Sie sind auf Seite 1von 2

Maintenance Checklist - cobas® 311

Site: .................................................. Site Contact: ……............................................

S/N: .................................................. SW Version: ...................................................


U C R A L T Remarks Done
Cover / housing X
Reaction bath / disc
Reaction bath filling , circulation X
Reaction bath temperature X
Reaction bath X
Reaction bath filter ( metal) X
Reaction cuvette X
Ultra Sonic Mixer
Check USM X
Photometer
Lamp cooling X
Photometer lens clean X
Photocurrent monitor
Heat Cut Filter X
Photocurrent values (specially
for 340 X
nm) blank
Cell X
Sample Pipettor
Needle X
Wash bath Contamination X
Rinse water adjustment
Drying cylinder contamination X
Sample probe adjustment X
Shaft X X X
Function and movement X
Pressure sensor (Clot- X
detection)
Reagent Pipettor
Needle contamination X
Wash bath Contamination X
Rinse water adjustment X
Drying cylinder contamination X
Probe adjustment X
Shaft, linear guides X X
Function and movement X
Pressure sensor ( Init. Vol. X
check)
Piercer
Contamination and condition X
Reagent disk
Reagent disk contamination X
check disk sensors X

Version 1.0_07.12.2012
Rinse unit
Tubing, connections + photo
interrupter X
Rinse nozzle X
Nozzle tip X
Linear guides X
Rinse water X
Sampling Rotor
Contamination + photo X
interrupter
Barcode reader X
Belt tension X
Syringes: Sample, Reagent,
ISE
Syringe seals X
Gear, pinion, shaft, steel plate X
Bubbles, leakages X
Syringe movement + X
photointerrupter
ISE Unit
Sipper Nozzle X
Sipper Tube and Pinch Valve X
Tube Nozzle / contamination /
Sipper
function X
Sipper waste pipe X
Airbubbles in tubing X
ISE electrodes X
Water pumps
Gearpump pressure a(300 X
kPa) pump (80-100 kPa)
Main X
Power supplies, electronics
Fan, contamination X
Boards X
Adjustments power supply
(5V, 12V, 24V) X X
Vacuum system
Diaphragm a X
Tubings / vacuum bottles X
Tank X
Hepa Filter X
Cooling unit
Reagent cooling function X
Fan, dust filter X
Condenser X
Miscellaneous
Greasing of probe guard X
Mechanism check
(incl. Photointerrupter) X
Calibration & QC Checks X

C: Check/ Clean R: Replace A: Adjust L: Lubricate T: To do

Comments: ……………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
FSR: ................................................................ Date: ................................................................

Version 1.0_07.12.2012

Das könnte Ihnen auch gefallen