Sie sind auf Seite 1von 6

NEWSLETTER NO.

909
October 29, 2002
Dear Sirs,

Information on
New PCB connector and New Casters of SSD-5500

From the following serial number, connectors of PCBs (Printed Circuit Board) in the unit and
the casters will change. However, software version (Ver. 5.1) and the performance of the
system remain unchanged.

M02220 and higher

With this change, some optional units will also change to cope with the change of the
connector of PCBs.
Detailed explanation follows.

1. New PCB connectors

1.1 SSD-5500 main unit

To reinforce connection of PCBs, we have taken a drastic measure. The connectors of


some PCBs have been changed to new ones. For the picture of the new connector,
please refer to the attached papers. Not only the connectors of the PCBs but also the
sockets inside the unit will change to accept the PCBs with new connectors.

We believe this alteration will increase reliability of our products further.

The change of the PCB connectors causes no change in performance.


-2-

2.2 Optional units


Please note that the following optional units have been altered to adapt to the new PCB
connectors of the main unit. Please use the new model numbers when you order the new
optional units. Please put suffix “Q” to the last of the former model number.

Example) Former model number: EU-9078


New model number: EU-9078Q

NOTE: The former and new optional units have no interchangeability.


The former optional units can be installed only to the former SSD-5500 main units.
The new optional units can be installed only to the new SSD-5500 main units.

Model numbers:
New model numbers Former model numbers Description
EU-9078BQ EU-9078B A-SMA
EU-9079CQ EU-9079C Volume Mode
PEU-5500Q PEU-5500 Physiological signal
SCU-5500Q SCU-5500 Mechanical probe connection
DAS-5500-1Q DAS-5500-1 eDMS (100V-120V)
DAS-5500-1VQ DAS-5500-1V eDMS (200-240V)
Others, such as EU-3040, DMS-5500B, and software options SOP-5500-*, remain unchanged.

NOTE:
Example)
1. Please be advised that the description (in
Former model MODEL EU-9078B
the “MODEL” section) on the label put on
New model EU-9078BQ
the case of the optional units remain
unchanged (the model numbers without Q).
The new model numbers will be written
under the “MODEL” section.

2. The model numbers of PCBs with new connectors differ from the former ones on the
following point.

If the former model number is EP XXXX 00 AA, (The 00 might be 01, 02, or higher.)

The new model number is EP XXXX 50 AA. (The 50 might be 51, 52, or higher.)
-3-

2. Casters

Because of terminated production of the former casters, the casters have been changed to
new ones. With the change of the casters, the height of the unit increases a little and
becomes 132 cm to 157 cm (changes with the monitor position).
Former caster is double-wheel type and the new caster is single-wheel type.

Pictures:

New caster Former caster

3. Sales materials
3.1 Brochures
From the document numbers E169-2 (for radiology) and E170-2 (for cardiology) the picture
of SSD-5500 have been changed to those with new casters.
If you need pictures of the units with new casters, please ask our staff in charge.
We are able to supply the pictures in digital form (J-PEG and TIFF) and positive films. See
attached papers for the pictures.

3.2 Specification pamphlet


Please use the attached revised specification pamphlet with document number SP-5500-
V5.1-2.0.

Yours very truly


ALOKA CO., LTD.

Y. Eihara
Director & General Manager
International Sales and Marketing Dept.
YE/Ss
Encl. Specifications pamphlet SP-5500-V5.1-2.0
PCB connectors

New connector

A
B

A B
Double sided contact

Former connector

A
B

A B
Single sided contact
Photo of the unit with new casters. Difference between the radiology version and the
cardiology version is the connected probes.

R-1 R-2
Radiology version

C-1 C-2

Cardiology version

Das könnte Ihnen auch gefallen