System Specifications

2-2-6-1 Physical Dimensions of Voluson® 730Pro/ProV


The physical dimensions of the Voluson® 730Pro/ProV unit are summarized in Table 2-5.
Table 2-6 lists the size of Voluson® 730Pro/ProV, with monitor and without on-board peripherals.

Operating Operating Heat Dissipation Storage Storage Humidity


Temperature Humidity Temperature

10 to 30oC 30 to 80% rH 3446 BTU pr hour -10 to 40 oC < 90% rH


(50 to 104oF) non-condensing (14 to 104oF) noncondensing

Power Consumption nominal 920 VA including all options.


Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories.
All mains outlets are co-switched by the unit’s mains switch via built-in isolation transformer.
Output voltage for: ST1 - ST5: 115V or 230V.

Electrical Specifications for Voluson® 730Pro/ProV


Voltage Tolerances Current Frequency
100 VAC ±10% 9,2 A 50, 60 Hz (±2%)
115 VAC ±10% 8.00 A 50, 60 Hz (±2%)
130 VAC ±10% 7.10 A 50, 60 Hz (±2%)
230 VAC ±10% 4.00 A 50, 60 Hz (±2%)
240 VAC ±10% 3,85 A 50, 60 Hz (±2%)

Voluson® 730Pro/ProV Power Requirements


System Specifications
2-2-6-1 Physical Dimensions of Voluson® 730Pro/ProV
The physical dimensions of the Voluson® 730Pro/ProV unit are summarized in Table 2-5.
Table 2-6 lists the size of Voluson® 730Pro/ProV, with monitor and without on-board peripherals.
System Specifications
2-2-6-1 Physical Dimensions of Voluson® 730Pro/ProV
The physical dimensions of the Voluson® 730Pro/ProV unit are summarized in Table 2-5.
Table 2-6 lists the size of Voluson® 730Pro/ProV, with monitor and without on-board peripherals.

Possible EMI sources should be identified before the unit is installed.

Electrical and electronic equipment may produce EMI unintentionally as the result of a defect.
These sources include:
• medical lasers
• scanners
• cauterizing guns
• computers
• monitors
• fans
• gel warmers
• microwave oven
• light dimmers
• portable phones
• broadcast stations and mobile broadcasting machines

EMI Prevention/Abatement

EMI Rule Details


Be aware of RF sources.
Keep the unit at least 5 meters (16.4 feet) away from other EMI sources. Special shielding may be
required to eliminate interference problems caused by high frequency, high powered radio or video
broadcast signals.
Ground the unit. Poor grounding is the most likely reason a unit will have noisy images.
Check grounding of the power cord and power outlet.
Replace and/or reassemble
all screws, RF gaskets,
covers and cores.
After you finish repairing or updating the system, replace all covers and tighten all screws.
Any cable with an external connection requires a magnet wrap at each end.
Install the shield over the front of card cage. Loose or missing covers or RF gaskets allow radio
frequencies to interfere with the ultrasound signals.
Replace broken RF gaskets. If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Do not turn ON the unit until any loose metallic part is removed and replaced, if required.
Do not place labels where
RF gaskets touch metal.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit RF
leakage. In case a label has been found in such a location, move the label to a different,
appropriate location.
Use GE- specified
harnesses and peripherals.
The interconnect cables are grounded and require ferrite beads and other shielding.
Cable length, material, and routing are all important; do not make any changes that do not meet
all specifications.
Take care with cellular
phones. Cellular phones may transmit a 5 V/m signal that causes image artifacts.
Properly dress peripheral
cables.
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays.
Loop the excess length for peripheral cables inside the peripheral bays.
Attach the monitor cables to the frame.
System Specifications
2-2-6-1 Physical Dimensions of Voluson® 730Pro/ProV
The physical dimensions of the Voluson® 730Pro/ProV unit are summarized in Table 2-5.
Table 2-6 lists the size of Voluson® 730Pro/ProV, with monitor and without on-board peripherals.

Physical Dimensions of Voluson® 730Pro/ProV


Height Width Depth
142 cm / 55.9 inches 68 cm / 26.8 inches 100 cm / 39.4 inches

Weight of Voluson® 730Pro/ProV with Monitor and without other Peripherals


Model Weight [kg] Weight [lbs.]
Voluson® 730Pro/ProV 136 300