Beruflich Dokumente
Kultur Dokumente
2 Main Components
3 Electrical System
6 Electrical Documentation
Service Manual for Centrifuges
Section 1: General Safety Instructions/Norms
1.2 Norms......................................................................................................................2
It is obvious that not all the safety precautions which have to be observed by users
can always be adhered to by service personnel (e.g. testing with lid open).
1.2 Norms
Please refer to the respective operating instructions as regards the valid norms and
guidelines.
Section 1.1 (technical data) of the respective operating instructions will provide
information as to whether the centrifuge is subject to a mandatory examination (e.g.
according to UVV VBG 7z).
Should this be the case, the centrifuge must be tested in accordance with the
detailed instructions as laid down in the relevant regulations.
Sigma offers its customers maintenance service contracts. It is the obligation of the
after sales service to carry out maintenance work in such a way that all the valid
accident prevention regulations are strictly adhered to.
Type plate
Arrow indicating direction of rotation Rotor, buckets, suspension gear
Frame
Mechanical fastening Rotor
Lid, panels, seals Check rotors for material fatigue (cracks); swing
rotors need checking whether all parts move
Lid lock easily; clean suspension and grease, if
necessary
For motor driven lid locks:
Check whether drive belt is dry and clean and Rotor accessories
regulate the tension, if necessary.
Rotor inserts
Check all moving parts of the lid lock and lid
Welding connections
hooks to see whether they move easily and
grease, if necessary Fastening of drum
Lid hooks Rotor lid screw
Product supply and outflow
Panels Rotor bearing
(Splash boards etc. inside) Seat of drum
Rotor chamber Bearing seat
Vibration damper: Gasket
Confirmation: Returned part is not contaminated. Confirmation: Returned part is not contaminated.
Signature:_________________________ Signature:_________________________
Service Information Laboratory Centrifuges Service Information Laboratory Centrifuges
Confirmation: Returned part is not contaminated. Confirmation: Returned part is not contaminated.
Signature:_________________________ Signature:_________________________
Service Information Laboratory Centrifuges Service Information Laboratory Centrifuges
Confirmation: Returned part is not contaminated. Confirmation: Returned part is not contaminated.
Signature:_________________________ Signature:_________________________
3.6 Declaration of Decontamination / Return Declaration
Following declarations serve for keeping safety and health of our employees. Fill in the
forms and attach them when returning centrifuges, spare parts and accessories.
Please understand that we cannot carry out any work before we have the declarations.
(We recommend to make several copies of this page.)
!!!! Attention – This form must be glued on outside of the packing !!!!
Return Declaration
YES NO
Decontamination declaration inside :
!!!! Attention – This form must be glued on outside of the packing !!!!
Declaration of Contamination of Centrifuges, Accessories and
Spare Parts
This declaration may only be filled in and signed by authorised staff.
Order No. :
Accessories :
If not, which substances have come into contact with the equipment?
Prior to repair, radioactively contaminated components must be decontaminated according to the valid
regulations for radiation protection.
I / we hereby declare that the information on this declaration are correct and complete.
Company / Institute :
Street :
Postcode, City :
Tel. : FAX :
Name :
Date : Stamp :
Signature :
Service Manual for Centrifuges
Section 4: Error Messages / Malfunctions
4.1.3 Error List Centrifuges 1-13, 1-16, 1-15, 1-15K, 2-3, 2-4, 2-5, 2-16 ............................................ 9
4.2 Malfunctions........................................................................................................ 13
4.2.1 Display dark, Zent3................................................................................................................. 13
1 • • CPU test indicates fault after power up Replace EPROM/replace microcontroller board
2 • • Stack overrun
7 • • Communication to display microcontroller faulty Check connection between microcontroller and display
board / Replace display board / replace microcontroller
board
8 • • Internal watchdog activated too frequently Check power supply of microcontroller board / replace
EPROM / replace microcontroller board
18 • Internal watchdog of motor microcontroller Check power supply of microcontroller board / replace
activates too frequently EPROM / replace microcontroller board
20 • • Interrupt signal of speed sensor faulty Check connection between speed sensor and
microcontroller board (particularly shielding) / check
distance between Hall sensor and magnet (distance 1+/-0,5
mm) / replace speed sensor
24 • • Final speed not reached Check input voltage of centrifuge (see type plate) / check for
stiffness / replace power board / replace motor
25 • • Converter indicates fault Check input voltage of centrifuge (see type plate) / replace
power board / if error appears during braking, check braking
resistor
26 • • MFR signal non-existent or faulty although Replace EPROM / replace microcontroller board
speed signal available
28 • • Start without speed signal Check connection between speed sensor and
microcontroller board (particularly shielding) / check
distance between Hall sensor and magnet (distance 1+/-0,5
mm) / replace speed sensor
29 • • No speed signal during run Check connection between speed sensor and
microcontroller board (particularly shielding)
31 • Motor microcontroller exceeded maximum speed Check connection between speed sensor and
microcontroller board (particularly shielding)
34 • Final speed not reached by motor Check input voltage of centrifuge / check for stiffness
microcontroller
Replace power board / replace motor
36 • MFR signal non-existent or faulty although Replace EPROM / replace microcontroller board
speed signal available
38 • Motor microcontroller starts without rotor Check connection between speed sensor and
microcontroller board (particularly shielding)
40 • Hardware for speed control faulty Jumper on microcontroller board connected correctly? /
replace microcontroller board
41 • Hardware speed control activated at maximum Jumper list see section 6.2.1.6
rotor speed
42 • • Hardware for imbalance detection faulty Check connection between imbalance sensor and
microcontroller board / imbalance sensor not in the centre
between the two magnets? / replace imbalance sensor
43 • • Signal of imbalance sensor has unrealistic value (Check position, service mode No. 2, values between 300
and 400, for model 8K10 between 400 and 600)
50 • • Error during floating point calculation Replace EPROM / replace microcontroller board
69 • • EEPROM data not compatible with type of Erase EEPROM / replace EPROM / replace microcontroller
centrifuge board
70 • • EEPROM version not compatible with SW This error occurs the first time the microcontroller board is
version used after having been used with other types of
centrifuges / erase by pressing any key or mains off/on (also
applicable for error 77)
73 • • Checksum error when reading program data Erase EEPROM / replace EPROM / replace microcontroller
from EEPROM board
76 • • Checksum error when reading offset values from Erase EEPROM / replace EPROM / replace microcontroller
EEPROM board
78 • • Lid not closed properly Mechanism is jammed / readjust / motor has not been
activated / replace board for motor control
79 • • Time difference between right and left hook is Check for foreign objects / close lid again / delete by
too great when closing lid pressing any key
80 • • Board for lid control defective Replace board / check board for light barrier
81 • • Rotor turns (MFR) and lid is open (rotor has Check lid contact / check speed sensor
been turned very quickly manually / emergency
release was opened while motor was turning)
82 • • Test signal for lid locking is faulty Check lid switch and connection to power board / replace
power board / replace microcontroller board
84 • • Heat sink above 70 °C (switch) Heat sink too hot? -> improve ventilation / check
overtemperature switch and cable
85 • • Temperature inside rotor above 50° C Improve ventilation / check refrigeration system / check
power board
87 • • Temperature sensor signal faulty Check connection between temperature sensor and
microcontroller board / replace temperature sensor / replace
microcontroller board
88 • • Overtemperature switch of rotor chamber has Improve ventilation / check refrigeration system / check
been activated connection between overtemperature switch and power
board / replace overtemperature switch
94 • • Signal of static imbalance sensor faulty Check connection between imbalance sensor and
microcontroller board / imbalance sensor not in the centre
between the two magnets? / replace imbalance sensor
95 • • Signal of dynamic Imbalance sensor faulty Replace EPROM / replace microcontroller board
98 • • Rotor identification not possible Distance between rotor and rotor code sensor too great? /
rotor used without magnets / check cable to rotor code
sensor / check rotor code sensor
99 • • Rotor has been identified but does not Insertion of incorrect rotor / distance between rotor and rotor
correspond to selected rotor code sensor too great? / rotor used without magnets / check
cable to rotor code sensor / check rotor code sensor
8 Internal watchdog activated too frequently Check power supply of microcontroller board / replace
EPROM / replace microcontroller board
20 Interrupt signal of speed sensor faulty Check connection between speed sensor and microcontroller
board (particularly shielding) / check distance between Hall
sensor and magnet (distance 1+/-0,5 mm) / replace speed
sensor
24 Rotor does not rotate - Check for foreign bodies inside the centrifuge → remove
them
- Mechanical damage or fault in the motor winding → replace
motor
- Check connecting cable and plug between motor and power
board → renew connection between cable and plug and re-
establish contact
- Power board defective →replace power board
- Check all connecting cables → re-establish contact
- Microcontroller board defective (No control signals for power
board) → replace microcontroller board
25 Converter indicates fault Check input voltage of centrifuge / replace power board / if
error appears during braking, check braking resistor
26 MFR signal non-existent or faulty although speed signal Replace EPROM / replace microcontroller board
available
28 No speed signal or rotor does not rotate Correct faults no. 23 or 24 (fault exists only in centrifuges
without rotor identification)
29 No speed signal during run Check connection between speed sensor and microcontroller
board (particularly shielding) / Check distance between Hall
sensor and magnet (distance 1+/-0,5 mm) / replace speed
sensor
40 Test of system: Overspeed not detected / overspend Replace overspeed controller or EPROM
controller not compatible with EPROM
42 Test of system: imbalance detection faulty Check grey wedge of the light barrier of the imbalance system
43 Signal of imbalance sensor has wrong offset value Recalibrate imbalance detection system
50 Error during floating point calculation Replace EPROM / replace microcontroller board
69 EEPROM data not compatible with type of centrifuge Erase EEPROM / replace EPROM / replace microcontroller
board
72 Error during data transfer with EEPROM. EEPROM is Replace EPROM / replace microcontroller board
defective of communication is disrupted
73 Checksum error when reading program data from Erase EEPROM / replace EPROM / replace microcontroller
EEPROM board
78 Lid not closed properly Mechanism is jammed / readjust / motor has not been
activated / replace board for motor control
79 Time difference between right and left hook is too great Check for foreign objects / close lid again / delete error by
when closing lid pressing any key.
80 Board for lid control defective Replace board / check light barrier board
81 Rotor turns (MFR) and lid is open (rotor has been turned Check lid contact / check lid sensor
very quickly manually / emergency release was opened
while motor was turning)
82 Test signal for lid locking is faulty Check lid switch and connection to power board / replace
power board / replace microcontroller board
83 Lid does not open after activating the lid release device
84 Heat sink above 70 °C (switch) Heat sink too hot? -> improve ventilation / check
overtemperature switch and cable
85 Temperature inside rotor above 50° C Improve ventilation / check refrigeration system / check power
board
87 Temperature sensor signal faulty Check connection between temperature sensor and
microcontroller board / replace temperature sensor / replace
microcontroller board
88 Overtemperature switch of rotor chamber has been Improve ventilation / check refrigeration system / check
activated connection between overtemperature switch and power
board / replace overtemperature switch
94 Signal of static imbalance sensor faulty Check connection between imbalance sensor and
microcontroller board / imbalance sensor not in the centre
between the two magnets? / replace imbalance sensor
95 Signal of dynamic imbalance sensor faulty Replace EPROM / replace microcontroller board
98 Rotor identification not possible Distance between rotor and rotor code sensor too great? /
rotor used without magnets / check cable to rotor code
sensor / check rotor code sensor
99 Rotor has been identified but does not correspond to Insertion of incorrect rotor / distance between rotor and rotor
selected rotor code sensor too great? / rotor used without magnets / check
cable to rotor code sensor / check rotor code sensor
4.1.3 Error List Centrifuges 1-13, 1-16, 1-15, 1-15K, 2-3, 2-4, 2-5, 2-16
No. valid Description Corrective Measures
for
1 Speed sensor signal faulty Check connection between speed sensor and
microcontroller board (particularly shielding) / check
distance between Hall sensor and magnet (distance 1+/-
0,5mm) / replace speed sensor
3 Centrifuge does not start up Check speed sensor / check inverter / check connection
between motor and microcontroller board / check motor
4 Lid does not open after activating the lid release device Check lid switch and connection to power board / replace
power board / replace microcontroller board
5 Rotor turns and lid is open (rotor has been turned very Check lid switch and speed sensor
quickly manually / emergency release has been used while
rotor has been turning)
all
6 MFR signal non-existent or faulty although speed signal Replace microcontroller board / replace rotor
available / same poles of rotor magnets following one
another
7 Maximal um speed has been exceeded Check connection between speed sensor and
microcontroller board (particularly shielding) / check
distance between Hall sensor and magnet (distance 1+/-
0,5mm) / replace speed sensor
11 Wrong setting of jumpers (not possible in 1-15K, 2-16) Check setting, see table 6.5 “Setting of jumpers”
13 1-15K Malfunction of temperature sensor Check temperature sensor / not connected of defective
14 1-15K Overtemperature (>50°C) Let cool down / check refrigeration unit / check switching
function of relay K2
15 2-16 Rotor code not recognised or read error Check rotor code sensor / check magnet codification of
rotor
16 2-16 Rotor not allowed / rotor magnets too weak Check rotor / replace rotor
In these models, the number of the error that has occurred is indicated by the
position of the hand in the time display. At the same time the three keys and all
segments of the speed display light up.
In models 1-13, 2-3 and 2-4, an error is indicated by the “Start” LED flashing on and
off quickly. The number of flashes of the “Lid ” LED indicates the number of the error.
The error mode is indicated by "ERR". The error number is shown in the time display.
When "ERR" and "Safe" appear alternately in the display, operate power on/off not
until "OFF" has been displayed.
7 SPD_OVERSPD Current speed exceeds max. allowed speed for - Check speed sensor function
more than 200 rpm. Also possible when rotor- - Check speed sensor plug, socket and
code identification don’t works correctly cable
- Check rotor-code sensor
8 CPU_FAIL Internal micro-controller failure - Replace control-board
9 DOOR_TRANSISTOR Lid-transistor V10 or V11 resp. V27 defective - Check V10/V11/V27 on power-board
- Check relays K2
- Replace power-board
10 SW_PROBLEMS Undefined software state - Replace control-board
- Firmware-Update
18 All LPCOM_ERROR Communication error between power-board and - Check 10 pole flat connection cable
control-board - Check compatibility between power-board
version and control-board version
- Replace control-board /power-board
3-16K Drive temperature >140°C - Check drive cables, plugs and sockets
- Replace power-board
When the error is released by the powerboard, „Pb“ is additionally displayed in the setspeed area.
4.2 Malfunctions
NOTE: The imbalance and offset values have to be entered again. If the sticker
on the cover in the front is missing, a parameter setting run has to be
carried out to determine the imbalance values anew.
Corrective measure: Replace grey wedge, carry out parameter setting run.
1 MAIN COMPONENTS...........................................................................2
5.1 Panels.............................................................................................................................................. 2
5.1.1 Dismantling of Panels in Model 1-13..........................................................................................................2
5.1.2 Dismantling of Panels in Models 1-6 and 1-15...........................................................................................2
5.1.3 Dismantling of Panels in Models 2-3, 2-4 and 2-15....................................................................................3
5.1.4 Dismantling of Panels in Models with Steel Sheet Housing......................................................................3
5.1.5 Dismantling of Panels in Model 8K10........................................................................................................5
5.1.6 Dismantling of Panels in Model 2-5............................................................................................................6
5.2 Lid 8
5.2.1 Lid Emergency Release..............................................................................................................................8
5.2.2 Lid Locks....................................................................................................................................................9
5.2.2.1 Removal of Lid Locks........................................................................................................................9
5.2.3. Lid Gasket...............................................................................................................................................11
5.2.3.1 Replacement of Lid Gasket in Models 1-15K, 1K15, 2K15,
4-15, 4K15, 6-15 and 6K15...........................................................................................................................11
5.2.3.2 Replacement of Lid Gasket in Model 8K10.....................................................................................11
5.2.4 Relief Springs...........................................................................................................................................11
5.2.4.1 Removal of Relief Springs in Models 1-6 and 1-15.........................................................................11
5.2.4.2 Removal of Relief Springs in Models 2-3, 2-4 and 2-15..................................................................12
5.2.4.3 Removal of Relief Springs in Centrifuges with Steel Sheet Housing..............................................12
5.2.4.4 Gas Pressure Springs........................................................................................................................13
5.2.4.5 Gas Pressure Springs (Models 1-15K, 2-5, 2-16).............................................................................13
Loosen lower fastening screw of the gas pressure spring at the bottom side of the centrifuge (see fig. 5.11).
Remove pressure spring in upward direction. ...................................................................................................14
5.2.5 Lid of Model 8K10...................................................................................................................................14
5.3 Drive System of Centrifuge........................................................................................................ 15
5.3.1 Motor Suspension.....................................................................................................................................15
5.3.2 Rubber Boot..............................................................................................................................................15
5.3.3 Vibration Damper.....................................................................................................................................15
5.3.3.1 Vertical Vibration Damper...............................................................................................................16
5.3.4 Removal of Motor...................................................................................................................................17
5.3.4.1 Preparatory Work for all Models.....................................................................................................17
5.3.4.2 Removal of Motor in Model 1-13......................................................................................................18
5.3.4.3 Removal of Motor in Models 1-6 and 1-15.......................................................................................18
5.3.4.4 Removal of Motor in Models 2-3 and 2-4.........................................................................................18
5.3.4.5 Removal of Motor in Model 2-15.....................................................................................................19
5.3.4.6 Removal of Motor in Models 2K15 (starting with serial no. 51003) and models of the 3-range .. .19
5.3.4.7 Removal of Motor in Models 4-15, 4K15, 6-15 and 6K15..............................................................19
5.3.4.8 Removal of Motor in Model 8K10...................................................................................................19
5.3.4.9 Removal of Motor in Models 1-5K and 2-5.....................................................................................20
5.3.4.10 Removal of Motor in Model 2-16...................................................................................................20
5.3.5 Installation of Motor.................................................................................................................................21
5.3.5.1 Special Information for Models 2-3 and 2-4....................................................................................21
5.4 Fuses............................................................................................................................................. 21
5.5 Refrigeration System................................................................................................................. 22
5.5.1 Air Cooling of Centrifuge........................................................................................................................22
5.5.2 Cooling of Centrifuge by Refrigeration...................................................................................................23
5.5.2.1 Compressor.......................................................................................................................................23
5.5.2.2 Cooling Agent Condensor................................................................................................................23
5.5.2.3 Expansion Device.............................................................................................................................23
5.5.2.4 Vaporizer..........................................................................................................................................24
5.5.2.5 Working on the Refrigeration System..............................................................................................24
5.5.2.6 Quantity of Cooling Agent...............................................................................................................24
1 Main Components
5.1 Panels
CAUTION !
Prior to working on the centrifuge disconnect mains power plug
and make sure it cannot be reconnected by mistake (e.g. by means
of a warning sign)!
• Remove rotor
• Unscrew the four screws of the metal motor cover in the middle of the rotor
chamber and remove cover
• Disconnect cables
• Unscrew the two screws on the left and right of the front panel of the housing
• Unscrew and remove bottom plate of centrifuge; all components inside the
centrifuge are now readily accessible
Front and back panels of centrifuge must be removed completely to gain access to
the electronic and mechanical components and refrigeration system.
After its removal the front panel can be inserted into the groove in the housing (see
fig. 5.2). It is now possible to work on the front part of the centrifuge and it is no
longer necessary to disconnect the cables. It is important to ensure that the front
panel rests on a firm surface (the distance between the front lower edge of the
centrifuge and the edge of the table should be at least 10 cm).
In addition it is necessary with some models to remove the housing when carrying
out other service tasks.
• Disconnect mains power plug and secure against re-connection by mistake (e.g.
by means of a warning sign)
• Remove rotor
• Unlock clip of the gas pressure spring and draw it out of the connection between
lid and gas pressure spring (see section 5.2.4.5, fig. 5.10)
• Loosen two lateral (1) and the lower (2) screw of the
front panel (see fig. 5.4)
(1)
• Loosen the two upper housing fastening screws (1) Fig. 5.4: of
Position of the screws
the front panel
at the metal angles at the front right/left (see fig. 5.5)
(1)
• Disconnect flat cable between microcontroller board
and power board
• Loosen the two recessed head screws (1) at the Fig. 5.5: Position of the housing
back panel right/left (see fig. 5.6) fastening screws
(1)
• Disconnect mains power plug and secure against re-connection by mistake (e.g.
by means of a warning sign)
• Loosen two lateral (1) and the lower (2) screw of the front panel (see fig. 5.4)
• Remove front panel in downward direction and to the front (tip centrifuge slightly
back)
• Loosen the two upper housing fastening screws (1) at the metal angles at the
front right/left (see fig. 5.5)
• Unlock clip of the gas pressure spring and draw it out of the connection between
lid and gas pressure spring (see section 5.2.4.5, fig. 5.10)
• Disconnect mains power plug and secure against re-connection by mistake (e.g.
by means of a warning sign)
• Loosen two lateral (1) and the lower (2) screw of the front panel (see fig. 5.4)
• Remove front panel in downward direction and to the front (tip centrifuge slightly
back)
• Loosen the two upper housing fastening screws (1) at the metal angles at the
front right/left (see fig. 5.5)
• Unlock clip of the gas pressure spring and draw it out of the connection between
lid and gas pressure spring (see section 5.2.4.5, fig. 5.10)
• Remove rotor
• Loosen the two recessed head screws (1) at the back panel right/left (see fig. 5.6)
5.2 Lid
CAUTION!
Release the lid lock only if the rotor does not turn and
any possible danger to other persons can be excluded!
• Drill a hole (Ø 4.8 mm) into both sides of the housing (see fig. 5.8).
• Push the screw drivers carefully in upward direction. Thus the closing spring of
the lid locks are pressed down and the locks are released
• Open lid
• Remove key pad (models 1-13, 1-15K, 2-3, 2-4, 2-5, 2-15 and 2-16) or
front panel (section 5.1)
• Disconnect power supply of lids from the power board or power supply
unit (6 pin plug respectively 4 pin plug in model 1-13; 4 pin plug and 2 pin
plug up to model 2-16)!
• Loosen screws from the top. In models 1-15K, 2-5 and 2-16 from the
front.
• All movable parts of the lock and the lid hooks must be greased sufficiently
to prevent increased wear and stiffness!
• Particular attention must be paid to the drive belt of the motor driven locks.
The belt must be dry and clean and its tension may have to be adjusted.
Note: If it is necessary to carry out an operational test of the centrifuge with the lid
open, block the locks from above by inserting a separate lid hook into
the rectangular opening in the lock. The service department at Sigma
must be consulted when working on centrifuges with a motor driven
lock.
CAUTION!
There is considerable danger for people in the vicinity of the
centrifuge if the centrifuge is operated with the lid open!
Fig. 5.9: Lid gasket (the lid gasket of model 6-15 does not correspond to the drawing)
• Loosen both screws on the lid from outside, hold pin from inside while doing so
(e.g. with pliers)
• Remove lid
• Pull out the pins which are inserted into the lid and to which the springs are
attached
• Remove springs
• Bend springs towards the outside with pliers and remove lid
• Press ends of the hinge axle together and remove together with springs
• Open lid
• Loosen both screws on the lid from outside, hold pin from inside while doing so
(e.g. with pliers)
• Remove lid
• Pull out the pins which are inserted into the lid and to which the springs are
attached.
• Remove springs
Gas pressure springs are used in these models to relieve the lid. The lower fastening
point is a metal angle at the back of the armoured rotor chamber or a mounting point
at the bottom of the centrifuge.
• Secure lid
• Open clamping device on the fastening point of the pressure spring manually
• Secure lid
1.1.1
CAUTION!
The lid is heavy due to its construction and size.
• Open lid.
• Loosen the five screws on the front edge on the top of the lid
• Pull top side of lid towards you (hold firmly on both right and left side). The lid is
clamped in tightly at the other side and therefore does not move easily. Secure
the lid by a another person, as the lid becomes lighter once the top side has been
removed and can shoot up.
• Only when the lid is closed is the motor for locking the lid supplied with power via
contact switches in the hinge.
• Microswitches on one of the hooks provide information on the “lid closed” position
Lid gasket:
• On the inside of the lid there are 12 screws holding the insulating plate in position
from above. The screws are located in two circles. Eight screws in the outer circle
and 4 in the inner circle. The inner screws can only be loosened once the excenter
wheel of the locking mechanism has been turned slightly. The insulating plate then
falls down onto the supporting rim.
Removal of lid:
Removing the lid requires a great amount of time and effort and should therefore be
avoided.
The rubber boot seals the rotor chamber against the motor. It is inserted into the
respective opening at the bottom of the rotor plate.
The connecting cable of the rotor identification which is screwed to the cover plate
(see section 7) runs through the rubber boot.
Due to the safety relevant function of the vibration dampers, special care must be
taken with respect to them during maintenance. Please observe the instructions of
section 3 and the respective accident prevention regulations in their valid versions (in
Germany according to VBG 7z).
• Rubber buffer (models 1-6, 1-13, 1-15, 1-15K, 4-15, 4K15, 6-15, 6K15 and
8K10)
The base plate of the motor is suspended freely beneath the armoured chamber. In
nearly all models this connection is dampened with rubber buffers.
Removal:
The motor in models 1-6, 1-13, 1-15, 1-15K, 1K15, 2-5, 2-16, 4-15, 4K15, 6-15, 6K15
and 8K10 is supported on vertical metal rubber elements. The only difference to be
observed is that the motor is mounted either standing on the bottom of the centrifuge
or hanging underneath the armoured chamber.
Model 3K30 is fitted with oil vibration dampers. The vertical vibration dampers are
attached to the motor base plate with a hexagonal lock nut and to the frame by a
plastic clamping device, the horizontal vibration damper are attached to the
motor/motor flange with a horizontally oriented threaded bolt and to the frame by a
plastic clamping device.
Note: The horizontal oil vibration damper is filled with thin oil and the vertical oil
vibration damper with viscous oil. Vibration dampers with thin oil are
marked with a ’3’ inside a circle. Do not mix them up despite their
identical appearance!
CAUTION!
Disconnect mains power plug and make sure it cannot be
reconnected by mistake (e.g. by means of a warning sign)!
• Remove panels: depending on model front and back panel, in some cases also
side panels (8K10) or all panels (for microphages see 5.1).
• Important! The rotor has to be removed first. Loosen the collet chuck screw and
take out the rotor. On models of the 4- and 6-range, there is a cone mounted on
the drive shaft which has to be taken off additionally. A suitable extractor tool has
to be employed for this. Another screw should be screwed back into the thread in
such a way that the extracting tool does not press into the collet chuck and widen
it.
• Unscrew the cover plate with the rotor identification signal generator and take out
the rubber boot which seals the rotor chamber at the bottom (not in all models).
• Disconnect all connections from the drive to the speed sensor board, the power
board and the microcontroller board. Mark all connections clearly prior to
disconnecting.
• Loosen the three screws on the socket of the drive from the top
• Disconnect cables
• Unscrew the 4 screws on the bottom of the centrifuge and take off bottom
plate.
• 4 screws hold the motor and the speed sensor board in place. Unscrew 3 of
the 4 screws on the bracket holding the speed sensor board and remove
speed sensor board. The last screw holds the motor in place.
• Unscrew the last screw and pull motor out through the rotor chamber.
• Unscrew the 3 respectively 6 fastening nuts with a long socket spanner between
housing and motor from above.
CAUTION!
Disconnect mains power plug and make sure it cannot be reconnected
by mistake (e.g. by means of a warning sign)!
• Remove front grid and, if necessary, the lower panel on the left-hand side (see
5.1.3)
• Take whole unit out towards the top (if necessary using hoisting equipment due
to its weight)
• Loosen the four screws of the motor at the motor flange (see fig. 5.16)
Fig. 5.16: Position of the screws of
• Remove motor through the rotor chamber the motor at the motor flange
For the imbalance switch, it is particularly important to note the position of the switch
and contact plate, respectively Hall sensor. If necessary, mark them prior to starting
work. The Hall sensor is very sensitive and must be handled with care. After
installation check the value of the static imbalance (300-400) in service mode 5
(Zent1/Zent2) respectively in the sensor mode (Zent3).
CAUTION!
Disconnect mains power plug and make sure it cannot be
reconnected by mistake (e.g. by means of a warning sign)!
• Be sure to reinstall the friction discs. They will help to prevent the rubber
dampers from being twisted during installation.
• After installation the imbalance switch has to be checked and adjusted (see
section Sensors).
• Place centrifuge on its side and adjust the motor to correct height from above.
The distance from the upper edge of the motor (end plate) to the flange of the
rubber boot has to measure 58 mm for model 2-3 and 62 mm for model 2-4.
• Make sure that the motor is mounted vertically and does not tilt.
• Tighten clip
5.4 Fuses
Maintenance-free thermal fuses are used in models 3K30, 4-15, 4K15, 6-15 and
6K15. The bimetal switch warms during overvoltage or overcurrent and cuts the
centrifuge off from the mains power supply. After the fuse has been activated it is
necessary to wait a few minutes for the bimetal to cool down before the centrifuge
can be switched back on again.
For models 1-6, 1-13, 1-15, 1-15K, 1K15, 2-3, 2-4, 2-5, 2-15, 2-16, 2K15, 3-15, 3K15
and 8K10 safety fuses are used which are located above the mains power
connection and can be replaced, if necessary.
Air
outflow Drive
Slits in the
panels
Holes in bottom plate
Onlyininmodel
Only model
6-10
6-10
Fig. 5.18: Air cooling, schematic diagram of air flow
Issue 09/98 Only in model Section 5 Page 22
Rev. 10.01.2000 6-10
Service Manual for Centrifuges
Section 5: Main Components
5.5.2.1 Compressor
Only hermetically sealed and maintenance-free compressors are used. The
maintenance of the compressors is limited to occasional cleaning of the surface and
checking of the elastic rubber elements.
There is a start relay in the terminal box of the compressor which switches on an
auxiliary winding in the compressor for starting-up. The RC component limits the
inductive voltage peaks when switching on and off. Up to model 3K18 it is located in
the terminal box. Starting with model 3K30 it is attached to the relay on the
compressor.
Both the start relay and the RC component may become defective and must
therefore be checked during maintenance work.
The lamellae can become bent by mechanical force thus hindering the air flow.
5.5.2.4 Vaporizer
The vaporizer consists of a copper tube closely wound around the rotor chamber and
soldered to the chamber for better heat conductance. The vaporizer is thermally
shielded against the ambient air by means of an insulation layer. Therefore only the
heat generated in the chamber is removed.
CAUTION!
Cooling circuit repairs should only be performed by a qualified
refrigeration engineer!
Please refer to the type plate of the compressor for information on the quantity
and type of cooling agent to be used!
6 ELECTRICAL SYSTEM........................................................................ 1
6.1 Zent1-Control Unit.................................................................................................1
6.1.1 Microcontroller Board.................................................................................... 1
6.1.1.1 Removal of Microcontroller Board............................................................. 1
6.1.1.2 Exchanging Memory Components.............................................................1
6.1.1.3 Pin Assignment.......................................................................................... 2
6.1.1.4 Control Electronics.....................................................................................4
6.1.1.5 Test Points ................................................................................................ 6
6.1.2 Power and Relay Board 8K10........................................................................6
6.1.2.1 Position, Removal and Installation.............................................................6
6.1.2.2 Pin Assignment.......................................................................................... 7
6.1.2.3 Relay Board............................................................................................... 8
6.2 Zent2-Control Unit.................................................................................................9
6.2.1 Microcontroller board.................................................................................... 9
6.2.1.1 Removal of Microcontroller board..............................................................9
6.2.1.2 Exchanging Memory Components.............................................................9
6.2.1.3 Pin Assignment ......................................................................................... 9
6.2.1.4 Control electronics................................................................................... 11
6.2.1.5 Test Points ..............................................................................................11
6.2.1.6 Jumper list ...............................................................................................12
6.2.2 Power Board Zent.........................................................................................12
6.2.2.1 Position, Removal and Installation...........................................................12
6.2.2.2 Pin Assignment........................................................................................13
6.2.2.3 Power Supply of Power board Zent.........................................................14
6.3 Zent3-Control Unit...............................................................................................15
6.3.1 Microcontroller Board..................................................................................15
6.3.1.1 Removal of Microcontroller board............................................................15
6.3.1.2 Replacing EPROM and Overspeed Controller........................................15
6.3.1.3 Pin Assignment ....................................................................................... 17
6.3.1.4 Control Electronics...................................................................................19
6.3.1.5 Function of the LEDs on the Power Board Zent3 ....................................19
6.3.2 Power board Zent ........................................................................................ 20
6.4 Control Unit Microfuges..................................................................................... 21
6.4.1 Microcontroller and Power board of Models
1-13, 2-3 and 2-4.....................................................................................................21
6.4.1.1 Jumper List...............................................................................................22
6.4.2 Control Unit of Models 1-6, 1-15, 1-15K, 2-5 and 2-16...............................23
6.4.2.1 Microcontroller Board...............................................................................23
6.4.2.2 Jumper list................................................................................................24
6.4.2.3 Power board.............................................................................................24
6.5 Control Unit Zent4...............................................................................................28
6.5.1 Microcontroller Board..................................................................................28
6.5.1.1 Removal of Microcontroller Board........................................................... 28
6.5.1.2 Pin assignment........................................................................................ 29
6.5.2 Power Board Zent4.......................................................................................29
Issue 09/98 Section 6
Rev. 19.03.2003
Service Manual for Centrifuges
Section 6: Electrical System
CAUTION!
Disconnect mains power plug prior to starting work and make sure
it cannot be reconnected by mistake
(e.g. by means of a warning sign)!
• Prior to touching the component again, equalise the potential by touching the
conductive surface on which the component rests.
• Do not bend or break a pin when inserting the component into a socket.
• Make sure the notch is in the correct position (on the left hand side).
6.1.1.3 Pin Assignment
• Microcontroller Board
The following diagram shows the pin assignment on the microcontroller board
(Layout). Table 6.1 contains the assignment of the 50 pin plug.
VS11 Overtemperature.
Heat sink
VS16 free
VS10 Imbalance
switch
VS9 Lid switch
VS8 Cooling at
standstill
VK1 free
VS15 free
VS12 Ext. keyboard
Fig. 6.2: Display board of Zent1 control unit (with pin assignment of microcontroller board)
• Assignment of 50 pin plug of Zent1 control unit
The signal directions for the microcontroller board are identical to those of the power
board and are listed below. The assignment of the plugs on the microcontroller and
power board are also identical.
>: Signal direction from microcontroller to power board
<: Signal direction from power board to microcontroller
H: 5V in relation to microcontroller ground (High)
L: 0V in relation to microcontroller ground (LOW)
If components of the microcontroller board are faulty, the whole board must be
replaced. The replacement of individual components is not advisable and also in
most cases not possible.
• The power board together with a heat sink and two fans are installed on a
supporting plate which is held in position by three screws.
• Loosen the screws at the bottom and the screw with a sheet metal bracket.
Relay board:
• The relay board is screwed to the bottom of the centrifuge and can be
removed after loosening the screws.
6.1.2.2 Pin Assignment
Circuit diagram and layout are to be found in the appendix.
VS8
VS26
VS28 VS27
The assignment of the plug and the signal directions can be taken from table 6.1.
• Pin assignment
The circuit diagram and the layout are to be found in the appendix.
The assignment of the 20-pin plug is identical to the 50 pin plug, pin 21 - pin 40 (see
table 6.1).
X/R1 Power
supply
X/R2 Fan
X/R3 Fan
X/R4 Star/
Delta
X/R5
X/R13 Lid Contactor KM
motor
VS10 free
The signal directions are listed below refer to the microcontroller. The assignment of
the plugs on the microcontroller and power board are identical.
16 Spalten 8 Spalten
Tasta- 7 Spalten Spalten I/O Port mit
Folienstecker tur-
Inter-
Tastatur face
Ansteuerung Segmenttreiber
Service
µC-Bus RS232
scr.#2
(o. HCMOS)
4 Reihen PWM Check
µ
C-Kern Inter-
face/
Ver-
PWM Aus knüp-
fung
Kammertemperatur -µ C 517A 5V Wechsel-
NTC WRF richter
UNWSTAT
- Pwr.Mon/Reset-Gen.
WRB
Unwuchtsensor - Analog Referenz
UNWDYN
UNWTEST
- EPROM 512kB 1MB Relais
- EEPROM
parallel 28 DIP
alternativ Deckel- DE Frei
Relais
seriell Verriegelung
Karte
Regel- DE Test
entsprechend ZENT1 Anpas-
sung
Übertemp.
Unwucht dig.
Deckel offen
12V
+12V Linear 15V
Regler Service
Stern/Dreieck MFR 8V
8V Pwr
Rotorcodesensor Signal2/Reset MFR
Dreh S↑ Signal1 Monoflop Über- 5V
Linear
spannungs-
5V Regler
Drehzahlsensor Drehzahl PWM aus schutz
GND
Entsprechend Schaltungs-
vorschlag Sigma
HW - Drehzahlüberwachung optional für Tests etc.
Test points can only be defined with difficulty due to the high complexity of the circuit
and the SMD technology. The signals can be measured best on the 50 pin plug
already described above. (see Table 6.2).
If components of the microcontroller board are faulty, the whole board must be
replaced. The replacement of individual components is not advisable and also in
most cases not possible.
6.2.1.6 Jumper list
The jumpers provide additional hardware protection. The centrifuge is switched off if
the maximum speed is exceeded.
In all centrifuges with a sheet steel housing, the power board is installed at the
bottom of the centrifuge. The front or the back panel has to be removed to gain
access to the board.
Caution! The electrical connections to the board should first be marked
and then disconnected.
2. The power board is attached to the heat sink by printed circuit board fastening
devices.
To remove the power board either unscrew the four screws from inside or the nuts at
the bottom. Detachment of the printed circuit board fastening devices is not
recommended.
X2 Brake resistor
X10 Motor
X4
Lid locks
X5
Fan
Electronics
X6
Fan
Refrigerator
X7
Refrigerator
X2 Brake resistor
X10 Motor
X4
Lid locks
X5
Fan
electronics
X6
Fan
refrigerator
X7
Refrigerator
X3 Mains power X1
Intermediary circuit
For measuring the intermediary circuit see Fig. 6.5.
Power supply unit
AC voltage from the transformer T1 is rectified by the rectifier V37, smoothed by C1
and stabilised to 15 Volts by the stabiliser N4. A voltage of approx. 11V is supplied to
the microcontroller board via the pins X1/37-40 and stabilised there to 5 volts by the
stabiliser N1. These 5V represent the main supply of operating voltage for the
microcontroller board and for some of the sensors.
Note: When replacing the EPROM and the overspeed controller make sure that the
correct polarity is used.
24573
6k15C / Z3
Version 1.16 / RK1
Fig. 6.11: Example
Replace EPROM and overspeed controller only together as data
contained in the EPROM and the overspeed controller have to
correspond. (Pay attention to the correct RK number, see Fig. 6.11)!
Changes in the type (to be distinguished from version) of software can erase data in
the EEPROM such as run time, cycles etc. when replacing an EPROM. This is not a
software error.
Please compare the software type and, if necessary, write down important data prior
to changing the EPROM or the software!
When replacing the EPROM and the overspeed controller please observe the
following:
Prior to touching the component again, touch the conductive surface to achieve
potential equalisation.
Do not bend or break a pin when inserting the component into the socket.
A test of the overspeed controller should be carried out after replacing the EPROM
and/or the overspeed controller. This is done using a service mode (see section
7.5.2.6).
6.3.1.3
recognition
VS3 Chamber overtemp.
VS10
The display is connected to the microcontroller board via a 20 pin flat cable. The
speed sensor is connected to the microcontroller board via a 6 pin flat cable.
Prior to each centrifuge run, a run will be simulated to check, whether the overspeed
controller is working correctly. This is done by first testing with the maximum
permissible speed for the rotor in use minus 100 revolutions per minute and then by
using a maximum permissible speed plus 100 revolutions per minute. If the
overspeed controller reacts correctly, then the centrifuge run can be started. If it
reacts incorrectly, an error message is displayed and the centrifuge cannot be
operated.
All LEDs off, overcurrent in the intermediate: Occurs during braking, if the brake
resistor is interrupted.
6.3.2 Power board Zent
See 6.2.2.
6.4 Control Unit Microfuges
The board situated behind the keyboard on the front of the centrifuge can be
removed towards the front after undoing the four screws visible from the front.
Motor
Brake
resistor
Lid coil
and mains
power
Speed
sensor
It is essential to observe the positions of the jumpers when replacing the control
electronics (see marking on the board):
Centrifuge J1 J2 J3 J4
1-13 X X
2-3 X X
2-4 X X X X
6.4.2 Control Unit of Models 1-6, 1-15, 1-15K, 2-5 and 2-16
The microcontroller board is situated behind the liquid crystal display on the front of
the centrifuge. In order to remove it, first take off the panel (see 5.1.2) and then
loosen the screws holding the board.
CAUTION!
Prior to working on the electronics disconnect mains power plug
and make sure it cannot be reconnected by mistake!
It is necessary to place the board on a clean, anti static support when working on it.
Connection to the
power board
Jumper
Fig. 6.14: Microcontroller board, here: 1-15, V2.6 (not existing in 1-15K
and 2-16)
6.4.2.2 Jumper list
It is essential to observe the positions of the jumpers when replacing the control
electronics (see marking on the board).
Centrifuge J1 J2 J3 J4
1-6 (from SW-version 3.4) X X
1-6 PFC (from SW-version 3.5) X
1-15 (from SW-version 3.5) X X
1-15 PFC (from SW-version 3.5) X
2-5 (with power board 70667/70668, from SW-version 3.1 – 3.4)
2-5 PFC (with power board 70740, from SW-version 3.5 – 3.7) X
2-5 230V / 120V (with power board 70667/70668, from SW-version 3.5) X
2-5 230V /120V (with power board 70850/70851, from SW-version 3.5) X
The power board is situated in the front part of the housing behind the panels and is
fastened with two screws and circuit board clips. It is necessary to remove the panel
when working on the board (see 5.1.2).
CAUTION!
Prior to working on the electronics disconnect mains power plug
and make sure it cannot be reconnected by mistake!
It is necessary to place the board on a clean, anti static support when working on it.
Pin assignment Power board of Models 1-6 and 1-15
Fig. 6.15: Power board
Speed Microcontroller
sensor board
Mains power
Lid
Brake
resistor
Motor
Pin assignment Power board of Models 1-6 from construction year 2000
and 1-15 from serial no. 71140
Micro-
Speed Microswitch
controller
sensor lid
board
Lid lock 1
Mains
power
Lid lock 2
X3 Microcontroller-
X2 Tacho- board
generator
X1 no X7 Brake
function resistor
Ground
Refrigerator
Fan
refrigerator
Mains power
Lid lock 1
Lid lock 2
Motor Brake
resistor
Fig. 6.18: Power board
6.5 Control Unit Zent4
CAUTION!
Prior to working on the electronics disconnect mains power plug
and make sure it cannot be reconnected by mistake!
It is necessary to place the board on a clean, anti static support when working on it.
6.5.1.2 Pin assignment
X4 Serial interface
X5 Chamber overtemperature
X6 Motor overtemperature
X8 Chamber temperature
X7 Imbalance sensor
X3 Right encoder
X1 Power board
X2 Left encoder
The boards are fastened with screws on distance bolts and with touch panel holders.
In order to remove the board, loosen the screws. Press the locking pins of the
holders and remove the board. Install in reverse order.
The power board in models 3-16 and 3-16K is plugged onto a cooling plate. This
plate is screwed on four distance bolts which are screwed at the bottom of the
centrifuge. In order to remove the board, loosen the two screws in the rear and
remove the screws in the front. Do not loosen the touch panel holders as the IGBT
module and the cooling plate are screwed together.
X4 Temperature sensor
X3 Rotor code sensor
K2 Relay refrigerator
X1 Imbalance switch
X6 Microswitch lid
X7 Microswitch lid
X2 Speed sensor
K1 Relay fan
X8 Fan
X9
Refrigerator
X10
Power input
X1 Microcontroller board
X6 Fan
refrigerator
X7 Refrigerator
X1 Rechnerplatte
X5/5 = GND
X5/6 = VCC
X5/7 = GND
X5/8 = VCC
X5/9 = GND
Assignment microcontroller board Zent4, P/N 70796, 70797, 70798, 70799, 70800:
X1 = Serial interface, power board
X2 = Encoder "select"
X3 = Encoder "value"
X4 = Serial update plug
X5 = Chamber overtemperature
X6 = Motor overtemperature
X7 = Imbalance switch
X8 = Temperature sensor
H3 orange brake transistor off: Extingiushes when the brake resistor is switched on.
Is flaring during braking.
A faulty cable or a short circuit in the sensor are displayed as ERROR 90, in models
1-15K, 2-16, 2-16K, 2-16KC, 3-16, 3-16K as ERROR 13.
7.1.2.2 Motor
The motor of models 3-16 and 3-16K has a temperature switch that protects the
motor from overheating. The temperature switch opens at approx. 140°C and the
microcontroller displays ERROR 26.
There is a temperature switch attached to the cooling coil for additional protection
against dangerous overtemperatures, e.g. when there is a fault in the electronic
temperature control unit. A bimetallic switch is used which is located inside the foam
of the insulating layer between the armoured chamber and the rotor chamber.
The switch is connected to the microcontroller board via the plug VS3. If the
permissible temperature is exceeded, ERROR 88 is displayed, in model 3-16K
ERROR 24.
It is not possible to replace this temperature switch if it gets damaged. Contact Sigma
directly to discuss instructions for a possible repair.
Small bar magnets of alternating polarity are located in the rotor at a distance of 0.8 -
2 mm in front of the Hall sensor. There are a minimum of 2 and a maximum of 12
magnets located in a circle which is divided into sectors of 30°. The number of
magnets is always even. The rotor code depends on the angular position and
number of magnets (see service mode No.1 for Zent1/Zent2 respectively sensor
mode for Zent3 control unit).
In models 1-6, 1-13, 1-15, 1-15K, 1K15*, 2-3, 2-4, 2-5, 2-15*, 2-16, 2K15*, 2-16K,
3-15*, 3-16, 3K15*, 3-16K, 3K18*, 3-18K and 3K30*, the plastic casing containing the
Hall sensor is mounted on a small circuit board which is located below the motor. In
models marked with a `*’, the sensor together with a small circuit board and passive
components is encapsulated in a small plastic housing as protection against shock
and humidity. This unit cannot be repaired, only replaced. One plug connection leads
Issue 09/98 Section 7 Page 5
Rev. 23.02.2005
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
to the rotor detection, the other to the power board. Both plug connections are equal
i. e. they can be interchanged without influencing their function.
The connection in all other models is of a permanent nature. If the speed sensor is
defective, the whole speed sensor board should be replaced.
Disconnect cables.
2 Disconnect cables.
Window in Lid
For models 1-6, 1-13, 1-15, 1-15K, 2-5 and 2-6:
Start centrifuge
Fix the lid in the openings of the gas pressure dampening elements using the locking
bars in the back. The lid hooks must be inserted into the center of the openings of the
lock. This is important as otherwise the optical hook detection does not recognise
the lid, leading to error messages.
Switch on centrifuge and press the testing lid downwards with both hands so that the
motor driven locks can close
Start centrifuge
Check the speed using a stroboscope through the viewing window in the testing lid.
After checking, the lid can be opened using the lid key. Dismantle in reverse order.
Service Manual for Centrifuges
Section 7: Sensors, Service mode and Adjustments
Start centrifuge.
Contact plate,motor
Attention:
Fastening bracket
Fig.7.7: Mechanical imbalance switch
Adjustment: After removing the front panel, a brass screw (1) becomes visible. This
screw is used to adjust the imbalance cut-off threshold.
The brass screw should be turned carefully until it touches the contact plate.
For model 1K15: Retract the screw by turning it between 1/2 and 3/4 of a revolution
in the other direction.
Check setting by carrying out a test run using a switch-off imbalance (see table 7.2).
Service Manual for Centrifuges
Section 7: Sensors, Service mode and Adjustments
Model 1-15K
To adjust the imbalance switch, loosen the six screws of the back panel. After
removal of the back panel, the imbalance switch is accessible.
Retract the screw by turning it between 1/2 and 3/4 of a revolution anti clockwise.
Check setting by carrying out a test run using a switch-off imbalance (see table 7.2).
Model 2-15
The imbalance switch can be adjusted from the outside without removing the
housing. The opening for this is located at the lower part of the panel on the right-
hand side.
Remove cover plug.
Check setting by carrying out a test run using a switch-off imbalance (see table 7.2).
Model 2-16
To adjust the imbalance switch, the front panel must be removed (see 5.1.8). The
adjusting screw is behind the second opening from the right side at the lower part of
the front panel.
Turn screw carefully until it touches the contact plate.
Check setting by carrying out a test run using a switch-off imbalance (see table 7.2).
Service Manual for Centrifuges
Section 7: Sensors, Service mode and Adjustments
Zent1/Zent2-control unit:
Adjustment:
The value of the static imbalance must be between 300 and 400. In case of
deviations:
Adjust grey wedge in the middle of the light barrier. The potentiometer
on the circuit board (see diagram) of the grey wedge is in a middle
position (Notch respectively flattened part point vertically upwards.) Fig.7.9:
Grey wedge circuit board
Set a value of 350 by turning the potentiometer.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Table 7.2: Settings of imbalance values. We reserve the right to technical modifications!
The imbalance values in the above table, both for cut-off imbalance and max.
permissible imbalance at critical speed, depend on the rotor type and mass. Water
filled centrifuge glasses are suitable to simulate an imbalance (1 ml ≈ 1g).
Special rotors:
The learn run for special rotors must be done with particular weights.
The service mode is switched on by pressing the ← key once and then the ENTER
key three times, holding it for more than three seconds the third time. The centrifuge
indicates that the service mode has been activated by displaying ‘SER’ in the time
display, the menu number in the program display and the menu options as full text in
the speed display.
By pressing EDIT, the normal operating mode is reactivated.
Important: In order to follow the instructions given below correctly, you should first
read out the number of the software version by using service mode No. 0. The table
„Menu options in the service mode of the Zent control units“ should be observed!
The software version can also be determined by pressing the key „FSTOP“ before or
up to three seconds after switching the centrifuge on.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Menu options
No. SWV Text Description
0 ≥1.78 ver Display of version number and system information
1 >1.79 offset
1 ≥1.63 sen Display of sensor values
Set offset value of temperature sensor after pressing the ENTER
key
2 ≥1.75 i-fak DETERMINATION OF IMBALANCE FACTOR (FROM SW VERSION 1.75 ONWARD)
3 error Display/deletion of error table
4 eep-- Erase EEPROM and ‘lid lock’
5 ≥1.75 imb w Parameter setting run for measuring imbalance warning
limit/Display of static imbalance
6 ≥1.75 imb f From this software version onwards a parameter setting run for
measuring the imbalance cut-off limit should no longer be
carried out ⇒ Carry out service mode No. 2 instead.
7 led88 Test of LEDs
8 time Operating hour counter
9 Cycle Switching-on of cyclic operation; display/deletion of cycle
counter
Table 7.3: Menu options of service mode of Zent1 and Zent2 control units
Procedure:
• Press ← key and then ENTER key three times; hold ENTER key when
pressing for the third time.
• Use ↑ and ↓ keys to select the required function and activate by pressing
ENTER key. If the chosen menu option has a substructure, it is possible to return to
the main menu by pressing EDIT.
The version sub-number is valid for every centrifuge type. Its number is increased in
the current software version after general modifications. The version number of the
centrifuge is type specific and is increased in the centrifuge configuration file after
every modification.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Display
Note! Service mode No. 1 should be used to check the sensors of the centrifuge.
This is achieved by changing the output signals of the sensors, for instance,
by opening and closing the lid ⇒ Signal switches 1-0-1-0…
The rotor must be turned manually in order to check the imbalance sensor, the
rotor code sensor and the monoflop sensor. If the sensors are in order, the
numbers in the displays alternate 0-1-0-1…
Caution! This correction value is valid for all rotors of the type of centrifuge in
use!
The following table shows the conversion of correction values into their
corresponding correction factor using the formula 1/(1+x) for negative and 1+(x/10)
for positive values:
Proceed as follows:
• Select suitable rotor and load with test weight (see table 7.2).
• Press ENTER in service mode No. 2 until imbalance LED lights up.
• The acceleration curve 19, the maximum speed of the rotor used and a run
time of 10 minutes are used automatically after activating the parameter
setting run under point to of the service mode. The run can be started by
pressing START without leaving the service mode.
• After the run the imbalance LED is turned off and the calculated factor is
stored in the EEPROM and is displayed.
• Do not erase EEPROM completely after this service mode has been carried
out as the imbalance correction value would be lost in this case. If the
EEPROM has been erased accidentally, enter the correction values in point 2
of the service mode by pressing the keys ‘↑‘ and ‘↓‘.
• Carry out a test run with cut-off weights to verify the correction value that has
been determined.
The value can be set manually in the service mode if the customer finds the
determined value to be too sensitive/insensitive. The above table should be observed
and the value altered accordingly: positive (less sensitive) and negative (more
sensitive).
Proceed as follows:
• Activate service mode and select menu No. 2
• Press ENTER once to activate the manual input mode
• Measured correction value flashes in temperature display
• Set desired correction value by using ↑ and ↓ keys
• Confirm and leave service mode by pressing EDIT twice
Note: Starting with SW version 1.78 the errors 79, 98 and 99 are no longer stored in
the EEPROM.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Display
Possible options
The user data in the EEPROM, which include programs, user defined curves (only in
ZENT1) and stored error messages, are erased by pressing ENTER. Total run time,
cycle counter, temperature offset and imbalance values are not erased.
All EEPROM data (except total run time and the cycle counter) are erased by
pressing [←]. These two values can be erased by using the options on the
corresponding menu in the service mode.
The warning and cut-off imbalance limits are also erased and substituted by the
default values in the EPROM.
If the speed signal fails during a run or the centrifuge starts up without a speed
signal, errors 28 or 29 appear on the display. In this case the control unit of the
centrifuge does not know whether or how fast the rotor is turning. To prevent the
operator from opening the lid while the rotor is turning, the lid is locked for a set time.
The error is stored even if the centrifuge has been switched off. After switching the
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
centrifuge back on, error 28 is displayed immediately and remains until the set
waiting period, during which the centrifuge must stay switched on, has passed. The
waiting time is set to 12 minutes for the centrifuge model ‘1K15’.
Possible options
The EEPROM is initialised when the main microcontroller board is switched on for
the first time. The following data are transferred from the EPROM to the EEPROM:
• Centrifuge type
These three identification numbers in the EPROM and EEPROM are compared every
time the centrifuge is switched on.
In all other cases the error code ERROR 69/70 is used to indicate that the content of
the EEPROM has been altered. This message can be cancelled by pressing any key.
The imbalance limits determine at which imbalance intensity the warning LED lights
up, respectively when the centrifuge is stopped by means of a quick stop. The
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
EPROM contains imbalance default values for every type of centrifuge and group of
rotors. These default values are appropriate for standard operations.
If, however, other limits are required, these can be determined by means of
parameter setting runs and stored in the EEPROM. The default values in the EPROM
are then no longer used. By erasing all the EEPROM data, the limits set in the
parameter setting runs are erased and the original default values used again.
See service mode No. 2 for instructions on conducting a parameter setting run.
When a parameter setting run has been carried out, the values in the
EEPROM should be erased and a parameter setting run conducted
according to service mode No. 2!
Display
Possible options
The number of cycles recorded on the cycle counter since it was erased last is
displayed in the speed display. The cycle counter is stored permanently in the
EEPROM and updated after the completion of every run.
The time display shows the waiting time between stopping and restarting the
centrifuge in minutes. Exception: If the display shows 0, the waiting time is 10 sec. By
pressing the ↑ and ↓ keys the waiting time can be altered. The key [←] erases the
cycle counter. By pressing the START key the cyclic operation is activated and the
centrifuge started. Press S-STOP to leave cyclic operation.
Should an error occur under cyclic operation, the control unit of the centrifuge checks
by using a table, whether further operation of the centrifuge is possible. If this is the
case, the error is stored in the EEPROM and the centrifuge continues cyclic
operation.
Display
Possible options
Important! To be able to follow the instructions below correctly, you should first
read out the software version using the menu option configuration
„Info“.
• Select the configuration mode (→) by turning the control knob and activate the
mode by pressing the control knob.
• Select the menu „Info“ by turning the control knob and activate by pressing the
control knob.
→ The number of the software version can be found under point „Zent3
Version“.
By entering the code the service mode is activated. Please contact the service
department of Sigma if the code for activating the service mode has been lost.
In order to go to page 2 of the service mode, select the display field „page 2“ by
turning the control knob, press the control knob.
• Select display field „Leave service mode“ by turning the control knob.
Service functions
By entering the user code the user can store the centrifuge functions „Store
programme“, „Modify parameters“, „Load programme“ and lock the ‘Start’ key.
Call the service of Sigma, if the code to lift the lock has been lost.
By entering this code the stored programmes, contrast, the configuration setting
and the user defined curves are erased. After entering this code the centrifuge
returns to its factory setting.
• Turn control knob from this position clockwise (right) by one segment
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
• Turn control knob from this position anti clockwise (left) by one segment
Designation Description
Page 1
Imbalance Adjustment Correction value for imbalance
Learn run Determination of imbalance factors
Temp Offset Setting of offset value for temperature
sensor
EEPROM Delete time + cycles Erase total run time and cycle numbers
stored in the EEPROM
Attention! The correction factor is valid for all rotors for this type of centrifuge!
The following table shows the conversion of the correction values into their
corresponding correction factors (for negative values 1/(1+x) and for positive values
1+(x/10):
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Proceed as follows:
• Select display field „Learn run“ by turning and then pressing the control knob →
dialog window appears.
• Select display field „Yes“ by turning and then pressing the control knob → dialog
window appears.
• Close lid.
→ The centrifuge accelerates to its maximum speed and then slows down again.
During this run the correction value is determined and displayed under menu option
„adjustment“.
Note: There must not be any vibrations during the acceleration phase of the
parameter setting run. If in doubt repeat the parameter setting run.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Do not erase the EEPROM completely after the imbalance correction value
has been determined, as otherwise the imbalance correction value will be lost.
If the EEPROM has been erased by mistake, the correction value must be
entered again.
The value can be corrected in the service mode manually if the user considers the
measured value to be too sensitive or insensitive. The above table should be
observed and the value altered accordingly to more positive (less sensitive) or more
negative (more sensitive) values.
Proceed as follows:
• Select display field „Adjustment“ and activate the manual input mode by pressing
the control knob (SET lights up).
• Set the imbalance correction value by turning the control knob. Values between
-30 and +30 can be set.
• Select display field „Offset“ by turning the control knob and then activate the
manual input mode by pressing the control knob (SET lights).
• Set offset value by turning the control knob. The offset value is increased in steps
of 1/10 K.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Note: A positive offset value decreases the temperature of the chamber and a
negative value increases the temperature.
This menu option allows deletion of the total run time and number of cycles since the
last reset.
Note: The menu option configuration „Info“ (see section 7.5.2) provides data on the
total run time „Runtime“ and number of cycles „Cycles“ since the last reset.
This menu option allows all data stored in the EEPROM except total run time and
number of cycles to be deleted. The data includes user data, programmes, setting of
contrasts, configuration of system, user defined curves and user codes as well as
service data, temperature offset and imbalance correction value.
• Select display field „Delete time + cycles“ or „Delete“ by turning the control knob
and pressing it → dialog window appears.
• Select „Yes“ by turning the control knob and acknowledge by pressing the control
knob.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
The menu option „Display list“ shows an overview of all the errors that have been
stored in the EEPROM. As many as 14 error numbers are stored. If more than 14
errors have occurred, the oldest entry is overwritten.
The menu option „Delete“ allows deletion of all errors stored in the EEPROM.
• Select display field „Delete“ by turning the control knob and pressing it → dialog
window appears.
• Select „Yes“ by turning the control knob and acknowledge by pressing the control
knob.
Using the menu option „Rest time“ allows the user to set the time in h:min:sec
between the rotor coming to a standstill and re-starting it:
• Select display field „Standstill“ by turning the control knob and activate the manual
input mode by pressing the control knob (SET lights up).
• Set the desired value by turning the control knob and enter the value by pressing
the control knob.
• Select „Start“ by turning the control knob and pressing it → dialog window
appears.
• Select „Yes“ by turning the control knob and acknowledge by pressing the control
knob.
During the rest time that has been set the key „Start“ flashes and the count down of
the remaining rest time is to be seen in the time display.
The run can be started or stopped ahead of time without ending the mode for cyclic
operation.
• Press ‘Stop’ key for more than 1 sec. → dialog window appears.
All rotors which can be used for a certain type of centrifuge are stored in the
overspeed controller. It monitors the speed of the rotor independently from the main
controller and switches off the drive system once the maximum speed which is
permissible for the rotor has been exceeded.
Prior to each centrifuge run, a run will be simulated to check, whether the overspeed
controller works correctly (see section 6.3.1.4). If the overspeed controller reacts
incorrectly, an error message is displayed and operation of the centrifuge is
impossible.
A test of the overspeed controller should be carried out after replacing the
EPROM and/or the overspeed controller as both components must match each
other.
During this test the data of all rotors which can be used together with a certain type of
centrifuge are checked using simulated runs This is done by first testing with the
maximum permissible speed of the rotor used minus 100 revolutions per minute and
then by using a maximum permissible speed plus 100 revolutions per minute.
Proceed as follows:
• Select the display field „Test controller“ by turning the control knob and activate
control unit by pressing the control knob.
After the end of the test another dialog window appears depending on the test result:
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
12 sensor values belonging to two categories are indicated in the sensor mode:
This allows an immediate diagnoses and fast repair of faults should there be a
malfunction during operation. Values can neither be entered nor modified.
SPEED (digital)
Speed signal
ROTORID (digital)
Rotor code
MFR (digital)
(Monoflop signal)
1: Rotor is turning
0: Rotor is stationary
The MFR signal switches from 0 to 1 after the 1→ 0 slope of the speed
signal. Approx. 2 seconds later the signal returns to 0.
WRB (digital)
1: ready
0: not ready
DOOR (digital)
Lid switch
1: Lid closed
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
0: Lid open
HOOK (analogue)
RES2 (analogue)
RES4 (analogue)
IMB-STA (analogue)
IMB-DYN (analogue)
During standstill the value should be above 1000. Shake the rotor if an
imbalance malfunction has occurred. The value should become
< 1000 for a short period of time and return to a value > 1000 when the
rotor is again stationary.
IMB-DIG (digital)
Note This service mode is intended to check the sensors of the centrifuge.
To do so, the signals must be altered, e.g. close/open lid → signal
changes 1-0-1-0...
To check the imbalance sensor, the rotor code sensor, the speed
sensor and the monoflop sensor, the rotor must be turned manually. If
a sensor works correctly, the display alternates 0-1-0-1...
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
7.5.4.1 Outline
With the service mode of Zent4 following settings and trouble shooting are possible:
Close the lid. Press the “Lid open” key and hold it before and during switching on the
centrifuge. The display will show following data:
CentrifugeZentrifugentyp
model SW-Bestellnummer
SW order number
Right-justified in the “Speed” field, marked as “Pb” like “Power Board”. This is
displayed with a lag of approx. 1 second. Example: “27” means software version 2.7.
Centrifuge model
The centrifuge model is displayed in the lower field “Rotor”, e.g. “2-16K”.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
The order number (part number) of the software is shown on the right-hand side in
the “rcf” field. It is marked as “od no” like “Order No.” In case of software updates the
new software file must have the same order number.
Quit the “Display Software Information” by pressing the “Stop” or “Lid” key.
The status of the digital sensors and switches can be checked out in the sensor
mode. But only the sensors connected directly to the control board and the speed
and rotor signals are displayed. The sensors connected to the power board are not
indicated.
Proceed as follows:
Rotor code
Rotor code sensor (over X1)
Motor overtemperature X6
Speed sensor (over X1)
Chamber overtemperature X5
Imbalance sensor X7
Rotorcode:
Measured rotor code for checking the magnet coding of the inserted rotor.
Motor overtemperature:
Status of the overtemperature sensor at the motor (only in 3-16/3-16K from SW 2.1)
0 = 0V at X6/2 = no error message
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Chamber overtemperature:
Status of the overtemperature sensor in the chamber (only in 3-16K)
0 = 0V at X5/2 = no error message
1 = 5V at X5/2 = overtemperature or no sensor connected (Error 25)
Imbalance sensor:
Status of the mechanical imbalance switch. Even short imbalance impulses are
stored for at least 100 ms and output.
0 = 5V at X7/2 = imbalance switch open or not connected
1 = 0V at X7/2 = imbalance impulse stored
This display only works with the imbalance sensor that is directly connected to the
control board, only in models 3-16/3-16K with power board 70783/70784.
The time-, speed- and temperature scale keeps its function and the centrifuge can be
operated. To quit the sensor mode “digital inputs and outputs”, the centrifuge must be
switched off and on again.
To quit the sensor mode “exact temperature display”, the centrifuge must be
switched off and on again.
Proceed as follows:
The offset value is displayed in the rcf-field (blinking) and can be changed by turning
the -/+ turnign knob.
By pressing any key the offset value is stored and the enter mode is left.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
The Zent4 electronics have an internal cycle counter that is increased after every
completed run, and an error list in which the last 30 malfunctions together with cycle
status and used rotor are stored.
To see the cycle status and the error list, proceed as follows:
First, the cycle status ist displayed, marked as “CYCLE” in the speed-field. The error
list is marked as “ErrLi”.
The cycles are displayed right-justified in the set value fields “rcf” and “speed”.
Having a cycle number with more than 5 digits, the upper speed field is used for the
higher-order digits.
By turning the right knob and after displaying the cycle status, the last 30
malfunctions – if existing – are displayed. The order of malfunctions is shown in the
time field (0=latest malfunction, 30=oldest malfunction). In case of overrun, the list is
updated according to the FIFO principle. The oldest error is overwritten.
The ascertained number of cycles at the time of malfunction is displayed in the upper
set value field, as described under 7.5.4.6.1. The assigned rotor is displayed in the
rotor field and the error number in the rcf-field.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Example: In this figure the last malfunction, Err12, is displayed. It occured at cycle
no. 267 with rotor 11122.
You can swap the list with the “Lid” and “Stop” key, too. By pressing the “Start” key,
the error list is left.
This mode is for internal parameter optimization and error diagnostics and is for
information only. For activation, proceed as follows:
After activation, you can switch between normal display and display of motor
parameter (level 1) by pressing the inactive “Lid” key. In centrifuges having the
4 Amp power board, further data can be displayed in the 2. level. The current level is
signaled in the temperature field and can be changed by pressing the “Lid” key. To
quit this mode, the centrifuge must be switched off and on again.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
ATTENTION: Temperature sensor offset, cycle list and error list are deleted!
1
Only in centrifuges with 4A power electronics 70783/70784 (3-16/3-16K)
2
Old limit value=380 in 3-16 TypVer.≤ 9 and 3-16K TypVer. ≤ 14.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
The lock mode saves the set centrifugation parameter from unintentional or malicious
changes.
Open the lid, press the “Start” key three times and hold it when pressing it for the
third time. Now the “Lock” symbol is blinking and the centrifuge parameter cannot be
changed directly using the encoders. The softstart/softstop settings cannot be
changed, either.
Deactivation of the lock function is carried out in the same way as its activation.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Cycle
Error number status
Order of
errors
The lower speed scale indicates the order of errors (0 ≙ latest error, 14 ≙ oldest
error in 2-16 resp. 13 ≙ oldest error in 1-15K), and the upper speed scale shows the
ascertained cycle status in binary notation.
- By pressing the ‛Lid‛ or ‛Stop‛ key it is possible to select the different errors in the
list.
If "15" in 2-16 resp. "14" in 1-15K is selected, the upper speed scale shows the
current cycle status only.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
Example: The upper scale (fig. 7.18) shows this binary value: 1000011.
The decimal value is calculated as follows:
1∗26 + 0 ∗ 25 + 0 ∗ 24 + 0 ∗ 23 + 0 ∗ 22 + 1 ∗ 21 + 1 ∗ 20 = 64 + 2 +1 = 67 cycles
i.e. after 67 cycles error no. 6 occured. This was the latest error as the lower scale
indicates "0".
Monoflop signal
Refrigerator
Fan
Monoflop signal
Press the ‛Start‛ key 3 times and hold it when pressing it for the third time.
Moreover, the set value segment is blinking in this mode as long as the pre-selected
temperature is not reached. The set value segment blinks slower than the segment of
the odd actual value temperature.
By switching the centrifuge off and on again the sensor mode is left.
Press the ‛Start‛ key 4 times and hold it when pressing it for the fourth time.
By pressing any key the offset value is stored and the entering mode is left.
Service Manual for Centrifuges
Section 7: Sensors, Service Mode and Adjustments
RETROFITTING..........................................................................................................1
8.1 General Remarks...................................................................................................1
8.2 Retrofitting which should, without exception, be carried out during the visit
of a service engineer (unless previously carried out) .................................7
8.4 Retrofitting which should, without exception, be carried out during visit of a
service engineer when using diverse rotors (unless previously carried
out)...................................................................................................................23
Retrofitting
8.1 General Remarks
It is necessary to check that the ground connection to the front panel is in good
working order to prevent drifting of the microprocessor as a result of static
discharging. Otherwise a destruction of an IGBT module could occur as a result.
Therefore never operate the centrifuge without a proper ground connection to the
front panel!
Applicable for: Old power boards (70580, 70581) of models 1-6 and 1-15
Cause: The lid lock plugs on the power board were changed from one 6-
channel connection to two 4-channel connection.
Corrective Measures:
Connect the existing lid locks to the new power board using an
adapter plug (70751).
X2 Speed sensor
X3 Control board
X8 Power supply
X5 Lid lock
X6 Brake resistor
X2 Speed sensor
X4 + X5 free
X5 Lid locks
X8 Power supply
NEW VERSION
Cause: The lid hook is not aligned vertically to the lower side of the lid.
Corrective Measures:
Adjust the lid lock again.
- Open lid, power switch OFF, remove the front panel (see 5.1.4)
- Remove the lid hook using a 17er socket wrench
- Remove lacquer residues in the cut-out, if necessary
- Align the hook vertically to the lower side of the lid. Replace the hook disc with
the modified one (part no. 25694).
- Loosen the two screws of the lid lock from above (see fig. 8.1.2) and fasten
them so tight again that adjustment is possible.
- For adjustment, press down the lid on the threaded spindle of the lid lock.
- Align the lock in a way that its threaded spindle is gripping in the thread of the
hook without canting.
- Put the socket for emergency lid release through the boring provided for this
release (see fig. 8.1.3)
- If the socket is gripping smoothly into the hexagon socket, open the lid and
tighten the screws of the lid. Otherwise, adjust again.
Side view with opened lid Top view with opened lid
Fig. 8.1.1: vertical position of the hook to the lower side of the lid
Fig 8.1.2: Fastening screws of lid lock Fig. 8.1.3: Socket for emergency lid release
Notice
All moving parts of the lid lock (see fig.
8.1.4) and the inside thread of the hook
should be cleaned and greased
regularly every month.
Parts to be greased
Fig. 8.1.4: Motor of lid lock
Functional Characteristics:
When the lid is open, the lid switches S02, S03, S04 and S05 are not active. There is
no voltage impressed to the power electronics. The optocoupler U8 does not have
voltage at its input and signalizes to the control board that the lid is open. The switch
S06 is active.
Lid Closing:
When closing the lid, it contacts the motorized lid lock. The lock shuts down the lid.
The switch S06 opens and signalizes to the control board that the process "Lid
closing" has begun. Before the lid is closed completely, the switches S02 and S03 are
activated. This impresses voltage to the intermediate circuit of the power board. The
control board gets the signal "Voltage impressed to the Power Board" via optocoupler
U8. After further closing, the switch S05 is activated and signalizes "Lid is closed".
After that, the switch S05 is activated and interrupts the circuit of the lid motor. The
centrifuge can be started.
The control board monitors the closing time of the lid via the switches S06 and S05. If
the time limit is exceeded, error message "ERROR 78" occurs and the lid re-opens.
During state "Lid closed" the plausibility of the switches S02, S03, S05 and S06 are
checked. When one of these switches is activated falsely, error message "ERROR 78"
is shown.
Lid Opening:
The models 2-16KC, 3-18 and 3-18K have following electronical safety devices that
control opening of the lid.
The microcontroller monitors the closing time of the lid via the switches S05 and S06.
If the time limit is exceeded, error message "ERROR 83" occurs.
When opening is completed correctly, the plausibility of the states of switches S02,
S03 and S05 are checked. In case of malfunction, "ERROR 83" is displayed.
Problem: Lid does not open after pressing the lid key
ERROR 78, ERROR 83
Cause: The hook is not aligned correctly and not screwed in deep enough. S04
switches before S05. "ERROR 78" is shown.
The hook is not aligned correctly and screwed in too deep. The closing force is
too high. S04 does not switch. Thus, the motor does not switch off and
the PTC-resistor overheats. The lid lock motor loses power.
The operating plate does not operate microswitch S04. Thus, the motor does
not switch off and the PTC-resistor overheats. The lid lock motor loses
power.
Overheating of the PTC-resistor caused by stiff threaded spindle in the lid hook.
The lid lock motor loses power.
Smooth running of the threaded spindle in the hook. If necessary, grease the
thread.
In case of erosion at the hook and threaded spindle, exchange the hook and the
lid lock.
Corrective Measures:
Adjust the hook. The operating plate must operate the microswitch S04 when
the lid is closed.
Operating plate
PTC
Applicable for: 3K30, 4-15, 4K15, 6-15 and 6K15 up to serial No. 61466
Cause: Overheating
Corrective Measures:
- Connect adapter "Wiring of fan, article No. 70646" to the AC
voltage side of the rectifier according to the wiring diagram (see
Fig. 8.1).
R58
Fig. 8.1
The temperature switch (see Fig. 8.2) is now no longer in use.
Temperature switch
Fig. 8.2: Position of temperature switch on the cooling plate of the power board
Cause: In refrigerated centrifuges, the fan in the top of the housing starts up
at the same time as the fan for the compressor.
Corrective measure:
Radio interference suppression of fan for compressor by means of
a varistor (solder on across the fan plug, see Fig. 8.3)
Varistor
8.2.3
Replacing the Brake Resistor
Corrective measure:
Replace brake resistor at plug X2 on the power board using the
retrofit kit (Order No. 70659) with two brake resistors (see diagram).
The existing hole can be used to fasten the new brake resistors of
the retrofit kit. Position: right hand side in the back of the centrifuge
at the gas pressure damper fastening.
Corrective measure:
Replace brake resistor at plug X2 on the power board using the
retrofit kit (Order No. 70652) with two brake resistors (see diagram).
The existing holes in the chamber can be used to fasten the new
brake resistors of the retrofit kit.
Connection between the outer soldering point of J19 and the Pin5 of U5
(7702AC) which is best accessible via the soldering point near the processor
U4.
Connection between soldering point J20 which is nearer to the fastening screw
and Pin6 of U5.
These two connections are reducing the influence of trouble signals when switching
the centrifuge on/off or in case of voltage drops.
Applicable for: 3K30, 4-15, 4K15, 6-15 and 6K15 up to serial no. 66420
Corrective measure:
Interference suppression with interference suppression kit
part no. 70661
Ground connection
3K30
5. Add a ferrite bead near the plug of the motor power cable.
Applicable for: Centrifuges with Zent2 and Zent3 control that are produced
before September 2001
Corrective measure:
The cable lines “1” and “2” – as marked in the figures below –
must be laid separately. The cable line “1” contains the
connection cables of the tachogenerator and the rotor
identification.
1
1
2 2
Software type
ERROR 42 occurs after Zent3 The build up of the display Upgrade with new EPROM with
dialogue “Cut / permit Up to blocks the hardware test. software version 1.16
curve” software
version
If ‘Start‘ and ‘Stop‘ are 1.15 Software related causes
pressed alternatively (3K30,
very quickly, the 4-15,
centrifuge accelerates 4K15,
up to its nominal speed; 6-15,
Pressing ‚Stop‘ does not 6K15)
prevent this from
happening any more.
Damage to motor 4-15, Motor is subject to increased Installation of an bearing
(corrosion, noise when 4K15, wear and tear when operated protection disc on the upper
running, etc.) in 6-15, under high humidity bearing flange of the motor (Part
installation sites and 6K15 No. 24752, see Fig. 8.9.1 and
areas with high humidity 8.9.2). Grease the upper bearing
before.
Error 25 3K30, IC D1 and D2 are of type HCT Check power boards 70551 and
Relay K4 of the power 4-15, instead of HC (see Fig. 8.10) 70568. If the IC of type HCT is
board switches with 4K15, existent, exchange for type HC
"click"-sounds 6-15, (only possible with SMD
6K15 with soldering equipment). Otherwise,
power send to Sigma for exchange.
board From January 1999 this
Part No. modification is already carried
70551 out.
and
70568
Does not stop after the 1-6, Malfunction during preselection Solder a ceramic capacitor 0,1μF
preselected time and 1-15, of time (part no. 210012) on speed
starts again 2-5 potentiometer (see fig. 8.6). For
interference suppression of the
analog output for speed
preselection.
Speed
potentiometer
RC
combination
Fig. 8.9.2: Motor without collect chuck with visible upper bearing protection disc
Problem: Bad running when using microtiter rotor part no. 11121
9 ELECTRICAL DOCUMENTS
9.1 Electrical Documents 1-13, 1-6, 1-15, 1-15K, 1K15, Microfuge R, 2-3, 2-4, 2-5
and 2-16
9 ELECTRICAL DOCUMENTS
Circuit diagram mains power supply 1-13, Microfuge Lite, 2-3, 2-4
100V – 230 V 24755
Circuit diagram mains power supply 1-6 (from construction year 2000), 1-15 (from serial
no. 71140), 2-5, Microfuge 18 (from serial no. MFA99J01)
100V – 230V 24798
Circuit diagram mains power supply 2-16 (Zent3) 100V – 230V 24851
Circuit diagram mains power supply 2-16 (Zent4) 100V – 230V 25576
Circuit diagram mains power supply 1-15K (Zent3) 120V – 230V 24852
Circuit diagram mains power supply 1-15K (Zent4) 120V 25584 Page 1
Circuit diagram mains power supply 1-15K (Zent4) 120V 25584 Page 2
Circuit diagram mains power supply 2K15, PFC 230V 25161 Page 1
Circuit diagram mains power supply 2K15, PFC 230V 25161 Page 2
Circuit diagram mains power supply 3-15, 3K15 230V 24150 Page 1
Circuit diagram mains power supply 3-15, 3K15 230V 24150 Page 2
Circuit diagram mains power supply 3-15, 3K15 100V 24204 Page 1
Circuit diagram mains power supply 3-15, 3K15 100V 24204 Page 2
Circuit diagram mains power supply 3-16, 3-16K 230V 25577 Page 1
Circuit diagram mains power supply 3-16, 3-16K 230V 25577 Page 2
Circuit diagram mains power supply 3-16, 3-16K 120V 25578 Page 1
Circuit diagram mains power supply 3-16, 3-16K 120V 25578 Page 2
Circuit diagram mains power supply 3-16, 3-16K 100V 25579 Page 1
Circuit diagram mains power supply 3-16, 3-16K 100V 25579 Page 2
Circuit diagram mains power supply 4-15 230V – 120V 24605 Page 1
Circuit diagram mains power supply 4-15 230V – 120V 24605 Page 2
Circuit diagram mains power supply 4K15, 3K30z 230V 24604 Page 1
Circuit diagram mains power supply 4K15, 3K30z 230V 24604 Page 2
Circuit diagram mains power supply 6-15 230V – 120V 24339 Page 1
Circuit diagram mains power supply 6-15 230V – 120V 24339 Page 2
Circuit diagram mains power supply 6-15 with heating 230V 24742 Page 1
Circuit diagram mains power supply 6-15 with heating 230V 24742 Page 2