect a testlung (for example a respiration bag) to the Y-piece and start the ventilator with
default settings and tidal volume set to 200 ml. Fill bellows by pressing the O2 flush.

Disconnect the drivegas (AIR or O2) supply hose to the machine and check that the »drivegas low«
alarm starts when the pressure of the drivegas is 2.0 - 2.8 (x100 kPa), by reading the pressure gauge
on front of the machine.

Oct. 08 7-27
Service Manual for Dameca MRI 508 Dameca

Connect the drivegas (AIR or O2) supply hose to the machine to the central gas outlet and check
that the ventilator starts to function again, and that all alarms disappear.

If this test indicates a malfunction, the drivegas alarm must be re-calibrated (see section 8.7).

7.8.13 Fans
Check that lower fan is running.

7.8.14 AGSS
Activate the AGSS flow and check that the yellow ball in the flowmarker is visble.

7.8.15 O2 flush
Activate the auxiliary freshgas outlet and connect a reference flow measuring instrument to
the outlet.
If the machine is not equipped with an auxiliary freshgas outlet, connect the reference flow
measuring instrument to the insp. cone on the IBS, set APL valve to max. and block the cone
for the respiration bag on the IBS.

Activate the O2 flush and check that the flow is 35 to 50 L/min.

Connect an oxygen analyzer and check that the flow is 100 % O2.

Set the freshgas switch to »circle system«.

7.8.16 MPL valve


If the machine is equipped with an auxiliary freshgas outlet, the following test must be carried out.

Activate the auxiliary freshgas outlet and connect a reference pressure measuring instrument
(0-100 cm H2O) to the outlet.

Set O2 freshgas flow to 2 L/min and check that the pressure is 90-100 cm H2O when the outlet is
blocked.

If not, the MPL valve placed on the auxiliary freshgas outlet block (facing the inside of
the machine) must be adjusted.

Set the freshgas switch to »circle system« and close the freshgas flow.

7-28 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.8.17 Minoxyguard

Activate the auxiliary freshgas outlet and connect a reference instrument for measurement of
O2 % to the outlet.

If the machine is not equipped with an auxiliary freshgas outlet, connect the reference instrument
to the insp. cone on the IBS, set APL valve to max. and block the cone for the respiration bag
on the IBS.

Set change-over valve (if present) to N2O and set N2O freshgas flow regulator knob to max.

Set O2 freshgas flow to 0.5 L/min and wait for stable O2 % reading on the reference instrument.
The O2 % must be 23 - 29. If not re-adjust the minoxyguard as described in section 8.8.

Set O2 freshgas flow to 2.0 L/min and wait for stable O2 % reading on the reference instrument.
The O2 % must be 22 - 28. If not re-adjust the minoxyguard as described in section 8.8.

Set O2 freshgas flow to 4.0 L/min and wait for stable O2 % reading on the reference instrument.
The O2 % must be 24 - 30. If not re-adjust the minoxyguard as described in section 8.8.

7.8.18 Rotameters
Check that all freshgas flows for O2, AIR and N2O can be adjusted from 0 to 10 L/min, and
that the rotors of the rotameters rotate freely, from 0 to 10 L/min.

7.8.19 Integrated patient suction


If the machine is equipped with integrated patient suction, the following test must be carried out:

Activate the patient suction and set it to max.


Block the suction outlet (placed on the side of the machine) and check that the patient suction
vacuum gauge indicates a vacuum of at least -0.7 (x100 kPa).

If not, and the patient suction is powered by AIR or O2, the suction unit placed under the stand
of the machine could be defective.

Turn off the patient suction.

Oct. 08 7-29
Service Manual for Dameca MRI 508 Dameca

7.8.20 Auxiliary O2 flowmeter

If the machine is equipped with an auxiliary O2 flowmeter, the following test must be carried out.

Connect a reference flow measuring instrument to the outlet of the auxiliary O2 flowmeter.
Set the flowmeter to 5 L/min O2 (reading is middle of the ball) and check that the measured
flow is 4.3 to 5.7 L/min.

Connect an oxygen analyzer and check that the flow is 100 % O2.

Close the auxiliary O2 flowmeter.

7.8.21 IBS
Place a testlung (f.ex. a respiration bag) at the Y-piece.
Start the ventilator with default settings, fill bellows by pressing the O2 flush and set
freshgas flow to 0.3 L/min.
Disconnect the CO2 absorber from the IBS and check that both absorber valves close.
Check that the bellows still goes to the top of the bag in bottle chamber, during expiration.
Connect the CO2 absorber to the IBS.

7.8.22 Ventilator function


Set ventilator to STBY.
With the Y-piece open, enter the main menu, go to spirometry sub-menu and set the sensor to
»Paed.« <ETR> and then set the sensor to »Adult« <ETR>.

Connect a testlung (for example a respiration bag) to the Y-piece and start the ventilator with
default settings, incl. tidal volume 500 ml.
Set high alarm to 80 and fill the bellows by pressing the O2 flush.

After 5 respirations read the TV exp shown on the display.


TV exp before:_________ml
Set respiration rate to 4, enter the main menu, go to the spirometry sub-menu and set sensor to
»Paed.« <ETR>.
Then set sensor back to »Adult« and press <ETR> WHILE the ventilator is doing an inspiration.

Set respiration rate to 12, wait 3 respirations and check that the measured TV exp shown on the
display is the same as before (within +/- 30 ml).
TV exp after:_________ml

If not, the zero valves for the volume measurement might be leaking.

Set ventilator to STBY

7-30 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.8.23 Tidal volume

Connect a reference volume measuring instrument to the INSP cone on the IBS.
If the drivegas for the ventilator is AIR, the volume measuring instrument must be set to AIR.
If the drivegas for the ventilator is O2, the volume measuring instrument must be set to O2.
If a Mallinckrodt PTS 2000 is used, the trigger flow for volume measurement should be set to 5 L/min.

Remove the bellows in the bag in bottle and mount the chamber.

Start the ventilator with default settings and set freshgas flow to zero.

Set tidal volume to 200 ml and wait 3 respirations.


Read the measured tidal volume on the instrument:_________ml
The volume should be 180 to 220 ml.
If not, the inspiratory flow must be re-calibrated (see section 8.1).

Set tidal volume to 500 ml and wait 3 respirations.


Read the measured tidal volume on the instrument:_________ml
The volume should be 475 to 525 ml.
If not, the inspiratory flow must be re-calibrated (see section 8.1).

Set tidal volume to 700 ml and wait 3 respirations.


Read the measured tidal volume on the instrument:_________ml
The volume should be 675 to 725 ml.
If not, the ventilator insp. flow must be re-calibrated (see section 8.1).

Set ventilator to STBY and mount the bellows in the bag in bottle.

7.8.24 Paed. volume measurement


Connect a reference volume measuring instrument between the Y-piece and the EXP cone on the
IBS (flow direction: from Y-piece to EXP cone).

Set the volume measuring instrument to O2.


If a Mallinckrodt PTS 2000 is used, the trigger flow for volume measurement should be set to
5 L/min.

Place a testlung (NOT a respiration bag) at the Y-piece. Place a PAED. volume sensor in the patient
system. Remember to place a filter between the testlung and the volume sensor.

Go to main menu and to sub-menu »spirometry« and set sensor to PAED. <ETR>.

Start the ventilator with default values and respiration rate set to 10.

Oct. 08 7-31
Service Manual for Dameca MRI 508 Dameca

Fill the bellows by pressing the O2 flush and set freshgas flow to 10 L/min O2.

Connect an O2 sensor to exp. cone on IBS and wait until O2% is higher than 95.
Close freshgas flow, wait 3 respirations and check that O2% is still higher than 95.

Set tidal volume to 200 ml, wait 3 respirations and read the measured expired tidal volume
on the ventilator:_________ml
on the instrument:_________ml

These volumes should be identical, within +/- 20 ml.

If not, the Paed. volume sensor must be re-calibrated (see section 8.3.2).

Set ventilator to STBY.

7.8.25 Adult volume measurement


Place an ADULT volume sensor in the patient system Remember to place a filter between the
testlung and the volume sensor.

Go to main menu and to sub-menu "spirometry" and set sensor to ADULT <ETR>.

Start the ventilator with default values and respiration rate set to 10.
Set high alarm to 80.

Fill the bellows by pressing the O2 flush and set freshgas flow to zero.

Check that O2% at exp. cone on IBS is still higher than 95. If not, open O2 freshgas flow
and after a while, close the flow and check that O2% is still higher than 95.
Set tidal volume to 500 ml, wait 3 respirations and read the measured expired tidal volume;
on the ventilator:_________ml
on the instrument:_________ml
These volumes should be identical, within +/- 50 ml.
If not, the Adult volume sensor must be re-calibrated (see section ).

Set tidal volume to 700 ml, wait 3 respirations and read the measured expired tidal volume;
on the ventilator:_________ml
on the instrument:_________ml
These volumes should be identical, within +/- 70 ml.
If not, the Adult volume sensor must be re-calibrated (see section 8.3.1).

Set ventilator to STBY and disconnect the reference volume measuring instrument.

7-32 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.8.26 PEEP

Place a testlung (NOT a respiration bag) at the Y-piece. Connect a reference pressure measuring
instrument (0-100 cm H2O) to the patient system.

Start the ventilator with default settings, set respiration rate to 6 and set high alarm to 80.

Set PEEP to 5, 10 and 20 and at each setting check that the pressure measured by the reference
pressure instrument, at the end of the expiration phase, matches the set PEEP within +/- 2 cm H2O.
If not, the Pat. and BiB pressure sensors must be re-calibrated (see section 8.2).

Also, check that the airway pressure graph shown on the display indicates a stable pressure during
the expiratory phase.

Set ventilator to STBY and disconnect the pressure gauge.

7.8.27 O2 fuel cell sensor


If the machine is equipped with an external O2 fuel cell sensor (placed at the INSP cone on the
IBS, the following test must be carried out:
Place the O2 sensor in room air and check that the O2 % shown on the display is 20-22.

Set ventilator to STBY, place the O2 sensor on the INSP cone on the IBS and set freshgas flow
to 10 L/min O2. Check that the O2 % shown on the display is 98-102.

If one of these two tests indicates an inaccuracy, the O2 sensor must be re-calibrated, as described
in the user manual (via the main menu).

Close the freshgas flow.

7.8.28 PCV
Place a testlung (NOT a respiration bag) at the Y-piece.

Start the ventilator with default settings, set respiration rate to 10 and set ventilation mode to PCV.
Fill the bellows by pressing the O2 flush and set freshgas flow to app. 1.0 L/min.

Set Insp. Pressure to 17, wait 3 respirations and check that the measured PEAK pressure shown on
the display is 18-22 hPa.

Also, check that the airway pressure graph shown on the display indicates a stable pressure during
the inspiratory phase.

Oct. 08 7-33
Service Manual for Dameca MRI 508 Dameca

Set Insp. Pressure to 27, wait 3 respirations and check that the measured PEAK pressure shown on
the display is 28-32 hPa.

Also, check that the airway pressure graph shown on the display indicates a stable pressure during
the inspiratory phase.

Set Insp. Pressure to 47, wait 3 respirations and check that the measured PEAK pressure shown on
the display is 48-52 hPa.

Also, check that the airway pressure graph shown on the display indicates a stable pressure during
the inspiratory phase.

If this test indicates an unstable airway pressure during inspiratory phase, the PEEP valve must be
re-calibrated (see section 8.5)

Set ventilator to STBY.

7.8.29 System pressure and disconnection alarm


Place a testlung (for example a respiration bag) at the Y-piece and start the ventilator with default
settings. Fill the bellows by pressing the O2 flush.

Remove the testlung and check that »system pressure high« alarm start when the bellows goes to
the bottom of the chamber.

Check that »disconnection« alarm starts after app. 15 sec.

Connect the testlung, activate O2 flush until bellows is in top position, and check that both alarms
stop after a couple of respirations.

If this test is not OK, the Pat. and BiB pressure sensors must be re-calibrated (see section 8.2).

7.8.30 High pressure alarm


Set high pressure alarm to 15 hPa and check that the »high pressure« alarm starts.

Check also, that the PEAK pressure shown on the display is 14-16 hPa.

Set high alarm to 80 and check that the alarm stops after one respiration cycle.

7-34 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.8.31 Mains alarm

Disconnect the mains cable to the machine from the mains outlet.

Check that »mains failure« alarm starts and that the machine functions as before.

Let the machine run on batteries for 10 minutes.

If a »battery low« alarm starts during these 10 minutes, charge the batteries for at least 8 hours.

If the problem still exists, the batteries should be changed.

Connect the mains cable to the machine.

7.8.32 Leakage current


Check the leakage current according with a leakage current test instrument, according to
Class I, Type B.

Consult the user manual of the leakage current test instrument.

Oct. 08 7-35
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

7-36 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.9 Check Scheme for Post Maintenance / Repair


Refer to section 7.8 for explanation on each check point.

Technician Date Machine S/N IBS S/N APL valve S/N Signature

7.9.1 Internal reduction valves † OK † Not OK


Valve at O2 inlet manifold: (x100 kPa). Should be 4 - 4.5 (x100 kPa)
Valve at drivegas selector block: (x100 kPa). Should be 4 - 4.5 (x100 kPa)
Valve at ventilator manifold: (x100 kPa). Should be 2.8 - 3.2 (x100 kPa)

7.9.2 O2 alarm adjustment † OK † Not OK


Alarm (whistle) starts at app. 2.5 (x100 kPa)

7.9.3 High pressure leakage O2 and AIR † OK † Not OK


Disconnect N2O supply, open O2- and N2O-flow until N2O-rotor goes to zero.
Close O2-flow and set change-over valve to AIR.
O2 leakage: ml/min. Should be < 25 ml/min
AIR leakage: ml/min. Should be < 25 ml/min

7.9.4 High pressure leakage N2O † OK † Not OK


Disconnect AIR supply, open AIR-flow until AIR-rotor goes to zero.
Set change-over valve to N2O.
N2O leakage: ml/min. Should be < 25 ml/min

7.9.5 Dose system leakage † OK † Not OK


Leakage: ml/min. Should be < 25 ml/min

7.9.6 Selftest † OK † Not OK


No error messages during selftest.
Leakage in VENT system: ml/min. Should be < 25 ml/min

Oct. 08 7-37
Service Manual for Dameca MRI 508 Dameca

7.9.7 Auxiliary freshgas outlet † OK † Not OK

Set flow 10 L/min


Flow at BAG 0.0 L/min when activated.
Flow at AUX FG outlet 0.0 L/min when not activated.

7.9.8 Change-over valve AIR (if installed) † OK † Not OK


Set O2-flow and N2O-flow 5 L/min, set AIR-flow to 0.0 L/min
Check AIR/N2O change-over function
Set change-over valve to N2O and check AIR leakage: ml/min. Should be < 10 ml/min

7.9.9 Change-over valve N2O (if installed) † OK † Not OK


Set O2-flow and AIR-flow 5 L/min, set N2O-flow to 0.0 L/min
Check AIR/N2O change-over function
Set change-over valve to AIR and check N2O leakage: ml/min. Should be < 10 ml/min

7.9.10 N2O slave function † OK † Not OK


Set O2-flow and N2O-flow 1 L/min, disconnect N2O and check slave function.

7.9.11 O2 alarm adjustment † OK † Not OK


Alarm (whistle) starts at app. 2.5 (x100 kPa)

7.9.12 Drivegas alarm † OK † Not OK


Alarm starts when drivegas pressure is 2.0 - 2.8 (x100 kPa)

7.9.13 Fans † OK † Not OK


Lower fan, incl. filter

7.9.14 AGSS † OK † Not OK


Yellow ball visible (25 L/min)

7.9.15 O2 flush † OK † Not OK


O2-flow: L/min. Should be 35 - 50 L/min
O2 %: % O2. Should be 100 % O2

7-38 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.9.16 MPL valve † OK † Not OK

Pressure at 2 L/min: (x100 Pa). Should be 90 - 100 (x100 Pa)

7.9.17 Minoxyguard † OK † Not OK


Set N2O flow regulator knob to max. and change-over valve to N2O.
O2 % at 1.5 L/min O2: % O2. Should be 23 - 29 % O2
O2 % at 2.0 L/min O2: % O2. Should be 22 - 28 % O2
O2 % at 4.0 L/min O2: % O2. Should be 24 - 30 % O2

7.9.18 Rotameters † OK † Not OK


Check rotor for O2, AIR and N2O rotameters, up to 10 L/min.

7.9.19 Integrated patient suction † OK † Not OK


Max. vacuum: (x100 kPa). Should at least -0.7 (x100 kPa)

7.9.20 Auxiliary O2 flowmeter † OK † Not OK


O2 flow at 5 L/min: L/min. Should be 4.3 - 5.7 L/min
O2 %: % O2. Should be 100 % O2

7.9.21 IBS † OK † Not OK


Ventilator with default settings, freshgas flow 0.3 L/min.
Remove absorber, absorber valves close, bellows goes to top.

7.9.22 Ventilator function † OK † Not OK


With open Y-piece, set sensor to Paed. and then to Adult.
Default parameters, TV exp on display: ml
Set rate = 4, set sensor to Paed. Set sensor to Adult DURING inspiration.
Set rate = 12, TV exp on display: ml. Should be the same, within +/- 30 ml

7.9.23 Tidal volume † OK † Not OK


Ventilator with default settings and freshgasflow = 0.0 L/min
Set reference instrument to same gas as actual drivegas (AIR or O2)
Tidal volume at 200 ml: ml. Should be 180 - 220 ml.
Tidal volume at 500 ml: ml. Should be 475 - 525 ml.
Tidal volume at 700 ml: ml. Should be 675 - 725 ml.

Oct. 08 7-39
Service Manual for Dameca MRI 508 Dameca

7.9.24 Paed. volume measurement † OK † Not OK

Ventilator with default settings, rate = 10 and freshgasflow = 10.0 L/min O2


Set reference instrument to O2 and wait until measured O2 % in patient system is 95 - 100.
TV exp on display at 200 ml: ml.
TV exp on reference instrument: ml. Should be the same, within +/- 20 ml

7.9.25 Adult volume measurement † OK † Not OK


Ventilator with default settings, rate = 10, freshgasflow = 0.0 L/min and high alarm = 80
Set reference instrument to O2 and wait until measured O2 % in patient system is 95 - 100.
TV exp on display at 500 ml: ml.
TV exp on reference instrument: ml. Should be the same, within +/- 50 ml
TV exp on display at 700 ml: ml.
TV exp on reference instrument: ml. Should be the same, within +/- 70 ml

7.9.26 PEEP † OK † Not OK

Ventilator with default settings, rate = 6 and high alarm = 80


PEEP at 5 (x100 Pa) (x100 Pa). Should be 3 - 7 (x100 Pa)
PEEP at 10 (x100 Pa) (x100 Pa). Should be 8 - 12 (x100 Pa)
PEEP at 20 (x100 Pa): (x100 Pa). Should be 18 - 22 (x100 Pa)
Check for stable airway pressure graph during expiration.

7.9.27 O2 fuel cell sensor (if installed) † OK † Not OK


O2 % in room air: % O2. Should be 20 - 22 % O2
O2 % in O2 flow: % O2. Should be 98 - 102 % O2

7.9.28 PCV † OK † Not OK


Ventilator with default settings, rate = 10
PEAK pressure at 17 (x100 Pa) (x100 Pa). Should be 18 - 22 (x100 Pa)
PEAK pressure at 27 (x100 Pa) (x100 Pa). Should be 28 - 32 (x100 Pa)
PEEP pressure at 47 (x100 Pa): (x100 Pa). Should be 48 - 52 (x100 Pa)
Check for stable airway pressure graph during inspiration.

7.9.29 System pressure and disconnection alarm † OK † Not OK


Disconnect testlung and check both alarms start (»Disconnection« after app. 15 sec.)
Connect testlung, fill bellows and check both alarms stop.

7-40 Oct. 08
Dameca Service Manual for Dameca MRI 508

7.9.30 High pressure alarm † OK † Not OK

Set high alarm to 15 (x100 Pa) and check alarm starts.


PEAK pressure: (x100 Pa). Should be 14 - 16 (x100 Pa).
Set high alarm to 80 (x100 Pa) and check alarm stops after one respiration cycle.

7.9.31 Mains alarm † OK † Not OK


Disconnect mains, check alarm starts and machine functions as before.
Check battery functionality for app. 10 minutes.

7.9.32 Leakage current † OK † Not OK


Perform with test instrument, according to class I, type B

Oct. 08 7-41
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

7-42 Oct. 08
Dameca Service Manual for Dameca MRI 508

8. Calibration
Before calibration can be carried out, the tests If the primary drivegas to the ventilator is AIR,
described in section 7.8.1-7.8.9 must be carried the external adult patient flow sensor used by
out. the integrated volume monitor can be calibrated
at the same time.
The calibration requires a reference measuring
instrument for pressure (0-80 hPa) and flow If the primary drivegas to the ventilator is O2,
(0-80 L/min.) the external adult patient flowsensor must
NOT be calibrated at the same time.
Activate MAIN menu, select service menu and
enter the service technician password (91-34-80). Calibrate only inspiratory flow sensor:
Select Analog calib. <ETR>. In the installation menu (see section 6.7),
set »Volume sensor« to »No vol. sensor«.
When calibration of the inspiratory flow sensor
In the calibration menu you get is completed, set »Volume sensor« as desired
access to: (see section 6.7).

Insp. Flow & Adult pat. Flow Remove bellows in Bag in Bottle, and connect
BiB & Pat. Pressure inspiration port of the integrated patient system
Pat. Flow to flow measuring reference instrument.
Freshgas Flow
PEEP valve Press enter to confirm <ETR>.
Temperature calib.
Calibrate both inspiratory flow sensor and
adult patient flow sensor:
Note: Remove bellows in Bag in Bottle, and connect
inspiratory port of the integrated patient
The accuracy of the ventilator depends
breathing system to external adult patient
on the accuracy of the calibration.
flow sensor (MUST be mounted with patient
Therefore the ventilator must have
port facing flow instrument) and then to flow
room temperature, and must be switched
measuring reference instrument.Press enter
on for at least 30 min., before
to confirm <ETR>.
calibration is carried out.

Note:
When calibrating the patient flow
8.1 Insp. Flow & Adult Pat. Flow sensor, it must be placed with the
port labeled »PATIENT« facing the
Select »Insp. Flow & Adult pat. Flow« to cali- flow measuring reference instrument.
brate the sensor that measures the inspiratory
flow from the ventilator to the Bag in Bottle.

Oct. 08 8-1
Service Manual for Dameca MRI 508 Dameca

Step 1
Check that the insp. flow pressure regulator
on the pneumatic block inside the ventilator $GMXVWVFUHZRQ
LQVSLUDWRU\IORZYDOYH
is adjusted to 2.8-3.2 (x100kPa).

Use pressure gauge with hose (P/N72231)


and adjust if necessary.

The ventilator automatically generates a


flow, which makes adjustment possible.

Press enter to confirm <ETR>.

Note:
During calibration, the inspiratory flow qusjob42.10987-25 manual 1.pdf

measured by the ventilator according


to data from last calibration, is shown Step 3
on the display. Check that inspiratory flow reading on flow
measuring instrument is 0.0 L/min.
The electrical signal to the insp. Press enter to calibrate at 0.0 L/min <ETR>.
valve is also shown on the display
as »PWM«. Step 4
Adjust inspiratory flow from ventilator, until
flow reading on flow measuring instrument
is 0.95-1.05 L/min.
Step 2 Adjust by turning the control wheel.
Check that inspiratory flow reading on flow Press enter to calibrate at 1.0 L/min <ETR>.
measuring instrument is 78-82 L/min.
Step 5
If necessary, adjust screw on top of inspiratory Adjust inspiratory flow from ventilator, until
valve on the pneumatic block inside the flow reading on flow measuring instrument
ventilator. is 1.95-2.05 L/min.
Adjust by turning the control wheel.
Press enter to confirm <ETR>. Press enter to calibrate at 2.0 L/min <ETR>.

8-2 Oct. 08
Dameca Service Manual for Dameca MRI 508

Step 6 Step 12
Adjust inspiratory flow from ventilator, until Adjust inspiratory flow from ventilator, until
flow reading on flow measuring instrument flow reading on flow measuring instrument
is 4.95-5.05 L/min. is 39.9-40.1 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 5.0 L/min <ETR>. Press enter to calibrate at 40.0 L/min <ETR>.

Step 7 Step 13
Adjust inspiratory flow from ventilator, until Adjust inspiratory flow from ventilator, until
flow reading on flow measuring instrument flow reading on flow measuring instrument
is 9.9-10.1 L/min. is 49.9-50.1 .0 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 10.0 L/min <ETR>. Press enter to calibrate at 50.0 L/min <ETR>.

Step 8 Step 14
Adjust inspiratory flow from ventilator, until Adjust inspiratory flow from ventilator, until
flow reading on flow measuring instrument flow reading on flow measuring instrument
is 14.9-15.1 L/min. is 59.9-60.1 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 15.0 L/min <ETR>. Press enter to calibrate at 60.0 L/min <ETR>.

Step 9 Step 15
Adjust inspiratory flow from ventilator, until Adjust inspiratory flow from ventilator, until
flow reading on flow measuring instrument flow reading on flow measuring instrument
is 19.9-20.1 L/min. is 79.8-80.2 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 20.0 L/min <ETR>. Press enter to calibrate at 80.0 L/min <ETR>.

Step 10 Step 16
Adjust inspiratory flow from ventilator, until Press enter to save calibration <ETR>.
flow reading on flow measuring instrument Press »Normal screen« to cancel the calibration.
is 24.9-25.1 L/min.
Adjust by turning the control wheel.
Press enter to calibrate at 25.0 L/min <ETR>. Note:
If the ventilator does not accept the
Step 11 calibration, the filter inside the venti-
Adjust inspiratory flow from ventilator, until lator pneumatic manifold could be
flow reading on flow measuring instrument defective.Change filter (P/N36709),
is 29.9-30.1 L/min. gaskets (2 pcs. P/N36498 and 1 pcs.
Adjust by turning the control wheel. P/N36499), O-ring (P/N36423) and
Press enter to calibrate at 30.0 L/min <ETR>. redo the calibration.

Oct. 08 8-3
Service Manual for Dameca MRI 508 Dameca

8.2 Bag in Bottle and Patient


Pressure
Select »BiB & pat.pressure« to calibrate the pressure measuring reference instrument is 80 hPa.
sensor that measures the pressure between Press enter to re-calibrate at 80 hPa <ETR>.
chamber and bellows in the Bag in Bottle system,
and the sensor that measures the airway pressure Step 5
in the patient breathing system. Press enter to save calibration <ETR>.

Assemble the integrated patient breathing Press »Normal screen« to cancel the calibration.
system, incl. hoses and Y-piece.

Connect the Y-piece to pressure measuring 8.3 Patient Flow


reference instrument.
If the primary drivegas to the ventilator is O2,
Step 1 the O2 supply hose to the machine MUST be
Remove the Bag in Bottle chamber and bellows. connected to AIR during calibration of the
Press enter to calibrate at 0 hPa <ETR>. volume sensor, as these sensors MUST be
calibrated with AIR-flow.

Note: 8.3.1 Calibrate adult patient flow


During calibration, the Bag in Bottle sensor
pressure and the patient breathing
system pressure measured by the Select »Adult sensor« to calibrate the external
ventilator according to data from last adult patient flow sensor.
calibration, is shown on the display
as »BiB« and »Pat.«. Remove bellows in Bag in Bottle, and connect
inspiratory port of the integrated patient
breathing system to external adult patient flow
sensor (should be mounted with patient port
Step 2 facing flow instrument) and then to flow
Re-mount only the Bag in Bottle chamber and measuring reference instrument.
open freshgas flow until stable reading on the
pressure measuring reference instrument is 80 hPa. Press enter to confirm <ETR>.
Press enter to calibrate at 80 hPa <ETR>.

Step 3 Note:
Remove the Bag in Bottle chamber. When calibrating the patient flow
Press enter to re-calibrate at 0 hPa <ETR>. sensor, it must be placed with the
port labeled »PATIENT« facing the
Step 4 flow measuring reference instrument.
Re-mount only the Bag in Bottle chamber and
open freshgas flow until stable reading on the

8-4 Oct. 08
Dameca Service Manual for Dameca MRI 508

Step 1 Step 7
Check that reading on flow measuring Adjust flow from ventilator, until flow reading
instrument is 0.0 L/min. on flow measuring instrument is 19.9-20.1 L/min.

Press enter to calibrate at 0.0 L/min <ETR>. Adjust by turning the control wheel.
Press enter to calibrate at 20.0 L/min <ETR>.

Note: Step 8
During calibration, the patient flow Adjust flow from ventilator, until flow reading
measured by the ventilator according on flow measuring instrument is 24.9-25.1 L/min
to data from last calibration, is shown .
on the display. Adjust by turning the control wheel.
Press enter to calibrate at 25.0 L/min <ETR>.

Step 9
Step 2 Adjust flow from ventilator, until flow reading
Adjust flow from ventilator, until flow reading on flow measuring instrument is 29.9-30.1 L/min.
on flow measuring instrument is 0.95-1.05 L/min. Adjust by turning the control wheel.
Adjust by turning the control wheel. Press enter to calibrate at 30.0 L/min <ETR>.
Press enter to calibrate at 1.0 L/min <ETR>.
Step 10
Step 3 Adjust flow from ventilator, until flow reading
Adjust flow from ventilator, until flow reading on flow measuring instrument is 39.9-40.1 L/min
on flow measuring instrument is 1.95-2.05 L/min. .
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 2.0 L/min <ETR>. Press enter to calibrate at 40.0 L/min <ETR>.

Step 4 Step 11
Adjust flow from ventilator, until flow reading Adjust flow from ventilator, until flow reading
on flow measuring instrument is 4.95-5.05 L/min. on flow measuring instrument is 49.9-50.1 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 5.0 L/min <ETR>. Press enter to calibrate at 50.0 L/min <ETR>.

Step 5 Step 12
Adjust flow from ventilator, until flow reading Adjust flow from ventilator, until flow reading
on flow measuring instrument is 9.9-10.1 L/min. on flow measuring instrument is 59.9-60.1 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 10.0 L/min <ETR>. Press enter to calibrate at 60.0 L/min <ETR>.

Step 6 Step 13
Adjust flow from ventilator, until flow reading Adjust flow from ventilator, until flow reading
on flow measuring instrument is 14.9-15.1 L/min. on flow measuring instrument is 79.8-80.2 L/min.
Adjust by turning the control wheel. Adjust by turning the control wheel.
Press enter to calibrate at 15.0 L/min <ETR>. Press enter to calibrate at 80.0 L/min <ETR>.

Oct. 08 8-5
Service Manual for Dameca MRI 508 Dameca

Step 14 Step 2
Press enter to save calibration <ETR>. Adjust flow from ventilator, until flow reading
Press »Normal screen« to cancel the calibration on flow measuring instrument is 0.95-1.05 L/min.
Adjust by turning the control wheel.
8.3.2 Calibrate paediatric patient Press enter to calibrate at 1.0 L/min <ETR>.
flow sensor
Select »Paed. sensor« to calibrate the Step 3
external paediatric patient flow sensor. Adjust flow from ventilator, until flow reading
on flow measuring instrument is 1.95-2.05 L/min.
Remove bellows in Bag in Bottle, and connect Adjust by turning the control wheel.
inspiratory port of the integrated patient
breathing system to the external paediatric Press enter to calibrate at 2.0 L/min <ETR>.
patient flow sensor (should be mounted with
patient port facing flow instrument) and then Step 4
to flow measuring reference instrument. Adjust flow from ventilator, until flow reading
on flow measuring instrument is 3.95-4.05 L/min.
Press enter to confirm <ETR>. Adjust by turning the control wheel.

Press enter to calibrate at 4.0 L/min <ETR>.


Note:
When calibrating the patient flow Step 5
sensor, it must be placed with the port Adjust flow from ventilator, until flow reading
labeled »PATIENT« facing the flow on flow measuring instrument is 6.9-7.1 L/min.
measuring reference instrument. Adjust by turning the control wheel.

Press enter to calibrate at 7.0 L/min <ETR>.

Step 1 Step 6
Check that reading on flow measuring Adjust flow from ventilator, until flow reading
instrument is 0.0 L/min. on flow measuring instrument is 9.9-10.1 L/min.
Adjust by turning the control wheel.
Press enter to calibrate at 0.0 L/min <ETR>.
Press enter to calibrate at 10.0 L/min <ETR>.

Note: Step 7
During calibration, the patient flow Adjust flow from ventilator, until flow reading
measured by the ventilator according on flow measuring instrument is 14.9-15.1 L/min.
to data from last calibration, is shown
on the display. Adjust by turning the control wheel.

Press enter to calibrate at 15.0 L/min <ETR>.

8-6 Oct. 08
Dameca Service Manual for Dameca MRI 508

Step 8 If the machine is not equipped with an auxiliary


Adjust flow from ventilator, until flow reading freshgas outlet, connect the reference flow
on flow measuring instrument is 19.9-20.1 L/min. measuring instrument to the insp. cone on the
Adjust by turning the control wheel. IBS, set APL valve to max. and block the cone
for the respiration bag on the IBS.
Press enter to calibrate at 20.0 L/min <ETR>.
Step 1
Step 9 Close the freshgas flow and check that the refe-
Adjust flow from ventilator, until flow reading rence instrument reads 0.0 L/min. Press <ETR>.
on flow measuring instrument is 24.9-25.1 L/min.
Adjust by turning the control wheel. Step 2
Adjust AIR freshgas flow until reference instru-
Press enter to calibrate at 25.0 L/min <ETR>. ment reads 10.0 L/min. AIR flow. Press <ETR>.

Step 10 Step 3
Adjust flow from ventilator, until flow reading Close the freshgas flow and check that the refe-
on flow measuring instrument is 29.9-30.1 L/min. rence instrument reads 0.0 L/min. Press <ETR>.
Adjust by turning the control wheel.
Step 4
Press enter to calibrate at 30.0 L/min <ETR>. Adjust AIR freshgas flow until reference instru-
ment reads 10.0 L/min. AIR flow. Press <ETR>.
Step 11
Adjust flow from ventilator, until flow reading Step 5
on flow measuring instrument is 34.9-35.1 L/min. Press enter to save calibration <ETR>.
Adjust by turning the control wheel. Press »Normal screen« to cancel calibration.

Press enter to calibrate at 35.0 L/min <ETR>.


8.5 PEEP Valve
Step 12
Press enter to save calibration <ETR>. Before the characteristics of the PEEP valve
Press »Normal screen« to cancel the calibration. can be calibrated, the ventilator insp. flow
(see section 8.1) and the Paed. flow sensor
(see section 8.3.2) must be calibrated previously.
8.4 Freshgas Flow
Select »PEEP valve« to calibrate the PEEP
Select »Freshgas Flow« to calibrate the sensor valve of the ventilator.
that measures the freshgas flow from the Remove insp. and exp. hoses from the IBS.
rotameterbox to the patient system. Remove bellows in Bag in Bottle and mount
Bag in Bottle chamber. Place the Paediatric
Activate the auxiliary freshgas outlet and flowsensor on insp. cone of the IBS. The port
connect a reference flow measuring instrument labeled PATIENT must face away from the
to the outlet. IBS outlet <ETR>.

Oct. 08 8-7
Service Manual for Dameca MRI 508 Dameca

Step 1 When the temperature of the coil has reached


The PEEP valve is now warmed up quickly 37-43OC, press enter to start next step of the
during a 4 min. period. calibration <ETR>.

During this period, the temperature of the PEEP If enter is not pressed within 20 min., the
valve coil must be measured, using an infrared calibration is automatically cancelled.
temperature sensor. If the temperature of the
coil reaches 35-40OC before the 4 min., press Step 4
enter to start next step of the calibration <ETR>. Now the ventilator checks the characteristics
of the PEEP valve.

6FUHZIRUDGMVXWHPHQW If the PEEP valve should be re-calibrated, a


RI3((3YDOYHFRLO
»difference flow« will appear on the display.
3((3YDOYHFRLO
Adjust the screw on top of the PEEP valve on
the ventilator manifold, until »difference flow«
is 0.0 L/min., and press enter <ETR>.

The calibration is now verified.

When the PEEP valve characteristics is within


tolerance limits, the display will show this, and
enter should be pressed to finish the PEEP valve
calibration.

Note:
The PEEP valve calibration can be
qusjob42.10987-25 manual 2.pdf cancelled at all times, by pressing
»Normal screen«.
Step 2
The PEEP valve temperature is now stabilised
during a 5 min. period.
8.6 Temperature Calibration
When the 5 min. period is ended, the next step
of the calibration is activated automatically. The«Temperature calib.« is for definition of
the temperature characteristics of the pressure
Step 3 sensors used in the ventilator.
The PEEP valve is now warmed up slowly, and This definition can NOT be changed in the field,
the temperature of the PEEP valve coil must be as it requires specific control of room tempera-
measured, using an infrared temperature sensor. ture for the electronic box of the ventilator.

8-8 Oct. 08
Dameca Service Manual for Dameca MRI 508

8.7 Drivegas Adjustments


8.7.1 Adjustment of ventilator drivegas
low alarm
Connect a testlung (f.ex. a respiration bag) Step 2
to the Y-piece and set status to VENT with Adjust the drivegas (AIR or O2) supply pressure
default setting and respiration rate set to 4 to 3.0 (x100kPa) and check that the »drivegas
and I:E ratio set to 3:1. low« alarm stops.

Step 1 If not, turn the screw on top of the pressure


Reduce the drivegas (AIR or O2) supply pres- switch on the ventilator manifold during
sure to 2.5 (x100kPa) and during inspiration, inspiration, until the »drivegas low« alarm
turn the screw on top of the pressure switch on stops.
the ventilator manifold,until the »drivegas low«
alarm starts. Then perform step 1 and 2 again.

6FUHZIRUDGMVXWHPHQW
RIGULYHJDVORZ
DODUP

qusjob42.10987-25 manual 3.pdf

Oct. 08 8-9
Service Manual for Dameca MRI 508 Dameca

8.8 Minoxyguard
Set ventilator to STBY, close O2 and AIR Step 4
freshgas flows, set change-over valve to Adjust O2 freshgas flow regulator knob until the
N2O and set N2O freshgas flow regulator O2 flow is 2 L/min.
knob to max.
Step 5
Step 1 Turn the adjustment screw on the rear side of
Slowly open O2 freshgas flow regulator knob the lower minoxyguard manifold (positioned
until the O2 flow is 250 ml/min. behind the N2O freshgas flow regulator) until
the N2O flow is 6 L/min.
Step 2
Turn the adjustment screw on the rear side of Step 6
the lower minoxyguard manifold (positioned Close the O2 freshgas flow and repeat step 3.
behind the O2 freshgas flow regulator) until If a re-adjustment is required, step 4 and 5
the N2O float (rotor) just begins to move. must also be performed again.

Step 3
Close O2 freshgas flow regulator knob and
slowly open it again. Check that the N2O flow
starts when the O2 flow is 250 - 300 ml/min. If
not, re-adjust as described in section 1 and 2.

8-10 Oct. 08
Dameca Service Manual for Dameca MRI 508

9. Spare Parts
Contents Page
Dameca MRI 508 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Stand with Integrated Suction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Gas- and Electrical Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Gasbox for Dameca MRI 508 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Gasbox Assy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10
Togggle Switch Valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
Turn Switch Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
Rotameterbox with 3-tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-14
Minoxyguard with 3 tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-16
Rotameterbox with 5-tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-18
Drive Gas Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-22
Backbar Selectatec . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-24
Backbar Dräger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-26
Inlet Manifold O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-28
Inlet Manifold AIR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-30
Inlet Manifold N2O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-32
Inlet Manifold VAC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-34
Inlet Manifold AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-35
Whispa Ventilator Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-36
E-Box for Whispa ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-38
Power Supply Box, Old version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-40
Valve Block Assy. for Whispa . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-42
Rear Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-48
Flowmarker Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-49
Suction Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-50
Mains Switch, standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-52
Vacuum Suction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-54
AGSS Ejector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-56
O2 Flush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-58
IBS System Complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-59
IBS Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-60
IBS System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-62

Spare parts for DAMECA products should be ordered through our local agents.
When ordering spare parts, identify the part, and use the part number stated in this manual.

Colour Code Colour


00 Black
Note:
02 Red Due to safety reasons always use
05 Green
original DAMECA spare parts.
Never use commercially available
06 Blue
O-rings and/or gaskets.
09 White

Oct. 08 9-1
Service Manual for Dameca MRI 508 Dameca

Dameca MRI 508


Item Qty Description Part No.
1 1 Stand for Siesta i MRI 10651M-01
2 4 Mounting pole for Siesta 508 MRI 10651-114
3 1 Side panel right hand side 10651-25
4 2 Rear corner profile 10651-11
5 2 Front corner profile 10651-10
6 1 Top plate for Siesta i, assembled 10656
7 1 Rotameter gas box for Siesta i Whispa MRI 10654M
8 1 Rear Panel for Siesta i Whispa MRI 10657M
9 1 Vaporizer gas box for Dameca MRI 508 10653M
10 1 Table plate 10655
11 1 Front cover for Siesta i, MRI 10651M-28
12 1 Ventilator for Siesta i Whispa MRI 10659M-100
13 1 Finger screw (optional) 10656-40
14 2 Mounting plate for Siesta i MRI 10651M-14
15 1 Bushing for infusion stand 10651-113
16 4 Star washer 70479
17 4 Screw DIN 912 M8X35 70028-0835
18 4 Screw M5x12 70414-0512
19 4 Screw M4x10 70414-0410
20 1 AGSS block 11013-120
21 1 Side panel assembly left hand side 10651-260
22 1 Mini ejector 11013
23 1 Aux. freshgas outlet 10651-270
24 1 Rear cover assembly for Siesta i MRI 10651M-12
25 1 External Powersupply for Dameca MRI 508 40098-30

9-2 Jan. 10
Dameca Service Manual for Dameca MRI 508

6
13

4
3
8

7
15

4
2
21

5 25

20
19

22 24

12
18
10
23

14

1 11

16
17

qusjob42.10651M-00 manual-1.pdf

Fig. 9-1 Dameca MRI 508

Jan. 10 9-3
Service Manual for Dameca MRI 508 Dameca

Stand with Integrated Suction


Item Qty Description Part No.
1 1 Stand for Siesta i 10651-011
2 1 Snap bushing 40058-26
3 6 Screw DIN 912 M5X10 70013
4 1 Mounting plate for suction unit 32530-05
5 1 Integrated suction unit (optional) 32530-100
6 1 Lid for suction unit 10651-012
7 2 Wheel with brake (gray) 36853-05
8 2 Wheel without brake (gray) 36854-05

qusjob42.10651M-01 manual.png

Fig. 9-2 Stand with integrated suction

9-4 Oct. 08
Dameca Service Manual for Dameca MRI 508

Gas- and Electrical Panel


Item Qty Description Part No.
1 1 Electrical & gas panel, MRI 10657M-01
2 4 Screw 70046-0410
3 1 Inlet manifold O2 31350
4 1 Inlet manifold AIR 31350C
5 1 Inlet manifold N2O 31350A
6 1 Inlet manifold VAC 31350B
7 1 Inlet manifold AGS 31350F
8 10 Screw 70046-0510
9 1 Cover for inlet manifold 10657-08
10 6 Screw M5x12 70414-0512
11 1 Rear cover for Dameca MRI 508 10657M-50



 

 


 

qusjob42.10657MS manual.pdf

Fig. 9-3 Gas- and electrical panel

Oct. 08 9-5
Service Manual for Dameca MRI 508 Dameca

Gasbox for Dameca MRI 508


Item Qty Description Part No.
1 1 Gasbox for Siesta i Whispa 10654-101
2 1 Light Armature 10654M-103
3 1 Whispa TFT MMI 10987FM-100
4 1 5-rotameter box 10774-750
5 1 O2 flush assembly 10658-30
6 1 Change over valve AIR/N2O for 10770 11981-10
7 1 Mains switch 40090-70
8 26 Nut DIN 6923 A2 70424
9 1 Code ring 11371-096
10 2 Screw DIN 912 M4X10 70008
11 4 Nut M4 DIN 934 70440
12 1 Front foil for Dameca MRI 508 10654-135
13 1 Acrylic plate for rotameter 10654-112
14 2 Sealing Strip 23825
15 1 Sealing Strip 23825
16 1 O2 tank w/flute 10654-30
17 10 Nylon washer 70468
18 4 Screw M5x12 70414-0512
19 1 Destributer block 10659-40
20 2 Screw DIN 912 M5X12 70014
21 1 Steady list (SS) 10654-03

9-6 Oct. 08
Dameca Service Manual for Dameca MRI 508

 






















 



,WHP 4W\ 'HVFULSWLRQ 3DUW1R


  *DVER[IRU6LHVWDL:KLVSD 
  /LJKW$UPDWXUH 0
  :KLVSD7)700,05, )0
  URWDPHWHUER[ 
  2IOXVKDVVHPEO\ 
  &KDQJHRYHUYDOYH$,512IRU 
  0DLQVVZLWFK 
  1XW',1$ 
  &RGHULQJ 
  6FUHZ',10; 
  1XW0',1 
  )URQWIRLO'DPHFD05, 
  $FU\OLFSODWHIRUURWDPHWHU 
  6HDOLQJ6WULS 
  6HDOLQJ6WULS 
  2WDQNZIOXWH 
qusjob42.10654MS manual.pdf

Fig. 9-4 Gasbox for Dameca MRI 508

Oct. 08 9-7
Service Manual for Dameca MRI 508 Dameca

Gasbox Assy.
Item Qty Description Part No.
1 1 Fordampergasbox Dameca MRI 508, complete 10653M-20109
2 1 Back Bar selectatec compatible (optional) 11080-650
3 1 Cone 30 mm female 13952
4 1 Nipple VAC 31094b
5 1 Mini Coupling O2, white 31081
6 1 Pneumatic control panel for Siesta i Whispa MRI 10653M-204
7 18 Spacer bushing. M4, l: 30mm 40098-77
8 1 Equipotential connector 40055-01
9 1 Potential equalising label (green/yellow) 11360
10 4 Nut DIN 6923 A2 70424
11 2 Screw RHR M4 X 8MM A2 70046-0408
12 2 Screw DIN 912 M4X12 70009
13 2 Nylon washer 70467
14 1 Negative pressure valve 10653-150
15 2 Ball chain end 36871-02
16 1 Ball chain 36871-01
17 2 Screw RHR M3 X 8MM A2 70046-0308
18 1 Plug for AGSS 13951-11
19 1 Code ring O2 (white) 11365
20 1 Code ring (blue) 11365A
21 1 Code ring (black/white) 11365C
22 1 Strain relief 40061-63
23 2 Nut M3 DIN 934 70447
24 4 Screw M5x12 70414-0512
25 1 Reservoir assembly for Siesta i 10653M-60
26 2 Spacer bush 10653-63
27 2 Nut M5 DIN 934 70441
28 1 Snap bushing 40058-26
29 1 Fresh Gas Switch Valve 10653-220
30 1 Vacuum Switch Valve 10653-230
31 1 Rear cover for Dameca MRI 508 10653M-021
32 3 Quick connector angled, ø6/M5 36690-22
33 3 Mounting ring for ø40 gauges 10653-52
34 3 Gauge 0-10 bar, plastic housing 34108
35 3 Mounting plate (SS) for ø40 gauge 10653M-53
36 6 Screw DIN 912 M5X10 70013
37 1 Pin-index pressure regulator N2O (optional) 33052-50
38 1 Pin-index pressure regulator O2 (optional) 33053

9-8 May 09
Dameca Service Manual for Dameca MRI 508



 











 


 
 







 
 
 

 




 






qusjob42.10653MS manual.pdf

Fig. 9-5 Gasbox assy.

May 09 9-9
Service Manual for Dameca MRI 508 Dameca

Control Panel
Item Qty Description Part No.
1 1 Front plate for control panel, SS 10653M-2041
2 1 O2 flow meter 0-10 L/min (optional) 32022
3 1 Gauge 1-0 bar, plastic housing 34109
4 1 Regulator forsuction unit (optional) 10995-155
5 1 Foil for flow meter (optional) 10653-06
6 1 Foil for suction unit (optional) 10653-107
7 1 Blind foil for short model 10653-09
8 1 Foil for outlet selector valve (optional) 10653-11
9 1 Acrylic plate for vacuum gauge 10653-12
10 2 Screw DIN 912 M5X8 70017-0508
11 2 Screw DIN 912 M4X8 70007
12 1 Clamp for flowmeter 10653M-05
13 2 Support for flowmeter, MRI 10653M-055
14 16 Nut DIN 6923 A2 70424
15 1 Knob for O2 flow meter (optional) 10653-044
16 1 Allen screw 70107
17 1 Quick connector angled, ø6/1/8" 36690-20
18 1 Label O2 White 11336
19 1 Toggle switch valve 11951
20 1 Turn switch valve 11982
21 1 Quick connector angled, ø6/M5 36690-22
22 1 T-connector ø6/1/8"RG 36690-60
23 1 Lynkobling, T-stykke, Ø6 36690-18
24 1 Silencer 10653-208
25 2 Spacer bushing. l: 28mm 10653-78
26 1 Mounting plate (SS) for ø40 gauge 10653M-53
27 1 Spacer for vacuum gauge, MRI 10653M-043

9-10 Oct. 08
Dameca Service Manual for Dameca MRI 508




 

*








 

 




 


 









qusjob42.10653M-204S manual.pdf

Fig. 9-6 Control panel

Oct. 08 9-11
Service Manual for Dameca MRI 508 Dameca

Toggle Switch Valve


Item Qty Description Part No.
1 1 Mounting ring for PXB-valve 11951-04
2 1 Switch for PXB-valve 11951-01
3 1 Mounting block 11926-02
4 1 PXB-valve 11926-07
5 2 Quick connector angled, ø6/M5 36690-22

qusjob42.11951.png

Fig. 9-7 Toggle switch valve

9-12 Oct. 08
Dameca Service Manual for Dameca MRI 508

Turn Switch Valve


Item Qty Description Part No.
1 1 Mounting ring for PXB-valve 11951-04
2 1 Tuggle handle for PXB-valve 11926-01
3 1 Mounting block 11926-02
4 1 PXB-valve 11926-07
5 2 Quick connector angled, ø6/M5 36690-22
6 1 Alu. washer ø8/ø5 x1 36596
7 1 Silencer 10653-208

qusjob42.11982.png

Fig. 9-8 Turn switch valve

Oct. 08 9-13
Service Manual for Dameca MRI 508 Dameca

Rotameterbox with 3-tubes


Item Qty Description Part No.
1 1 3-rotameter assembly 10654-75
2 1 Mounting rod for rotameter 10774-181
3 1 Mounting rod for rotameter 10654-81
4 2 Screw 70058-0420
5 2 Screw 70058
6 4 Spacer bushing 10654-82
7 4 Screw DIN 912 M4X16 70010
8 1 Nipple 10894-71
9 1 Coubling Ø8/1/8" RG 36690-14
10 3 Retainer nut 10774-71
11 1 Allen screw 70101-0408
12 1 Opal pane 10654-76

9-14 Oct. 08
Dameca Service Manual for Dameca MRI 508

qusjob4210654-80 manual.png

Fig. 9-9 Rotameterbox with 3-tubes

Oct. 08 9-15
Service Manual for Dameca MRI 508 Dameca

Minoxyguard with 3-tubes


Item Qty Description Part No.
1 1 Lower manifold for 3-tubes 10887-31
2 1 Side rail, right 11025-03
3 1 Side rail, left 11024-03
4 1 Top manifold for 3-tubes 10887-32
5 1 Proportional reglator 11780
6 4 Screw 11024-05
7 1 O-ring Ø1,78x1,78 EPDM 36277
8 6 Allen point screw M4 X 5,5MM 70135-04
9 2 Allen point screw 70135-06
10 2 Allen screw 70120
11 2 Quick connector 36691-0405
12 4 Quick connector Ø6/M5 Viton 36690-12
13 1 Alu. washer ø8/ø5 x1 36596
14 1 Nozzle 10894-485
15 6 Silicone gasket Ø16,1xØ7x2mm 36509
16 6 Silicone gasket Ø15,5xØ11x2mm 36505
17 3 Filter bronze Ø9,8xØ3x21mm 36704
18 3 Guide for 9" tubes 11031-52
19 1 Tube 9", O2, 1-10 lpm 11082-09
20 3 Stop, 9" rotameter tube 11031-51
21 3 Nylon washer 70467
22 3 Spring 11004-44
23 3 Top screw 11004-46
24 3 Gasket Ø21xØ16x1mm EPDM 36513
25 1 Tube 9", AIR, 1-10 lpm 11089-09
26 1 Tube 9", N2O, 1-10 lpm 11085-09
27 3 Valve seat 32015-02
28 6 O-ring Ø8,1x1,6 EPDM 36304
29 5 O-ring ø6,07x1,78 teflon 36311
30 5 Spindle guide M6x0.5 10887-24
31 3 Spindle 10887-25
32 3 Thrust pad large 11004-67
33 3 Thrust pad small 11004-69
34 3 Allen screw 70108
35 3 Spring 11034-09
36 1 Button 11004-60
37 1 Label O2 White 11336
38 2 Knob, gray 11034-60
39 1 Label AIR (black/white) 11336C
40 1 Label N2O (blue) 11336A
41 2 Spindle short M6x0.5 10887-26
42 2 Retainer w/slot 11004-18
43 2 Spring 5,0/3,4/8,0 10887-27
44 1 Tube, Ø6/ø4, white 23209
45 1 Tube, Ø6/Ø4, blue 23206
46 1 Tube, Ø6/Ø4, blue 23206

9-16 Oct. 08
Dameca Service Manual for Dameca MRI 508











     





 




 
 



 









 
 



 

 

 








qusjob4210654-75 manual.pdf

Fig. 9-10 Minoxyguard with 3-tubes

Oct. 08 9-17
Service Manual for Dameca MRI 508 Dameca

Rotameterbox with 5-tubes


Item Qty Description Part No.
1 1 Lower manifold for 5-tubes 10774-751
2 1 Top manifold for 5-tubes 10774-752
3 2 Side rail for 5-tubes 10774-754
4 1 Opal pane 10774-755
5 1 Lid for 5-tubes 10774-756
6 2 Allen screw 70120
7 5 Allen screw 70135-04
8 2 Spring 5,0/3,4/8,0 10887-27
9 2 Spindle short M6x0.5 10887-26
10 5 Spindle guide M6x0.5 10887-24
11 5 O-ring teflon Ø6,07x1,78 36311
12 2 Retainer w/slot 11004-18
13 5 Screw DIN 912 M5X25 70017-0525
14 1 Proportional reglator 11780
15 10 Quick connector 36690-12
16 2 Washer ø5,3xø16x4 70472
17 2 Screw 70060
18 4 Screw 70061
19 4 Screw DIN 912 M3X8 70001-0308
20 1 Fitting Ø8/Ø6-1/8" RG 36680-8618
21 2 Alu. washer 36594
22 1 Allen point screw 70135-06
23 1 Stopper 36670-18
24 3 Valve seat 32015-02
25 10 O-ring Ø8,1x1,6 36304
26 3 Retainer nut 10774-71
27 3 Spindle 10887-25
28 3 Washer for spring, big 11004-67
29 3 Washer for spring, small 11004-69
30 3 Allen screw 70108
31 2 Knob, gray 11034-60
32 3 Spring 11034-09
33 1 Label AIR (black/white) 11336C
34 1 Label N2O (blue) 11336A
35 1 Knob, black 11004-60

9-18 Oct. 08
Dameca Service Manual for Dameca MRI 508

    













 

 

 
  
 

 

 

 



  
 
  







 

 
 


 
 


 





 
 

 

 

 
 
 





 

qusjob42.10774-750 manual.pdf

Fig. 9-11 Rotameterbox with 5 tubes

Oct. 08 9-19
Service Manual for Dameca MRI 508 Dameca

Rotameterbox with 5-tubes (continued)

Item Qty Description Part No.


36 1 Label O2 White 11336
37 4 Silicone gasket Ø10,9xØ5x2mm 36503
38 2 Guide for 0.05-0.95 lpm tubes 11031-53
39 4 Silicone gasket Ø15,5xØ11x2mm 36502
40 1 Tube O2, 0.05-0.95 lpm, 210 mm 11085-270
41 2 Stop for 210 mm tube 11031-54
42 2 Washer 70604
43 2 Spring 11004-45
44 2 Connector bushing for Ø11 tubes 10774-758
45 1 Tube O2, 0.05-0.95 lpm, 210 mm 11082-270
46 3 Silicone gasket Ø16,1xØ7x2mm 36509
47 3 Filter bronze Ø9,8xØ3x21mm 36704
48 1 Guide for 9" tubes 11031-52
49 1 Silicone gasket Ø15,5xØ11x2mm 36505
50 1 Tube AIR, 1-10 lpm, 210 mm 11089-275
51 5 Silicone gasket Ø16,8xØ10,9x2mm 36517
52 1 Stop for Ø17 tubes 11031-56
53 3 Silicone gasket Ø16,9xØ7x2mm 36508
54 3 Washer brass Ø16XØ9,5X1,25 70603
55 3 Spring 11004-44
56 3 Connector bushing for Ø17 tubes 10774-757
57 6 O-ring Ø14,1x1,6 36307
58 2 Guide for O2 and AIR tubes 11031-58
59 1 Tube N2O, 1-10 lpm, 210 mm 11085-275
60 2 Stop for O2 and N2O tube 11031-55
61 1 Tube O2, 1-10 lpm, 210 mm 11082-275
62 1 Alu. washer 36596
63 1 Nozzle 10894-485
64 1 Tube Ø6/Ø4, blue 23206
65 1 Tube Ø6/Ø4, blue 23206
66 1 Tube Ø6/Ø4, blue 23206
67 1 Tube Ø6/Ø4, white 23209
68 1 Tube Ø6/Ø4, white 23209

9-20 Oct. 08
Dameca Service Manual for Dameca MRI 508

    













 

 

 
  
 

 

 

 



  
 
  







 

 
 


 
 


 





 
 

 

 

 
 
 





 

qusjob42.10774-750 manual.pdf

Fig. 9-11 Rotameterbox with 5 tubes

Oct. 08 9-21
Service Manual for Dameca MRI 508 Dameca

Drive Gas Block


Item Qty Description Part No.
1 1 Distributer block 10659-41
2 1 Spring 33005-74
3 1 Valve 33005-73
4 1 Valve seat 33005-72
5 1 Membrane, black 33005-71
6 1 Friction ring 33005-75
7 1 Spring 33005-76
8 1 Nut 10967-29
9 1 Nut 36083
10 1 Adjustment screw 10967-291
11 2 Quick connector straight 36690-10
12 2 Quick connector angled, ø6/1/8 36690-20
13 1 Quick connector 31405
14 1 Alu. washer 36594
15 1 Stopper 36670-18

9-22 Oct. 08
Dameca Service Manual for Dameca MRI 508

qusjob42.10659-40 manual.png

Fig. 9-12 Drive gas block

Oct. 08 9-23
Service Manual for Dameca MRI 508 Dameca

Backbar Selectatec
Item Qty Description Part No.
1 1 Selectatec back bar block 11080-651
2 4 Valve 11080-70
3 2 Cover plate 11080-66
4 2 Code pin 11080-53
5 2 Spring 11080-59
6 4 Philips screw M2.5x6mm 70273-256
7 2 Allen point screw 70135-06
8 1 Label 11338
9 2 Quick connector angled 36690-20
10 1 Quick connector T-piece 36690-18
11 1 Tube connector 36690-60
12 4 Philips screw M3x8mm 70276
13 4 Screw 70056

9-24 Oct. 08
Dameca Service Manual for Dameca MRI 508

12

4 7

2
8

13

Use for 10651-00 (Rotameter) only


9

11 Use for 10650-00 (TS) only

10

qusjob42.11080-650s manual.pdf

Fig. 9-13 Backbar Selectatec

Oct. 08 9-25
Service Manual for Dameca MRI 508 Dameca

Backbar Dräger
Item Qty Description Part No.
1 1 Back bar block Dräger 11060-651
2 4 Screw 70056
3 2 Quick connector angled 36690-20
4 2 Screw DIN 912 M5X30 70017
5 2 Allen point screw 70135-06
6 4 Valve 11080-70
7 2 Lock 11066-62
8 4 Screw "70273-02,505"
9 1 Tower for back bar 11060-653
10 2 Pin 11060-654
11 1 Arm 11060-652
12 2 Screw DIN 912 M4X12 70009
13 2 Cover cap 36653-09
14 1 Spring 11060-659
15 1 Ball ø4 mm 30122-08
16 1 Lock pin 11060-658
17 1 Handle 11060-655
18 1 Label 11338
19 1 Quick connector T-piece 36690-18
20 1 Tube connector 36690-60

9-26 Oct. 08
Dameca Service Manual for Dameca MRI 508

13

12
16

11
17
14

9
10 15

18

2
5

4
3
Use for 10651-00 (Rotameter) only

20
Use forl 10650-00 (TS) only
19

qusjob42.11060-650s manual.pdf

Fig. 9-14 Backbar Dräger

Oct. 08 9-27
Service Manual for Dameca MRI 508 Dameca

Inlet Manifold O2
Item Qty Description Part No.
1 1 Inlet manifold O2 31350-05
2 1 Filter nut 31350-06
3 1 Nut for check valve 31350-07
4 1 Filter housing 11213
5 1 Valve housing w/filter w/non-return valve 11215
6 1 Check valve 11214
7 1 Zeeger circlip ring 36735-09
8 3 Quick connector 31405
9 5 Quick connector angled 36690-20
10 1 Label O2 White 11336
11 1 Retainer nut 31381-03
12 1 Hose connector NIST O2 31381-02
13 1 Spring 33005-74
14 1 Valve seat 33005-72
15 1 Membrane, black 33005-71
16 1 Friction ring 33005-75
17 1 Spring 33005-76
18 1 Nut 10967-29
19 1 Adjustment screw 10967-291
20 1 Nut 36083
21 1 Valve 33005-73
22 1 O-ring Ø16,1x1,6 36256
23 3 Allen point screw 70135-05
24 1 O-ring 36305
25 1 O-ring Ø9,1x1,6 36271
26 1 Filter bronze, Ø8x2mm 36701
27 1 O-ring Ø5,1x1,6 36269
28 1 Tube connector 36690-60
29 1 Quick connector T-piece 36690-18

9-28 Oct. 08
Dameca Service Manual for Dameca MRI 508

19

20

18
8
11

12
17
Loctite 270
24 16

7 15
29
4 14

8
28 23
21
Filter facing inwards
13 Loctite 270
Loctite 270
27

9 Loctite 270
26
8

10
6
22
2
1 5
25
9 3

8-2006 11:51 31350 Manual.SLDDRW

qusjob42.31350 manual.pdf

Fig. 9-15 Inlet manifold O2

Oct. 08 9-29
Service Manual for Dameca MRI 508 Dameca

Inlet Manifold AIR


Item Qty Description Part No.
1 1 Inlet manifold AIR 31350C-05
2 1 Filter housing 11213
3 1 Check valve 11214
4 1 Zeeger circlip ring 36735-09
5 1 Quick connector 31405
6 2 Quick connector angled 36690-20
7 1 Label AIR (black/white) 11336C
8 1 Filter nut 31350-08
9 1 Retainer nut 31381-03
10 1 Hose connector NIST AIR 31381C-02
11 1 O-ring Ø16,1x1,6 "O-ring 1,6mm"
12 1 Allen point screw 70135-05
13 1 O-ring O-ring 2,4mm
14 1 O-ring Ø5,1x1,6 O-ring 1,6mm
15 1 Filter bronze, Ø8x2mm 36701
16 1 Tube connector 36690-60
17 1 Quick connector T-piece 36690-18

9-30 Oct. 08
Dameca Service Manual for Dameca MRI 508

10

13
Loctite 270
4 1 6
12
2

Filter facing inwards

Loctite 270
6

3
5
11
15 8

14

16

17

qusjob42.31350C manual.pdf

Fig. 9-16 Inlet manifold AIR

Oct. 08 9-31
Service Manual for Dameca MRI 508 Dameca

Inlet Manifold N2O


Item Qty Description Part No.
1 1 Inlet manifold N2O 31350A-05
2 1 Filter housing 11213
3 1 Valve housing w/filter w/non-return valve 11215
4 1 Check valve 11214
5 1 Zeeger circlip ring 36735-09
6 2 Quick connector 31405
7 1 Quick connector angled 36690-20
8 1 Label N2O (blue) 11336A
9 1 Filter nut 31350-06
10 1 Retainer nut 31381-03
11 1 Hose connector NIST N2O 31381A-02
12 1 O-ring Ø16,1x1,6 36256
13 1 O-ring 36305
14 1 Nut for check valve 31350-07
15 1 O-ring Ø9,1x1,6 36271
16 1 Filter bronze, Ø8x2mm 36701
17 1 O-ring Ø5,1x1,6 36269
18 1 Quick connector T-piece 36690-18
19 1 Tube connector 36690-60

9-32 Oct. 08
Dameca Service Manual for Dameca MRI 508

10

11 Loctite 270 18

19
13
17
5
16
2

Filter facing inwards


Loctite 270

8
15
14
7
1 4
12
9
3

18-08-2006 12:42 31350A Manual.SLDDRW

qusjob42.31350A manual.pdf

Fig. 9-17 Inlet manifold N2O

Oct. 08 9-33
Service Manual for Dameca MRI 508 Dameca

Inlet Manifold VAC


Item Qty Description Part No.
1 1 Inlet manifold N2O 31350B-05
2 1 Label Red VAC 11336B
3 1 Retainer nut 31381-03
4 1 Hose connector NIST VAC 31381B-02
5 1 O-ring 36305
6 1 Quick connector angled 36690-20

1 2

qusjob42.31350B manual.pdf

Fig. 9-18 Inlet manifold VAC

9-34 Oct. 08
Dameca Service Manual for Dameca MRI 508

Inlet Manifold AGSS


Item Qty Description Part No.
1 1 Inlet manifold AGSS 31350F-05
2 1 Hose connector 31381F-02
3 1 Hose connector 31381F-03
4 1 Alu. washer 36594
5 1 Label VAC (red) 11353
6 1 Hose connector 36170
7 1 Retainer nut 31381F-04
8 1 O-ring 36310

5
3

qusjob42.31350F manual.pdf

Fig. 9-19 Inlet manifold AGSS

Oct. 08 9-35
Service Manual for Dameca MRI 508 Dameca

Whispa Ventilator Spare Parts


Item Qty Description Part No.
1 1 Mounting plate 10659M-01
2 2 Spacer Plate 10652-014
3 2 Rail for drawer 10652M-21
4 1 Electronic box for Dameca MRI 508 10987M-50
5 1 Valve manifold for Dameca MRI 508 10987M-25
6 1 MRI ventilator box 10659M-101
7 1 Lid assembly for Dameca MRI 508 10659M-102
8 1 Flow sensor assembly 10659-24
9 4 Distance bush for flowmeter 10659-27
10 1 Mounting plate for flowsensor, MRI 10659M-28
11 50 Nut DIN 6923 A2 70424
12 2 Battery 12V 2,2Ah lead-acid 40099-62
13 1 Battery holder for Dameca MRI 508 10659M-112
14 1 Battery holder (small) for Dameca MRI 508 10659M-113
15 4 Self sticking rubber knob 24030-03
16 4 EMC gasket 1,5x14x454 23926
17 4 EMC gasket 1,5x14x266 23926
18 4 Screw DIN 912 M4X6, RS 70006
19 2 Screw DIN 912 M5X10 70013
20 1 Strain relief 40061-64
21 2 Nut M3 DIN 934 70439
22 2 Tube 400 mm, ID 22 mm 82792
23 1 Flowmeter cabell for MRI 10659m-84
Not shown 1 Com/power/WD/ cabel between E-box and MMI for MRI 10659M-85
25 1 Angle coupling 31425

9-36 Oct. 08
Dameca Service Manual for Dameca MRI 508










 
 
 
 










 

,WHP 4W\ 'HVFULSWLRQ 3DUW1R

  0RXQWLQJSODWH 0

  6SDFHU3ODWH 

qusjob42.10659M-100 manual.pdf
Fig. 9-20 Whispa ventilator

Oct. 08 9-37
Service Manual for Dameca MRI 508 Dameca

E-Box for Whispa Ventilator


Item Qty Description Part No.
1 1 E-box, base part 10987M-512
2 1 E-box, lid 10987M-52
3 6 Screw ISO 7380 M4x8 RHR 70046-0510
4 3 Tube connector 36151-06
5 3 Nut M6 DIN 934 70451U
6 1 Label for E-box hoses 11420-03
7 1 Fuse box 10987-330
8 1 Spirometry zeroing valve block 10987-810
9 2 Screw DIN 912 M5X8 70017-0508

9-38 Oct. 08
Dameca Service Manual for Dameca MRI 508

 


 


qusjob42.10987M-50S manual.pdf
Fig. 9-21 E-Box for whispa Ventilator

Oct. 08 9-39
Service Manual for Dameca MRI 508 Dameca

Power Supply Box


Item Qty Description Part No.
1 1 Box for MRI power supply, machined 40098-33
2 1 Power box assembly for Dameca MRI 508 40098-32
3 1 Appliance 250VAC w. filter/fuse/switch 40075-70
4 4 Lock washer 5 mm 70477
5 3 Nut M5 DIN 934 70441
6 1 Label: WARNING Gauss/Fix power... 11333-13
7 1 Label: Power Supply Box for Dameca MRI 508 11472
8 1 LABEL: T2A 230VAC 11395-05
9 1 Label: Output 3A 24VDC 11395-22
10 2 Fuse T2A 40055-2AT
11 4 Screw RHR M3 X 8MM A2 70046-0308
12 8 Dual lock 3M, SJ356D, 50mm 23950
13 5 PI-filter for power cable 40098-52
14 1 Receptable assembly 40053-661
15 1 Mounting ring for AMP-plug 40098-25
16 5 Crimp contact. female, CPC plug 40053-6611
17 1 Filter box for MRI-power supply box 40098-341
18 2 Torx screw 70130-0306
19 1 Mains cable, Schuko 40045-16
20 2 Screw DIN 912 M3x18 70001-0318
21 2 Screw DIN 912 M3X6 70001-0306
22 4 Screw DIN 912 M3X8 70001-0308
23 1 Strain relief, M3 40061-62
24 1 Strain relief 40061-61
25 1 Strain relief plate 40098-02
26 1 Power supply cable 10651M-83

9-40 Jan. 10
Dameca Service Manual for Dameca MRI 508

7
6

19 2
8 23 20 10 3
5

24 22
13
25 15
18 14 17
21
9

11

1 26

12

qusjob42.40098-30 manual.pdf

Fig. 9-22 Power supply box

Jan. 10 9-41
Service Manual for Dameca MRI 508 Dameca

Valve Block Assy. for Whispa


Item Qty Description Part No.
1 1 Valve block for Whispa 10987-26
2 1 Cover 10967-406
3 1 Retaining clip 10967-401
4 1 Coil 10967-402
5 1 Solenoid base sub-assembly 10967-403
6 1 Core assembly & rider ring 10967-404
7 1 Spring, core 10967-405
8 1 O-ring 10967-407
9 1 Adjustment screw 10967-408
10 4 Screw M3x6 10967-409
11 1 Valve seat 10967-28
12 1 O-ring Ø8,1x1,6 EPDM 36304
13 1 O-ring Ø27,1x1,6 EPDM 36295
14 1 Adaptor for PEEP-valve 10987-506
15 1 Sol. base sub-assembly 10987-501
16 2 Spring washer 10967-609
17 1 Coil 10987-510
18 1 Plugnut assembly 10987-503
19 1 Stem 10987-507
20 1 Disc 10987-508
21 1 Spring 10987-509
22 1 O-ring 10987-511
23 1 Seal for PEEP-valve 10987-504
24 1 Washer for Peep-valve 10987-505
25 2 Core 10967-612
26 1 Screw for adjustment 10987-513
27 1 Zeeger circlip ring 10987-514
28 1 Bonnet 10967-613
29 1 Solenoid base sub-assembly 10967-610
30 1 Plugnut 10967-611
31 1 O-ring 10967-616
32 1 O-ring 10967-615
33 1 Coil 10967-601
34 1 Spacer 10967-607
35 1 Retainer clip 10967-608
36 1 Diaphragm/support assembly 10967-603
37 1 Disc holder assembly 10967-606
38 1 Disc holder spring 10967-605
39 1 O-ring 10967-614
40 4 Screw DIN 912 M6X16 70020
41 4 Nylon washer 70469

9-42 Oct. 08
Dameca Service Manual for Dameca MRI 508











 


 


 
 

 






  

 


 
/RFWLWH 
 
 

/RFWLWH  


 /RFWLWH




 
/RFWLWH

 
/RFWLWH 
 
 
 
 


 

 

  
 
 
 
 

 

7RUTXH1P/RFWLWH











qusjob42.10987M-25S
Fig. 9-23 Valve block assy. for Whispa

Oct. 08 9-43
Service Manual for Dameca MRI 508 Dameca

Valve Block Assy. for Whispa (continued)


Item Qty Description Part No.
42 1 Velve seat 10967-27
43 6 Allen point screw 70135-06
44 2 ISO22 female cone 11890-103
45 3 O-ring Ø24x1,6 EPDM 36242
46 3 Change over valve 10987-801
47 3 Screw M1,7x17 10987-802
48 3 Gasket 10987-803
49 3 Plug with cable for 10987-800 10987-804
50 1 Quick connector 31405
51 2 Silicone gasket 36498
52 1 Filter 36709
53 1 Teflon gasket 36499
54 1 Support for filter 10995-502
55 1 Pressure switch 40083-10
56 2 O-ring Ø10,1x1,6 EPDM 36306
57 6 Alu. washer ø8/ø5 x1 36596
58 6 Tube connector 36151-0405
59 1 O-ring Ø21,95x1,78 EPDM 36288
60 1 Friction ring 33005-75
61 1 Membrane, black 33005-71
62 1 Adjustment screw 10967-291
63 1 Nut 36083
64 1 Spring 33005-76
65 1 Valve seat 33005-72
66 1 Spring 33005-74
67 1 Valve 33005-73
68 1 Nut 10967-29
69 1 Stopper 11018-56
70 1 Label: Batch YY-WW-X 11385-02
71 1 MPL valve housing 10655-61
72 1 MPL valve nut 10655-62
73 1 Silicone gasket 51220-15
74 1 Valve flap 12350-06
75 1 Spring 51220-42
76 1 O-ring Ø18,1x1,6 EPDM 36267
77 2 Angle cone ISO22 male, with O-ring 13910
78 1 Quick connector straight, ø6/1/8" 36690-10
79 1 Filter 11213
Not shown 1 Label for valves 10987-81

9-44 Oct. 08
Dameca Service Manual for Dameca MRI 508











 


 


 
 

 






  

 


 
/RFWLWH 
 
 

/RFWLWH  


 /RFWLWH




 
/RFWLWH

 
/RFWLWH 
 
 
 
 


 

 

  
 
 
 
 

 

7RUTXH1P/RFWLWH











qusjob42.10987M-25S

Fig. 9-23 Valve block assy. for Whispa

Oct. 08 9-45
Service Manual for Dameca MRI 508 Dameca

Valve Block Assy. for Whispa


Item Qty Description Part No.
1 1 Valve block for Whispa 10987-26
2 1 Inspiratory valve 10967-400
3 1 Valve seat 10967-28
4 1 O-ring Ø8,1x1,6 EPDM 36304
5 1 O-ring Ø27,1x1,6 EPDM 36295
6 1 Adaptor for PEEP-valve 10987-506
7 1 PEEP valve 10987-500
8 1 Expiratory valve 10967-600
9 4 Screw DIN 912 M6X16 70020
10 4 Nylon washer 70469
11 1 Velve seat 10967-27
12 6 Allen point screw 70135-06
13 2 ISO22 female cone 11890-103
14 3 O-ring Ø24x1,6 EPDM 36242
15 1 Pilot valve 10987-800
16 1 Quick connector 31405
17 2 Silicone gasket 36498
18 1 Filter 36709
19 1 Teflon gasket 36499
20 1 Support for filter 10995-502
21 1 Pressure switch 40083-10
22 2 O-ring Ø10,1x1,6 EPDM 36306
23 2 Alu. washer ø8/ø5 x1 36596
24 2 Tube connector 36151-0405
25 1 O-ring Ø21,95x1,78 EPDM 36288
26 1 Friction ring 33005-75
27 1 Membrane, black 33005-71
28 1 Adjustment screw 10967-291
29 1 Nut 36083
30 1 Spring 33005-76
31 1 Valve seat 33005-72
32 1 Spring 33005-74
33 1 Valve 33005-73
34 1 Nut 10967-29
35 1 Stopper 11018-56
36 1 Label: Batch YY-WW-X 11385-02
37 1 MPL valve 10655-60
38 2 Angle cone ISO22 male, with O-ring 13910
39 1 Quick connector straight, ø6/1/8" 36690-10
40 1 Filter 11213
41 1 Label for valves 10987-81

9-46 Oct. 08
Dameca Service Manual for Dameca MRI 508



  


 

/RFWLWH

 

/RFWLWH
/RFWLWH  


/RFWLWH

 



 
/RFWLWH /RFWLWH




 

/RFWLWH 
 
/RFWLWH


/RFWLWH  



/RFWLWH

   

 /RFWLWH
 
 

/RFWLWH

/RFWLWH

/RFWLWH

/RFWLWH

7RUTXH1P

qusjob42.10987M-25 .pdf

Fig. 9-24 Valve block assy. for Whispa

Oct. 08 9-47
Service Manual for Dameca MRI 508 Dameca

Rear Cover
Item Qty Description Part No.
1 1 Rear cover for Dameca MRI 508 10651M-13
2 1 Fan filter complete 36715
3 4 Screw 70058
4 4 Toothed lock washer ø4 70476
5 4 Nut M4 DIN 934 70440









qusjob42.10651M-12 manual.pdf

Fig. 9-25 Rear cover

9-48 Oct. 08
Dameca Service Manual for Dameca MRI 508

Flowmarker Assembly
Item Qty Description Part No.
1 1 Housing for flow marker 13680-01
2 1 Holder for flow marker 13680-02
3 1 Flowmarker 25 L/min 13676
4 1 Piston for valve, green 30325-10
5 1 Piston for valve, red 11014-11
6 2 Quick connector 36690-12
7 2 Nipple for hose 10894-62
8 3 O-ring ø7,1x1,6 36268
9 1 O-ring ø12,29x3,53 36247
10 1 O-ring ø10,69x3,53 36246
11 2 O-ring ø17,17x1,78 36248

Loctite 242

5 8 6 7 1 8 8 4

Loctite 242

3
6

Loctite 242

10

11

11

qusjob42.13680 manual.pdf

Fig. 9-26 Flowmarker assembly

Oct. 08 9-49
Service Manual for Dameca MRI 508 Dameca

Suction Unit
Item Qty Description Part No.
1 1 Body for integrated suction 32525-03
2 1 Diaphrame house 32525-02
3 1 Diaphrame for suction, red 32515-13
4 1 O-ring 36288
5 1 House for valve 32515-12
6 1 Bottum nozzle 32515-21
7 1 O-ring Ø13,1x1,6 36272
8 1 O-ring 36281
9 1 Top nozzle 32515-17
10 1 O-ring Ø9,1x1,6 36271
11 1 Middle nozzle 32515-19
12 1 Bushing for silencer 32515-22
13 1 House for silencer 32525-04
14 2 Silencer 32010-22
15 2 Screw DIN 912 M5X8 70017-0508
16 1 O-ring Ø14,1x1,6 36307
17 1 Diaphram, yellow 32515-14
18 2 Quick connector angled, ø6/M5 36690-22
19 1 Quick connector angled, ø6/1/8 36690-20
20 2 Washer 70468

9-50 Oct. 08
Dameca Service Manual for Dameca MRI 508

19

2
18

17

16

1
20
15

8
9
10
11
7
6
12

14

14

13

qusjob42.32530-100 manual.pdf

Fig. 9-27 Suction unit

Oct. 08 9-51
Service Manual for Dameca MRI 508 Dameca

Mains Switch, standby


Item Qty Description Part No.
1 1 Pushbutton 40090-71
2 1 Coubling 40090-72
3 1 LED, 24V 40090-75
4 2 Switch, NO 40090-73
5 1 Holder for LED 40090-74

9-52 Oct. 08
Dameca Service Manual for Dameca MRI 508

qusjob42.40090-70 manual.png

Fig. 9-28 Mains switch, standby

Oct. 08 9-53
Service Manual for Dameca MRI 508 Dameca

Vacuum Suction
Item Qty Description Part No.
1 1 House for valve, front part 10653-41
2 1 Housing for vacuum suction 10653-72
3 1 Screw for regulation 10653-43
4 1 Vacuum diaphragm 32550-08
5 1 Cone for diaphragm 32550-07
6 1 Shaft for valve 10653-73
7 1 O-ring 36294
8 1 Zeeger ring 36738
9 1 Spring 32550-15
10 1 Nylon gasket 32550-10
11 1 Button for vacuum 10653-48
12 1 Shaft for valve 10653-44
13 1 Allen screw 70101
14 2 Washer 70467
15 1 Nut M3 DIN 934 70439
16 1 Label VAC, red 11336B
17 1 Label VAC, yellow 11336B-S
18 1 Label Yellow 11361-04
19 1 Nut 10653-47
20 1 Quick connect Ø8-1/8" Viton 36690-14
21 1 Quick connector angled, ø6/M5 36690-22
22 1 Quick connector 90 deg. angel, Ø8x1/8" 36690-21

9-54 Oct. 08
Dameca Service Manual for Dameca MRI 508

qusjob42.10653-70s.png

Fig. 9-29 Vacuum suction

Oct. 08 9-55
Service Manual for Dameca MRI 508 Dameca

AGSS Ejector
Item Qty Description Part No.
1 1 Ejector body 11014-15
2 1 Nozzle 0.6 mm 11014-60
3 1 Filter bronze 36132
4 1 Manifold for ejector 11013-01
5 2 Screw DIN 912 M3X20 70001-0320
6 1 Hose nipple w/1/2-26 10894-63
7 1 O-ring ø7,65x1,78 36309
8 1 Nipple w/M10x1 11014-16
9 1 Spring 5,0/3,4/8,0 10887-27
10 1 O-ring teflon Ø6,07x1,78 36311
11 1 Spindle short M6x0.5 10887-26
12 1 Spindle guide M6x0.5 10887-24
13 1 Retainer w/slot 11004-18

9-56 Oct. 08
Dameca Service Manual for Dameca MRI 508

13 Loctite 270

1
12 8
2
11
3
10 7

Loctite 270

qusjob42.11013 manual-2.pdf

Fig. 9-30 AGSS ejector

Oct. 08 9-57
Service Manual for Dameca MRI 508 Dameca

O2 Flush
Item Qty Description Part No.
1 1 House for O2 flush 10658-31
2 1 Valve shaft for O2 flush 10658-32
3 1 Pushbutton for O2 flush 10658-33
4 1 Label O2+, white 11356
5 1 Spring for O2 flush 10784-07
6 1 O-ring ø4,48x1,78 36281
7 1 O-ring ø2,57x1,78 36279
8 1 Quick connector straight 36690-10
9 1 Label QC 11385-11
10 1 Label: BATCH LABLE YY-WW 11385-01
11 1 Quick connector angled, ø6/M5 36690-22

qusjob42.10658-30 manual.png
Fig. 9-31 O2 flush

9-58 Oct. 08
Dameca Service Manual for Dameca MRI 508

IBS System Complete


Item Qty Description Part No.
1 1 IBS console 11960-80
2 1 IBS system 11960-100


'HIDXOW
47<




qusjob42.11960 manual.pdf
Fig. 9-32 IBS system complete

Oct. 08 9-59
Service Manual for Dameca MRI 508 Dameca

IBS Console
Item Qty Description Part No.
1 1 Absorber lock 11960-60
2 1 IBS base 11960-81
3 3 Friktion washer 11960-82
4 1 Freshgas connector 11960-84
5 1 Connector M18x1 11960-85
6 1 Hinge for absorber 11960-13
7 1 Mount block for switch 11960-93
8 1 Quick connector straight, ø6/1/8" 36690-10
9 1 Piston for changeover valve 11960-87
10 1 Cylinder for changeover valve 11960-88
11 1 Connector Ø22 12064-02
12 2 Dowel pin for base 11960-105
13 2 Lock bolt for base 11960-102
14 1 Mount plate for base 11960-101
15 4 Dowel pin Ø3x10 10991-256
16 1 Banjo bolt M5x14 36681-05
17 2 Alu. washer ø8/ø5 x1 36596
18 1 Freshgas connector 11960-841
19 3 Screw DIN 7991 M4x10 70056
20 2 Screw DIN 912 M4X25 70012
21 2 O-ring, silicone red, ø9,19x2,62 36343
22 1 O-ring, silicone red, ø15,08x2,62 36344
23 2 Screw DIN 912 M3X8 70001-0308
24 2 Screw DIN 912 M4X16 70010
25 2 O-ring Ø27,1x1,6 EPDM 36295
26 2 O-ring Ø24x1,6 EPDM 36242
27 2 Screw DIN 912 M8X25 70026
28 4 Screw DIN 912 M5X10 70013
29 1 Banjo with hose connector 36692-7405
30 1 Quick connect Ø8-1/8" Viton 36690-14
31 1 Lock ring w/handle 11960-140
32 2 Dowel pin 3mm 36753-20
33 2 Spring 5,0/3,4/8,0 10887-27
34 1 lock nut DIN 985 M6 70436
35 1 Pin M6 11960-104
36 4 Screw DIN 912 M2X10 70001-0210
37 1 Mount block for switch 11960-92
38 1 Switch for IBS 40090-330
39 1 Switch for absorber 40090-325

9-60 Oct. 08
Dameca Service Manual for Dameca MRI 508



 

 


  


 


 





 


 /XEULFDWHZLWK0RO\FRWH';

/RFWLWH

 

  
 

 

 


 




/RFWLWH



qusjob42.11960-80 manual.pdf

Fig. 9-33 IBS console

Oct. 08 9-61
Service Manual for Dameca MRI 508 Dameca

IBS System
Item Qty Description Part No.
1 1 IBS body 11960-01
2 1 ISO 22/15 Cone 11960-02
3 5 Plug M18x1 11960-03
4 1 Plug 7/8"-26wg 11960-13
5 1 Bellows 11960-05
6 1 Chamber 11960-09
7 1 Valve pin exp. 11960-11
8 1 Valve pin INSP. 11960-12
9 1 Inspiratory valve for absorber 11960-230
10 1 Expiratory valve for absorber 11960-460
11 1 Changeover valve for IBS 11960-550
12 1 Bag connector 11960-70
13 1 Lid for overflow membrane 51525-10
14 1 APL valve 11960-330
15 5 O-ring Ø15,1x1,6 EPDM 36240
16 2 O-ring Ø19,1x1,6 EPDM 36244
17 8 Screw BN 9995, (ISO 14585C), 2,9x16 70128-2,9x16
18 4 Screw BN 9995, (ISO 14585C), 2,9x9,5 70128-2,9x9,5
19 2 One-way valve 11890-31
20 1 O-ring ø49,6x2,40 36261-4924
21 3 Dowel pin Ø2 x 10 36757-10
22 1 Release button for APL valve 11960-77
23 6 O-ring Ø18,1x1,6 EPDM 36267
24 1 Membrane 51025-08
25 1 Membrane weight 51025-07
26 2 Spring 4,3/3,3/15mm 11080-62
27 2 Valve dome with o-ring 11890-42
28 1 ISO 22/15 cone 11960-10
29 1 O-ring Ø27,1x1,6 EPDM 36295
30 1 O-ring, silicone blue, ø129,77x3,53 36330

9-62 Jan. 10
Dameca Service Manual for Dameca MRI 508


 


 

 





 





 
 



  



 
   






 

qusjob42.11960-100 manual.pdf

Fig. 9-34 IBS system

Oct. 08 9-63
Service Manual for Dameca MRI 508 Dameca

Left blank intentionally

9-64 Oct. 08
Dameca Service Manual for Dameca MRI 508

10. Schematics
Contents Page

Fig. 10-1 Wiring diagram for ventilator, Old version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2


Fig. 10-2 Wiring diagram for ventilator, New version. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Fig. 10-3 Wiring diagram for ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Fig. 10-4 Wiring diagram for ventilar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Fig. 10-5 Wiring diagram for ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
Fig. 10-6 Pneumatic diagram Dameca MRI 508 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7

Oct. 08 10-1
10-2
Wall Power Supply Unit External cable Internal cable Ventilator
5m
20
24V DC B1
-B1
PIN 5 F5
A (pin 5) 13 16 21
+24V DC +24V DC B2
2.3 2.4
2AT 4AT
Service Manual for Dameca MRI 508

Battery F6
Charger 4AT
MAINS 2AT
24V DC 18

+
24V Battery Pack

PIN 4 B (pin 4) 15 17
0V DC 0V DC

-B1
PIN 3 C (pin 3) F4
4 5 26
+24V DC +24V DC
22
4AT 84
PSU
24V DC
PIN 2 D (pin 2) 85
23 23
0V DC 0V DC 3

24
0V DC 4

25
0V DC 2

PIN 1 GND (pin 1)

Fig. 10-1 Wiring diagram for ventilator, Old version


GND GND

XLR FEMALE XLR MALE

qusjob3.siesta i whispa mri wiring a3.pdf

Oct. 08
Dameca
Oct. 08
Dameca

Wall Power Supply Unit External cable Internal cable Ventilator


5m
20
24V DC B1
-B1
PIN 1 F5
PIN 1 13 16 21
+24V DC +24V DC B2
2.3 2.4
2AT 4AT
Battery F6
Charger 4AT
MAINS 2AT
24V DC 18

+
24V Battery Pack

PIN 3 PIN 3 15 17
0V DC 0V DC

-B1
PIN 2 PIN 2 F4
4 5 26
+24V DC +24V DC
22
4AT 84
PSU
24V DC
PIN 5 PIN 5 85
23 23
0V DC 0V DC 3

24
0V DC 4

25
0V DC 2

PIN 7 GND (PIN 7)

Fig. 10-2 Wiring diagram for ventilator, New version


GND GND

CPC CPC

10-3
qusjob3.siesta i whispa mri wiring a3 new.pdf
Service Manual for Dameca MRI 508
Service Manual for Dameca MRI 508 Dameca

Fig. 10-3 Wiring diagram for ventilator

10-4 Oct. 08
Dameca Service Manual for Dameca MRI 508

Fig. 10-4 Wiring diagram for ventilator

Oct. 08 10-5
Service Manual for Dameca MRI 508 Dameca

Fig. 10-5 Wiring diagram for ventilator

10-6 Oct. 08
Dameca Service Manual for Dameca MRI 508

48
61
22 46 45
58
56 59 68
60 70 49
57 67

47

55 69
54 66 77
52
44
53 50
51
72 73

25 71
26 65 76 82

24 64
63
74 78
23
8 62
4
1
7 9 79 81

13
11 80 75
12 20
11 85
28 17 32 40
10 43 21 83 84
3
7 9 16 31 39 18
33 15 38
34
37

5 14 30 36
7 9
19

35 29

qusjob42.10650-whispa-mri-pne-manual.png
Fig. 10-6 Pneumatic diagram, Dameca MRI 508

Oct. 08 10-7
Service Manual for Dameca MRI 508 Dameca
Dameca Service Manual for Dameca MRI 508

11. Technical Data

Note:
All flow and volume measurements are related to STPD conditions
(Standard Temperature Pressure Dry).

11.1 General
Dimensions:
Height : 1510 mm
Width : 760 mm (including IBS)
Depth : 670 mm
Weight: : Approx. 110 kg (depending on configuration)

MRI Compatibility:
Max. magnetic field : 1000 Gauss

Electrical:
Class :I
Type :B
Mains voltage : 100/115/230 volt 50/60 Hz.
Power consumption (typical) : 80 VA
Battery capacity : 2.2 Ah (ventilator)
Battery operation time : App. 60 minutes (only by mains failure)
Battery type : Pb rechargeable
Charge time : Approx. 12 hour

Note:
There are no changes in performance when the ventilator is battery powered.

11.2 Gases

Note:
All gases and anaesthetic agents must conform to the European Pharmacopeia.

Oct. 08 11-1
Service Manual for Dameca MRI 508 Dameca

Central gas supply:


Inlet pressure : 4-6 (x100kPa) for O2, N2O and AIR

Spare Gas Supply (optional): : ONLY MRI COMPATIBLE


Pin index for 4 ltr cylinders (O2 and N2O) : Gas cylinders must be used

Pressure gauges:
Central gas supply : 0-10 (x100kPa) for O2, AIR and N2O
Gas cylinder (optional) : 0-315 (x100kPa) for O2
: 0-100 (x100kPa) for N2O

Auxiliary outlets (optional):


Flow for O2 or AIR : 100 L/min at 4 (x100kPa) static pressure
: 140 L/min at 6 (x100kPa) static pressure
: >300 L/min at 12 (x100kPa) static pressure

AGSS:
Scavenging flow : 30 to 40 L/min

Gas Alarm:
Alarm start-pressure for O2 : 2.5 (x100kPa)

Auxiliary freshgas outlet (optional):


Max. pressure limiting valve : 90 (x100Pa)

Auxiliary O2 flowmeter (optional):


Flow range : 0 to 10 L/min
Accuracy : 7.5% actual flow + 2.5% full scale

Patient suction unit (optional):


Max. vacuum : -0.8 (x100kPa)
Max. suction flow : 35 L/min

11.3 Rotameterbox
Rotameters:
Measuring range : 0.1-10 L/min for O2, AIR and N2O
optional: 0.05-1 L/min for O2 and N2O
1-10 L/min for O2 and N2O

Accuracy : +/-(1.9% actual reading +0.6% full scale)

11-2 Oct. 08
Dameca Service Manual for Dameca MRI 508

Minoxyguard:
Min. O2 concentration in N2O/O2 mix : 25% O2
Min. O2-flow in order for N2O-flow to open : app. 0.3 L/min O2

O2 flush flow:
Fixed flow : App. 45 L/min

11.4 Ventilator
Driving gas (AIR or O2):
Pressure : Min. 3.0 (x100kPa) at 80 L/min.
Max. consumption (peak flow) : 120 L/min
Average consumption : Max. 80 L/min at 2.8 (x100kPa)

Pressure range:
Pressure limitation (Plim max) : 90 (x100Pa) (mechanical pressure limitation valve)
Max. adjustable working pressure : 80 (x 100Pa)
High pressure alarm : 10 to 80 (x100Pa)
Min. expiration pressure (Plim min) : +1 (x100Pa)
Cycling pressure : 1 to 80 (x100Pa)

Note:
Negative airway pressures are possible if the patient is breathing spontaneously only.

Set parameters:
Tidal Volume : 20 to 1500 ml.
Accuracy : 20 to 250 ml: +/-10%, min. 10 ml
: 250 to 1500 ml: +/-5%, min. 25 ml
Minute volume : 0.2 to 60 L/min.
Frequency : 4 to 80 resp./min.
Accuracy : +/- 5 %
I:E Ratio : 3:1 and up to 1:9,9
PEEP : OFF, 4 to 20 (x100Pa)
Inspiratory pressure : 10 to 70 (x100Pa)
Inspiratory pause : 0 to 70 %
Ventilation mode (controlled) : VCV, SIMV (option), PCV, PSV (option), PRVT (option)

Oct. 08 11-3
Service Manual for Dameca MRI 508 Dameca

SIMV settings:
Trigger point : -0.5 to -10.0 (x100Pa)

PSV settings:
Support pressure : 4 to 50 (x100Pa)
Inspiratory trigger : 1 to 10 L/min.
Expiratory trigger : 10 to 80 %
PSV backup : 10 to 40 sec.

Test mode : LC TEST, FULL TEST

Sigh settings:
Interval : 50 to 150 respirations
Added tidal volume : 30 to 80 %

Note:
The sigh function is active in VCV and SIMV mode only.

Airway pressure monitor:


Measured parameters: : PEAK, Plateau, PEEP, Compliance
(if vol. monitor is installed)
Pressure range : -10 to 99 (x100Pa)
Accuracy : +/-2 (x100Pa)

O2 % fuel cell monitor (optional):


Measuring range : 0 to 100 % O2 (V/V)
Accuracy : +/- 2 % O2 (V/V) at constant temperature and pressure
High O2% alarm : 19 to 100% and OFF
Low O2% alarm : 18 to 100%
Expected O2 sensor lifetime : Over 500.000 O2 percent hours under normal
conditions (equivalent to 33 months, when placed
in AIR at 25OC)
Cross interference : less than 1.25% O2 response to anaesthetic agents
Drift : less than 1% O2 over 24 hours
Rise time : less than 6 seconds for 90% of final value

Volume monitor (optional):


Measuring range Paed. sensor : 20 to 300 ml (insp./exp. flow: 2-35 L/min.)
Measuring range Adult sensor : 200 to 1500 ml (insp./exp. flow: 10-120 L/min)
Accuracy Paed. sensor : 20 to100 ml: +/- 10 ml
100 to 300 ml: +/- 10% of actual value
Accuracy Adult sensor : 200 to 500 ml: +/- 50 ml
500 to 1500 ml: +/- 10% of actual value
High expired minute volume alarm : 0.1 to 80.0 L and OFF
Low expired minute volume alarm : 0.0 to 79.9 L and OFF

11-4 Oct. 08
Dameca Service Manual for Dameca MRI 508

11.5 Integrated Breathing System


Dimensions:
Size : ø160 x 360 mm
Weight : 4 kg (complete system excl. i-SORB CO2 absorber)
Respiration valve:
Setting : SP, 5 to 75 (x100 Pa)
Accuracy : +/-7 (x100Pa) at 4 L/min.
i-SORB CO2 absorber:
Capacity (sodalime) : Approx. 900 gr.
Volume (empty) : 1420 ml.

11.6 Environment
Storage temperature : -20 to + 50OC (optional O2-sensor: 0OC to +50OC)
Working temperature : 15 to 35OC
Relative humidity : 10 to 90% RH (non condensing)
Storage pressure : 630 to 1060 (x 100Pa)
Working pressure : 960 to 1060 (x 100 Pa)

11.7 Standards
Standards fulfilled:
Software : EN 60601-1-4
EMC : EN 55011; Group 1; Class B
Electrical safety : EN 60601-1
Function safety : EN ISO 10524-1, EN ISO 5359, EN 740,
EN 837-1, EN 980, EN ISO 5356-1
Oxygen monitor (optional) : EN ISO 19054, EN ISO 21647, EN ISO 15002

Classification according to EN 60601-1:


Class I Equipment : according to the type of protection against
electrical shock.
Type B Equipment : according to the degree of protection against
electrical shock.
Continuous Operation : according to the mode of operation

Classification according to EU Medical Device Directive 93/42/EEC:


The Dameca anaesthesia machine model Dameca MRI 508 is classified as IIb