Beruflich Dokumente
Kultur Dokumente
Water Technology
Table of Contents
Page
1. Monitor description ..................................................................................................................... 3
1.1. General ..............................................................................................................................................3
1.2. Scattered light measuring method .....................................................................................................4
1.3. Features.............................................................................................................................................4
1.4. Adjustment.........................................................................................................................................4
1.5. Displays and Alarms ..........................................................................................................................5
1.6. Design of the Monitor.........................................................................................................................5
2. Technical Specifications............................................................................................................. 6
3. Installation of the Monitor OMD 21 ............................................................................................ 7
3.1. With an RWO oily water separator of 0.1 or 0.25 m3/h capacity......................................................7
3.2. With an RWO oily water separator from 0.5 to 10.0 m3/h capacity...................................................7
4. Piping connections...................................................................................................................... 8
4.1. With an RWO oily water separator of 0.1 or 0.25 m3/h capacity.......................................................8
4.2. With an RWO oily water separator from 0.5 to 10.0 m3/h capacity...................................................9
5. Electrical connections............................................................................................................... 10
5.1. General wiring information ...............................................................................................................10
5.2. Connection diagram, RWO - OMD 21 .............................................................................................11
5.3. Wiring with an RWO oily water separator of 0.1 or 0.25 m3/h capacity...........................................12
5.4. Wiring with an RWO oily water separator from 0.5 to 10.0 m3/h capacity.......................................12
6. Operation .................................................................................................................................... 13
6.1. Important Notes ...............................................................................................................................13
6.2. Commissioning ................................................................................................................................13
6.3. Functional Tests ..............................................................................................................................14
6.4. Alarm ...............................................................................................................................................14
6.5. Programming Mode .........................................................................................................................15
6.6. Alarm Set Points and Time Delay ....................................................................................................15
6.7. Signal output....................................................................................................................................15
6.8. Zero adjustment...............................................................................................................................15
6.9. Default values ..................................................................................................................................16
6.10. Recorder ..........................................................................................................................................16
7. Calibration .................................................................................................................................. 16
7.1. Calibration check .............................................................................................................................17
7.2. Calibration........................................................................................................................................17
8. Monitor operation ...................................................................................................................... 18
8.1. with an RWO oily water separator of 0.1 or 0.25 m3/h capacity ......................................................18
8.2. with an RWO oily water separator from 0.5 to 10.0 m3/h capacity ..................................................18
8.3. Before commissioning .....................................................................................................................18
8.4. Before each start .............................................................................................................................18
8.5. Monitoring operation and alarm events during operation.................................................................18
9. Storage under frost conditions ................................................................................................ 19
10. Maintenance ............................................................................................................................... 19
11. Troubleshooting ........................................................................................................................ 20
11.1. Fault finding .....................................................................................................................................20
11.2. Sample Cell Tube Replacement ......................................................................................................21
11.3. PCB Replacement ...........................................................................................................................22
11.4. Display PCB Replacement...............................................................................................................22
11.5. Optic Block PCB Replacement ........................................................................................................23
12. Spare Parts ................................................................................................................................. 24
12.1. Drawing showing spare part positions .............................................................................................24
12.2. List of spare parts ............................................................................................................................24
12.3. List of RWO-recommended on board spares ..................................................................................24
13. Exchange Parts Service ............................................................................................................ 25
13.1. General ............................................................................................................................................25
13.2. Exchange Equipment Service..........................................................................................................25
13.3. Exchange of electronic components ................................................................................................25
1. Monitor description
1.1. General
The OMD 21 Bilge Alarm Unit has been designed specifically for use in conjunction with
15 ppm oil-water separator units and has a specification and performance which meets
the requirements of the IMO (International Maritime Organisation) specifications for Bilge
Alarm Units contained in Resolution MEPC. 60 (33).
The unit is supplied with 2 works-adjusted alarms at 15 ppm. Other set points (10 ppm or
5 ppm) are possible and can be adjusted on site at any time by using the buttons at the
front panel.
If an alarm set point is exceeded, the alarms are visible at the front panel and the appro-
priate relays are switched.
Furthermore a 0(4) - 20 mA (equal to 0 - 30 ppm) signal output is available for driving a re-
corder or external meter.
The water sample is taken from the flowing water upstream of an orifice.
The following National IMO-type test certificates in accordance with MEPC. 60(33) are
available, issued by the national authorities of the countries listed below
(Date: Aug. 2002).
The list is an interim one and necessarily incomplete, as further national certification has
been applied for.
0736
reference beam
light receiver
light emitter
IR LED
This measuring principle is the most frequently used method for the continuous meas-
urement of oil in water and has proved itself in practice.
1.3. Features
• Compact and robust construction
• Automatic voltage selection
• Low maintenance
• Small dimensions
• Low weight
• Easy installation
• Constant readiness
• Low spare part stock holding
• Watertight Housing
• Works adjustment
• Easy zero adjustment via push buttons
1.4. Adjustment
The unit is delivered with a works calibration according the IMO-requirements. The alarm
points are set to 15 ppm.
The "Zero" point is also works calibrated and can be re-adjusted on site by using the pro-
gramming mode and clean water. See Section 6.8 “Zero adjastment”.
Fig. 1
Power on ppm
Alarm 1
Alarm 2
Error
SetMode
OMD-21
Oil Monitoring Device
RWO
Water Technology
2. Technical Specifications
Range: 0 – 50 ppm
Accuracy Better than IMO requirements
Linearity Up to 30 ppm better than ± 2 %
Display 2 digit 7 segment LED's
Power Supply: 24 V – 240 V AC or DC
Automatic Voltage Selection
Consumption: < 6 VA
Alarm Points 1 + 2: Adjustable between 1 - 15 ppm
(Works adjustment 15 ppm)
Alarm 1 Operating Delay: Adjustable between 1 – 540 sec.
(for annunciation purpose) (Works adjustment 2 sec)
Alarm 2 Operating Delay: Adjustable between 1 – 20 sec.
(for control purposes) (Works adjustment 10 sec)
System Fault Alarm: LED
Alarm Contact Rating: Potential free 1 pole change over contacts, 3
A / 240 V
Alarm Indication: Red LED's
Output Signal: 0 – 20 mA or 4 – 20 mA, reversible
ext. Load < 150 Ω
Sample Water Pressure: 0,1 – 10 bar
Sample Flow: Approx. 0,1 - 4 l/min depend. to pressure
Ambient Temperature: + 1 to + 55° C
Sample Water Tempera- + 1 to + 65° C
ture:
Roll: Up to 22,5°
Size (over all): 152 mm W x 155 mm H x 91 mm D
Degree of Protection: IP 65
Weight: 2,6 kg
Pipe Connections: R 1/4" Female
IMPORTANT
The Monitor OMD 21 should be located close to the oily water separator in order
to minimise response delays. The distance between monitor and separator should
never be more than 8 metres; any greater distance would result in a response
time of over 20 seconds which is the max response time in accordance with IMO
regulations.
Fig. 2
overall dimensions
Fix the Monitor OMD 21 with four M5 bolts on a rigid vertical surface, if possible with the
display at eye level.
Connect the 3-way cock on the Monitor OMD 21 with the separator’s clean water outlet
pipe and the freshwater supply, using 8 mm copper piping and ERMETO connections.
The sampling pipe should be installed horizontal or in a continuous upward flow.
The outlet pipe of the Monitor OMD 21 has to be connected to the 3-way diverting valve.
In the alarm state, the effluent water is recirculated to bilge.
When assembling the housing and connecting the piping, it is important to avoid stressing
the housing so as to prevent possible damage.
3.2. With an RWO oily water separator from 0.5 to 10.0 m3/h capacity
The monitor is supplied ready-installed on the oily water separator with all necessary pip-
ing and wiring.
4. Piping connections
Installation of the Monitor OMD 21 and the 3-way recirculation valve including all accessories for
automatic switching to recirculation when the limit value (normally 15 ppm) in cleaned bilgewater
is exceeded.
4.1. With an RWO oily water separator of 0.1 or 0.25 m3/h capacity
The Monitor OMD 21 should be mounted on a vertical surface in the immediate neigh-
bourhood of the RWO oily separator and connected below as shown in the piping dia-
gram. The piping connections for the equipment are R 1/4" internal thread. We recom-
mend 8 mm copper piping x 1 mm wall thickness with clamp ring screw connectors for the
pipe connections.
The following connections should be made:
1. The sample water inlet (Pos 25, 3-way ball-valve) on the monitor should be connected
by pipes to
• the clean water outlet on the separator (Pos. 13 on the pump) and
• a line with freshwater or clear seawater for cleaning and zero-balancing.
2. The sample water outlet is at the upper right of the transmitter head, below the screw
cap. It shall be connected to the 3-wayrecirculation valve, Pos. 23.
IMPORTANT!
3
For separators with a capacity of 0.1 m / h, the pump is not installed on the
separator but is supplied as a separate item.
4.2. With an RWO oily water separator from 0.5 to 10.0 m3/h capacity
The inlet pipe is screwed into the base of the vessel and for oily water separators with a
3
capacity of 0.5 to 10 m / h is already fitted and piped up, including all accessories or-
dered. The outlet is top right at the transmitter head, just below the screw cap. The sam-
ple water is fed to a funnel. Samples can be taken here for test purposes.
5. Electrical connections
DANGER
At electrical voltages > 42 V
Direct contact with live parts may cause serious injury or death.
Work on live parts should only be carried out by qualified electricians.
M 16x1,5 M 16x1,5
Ground Connection
M6
Power on ppm
Alarm 1
Alarm 2
Error
SetMode
OMD-21
O il M onitoring Dev ice
RWO
Water Technology
X4 Terminal
Signal Output
0(4)-20 mA
Fuse F2
(T2A) F2: T2A A la rm1 Al arm 2
0(4)-20mA
Display -
PE
X5 X4 + X6
Connector 1,2: PO WER SUPPLY INPUT SIGNAL
4,5: PILOT VOLTAGE O UTPUT OU TPUT
24V/115V/240V AC /DC R i <=150 Ω
ALAR M
CON TACTS
max. 3A
X5 Terminal X6 Terminal
Power Supply Alarm Contacts
24-240 V Ölalarm Ölalarm E
oil alarm oil alarm e
A C / DC RL 1 RL 2
Ground
Ground Connection
Connection
1 2 3 4 5 6 7 8 E
X5 X4 X6
Signal Output
0(4)-20 mA
To Alarmsystem
Solenoid Valve
The relay contacts are shown in the alarm state (not energised).
5.3. Wiring with an RWO oily water separator of 0.1 or 0.25 m3/h capacity
The monitor, including all accessories ordered for the automatic switchover of recircula-
tion to bilge, will be supplied as individual items and not assembled.
As can be seen from the wiring diagram in Section 5.2, the alarm relay contacts are po-
tential-free and shown in the alarm state.
The monitor is supplied complete with a 9-core, 2 meter-long connecting cable.
This cable conveys the alarm signals to the terminal strip of the oily water separator control and
carries the power supply to the monitor.
IMPORTANT!
If the connecting cable is not long enough, be careful, if you install a longer cable.
There is a danger of damaging the sensitive electronic parts and invalidating the
guarantee.
You will find details of the connections in the wiring diagram accompanying of the installa-
tion instructions of the oily water separator.
The following electrical connections should be made:
• Power supply 24V AC via cores 1 and 2 of the RWO oily water separator control, terminals
13 and 14 to Monitor OMD 21.
• Oil - alarm output circuit 2 via cores 6 and 7 from monitor to RWO oily water separator,
terminals 13 and 15 and from there further to the 3-way-recirculation valve.
• Connection to an alarm indicator in the control room, if required.
We recommend you to connect the oil alarm output circuit 1 via the cable cores 3 and 4
from the monitor to the RWO separator, terminals 17 and 18, and from there further to the
control room.
• Connection of an external recorder etc., if required
For the routing of this signal you connect the recorder with the terminals
X4/1 and X4/2.
5.4. Wiring with an RWO oily water separator from 0.5 to 10.0 m3/h ca-
pacity
The monitor is supplied ready-to-use, including all accessories ordered, for the automatic
switching of recirculation to bilge and mounted on the RWO separator.
6. Operation
6.1. Important Notes
a) This equipment must be installed and operated in strict accordance with the instruc-
tions contained in this manual. Failure to do so will impair the protection provided.
b) Installation and servicing must be undertaken by a competent and suitably skilled
person.
c) The equipment must be earthed according to the relevant requirements.
d) The unit must be isolated from the electrical supply before any maintenance of the
equipment is attempted.
e) All national or local codes of practice or regulations must be observed and, where
applicable, are deemed to take precedence over any directive or information con-
tained in this manual.
f) To avoid wrong measurements, only quick separating detergents / cleaner should be
used.
g) In case of freezing conditions the measuring cell should be emptied completely.
6.2. Commissioning
Once the installation, wiring and piping connections have been completed, carry out the
following checks:
a) Switch the oily separator to „ON“. Check whether the Power LED lights up. Check the
power supply to the oily separator if necessary.
DANGER
At electrical voltages > 42 V
Direct contact with live parts can cause serious injury or death.
Only qualified electricians should carry out work on live parts.
b) Check all piping connections for leaks and rectify as appropriate.
c) Check the flow rate.
The equipment is set up in the factory for a flow rate 0.5 up to 4 l/min. Adjust the flow rate
through the unit by using the small screws in the cell cap (Fig. 1, Pos. 24). Turning
the screw anticlockwise will increase the flow rate.
Note: The flow rate should be checked on both the clean water supply and the separator
sample supply. If the clean water supply is obtained from a high pressure source,
the flow rate will be higher than from the sample point.
The flow rate does not influence the accuracy of the instrument. The adjustment is only
important for the time delay between the sample point and the monitor.
6.4. Alarm
The unit is delivered with a works adjustment according the IMO-requirements. The alarm
points are set to 15 ppm. Other alarm levels (for 5 or 10 ppm) can be set.
In RWO oily separators with Monitor OMD 21, the alarm 1 signal is taken to the ship’s
alarm system. The signal cable should be connected to terminals 17 and 18 of the sepa-
rator’s control system.
In RWO oily separators with Monitor OMD 21, the alarm 2 signal is used to control the
3-way recirculation valve. If the limiting value for residue oil is exceeded, it is switched
over from "overboard discharge“ to "recirculation to bilge or bilgewater tank“. The valve
will also be switched to this position if the bilge water oil separator is switched to the
„OFF“ position or the pneumatic pressure fails.
Please note for better understanding:
Alarm output of the Monitor OMD 21
• in alarm status the position of the alarm switch is open (potential-free, de-energised)
• if the power supply is interrupted or the system is not in service, the position of the
alarm switch is also in alarm position open, (potential-free, de-energised)
This fact sometimes causes trouble because the ship’s alarm is initiated when
the oily water separator is off.
Because of this fact, we recommend to connect the alarm 1 output to the control box of
the oily water separator. Here the alarm circuit will be closed as soon as the main switch
of the control box is in „off“ position i. e. no alarm signal will be transmitted as long as the
oily water separator is not in service.
NB: All changed values have to be confirmed by pressing the " " button. Otherwise
the existing values are valid.
6.10. Recorder
If a recorder is necessary for continuous measured value data acquisition, RWO can offer
its own single-channel recorder of type DEGU. The Monitor OMD 21 has an output signal
(either 0 - 20 mA or 4 - 20 mA) available for this purpose.
7. Calibration
(only necessary if calibration check fails)
DANGER
At electrical voltages > 42 V
Direct contact with live parts may cause serious injury or death.
Work on live parts should only be carried out by qualified electricians.
IMPORTANT!
Before opening the device, make sure that any water that might be lying on the
housing is wiped off. If you don’t take this precaution, water can run over the edge
into the device and damage the electronic parts.
The OMD-21 is works adjusted and a calibration on site is normally not necessary. Before
starting the calibration procedure carry out a calibration check by following the procedure
as described below.
For both actions a definite fluid for comparison needs to be available. With the OMD-21
Monitor a formazin in water solution with the concentration of 100 FTU (Formazin Turbid-
ity Units) has to be used. A suitable calibration set can be ordered specially under the Ar-
ticle No. 51108EM002.
7.2. Calibration
(only necessary if calibration check fails)
a) Switch off the power supply and stop the water inflow.
b) Open the instrument and set the jumper X5 on the backside of the display to its op-
posite position, so that the other 2 pins than before are linked.
c) Mount the front panel and switch on the power supply. The display will show "CA"
and the instrument is in alarm condition.
d) Clean the sample tube accurate by using a suitable bottle brush as described under
Section 10. Make sure, that the Zero adjustment is correct.
e) Empty the sample tube and fill it with a solution of 100 FTU as described with the
calibration set. It is necessary, to fill the sample tube twice with the formazin solution
to avoid any dilution with a rest of the oil free water from cleaning.
f) If it is sure, that the correct formazin solution is in the instrument, press the buttons "
" and " " simultaneously. The display will show 12 ± 1 and the instrument is cali-
brated.
g) Continue as described under Section 8.4. Do not change the position of the jumper
again!
h) If it is required to recall the works calibration, refer to section 6.9, “Default Values”
8. Monitor operation
8.1. with an RWO oily water separator of 0.1 or 0.25 m3/h capacity
IMPORTANT!
With these low capacity separators, the complete water flow of the separator is
led through the monitor. For this, the holes in the head screw of the cell are
opened by removing the flow adjustment screws in the factory.
8.2. with an RWO oily water separator from 0.5 to 10.0 m3/h capacity
IMPORTANT!
With these separators, only a part of the water flow through the separator is led
via a bypass through the monitor. For this, a back-pressure orifice plate (separa-
tor, Pos. 11) is built into the main flow between the pump and the 3-way recircula-
tion valve (separator Pos. 23). The amount of the flow through the monitor is, as a
rule, already set up in the factory.
The flow rate can be increased in special installation cases by the removal of an
adjusting screw, or even both if required, Pos. 24, in the cell head screw (see
Section 1.6, Fig. 1). The flow rate should be at least 0.5 litre/min and for these
separator duties, usually not greater than 2 litre/min.
IMPORTANT!
During standstill when there is a danger of freezing, the measurement cell must be com-
pletely emptied in order to prevent the glass tube breaking. To do this, loosen and then
remove both water supply pipes at 3-way ball valve (separator, Pos 25),. switch the 3-
way ball valve to the flushing water supply position. The water can then run out of the
measuring cell.. Remove the measuring cell head screw and carefully blow out any water
remaining.
10. Maintenance
The maintenance effort needed to keep the device functioning is relatively low.
The following maintenance work should be carried out each week:
a) Switch the separator off.
b) Flush out the measuring cell with oil-free water.
c) Close the sample water and clean water supplies.
d) Remove the measuring cell head screw (Fig. 1, Pos. 13).
e) Use a brush in the measuring cell to clean the glass tube by using an up and down motion
over the whole of its length.
f) Remove the brush and replace the measuring cell head screw.
g) Open the clean water inlet and flush out the measuring cell for several minutes.
h) Switch the separator on.
i) The Power LED is illuminated and the display is changing between "dh" or “- -“ and num-
bers, decreasing from "20". During the start also the “Set Mode” LED is flashing to indicate
the initiation. After a short time the monitor will display "0" with clean water.
j) If "0“ does not appear, carry out a zero balance as in Section 6.8.
k) Open the process water inlet.
l) Check whether the desiccator cartridge (Section 1.6, Fig. 1, Pos. 15) is blue in colour. If it
is light blue or white, the cartridge is exhausted and has to be replaced.
IMPORTANT!
The desiccator cartridge ensures a low humidity inside the measuring cell to prevent erro-
neous readings which could arise from condensation on the glass tube. Changing the
cartridge is done by unscrewing the unit from the front plate and screwing in again. This
can be done without opening up the equipment. The protective shell of the new cartridge
serves also as a tool.
11. Troubleshooting
11.1. Fault finding
See Section 13 for important notes.
A fault is indicated, when the red "Error"-LED is illuminated. Additional following messa-
ges maybe shown for at least 15 sec at the display for some special functions of the unit:
Display Reason Servicing
Meter out of range, oil content too Clean sample tube, wait until oil
high, dirty sample tube content is within the range
Different LED’s on the PCB’s indicate also the status of the unit if it is powered on.
They are visible as shown in Fig. 6 after removing the front cover and the desiccator.
Optic Block Optic Block PCB defective Replace Optic Block PCB
PCB LED’s
not on Connection interrupted Check Wiring
IMPORTANT!
Before opening the device, make sure that any water that might be lying on the
housing is wiped off. If you don’t take this precaution, water can run over the edge
into the device and damage the electronic parts.
a) Switch off the Power supply and disconnect all wiring from the terminals.
b) Remove all piping from the instrument.
c) Dismount the unit and take it to a clean working place.
d) Open the instrument from the back and
remove the upper cover. Remove
carefully also the lower cover with the
power supply PCB and disconnect the
plug connection from the optic block
PCB.
e) Loose the 2 screws (5) from the outlet
block (2) and remove this carefully.
Loose the 1 (6) screw from the inlet
block and remove this together with the
sample tube (4).
f) Clean the inner part of the instrument
and make sure, that everything is dry.
Check also the optic block PCB visual
g) Replace the O-Rings, the PTFE washer
and the sample tube in the inlet block
(3). Replace the O-Rings and the PTFE
washer in the outlet block (2). Check
and replace the seals of the housing as
necessary.
h) Mount the outlet block (2) and secure it
with the 2 screws (5) when it is in the
correct position. Make sure, that the
optic block PCB is in the correct position
Fig. 7
and mount the inlet block (3) with the
sample cell tube (4). Make sure, that the
sample tube is absolutely clean from outside. Secure it with the screw (6) when it is
in the correct position.
i) Connect the plug connection from the optic block PCB to the power supply PCB and
make sure that the cable is sealed. Mount the lower cover with the power supply
PCB.
j) Close the upper cover.
k) Do a pressure test of appr. 8 bar with oil free water.
l) Examine the colour of the desiccator (Fig. 1, Pos. 15). If the colour is light blue or
white, the desiccator need to be replaced.
m) Mount the unit at its place and connect the piping and wiring.
n) Follow Section 8.4 and carry out a calibration check according Section
IMPORTANT!
The desiccator cartridge ensures a low humidity inside the measuring cell to pre-
vent erroneous readings which could arise from condensation on the glass tube.
Changing the cartridge is done by unscrewing the unit from the front plate and
screwing in again. This can be done without opening up the equipment. The pro-
tective shell of the new cartridge serves also as a tool.
The monitor is now ready for use.
IMPORTANT!
Before opening the device, make sure that any water that might be lying on the
housing is wiped off. If you don’t take this precaution, water can run over the edge
into the device and damage the electronic parts.
a) Switch off the Power supply and disconnect all wiring from the terminals.
b) Remove all piping from the instrument.
c) Dismount the unit and take it to a clean working place.
d) Open the instrument from the back and remove carefully the lower cover with the
power supply PCB. Disconnect the plug connection from the optic block PCB.
e) Check the seals and mount the new lower cover with the replacement power supply
PCB.
f) Examine the colour of the desiccator (Fig. 1, Pos. 15). If the colour is light blue or
white, the desiccator need to be replaced.
g) Mount the unit at its place and connect the piping and wiring.
h) Follow Section 8.4 and carry out a calibration check according Section 7.1.
IMPORTANT!
Before opening the device, make sure that any water that might be lying on the
housing is wiped off. If you don’t take this precaution, water can run over the edge
into the device and damage the electronic parts.
a) Switch off the Power supply and open the front cover.
b) Disconnect the plug connection from the power supply PCB.
c) Make sure, that the jumpers of the new PCB are in the same position than at the old
one and replace the display PCB.
d) Connect the plug connection to the power supply PCB.
e) Close the front cover and continue as described under Section 8.4.
IMPORTANT!
Before opening the device, make sure that any water that might be lying on the
housing is wiped off. If you don’t take this precaution, water can run over the edge
into the device and damage the electronic parts.
It is strongly recommended to send the unit for repair to the manufacturer. Chang-
ing of the Optic Block PCB may influence the accuracy of the instrument!
a) Switch off the Power supply and disconnect all wiring from the terminals.
b) Remove all piping from the instrument.
c) Dismount the unit and take it to a clean working place.
d) Open the instrument from the back and remove the upper cover. Remove carefully
also the lower cover with the power supply PCB and disconnect the plug connection
from the optic block PCB.
e) Loose the 2 screws from the outlet block and remove this carefully. Loose the 1
screw from the inlet block and remove this together with the sample tube.
f) Clean the inner part of the instrument and make sure, that everything is dry. Replace
the optic block PCB.
g) Replace the O-Rings, the PTFE washer and the sample tube in the inlet block. Re-
place the O-Rings and the PTFE washer in the outlet block.
h) Mount the outlet block and secure it with the 2 screws when it is in the correct posi-
tion. Make sure, that the optic block PCB is in the correct position and mount the inlet
block with the sample cell tube. Make sure, that the sample tube is absolutely clean
from outside. Secure it with the screw when it is in the correct position.
i) Connect the plug connection from the optic block PCB to the power supply PCB and
make sure that the cable is sealed. Mount the lower cover with the power supply
PCB.
j) Close the upper cover.
k) Do a pressure test of app. 8 bar with oil free water.
l) Examine the colour of the desiccator (Fig. 1, Pos. 15). If the colour is light blue or
white, the desiccator need to be replaced.
m) Mount the unit at its place and connect the piping and wiring.
n) Follow Section 8.4 and carry out a calibration check according Section 7.1.
Power on ppm
Alarm 1
Alarm 2
Error
SetMode
OMD-21
Oil M onitoring Device
RWO
Water Technology
Optional item
1 off Manual Cell Clean Unit 51108EM026
IMPORTANT!
When you are ordering spare parts or exchange units, please give us the serial no.
or construction no. of the oily water separator number! (see cover sheet of the op-
erating instructions or the name plate on the bilge water separator).