Specifications
Hitachi, Ltd.
SN-U391E
Trademarks or the registered trademarks
The following is the trademarks or the registered trademarks in this manual or in this equipment.
HI VISION Avius, Avius, Real-time Tissue Elastography, HdTHI, HI REZ, Ultra BE/Ultrasound
Broadband Engine, HINT-NAVI, Tool Bar, Fine Flow are registered trademarks of Hitachi in Japan.
HI VISION is registered trademarks of Hitachi in the U.S.A. and Europe.
Windows, Windows XP, Windows Embedded Standard 2009, Windows Vista, Windows 7, Windows 8,
Windows 8.1, Windows NT, Windows 2000, Windows 98, Microsoft Excel, Office, PowerPoint, Windows
Media, DirectX are trademarks or the registered trademarks of Microsoft Corporation in the U.S.A. and
other countries.
Linux® is a trademark registered to Linus Torvalds.
CinePlayer Decoder Pack is used in Digital Video Archiving Software (Option). CinePlayer Decoder Pack
is trademark of Sonic Solutions.
In this manual or in this equipment, other brand names and product names are trademarks or registered
trademarks of their respective companies.
Abbreviations
No part of this publication may be transmitted into without the prior written
permission of HITACHI.
This publication is subject to change without notice.
Although HITACHI has attempted to ensure accuracy throughout this publication, HITACHI
welcomes customer input on corrections and suggestions for improvements to this publication.
SN-U391E
Contents
1. General Description ...................................................................................................... 1
1.1. General ........................................................................................................................................ 1
1.2. Features ...................................................................................................................................... 1
2. System Composition ..................................................................................................... 6
2.1. Main System Composition ........................................................................................................ 6
2.1.1. Ultrasound Scanner Main Unit .......................................................................................... 6
2.1.2. Accessories ........................................................................................................................... 6
2.2. Probes ......................................................................................................................................... 6
2.2.1. Convex array probes ............................................................................................................ 6
2.2.2. Linear array probes ............................................................................................................. 7
2.2.3. Phased array probes ............................................................................................................ 7
2.2.4. Other probes ........................................................................................................................ 8
2.3. Options ....................................................................................................................................... 8
2.4. Peripheral Devices ................................................................................................................... 10
3. Specifications .............................................................................................................. 11
3.1. General ...................................................................................................................................... 11
3.1.1. System ................................................................................................................................ 11
3.1.2. Basic Function ................................................................................................................... 11
3.2. Options ..................................................................................................................................... 22
3.2.1. 17-inch Digital LCD Monitor (EZU-MT29-S1) ................................................................. 22
3.2.2. 17-inch Digital LCD Monitor (Flexible arm) (EZU-MT30-S1) ........................................ 22
3.2.3. Biomedical Waveform Display Unit (EZU-EK28) ............................................................ 22
3.2.4. CW Doppler Unit (EZU-ST8) ............................................................................................ 22
3.2.5. Real-time Tissue Elastography Unit (EZU-TE6) ............................................................. 23
3.2.6. Real-time Tissue Elastography software (Strain Histogram) (EZU-TESH1) ................. 24
3.2.7. Digital Video Archiving Software (EZU-DV4) .................................................................. 25
3.2.8. Logon Management Software (EZU-LG2) ........................................................................ 26
3.2.9. DICOM Software(Transfer and Media Storage) (EZU-FC11) ......................................... 27
3.2.10. DICOM Software (WorkList) (EZU-FC11W) / DICOM Software (Printer) (EZU-FC11P)31
3.2.11. DICOM Software (Query / Retrieve) (EZU-FC11Q) ......................................................... 31
3.2.12. Omni-directional M mode display function (EZU-OD3) .................................................. 32
3.2.13. Advanced Measurement Software (EZU-AM7) ................................................................ 32
3.2.14. IMT Measurement Software (EZU-IM2) .......................................................................... 33
3.2.15. Software Specified for Contrast Agents (EZU-CH8) ........................................................ 33
3.2.16. WideView Software (EZU-WS8) ........................................................................................ 35
3.2.17. Stress Echo Software (EZU-SE6)...................................................................................... 36
3.2.18. Three-dimensional display unit (Real-Time) (EZU-4D7)................................................. 36
3.2.19. Three-dimensional Display Software (STIC) (EZU-4DST2) ............................................ 39
<i> SN-U391E
3.2.20. Three-dimensional Display Software (Freehand) (EZU-3D8) ......................................... 40
3.2.21. Remote Control Unit (EZU-RH4) ..................................................................................... 42
3.2.22. Remote Control Interface Unit (EZU-RiF5) ..................................................................... 42
3.2.23. Picture in Picture Software (EZU-PP5) ............................................................................ 42
3.2.24. Network Interface Unit (EZU-NW1) ................................................................................ 43
3.2.25. Analog Capture Card Unit (EZU-ACP2) .......................................................................... 43
3.2.26. Digital Capture Card Unit (EZU-DCP3) .......................................................................... 43
3.2.27. Video Interface Unit (EZU-ViF2) ...................................................................................... 43
3.2.28. Expanded USB Unit (EZU-UP2) ...................................................................................... 43
3.2.29. External LCD Monitor Unit (EZU-EL2)........................................................................... 44
3.2.30. Jelly Warmer Unit (EZU-JW1) ......................................................................................... 44
3.2.31. Alpha numeric keyboard Unit (EZU-AK1E) .................................................................... 44
3.2.32. High Neck Unit (EZU-HN1) ............................................................................................. 44
3.2.33. Probe Hanger (EZU-TH7) ................................................................................................. 44
3.2.34. Processor Tray (EZU-TY3) ................................................................................................ 44
3.2.35. Junction Box (PUN-JBP1) ................................................................................................ 44
3.3. Power requirements ................................................................................................................. 44
3.4. Environment conditions .......................................................................................................... 45
3.5. Exterior view ............................................................................................................................. 45
<ii> SN-U391E
1. General Description
1.1. General
Model HI VISION Avius is a multi-functional ultrasound diagnostic scanner in which Doppler, ColorFlow
Mapping, etc. are provided and all circuits related to image quality are fully digitalized.
As not only linear and convex scan type probes, but also phased array scan type probe can be operated with
HI VISION Avius, it is possible to use it for any field of application.
Furthermore, High-definition dTHI (HdTHI) mode (Hitachi unique technology) are provided in standard
function.
This mode provide “Clear image” to reduce artifact such as vibration echo, and to improve spatial resolution
and contrast resolution.
HI Com function, HI REZ+ function, Trapezoidal display function, Oblique B-mode image function,
real-time Doppler trace, cine memory, patient data management function, and Real-time Archiving function
can be used on the standard model.
Wide view software, Stress echo software, Software specified Contrast Agents, cardiac measurement
software, Real-time Tissue Elastography, omni-directional M (ODM) mode, Remote control unit, Picture in
Picture software can be used optional function.
It is convenience for user to adopt Ergonomic design and 17-inch digital LCD Monitor and more improved
High-definition screen-displayed format.
1.2. Features
True digital architecture
Due to beamformer which is decisively concerned to the image quality and all image processing are
digitalized, high spatial resolution, wide dynamic range and high contrast resolution are realized.
Ultrasound transmit circuit is improved by high performance transmit amplifier, and ultrasound receiving
circuit is improved by high performance pre-amplifier.
Frame rate is improved much more faster comparing with conventional system.
High quality image is digital image processing application.
This architecture that is achieved high quality image is good for new generation.
Quick response
Mode changing, image condition settings, probe changing and application selection can be performed
quickly so that operator can pay attention to the patients.
-1- SN-U391E
17-inch digital LCD Monitor
It is adopted 17-inch digital LCD Monitor. LCD Monitor provides high quality image.
The LCD simple arm that optimizes the position of the LCD monitor and the operational panel provides ease
of use. The LCD flexible arm that can adjust height and four-way position provides freely position of
monitor.
Ergonomic design
- Design
Design is adopted to focus on user’s “easiness to use”, like front layout of probe connector, based on
ergonomics.
- Easily operable keyboard
As key arrangement is taken in consideration of frequency of use and operational has improved. An
optional alphanumeric keyboard that is adopted backlight keyboard improves key response and
visibility.
- Excellent mobility with four swivel casters
As all the four casters are of swivel type, it is easy to move in narrow space.
Due to reduced friction force to floor it is easy to move from ward to ward.
- Probe rest
It connects 3 probes simultaneously and has 1 dummy connector.
Use interface
Workflow is improved because of new GUI.
It is a few procedures for many functions to use screen buttons.
Online Manual
It is applicable to browse instruction manual on your HI VISION Avius.
-2- SN-U391E
Various B-mode scan
- HI Com function
Compound function provides high spatial resolution, wide dynamic range and high contrast resolution.
- Trapezoidal display function
Trapezoidal display is applicable for linear probe.
It is improved diagnosis area for linear probe.
- Oblique B-mode image function
This function enables you to use oblique function not only CFI/CFA mode but B mode.
-3- SN-U391E
Filing function.
- Examination data management
Examination data (images measurement results and other data) can manage and operate together, so it is
easy to refer data before an examination and after an examination.
- Multiple functions to save data
No image degradation because of digital image recording.
This filing function can record still images and multi images not only HDD include system) but also
DVD-RAM, DVD-R, USB memory and USB HDD.
- Easy-to-use filing function
In the Filing window, you can view examination data.
It is consideration for easy-to-use in the routine examination by filtering Patient ID or Patient name or
keyword search.
Thumbnail view provides easy access to destination image.
Review images can display 1/4/9/16 screen depend on choosing images.
- Patient data management
You can select the patient in the list view who had examination once.
-4- SN-U391E
DICOM function
This software can transfer ultrasound images and measurement reports to a storage server.
It has a Storage Commitment function for ultrasound images.
This software can output and read ultrasound images in DICOM file format.
This software has a function to obtain WorkLists from the MWM server.
It has a function to send examination start and completion notifications for ultrasound images to the MPPS
server.
And it can be sent Ultrasound images to grayscale or color DICOM printers.
4D function
3D ultrasound images can be viewed in almost real-time. As the scanning part of the probe swings
automatically in a fan-like fashion, it sends signals to the three-dimensional display unit, which rapidly
builds a 3D image based on the probe’s signal.
STIC function
Multiple 3D images representing a single heartbeat can be displayed by automatically swinging the
scanning part of the probe in a fan-like fashion and estimating the cardiac cycle from the obtained images.
3D function
The software is able to collect sequential 2D image data by using a probe to scan at a steady rate and then
display the collected data as 3D images.
-5- SN-U391E
2. System Composition
2.1. Main System Composition
2.1.1. Ultrasound Scanner Main Unit
Ultrasound transmitting / receiving unit
Digital scan converter unit
Pulsed Doppler unit
Color flow mapping unit
HI Com function
Adaptive Imaging function (HI REZ+)
TDI (Tissue Doppler imaging) function
Filing/Patient data management function
Tissue harmonic imaging function
Cine-memory unit
Automatic image optimization (HI Support)
High density zoom function (HI Zoom)
Power supply unit
DVD (DVD-RAM, DVD-R) drive
2.1.2. Accessories
System CD 2
Hexagonal wrench bar 1 set
Spare fuses 1 set
Cords 1 set
Instruction manual (Include CD) 1 set
DVD-RAM (4.7GB) 1
2.2. Probes
2.2.1. Convex array probes
4-2MHz 10R EUP-C511
5-2 MHz 40R EUP-C514
5-1 MHz 50R EUP-C715
5-2 MHz 60R EUP-C516
6-3 MHz 40R EUP-C524
5-2 MHz 40R (Real-time 3D) EUP-CV714
7-2MHz 40R (Real-time 3D) EUP-CV724
6-3 MHz 40R (Real-time 3D) EUP-CV524
8-4 MHz 20R EUP-C532
-6- SN-U391E
5-2 MHz 20R (Biopsy) EUP-B512
6-1 MHz 20R (Biopsy) EUP-B712
5-2 MHz 40R (Biopsy) EUP-B514
5-1MHz 50R(Biopsy) EUP-B715
8-4 MHz 10R (Transvaginal wide view) EUP-V53W
10-2 MHz 10R (Transvaginal wide view) EUP-V73W
8-4MHz 10R (Transvaginal, Real-time 3D) EUP-VV531
8-4 MHz 10R (Transvaginal, Real-time 3D) EUP-VV731
8-4 MHz 10R Convex / 8-4 MHz 10R Convex (Intracavital) EUP-CC531
8-4 MHz 10R Convex / 8-4 MHz 10R Convex (Intracavital) EUP-CC531S
8-4 MHz 10R Convex / 10-5 MHz Linear (Intracavital) EUP-U533
8-4 MHz 10R (Transrectal) EUP-U531
8-4 MHz 10R (Finger top) EUP-F531
7.5 MHz 40R (Finger top) EUP-F334
10-5 MHz 10R (Laparoscopic) EUP-OL531
7.5 MHz 40R (Laparoscopic) EUP-OL334
10-4MHz 20R (Intraoperative) EUP-O732T
10-5 MHz 6R (Transrectal, 19cm, 360°) EUP-R54AW-19
10-5 MHz 6R (Transrectal, 33cm, 360°) EUP-R54AW-33
-7- SN-U391E
2.2.4. Other probes
7.5 MHz 10R (Upper G.I. Fiberscope, made by HOYA) FG-34UX
7.5 MHz 10R (Upper G.I. Fiberscope, made by HOYA) FG-36UX
7.5 MHz 10R (Upper G.I. Fiberscope, made by HOYA) FG-38UX
7.5 MHz 10R (Upper G.I. Videoscope, made by HOYA) EG-3630U
10-5MHz 6R (Upper G.I. Videoscope, made by HOYA) EG-3270UK
10-5 MHz 6R (Upper G.I. Videoscope,made by HOYA) EG-3630UR
10-5 MHz 10R (Upper G.I. Videoscope, made by HOYA) EG-3830UT
10-5 MHz 6R (Upper G.I. Videoscope, made by HOYA) EG-3670URK
10-5 MHz 10R (Upper G.I. Videoscope, made by HOYA) EG-3870UTK
10-5 MHz 6R (Videobronchoscope, made by HOYA) EB-1970UK
10-5 MHz 6R (Videobronchoscope, made by HOYA) EB19-J10U
2.3. Options
17-inch Digital LCD Monitor Unit EZU-MT29-S1
17-inch Digital LCD Monitor Unit (Flexible arm) EZU-MT30-S1
Biomedical Waveform Display Unit EZU-EK28
CW Doppler Unit EZU-ST8
Software Specified for Contrast Agents EZU-CH8
DICOM Software (Transfer and Media Storage) EZU-FC11
DICOM Software (Printer) EZU-FC11P
DICOM Software (Worklist) EZU-FC11W
DICOM Query/Retrieve Software EZU-FC11Q
Three-dimensional display unit (Real-Time) EZU-4D7
Three-dimensional Display Software (STIC) EZU-4DST2
Three-dimensional Display Software (Freehand) EZU-3D8
Real-time Tissue Elastography Software EZU-TE6
Real Time Tissue Elastography Software (Strain Histogram) EZU-TESH1
ODM Software EZU-OD3
Advanced Measurement Software EZU-AM7
IMT Measurement Software EZU-IM2
WideView Software EZU-WS8
Stress Echo Software EZU-SE6
Digital Video Archiving Software EZU-DV4
Remote Control Unit EZU-RH4
Remote Control I/F Unit EZU-RiF5
Picture in Picture Software EZU-PP5
-8- SN-U391E
Logon Management Software EZU-LG2
Alpha numeric keyboard Unit EZU-AK1E/F
Network Interface Unit EZU-NW1
Digital Capture Card Unit EZU-DCP3
Analog Capture Card Unit EZU-ACP2
Video Interface Unit EZU-ViF2
Foot Switch (for Freeze) EZU-FS1
Foot Switch (for Freeze / User Define) EZU-FS2
Expanded USB Unit EZU-UP2
External LCD Monitor Unit EZU-EL2
Jelly Warmer Unit EZU-JW1
High Neck Unit EZU-HN1
Probe Hanger EZU-TH7
Processor Tray EZU-TY3
5MHz Pencil CW Probe Adapter EZU-AC1
- 5MHz Pencil CW Probe
Sonde Crayon 5MHz P050CN20W00+AD (with BNC cabling)
(by AXON’CABLE S.A.S)
10MHz Pencil CW Probe Adapter EZU-AC2
- 10MHz Pencil CW Probe
Sonde Crayon 10MHz 10D4 (with BNC cabling)
(by AXON’CABLE S.A.S)
5MHz/10MHz pencil CW probe Interface EZU-AC3
Junction Box PUN-JBP1
-9- SN-U391E
2.4. Peripheral Devices
Peripheral device Manufacturer Manufacturer’s model, type
Black & white printer SONY UP-D897
UP-897MD (Notice II)
UP-X898MD
MITSUBISHI P93DW
P93E/W (Notice II)
P95DE/DW
Color printer SONY UP-D23MD
UP-D25MD
MITSUBISHI CP30DW
DVD Video Recorder SONY DVO-1000MD (Notice I)
PC printer HP HP Photosmart D5460 *
HP OFFICEJET 6000 *
Wireless LAN adapter CISCO Aironet 1130AG *
* These devices meet IEC60950 compliance.
(Notice I): EZU-ViF2 is required to record. EZU-ViF2 and EZU-ACP2 is required to playback.
(Notice II): EZU-ViF2 is required to print.
(NoticeⅢ): It is necessary to detach a rear handle to carry out simultaneous loading of a color printer and a
DVD video recorder.
- 10 - SN-U391E
3. Specifications
3.1. General
3.1.1. System
OS Windows Embedded Standard 2009
DVD DVD-RAM DVD-R
USB memory USB HDD Connectable
Network 10/100 BaseT (EZU-NW1 is required to network connection)
Starting time About 100 seconds (About 70 seconds in Hibernation mode)
3.1.2.4 Focus
Transmission Max 10 steps (applicable to the setting of Focus width at multiple focus)
- 11 - SN-U391E
Reception Continuous dynamic focusing with variable aperture
Other 2-dimentional apodization
Dual image Different probe and different depth can be displayed. Up-and-down can be selected for
linear probe.
- 12 - SN-U391E
(Notice) Biopsy guideline can not be displayed, when an application that image split setting of B-mode is
up-and-down is selected.
3.1.2.16 Orientation
Flip vertical and horizontal B mode, Doppler mode
Flip vertical B mode: 90 deg steps rotation
- 13 - SN-U391E
3.1.2.19 Doppler function
Display mode B/D (B-Freeze, B/D real), D
Display area Side-by-side or up-and-down (can be changed in real-time and frozen),
split ratio: 1:1 (side-by-side), 1:1, 1:2, 2:1 (up-and-down)
Analysis FFT method, 3 steps, max. 128 points
Pulse repetition frequency (PRF)
1, 1.3, 1.5, 1.8, 2, 2.2, 2.5, 3, 3.5, 4, 4.5, 5, 5.5, 6, 6.5, 7, 7.5, 8, 9, 10, 11, 12,
14, 16, 18, 20 [kHz], total 26 steps
Reference frequency (REF) 3 frequencies: (Depend on probes)
Wall filter 12 steps (maximum value : 1/16 of PRF)
Baseline shift Applicable (in real-time and frozen)
Detectable depth 0mm ~360 mm
Sample length 0.5 mm ~ 5.0 mm: 0.5 mm step,
5.0 mm ~ 15.0 mm: 1.0 mm step
Measurement
- Real-time Velocity
- Frozen Velocity, Acceleration, Index for stenosis, Pressure gradient, Flow volume,
Time average velocity, Ratio-time, Ratio-velocity, Average velocity, PHT
(Pressure Half Time), HR (Heart Rate), TVI (Time Velocity Integral), Auto
trace is available
- Limitations HR: 25 ~ 300, measurement is not applicable in cine playback.
Audio input / output 2ch speaker mounted (for forward / reverse flow), VCR audio input / output
(for forward / reverse flow)
Maximum detectable velocity
- 14 - SN-U391E
- Down trace Trace bottom the baseline
- Both trace comparing for peak of both upper and bottom baseline, and trace either one
higher
- Display method
Real-time Moving bar
Cine playback Scroll
- Sweep speed 1, 1.5, 2, 3, 4, 6, 8, 10 s /screen (can be changed in real-time and frozen)
Color display
120, 250, 350, 400, 500, 650, 800, 1k, 1.3k, 1.5k, 2k, 2.5k, 3k, 3.5k, 4k, 4.5k,
5k, 6k, 8k, 10k, 12k, 16k, 20k [Hz], total 23 steps
Persistence 8 steps
Motion artifact suppression OFF + 3 steps
Scan line density 8 steps
- 15 - SN-U391E
Color preference Selectable 2 modes
- Auto Brighter signal is displayed between B and color
- Manual (16 steps) B image is displayed where B image intensity is bigger than preference bar
level on the gray scale.
Peak hold [Not ECG display]
Overlay color data in real-time, Refresh time: 1s, 2s, 4s, 8s, ∞
[ECG display]
Overlay color data in real-time, Refresh time: 1beat, 2beat, 4beat, 8beat, ∞
Velocity tagging Applicable
Color filter 3 steps
Interleave Max.32 lines (depending on PRF)
Packet size 3 steps
CFA dynamic range 8 steps (available to Fine flow)
Reference frequency (REF) Max.3 frequencies (Depend on probes)
- 16 - SN-U391E
Automatic optimization parameter
- 17 - SN-U391E
B-image Frame storage capacity
Applicable (M/D)
3.1.2.28 Preset
Application Max 22 sets of operating parameters can be registered for each probes
Max 10 set of user registered
PSS function Registered for each application
Max 8 set of preset
Region Data Setting Max 37 set of region data
Max 30 set of user registered
Data and time (year, month, day, hour, minute, and second), Physician, age,
sex, Probe name, frequency of the probes, gain settings, acoustic power level,
post-processing parameters, dynamic range (DYN), delay times from
R-wave, heart rate, measurement values, Doppler function display, focus
point, view of depths display.
Annotation Preregistered 402 annotations (abbreviated medical terms)
- 18 - SN-U391E
3.1.2.30 Body marks
Display position Adjustable
Display size Large or Middle
Types
- Abdomen 8 types
- Breast 10 types
- Head/Neck 15 types
- Urology 10 types
- Cardiovascular 14 types
- Organ 11 types
- Upper limb 18 types
- Lower limb 10 types
- Obstetrics 21 types
- Gynecology 4 types
- 19 - SN-U391E
items, measurement orders, and formulas.
Applicable
⁃ Copy destination USB memory, USB HDD, DVD-RAM, Network folder
- Frame to Frame (F-F) All image or 10, 20, 30, 40, 50, 60, 70, 80, 90, 100 images from last image
- R-Wave to R-Wave (R-R)
All image or 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 beats from last image
(Mounted a Biomedical Waveform Display Unit)
- All
(Notice) : When the number of images in cine memory exceeds the record upper limit, all image
transfer is not available.
Image format BMP (Bitmap), TIFF, AVI, AVI (Compression), JPEG, DICOM,
DICOM JPEG
TIFF compression format None, Pack Bits
AVI compress Microsoft Video 1
Number of retrieve images 1, 4, 9, 16
Retrieve mode Frame by frame review Continuous review, Changing the time axis
Retrieve speed -1/1, -1/2, -1/4, -1/8, 0, 1/8, 1/4, 1/2, 1/1
Divided frame display Available with motion image review
Review range specification Applicable
Playback measurement Applicable excluding D-Profile, CFM-Velo, histogram, auto trace and PISA,
MVG-M, vascularity in MV Regurg and TV Regurg
Filing with Each image, Each patient
Masking of ID and patient name
- 20 - SN-U391E
Patient information window
- Keyword search Today's Study, Search All, Patient ID, Patient name, Study Date, Study Term,
Body Part Exam., Examined by, Ref. Physician, Additional Patient History,
Application, Comment, Transfer Status
- Thumbnail images sorting
Applicable
- Image Data copy, delete applicable
- Correction of the patient information for the saved examination data like an image.
Not applicable
(Notice) : Other patients’ examination data cannot be referred to during examination.
(Notice) : To record data onto media that was created by using equipment that uses different software
version, select the target location list in the Filing window, and then update the database before
recording the data.
OS of the network computer which can perform file sharing.
Windows XP Professional,
Windows Vista Ultimate/Home Premium
Windows 7 Ultimate/Professional/Home Premium
Windows 8/Windows 8 Pro
Windows 8.1/Windows 8.1 Pro
(Notice) : The simultaneous reference of network folder cannot be performed from two or more system.
(Notice) : File sharing is available when using a password-protected account.
3.1.2.35 Others
Brightness Setting Character 6 level
Background 6 level
Screensaver Applicable
Online Manual Applicable
Screen Keyboard Enter alphanumeric characters and symbols
(If the optional alphanumeric keyboard is not mounted)
- 21 - SN-U391E
3.2. Options
3.2.1. 17-inch Digital LCD Monitor (EZU-MT29-S1)
Standard component LCD monitor, LCD simple arm, cables
Screen 17 inch, Non-interlace
Input voltage DC12V
Power consumption Max.34VA
Video signal DVI-D
- 22 - SN-U391E
- EUP-ES52M 3.5 [MHz]
FFT pulse repetition frequency (PRF)
4, 4.5, 5, 6, 8, 10, 12, 16, 20, 25, 35, 50 [kHz], total 12 steps
Analysis FFT method, 3Steps, Max. 256point
Maximum detectable velocity
- 23 - SN-U391E
Rejection 65 steps
Blending 0% ~ 100%, 2% steps
Blend Type 2 types
Frame rate 5steps
Dual mode Applicable
Biplane mode Applicable (EUP-CC531S)
Biopsy guide line Applicable
Press Guide function
- Strain Graph Displays the variation over time of tissue strain caused by compression.
Available graph to exclude low strain area.
- Press Indicator Display the compression status
Frame select function Possible to select images under stable compression automatically
Strain ratio measurement
- Number of ROI Max. 10, possible to position the ROIs on B mode image
- ROI shape Circle, oval, rectangle and closed spline curve
- Averaging function Max. 5 frames
- Auto ROI Setting function
Clicking center of the tumor automatically sets the measurement ROI to the
tumor and the fatty layer (the mammary gland region and lesion are targeted).
- Measurements of retrieved images Applicable
- 24 - SN-U391E
3.2.7. Digital Video Archiving Software (EZU-DV4)
Components Software
- 25 - SN-U391E
(Notice) : When the number of images in cine memory exceeds the record upper limit, all image transfer is
not available.
(Notice) : Quality of AVI(uncompressed) or DICOM format and AVI(Motion JPEG Compression) or
DICOM JPEG format are same in real-time.
Applicable
Playback speed
- Play 1×
- Fast-forward 2×, 8×, 32×
- Fast-rewind -1×, -2×, -8×, -32×
- Slow-forward 1/8×, 1/4×, 1/2×
- Slow-rewind -1/8×, -1/4×, -1/2×
Playback to the start by chapter
Applicable
Playback function Only MPEG2 clip file that is captured this software
(Notice) : EZU-DCP3 is required for EZU-DV4.
- Service 1 (undeletable)
- Administrator max. 30
- General user max. 30
User licensing function
- 26 - SN-U391E
- Display of information Displaying a list of account
- Registration function New registration of account
- Delete function Deregistration of account
Logoff function
- 27 - SN-U391E
DICOM log acquisition Applicable (DICOM log in communication)
SR transfer operation DICOM SR is transferred by pushing "SR Transfer" button in the Filing
Window, "Save" button in the Report Winodw, Patient Key or End Exam
key.
Output measurement item to DICOM SR
- 28 - SN-U391E
L-Ovary measurement: L, AP, W, V
R-Ovary measurement: L,AP,W,V
- Follicular measurement
L-Ovary measurement: L, AP, W, V
R-Ovary measurement: L, AP, W, V
L-Follicular measurement :L,AP,W,V
R-Follicular measurement :L,AP,W,V
Uterus measurement: Uterus L, Uterus AP, Uterus W, V,
Endomet. T., Cervix L
- Vascular measurement Extracranial Arteries measurement: Bulb, CCA, ECA, ICA, PROX, ICA
MID, ICA DIST, ICA(RATIO), Subclav.A, Vertebral A
Lower Extremity Arteries measurement: CIA, ATA, CFA, DPA, EIA, IIA,
Pero.A, Pop.A, PTA, PFA, SFA
Lower Extremity Veins measurement: ATV, CFV, CIV, EIV, GV, GSV, LSV,
Pero.V, Pop.V, PTV, PFV, SFJ, SV, SFV, SPJ, Hunter, Cocket, Boyd
- Cardiac measurement (Cardiac_SR)
Left Ventricle measurement: HR, LVEdD, LVEsD, FS, LVEdV, LVEsV, EF,
Regurg Vol., Flow Rate, PISA, PISA-R, Vp, TVI, IVRT, e', E/e', A, S, SV, CO,
CI, IVSd, LVPWs, LVPWd, LVMass-d, IVSd/LVPWd
Right Ventricle measurement: HR, Regurg Vol., Flow Rate, PISA, PISA-R,
Vp, TVI, SV, CO, CI
Aortic Valve measurement: Avcusp Open, ET, Ao/ET, Regurg Vol., Flow
Rate, PISA, PISA-R, Vp, TVI
Mitral Valve measurement: A, E, E/A, EFslope, EPSS, Adur, Regurg Vol.,
Flow Rate, PISA, PISA-R, Vp, TVI, SV, CO, CI, dp/dt
Pulmonic Valve measurement: Regurg Vol., Flow Rate, PISA, PISA-R, Vp,
TVI
Tricuspid Valve measurement: Regurg Vol., Flow Rate, PISA, PISA-R, Vp,
TVI
Pulmonary Veins measurement: S, D, A, Adur, HR
Congenital measurement: HR, Regurg Vol., Flow Rate, PISA, PISA-R, Vp,
TVI, Qp/Os
- Cardiac measurement (Cardiac_SR_II)
Echo LV measurement : B-LV, M-LV, M.Simpson, Dia.Func,
Dop Index, CO, BI-Plane, Area Length, LVMass-AL,
LVMass-TE, HR
Echo LA measurement : B-LA, M-LA, TEE-LA
Echo RV measurement : CO, B-RV, M-RV, HR
Echo RA measurement : B-RA
Echo AV measurement : AV Regurg, Velocity, PHT, D-Trace,
- 29 - SN-U391E
B-Area, Area-Vp, Area-Vm, Area-VTI, HR
Echo MV measurement : Dia.Func, dP/dt, MV Regurg, Velocity,
PHT, D-Trace, B-Area, Area-Vp, Area-Vm, Area-VTI, MV-RegVol, HR
Echo PV measurement : PV Regurg, Velocity, PHT, D-Trace, HR
Echo TV measurement : TV Regurg, Velocity, PHT, D-Trace, HR
Echo Ao measurement : B Ao, Ao, HR
Echo PArtery measurement : B-PArtery, D-PArtery, HR
Echo IVC measurement : B-IVC, HR
Echo PVein measurement : D-PVein, HR
Echo CS measurement : Qp/Qs, HR
Echo Cong measurement : Cong Regurg, HR
- 30 - SN-U391E
3.2.10. DICOM Software (WorkList) (EZU-FC11W) / DICOM Software
(Printer) (EZU-FC11P)
Components Software
Functions
(Notice) : EZU-NW1 and EZU-FC11 are required for EZU-FC11P and EZU-FC11W.
- 31 - SN-U391E
- Measurement Caliper, distance ratio, angle, area, average velocity, acceleration, heart rate
(Notice: If there is no tag for calibrations, average velocity, acceleration and
heart rate cannot be measured.)
- Distance calibration Applicable in caliper and area
- Annotation Applicable (text and rectangle on display)
- Hard disk storage Images acquired from DICOM server can be preserved up to 3GB.
- Real-time ODM mode B (Real) / ODM, B (Real) / ODM / ODM, ODM,ODM / ODM
- Cine ODM mode B (Cine) / ODM, B (Cine) / ODM / ODM
Settings
- 32 - SN-U391E
- Data export Applicable (CSV format)
Specifications - IMT (Intima-Media Thickness) measurement
- Image format Still image (Frozen, playback image)
- Measurement results
Trace method Max, mean, value of 3 points, average value of 3 points
Simple 3 points method
Value of 3 points, average value of 3 points
- Manual correction of trace line
Applicable (in trace method)
- Detection position correction
Applicable (in simple points method)
Specifications - Color tissue tracking measurement
- Image format AVI, Motion JPEG (B/W B mode image)
- Specified measurement segments
APICAL, MID, BASAL
- Graphs for transition Applicable
- Data export Applicable (CSV format)
- Creating Bull’s eye Applicable
Specifications - LA tracking measurement
- Image format AVI, Motion JPEG (B/W B mode image)
- Measurement results Volume, dV/dt, area, circulation, long axis
- Volume measurement Modified Simpson method
- Biplane measurement Applicable
- Graphs for transition Applicable
- Data export Applicable (CSV format)
- 33 - SN-U391E
1.5s ~ 10.0s: 0.5s step
11.0s ~ 14.0s: 1s step
Cardiac 31 steps
0.2s ~ 1.0s: 0.1s step
1.5s ~ 10.0s: 0.5s step
11.0s ~ 14.0s: 1s step
- ECG trigger timing (ECG display on) 32 steps
1 ~ 32 beats, 1 beat step
Acoustic pressure %Control and MI Control are selectable
MI Control : 0~0.5 0.01 step
0.5~ 0.05 step
(Max MI value: Depend on probes and focus setting and so on)
%Control : 0~100% 34 steps
Automatic acoustic pressure control on fundamental image in monitor mode
and alternate mode is possible.
Reference frequency (REF) 2 frequencies
Stopwatch function -30s ~ 59m59s, for monitoring the time after injection of contrast agent
For monitoring the time of image recorded in cine memory.
Application function Possible to independently set up each application parameter for contrast
imaging
Multi-step trigger (MST) mode
This function uses the technology of a different aspheric focusing from the
conventional focal technique, it can generate uniform acoustic field in the
body
- 34 - SN-U391E
Alternate mode In dual screen displays in alternate mode, contrast image (left screen) and
fundamental image (right screen) are transmitted and displayed alternately,
ultrasound contrast agents and anatomy can be observed on real time
(applicable in middle and low acoustic pressure transmission mode)
MTI (microbubble trace imaging) mode
(Notice) :Up and Down display of B mode can not be selected, when an applicable application of
EZU-CH8 is selected.
- 35 - SN-U391E
Measurement function B mode measurement (except for histogram and CFM VELO)
- ROI setting mode Displays the ROI that specifies the 3D image creation area on the 2D image
- 3D mode Real-time 3D image is displayed on 1 screen.
- 36 - SN-U391E
- 2D-3D mode Screen is vertically divided into two sides. The ROI setting mode is
displayed on the left side and the 3D mode on the right side
- MPR 4 screen mode Screen is vertically and horizontally divided into four. 3D image is displayed
on the lower right, while the random orthogonal 3 cross-section surface
images are displayed on the other sections
- MPR 1 screen mode One of the orthogonal 3 cross-section surface image in MPR 4 screen mode
is displayed on 1 screen
- Multi Slice multi screen mode
Screen is divides into multi (2, 4, 9, 16) sections. A reference image is
displayed on the upper left section, and the slice images are displayed on
the other sections.
- Multi Slice 1 screen mode
The selected slice image from Multi Slice multi screen mode is displayed
on 1 screen.
HI Definition 3D By one swing scanning at a speed that is slower than real time 3D, and
high-definition 3D image or MPR image is displayed.
Applicable to CFM mode (CFI, CFA and Fine Flow except for color maps
for displays of velocity variance)
Rendering mode
B/W
- Opacity
- MIP (Maximum Intensity Projection)
- mIP (minimum Intensity Projection)
- Transparency
- Inversion
- Orthogonal
- Opac. Advanced
- 4Dshading
Color
- Opacity
- MIP (Maximum Intensity Projection)
- Transparency
- Orthogonal
Quality settings
- 37 - SN-U391E
- Color
3D color 37 types (except rendering mode: Orthogonal, Opac.Advanced, and
4Dshading)
MPR color 37 types
Depth color 14 types (rendering mode: Opac.Advanced)
- Shading 6 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
- Filter 20 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced,, or
4Dshading)
- Threshold 256 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced, or
4Dshading)
- Smoothing 8 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
- P-Filter 8 steps (rendering mode: MIP or mIP)
- BW 3D Smooth. 3 types (rendering mode: 4Dshading)
- 3D Smooth Coef 11 types (rendering mode: Opac.Advanced, or 4Dshading)
- Depth Contrast 40 types (rendering mode: Opac.Advanced)
- Depth Color Shift 127 types (rendering mode: Opac.Advanced)
- Region Map 9 types (rendering mode: 4Dshading)
- Absorption 10 types (rendering mode: 4Dshading)
- Hue 4 types (rendering mode: 4Dshading)
- Shadow Smooth 10 types (rendering mode: 4Dshading)
- Light Intensity 5 types (rendering mode: 4Dshading)
- Lighting Reso. 4 types (rendering mode: 4Dshading)
- Transp. Level 8 types (rendering mode: 4Dshading)
Color
- Combination B/W and CFM image combination / only B/W image / only CFM image
(only CFM image is impossible when rendering mode is Orthogonal)
- Blend 100 steps (except rendering mode: Orthogonal)
- Brightness
3D image 17 steps (-8 ~ 8, except rendering mode: Orthogonal)
MPR image 17 steps (-8 ~ 8) (except CFI mode)
- Shading 6 steps (rendering mode: opacity except CFI mode)
- Filter 5 steps (rendering mode: only opacity)
- Threshold 256 steps (rendering mode: only opacity )
- Smoothing 8 steps (rendering mode: only opacity)
- P-Filter 8 steps (rendering mode: only MIP)
Operations
- 38 - SN-U391E
Image operations: both B/W and CFM images / only B/W image
- Cut function The arbitrarily specified part of 3D image can be dug (only in frozen, except
rendering mode: Orthogonal)
Image operations: both B/W and CFM images / only B/W image
- Zoom 1.0~4.0 (0.25 steps)
- Z Adjust 10 steps
- Image filing 3D volume data, Real-time 3D mode single images (such as BMP) and multi
images (such as AVI) can be saved
Saved images can be played back from the image filing screen, as with
regular 2D images
Image pre-cutted by ViewLine and full image can be switched
- Cine memory 3D images can be saved in the cine memory and played back
- PSS function Using the 3D PSS (3D Preset Scanning Selector) function
- Intelligent Surface Tracking
Set ViewLine automatically
- Volume Slice Imaging Display a thick cross-sectional image
From 0mm (OFF) to 20mm. 24 steps.
Measurement Distance, area, circumference, volume etc. can be measured on 2D image,
MPR image or Multi Slice image.
User-defined screen button The items of 3D screen button menu and image adjustment screen button
menu are registerable
- 39 - SN-U391E
3.2.20. Three-dimensional Display Software (Freehand) (EZU-3D8)
Components Software (Freehand 3D software)
Applicable probe Refer to Appendix
Display mode
- ROI setting mode Displays the ROI that specifies the 3D image creation area on the 2D image
- 3D mode 3D image is displayed on 1 screen.
- 2D-3D mode Screen is vertically divided into two sides. The ROI setting mode is
displayed on the left side and the 3D mode on the right side
- MPR 4 screen mode Screen is vertically and horizontally divided into four. 3D image is displayed
on the lower right, while the random orthogonal 3 cross-section surface
images are displayed on the other sections
- MPR 1 screen mode One of the orthogonal 3 cross-section surface image in MPR 4 screen mode
is displayed on 1 screen
Rendering mode
B/W
- Opacity
- MIP (Maximum Intensity Projection)
- mIP (minimum Intensity Projection)
- Transparency
- Inversion
- Orthogonal
- Cube
- Opac. Advanced
- 4Dshading
Color
- Opacity
- MIP (Maximum Intensity Projection)
- Transparency
- Orthogonal
- Cube
Quality settings
B/W
- Brightness
3D image 17 steps (-8 ~ 8, except rendering mode: Orthogonal and Cube)
MPR image 17 steps (-8 ~ 8)
- Color
3D color 37 types (except rendering mode: Orthogonal, Cube and, Opac.Advanced,
4Dshading)
- 40 - SN-U391E
MPR color 37 types
- Shading 6 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
- Filter 20 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced,, or
4Dshading)
- Threshold 256 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced, or
4Dshading)
- Smoothing 8 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
- P-Filter 8 steps (rendering mode: MIP or mIP)
- BW 3D Smooth. 3 types (rendering mode: 4Dshading)
- 3D Smooth Coef 11 types (rendering mode: Opac.Advanced, or 4Dshading)
- Depth Contrast 40 types (rendering mode: Opac.Advanced)
- Depth Color Shift 127 types (rendering mode: Opac.Advanced)
- Region Map 9 types (rendering mode: 4Dshading)
- Absorption 10 types (rendering mode: 4Dshading)
- Hue 4 types (rendering mode: 4Dshading)
- Shadow Smooth 10 types (rendering mode: 4Dshading)
- Light Intensity 5 types (rendering mode: 4Dshading)
- Lighting Reso. 4 types (rendering mode: 4Dshading)
- Transp. Level 8 types (rendering mode: 4Dshading)
Color
- Combination B/W and CFM image combination / only B/W image / only CFM image
(only CFM image is impossible when rendering mode is Orthogonal or
Cube)
- Blend 100 steps (except rendering mode: Orthogonal and Cube)
- Brightness
3D image 17 steps (-8 ~ 8, except rendering mode: Orthogonal and Cube)
MPR image 17 steps (-8 ~ 8) (except CFI mode)
- Shading 6 steps (rendering mode: opacity except CFI mode)
- Filter 5 steps (rendering mode: only opacity)
- Threshold 256 steps (rendering mode: only opacity )
- Smoothing 8 steps (rendering mode: only opacity)
- P-Filter 8 steps (rendering mode: only MIP)
Operations
- 41 - SN-U391E
rendering mode: Orthogonal and Cube)
Image operations: both B/W and CFM images / only B/W image
- Zoom 1.0~4.0 (0.25 steps)
- Image filing 3D volume data, Real-time 3D mode single images (such as BMP) and multi
images (such as AVI) can be saved
Saved images can be played back from the image filing screen, as with
regular 2D images
Image pre-cutted by ViewLine and full image can be switched
- PSS function Using the 3D PSS (3D Preset Scanning Selector) function
Measurement Not applicable
User-defined screen button The items of 3D screen button menu and image adjustment screen button
menu are registerable
- 42 - SN-U391E
Input signal Y/C
Applicable probes All probes
Display mode
- 43 - SN-U391E
3.2.29. External LCD Monitor Unit (EZU-EL2)
Components DVI Connector Unit
Specifications
- 44 - SN-U391E
3.4. Environment conditions
Temperature 10 ~ 35°C
Relative humidity 30 ~ 85% RH (No dew condensation)
Atmospheric pressure 700 ~ 1060hP
- 45 - SN-U391E
1350
820
450 700
1430
900
450 700
Unit: mm
Body dimension (with EZU-MT29-S1 and EZU-HN1)
- 46 - SN-U391E
1400
820
450 700
1480
900
450 700
Unit: mm
Body dimension (with EZU-MT30-S1 and EZU-HN1)
E