General Data
Electrical:
Circuit:
High Voltage Circuit ..................................................... Constant Potential High-Voltage Generator
Grounding ............................................................................................................ Center-Grounded
Nominal X-ray Tube Voltage (IEC60613:2010) ........................................................................ 110 kV
Nominal Focal Spot Value (IEC60336:2005) ................................................................................. 0.6
Nominal Anode Input Power (at 1.0s) ..................................................................................... 2020 W
Nominal Radiographic Anode Input Power (IEC60613:2010) ................................................. 2100 W
Exposure Duty Cycle ....................................................................................................... 1:60 or more
(Exposure Time : Interval Time)
★The information contained herein is presented only as a guide for the application of our products. No responsibility is assumed by Toshiba Electron
Tubes & Devices Co., Ltd. (TETD) for any infringements of patents or other rights of the third parties which may result from its use.
No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore, advisable to contact to TETD before processing with the design of
equipment incorporating this product.
D-063 / D-063R / D-063S / D-063SR
Mechanical:
Dimensions:
Overall Length ........................................................................................... See dimensional outline
Maximum Diameter ................................................................................... See dimensional outline
Target:
Anode Angle .................................................................................................................... 8 degrees
Material ............................................................................................................................. Tungsten
Inherent Filtration ..................................................................................... At least 0.8 mm Al at 50 kV
X-ray Coverage .......................................................................................... φ245 mm at SID 970 mm
Weight:
D-063 ......................................................................................................................... Approx. 320 g
D-063R ...................................................................................................................... Approx. 440 g
D-063S ...................................................................................................................... Approx. 450 g
D-063SR .................................................................................................................... Approx. 570 g
Cooling Method ............................................... Oil immersed (60℃ Max.) and convection oil cooling.
Tube Holding:
D-063 / D-063R ............................ Holding the glass envelope of the anode end and cathode end,
or the screw of the anode shank.
D-063S / D-063SR ........................................................................... Holding the insulation cylinder.
-2-
D-063 / D-063R / D-063S / D-063SR
Environmental Limits
-3-
D-063 / D-063R / D-063S / D-063SR
18 40kV
TUBE CURRENT [mA]
16
14
12
10
0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10
EXPOSURE TIME [s]
Refer to IEC60613:2010
110kV
TUBE CURRENT [mA]
12 3.5
Ef
8 3.0
4 2.5
0 2.0
2.9 3.1 3.3 3.5 3.7
-4-
D-063 / D-063R / D-063S / D-063SR
D-063 / D-063S
28
265W
25
HEAT STORAGE [kJ]
20
150W
15
10
COOLING
HEATING
0
0 100 200 300 400 500 600 700 800 900
TIME [s]
D-063R / D-063SR
35
600W
30
HEAT STORAGE [kJ]
25
300W
20
15
COOLING
10
100W
5
HEATING
0
0 60 120 180 240 300 360 420 480 540 600
TIME [s]
-5-
D-063 / D-063R / D-063S / D-063SR
CENTRAL X-RAY
φ16
M5 10DEEP M3
65±1
8°
120±5
CENTRAL X-RAY
φ52MAX
FOCAL SPOT
TERMINAL SLEEVE
12MAX
C2 F )
.4
φ20
(
N
CENTRAL X-RAY
SHORT PIN
C1 F
F :FILAMENT
C1,C2:CATHODE
N :NOT FOR USE
-6-
D-063 / D-063R / D-063S / D-063SR
5
0
.
0
±
1
.
4
φ
-
4
CENTRAL X-RAY
1
.
0
±
8
2
φ
φ40
M5 10DEEP
RADIATOR
21
26
86±1
141±5
8°
CENTRAL X-RAY
φ52MAX
FOCAL SPOT
TERMINAL SLEEVE
12MAX
C2 F )
0 .4
( φ2
N
CENTRAL X-RAY
SHORT PIN
C1 F
F :FILAMENT
C1,C2:CATHODE
N :NOT FOR USE
-7-
D-063 / D-063R / D-063S / D-063SR
CENTRAL X-RAY
φ16
M5 10DEEP
M3
INSULATION CYLINDER
65±1
60
φ12
8°
130±3
125
CENTRAL X-RAY
FOCAL SPOT
65
4MAX
TERMINAL SLEEVE φ60
C2 F )
.4
φ20
(
N
CENTRAL X-RAY
SHORT PIN
C1 F
F :FILAMENT
C1,C2:CATHODE
N :NOT FOR USE
-8-
D-063 / D-063R / D-063S / D-063SR
5
0
.
0
±
1
.
4
φ
-
4
CENTRAL X-RAY
1
.
0
±
8
2
φ
φ40
M5 10DEEP
RADIATOR
7
21
26
INSULATION CYLINDER
86±1
60
φ12
151±3
8°
125
CENTRAL X-RAY
FOCAL SPOT
65
4MAX
TERMINAL SLEEVE φ60
C2 F )
.4
φ20
(
N
CENTRAL X-RAY
SHORT PIN
C1 F
F :FILAMENT
C1,C2:CATHODE
N :NOT FOR USE
-9-
Marketing Engineering Group, Sales Department
1385,Shimoishigami,Otawara-shi,Tochigi 324-8550,Japan
Tel: +81-287-26-6666 Fax: +81-287-26-6060
http://www.toshiba-tetd.co.jp/
・The head office of Toshiba Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Environmental Management System ISO14001.
・Toshiba Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Quality Management Systems ISO9001 and ISO13485.
Product scope is referred to the following URL. http://www.toshiba-tetd.co.jp/tetd/eng/company/quality.htm.
PE-D-063/R/S/SR 2017-02