Sie sind auf Seite 1von 51

Ultrasound Diagnostic Scanner

Specifications

Step5.5 (Jun, 2017 Revised)


Rev5.5

Hitachi, Ltd.

CONFIDENTIALITY: This material contains proprietary information of Hitachi and it


shall not be disclosed to the third party or reproduced without the prior written
permission of Hitachi.
Copyright ©Hitachi, Ltd. 2017. All rights reserved.

SN-U391E
Trademarks or the registered trademarks

The following is the trademarks or the registered trademarks in this manual or in this equipment.
 HI VISION Avius, Avius, Real-time Tissue Elastography, HdTHI, HI REZ, Ultra BE/Ultrasound
Broadband Engine, HINT-NAVI, Tool Bar, Fine Flow are registered trademarks of Hitachi in Japan.
 HI VISION is registered trademarks of Hitachi in the U.S.A. and Europe.
 Windows, Windows XP, Windows Embedded Standard 2009, Windows Vista, Windows 7, Windows 8,
Windows 8.1, Windows NT, Windows 2000, Windows 98, Microsoft Excel, Office, PowerPoint, Windows
Media, DirectX are trademarks or the registered trademarks of Microsoft Corporation in the U.S.A. and
other countries.
 Linux® is a trademark registered to Linus Torvalds.
 CinePlayer Decoder Pack is used in Digital Video Archiving Software (Option). CinePlayer Decoder Pack
is trademark of Sonic Solutions.
 In this manual or in this equipment, other brand names and product names are trademarks or registered
trademarks of their respective companies.

Abbreviations

 In this document, Microsoft Windows Operating System is called Windows.


 In this document, the Windows® XP operating system is called Windows XP.
 In this document, Microsoft® Windows® Vista Operating System is called Windows Vista.
 In this document, Microsoft® Windows® 7 Operating System is called Windows 7.
 In this document, Microsoft® Windows® 8 Operating System is called Windows 8.
 In this document, Microsoft® Windows® 8.1 Operating System is called Windows 8.1.
 In this document, the Microsoft(R) Windows(R) NT Workstation Operating System
or Microsoft(R) Windows(R) NT Sever Network is called Windows NT.
 In this document, Microsoft(R) Windows(R)2000 Professional Operating System or Microsoft(R)
Windows(R)2000 Server Operating System or Microsoft(R) Windows(R)2000 Advanced Server
Operating is called Windows NT.
 In this document, Microsoft® Windows® 98 Operating System is called Windows 98.

 No part of this publication may be transmitted into without the prior written
permission of HITACHI.
 This publication is subject to change without notice.
 Although HITACHI has attempted to ensure accuracy throughout this publication, HITACHI
welcomes customer input on corrections and suggestions for improvements to this publication.

SN-U391E
Contents
1. General Description ...................................................................................................... 1
1.1. General ........................................................................................................................................ 1
1.2. Features ...................................................................................................................................... 1
2. System Composition ..................................................................................................... 6
2.1. Main System Composition ........................................................................................................ 6
2.1.1. Ultrasound Scanner Main Unit .......................................................................................... 6
2.1.2. Accessories ........................................................................................................................... 6
2.2. Probes ......................................................................................................................................... 6
2.2.1. Convex array probes ............................................................................................................ 6
2.2.2. Linear array probes ............................................................................................................. 7
2.2.3. Phased array probes ............................................................................................................ 7
2.2.4. Other probes ........................................................................................................................ 8
2.3. Options ....................................................................................................................................... 8
2.4. Peripheral Devices ................................................................................................................... 10
3. Specifications .............................................................................................................. 11
3.1. General ...................................................................................................................................... 11
3.1.1. System ................................................................................................................................ 11
3.1.2. Basic Function ................................................................................................................... 11
3.2. Options ..................................................................................................................................... 22
3.2.1. 17-inch Digital LCD Monitor (EZU-MT29-S1) ................................................................. 22
3.2.2. 17-inch Digital LCD Monitor (Flexible arm) (EZU-MT30-S1) ........................................ 22
3.2.3. Biomedical Waveform Display Unit (EZU-EK28) ............................................................ 22
3.2.4. CW Doppler Unit (EZU-ST8) ............................................................................................ 22
3.2.5. Real-time Tissue Elastography Unit (EZU-TE6) ............................................................. 23
3.2.6. Real-time Tissue Elastography software (Strain Histogram) (EZU-TESH1) ................. 24
3.2.7. Digital Video Archiving Software (EZU-DV4) .................................................................. 25
3.2.8. Logon Management Software (EZU-LG2) ........................................................................ 26
3.2.9. DICOM Software(Transfer and Media Storage) (EZU-FC11) ......................................... 27
3.2.10. DICOM Software (WorkList) (EZU-FC11W) / DICOM Software (Printer) (EZU-FC11P)31
3.2.11. DICOM Software (Query / Retrieve) (EZU-FC11Q) ......................................................... 31
3.2.12. Omni-directional M mode display function (EZU-OD3) .................................................. 32
3.2.13. Advanced Measurement Software (EZU-AM7) ................................................................ 32
3.2.14. IMT Measurement Software (EZU-IM2) .......................................................................... 33
3.2.15. Software Specified for Contrast Agents (EZU-CH8) ........................................................ 33
3.2.16. WideView Software (EZU-WS8) ........................................................................................ 35
3.2.17. Stress Echo Software (EZU-SE6)...................................................................................... 36
3.2.18. Three-dimensional display unit (Real-Time) (EZU-4D7)................................................. 36
3.2.19. Three-dimensional Display Software (STIC) (EZU-4DST2) ............................................ 39

<i> SN-U391E
3.2.20. Three-dimensional Display Software (Freehand) (EZU-3D8) ......................................... 40
3.2.21. Remote Control Unit (EZU-RH4) ..................................................................................... 42
3.2.22. Remote Control Interface Unit (EZU-RiF5) ..................................................................... 42
3.2.23. Picture in Picture Software (EZU-PP5) ............................................................................ 42
3.2.24. Network Interface Unit (EZU-NW1) ................................................................................ 43
3.2.25. Analog Capture Card Unit (EZU-ACP2) .......................................................................... 43
3.2.26. Digital Capture Card Unit (EZU-DCP3) .......................................................................... 43
3.2.27. Video Interface Unit (EZU-ViF2) ...................................................................................... 43
3.2.28. Expanded USB Unit (EZU-UP2) ...................................................................................... 43
3.2.29. External LCD Monitor Unit (EZU-EL2)........................................................................... 44
3.2.30. Jelly Warmer Unit (EZU-JW1) ......................................................................................... 44
3.2.31. Alpha numeric keyboard Unit (EZU-AK1E) .................................................................... 44
3.2.32. High Neck Unit (EZU-HN1) ............................................................................................. 44
3.2.33. Probe Hanger (EZU-TH7) ................................................................................................. 44
3.2.34. Processor Tray (EZU-TY3) ................................................................................................ 44
3.2.35. Junction Box (PUN-JBP1) ................................................................................................ 44
3.3. Power requirements ................................................................................................................. 44
3.4. Environment conditions .......................................................................................................... 45
3.5. Exterior view ............................................................................................................................. 45

<ii> SN-U391E
1. General Description
1.1. General
Model HI VISION Avius is a multi-functional ultrasound diagnostic scanner in which Doppler, ColorFlow
Mapping, etc. are provided and all circuits related to image quality are fully digitalized.
As not only linear and convex scan type probes, but also phased array scan type probe can be operated with
HI VISION Avius, it is possible to use it for any field of application.
Furthermore, High-definition dTHI (HdTHI) mode (Hitachi unique technology) are provided in standard
function.
This mode provide “Clear image” to reduce artifact such as vibration echo, and to improve spatial resolution
and contrast resolution.
HI Com function, HI REZ+ function, Trapezoidal display function, Oblique B-mode image function,
real-time Doppler trace, cine memory, patient data management function, and Real-time Archiving function
can be used on the standard model.
Wide view software, Stress echo software, Software specified Contrast Agents, cardiac measurement
software, Real-time Tissue Elastography, omni-directional M (ODM) mode, Remote control unit, Picture in
Picture software can be used optional function.
It is convenience for user to adopt Ergonomic design and 17-inch digital LCD Monitor and more improved
High-definition screen-displayed format.

1.2. Features
 True digital architecture
Due to beamformer which is decisively concerned to the image quality and all image processing are
digitalized, high spatial resolution, wide dynamic range and high contrast resolution are realized.
Ultrasound transmit circuit is improved by high performance transmit amplifier, and ultrasound receiving
circuit is improved by high performance pre-amplifier.
Frame rate is improved much more faster comparing with conventional system.
High quality image is digital image processing application.
This architecture that is achieved high quality image is good for new generation.

 Quick response
Mode changing, image condition settings, probe changing and application selection can be performed
quickly so that operator can pay attention to the patients.

 Usable for wide field of application


Linear, convex and phased array scan type probe are usable. It is possible to cover from abdominal and
obstetric to cardiovascular fields.

-1- SN-U391E
 17-inch digital LCD Monitor
It is adopted 17-inch digital LCD Monitor. LCD Monitor provides high quality image.
The LCD simple arm that optimizes the position of the LCD monitor and the operational panel provides ease
of use. The LCD flexible arm that can adjust height and four-way position provides freely position of
monitor.

 High-definition screen-displayed format


The resolution of the monitor is increased and the layout of the screen is displayed to the high-definition. The
inspection efficiency is improved by additional information and the operation assistance on the screen.
- Design
The screen design which looks big ultrasound image.
- Tab menus
It is easy to use the tab menus to switch applications, perform measurements, and activate annotations.
Because it is positioned out of diagnosis area.
- Brightness Setting
It is adjustable character and background brightness level on the screen depends on setting condition.
- Image size change function
The image size can be changed under certain conditions.

 Ergonomic design
- Design
Design is adopted to focus on user’s “easiness to use”, like front layout of probe connector, based on
ergonomics.
- Easily operable keyboard
As key arrangement is taken in consideration of frequency of use and operational has improved. An
optional alphanumeric keyboard that is adopted backlight keyboard improves key response and
visibility.
- Excellent mobility with four swivel casters
As all the four casters are of swivel type, it is easy to move in narrow space.
Due to reduced friction force to floor it is easy to move from ward to ward.
- Probe rest
It connects 3 probes simultaneously and has 1 dummy connector.

 Use interface
Workflow is improved because of new GUI.
It is a few procedures for many functions to use screen buttons.

 Online Manual
It is applicable to browse instruction manual on your HI VISION Avius.

-2- SN-U391E
 Various B-mode scan
- HI Com function
Compound function provides high spatial resolution, wide dynamic range and high contrast resolution.
- Trapezoidal display function
Trapezoidal display is applicable for linear probe.
It is improved diagnosis area for linear probe.
- Oblique B-mode image function
This function enables you to use oblique function not only CFI/CFA mode but B mode.

 High quality image technique.


- High-definition dynamic tissue harmonic imaging (HdTHI) Mode
High-definition dynamic tissue harmonic imaging (Hitachi unique technology) provide “Clear image” to
reduce artifact such as vibration echo, and to improve spatial resolution and contrast resolution.
- Adaptive Imaging (HI REZ+)
Adaptive imaging is the technique to generate optimum image with changing filter characteristic
depending on input signal.
This technique which installed this ultrasound diagnostic scanner reduces speckle noise on B-mode
images and provides uniformly and high contrast images.
- Fine Flow function
It is the function to display high frame rate and fine high-definition color Doppler image.
It is possible to display small vessel less blooming than original system.
- Automatic image optimization (HI Support)
It is the function to adjusts B mode parameters (Gain, TGC and Depth-specific PDC) and PW mode
parameters (Gain, PRF and Baseline) so that it may become the optimal image automatically, and leads
to shortening of inspection time.

 Real time Doppler measurement function


Automatic tracing function and measuring function in Doppler-mode have been allowed to apply only to
frozen images, but with this function it is possible to perform it to real time images which contributes to
shorten the time for examination.
It is displayed position of Doppler waveform and Vp, Vd values of Doppler waveform in real time.
After Freeze, you can select the trace and measuring results of any past waveform by playing cine
memory images frame-by-frame and turning the PW Gain encoder.

-3- SN-U391E
 Filing function.
- Examination data management
Examination data (images measurement results and other data) can manage and operate together, so it is
easy to refer data before an examination and after an examination.
- Multiple functions to save data
No image degradation because of digital image recording.
This filing function can record still images and multi images not only HDD include system) but also
DVD-RAM, DVD-R, USB memory and USB HDD.
- Easy-to-use filing function
In the Filing window, you can view examination data.
It is consideration for easy-to-use in the routine examination by filtering Patient ID or Patient name or
keyword search.
Thumbnail view provides easy access to destination image.
Review images can display 1/4/9/16 screen depend on choosing images.
- Patient data management
You can select the patient in the list view who had examination once.

 Omni-Directional M mode (ODM) function


Usually the normal M-mode image is controlled by the beam direction.
With this function, it is possible to display the M-mode image on an optional line on the B-mode image, so it
enables to perform more accurate left ventricle volume measurement.

 Real time Tissue Elastography


Real time Tissue Elastography imaging is a qualitative imaging technique that displays the relative tissue
stiffness by detecting strain.

 Digital video archiving function


It is the function to record a B-mode and a color mode image as a digital movie file on real time. The
operational of the monitoring of contrast agents and recording of digital motion picture is improved by this
function.
It is possible to use Clip capture mode that is suitable for clip analyses and Long time recording mode.

 Contrast harmonic imaging function


With this function enabling to produce images further effectively from signal coming out of the contrast
agent it is possible to improve to diagnose the existence of tumor with ease. Two or more ROIs can be set up
to the video recorded by contrast echo inspection, the Time-Average (Time-Intensity) value in ROI can be
indicated by graph, and quantitative analysis is possible.

-4- SN-U391E
 DICOM function
This software can transfer ultrasound images and measurement reports to a storage server.
It has a Storage Commitment function for ultrasound images.
This software can output and read ultrasound images in DICOM file format.
This software has a function to obtain WorkLists from the MWM server.
It has a function to send examination start and completion notifications for ultrasound images to the MPPS
server.
And it can be sent Ultrasound images to grayscale or color DICOM printers.

 4D function
3D ultrasound images can be viewed in almost real-time. As the scanning part of the probe swings
automatically in a fan-like fashion, it sends signals to the three-dimensional display unit, which rapidly
builds a 3D image based on the probe’s signal.

 STIC function
Multiple 3D images representing a single heartbeat can be displayed by automatically swinging the
scanning part of the probe in a fan-like fashion and estimating the cardiac cycle from the obtained images.

 3D function
The software is able to collect sequential 2D image data by using a probe to scan at a steady rate and then
display the collected data as 3D images.

 Advanced Measurement function


- %WT measurement
Measurement useful to heartbeat diagnosis of thickness change of heart muscle etc. is enabled by tracing
automatically the movement toward a heart tissue in arbitrary positions.
- Color tissue tracking measurement
The changes of myocardium thickness are automatically tracked and their strain amount is displayed as
colored movies on the screen. The graph of the changes of myocardium, caliper on the graphs, movie
outputs, tracking data outputs can be possible. Bull’s eye can be created by loading the tracking data which
three cross-sectional surface of APICAL, MID and BASAL.
- LA tracking measurement
The motion of left atrium wall is automatically tracked so that the volume of left atrium and other
parameters against time are measured.
- IMT measurement
IMT (Intima-Media Thickness) measurement is enabled by detecting the position of intima and adventitia
of carotid artery.

 Remote control function


Freeze, REF, mode and etc. keys of the operation panel can be switched by remote control.

-5- SN-U391E
2. System Composition
2.1. Main System Composition
2.1.1. Ultrasound Scanner Main Unit
 Ultrasound transmitting / receiving unit
 Digital scan converter unit
 Pulsed Doppler unit
 Color flow mapping unit
 HI Com function
 Adaptive Imaging function (HI REZ+)
 TDI (Tissue Doppler imaging) function
 Filing/Patient data management function
 Tissue harmonic imaging function
 Cine-memory unit
 Automatic image optimization (HI Support)
 High density zoom function (HI Zoom)
 Power supply unit
 DVD (DVD-RAM, DVD-R) drive

2.1.2. Accessories
 System CD 2
 Hexagonal wrench bar 1 set
 Spare fuses 1 set
 Cords 1 set
 Instruction manual (Include CD) 1 set
 DVD-RAM (4.7GB) 1

2.2. Probes
2.2.1. Convex array probes
 4-2MHz 10R EUP-C511
 5-2 MHz 40R EUP-C514
 5-1 MHz 50R EUP-C715
 5-2 MHz 60R EUP-C516
 6-3 MHz 40R EUP-C524
 5-2 MHz 40R (Real-time 3D) EUP-CV714
 7-2MHz 40R (Real-time 3D) EUP-CV724
 6-3 MHz 40R (Real-time 3D) EUP-CV524
 8-4 MHz 20R EUP-C532

-6- SN-U391E
 5-2 MHz 20R (Biopsy) EUP-B512
 6-1 MHz 20R (Biopsy) EUP-B712
 5-2 MHz 40R (Biopsy) EUP-B514
 5-1MHz 50R(Biopsy) EUP-B715
 8-4 MHz 10R (Transvaginal wide view) EUP-V53W
 10-2 MHz 10R (Transvaginal wide view) EUP-V73W
 8-4MHz 10R (Transvaginal, Real-time 3D) EUP-VV531
 8-4 MHz 10R (Transvaginal, Real-time 3D) EUP-VV731
 8-4 MHz 10R Convex / 8-4 MHz 10R Convex (Intracavital) EUP-CC531
 8-4 MHz 10R Convex / 8-4 MHz 10R Convex (Intracavital) EUP-CC531S
 8-4 MHz 10R Convex / 10-5 MHz Linear (Intracavital) EUP-U533
 8-4 MHz 10R (Transrectal) EUP-U531
 8-4 MHz 10R (Finger top) EUP-F531
 7.5 MHz 40R (Finger top) EUP-F334
 10-5 MHz 10R (Laparoscopic) EUP-OL531
 7.5 MHz 40R (Laparoscopic) EUP-OL334
 10-4MHz 20R (Intraoperative) EUP-O732T
 10-5 MHz 6R (Transrectal, 19cm, 360°) EUP-R54AW-19
 10-5 MHz 6R (Transrectal, 33cm, 360°) EUP-R54AW-33

2.2.2. Linear array probes


 7-3 MHz EUP-L52
 9-4 MHz EUP-L73S
 10-5 MHz EUP-L53
 10-5 MHz (Wide view) EUP-L53L
 13-6 MHz EUP-L54MA
 13-5MHz EUP-L74M
 18-5MHz EUP-L75
 14-6 MHz EUP-L65
 13-5MHz (Real-time 3D) EUP-LV74
 10-5 MHz (Intraoperative) EUP-O53T
 13-7MHz (Intraoperative) EUP-O54J

2.2.3. Phased array probes


 4-2 MHz (Steerable CW) EUP-S50A
 5-1 MHz (Steerable CW) EUP-S70
 7-3 MHz (Steerable CW) EUP-S52
 9-2MHz (Steerable CW) EUP-S72
 8-3 MHz (Steerable CW, Electric multiplane transesophageal) EUP-ES52E
 6-3.5 MHz (Steerable CW, Multiplane transesophageal) EUP-ES52M

-7- SN-U391E
2.2.4. Other probes
 7.5 MHz 10R (Upper G.I. Fiberscope, made by HOYA) FG-34UX
 7.5 MHz 10R (Upper G.I. Fiberscope, made by HOYA) FG-36UX
 7.5 MHz 10R (Upper G.I. Fiberscope, made by HOYA) FG-38UX
 7.5 MHz 10R (Upper G.I. Videoscope, made by HOYA) EG-3630U
 10-5MHz 6R (Upper G.I. Videoscope, made by HOYA) EG-3270UK
 10-5 MHz 6R (Upper G.I. Videoscope,made by HOYA) EG-3630UR
 10-5 MHz 10R (Upper G.I. Videoscope, made by HOYA) EG-3830UT
 10-5 MHz 6R (Upper G.I. Videoscope, made by HOYA) EG-3670URK
 10-5 MHz 10R (Upper G.I. Videoscope, made by HOYA) EG-3870UTK
 10-5 MHz 6R (Videobronchoscope, made by HOYA) EB-1970UK
 10-5 MHz 6R (Videobronchoscope, made by HOYA) EB19-J10U

2.3. Options
 17-inch Digital LCD Monitor Unit EZU-MT29-S1
 17-inch Digital LCD Monitor Unit (Flexible arm) EZU-MT30-S1
 Biomedical Waveform Display Unit EZU-EK28
 CW Doppler Unit EZU-ST8
 Software Specified for Contrast Agents EZU-CH8
 DICOM Software (Transfer and Media Storage) EZU-FC11
 DICOM Software (Printer) EZU-FC11P
 DICOM Software (Worklist) EZU-FC11W
 DICOM Query/Retrieve Software EZU-FC11Q
 Three-dimensional display unit (Real-Time) EZU-4D7
 Three-dimensional Display Software (STIC) EZU-4DST2
 Three-dimensional Display Software (Freehand) EZU-3D8
 Real-time Tissue Elastography Software EZU-TE6
 Real Time Tissue Elastography Software (Strain Histogram) EZU-TESH1
 ODM Software EZU-OD3
 Advanced Measurement Software EZU-AM7
 IMT Measurement Software EZU-IM2
 WideView Software EZU-WS8
 Stress Echo Software EZU-SE6
 Digital Video Archiving Software EZU-DV4
 Remote Control Unit EZU-RH4
 Remote Control I/F Unit EZU-RiF5
 Picture in Picture Software EZU-PP5

-8- SN-U391E
 Logon Management Software EZU-LG2
 Alpha numeric keyboard Unit EZU-AK1E/F
 Network Interface Unit EZU-NW1
 Digital Capture Card Unit EZU-DCP3
 Analog Capture Card Unit EZU-ACP2
 Video Interface Unit EZU-ViF2
 Foot Switch (for Freeze) EZU-FS1
 Foot Switch (for Freeze / User Define) EZU-FS2
 Expanded USB Unit EZU-UP2
 External LCD Monitor Unit EZU-EL2
 Jelly Warmer Unit EZU-JW1
 High Neck Unit EZU-HN1
 Probe Hanger EZU-TH7
 Processor Tray EZU-TY3
 5MHz Pencil CW Probe Adapter EZU-AC1
- 5MHz Pencil CW Probe
Sonde Crayon 5MHz P050CN20W00+AD (with BNC cabling)
(by AXON’CABLE S.A.S)
 10MHz Pencil CW Probe Adapter EZU-AC2
- 10MHz Pencil CW Probe
Sonde Crayon 10MHz 10D4 (with BNC cabling)
(by AXON’CABLE S.A.S)
 5MHz/10MHz pencil CW probe Interface EZU-AC3
 Junction Box PUN-JBP1

-9- SN-U391E
2.4. Peripheral Devices
Peripheral device Manufacturer Manufacturer’s model, type
Black & white printer SONY UP-D897
UP-897MD (Notice II)
UP-X898MD
MITSUBISHI P93DW
P93E/W (Notice II)
P95DE/DW
Color printer SONY UP-D23MD
UP-D25MD
MITSUBISHI CP30DW
DVD Video Recorder SONY DVO-1000MD (Notice I)
PC printer HP HP Photosmart D5460 *
HP OFFICEJET 6000 *
Wireless LAN adapter CISCO Aironet 1130AG *
* These devices meet IEC60950 compliance.
(Notice I): EZU-ViF2 is required to record. EZU-ViF2 and EZU-ACP2 is required to playback.
(Notice II): EZU-ViF2 is required to print.
(NoticeⅢ): It is necessary to detach a rear handle to carry out simultaneous loading of a color printer and a
DVD video recorder.

- 10 - SN-U391E
3. Specifications
3.1. General
3.1.1. System
 OS Windows Embedded Standard 2009
 DVD DVD-RAM DVD-R
 USB memory USB HDD Connectable
 Network 10/100 BaseT (EZU-NW1 is required to network connection)
 Starting time About 100 seconds (About 70 seconds in Hibernation mode)

3.1.2. Basic Function


3.1.2.1 Type of probes
 Convex array probes 1.0~10 [MHz]
 Linear array probes 3.0~18 [MHz]
 Phased array probes 1.0~9.0 [MHz]

3.1.2.2 Image examination modes


 B, M mode B, B/B, B/M, M
 Doppler mode B/D,D
 CFM mode CFM-B, CFM-B/CFM-B, CFM-B/CFM-M, CFM-M, CFM-B/D
 Tissue Doppler mode TDI-B, TDI-M, TDI-PW
 Biplane mode B/B (Real-time biplane display)
 Harmonic dTHI-R,G,P : Filter method
dTHI-W-R,P : WPI (Wideband Pulse Inversion) method
HdTHI-R,P : HdTHI (High-definition dynamic tissue harmonic imaging)
 Compound HI Com
Applicable in Doppler and CFM mode
 Trapezoid Max display angle ±25°(Depend on probes)
 Oblique B-mode image Right and lest each 3 steps (Depend on probes)

3.1.2.3 Gain adjustment


 Depth Gain Control (DGC) 8 steps, 30dB variable, selectable Fixed depth/Variable
 B mode gain 60dB (adjustable in real-time and frozen)
 M mode gain 60dB (adjustable in real-time and frozen)
 Doppler mode gain 60dB (adjustable in real-time and frozen)
 CFM mode gain 60dB

3.1.2.4 Focus
 Transmission Max 10 steps (applicable to the setting of Focus width at multiple focus)

- 11 - SN-U391E
 Reception Continuous dynamic focusing with variable aperture
 Other 2-dimentional apodization

3.1.2.5 Frequency selection


 Fundamental Max. 5 steps
 Harmonic Max 6 steps (Depend on probes)
 HI Com: fundamental Max. 5 steps (Depend on probes)
 HI Com: harmonic Max. 6 steps (Depend on probes)

3.1.2.6 Acoustic power control


 Switching steps 5 ~ 100% (5% steps)
3 steps (CW mode)
 Applicable independent setup

3.1.2.7 B mode image display


 Single image

 Dual image Different probe and different depth can be displayed. Up-and-down can be selected for
linear probe.

3.1.2.8 Depth range


 20mm~360 mm, (5mm steps) (Depend on probes)

3.1.2.9 Display format change in frozen


Pan zoom, HI zoom, orientation (flip horizontal, flip vertical), 90 deg rotation, Image size change

3.1.2.10 Zoom function


 Pan zoom Adjustable position and size, display magnification: max. 8 times
 HI zoom Adjustable position and size depend on the probe shape

3.1.2.11 Image Size


80~100% (1% steps)

3.1.2.12 Display angle (width)


6 steps

3.1.2.13 Scan line density


4 steps

3.1.2.14 Biopsy guideline


6 selections (color: 2 types, size: 3 types)

- 12 - SN-U391E
(Notice) Biopsy guideline can not be displayed, when an application that image split setting of B-mode is
up-and-down is selected.

3.1.2.15 Marking assist display function


 Applicable Probe EUP-L75
 Condition of use A B-mode image or a CFM-mode image appears in single display
 Display method The Screen button pushing
 Displayed line A solid line and a dash line

3.1.2.16 Orientation
 Flip vertical and horizontal B mode, Doppler mode
 Flip vertical B mode: 90 deg steps rotation

3.1.2.17 Image processing (B mode and M mode)


 Dynamic range Max. 90dB 45~ 90 dB, 5 dB steps (can be changed in real-time and frozen)
 Enhancement

- Enhance 4 steps (can be changed in real-time and frozen without M mode)


- Weak-signal enhance Echo enhancement function, 8 steps
 Texture 2 steps (without M mode)
 Persistence 8 steps, including OFF
 AGC (auto gain control) 4 steps
 Gray map 8 steps (can be changed in real-time and frozen)
 Adaptive imaging HI REZ+ function, ON (8 steps)/OFF (can be changed in real-time and
frozen)
 B(M)color 29 maps (include B/W)
 PDC 29 steps
 Receiving Filter 3 steps
 Post processing γ-correction, rejection and saturation (can be changed in real-time and
frozen)

3.1.2.18 M mode image display


 Display method

- Real-time Moving bar


- Cine playback Scroll
 Sweep speed 1,1.5,2,3,4,6,8,10 s/screen (can be changed in real-time and frozen)
 Vertical shift Applicable
 Display area Side-by-side or up-and-down (can be changed in real-time and frozen),
split ratio: 1:1 (side-by-side), 1:1, 1:2, 2:1 (up-and-down)

- 13 - SN-U391E
3.1.2.19 Doppler function
 Display mode B/D (B-Freeze, B/D real), D
 Display area Side-by-side or up-and-down (can be changed in real-time and frozen),
split ratio: 1:1 (side-by-side), 1:1, 1:2, 2:1 (up-and-down)
 Analysis FFT method, 3 steps, max. 128 points
 Pulse repetition frequency (PRF)

1, 1.3, 1.5, 1.8, 2, 2.2, 2.5, 3, 3.5, 4, 4.5, 5, 5.5, 6, 6.5, 7, 7.5, 8, 9, 10, 11, 12,
14, 16, 18, 20 [kHz], total 26 steps
 Reference frequency (REF) 3 frequencies: (Depend on probes)
 Wall filter 12 steps (maximum value : 1/16 of PRF)
 Baseline shift Applicable (in real-time and frozen)
 Detectable depth 0mm ~360 mm
 Sample length 0.5 mm ~ 5.0 mm: 0.5 mm step,
5.0 mm ~ 15.0 mm: 1.0 mm step
 Measurement

- Real-time Velocity
- Frozen Velocity, Acceleration, Index for stenosis, Pressure gradient, Flow volume,
Time average velocity, Ratio-time, Ratio-velocity, Average velocity, PHT
(Pressure Half Time), HR (Heart Rate), TVI (Time Velocity Integral), Auto
trace is available
- Limitations HR: 25 ~ 300, measurement is not applicable in cine playback.
 Audio input / output 2ch speaker mounted (for forward / reverse flow), VCR audio input / output
(for forward / reverse flow)
 Maximum detectable velocity

±3.65m/s (2.0 [MHz], 20 [kHz])


 B/D simultaneous Up to 20 [kHz]
 Display method

- Real-time Moving bar


- Cine playback Scroll
 Sweep speed 1, 1.5, 2, 3, 4, 6, 8, 10 s /screen (can be changed in real-time and frozen)
 Angle correction Applicable (in real-time and frozen)
 FFT-γ 8 steps (can be changed in real-time and frozen)
 Oblique function Linear probe 0°,±10°,±15°,±20° (Depend on probes)
 D-Color 29maps (can be changed in real-time and frozen)
 Post-processing γ-correction, rejection and saturation (can be changed in real-time and
frozen)
 Unit 3types (cm/s, m/s, kHz)
 Smoothing PW simultaneous mode only
 Doppler real-time measurement

- Upper trace Trace upper the baseline

- 14 - SN-U391E
- Down trace Trace bottom the baseline
- Both trace comparing for peak of both upper and bottom baseline, and trace either one
higher

3.1.2.20 Color flow mapping function


 Display mode CFM-B, CFM-M, CFM-B/D (B-freeze, ALT), CFM-B/CFM-M
Applicable dual simultaneous display of CFM and B/W
 CFM-B

- Color box angle Continuously selectable in B-mode image display


- Color box depth Continuously selectable in B-mode image display
- Color ROI Color ROI guideline display on off selectable
- Oblique scan Linear probes: 0, ±10, ±15, ±20 deg (Depend on prbes)
Operations that are linked with or independent of Doppler cursor angles can
be performed.
 CFM-M

- Display method
 Real-time Moving bar
 Cine playback Scroll
- Sweep speed 1, 1.5, 2, 3, 4, 6, 8, 10 s /screen (can be changed in real-time and frozen)
 Color display

- Velocity - Velocity variance


The direction of blood and velocity is displayed in brightness of red and
blue, and add in brightness of green in the place of velocity variance.
- Velocity The direction and velocity is displayed in contrast of red and blue variation.
- CFA Angle dependence is small and color is displayed uniformly in blood vessels.
- Directional CFA The direction is displayed red and blue and angle dependence is small and
color is displayed uniformly in blood vessels.
- Fine Flow It displays blood flow power with higher definition than CFA mode. It is
possible to display microscopic vessels. (Include directional mode)
- Velocity variance The velocity variance is displayed in brightness of yellow.
- Color map CFI : 22maps
CFA : 13maps
Fin Flow : 9maps
- Color invert Applicable (can be changed in real-time and frozen)
 Pulse repetition frequency (PRF)

120, 250, 350, 400, 500, 650, 800, 1k, 1.3k, 1.5k, 2k, 2.5k, 3k, 3.5k, 4k, 4.5k,
5k, 6k, 8k, 10k, 12k, 16k, 20k [Hz], total 23 steps
 Persistence 8 steps
 Motion artifact suppression OFF + 3 steps
 Scan line density 8 steps

- 15 - SN-U391E
 Color preference Selectable 2 modes
- Auto Brighter signal is displayed between B and color
- Manual (16 steps) B image is displayed where B image intensity is bigger than preference bar
level on the gray scale.
 Peak hold [Not ECG display]
Overlay color data in real-time, Refresh time: 1s, 2s, 4s, 8s, ∞
[ECG display]
Overlay color data in real-time, Refresh time: 1beat, 2beat, 4beat, 8beat, ∞
 Velocity tagging Applicable
 Color filter 3 steps
 Interleave Max.32 lines (depending on PRF)
 Packet size 3 steps
 CFA dynamic range 8 steps (available to Fine flow)
 Reference frequency (REF) Max.3 frequencies (Depend on probes)

3.1.2.21 Compound function (HI Com function)


 Applicable probes Refer to Appendix
 Display mode B, dTHI, B/M, CFM-B, CFM-B/M, CFM-B/D, B/D, Elasto, Fine Flow
 Number of compound 2 steps (Max. 5 images)
 Oblique angle Max. 5 directions (include 0 deg)
 Scan line density 4 steps (Applicable independent setup)
 Steering HI Com Right and left (each 3steps) (can be changed to 3 step on linear array probes
only.)

3.1.2.22 Trapezoid display function


 Applicable probes Refer to Appendix
 Display mode B, CFM,PW ,Fine Flow
 Display angle Max. ±25°(Depend on probes)
 HI Zoom Applicable
 Prohibit mode HI Com, B oblique, PW/CFM oblique

3.1.2.23 Oblique B mode image function


 Applicable probes Refer to Appendix
 Display mode B, dTHI, B/M, CFM-B, CFM-B/M, CFM-B/D, B/D mode
 Oblique angle Right and left (each 3 steps) Angle is depend on probes
 HI Zoom NA

3.1.2.24 Automatic Image Optimization (HI Support)


 Applicable probes All probes(*1)
 Display mode All mode (Except as M, CFI/CFA/Fine Flow, TDI-B and TDI-M mode)

- 16 - SN-U391E
 Automatic optimization parameter

- B mode Gain, TGC, Depth-specific PDC


- PW mode Gain, PRF, Baseline
 Display

- B mode(Gain,TGC) ON : Reverse display of BG


- B mode(Depth-specific PDC) ON : Display “#” on PDC Screen button
- PW mode ON : Reverse display of DG and PRF
 Reset Available B mode only
 Brightness offset 21 steps (available B mode only)
*1) HI Support for PDC is available at linear probes only.

3.1.2.25 TDI (Tissue Doppler Imaging) function


 Applicable probes Refer to Appendix
 Display mode TDI-B, TDI-M, TDI-PW
 TDI-B, TDI-M mode

- Color map 8 maps (for TDI)


- Pulse repetition frequency (PRF)
120, 250, 350, 400, 500, 650, 800, 1k, 1.3k, 1.5k, 2k, 2.5k, 3k, 3.5k, 4k, 4.5k,
5k, 6k, 8k, 10k, 12k, 16k, 20k [Hz], total 23 steps
- Persistence 8 steps
- Smoothing 3 steps
- Blending 8 steps
- Velocity Tagging Applicable
 TDI-PW mode

- Analysis FFT method, 3 steps, Max. 128 points


- Pulse repetition frequency (PRF)
400, 500, 650, 800, 1k, 1.3k, 1.5k, 1.8k, 2k, 2.2k, 2.5k, 3k, 3.5k, 4k, 4.5k, 5k,
5.5k, 6k, 6.5k, 7k, 7.5k, 8k, 9k, 10k, 11k, 12k, 14k, 16k, 18k, 20k [Hz], total
30 steps
- Baseline shift Applicable (in real-time and frozen)
- Detectable depth 0mm ~360 mm
- Sample length 0.5 mm ~ 5.0 mm : 0.5 mm step,
5.0 mm ~ 15.0 mm : 1.0 mm step
- FFT-γ 8 steps (can be changed in real-time and frozen)

3.1.2.26 Real-time Biplane mode


 Applicable probe Convex type probe EUP-CC531S only

3.1.2.27 Cine memory function


 Recording mode B, M/D (include CFM)

- 17 - SN-U391E
 B-image Frame storage capacity

Max. 19,000 frames


 M/D recording time Max. 300s
 ODM recording time Max. 150s
 Cine memory split B/B split, B/MD split
 Playing images Continuous playback (B)
Rewinding frame by frame (B, M/D)
Automatically detected heartbeats (B)
(Mounted a Biomedical Waveform Display Unit)
 Play-back speed B : ±1/1, ±1/2, ±1/4, ±1/8, 0
M/D : Continuous playback is not applicable
 Phase bars Applicable (B)
 Changing the time axis (Freeze on)

Applicable (M/D)

3.1.2.28 Preset
 Application Max 22 sets of operating parameters can be registered for each probes
Max 10 set of user registered
 PSS function Registered for each application
Max 8 set of preset
 Region Data Setting Max 37 set of region data
Max 30 set of user registered

3.1.2.29 Character display


 Comments 102 characters×30 lines maximum.4 kinds of character size, 4 character style
(Normal / Bold / Italic / Bold italic) , 3 entry mode (Insert / Letter Overwrite
/ Single Word Overwrite), Auto Complete Entry, 2 initial cursor position
( Home Position / Previous Position)
 ID name Regular 15 characters
 Patient name Regular 26 characters
 Hospital name Regular 29 characters
 Automatically displayed items

Data and time (year, month, day, hour, minute, and second), Physician, age,
sex, Probe name, frequency of the probes, gain settings, acoustic power level,
post-processing parameters, dynamic range (DYN), delay times from
R-wave, heart rate, measurement values, Doppler function display, focus
point, view of depths display.
 Annotation Preregistered 402 annotations (abbreviated medical terms)

- 18 - SN-U391E
3.1.2.30 Body marks
 Display position Adjustable
 Display size Large or Middle
 Types

- Abdomen 8 types
- Breast 10 types
- Head/Neck 15 types
- Urology 10 types
- Cardiovascular 14 types
- Organ 11 types
- Upper limb 18 types
- Lower limb 10 types
- Obstetrics 21 types
- Gynecology 4 types

3.1.2.31 Measurement function


 General measurements Distance, area, circumference, volume, velocity, time,Doppler trace, Accel.,
Decel., velocity, average velocity, histogram, RI, PI, FlowVolume, Doppler
profile, velocity ratio, time ratio, distance ratio, area ratio, heart rate, and
CFM velocity,Angle
 Abdomen, urology, Hip joint angle, urology, SVolume, and Vascularity
and others
 Cardiovascular Left ventricle measurements, LV, MV, Ao, TV, CO, EF, regurgitant volume,
measurements dP/dt, Qp/Qs, blood flow measurements, MVG, and ventricular mass
 Vascular measurements Carotid artery, upper extremities, lower extremities, and %Steno
 Obstetrics and gynecology Fetal measurements, blood flow measurements, fetal weight estimates,
measurements gestational age estimates, fetal growth evaluations, ovarian follicle
measurements, and gynecology measurements
 Real-time Doppler Blood flow waveforms are traced in real time and the measurement results
measurements are displayed. The average heart rate can be calculated based on several
measurements.
 Measurements of retrieved Measurements can be performed only for images recorded using the image
images filing function. (Video playback measurements are not supported.)
 Obstetric tables Data specific to the United States, Europe, and Japan are incorporated. You
can create a new table or edit an existing table. Tables can be loaded to or
from external media.
 Report function You can display a list of measurement results, edit results, enter comments,
print reports, or display obstetrics graphs. You can record the measurement
results to a hard disk or other supported media.
 User-defined functions You can register unique measurement packages by combining measurement

- 19 - SN-U391E
items, measurement orders, and formulas.

3.1.2.32 Media Management


 Folder or file list display Applicable
 Free space display Applicable
 Delete and copy of deleted image data

Applicable
⁃ Copy destination USB memory, USB HDD, DVD-RAM, Network folder

3.1.2.33 Storage to hard disk


 Capacity Maximum 60GB
 Number of images B/W : approx. maximum 72,000
Color : approx. maximum 24,000

3.1.2.34 Filing Function / Patient information window


 Image transfer Digital
 Transferring image Motion image / Still image
 Motion image transfer method

- Frame to Frame (F-F) All image or 10, 20, 30, 40, 50, 60, 70, 80, 90, 100 images from last image
- R-Wave to R-Wave (R-R)
All image or 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 beats from last image
(Mounted a Biomedical Waveform Display Unit)
- All
(Notice) : When the number of images in cine memory exceeds the record upper limit, all image
transfer is not available.
 Image format BMP (Bitmap), TIFF, AVI, AVI (Compression), JPEG, DICOM,
DICOM JPEG
 TIFF compression format None, Pack Bits
 AVI compress Microsoft Video 1
 Number of retrieve images 1, 4, 9, 16
 Retrieve mode Frame by frame review Continuous review, Changing the time axis
 Retrieve speed -1/1, -1/2, -1/4, -1/8, 0, 1/8, 1/4, 1/2, 1/1
 Divided frame display Available with motion image review
 Review range specification Applicable
 Playback measurement Applicable excluding D-Profile, CFM-Velo, histogram, auto trace and PISA,
MVG-M, vascularity in MV Regurg and TV Regurg
 Filing with Each image, Each patient
 Masking of ID and patient name

Applicable (in transferring to external media)

- 20 - SN-U391E
 Patient information window

- Patient data list display Applicable


- Patient ID assigned automatically (Emergency situation)
Applicable
- Pre-input patient information
Applicable
- Single Image recording Applicable
- Auto complete setting Applicable
- The number of the patients who can register
10,000 (It is assumed that for 1 year and half.)
 Filing window

- Keyword search Today's Study, Search All, Patient ID, Patient name, Study Date, Study Term,
Body Part Exam., Examined by, Ref. Physician, Additional Patient History,
Application, Comment, Transfer Status
- Thumbnail images sorting
Applicable
- Image Data copy, delete applicable
- Correction of the patient information for the saved examination data like an image.
Not applicable
(Notice) : Other patients’ examination data cannot be referred to during examination.
(Notice) : To record data onto media that was created by using equipment that uses different software
version, select the target location list in the Filing window, and then update the database before
recording the data.
 OS of the network computer which can perform file sharing.

Windows XP Professional,
Windows Vista Ultimate/Home Premium
Windows 7 Ultimate/Professional/Home Premium
Windows 8/Windows 8 Pro
Windows 8.1/Windows 8.1 Pro
(Notice) : The simultaneous reference of network folder cannot be performed from two or more system.
(Notice) : File sharing is available when using a password-protected account.

3.1.2.35 Others
 Brightness Setting Character 6 level
Background 6 level
 Screensaver Applicable
 Online Manual Applicable
 Screen Keyboard Enter alphanumeric characters and symbols
(If the optional alphanumeric keyboard is not mounted)

- 21 - SN-U391E
3.2. Options
3.2.1. 17-inch Digital LCD Monitor (EZU-MT29-S1)
 Standard component LCD monitor, LCD simple arm, cables
 Screen 17 inch, Non-interlace
 Input voltage DC12V
 Power consumption Max.34VA
 Video signal DVI-D

3.2.2. 17-inch Digital LCD Monitor (Flexible arm) (EZU-MT30-S1)


 Standard component LCD monitor, LCD flexible arm, cables
 Screen 17 inch, Non-interlace
 Input voltage DC12V
 Power consumption Max.34VA
 Video signal DVI-D

3.2.3. Biomedical Waveform Display Unit (EZU-EK28)


 Components BIO PCB, ECG panel
 Display signal ECG, Ext-ECG
 Display method

- Real-time Moving bar


- Cine playback Scroll
 Display control Vertical movement, gain control, auto gain: 3 steps
 Sweep speed 1, 1.5, 2, 3, 4, 6, 8, 10 s/screen (can be changed in real-time and frozen)
 R-R Detection RRp/RRpp≒1 interval is detected automatically in freeze mode. (RRp:
preceding R-R interval, RRpp: prepreceding R-R interval)
(Notice) Biomedical waveform cannot be displayed, when an application that image split setting of B-mode
is up-and-down is selected.

3.2.4. CW Doppler Unit (EZU-ST8)


 Components CW PCB, total 1pcbs
 Applicable probes EUP-S50A, S70, S52, S72, ES52E, ES52M, Single Element CW Doppler
probe
 Display mode CW, B/CW (B, B-freeze), CFM-B/CW (CFM-B, B-freeze)
 Reference frequency (REF)

- EUP-S50A 2.0 [MHz]


- EUP-S70 2.0 [MHz]
- EUP-S52 3.5 [MHz]
- EUP-S72 3.0 [MHz]
- EUP-ES52E 3.5 [MHz]

- 22 - SN-U391E
- EUP-ES52M 3.5 [MHz]
 FFT pulse repetition frequency (PRF)

4, 4.5, 5, 6, 8, 10, 12, 16, 20, 25, 35, 50 [kHz], total 12 steps
 Analysis FFT method, 3Steps, Max. 256point
 Maximum detectable velocity

8.91m/s (2.0 [MHz], 50 [kHz])


 Sweep speed 1, 1.5, 2, 3, 4, 6, 8, 10 s/screen (can be changed in real-time and frozen)
 Wall filter 12 steps (maximum value : 1/8 of PRF)
 Transmission-reception focus

8 steps (automatic follow-up the sampling gate)


 Baseline shift Applicable (in real-time and frozen)
 Angle correction Applicable (in real-time and frozen)
 Post-processing γ-correction, rejection and saturation (can be changed in real-time and
frozen)
 D-Color 29maps (can be changed in real-time and frozen)
 Echo Enhancement 4 steps, including OFF

3.2.5. Real-time Tissue Elastography Unit (EZU-TE6)


 Components Stabilizers for compression assistance
 Applicable probes Refer to Appendix
 Display mode

- Elasto mode Overlapping display of Elastography on B-mode image


- Dual Elasto mode Dual image display of Elasto mode, possible to select different color map on
each side, also possible to display B-mode image on the right and Elasto
image on the left side
- Real-time biplane
Elasto mode Overlapping display of Elastography on B-mode image of either
cross-section, in the function to display B-mode image with the
one of different cross-section simultaneously
Specifications
 Reference frequency (REF) 2 frequencies (L65, L54MA, L75 and O54J: 1 frequency)
 Persistence 8 steps
 Smoothing 4 steps
 Dynamic range 8 steps
 Frame rejection 8 steps
 Noise rejection 8 steps
 Scan line density 4 steps
 Brightness Control 11steps
 Color map 8 maps(red color and blue color can be inverted)
 Tagging Max. 65 steps in both upper and lower limit values

- 23 - SN-U391E
 Rejection 65 steps
 Blending 0% ~ 100%, 2% steps
 Blend Type 2 types
 Frame rate 5steps
 Dual mode Applicable
 Biplane mode Applicable (EUP-CC531S)
 Biopsy guide line Applicable
 Press Guide function

- Strain Graph Displays the variation over time of tissue strain caused by compression.
Available graph to exclude low strain area.
- Press Indicator Display the compression status
 Frame select function Possible to select images under stable compression automatically
 Strain ratio measurement

- Number of ROI Max. 10, possible to position the ROIs on B mode image
- ROI shape Circle, oval, rectangle and closed spline curve
- Averaging function Max. 5 frames
- Auto ROI Setting function
Clicking center of the tumor automatically sets the measurement ROI to the
tumor and the fatty layer (the mammary gland region and lesion are targeted).
- Measurements of retrieved images Applicable

3.2.6. Real-time Tissue Elastography software (Strain Histogram)


(EZU-TESH1)
 Components Software
 Applicable Probes Same probe as Real-time Tissue Elastography
Specifications
 Strain Histogram Measurement
- Strain Graph 1 point
- ROI Shape rectangle
- Measurements of retrieved images Applicable
 Data export Applicable (CSV format)

(Notice) : EZU-TESH1 is required for EZU-TE6.


(Notice) : If B-mode images are displayed at the top and bottom part of the screen in dual display,
Real Time Tissue Elastography software (Strain Histogram) cannot be used.

- 24 - SN-U391E
3.2.7. Digital Video Archiving Software (EZU-DV4)
 Components Software

3.2.7.1 Clip capture mode


 Common specification
- Preset 8 steps
- Compression level 2 steps (Standard/HI Compression)
- Image format AVI (uncompressed, Motion JPEG Compression), DICOM, DICOM JPEG
- Playback speed
・ Play 1x
・ Rewind -1x
・ Slow-forward 1/8x, 1/4x, 1/2x
・ Slow-rewind -1/8x, -1/4x, -1/2x
 Specifications – in real-time

- Transfer mode Free capture, Beat Duration


- Capture rate Free capture : 6, 8, 10, 15, 20, 30 fps 6steps
Beat Duration : 30 fps
- Capture time Free capture : 10 sec (30 fps), 90 sec (20 fps), 120 sec (15 fps),
180 sec (10 fps), 225 sec (8 fps), 300 sec (6 fps)
Beat Duration : 1, 2, 3, 4, 5, 6, 7, 8, 9 and 10 heartbeat
- Measurements on clip capture
Applicable
However, there are following limitations for a trimmed clip.
・ Necessary to set all image parameter before measurement
・ Impossible to perform the following functions
%WT measurement, Color Tissue Tracking measurement,
LA Tracking measurement, Time Intensity Curve function
 Specifications – in frozen

- Transfer mode Frame to Frame (F-F) / R-Wave to R-Wave (R-R) / All


- Capture time Free capture :
AVI(Motion JPEG Compressiona) or DICOM JPEG format
10 sec (30 fps), 90 sec (20 fps), 120 sec (15 fps),180 sec (10 fps),
225 sec (8 fps), 300 sec (6 fps)
AVI(uncompressed) or DICOM format
10 sec (30 fps), 18 sec (20 fps), 24 sec (15 fps),
36 sec (10 fps), 45 sec (8 fps), 60 sec (6 fps)
R-Wave to R-Wave : ALL, 1, 2, 3, 4, 5, 6, 7, 8, 9 and 10 heartbeat/manual
(Notice) : EZU-DCP3 is required for EZU-DV4.
(Notice) : EZU-EK28 is required for R-wave to R-wave mode
(Notice) : EZU-FC11 is required for selecting DICOM JPEG as the image format.

- 25 - SN-U391E
(Notice) : When the number of images in cine memory exceeds the record upper limit, all image transfer is
not available.
(Notice) : Quality of AVI(uncompressed) or DICOM format and AVI(Motion JPEG Compression) or
DICOM JPEG format are same in real-time.

3.2.7.2 Long time recording mode


 Recording format MPEG2 (can be played on image filing function)
 Destination DVD-RAM / Hard Disk / USB HDD
 Continuous recordable time Maximum 1 examination
 Recordable time to DVD

- High-quality Approx. 1 hour (8Mbps)


- Standard Approx 2 hours (4Mbps)
- High-compression Approx 4 hours (2Mbps)
 Recording area Fundamental area
 Record and play of Doppler sound

Applicable
 Playback speed

- Play 1×
- Fast-forward 2×, 8×, 32×
- Fast-rewind -1×, -2×, -8×, -32×
- Slow-forward 1/8×, 1/4×, 1/2×
- Slow-rewind -1/8×, -1/4×, -1/2×
 Playback to the start by chapter

Applicable
 Playback function Only MPEG2 clip file that is captured this software
(Notice) : EZU-DCP3 is required for EZU-DV4.

3.2.8. Logon Management Software (EZU-LG2)


 Components software
 Registerable account number

- Service 1 (undeletable)
- Administrator max. 30
- General user max. 30
 User licensing function

- User ID Alphanumeric half size of character, symbols (except space), 1~16


characters
- Password Alphanumeric half size of character, 8~16 characters, (Not applicable to
input fewer than 8 characters and same letter string of user ID)
- Password changing Applicable
 Account management (permitted to administrator only)

- 26 - SN-U391E
- Display of information Displaying a list of account
- Registration function New registration of account
- Delete function Deregistration of account
 Logoff function

- Authentication release Applicable


- Auto logoff Applicable

3.2.9. DICOM Software(Transfer and Media Storage) (EZU-FC11)


 Components Software
 Functions

- Transfer Transfer to DICOM server


- Media transfer Save to the DVD-RAM, DVD-R, USB memory or USB HDD as DICOM
format
- Connection check C-ECHO function (carries out a connection check with server)
- Storage commitment function
The function to query whether to be preserved normally after transferring for
ultrasound images
- DICOM SR The function to transfer the measurement results to report server as DICOM
SR (Structured report)
A measurement item name is modifiable.
It is possible to copy a measurement which does not support DICOM SR for
the measurement using DICOM SR.
It is possible to add a measurement which does not support DICOM SR for
the measurement region using DICOM SR.
 Transfer image Single images, motion images (It is applicable the collectively transferring
by patient)
 Brightness adjustment The brightness of images can be adjusted when transferring them to the
storage server. The way of setting the corrected information to a DICOM
attribute, and the way of changing the brightness of the image pixel value to
selection is possible.
 Anonymize patient info. The patient information printed in image can be masked when transferring
them to the storage server.
 Motion images transfer Frame to frame transfer, R-R transfer [ECG is used], all images transfer in
cine memory
 Retrieve DICOM format images saved in the folder on network, DVD-RAM,
DVD-R,USB memory and USB HDD
It is applicable to play after DICOM server transfer by EZU-FC11Q.
 Image conversion BMP to DICOM, DICOM JPEG
AVI to DICOM, DICOM JPEG
 LAN 10/100BaseT

- 27 - SN-U391E
 DICOM log acquisition Applicable (DICOM log in communication)
 SR transfer operation DICOM SR is transferred by pushing "SR Transfer" button in the Filing
Window, "Save" button in the Report Winodw, Patient Key or End Exam
key.
 Output measurement item to DICOM SR

- OB-GYN measurement (OB_SR)


- OB measurement
GS, CRL, BPD, OFD, HC, LVW, HW, TCD, IOD, OD, APTD, TTD, AC,
FTA, HL, ULNA, RAD, FL, TIB, FIB, EFBW, CI, HC/AC, FL/AC, FL/BPD,
APTD*TAD
AFI measurement: Q1, Q2, Q3, Q4, AFI
- Gynecology measurement
Uterus measurement: Uterus L, Uterus H, Uterus W, Endomet. T., Cervix L
L-Ovary measurement: L, AP, W, V
R-Ovary measurement: L,AP,W,V
- Follicular measurement
L-Ovary measurement: L, AP, W, V
R-Ovary measurement: L, AP, W, V
L-Follicular measurement: L, AP, W, V
R-Follicular measurement: L, AP, W, V
Uterus measurement: Uterus L, Uterus AP, Uterus W, V, Endomet.
T.,Cervix L
- OB-GYN measurement (OB_SR_II)
- Vascular measurment
Fetal_Vasc measurment: Umbil.A. measurment, MCA measurment, OB-DV
measurment, OB-IVC measurment,
OB-Daorta measurment, OB-Aorta measurment,
OB-PA measurment, CIA measurment, L-RA measurment,
R-RA measurment, Umbil.V. measurment
Pelvic_Vasc measurment: L-Ovary.A. measurment,
R-Ovary.A. measurment, L-Uterine A. measurment,
R-Uterine A. measurment
- OB measurement
GS, CRL, BPD, OFD, HC, LVW, HW, TCD, IOD, OD,
APTD, TTD, AC, FTA, HL, ULNA, RAD, FL, TIB, FIB,
EFBW, CI, HC/AC, FL/AC, FL/BPD, APTD*TAD, Foot, NB, NT
AFI measurement: Q1, Q2, Q3, Q4, AFI
- Gynecology measurement
Uterus measurement: Uterus L, Uterus H, Uterus W,
Endomet. T., Cervix L

- 28 - SN-U391E
L-Ovary measurement: L, AP, W, V
R-Ovary measurement: L,AP,W,V
- Follicular measurement
L-Ovary measurement: L, AP, W, V
R-Ovary measurement: L, AP, W, V
L-Follicular measurement :L,AP,W,V
R-Follicular measurement :L,AP,W,V
Uterus measurement: Uterus L, Uterus AP, Uterus W, V,
Endomet. T., Cervix L
- Vascular measurement Extracranial Arteries measurement: Bulb, CCA, ECA, ICA, PROX, ICA
MID, ICA DIST, ICA(RATIO), Subclav.A, Vertebral A
Lower Extremity Arteries measurement: CIA, ATA, CFA, DPA, EIA, IIA,
Pero.A, Pop.A, PTA, PFA, SFA
Lower Extremity Veins measurement: ATV, CFV, CIV, EIV, GV, GSV, LSV,
Pero.V, Pop.V, PTV, PFV, SFJ, SV, SFV, SPJ, Hunter, Cocket, Boyd
- Cardiac measurement (Cardiac_SR)
Left Ventricle measurement: HR, LVEdD, LVEsD, FS, LVEdV, LVEsV, EF,
Regurg Vol., Flow Rate, PISA, PISA-R, Vp, TVI, IVRT, e', E/e', A, S, SV, CO,
CI, IVSd, LVPWs, LVPWd, LVMass-d, IVSd/LVPWd
Right Ventricle measurement: HR, Regurg Vol., Flow Rate, PISA, PISA-R,
Vp, TVI, SV, CO, CI
Aortic Valve measurement: Avcusp Open, ET, Ao/ET, Regurg Vol., Flow
Rate, PISA, PISA-R, Vp, TVI
Mitral Valve measurement: A, E, E/A, EFslope, EPSS, Adur, Regurg Vol.,
Flow Rate, PISA, PISA-R, Vp, TVI, SV, CO, CI, dp/dt
Pulmonic Valve measurement: Regurg Vol., Flow Rate, PISA, PISA-R, Vp,
TVI
Tricuspid Valve measurement: Regurg Vol., Flow Rate, PISA, PISA-R, Vp,
TVI
Pulmonary Veins measurement: S, D, A, Adur, HR
Congenital measurement: HR, Regurg Vol., Flow Rate, PISA, PISA-R, Vp,
TVI, Qp/Os
- Cardiac measurement (Cardiac_SR_II)
Echo LV measurement : B-LV, M-LV, M.Simpson, Dia.Func,
Dop Index, CO, BI-Plane, Area Length, LVMass-AL,
LVMass-TE, HR
Echo LA measurement : B-LA, M-LA, TEE-LA
Echo RV measurement : CO, B-RV, M-RV, HR
Echo RA measurement : B-RA
Echo AV measurement : AV Regurg, Velocity, PHT, D-Trace,

- 29 - SN-U391E
B-Area, Area-Vp, Area-Vm, Area-VTI, HR
Echo MV measurement : Dia.Func, dP/dt, MV Regurg, Velocity,
PHT, D-Trace, B-Area, Area-Vp, Area-Vm, Area-VTI, MV-RegVol, HR
Echo PV measurement : PV Regurg, Velocity, PHT, D-Trace, HR
Echo TV measurement : TV Regurg, Velocity, PHT, D-Trace, HR
Echo Ao measurement : B Ao, Ao, HR
Echo PArtery measurement : B-PArtery, D-PArtery, HR
Echo IVC measurement : B-IVC, HR
Echo PVein measurement : D-PVein, HR
Echo CS measurement : Qp/Qs, HR
Echo Cong measurement : Cong Regurg, HR

(Notice) : EZU-NW1 is required for EZU-FC11.

- 30 - SN-U391E
3.2.10. DICOM Software (WorkList) (EZU-FC11W) / DICOM Software
(Printer) (EZU-FC11P)
 Components Software
 Functions

- Printer function Printout to a DICOM printer on network


- WorkList Send/receive a patient data to/from a DICOM WorkList server
 Specifications for printer

- Print format B/W: MONOCHROME2, color: RGB


- Images in a film Max. 16 images
- Start of printing Each image : Press REC key
Each patient : Press End Exam key or click Patient key,
Click Image Print button on the Filing Dialog
- Film format setup Film size, image display format, film orientation, film destination
- Other setup Number of copies, magnification type, print priority
- Simultaneous transfer Simultaneous transfer to DICOM server and printer
 Specifications for WorkList Acquisition of patient data
Retrieval of patient data
Display of detailed patient data
The patient information which is indicating by the list using the mobile
function can be held, and use is possible at off-line
Setup is freely possible in the tag which quer(ies) to a work list server
It is possible automatic calculation of the age based on the birth date obtained
from Worklist
Possible to query automatically when the Patient Information window is
displayed.

(Notice) : EZU-NW1 and EZU-FC11 are required for EZU-FC11P and EZU-FC11W.

3.2.11. DICOM Software (Query / Retrieve) (EZU-FC11Q)


 Components Software
 Specifications

- Retrieve modality US, CT, MR, MG, EC


- Query key ID, Name, Date, Accession #, Modality
- Multi-display 1, 2, 4, 6 display
- Save format BMP, JPEG
- Display format Original size, window size
- Image size Zoom in, zoom out, panning
- Window level Window level, ROI window level, B/W inversion
- Rotation, inversion 90 deg rotation to right and left, 180 deg rotation, flip horizontal, flip vertical
- Review By frame, loop, synchronization

- 31 - SN-U391E
- Measurement Caliper, distance ratio, angle, area, average velocity, acceleration, heart rate
(Notice: If there is no tag for calibrations, average velocity, acceleration and
heart rate cannot be measured.)
- Distance calibration Applicable in caliper and area
- Annotation Applicable (text and rectangle on display)
- Hard disk storage Images acquired from DICOM server can be preserved up to 3GB.

(Notice) : EZU-NW1 and EZU-FC11 are required for EZU-FC11Q.

3.2.12. Omni-directional M mode display function (EZU-OD3)


 Applicable probes Refer to Appendix
 Display mode

- Real-time ODM mode B (Real) / ODM, B (Real) / ODM / ODM, ODM,ODM / ODM
- Cine ODM mode B (Cine) / ODM, B (Cine) / ODM / ODM
 Settings

- Number of ODM beam line


1 direction / 2 directions
- ODM beam line center control
Common center mode / Independent center mode
- Angle rotation 0 ~ 359 deg, 1 deg steps
- Center position movement
1 pixel steps
- Sweep speed 1, 1.5, 2, 3, 4, 6, 8, 10 s/screen (can be changed in real-time and frozen)
- ODM dynamic range 45~ 90 dB, 5 dB steps (can be changed in real-time and frozen)
- ODM enhancement 4 steps
- Gray map 8 steps (can be changed in real-time and frozen)
- B image scan density Fixes
- B image display angle Selectable (available for Real-time ODM mode)
 Display method

- Real-time Moving bar


- Cine Moving bar
 Measurement M mode measurement
 Others Not applicable with Doppler and CFM mode

3.2.13. Advanced Measurement Software (EZU-AM7)


 Components Software
 Specifications - %WT (wall thickness) measurement
- Image format AVI, Motion JPEG (B/W B mode image)
- Analysis point Max. 20 points
- WT graph display Applicable

- 32 - SN-U391E
- Data export Applicable (CSV format)
 Specifications - IMT (Intima-Media Thickness) measurement
- Image format Still image (Frozen, playback image)
- Measurement results
 Trace method Max, mean, value of 3 points, average value of 3 points
 Simple 3 points method
Value of 3 points, average value of 3 points
- Manual correction of trace line
Applicable (in trace method)
- Detection position correction
Applicable (in simple points method)
 Specifications - Color tissue tracking measurement
- Image format AVI, Motion JPEG (B/W B mode image)
- Specified measurement segments
APICAL, MID, BASAL
- Graphs for transition Applicable
- Data export Applicable (CSV format)
- Creating Bull’s eye Applicable
 Specifications - LA tracking measurement
- Image format AVI, Motion JPEG (B/W B mode image)
- Measurement results Volume, dV/dt, area, circulation, long axis
- Volume measurement Modified Simpson method
- Biplane measurement Applicable
- Graphs for transition Applicable
- Data export Applicable (CSV format)

3.2.14. IMT Measurement Software (EZU-IM2)


This is the same function as IMT measurement in EZU-AM7. Please refer to EZU-AM7 for details.

3.2.15. Software Specified for Contrast Agents (EZU-CH8)


 Components Software
 Applicable probes Refer to Appendix
 Display mode WPI (Wideband pulse inversion) mode
CWPI (Color wideband pulse inversion) mode
 Intermittent mode Monitor mode (possible to shot manually)
Multi-shot mode (2~16 shot, total 15 steps)
 Trigger timing

- System clock trigger timing (ECG display off)


 Abdominal 32 steps
0.1s ~ 1.0s: 0.1s step

- 33 - SN-U391E
1.5s ~ 10.0s: 0.5s step
11.0s ~ 14.0s: 1s step
 Cardiac 31 steps
0.2s ~ 1.0s: 0.1s step
1.5s ~ 10.0s: 0.5s step
11.0s ~ 14.0s: 1s step
- ECG trigger timing (ECG display on) 32 steps
1 ~ 32 beats, 1 beat step
 Acoustic pressure %Control and MI Control are selectable
MI Control : 0~0.5 0.01 step
0.5~ 0.05 step
(Max MI value: Depend on probes and focus setting and so on)
%Control : 0~100% 34 steps
Automatic acoustic pressure control on fundamental image in monitor mode
and alternate mode is possible.
 Reference frequency (REF) 2 frequencies
 Stopwatch function -30s ~ 59m59s, for monitoring the time after injection of contrast agent
For monitoring the time of image recorded in cine memory.
 Application function Possible to independently set up each application parameter for contrast
imaging
 Multi-step trigger (MST) mode

Possible to make a desirable protocol for contrast imaging by combining


freely an intermittent transmission interval and execute it.
- Number of registrable protocol
Max. 8 protocols
- Number of Steps for each protocol Max.
8 steps
- Items for registration Start time, Interval time, Image numbers
- Data Reading, Saving Applicable
 Flash function Breaking contrast agent for observing replenishment, Number of flash: 1, 2,
4, 8, 16, 32 frames, total 6 steps
 Intermittent flash mode

- System clock trigger timing (ECG display off)


5, 10, 15, 20, 25, 30, 35, 40s, total 8 steps
- ECG trigger timing (ECG display on) 32 steps
1 ~ 32 beats, 1 beat step, total 32 steps
 UCI (uniform contrast imaging) function

This function uses the technology of a different aspheric focusing from the
conventional focal technique, it can generate uniform acoustic field in the
body

- 34 - SN-U391E
 Alternate mode In dual screen displays in alternate mode, contrast image (left screen) and
fundamental image (right screen) are transmitted and displayed alternately,
ultrasound contrast agents and anatomy can be observed on real time
(applicable in middle and low acoustic pressure transmission mode)
 MTI (microbubble trace imaging) mode

To improve the visibility of running of the contrast agents, the maximum


intensity of each frame of the contrast image is overlaid
(applicable in middle and low acoustic pressure transmission mode)
 Color wideband pulse inversion mode

- Selectable Color overlay mode / Color stand-alone mode


- Color map 8 maps
- Smoothing 3 steps
- Dynamic range 45~80 dB, 5 dB steps (can be changed in real-time and frozen)
- Gain 60dB (adjustable in real-time and frozen)
- Scan line density 4 steps
- Receiving filter 3 steps
- Frame rate 3 steps
- Display mode Selectable triplet ON/OFF (in high acoustic pressure transmission mode)
 Time intensity curve function

- Motion image format AVI, Motion JPEG


- Applicable mode B, B/B
- Number of ROIs Max. 12
- Shapes of ROI Circle, ellipse, square, rectangle, banana, freehand
- Operation of ROI Copy, rotate, delete, interpolation
- Display magnification ×1, ×2, ×4
- Review of motion image Review (forward), frame review (forward / backward), review speed (5 steps
for each direction), move frame (start frame / end frame)
- Analysis result Display of ROI, graphs and data
- Graph display Averaged intensity, fitting curve (2 types)
- Data export Applicable (CSV format)

(Notice) :Up and Down display of B mode can not be selected, when an applicable application of
EZU-CH8 is selected.

3.2.16. WideView Software (EZU-WS8)


 Components Software
 Applicable probes Refer to Appendix
 Operating condition B / CFM-B single display
 Display mode WideView display, cine playback display
 Display functions Zoom, Panning, Rotation

- 35 - SN-U391E
 Measurement function B mode measurement (except for histogram and CFM VELO)

(Notice) : EZU-DCP3 is required for EZU-WS8.


(Notice) : Capture operation by foot switch can be done by connecting EZU-FS2.

3.2.17. Stress Echo Software (EZU-SE6)


 Components Software
 Specifications

- Applicable mode B mode (single image display)


- Database Patient Information, Stress examination information (Images, WMS report)
- Protocol Conformity with ASE (American Society of Echocardiography), user
programmable
- Capture functions Max. 12 stages, Max. 10 loop/time, selectable capture area, R-R capture
- Playback functions Multiple-display simultaneous playback (automatic synchronizer), standard
playback, frame by frame playback, reverse display, display of selected
images
- Full disclosure mode User can continuously acquire loops for a certain time (mostly 2 minutes)
Later, in select mode, the user can select the appropriate views
- Gain and ROI save function
Gain setting and the ROI position of the previous stage are saved and applied
for the next stage
- Image input signal Digital YUV
- Measurement function LV volume measurement
- Report function WMS report, LV volume measurement report, user comment
- Print B/W, color
- Protocol setting By protocol editor
- Image data export Applicable (BMP, AVI)
- Setting for destination of the image to save
By archive tool

(Notice) : EZU-DCP3 and EZU-EK28 are required for EZU-SE6.


(Notice): Not available to EZU-SE6 when selecting U/D the Image Split.
(Notice) : Capture operation by foot switch can be done by connecting EZU-FS2.

3.2.18. Three-dimensional display unit (Real-Time) (EZU-4D7)


 Components Software (Real time 3D software), MTDRV PCB, MTBP PCB
 Applicable probe Refer to Appendix
 Display mode

- ROI setting mode Displays the ROI that specifies the 3D image creation area on the 2D image
- 3D mode Real-time 3D image is displayed on 1 screen.

- 36 - SN-U391E
- 2D-3D mode Screen is vertically divided into two sides. The ROI setting mode is
displayed on the left side and the 3D mode on the right side
- MPR 4 screen mode Screen is vertically and horizontally divided into four. 3D image is displayed
on the lower right, while the random orthogonal 3 cross-section surface
images are displayed on the other sections
- MPR 1 screen mode One of the orthogonal 3 cross-section surface image in MPR 4 screen mode
is displayed on 1 screen
- Multi Slice multi screen mode
Screen is divides into multi (2, 4, 9, 16) sections. A reference image is
displayed on the upper left section, and the slice images are displayed on
the other sections.
- Multi Slice 1 screen mode
The selected slice image from Multi Slice multi screen mode is displayed
on 1 screen.
 HI Definition 3D By one swing scanning at a speed that is slower than real time 3D, and
high-definition 3D image or MPR image is displayed.
Applicable to CFM mode (CFI, CFA and Fine Flow except for color maps
for displays of velocity variance)
 Rendering mode

B/W
- Opacity
- MIP (Maximum Intensity Projection)
- mIP (minimum Intensity Projection)
- Transparency
- Inversion
- Orthogonal
- Opac. Advanced
- 4Dshading
Color
- Opacity
- MIP (Maximum Intensity Projection)
- Transparency
- Orthogonal
 Quality settings

- Image quality Real-time 3D 6 steps


HI Definition 3D 3 steps
B/W
- Brightness
 3D image 17 steps (-8 ~ 8, except rendering mode: Orthogonal)
 MPR image 17 steps (-8 ~ 8)

- 37 - SN-U391E
- Color
 3D color 37 types (except rendering mode: Orthogonal, Opac.Advanced, and
4Dshading)
 MPR color 37 types
 Depth color 14 types (rendering mode: Opac.Advanced)
- Shading 6 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
- Filter 20 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced,, or
4Dshading)
- Threshold 256 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced, or
4Dshading)
- Smoothing 8 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
- P-Filter 8 steps (rendering mode: MIP or mIP)
- BW 3D Smooth. 3 types (rendering mode: 4Dshading)
- 3D Smooth Coef 11 types (rendering mode: Opac.Advanced, or 4Dshading)
- Depth Contrast 40 types (rendering mode: Opac.Advanced)
- Depth Color Shift 127 types (rendering mode: Opac.Advanced)
- Region Map 9 types (rendering mode: 4Dshading)
- Absorption 10 types (rendering mode: 4Dshading)
- Hue 4 types (rendering mode: 4Dshading)
- Shadow Smooth 10 types (rendering mode: 4Dshading)
- Light Intensity 5 types (rendering mode: 4Dshading)
- Lighting Reso. 4 types (rendering mode: 4Dshading)
- Transp. Level 8 types (rendering mode: 4Dshading)
Color
- Combination B/W and CFM image combination / only B/W image / only CFM image
(only CFM image is impossible when rendering mode is Orthogonal)
- Blend 100 steps (except rendering mode: Orthogonal)
- Brightness
 3D image 17 steps (-8 ~ 8, except rendering mode: Orthogonal)
 MPR image 17 steps (-8 ~ 8) (except CFI mode)
- Shading 6 steps (rendering mode: opacity except CFI mode)
- Filter 5 steps (rendering mode: only opacity)
- Threshold 256 steps (rendering mode: only opacity )
- Smoothing 8 steps (rendering mode: only opacity)
- P-Filter 8 steps (rendering mode: only MIP)
 Operations

- Rotation By trackball and encoder


- View Line The cut-line of arbitrary curvature can be specified on 2D image
- Sliding, Plane cut Max. 14 planes by trackball (only in frozen, except rendering mode:
Orthogonal)

- 38 - SN-U391E
Image operations: both B/W and CFM images / only B/W image
- Cut function The arbitrarily specified part of 3D image can be dug (only in frozen, except
rendering mode: Orthogonal)
Image operations: both B/W and CFM images / only B/W image
- Zoom 1.0~4.0 (0.25 steps)
- Z Adjust 10 steps
- Image filing 3D volume data, Real-time 3D mode single images (such as BMP) and multi
images (such as AVI) can be saved
Saved images can be played back from the image filing screen, as with
regular 2D images
Image pre-cutted by ViewLine and full image can be switched
- Cine memory 3D images can be saved in the cine memory and played back
- PSS function Using the 3D PSS (3D Preset Scanning Selector) function
- Intelligent Surface Tracking
Set ViewLine automatically
- Volume Slice Imaging Display a thick cross-sectional image
From 0mm (OFF) to 20mm. 24 steps.
 Measurement Distance, area, circumference, volume etc. can be measured on 2D image,
MPR image or Multi Slice image.
 User-defined screen button The items of 3D screen button menu and image adjustment screen button
menu are registerable

3.2.19. Three-dimensional Display Software (STIC) (EZU-4DST2)


 Components Software (STIC software)
 Applicable probe Refer to Appendix
 Display mode Same as EZU-4D7 (except 2D-3D mode)
 Rendering mode Same as EZU-4D7
 Quality settings

- Scan time 4 steps (7.5, 10, 12.5, 15s)


- Color MAP Variance maps are available.
Same as EZU-4D7 excluding the above-mentioned
 Operations

- Display 3D Selectable display or hide 3D images


Same as EZU-4D7 excluding the above-mentioned
 Measurement Same as EZU-4D7
 User-defined screen button The items of image adjustment screen button menu and mode-specific screen
button menu are registerable
(Notice) : EZU-4D7 is required for EZU-4DST2.

- 39 - SN-U391E
3.2.20. Three-dimensional Display Software (Freehand) (EZU-3D8)
 Components Software (Freehand 3D software)
 Applicable probe Refer to Appendix
 Display mode

- ROI setting mode Displays the ROI that specifies the 3D image creation area on the 2D image
- 3D mode 3D image is displayed on 1 screen.
- 2D-3D mode Screen is vertically divided into two sides. The ROI setting mode is
displayed on the left side and the 3D mode on the right side
- MPR 4 screen mode Screen is vertically and horizontally divided into four. 3D image is displayed
on the lower right, while the random orthogonal 3 cross-section surface
images are displayed on the other sections
- MPR 1 screen mode One of the orthogonal 3 cross-section surface image in MPR 4 screen mode
is displayed on 1 screen
 Rendering mode

B/W
- Opacity
- MIP (Maximum Intensity Projection)
- mIP (minimum Intensity Projection)
- Transparency
- Inversion
- Orthogonal
- Cube
- Opac. Advanced
- 4Dshading
Color
- Opacity
- MIP (Maximum Intensity Projection)
- Transparency
- Orthogonal
- Cube
 Quality settings

B/W
- Brightness
 3D image 17 steps (-8 ~ 8, except rendering mode: Orthogonal and Cube)
 MPR image 17 steps (-8 ~ 8)
- Color
 3D color 37 types (except rendering mode: Orthogonal, Cube and, Opac.Advanced,
4Dshading)

- 40 - SN-U391E
 MPR color 37 types
- Shading 6 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
- Filter 20 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced,, or
4Dshading)
- Threshold 256 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced, or
4Dshading)
- Smoothing 8 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
- P-Filter 8 steps (rendering mode: MIP or mIP)
- BW 3D Smooth. 3 types (rendering mode: 4Dshading)
- 3D Smooth Coef 11 types (rendering mode: Opac.Advanced, or 4Dshading)
- Depth Contrast 40 types (rendering mode: Opac.Advanced)
- Depth Color Shift 127 types (rendering mode: Opac.Advanced)
- Region Map 9 types (rendering mode: 4Dshading)
- Absorption 10 types (rendering mode: 4Dshading)
- Hue 4 types (rendering mode: 4Dshading)
- Shadow Smooth 10 types (rendering mode: 4Dshading)
- Light Intensity 5 types (rendering mode: 4Dshading)
- Lighting Reso. 4 types (rendering mode: 4Dshading)
- Transp. Level 8 types (rendering mode: 4Dshading)
Color
- Combination B/W and CFM image combination / only B/W image / only CFM image
(only CFM image is impossible when rendering mode is Orthogonal or
Cube)
- Blend 100 steps (except rendering mode: Orthogonal and Cube)
- Brightness
 3D image 17 steps (-8 ~ 8, except rendering mode: Orthogonal and Cube)
 MPR image 17 steps (-8 ~ 8) (except CFI mode)
- Shading 6 steps (rendering mode: opacity except CFI mode)
- Filter 5 steps (rendering mode: only opacity)
- Threshold 256 steps (rendering mode: only opacity )
- Smoothing 8 steps (rendering mode: only opacity)
- P-Filter 8 steps (rendering mode: only MIP)
 Operations

- Rotation By trackball and encoder


- View Line The cut-line of arbitrary curvature can be specified on 2D image
- Sliding, Plane cut Max. 14 planes by trackball (only in frozen, except rendering mode:
Orthogonal)
Image operations: both B/W and CFM images / only B/W image (both B/W
and CFM image is only possible when rendering mode is Cube)
- Cut function The arbitrarily specified part of 3D image can be dug (only in frozen, except

- 41 - SN-U391E
rendering mode: Orthogonal and Cube)
Image operations: both B/W and CFM images / only B/W image
- Zoom 1.0~4.0 (0.25 steps)
- Image filing 3D volume data, Real-time 3D mode single images (such as BMP) and multi
images (such as AVI) can be saved
Saved images can be played back from the image filing screen, as with
regular 2D images
Image pre-cutted by ViewLine and full image can be switched
- PSS function Using the 3D PSS (3D Preset Scanning Selector) function
 Measurement Not applicable
 User-defined screen button The items of 3D screen button menu and image adjustment screen button
menu are registerable

3.2.21. Remote Control Unit (EZU-RH4)


 Components Remote control, battery charger
 Size 185 (H) × 76 (W) × 22.5 (T) mm
 Weight 220 g
 Continuous use time Approx.5 hours (on a full charge)
 Charging time Approx. 3 hours
 Battery Lithium-ion battery
 Transmission method Infrared communication
 Coverage range Within 3 m
 Waterproofing Applicable
 Plasma sterilization Applicable
 Remote control function B/CFM/PW mode changes, B/CFM/PW gain adjustment, REF changes,
focus position changes, PRF changes, depth changes, freeze ON/OFF, caliper,
Single Image recording, movie recording, print out, orientation, trackball
priority, display changes,Picture in Picture settings, Remote user key1,
Remote user key2 (User defined)

(Notice) : EZU-RiF5 is required for EZU-RH4.


(Notice) : EZU-PP5 is required for “Picture in Picture display change.”

3.2.22. Remote Control Interface Unit (EZU-RiF5)


This is the optical receiver installed in the ultrasound diagnostic scanner (main unit) for making to work
remote control unit EZU-RH4.
 Components KBIF PCB

3.2.23. Picture in Picture Software (EZU-PP5)


 Components Software

- 42 - SN-U391E
 Input signal Y/C
 Applicable probes All probes
 Display mode

- Full screen display


- Dual screen display Probe : FG series, EG series, EB series, EUP-OL531
Mode : B, CFM, Elasto
- Small screen display
 Display position

- Dual screen display L/R preset can be possible


- Small screen display Four corners preset can be possible, display position can be changed in
real-time
 Display ON/OFF Applicable
 Cine playback Applicable only US images
 Movie recording Applicable

(Notice) : EZU-ACP2 and EZU-ViF2 are required for EZU-PP5.

3.2.24. Network Interface Unit (EZU-NW1)


 Components LAN cable
 Specifications Ethernet 10/100 Base-T (IEEE802.3)

3.2.25. Analog Capture Card Unit (EZU-ACP2)


 Components Analog capture card, cables, Driver software

3.2.26. Digital Capture Card Unit (EZU-DCP3)


 Components Digital capture card, cables, Driver software

3.2.27. Video Interface Unit (EZU-ViF2)


 Components TVIF unit, cables
 Specifications RGC output
Y/C input and output (Only connector for input)
Brightness signal
Remote control output for BE printer
Audio input and output

3.2.28. Expanded USB Unit (EZU-UP2)


 Components Expanded USB Unit
 Specifications

- Number of USB ports 2 ports

- 43 - SN-U391E
3.2.29. External LCD Monitor Unit (EZU-EL2)
 Components DVI Connector Unit
 Specifications

- Video signal DVI-D


- Resolution SXGA(1280x1024)

3.2.30. Jelly Warmer Unit (EZU-JW1)


 Components Jelly Warmer, Power cable

3.2.31. Alpha numeric keyboard Unit (EZU-AK1E)


 Components Alphanumeric keyboard unit, Lock screw, Hexagon socket head bolt M6X30

3.2.32. High Neck Unit (EZU-HN1)


 Components High Neck Unit

3.2.33. Probe Hanger (EZU-TH7)


 Components Probe Hanger

3.2.34. Processor Tray (EZU-TY3)


 Components Processor tray
(Notice): EZU-HN1 is required for EZU-TY3.
(Notice): When you install EZU-TY3 (processor tray) onto the equipment, peripherals other than the
video processor made by HOYA cannot be installed.

3.2.35. Junction Box (PUN-JBP1)


 Components Junction box
 Outside dimension 150 (W) × 118.5 (D) × 110.5 (H) mm
 Weight Approx. 900 g

3.3. Power requirements


 Input voltage 100/110 120, 220~240VAC ± 10%
 Frequency 50 / 60 Hz
 Auxiliary AC outlet 350VA (peripheral devices)
 Consumption Max. 1.5kVA (include 350VA for AC outlet)
 Outlet The 3-wire power outlet which has a protective earth terminal (with less than
10Ω grounding resistance) of the grounding facility.

- 44 - SN-U391E
3.4. Environment conditions
 Temperature 10 ~ 35°C
 Relative humidity 30 ~ 85% RH (No dew condensation)
 Atmospheric pressure 700 ~ 1060hP

3.5. Exterior view


 Outside dimension
- 450 (W) x 700 (D) (with EZU-MT29-S1)
x 1350 (H) mm
- 450 (W) x 700 (D) (with EZU-MT29-S1 and EZU-HN1)
x 1430 (H) mm
- 450 (W) x 700 (D) (with EZU-MT30-S1)
x 1400 (H) mm
- 450 (W) x 700 (D) (with EZU-MT30-S1 and EZU-HN1)
x 1480 (H) mm
 Weight
- Approx. 125kg (standard configuration with EZU-MT29-S1)
- Approx. 127kg (standard configuration with EZU-MT29-S1 and EZU-HN1)
- Approx. 130kg (standard configuration with EZU-MT30-S1)
- Approx. 132kg (standard configuration with EZU-MT30-S1 and EZU-HN1)
 Monitor 17 inch digital LCD monitor
1280 x 1024 Non interlace
 LCD Monitor Arm
- simple arm Angle: adjustable 2 directions
- flexible arm Height: adjustable approx. 18cm
Angle: adjustable 4 directions
 Caster Swivel・lock /brake
 Probe connector 4 probes (3 probes available, 1 probe rest)
 Jelly wormer Applicable (Option)
 Handle for movement Rear side
.

- 45 - SN-U391E
1350
820

450 700

Body dimension (with EZU-MT29-S1)

1430
900

450 700
Unit: mm
Body dimension (with EZU-MT29-S1 and EZU-HN1)

- 46 - SN-U391E
1400
820

450 700

Body dimension (with EZU-MT30-S1)

1480
900

450 700
Unit: mm
Body dimension (with EZU-MT30-S1 and EZU-HN1)