Desktop Automatic Processor

9000
INSTRUCTION MANUAL
DAFTAR ISI

1. Perkenalan
2. Safety Issues (please keep follow all the instructions) 2
3.Safety Precautions
3-1 Power and Earth Cables
3-2 Environment
3-3 Films and Chemical Solutions
4. Parts Name
5. Optional Parts
6. Outer Measurement (mm)
7. Installation
7-1 Preparation for Installation
7-2 Installation
7-3 Handling the Waste Chemical Solution and Drain Hose
7-4 Connecting the Water Hose
7-5 Power Supply
8. Preparation for Operation
8-1 Cleaning the Developing and Fixing Racks
8-2 Mixing the Developing Solution
8-3 Mixing the Fixing Solution
8-4 Pouring the Chemical Solution
8-5 Setting the Racks
9. Operating Instructions
9-1 Power Switch and Water Supply
9-2 Developing the Films
9-3 Turning off the Power
10. Operating Explanation
11. Maintenance and Cleaning
11-1 Daily Inspection
11-2 Monthly Maintenance
11-3 Changing the Chemical Solution
11-4 Periodical Inspection
11-5 List of Consumables
11-6 Inspection Check Sheet
12. Damage and Inspection
13. Basic Spec
14. Option Parts
14-1 Stand
14-2 Water Solenoid Valve
14-3 Water Filter Kit
14-4 Water Pump Unit
15. Warranty and After Service

1
1. Introduction

Thank you very much for your purchase of desktop compact automatic film processor,
ECOMAT9000.
This manual includes the basic instructions of daily use, maintenance, and countermeasures
during the troubles of ECOMAT9000.

It is important that all the instruction, description and safety issues are reviewed thoroughly
before using the equipment. After you finish studying the manual, please keep this manual with
the warranty card.

2. Safety Issues (please keep follow all the instructions)

Please read the "safety issues" thoroughly before using the equipment.
The following "safety issue" includes information on proper use to avoid damage to the
equipment and human (users). There are many different diagrams and signs showing different
meanings and all of them have important meanings. Please follow them carefully.

Warning Please do not ignore this sign. It may cause serious injuries or
Warning death to the user.
Warning Please do ignore this sign. It may cause injuries and damage to
Warning the equipment.

Sample of signs and diagrams:


This shows that you will need to pay attention to the warnings. There are actual
photos showing samples of safety issues.
This shows that the action is prohibited. There are actual photos showing samples
of safety issues.
This shows that the action is required. There are actual photos showing the safety
issues (the left photo shows a plug being pulled from an outlet)

Warning
Do not open the top cover while the equipment is running. The equipment includes
motors which may cause injuries if you touch it. The chemical solution may get into
your eyes or clothe.
Please turn off the power during maintenance. It may cause electric shock.

Make sure the earth cable is inserted. It may cause electric shock if it is not inserted
properly.
Use only 200V. It may cause fire and electric shock.

For any trouble or damage, stop the operation, pull out the power cable and contact
your local sales agent or distributor. If operation continues during any trouble, it may
cause fire or electric shock.

2
Warning
Do not splash any liquid to the equipment. Do not install the equipment where water
is present. It may cause damage and electric shock.
During operation, please make sure the equipment is stable and do not shake the
unit. It may cause damage, electric shock or fire.
Only a authorized engineer should open and fix the unit. It may cause fire, electric
shock or injuries unless you are trained. Please contact your local agent or
distributor.
Please do not modify the unit. It may cause fire, electric shock and injuries.

Warning
Please use only a 200V, 10A cable. Avoid multiple plug connection. It may cause
fire, trouble or injuries.
For installation, find a flat location where there is no vibration. If it is not properly
installed, the different chemical solution may get mixed up and cause trouble.
Make sure to thoroughly wash off the chemical solution in case you come in contact.
After handling the chemical solution, please wash hands. If the chemical solution
gets inside the eyes, wash thoroughly with water and consult with a doctor.
Make sure there is enough ventilation. If there is not enough ventilation, it may
cause trouble.
Do not place any objects on top of the power cable. It may cause fire.

Do not place any objects on top of the unit. It may cause damage, fire or injuries.

Place the power cable so that it may not be accidently pulled out. It may cause
injuries or damage.
When the power cable is pulled, do not pull the cord. Hold the power plug to pull it
out. It may cause trouble or electric shock.
During transportation, turn off the power, pull out the power cable and take out all
the chemical solution. Make sure the unit does not turn over. It may cause damage
and injuries.
Do not use mobile phone or any other device which generates electromagnetic field.
It may cause damage to the unit.
The unit is not explosion and fire proof. Do not use the unit where there are
flammable products.
In case the equipment is not used for a long time, turn off the power and pull out the
power cable. It may cause damage or fire.
If the unit is used after the power has been turned off for a long time, make sure the
unit is operating properly. It may cause damage or fire.
Do not turn the unit side ways during transportation. It will cause damage.

This equipment weights about 30kg. Make sure to be careful when opening the
package and handling the unit. It may cause injury and damage to the unit.

3
4. Parts Name

3-1 Power and Earth Cables

Power ; 220/240V,4.5A ( or 110/120V ,9A )

Make sure to connect to the Type 3 ground cable.

4
%RH

3. Safety Precautions 30 80
3-2 Envirnment
Requested conditions: 10 30 5
Temp:10 30
Humidity:30 80 RH
Water Temp: 5 Above

The dark room will require a ventilation fan and exhaust pipe. Dark Room lamp should be diode
type S2 (ortho regular)

Warning
Do not place any unnecessary objects inside the dark room.

3-3 Proper Films and Chemical Solution

Film Ortho regular


No specific brand of films
(please contact your local dealer for more information)
Chemical Solution Dev solution
Fix solution
Mixing method 8-2 Mixing method
8-3 Fix Solution mixing method

Warning on chemical solution


The Developing Solution has a high percentage of alkali, and fixing solution contains acidic
liquid.
Make sure they do not come in contact with human skin and eyes. The used chemical solution
(waste) needs to be disposed accordingly the regulation of your country. Please contact your
waste disposal agent for handling the chemical solution waste.

5
4-1 Name of Parts A

Insertion case
Insertion case lid
Top cover
Drying cover
Rubber stand
Power switch POWER
Ready lamp READY
Front cover
4
B
l
D
E
C
C
r
D
r
R
D
R
r
D
e
r

Measuring Cup
1

R
e
p
l
e
n
i
s
h
i
n
g

t
a
n
k

(
1
0
L
)

2
F
o
r

D
E
V

S
o
l
u
t
i
o
n s
o
a l
n u
d t
i
F o
I n
X
5. Optional Parts

Spare parts

No.
1 2
1 117-3730005 O-ring ,P12 4
2 117-3730006 O-ring,P18 2

3 4 5 3 117-3270045 Gear,rightA 2
4 117-3270046 Gear,leftA 2
5 117-3270047 Gear,leftB 1
6 117-3850002 Spring,roller 4
6 7 8 7 117-3850001 Spring,roller 4
8 117-3220011 Support,shaft 2
9 117-3610051 Fook,right 1
9 10
10 117-3610052 Fook,left 1
565

109 614

723
Plane view

723

463

1013
655
550
570

Front view Side view


7-1 Preparation for Installation

This equipment is designed to be installed in a complete dark room. Check the measurement of
the equipment and keep enough room and space in the dark room, and also keep enough
space to work for changing the films. Make sure there is a power outlet for 100V - 200V, 10A.
Make sure there is water supply, and drainage. In case you purchase an optional water
solenoid valve, water filter unit, water pump unit, make sure there is enough space, installation
space and additional power outlet.
(For more details on the optional parts, please refer to page 23).

7-2 Installation
Please conduct the installation inside the dark room only. Please adjust the height position of
the equipment by turning the rubber of the stand to make sure it is parallel to the ground.

Ventilation fan

Dark room

Exhaust exit

Warning
If the main unit is tilted, the developing solution and fixing solution may get mixed up inside the
unit and over flow. Please make sure the unit is installed flat to the ground.
Please refer to the following recommended environment for installation
TEMP 10 30
HUMIDITY 3 0 80 RH
WATER TEMP 5 or above
7. Installation
Make sure there is ventilation fan and exhaust exit inside the dark room.

7-3 Handling the Waste Chemical Solution and Drain Hose

Connect each of the waste and water hose. Insert the Waste Hose to the waste drain. Collect
the chemical solution waste by connecting the waste hose to the tank. Please cut the hose to
the necessary length. Ask the waste disposal agent to collect the chemical solution waste.
Warning
Make sure the hoses do not bend in a U shape. Please check the flow of the waste is smooth. If
there is any clog inside the hose, it may cause damage to the main unit.

7-4 Connecting the Water Hose

Tap water

Water coupling

Hose joint

Loosen the hose joint and insert the water coupling to the tap water.
Use the 3 screws and hold the water coupling to the tap water
Check the length of the water hose and cut to the right length.
After completing the connection, cut to the hose to the necessary length and dispose the
remains.
After the hose is connected, run the water to check if there is no leak.

Warning
Do not tighten the hose joint to tightly. The bucking may break and cause water leakage.

Warning
The water coupling comes with a continuous flow valve. (less than 2 liter / min) In case a
different type of water hose is used, please use this water coupling.

7-5 Power Supply


Make sure to use 220V/240V,5A ( or 110/120V , 10A)
Make sure to connect to the Type 3 ground cable.

Warning
Make sure to hold onto the plug part when pulling out the power plug. Do not pull the cord.
8. Preparation for Operation

8-1 Cleaning the Developing and Fixing Racks


Clean each of the racks with water and wipe off the dirt.
Soak each of the rack and cross over guide into a bucket of warm water. Wipe off any dirt on
the rollers and guides with a piece of cloth.

8-2 Mixing the Developing Solution


Use the attached Measuring Cup and pour 4 liter of water
Take out the Solution A and add to the 10 liter tank.
Take out the Solution B (brown bin) and add to the 10 liter tank.
The total volume should be 10 liters. Make sure to mix the solution thoroughly.

10
Measuring Cup Solution A
Solution B

10 liter tank

Warning
When handling the chemical solution, make sure to use rubber gloves to avoid contact with the
solution. Clean your hands thoroughly afterwards.

Both Solution A and Solution B is highly concentrated solution, and may cause crystallization
when the room temperature is cold. In case the solution is crystallized, mix the solution with
warm water and pour into the 4 liter pot.

Warning
The chemical waste needs to be collected using a waste tank and dispose it accordingly to your
the country's regulation.

8-3 Mixing the Fixing Solution

Use the attached Measuring Cup and pour 4 liters of water.


Use the A bottle and add to the 10 liter tank.
Use the B bottle and add o the 10 liter tank..
Add more water to make the total volume of 10 liters and mix the solution thoroughly.

11
Measuring Cup
Solution A
Solution B

10 liter tank

Warning
When handling the chemical solution, make sure to use rubber gloves to avoid contact with the
solution. Clean your hands thoroughly afterwards.

If you mix the A solution with B solution, it will dissolve and become cloudy. Make sure to water
down the solution A and then mix the solution while adding the Solution B.

Warning
The chemical waste needs to be collected using a waste tank and dispose it accordingly to your
the country's regulation.

8-4 Pouring the Chemical Solution


Carefully pour each of the Developing and Fixing Solution into the racks in order using the 4L
pot. Check the solution level line and pour until the solution reaches the line. Both the
developing rack and fixing rack has a 2 liter capacity.

Valve
Insertion line

solution level
9. Operating Instructions

Warning
In case the fixing solution gets inside the developing rack, the final image will be effected.
Make sure to clean the developing rack to get clear image.

Both the developing and fixing solution contains a high percentage of alkali and acidic solution.
Make sure they do not get mixed up inside the unit.

8-5 Setting the Developing Rack

1) Set each racks, developing rack, fixing rack, water rack carefully in each location.
2) Slide each of the racks towards the left side until there is a stopper. The gear will hold the rack.
3) Place the Drying rack on top of the Water Rack, and the air case to the dryer exit.

Drying rack
Drying rack Hook
Washer Rack

Air case

Drying rack

Stay

Water rack Hook

Place the cross over guide to the ratchet in between each of the racks.
Check the arrow mark and place it to the same direction of the film. Listen carefully to the sound
of the gear when it connects. Make sure the cross over guide does not move.
9-1 Power Switch and Water Supply

Warning
Make sure the power plug and the earth plug are connected to avoid any power leakage.
Check inside to make sure the solution is filled. The unit may break if the power is turned on
without any solution.

Turn on the power ON

9-2 Developing the Films

1) Make sure the READY lamp is lit.


2) Use 1 to 2 films for cleaning purpose everyday before start using the equipment.
3) Turn off the light inside the dark room and turn on the dark room lamp.
4) Open the lid of the film insertion, and insert the film accordingly to the left side of the film
guide. In case the width of the film is less than 8 inches, place the film in the middle of the
case.

Warning
Make sure to use the appropriate film and dark room lamp during operation.

5) After inserting the films, the READY sign will start to flash and buzzer will go off. After closing
the lid, you may turn on the light inside the dark room.
6) The READY sign will disappear when the films are inserted.
7) When the READY sign starts to flash, the next film is ready to be inserted.
8) When inserting films continuously, please follow the above instructions, 3 to 7.
9) When the operation is completed, make sure to close the water supply.

9-3 Turning off the Power

1) Turn off the main Power (OFF ).


2) Close the water supply.
3) Pull out the power plug
4) Check the below diagram and open the insertion case, lid and the top cover for better
ventilation.

Insert Case

Warning
If you keep the lid, film insertion case and top cover, closed it will cause damage to the top
cuver unit ,or poor image due to excessive moisture and gas.
11. Maintenance and Cleaning

After Turning on the Power


When the power is turned on, the Developing / Drying Heater, Drying Blower, Drive Motor starts
to run and each and all rollers will turn.

Temperature Control
The heaters inside the developing rack and fixing tank will control the temperature of the
developing solution. Once it reaches the temperature, the ready sign will light and the buzzer
will go off three times. In order to maintain the temperature of the developing solution, the
heater will turn on to control the temperature. The temperature of the Dryer will also stop once it
reaches the set temperature. Once the temperature drops, the temperature control will start
again.

The Film Developing and READY


Status
When the film is being processed, open the insertion case, insert the films according to the film
guide to the right. The film sensor located at the middle of the film entrance will catch (sense)
the film which triggers the READY sign, turn on the drive motor, rollers and the drying blower.
Then it will automatically transport the films. When the film sensor catches the film, this will
trigger the developing and fixing replenishing pumps to run and they will stop when the
replenishing is completed. Once the film is completely inside the unit, the READY sign will flash
and the buzzer will go off 3 times. This means that the next film can be inserted. When the last
film is inserted and the READY sign starts to light, it will take 7 minutes for the drive motor and
drying blower to go into wait mode.

For Completing the Operation,


When completing the film processing, turn off the power, take out the insertion case. Turn off
the supply.

Please refer to page 14, Operation for more details.


10. Pengoperasian

11-1 Pemeriksaan Harian


Untuk menjaga peralatan dalam kondisi baik, pastikan untuk melakukan pemeliharaan dan
pembersihan secara berkala.
1) Pastikan untuk membersihkan dan memeriksa bagian penyisipan film.
a) Pastikan papan penyisipan bersih. Jika ada kotoran atau debu, pastikan untuk
membersihkan kotoran dengan menggunakan handuk atau kain basah.

Casing Penyisipan Film

Peringatan
Jikaada larutan developer yang terciprat ke dalam kotak penyisipan,maka gambar akan
terpengaruh.

B) Jika kotak penyisipan dibawa keluar, tangani dengan hati-hati karena casing dan
tutupnya akan terlepas dalam 2 bagian.
C) Setelah dibersihkan, pastikan kotak penyisipan terhubung dengan benar.
2) Memeriksa pembuangan limbah
Pastikan untuk memeriksa sisa pembuangan limbah

Peringatan
Jika Anda menggunakan tangki pembuangan ketika penuh dengan pembuangan limbah, limbah
akan bocor ke lantai.

3) Memeriksa isi ulang larutan


Pastikan untuk memeriksa jumlah isi ulang larutan
4) Memeriksa sambungan air
Pastikan untuk memeriksa tidak ada kebocoran air dari sambungan air.

11-2 Perawatan Bulanan:


Bersihkan setiap rak dan crossover.
Rendam setiap rak dan masing-masing crossover rak ke dalam air hangat dan gunakan handuk
bersih untuk menyeka kotoran (oksidasi, debu, komposisi kimia)

11-3 Mengganti Larutan Kimia


Kondisi operasi (menggunakan film pembersih dan penggunaan harian 3 film /
minggu):
Ubah larutan kimia setiap 2-3 bulan untuk menjaga kualitas kemampuan
pemrosesan. Perubahan larutan kimia secara berkala akan bervariasi
tergantung pada jumlah film yang diproses.
11-4 Pemeriksaan yang Direkomendasikan

Peralatan ini menggunakan bahan kimia dan air. Untuk menjaga kualitas satuan dalam kondisi baik, silakan
melakukan pemeriksaan keselamatan setiap tahun. Untuk rincian lebih lanjut tentang pemeriksaan
keselamatan, silakan hubungi agen penjualan / distributor lokal Anda.

Pembersihan Periode Metode Peringatan


Petunjuk Crossover 2~3 Rendam dalam air hangat dan bersihkan Keringkan lalu
Developer bulan tempatkan
Petunjuk Crossover Fixer

Rak Air Developer / Fixer 2~3 Rendam dalam air hangat dan bersihkan
bulan

Rak Developer 2~3 1) Bersihkan rol menggunakan larutan


bulan pembersih
2) Cuci menggunakan air
Rak Fixer 2~3 1) Bersihkan rol menggunakan larutan
bulan pembersih
2) Cuci menggunakan air
Rak Air 2~3 1) Rendam menggunakan pemutih selama 15
bulan menit
2) Gulung rol di dalam pemutih selama 2-3 menit
3) Cuci menggunakan air

Pemeriksaan dan Pembersihan


Pembersihan Periode Metode Peringatan
Tangki Developer 2~3 bulan Gunakan air hangat

Tangki Fixer 2~3 bulan Gunakan air hangat

Tangka Air 2~3 bulan Gunakan air hangat

Panduan Setiap hari Gunakan kain basah Pastikan plat sudah kering
Pemasangan plat

Film sensor Setiap Bulan Gunakan kain basah

Bagian eksterior Setiap hari Gunakan kain basah


lainnya

Peringatan
1) Saat membongkar rak, pastikan larutan kimia tidak bocor ke rak yang berdekatan.
2) Gunakan sikat untuk membersihkan kotoran yang mengkristal pada roda gigi.
3) Jangan gunakan sikat keras saat membersihkan rol. Ini akan merusak permukaan rol..

Nomor Bagian Nama Bagian Penggantian


117-3850002 Pegas roll Setelah 1.5th
11-5 List of Consumables
117-3730005 O ring 12 Setelah 2 th
Silakan ikuti suku cadang berikut 117-3730006 O ring 18 Setelah 2 th
(daftar dapat bervariasi tergantung
pada kondisi dan perawatan harian) 117-3710007 Isi Ulang Setelah 2 th
117-3270045 Roda gigi kanan A Setelah 2.5 th
117-3270047 Roda gigi kiri A Setelah 2.5 th
117-1190026 Kipas Setelah 2.5 th
11-6 Inspection and Check Sheet
No Description Content Check column

Is the developing solution Check if there is no crystallization of solution


1 clean? Check if the solution is clean
Is the solution crystallized?

Is the fixing solution clean? Check if there is no crystallization of solution


2 Is the solution cloudy? Check if the solution is cloudy

Is the water supply dirty? Check if there is no crystallization of water


3 Is the water clean? Check if the water does not contain any dirt

Check the DEV rack Is there any dirt on the DEV rack?
4
Are the gears and springs attached correctly?

Check the FIX rack Is there any dirt on the FIX rack?
5
Are the gears and springs attached correctly?

Check the Water rack Is there any dirt on the Water rack?
6
Are the gears and springs attached correctly?

Check the Dry rack Is there any dirt or crystallization of the solution?
7
Are the gears and springs attached correctly?

Check the cross over rack Check is there is any dirt or crystallization
8 (both)

Does the RRADY lamp Turn on the Power and check if the READY lamp will
9
turn on? turn on after temperature reaches the set temperature

Check the buzzer sound Check for the buzzer sound when the Power is turned
10
o and when the film is being processed.

Is the DEV temperature When the READY turns on, check the temperature of
11 set? DEV tank. The temperature should be 35 .

12 Use 1-2 films for testing Transportation test: check if the film is transported

13 Drying Test: check if the processed film is dried

14 Replenishing Test: check if the DEV / FIX tank is less

15 Check for any abnormal sound

16 Check for water supply Check if the water is running from the water tap

17 Check if the ground cable


is connected

18 Check if there is any water


leakage around the unit

Check the dark room Make sure there is not any unnecessary items inside
19 the dark room. Make sure there is no light leakage.
Make sure the dark room lamp is working.

All the above inspections and checking will help keep the unit working and for better results.
12. Damage and Inspection

In case of any trouble of operation, please conduct an inspection.


If the trouble is not fixed or do not understand the inspection, please contact your local agent / distributor.

Damage Inspection point Counter measure


Operation stopped Check the power cord properly Plug the power cord
(green)
Is the breaker working? Press the Reset button

Look for red lamp. If the lamp is lit, electricity is running.

The operation does not Check if the insertion case is inserted Push the insertion case tightly.
stop when READY

Check for any light leakage touching the film Make sure there is no light to the insertion.
sensor.

There is abnormal sound Check the DEV / FIX / DRY rack are set Set all the racks properly
properly.
The films are not Check if the cross over guides are set Set the cross over guide properly.
transported

Check if the unit inside the racks turns Clean the racks
smoothly by hand.
Check if there is any oxidation.

The films comes out wet Check if the amount of Fixing Use the proper amount
Solution is enough or too little. Check the amount in tank.

Check if the solution is dirty and old. Change with new solution.

Drying temperature is not set. Reset the safety thermo.

Image is light Check if the DEV and FIX solution is not Change the DEV solution
mixed.
Check the temperature of DEV solution Check the temperature of the Developer

Check if the amount of DEC solution Keep enough solution in Tank is enough.

Check if the DEV solution is not old. Change to new DEV Solution

Double image The film is exposed to the dark lamp Do not expose to the dark Change
Check if the dark room lamp filter is Correct. the light filter

Check if the insertion case is inserted Insert all the way to the end

Check if the insertion case lid is closed Close the lid completely

Irregular image Check if the level of the solution is correct Pour the correct amount

Is the level of solution even? Place the unit evenly to the ground.

Is the water running. Run the water during processing

Scratches on images Check for oxidation on insertion Clean off the oxidation
case and roller of each racks.

Are the films and cassettes used and handled Use them properly.
properly.

Image is too dark Check if the temperature of DEV is too high. Change to new solution
Was the solution mixed properly?

Films do not come out Check if the FIX solution is mixed properly. Change to new solution
smoothly Is the amount enough? Pour the correct amount

READY lamp does not turn The temperature of solution is not Rest the safety thermo

* flashing: temp solution is rising. Enough


* On: temp is set

18
13 Spesifiksi Dasar :
Transportasi Sistem transportasi rol diagonal

Kecepatan Processing 90 - 120 detik

Film Processing Lembaran Film 14" x 17" - 5" x 7"

Kapasitas Processing 100 films / jam

Solusi Processing Solusi bahan kimia regular (developer and fixer)

Kapasitas Tangki: DEV: 3.7L / FIX: 2.4L / WATER: 2.3L

Suhu Processing Developer: 35ᵒ


Pengering: 60ᵒ ± 5ᵒ

Kontrol Suhu Developer: kontrol oleh pemanas thermostat


Fixer kontrol oleh pemanas thermostat
Pengering kontrol oleh thermister

Penambahan Solusi Pengisian ulang film otomatis oleh sensor film

DEV Solusi Pencampuran Rotasi dengan mencampur piring

Persediaan Air Air mengalir

Mode Aman Perlindungan Pemanasan Tinggi , Keamanan thermo manual


(DEV/FIX/DRY) (DEV/FIX/DRY) Pemutus kabel

Mode Siap Daya menghemat, roller anti-degradasi, kontrol suhu, operasi


motor roller drive

Lingkungan Instalasi Suhu: 20C +/- 10C,Kelembaban: 55% +/- 25% Suhu
air: 5C lebih tinggi dari air keran

Daya 220/240V,4.5A ( or 110/120V ,9A )

Dimensi 565 x 722 x 464cm

Berat 27.5kg

Pilihan Stand
Pompa air
Katup Solenoida
Kit filter air

Spesifikasi dapat berubah tanpa pemberitahuan sebelumnya untuk kinerja yang lebih baik.

19
14.Bagian Pilihan

Berikut ini adalah rincian item opsional untuk ECOMAT9000

14-1 Stand
Bangun stand sendiri untuk digunakan untuk ECOMAT9000 untuk digunakan di ruangan gelap

14-2 Katup Air Solenoida


Solenoid Valve secara otomatis akan memasok air. Selama pemrosesan film, ini akan memasok
air dan ketika pemrosesan film selesai, pasokan air akan berhenti secara otomatis.
Instalasi akan membutuhkan penggantian pipa air dan ECOMAT9000.
Instruksi instalasi termasuk dalam unit katup solenoid. Silakan lihat instruksi dari katup solenoid.

14-3 Kit Penyaringan Air


Filter air dapat dipasang ketika kualitas air tidak baik. Disarankan untuk memasang filter air
bersama dengan katup solenoid. Anda perlu mengganti pipa air. Silakan lihat instruksi di dalam
unit filter air.

14-4 Unit Pompa Air


Jika tidak ada pasokan air, ini digunakan untuk memasok air. Tuangkan 20 liter air ke dalam
tangki dan gunakan air untuk mengisi air dari tangki air. Silakan ganti air tangki setelah 20-30
film olahan. Untuk pemasangan, silakan lihat instruksi di dalam unit pompa air.

15. Garansi dan Setelah Service

1. Setelah layanan
Jika ada masalah, silakan lihat halaman 22 Kerusakan dan Inspeksi. Hubungi agen / distributor
setempat Anda untuk instruksi lebih lanjut dan laporkan hal-hal berikut:
1) detail kerusakan (masalah)
2) nama model (ECOMAT 9000)
3) Tanggal pemasangan
4) Nomor seri peralatan
2. ECOMAT9000 diperiksa secara menyeluruh sebelum dikirim. Pastikan untuk menyimpan kartu
garansi dengan peralatan.
3. Setelah penggunaan normal peralatan, masa garansi 1 tahun setelah instalasi.
4. Garansi tidak berlaku jika peralatan tidak digunakan dengan benar.
5. Setiap penggantian atau biaya layanan setelah masa garansi akan dikenakan biaya. Silakan
hubungi agen / distributor lokal Anda untuk lebih jelasnya.
6. Pasokan suku cadang yang direkomendasikan akan dihentikan 7 tahun setelah pengumuman
penghentian. Namun, bahkan setelah 7 tahun pengumuman, masalah dapat diperbaiki. Sebaiknya
hubungi agen / distributor lokal Anda dalam hal demikian.

Agar unit tetap dalam kondisi baik, harap tinjau panduan ini untuk menjaga kinerja peralatan
dalam kondisi baik.

20