Beruflich Dokumente
Kultur Dokumente
0
User’s Manual
Contents
1 Safety ............................................................................................................................................... 6
1.1 General...................................................................................................................................... 6
1.2 Instrument symbols .................................................................................................................. 6
1.3 Warning and Caution Notices ................................................................................................... 6
1.3.1 Warning ............................................................................................................................. 6
1.3.2 Caution............................................................................................................................... 6
1.4 General safety instructions ....................................................................................................... 7
1.4.1 General safety measures for high voltage equipment ...................................................... 7
1.4.2 Training and local regulations ........................................................................................... 7
1.4.3 Grounding .......................................................................................................................... 7
1.4.4 User manual....................................................................................................................... 7
1.4.5 Pacemakers........................................................................................................................ 7
1.5 Instrument safety...................................................................................................................... 8
1.5.1 General instrument safety ................................................................................................. 8
1.5.2 Mains and grounding ......................................................................................................... 8
1.5.3 Test connections ................................................................................................................ 8
1.6 Maintenance ............................................................................................................................. 8
1.7 Environment ............................................................................................................................. 9
2 Introduction ................................................................................................................................... 10
2.1 Features and benefits ............................................................................................................. 10
2.2 Service and support ................................................................................................................ 10
2.2.1 Contact information ........................................................................................................ 10
2.2.2 Checklist before calling/emailing for support ................................................................. 11
2.2.3 Shipping ........................................................................................................................... 11
2.2.4 Data backup ..................................................................................................................... 11
2.3 Receiving Instructions ............................................................................................................. 11
2.4 Warranty ................................................................................................................................. 11
2.5 Warranty repair ...................................................................................................................... 11
3 System overview ............................................................................................................................ 12
3.1 IDAX 300.................................................................................................................................. 12
3.1.1 Front Panel ...................................................................................................................... 12
3.1.2 Back panel........................................................................................................................ 15
3.1.3 Operating position ........................................................................................................... 15
3.1.4 Standard Accessories ....................................................................................................... 15
3.2 IDAX 350.................................................................................................................................. 17
3.2.1 Front Panel ...................................................................................................................... 17
4 Installation ..................................................................................................................................... 20
4.1 Installation of IDAX 5.0 software ............................................................................................ 21
4.2 Installation of IDAX 4.1 software ............................................................................................ 21
4.3 Preparing instrument .............................................................................................................. 21
4.4 Preparing test object .............................................................................................................. 22
4.5 Configurations and connections ............................................................................................. 22
4.5.1 Measurement terminals .................................................................................................. 22
4.5.2 Test modes ...................................................................................................................... 22
4.5.3 Connection examples – Power transformers .................................................................. 22
1 Safety
1.1 General
IMPORTANT
Read and comply with the following instructions.
Always comply with local safety regulations.
WEEE, Waste Electrical and Electronic Equipment. Please utilize your local WEEE
collection facilities in the disposition of this product and otherwise observe all applicable
requirements.
CE marking indicates that the product conforms to the applicable EC directives for
products placed on the the market in the European Economic Area (EEA).
1.3.1 Warning
Warning, as used in this manual, is defined as a condition or practice which could result in personal
injury or loss of life.
1.3.2 Caution
Caution, as used in this manual, is defined as a condition or practice which could result in damage to
or destruction of the equipment or apparatus under test.
Persons actually engaged in the test must stand clear of all parts of the complete high-voltage circuit,
including all connections, unless the test set is de-energized and all parts of the test circuit are
grounded. Persons not directly involved with the work must be kept away from test activities by
suitable barriers, barricades, or warnings.
Treat all terminals of high-voltage power equipment as a potential electric shock hazard. There is
always the potential of voltages being induced at these terminals because of proximity to energized
high-voltage lines or equipment.
1.4.3 Grounding
Always use a safety ground stick to ground the high-voltage conductor. A safety ground jumper must
then be installed between all terminals of apparatus under test and ground. Always disconnect test
leads from power equipment before attempting to disconnect them at the test set. The ground
connection must be the first made and the last removed. Any interruption of the grounding
connection can create an electric shock hazard.
1.4.5 Pacemakers
High-voltage discharges and other sources of strong electric or magnetic fields may interfere with the
proper functioning of heart pacemakers. Persons with heart pacemakers should obtain expert advice
on the possible risks before operating this equipment or being close to the equipment during
operation.
This instrument operates from a single-phase power source. It has a three-wire power cord and
requires a two-pole, three-terminal, live, neutral, and ground type connector.
Before making connection to the power source, determine that the instrument rating matches the
voltage of the power source and has a suitable two-pole, three-terminal grounding connector.
The power input plug must be inserted only into a mating receptacle with a ground contact. Do not
bypass the grounding connection. Any interruption of the grounding connection can create an
electric shock hazard. Determine that the receptacle is properly wired before inserting the plug.
Before connecting this unit to the mains outlet user must verify that only a single protective ground
systems exists AND no measurable voltage potential exists between the mains outlet protective
ground and the test object protective ground. If a voltage potential is found between mains outlet
protective ground and the test object protective ground then additional measures described in local
safety standards should be applied to ensure personal safety.
The test voltage which is reported by the instrument depends on the voltage set in the software.
Default value is 140 Vrms and max value is 200V peak (equivalent to 141 V RMS). The oscilloscope
will indicate that test voltage is present. Never use the software only as the indicator. Use the
Generator On/Off switch to ensure that no voltage is present when connecting and disconnecting
test leads.
1.6 Maintenance
DISCONNECT the MAINS plug before any cleaning or maintenance. Maintenance should be
performed only by qualified persons familiar with the hazards involved with high-voltage test
equipment. Read and understand Safety in the User Manual before performing any service.
Routine maintenance is all that is required for these test sets. The cables and connector panel should
be inspected frequently to be sure all connections are tight and all ground connections intact.
The appearance of the test set can be maintained by occasional cleaning of the case, panel and cable
assemblies. The outside of the carrying case can be cleaned with detergent and water. Dry with a
clean, dry cloth. The panel can be cleaned with a cloth dampened with detergent and water. Do not
allow water to penetrate panel holes, because it may result in damage to components on the inside.
A household all-purpose spray cleaner can be used to clean the panel. Polish with a soft, dry cloth,
taking care not to scratch the display screen cover. The cables and mating panel receptacles can be
cleaned with isopropyl or denatured alcohol applied with a clean cloth.
1.7 Environment
General
IDAX is intended for both indoor and outdoor use but not in wet conditions (e.g. rain, snow or high
humidity, see Relative humidity below for more details). We also recommend avoiding dirty and
dusty environments, which eventually will affect accuracy of the unit.
Mains supply
100-240 V AC, 47-66 Hz
Mains supply voltage fluctuations up to ±10 % of the nominal voltage.
Relative humidity
Operating 0 to 90% non-condensing
Storage 0 to 95% non-condensing
Mechanical requirements
Dynamic IEC 61010-1
Static IEC 61010-1
Transport ISTA 2A
EMC EN 61 326-1:2006
2 Introduction
2.1 Features and benefits
IDAX 300 is a very compact instrument and is used together with an external PC. The IDAX 350 has a
built-in computer but can also be used with an external PC. IDAX 300/350 provides an accurate and
reliable condition assessment of insulation in transformers, bushings, generators and cables. The
IDAX system maximizes the outcome of maintenance activities allowing for load and service life
optimization.
IDAX 300/350 instruments are smaller, lighter and faster than their predecessor IDA200/IDAX 206. It
maintains better accuracy and ability to provide reliable data using true AC DFR (Dielectric Frequency
Response), also known as FDS (Frequency Domain Spectroscopy), for reliable test results in high
interference environments. The software makes testing both easier and faster, allowing transformer
moisture and oil assessment in about 22 minutes (20°C).
IDAX measures the capacitance and tan delta/power factor of the insulation between power
transformer windings at multiple frequencies. Analyzing the results using modelling technique makes
it possible to assess the moisture level in the solid insulation, oil conductivity/tan delta and power
frequency tan delta at reference temperature (20°C). The test can be performed at any temperature
as the temperature dependence of the dissipation factor can be estimated.
• Automated measurement and analysis of moisture content, tan delta/power factor and oil
conductivity
• Individual temperature correction (ITC) of tan delta/power factor and oil conductivity
• 40% faster measurements with new IDAX 5.0 SW
• Reliable measurements in high-interference environments
• Multi-function test set for transformer measurements
2.2.3 Shipping
If you need to ship the unit back to your local representative or to Megger Sweden, make sure that
the unit is well protected. Megger cannot take responsibility for any transport damages.
2.4 Warranty
• Products supplied by Megger are warranted against defects in material and workmanship for
a period of one year following shipment.
• Our liability is specifically limited to replacing or repairing, at our option, defective
equipment.
• This warranty does not include batteries, lamps or other expendable items, where the
original manufacturer’s warranty shall apply.
• We make no other warranty. The warranty is void in the event of abuse (failure to follow
recommended operating procedures) or failure by the customer to perform specific
maintenance as indicated in this manual.
3 System overview
IDAX is an insulation diagnostic system for investigations/analysis of dielectric materials, normally
insulators. The measurement method used is dielectric frequency response or dielectric
spectroscopy, i.e. measurement of the dielectric properties of the material as a function of the
frequency, and in some cases also as a function of the voltage. By studying the dielectric material
properties as functions of frequency it is possible to make a distinction between different types of
phenomena. For example, it is possible to separate polarization loss from leakage currents.
The system applies a sinusoidal voltage with desired frequency over the sample. This voltage will
generate a current in the sample. By accurately measuring the voltage and the current, the sample
impedance can be calculated. Depending upon the actual sample model, various parameters can be
calculated from the impedance, such as such as tan delta/power factor, capacitance and resistance
etc. This procedure can then be repeated at the specified frequencies and voltage levels and even
more information on the sample can be revealed.
The system is fully computerized and most of the measurements can be performed automatically.
Interface
d) USB Port
for connecting computer
e) ETHERNET
RJ45 connector for Ethernet connection via twisted
pair cable.
Output
f) Enable switch
1 for activating the generator output.
0 for deactivating the output of voltage.
g) Generator
Generator output triaxial connector
h) Active
Red high intensity LED. Is lit when output is
enabled, voltage may be generated.
External
Connector for external amplifier, e.g. VAX020
and VAX230.
Input
Input 1 (red) and Input 2 (blue)
First and second input channel.
Do not connect to mains or energized test
objects.
Do not connect directly to Generator output.
Protective ground
Connector for connection to test object ground.
This is also used as measurement input.
Test
Test inputs.
Can be used for controlling basic functionality
and accuracy.
Connect only to Generator, and Input 1 or 2.
Mains
Mains connector
For connecting to mains outlet.
When connecting the mains cord, the
internal computer will start.
Power ON/OFF switch
Turn the measurement and generator
module on or off.
Fuse
Use small screwdriver to gently pry out to
change the fuse, F1, 2AT slow blow.
Interface
USB Port
for connecting to an external computer. Put
the switch in EXT position. To use the
internal PC, put the switch in INT position.
USB Ports
for connecting an external mouse, keyboard
or USB memory to the internal PC
Ethernet
RJ45 connector for Ethernet connection via
twisted pair cable to internal computer.
Output
Generator
Generator output triaxial connector
Active
Red high intensity LED. Is lit when output is
enabled, voltage may be generated.
Enable switch
1 for activating the generator output.
0 for deactivating the output of voltage.
External
Amplifier
Connector for external amplifier, e.g.
VAX020 and VAX230.
Inputs
Input 1 (red) and Input 2 (blue)
First and second input channel.
Do not connect too mains or energized test
objects.
Do not connect directly to Generator output.
Test
Test inputs.
Can be used for controlling basic functionality
and accuracy.
Connect only to Generator, and Input 1 or 2.
Before delivery, the serial number label will be attached to the back panel with a unique serial
number (YYXxxxx) and Megger’s article number (AG-19172).
4 Installation
The IDAX 350 comes with the IDAX software already installed on the internal computer. On the IDAX
300 and if you want to use the IDAX 350 with an external PC you have to install the software.
IMPORTANT: On the IDAX 350 make sure to close the IDAX software and the operating system
before turning off the unit using the on/off mains switch and disconnecting the mains cord.
IDAX350 can be used as a central computer controlling other test instruments e.g Megger FRAX and
TTR. Megger SW is checked for co-existence and can be used together with IDAX 4 and 5. Be careful if
installing other manufacturer’s software on the IDAX unit. Doing so may prevent the IDAX software
from working as intended.
When first starting the IDAX software after installing it make sure the IDAX unit is connected and
turned on. For the IDAX 350 check that the PC interface switch is on the EXT position. When the
software starts it will look for the IDAX hardware and connect to it automatically.
When first starting the MODS software you will be asked for user name and key. Be sure to have
them handy.
In case of a two winding transformer the set-up the leads need to be moved between test 2 and test
3. In case of a three winding transformer the IDAX 300 second input channel is utilized. The test leads
need to be moved between test 2 and test 3 and between test 4 and test 5.
If using IDAX 300S or IDAX350 with two separate current measurement channels, the two set-ups
using same cable set-up can be measured simultaneously, e.g. test 1 and test 2 (test 3 and test4; test
5 and test 6). Note that Test 6, CTH (Energize T and measure H), in many cases are replaced by CHT
(Energize H and Measure T)
2. Select “Single Test” tab, “Moisture” button (see left-hand side arrows)
3. Then choose suitable test object template, e.g. 2W-transformer (two-winding transformer)
4. Enter mandatory temperature data for ambient and object temperature (highlighted). Object
temp. (°C) refers to insulation temperature. Top oil or winding temperature is normally used
as insulation temperature for the transformer. Designation is the tag or identification code of
the UUT (unit under test) e.g. T2 (transformer No. 2). Suggested file name will be
“Location”+”Designation”+”Date”+Time”.
5. Press the green “Create”-button at the low right-hand side of the window.
6. Select the folder where the measurement will be saved and name the file. The default name
is composed of: “Location”+”Designation”+”Date”+Time”.
Note: At any later time you may access Nameplate and change/add information by pressing “Edit
Nameplate” button in the program main window.
Now you have a “File” with one “Test”, the CHL configuration, visible on the legend in the left-hand
side of the main window (see picture below). On the top you have several “Tabs” showing
Nameplate, The Test (Named Moisture), Hook-up, Graph (and table), Settings and Help.
1. Press the Hook-up tab to show the cable connections from the instrument to the object.
There are two pictures showing the two versions of IDAX300, IDAX 300 delivered 2011 or
earlier (to the left) and IDAX 300 delivered 2012 or later. Note that in case you have an older
IDAX300 with two input connector for each channel, the SENSE input must NOT be
connected
2. Ensure that the safety ground cable is connected to the IDAX 300 instrument and to the
station ground near the test object/transformer. This cable acts as safety ground AND test
input lead.
3. Make sure that all HV phase terminals and any exposed neutral terminal are connected
together. While connecting together the terminals with bare copper wire, the
phases/windings should be connected to ground.
4. Make sure that all LV phase terminals and any exposed neutral terminal are connected
together. While connecting together the terminals with bare copper wire, the
phases/windings should be connected to ground
5. Connect the Generator cable to the high voltage winding – according to the picture in Hook-
up tab.
6. Connect the Red input lead to the low voltage winding – according to according to the
picture in Hook-up tab.
7. Remove any grounding connection from the windings before you start the measurement.
8. Connect to the IDAX instrument using the “Connect” button located on the right-hand side of
the main window or simply press the “Start” button (that will connect to IDAX 300 if not
connected yet)
9. Press Start to initiate the measurement/test
5.1.1 Measurement
The Moisture Test will start a routine initiation procedure where the system measures all1
capacitances, including stray capacitances in the test object as well as the interference level. This
routine procedure sets up the instrument and the model ensures the highest possible accuracy over
the selected frequency range.
Note; A capacitance check is also performed and in case the measured capacitance to ground does
not match with the sum of individually measured capacitances, the measurement will be stopped. If
so, re-check that all connections are correct and start the measurement again.
Next, the Dielectric Frequency Response (DFR) measurement starts from the highest selected
frequency (usually 1000 Hz) and then steps down to the set “Stop frequency” of e.g. 1 mHz. After
measuring 40 Hz, power frequency capacitance and dissipation factor are calculated and presented.
5.1.2 Analysis
In a Moisture Test; when DFR sweep passes 1 Hz, the first automatic analysis is performed and
preliminary results are presented in the “Analysis results” section. After the test is completed (last
frequency point measured) the final analysis results are displayed:
• Measured capacitance and %DF
1
From 2012, IDAX has single input terminals and utilizing a new methodology where stray capacitance are
estimated and compensated for based on pre-measurements on all terminals, red, blue and ground. In
previous IDAX models with dual input terminals, “input” and “sense”, some stray capacitances were not
compensated for. Note: The compensation is only important for higher frequencies well above 50/60 Hz.
5.1.3 Disconnect/re-connect
It is possible to disconnect the computer and let the IDAX unit run standalone by pressing the “Run
Standalone” button and then disconnect your computer. The data will automatically be stored in the
IDAX instrument’s internal memory and uploaded to the IDAX 5.0 when it is re-connected (you may
even close IDAX 5.0 SW and turn off your computer but if you re-connect using a different computer,
you will only get test data which does not include general e.g. Nameplate data)
When no file is loaded the left section shows the latest used files for quick access and a button for
opening other saved files. If you open an old IDAX file in either IDF or IZF format you will be given the
option of either just open the file or convert it to the new IDAX format.
5.2.1.1 File conversion wizard
A file conversion wizard will be started if the choice is to convert the file. Here you will be required to
add necessary information that is missing and you can also correct faulty information.
When a file is loaded the wizard displays a list of the sweeps in the file. You can select what sweeps
to show clicking the X in the colored square to the left of the sweep name. If you right-click the
colored square you can change the sweep color. If you double-click the sweep name you can add a
short description of it.
5.2.1.2 Sweep (test) actions
A sweep with no tick to the right of the sweep name is not measured and you can start the
measurement with F9 or “start”
A ticked sweep is measured can be re-measured with “Remake”. Please note that the new
measurement will be performed as by the actual measurement settings and not necessarily exactly
as the old measurement.
By clicking the name at the top you open up a menu for handling the file where you can open the
folder it is save in or send the file via e-mail.
At the bottom there is a button for deleting the selected sweep as well as the button for opening a
new file.
Notes is intended for noting specific details for the particular measurements
Attachements may be photos or other files describing the test object/test conditions etc
5.2.2.2 New test
New test creates a new measurement/sweep where the type of measurement can be:
Moisture
DFR
%PF/%DF
DC
For every measurement the object and test mode can be defined as well as other data necessary for
the test.
5.2.2.2.1 Moisture
See section 5.1.2 for information about Moisture tab.
5.2.2.2.2 DFR
In the DFR Tab the measurement is performed at multiple frequencies in the same way as in the
Moisture Tab but no automatic analysis is performed. Analysis is started manually from the Analysis
button, see section 5.2.3.5. The results of the analysis will only be displayed after an analysis has
been performed (as “DFR” or “Moisture”)
5.2.2.2.3 %PF/%DF
In the %PF / %DF tab, 50 or 60 Hz values for capacitance, dissipation factor/tan delta and excitation
current can be measured. Multiple recordings can be made by repeating “start”. Results are
presented in table format.
5.2.2.2.4 DC
In the DC tab you can perform DC insulation measurements. Results are presented as polarization
current versus time as well as Insulation Resistance (IR) at 60 seconds, Dielectric Absorption Ratio
(DAR) and Polarization Index (PI). Please note that the measurement time needs to be sufficient (10
minutes) to display all values.
5.2.2.5.1 General:
• Language: Different languages can be selected form this menu. After selecting a different
language, the SW needs to be re-started in order of this change to apply.
• Line Frequency: Frequency selection for the power network where the actual component is
measured (not necessarily the AC input to IDAX). Will set frequency for presentation in Test
tabs, for Line Frequency %DF/%PF measurements and power frequency interference
measurements (in the initialization procedure).
• %DF/PF/TD: Selection of %DF (percentage Dissipation factor), %PF (percentage Power Factor)
or %TD (percentage Tan-delta = %DF).
• Bushings: Bushing terminals in Multiple Sweeps are labeled according to ANSI (H1, H2…), IEC
(1U, 1V), ABC (A, B …) or Customized labeling that is possible to edit.
• Temperature: Selection of Fahrenheit or Celsius for ambient temperature entry.
Note: Object/insulation temperature is always entered as degree Celsius.
• Enable Keypad: Enables or disables the pop-up keypad suitable for touch screen operation.
• Enable oscilloscope: Enables the Oscilloscope (see upper right corner in the main window)
5.2.2.5.4 Models:
Selection of what parameters to be available in graphs and tables. The options are as follows:
C: Capacitance
%DF: % Dissipation Factor (it will change to
%TD if %TD is selected in General)
%PF: % Power Factor
C’’: The imaginary part of the complex
capacitance (C is capacitance AND the real
part of the complex capacitance. 1/Z = Y =
j*w*(C-jC’’)).
conductivity (or %DF /%PF) @ 25°C is set. By pressing the Edit Ranges button, the Edit
Analysis window will pop up and modifications to default setting values are possible.
o Note that each type of liquid has its own limits.
o Note that in the Edit Analysis window, it is possible to enter only values for
conductivity; however, these values are automatically presented as e.g. %DF based
on settings in Insulation Liquid Loss
The suggested stop frequency value for Moisture Test is based on the object/insulation
temperature vs. stop frequency table that can be edited under the “Edit Stop Frequency”
button.
Edit materials
Edit materials will let you choose which materials to use in the analysis window for specific
objects. A transformer usually has cellulose while an instrument transformer typically has
Kraft paper as insulation material. Default materials are Cellulose and Kraft paper, more
materials can be added.
Insulation Liquid loss
This item represents the losses for material 2, normally mineral oil, in the analysis window,
the different options are:
Conductivity (S/m)
%TD, Tan delta value in percent
%DF, Dissipation Factor value in percent
%PF, Power Factor value in percent
When the IDAX system is connected, the button will change label to a Disconnect button and you
may use this button to disconnect the IDAX.
When a long and time consuming test is initiated, it is possible to disconnect the PC from the IDAX
and let the unit store the measurement data until it is connected again. In cases where this is
possible, the button is labelled Run Standalone.
5.2.3.9 Settings section
In this section you may change the settings for the specific test:
Change colour of e.g. graphs by pressing the color button.
Change Test mode by clicking the test mode button (currently set to measure CHL for this
test
Change test voltage level by changing test voltage.
Change temperature of object/insulation tested.
5.2.3.10 Start/Stop/Remake button
Using the button labelled Start, you may start a test, and by pressing the button labelled Stop you
will stop a test. If you have marked an executed test in the legend, you are not able to override this
test and therefore the button change function gets labelled Remake. The Remake guides you to
make a new test as a copy of the executed test.
Note: Using the Remake-button will make the settings used in the previous measurement but not
copy the settings from the 5.2.2.5 section “Settings tab” such as frequency, generator and cable
setting.
Limits for X/Y depend on the test object and default settings as in table 1.
5.2.6 Moisture
This is the Moisture tab described in section 3
5.2.7 DFR
The Dielectric Frequency Response (DFR) tab opens for a general DFR measurement of capacitance
and dissipation factor over a frequency range. In this tab, the user defines test object and setup,
frequency range, if and how the results should be analyzed etc.
5.2.8 %DF/%PF
In this tab IDAX performs standard capacitance and dissipation factor measurement as well as
excitation current measurements at a single frequency e.g. at power frequency for the selected test
object/test mode.
5.2.9 DC
In the DC tab you can perform DC insulation measurements. Results are presented as polarization
current versus time as well as Insulation Resistance (IR) at 60 seconds, Dielectric Absorption Ratio
(DAR) and Polarization Index (PI). Please note that the measurement time needs to be sufficient (10
minutes) to display all values.
1. Follow the general preparation procedures as described in Getting Started with IDAX.
2. Connect IDAX to ground/transformer tank using the external ground cable. Make sure
the connection point is free from paint/dirt/rust.
3. Turn on IDAX and external PC
4. Connect the USB cable for PC connection
5. Start IDAX SW
6. Open Results window by File-Windows-Results
7. Create your test object by selecting File-New-Object in the Results window
8. Browse/select type of object/object template (2-winding transformer) and enter test
object type and name. Save (Note: If no objects are found, Win 7 has automatically
changed the default directory from the original setting. Object template folder is My
Documents\IDAX\ObjectTemplates. Browse to the correct folder and select)
10. Create your test sequence by selecting measurement object and File-New-
Measurement sequence (or right-click on File/Object name). Enter test name (not
mandatory) – OK
11. Select test and enter test information. Most important is to enter insulation
temperature (Apparatus temperature). For a transformer insulation temperature is
assumed to be the same as winding/oil temperature. Save.
1. Prepare/confirm the transformer and IDAX for the test as described in the
"Measurement Plan" tab text and picture.
9. When MODS opens, confirm the insulation temperature for the measurement
(temperature data should normally have been entered before the measurement and
the value is automatically transferred to MODS)
10. Check that “Conductivity”, “Moisture” and Geometries are selected for automatic
analysis
11. Click “Auto match”.
12. MODS will start the iteration procedure and when finalized after about 30 seconds,
display moisture content, oil conductivity, tan delta at power frequency and
approximate insulation geometry
If you discover a measurement problem during the test, the measurement can be
stopped/aborted immediately. To measure the same setup again, right-click on the
measurement file and select “Remake measurement”. IDAX will automatically create a
new measurement file with a different numbering. (An IDAX measurement can be
aborted at any time. All data collected so far will be stored and can be used for
analysis).
An IDAX (true AC DFR) measurement do not require discharging to measure accurate
results, you can simply re-measure the same configuration several times. (This is NOT
true for DC-methods!)
It is recommended to start using IDAX by going through the procedure in this section in order to
become familiar with the instrument and the software. This procedure is also a good test to perform
whenever a confirmation of the proper functioning of the instrument is required. This test is
executed using the standard equipment delivered with IDAX.
6.2.1 Measurement
1. Make sure that IDAX is properly connected to mains outlet.
2. Turn on IDAX using the mains switch POWER on the front panel. The IDAX performs a
self-test and start-up procedure.
4. In Results create new object selecting Object from menu File / New. Click "Browse"
and select !!Test Capacitor.obj template and enter arbitrary name as Object Name, for
example "Test" and click OK.
7. Create new measurement sequence selecting Measurement Sequence from menu File
/ New. Arbitrary name can be entered as Measurement Sequence, for example
"Getting Started" and press OK. New measurement sequence with corresponding
Measurement Plan files will appear in Test Browser.
The test object to be measured using the built-in capacitance box built into the system that
terminates at the front panel. The respective capacitances are:
C10 = ~2.5 nF
C12 = ~4.7 nF
Warning!
The measurement system can generate hazardous and even lethal voltages.
Carefully read "Safety precautions".
In Test Browser select Measurement Plan file !!C12 (UST), carefully read and follow the instructions.
1. Make sure that preparation procedures (both IDAX and Test Object) have been followed.
4. Turn on the output switch activate the output and then start the measurement. The IDAX starts by
running a pre-test sequence to verify test conditions and the actual test is not started if certain
conditions are not fulfilled.
1. Make sure that preparation procedures have been followed (see also "Ungrounded Specimen Test
(UST) Measurement").
4. Turn on the output switch activate the output and then start the measurement.
5. Wait until measurement is finished (or abort earlier by stopping the measurement).
1. By pressing <F4> on the keyboard, Results window can be activated if not already active and the
measurement progress can be viewed when selecting Graph tab. The graph shows the dissipation
factor for a typical transformer in the frequency range 0.002 - 1000 Hz .
2. By pressing <F5> on the keyboard, an oscilloscope will be displayed on the screen. The Ch 0 (red)
signal in the oscilloscope displays the applied voltage over the sample, Ch 1 (white) is the output
voltage from the first electrometer and Ch 2 (turquoise) is the output voltage from the second
electrometer. The output voltage from the electrometer is a function of the current through the
sample.
3. The instrument stops automatically after the last measurement point is completed, however, the
measurement can be interrupted earlier by a stop command.
Depending on which parameters are of interest to the user, the measurement data can be presented
using different models. E.g. in the tanδ model the dissipation and/or power factor as well as
capacitance can be displayed in the graph as a function of frequency. See also how to select models
in sections "Configuration / Models" and "Sample Modelling".
The results can be viewed using three additional tabs: Graph, Table and 50/60 Hz data as shown in
pictures below.
The selection of 50 or 60 Hz is selected in the Configuration menu under Power Frequency Data.
2. In the Graph tab you can select the desired parameters. For example, the dielectric
parameter ε'.
3. In Results window menu the choices File / Export, File / Print and File / Report allow
for exporting the measurement data and creation of reports. See section File / Report.
All measurements are stored on the IDAX 350 system hard disk or the measurement computer's hard
disk on an IDAX 300 system and can be opened and analysed at any time. For comparison purposes,
many measurements can be loaded into the same Results window.
In addition to the IDAX System Control program window, a few more windows are accessible from
here.
Window Functions
Results Editing of measurement plans and displaying the results. Window is activated
after up-start by default
Comments Adding comments to the actual measurement.
Messages Displaying messages from the software during a measurement.
Oscilloscope Viewing the actual voltage and current curve forms.
6.2.5.3 Oscilloscope
By pressing <F5> on the keyboard, an oscilloscope will be displayed on the screen. The Ch 0 (red)
signal in the oscilloscope displays the applied voltage over the sample, Ch 1 (white) is the output
voltage from the first electrometer and Ch 2 (turquoise) is the output voltage from the second
electrometer. The output voltage from the electrometer is a function of the current through the
sample.
Horizontal control
The oscilloscope time base can be set either on auto scale or manual. On auto scale, the time base is
set from the generator signal from the DSP. In the manual mode, the time base is set by the Time/div
slide bar. The time above the slide bar refers to the time per division, the full oscilloscope window
width consists of ten divisions.
Roll
The Roll check box can be used for slow signals that the oscilloscope may otherwise have difficulties
to accurately lock.
Vertical control
The gain of the input channels can be adjusted with the vertical slide bars. The voltage range is
displayed beside the slider. The offset of the displayed signals can be adjusted by placing the mouse
pointer inside the oscilloscope window and dragging it up and down while holding the mouse button.
German
French
Spanish
More languages can be added on request or you could translate it yourself.
Language is changed selecting menu Configuration / Change Language in IDAX System Control,
which opens a file opening dialog. The selected language file, *.lng, is loaded and it will be used after
restarting the IDAX program. Each language has its own language file, *.lng, and all language files are
located in the directory C:\Program Files\Megger\IDAX\Language files\.
The Test Leads dialog is used to select and, possible edit the settings for different cable sets.
The presets drop-down is used to select what cable set you are to use for a measurement. If the
cable set you are planning to use is not included you can use the “Add” button to add a new preset.
There is a check-box you to leave blank for when not using the sense lead. It only applies if Sense is
connected. Sense was available on earlier version of IDAX300, but has been discontinued.
The settings for a cable set can be edited. What the different settings means is explained in this
table.
6.2.5.6 About
In "About" window the software versions of installed IDAX system, its program components and
information about the operating system are displayed.
Interaction with the program is done via windows and toolbar menu, the functions are described in
"Results Menus and Commands"
Menu Description
File Menu Operations with measurement objects, measurement files, etc.
Edit Menu Editing and changing properties of opened measurements.
Diagram Menu Zooming features
View Menu Selection of presentation form
Configuration Menu Configuration of Results window settings
Help Menu On-line Help
Icon menu
Icon menu is designed for easier navigation and operations with Test Browser.
Open measurement
Close measurement
Save
Go one Measurement Plan or data file Down in Measurement Sequence if there are more
than one
Go one Measurement Plan or data file UP in Measurement Sequence if there are more
than one
New Object Create new Object is created if the object to be measured is not among
those in Test Browser
New Test Sequence Create new Measurement Sequence for a selected Object
1. Select New / Object from File menu which opens New Object dialog
2. Select appropriate object template by using Browse button and press Open
3. Enter name for the new object, for example, "Test capacitor" and press OK.
2. From File menu select File / New / Measurement Template, enter a name, e.g. template.icf, for
new template and press Save. New empty template will appear in Text View mode for editing. See
also section "IDAX Commands and Variables".
2. From File menu select File / New / Object Template, enter a name for new template and press
Save. New empty template will appear in Text View mode for editing. See also section "IDAX
Commands and Variables".
Note!
Before creating new Measurement Sequence, at least one measurement template must be
associated with the test object.
2. New Measurement Sequence will be automatically assigned a name in form of date as shown
below (date has the format according to International Standard ISO 8601). The name for
Measurement Sequence (i.e. "First Test" as in picture below) is optional.
This dialog window can be also invoked by right mouse click on selected object and choosing
Add/Remove Templates from the menu which appears as shown below.
3. Right click and choose Remake Measurement. It will create a Measurement Plan, (C-file) from the
selected data file, i.e. all measurement settings from this data file are copied into the new
Measurement Plan, (C-file).
Note: The Object Templates are saved in Object Templates directory and
Measurement Templates under Measurement Templates directory
This command allows for temporary loading of single measurement files, D-files. During loading a
Temporary Object and Temporary Sequence are created as shown in picture below. All subsequent
data files are opened under the same Temporary Sequence.
The opened files can be then imported under an existing object by using File / Import / File To
Structure.
If Results program is opened as an independent program and the data files are temporary opened,
then closing Results program, user is asked if the opened files should be imported into the existing
data structure.
imported in a fashion which corresponds that they can be appropriately imported back into the same
or other data structure.
Depending on data to be imported type and structure the following options are available:
Import the data in from archived (*.izf) format. Data are inserted among existing
measurements preserving the structure of the archived file.
1. Choose File / Import / IDA Zip File to invoke regular Open dialog.
5. Select measurement(s) to be imported. By default all measurements located in given directory are
imported. Deselect those which not to import by unchecking appropriate check boxes.
6. Press OK button and the measurement will appear in the Test Browser.
1. Choose directory where template(s) to be imported are located by typing in the path manually or
using Browse button.
2. Select template(s) to be imported from available templates in the left hand side.
3. Press Add button or double click by left mouse button on chosen template.
4. Press OK button.
1. Choose directory where the template(s) to be imported are located by typing in the path manually
or using Browse button.
2. Select the template(s) to be imported from available templates in the left hand side.
3. Press Add button or double click by left mouse button on chosen template.
4. Press OK button.
It is also possible to copy and paste directly from the Table view.
Export of the data in an archived (*.izf) format. By exporting data into an IDA Zip File, the data
structure is preserved. It is possible to export:
If s measurement sequence is selected the selected measurement sequence and the object will be
exported. If an object is selected that object and all measurement sequences will be exported.
The "Export" command invokes an export-dialog with two different file type options, export to a file
as a graph or the measurement points:
If the measurement data is exported as a TXT-file one can also select appropriate Format and
Decimal separator. The Format option allows user to remove the header of the file, which makes it
easier to import the data into another program. Decimal numbers are written in the file using chosen
Decimal separator option - point or comma.
Export will include all selected parameters. For selecting parameters see Selecting Models.
Moreover, it is possible to export also the measurement file itself by checking checkbox "Include D-
File", which may be favourable when exporting to a removable media.
6.2.6.1.1.10.3 Header
Export with header
6.2.6.1.1.10.4 Point/Comma
Point as a decimal separator
1. Select templates to be exported side from available templates in the left hand.
4. Press OK button.
To remove a template:
1. Select templates to be exported side from available templates in the left hand.
4. Press OK button.
To remove a template:
The Print command allows user to print out a simple report of the measurement results in a text
format.
This command invokes a dialog window where two report generators are available. It requires that
you have selected either a measurement or a measurement sequence and at least one measurement
is loaded.
When invoking MS Word report generator, a dialog window allows for choosing between various
options as shown in the picture below. If there are already prepared report file(s) they will be listed
in the left side of the dialog window and can be selected to work with.
In addition to different data export options, the measurement data file, (D-file) can be sent to other
programs directly for further analysis. For this purpose Results are configured by providing search
path to the external program. The configuration is made from menu Configuration / Program...
where the appropriate executable file is selected in dialog box.
Now the data file loaded into Results can be sent to the chosen external program from File / Send
to... as shown below. If the configuration step has been omitted, the dialog shown above appears
and user is prompted to enter the search path at this stage.
When a executable application is selected, the name of the executable is added to the Send to...
menu as shown below, where MODS has been chosen as application where data are to be sent.
The data can also be sent to selected application by right mouse click and selecting Send to .. MODS
as shown in picture below.
2. Choose Delete File/Folder from menu or right-click on the file or folder. A pop-up window will
appear:
3. To confirm, select Yes and press Enter or left-click, otherwise select No.
1. Closes Results program, without quitting IDAX System Control program. Results can be restored by
pressing F4 or selecting from menu Windows / Results in IDAX System Control window.
2. If Results were launched as stand-alone application - File / Exit quits Results program without
saving changes.
Reverse the effect of the last action performed, such as typing, deleting or formatting text, etc.
Place a duplicate of the selected text or item(s) onto the Clipboard where it can be pasted elsewhere.
Graph
Copy Results window to Clipboard. i.e. make screen dump as using Alt+Print Screen.
Table
Copy content of the table to Clipboard
50/60 Hz Values
Copy measured values at 50/60 Hz to Clipboard.
Test Plan
Highlight all text in the current text frame (Text View) and place a duplicate of it onto the Clipboard
where it can be pasted elsewhere.
Delete the selected text from Measurement Plan or Template and place it on the Clipboard where it
can be pasted elsewhere.
Place the contents of the clipboard (the last text cut or copied) back into Test Plan.
Highlight all text in the current text frame (Text View) for subsequent copying.
By selecting Edit / Read Only this feature is removed and user can change the content of all fields.
1. The Zoom In command allows user to select a rectangular area of the by dragging a rectangle with
the mouse.
2. The marked area will then be magnified to cover the entire graph area as shown below.
Zoom out command fully displays all active curves similarly as turning on Autoscale in axis settings.
The axis settings in the axis settings (see section Configuration / Graph / Axis), however, remain
unchanged.
editing
Show Measurement Templates Show/Hide Measurement Templates
Displayed content of Information tab depends on selected place in Test Browser. There are several
possibilities:
Object selected
Measurement Sequence selected
Measurement Template selected
Measurement Plan selected
Measurement Data selected
Information - Object
Displays information about object corresponding to the chosen Object Template. Object information
can be edited using also Text view mode. The changes made here will, however, be unique for the
given object. Permanent changes must be made in Object Template(s).
Caution: The data files are Read-only and even the editing is possible after disabling Read-only
attribute, it is strongly discouraged. Any data loss or damage caused by editing of data files is user's
responsibility.
The IDAX System Control reads the Measurement Plan and each line is interpreted in subsequent
order. Normally, every variable or command is contained in a separate row. However, multiple
variables and commands can be entered in the same row if they are separated with semicolons.
When the measurement is started, current Measurement Plan is copied to a file LastC.icf in D:\IDA
catalogue, which is then read and executed by the IDAX System Control.
Displays graphical output of an ongoing measurement and/or previously made and opened
measurements. It is possible to display graphical representation of the measurements using different
parameters, i.e. loss tangent, power factor (PF) capacitance (C) and others (see also Configuration /
Models).
The labels LIN and LOG can be clicked to switch between lin and log display of the axis. Note that
negative values can not be displayed in log.
By clicking right mouse button on a measurement point the properties of this point are shown in
Properties box in left lower corner of Results window.
Ongoing measurement and/or previously made and opened measurements are displayed in a text
mode. In addition to measured values loss tangent, power factor, PF, capacitance, C and others, in
Table mode also hum current, DC offset and object's capacitance are shown for specified
configurations.
10 kV equivalents
The 10 kV equivalents are values that are calculated from the measured values. Using low voltage,
e.g. 200 Vpeak, will result in a certain current through the sample. A linear insulating material, such
as oil and/or paper, will have the same dissipation factor and capacitance also at a 10 kV test voltage.
Therefore, it is possible to calculate the current and power at 10 kV from the measured values.
2. Advanced help enables automatic appearance of alternatives for variables. When typing
a variable name followed by "=" sign, the alternatives appear as a drop-down menu (see
below) and user can select the appropriate one.
3. Highlighting settings allows user to change the color scheme for appearance of text editor (see
example below).
4. Font... allows user to select the fonts to be used in Text view of Measurement or Object Templates
and Measurement Plan.
Title
X-axis
Y-axis.
6.2.6.1.5.2.2.1 Autoscale
Autoscale automatically sets the scale so that all active curves are displayed in the graph.
Note!
Negative values are not possible to display in logarithmic scale. You can switch between lin and log
scales by clicking the text in the graph.
6.2.6.1.5.2.2.2 Grid
For grid settings, three options are available:
6.2.6.1.5.2.2.3 Label
Allows user to enter a label text for both X- and Y- axis.
6.2.6.1.5.2.2.4 Log/Lin
Makes it possible to change between logarithmic and linear scale for the X and/or Y axis.
Note!
Negative values are not possible to display in logarithmic scale.
6.2.6.1.5.2.2.5 Precision
Precision sets how many digits of the point is presented in the Graph/Table.
6.2.6.1.5.2.2.6 Prefix/Scientific
Changes between Prefix or Scientific notation in the Graph/Table.
6.2.6.1.5.2.2.7 Title
Here the title for the graph in Results can be written. If Auto is checked, the title of the measurement
will be displayed for the measurement. If more than one measurements are loaded in the results
window, with different titles, no title will be displayed.
6.2.6.1.5.2.2.8 Unit
Units for both X- and Y- axis.
Using this command the background of the graph can be changed. The selected color is used on
display and in files exported in *.wmf format, however, when printing the background color is set to
white.
3) showing interpolated 10 kV equivalents for current, dissipated power, Power Factor and
capacitance
Command Configuration / Models... invokes a window where available models for presentation of
measurement results are shown. Different models can be selected and the corresponding results in
graphical or text format will be simultaneously displayed.
Selected models appear at the bottom of Results window, where one can choose which parameters
are to be shown. Below the Dielectric and C, PF, Tan-Delta models are chosen to be used in Results
and ε′ as a parameter is displayed in the graph.
Topic Description
IDAX Overview of IDAX operating principle and functions
Results User Interface Preparation of measurements and presentation of results
Command and Variable Reference Reference of programming directives
IDAX Error Messages Error messages generated by IDAX and possible
countermeasures
6.2.6.2 Test Browser
Test Browser is located at the left hand side of the Results window and it works similarly to Microsoft
Windows Explorer. The main purpose of the Test Browser is to facilitate:
Objects
Each object is represented by an object template specified in a file Object.obj. This file contains the
information about the object which is constant such as owner, manufacturer, serial number, etc.
There are several pre-defined Object Templates for power transformers, bushings and cables.
Depending on user's needs these files can also be modified as well as new templates can be created.
When creating new objects it is recommended to select Measurement Templates which can be
associated to those objects.
Measurement Sequence
Measurement Sequence is created in order to make a measurement and therefore at least one Measurement
Template must be associated with the Object.
Measurement structure
In order to perform a measurement a Measurement Sequence must be available or a new one must
be created. New sequence is created by selecting desired object and press File / New /
Measurement Sequence. This will result in:
A dialog will appear where a test date and an optional name of Measurement
Sequence can be defined
All Measurement Templates associated with the object will be copied to the folder of
new Measurement Sequence (inclusive the object information) as command or C-files (*.icf).
In order to run a measurement, a command file from those under Measurement
Sequence is chosen and start the measurement.
When measurement is started, the command file is replaced by a data file containing
both information from the command file and measurement data. Finished measurement can be
repeated by selecting data file and choosing File / Remake Measurement. This will create a new
command file copy from data file, which is possible to start again. (The name of command file will
remain the same as original data file with additional number at the end, for example xxxxx-2 or yyyy-
3, where number indicates times the measurement has been repeated.
When measurement is finished the next command file in given Measurement
Sequence can be selected and run until the entire Measurement Sequence is done. Unwanted
command files can be deleted or simply left as they are.
In all Measurement Templates it is provided that user can choose to perform either:
1) frequency sweep down to a specified stop frequency, or
2) measurement only at power frequency (50 or 60 Hz). This last option is not available for XLPE
cable measurements and in limited extent in general measurement plans.
Bushings
Available Measurement Templates are (for each phase):
C1, H1 - measurement of bushing H1* main insulation, C1
C1, H2 - measurement of bushing H2 main insulation, C1
C1, H3 - measurement of bushing H3 main insulation, C1
C2, H1 - measurement of bushing H1 insulation between test tap and ground sleeve,
C2
C2, H2 - measurement of bushing H2 insulation between test tap and ground sleeve,
C2
C2, H3 - measurement of bushing H3 insulation between test tap and ground sleeve,
C2
PILC Cables
PILC Cable, A - measurement of insulation of phase A*
PILC Cable, B - measurement of insulation of phase B
PILC Cable, C - measurement of insulation of phase C
XLPE Cables
XLPE Cable, A - measurement of insulation of phase A*
XLPE Cable, B - measurement of insulation of phase B
XLPE Cable, C - measurement of insulation of phase C
General measurements
Available Measurement Templates are:
General (PF Data) - generic measurement plan for measuring only at power frequency
(50/60 Hz)
General (Sweep) - generic measurement plan for measuring of frequency sweep
General HV - generic measurement plan for measurements using IDA High Voltage
Unit
Rotating machines
Available Measurement Templates are:
ABC-GND
A-GND
B-GND
C-GND
C-A
B-C
A-B
Reactor
Available Measurement Templates are:
CH
CT
Available Measurement Templates are:
Ph1, CH
Ph2, CH
Ph3, CH
CT with tap
Available Measurement Templates are:
Ph1, CH-Tap
Ph2, CH-Tap
Ph3, CH-Tap
Ph1, Tap-GND
Ph2, Tap-GND
Ph3, Tap-GND
!!!Quick Test
!!Test Capacitor
!General Specimen
Autotransformer with Tertiary + Bushings
Autotransformer with Tertiary
Autotransformer without Tertiary + Bushings
Autotransformer without Tertiary
Bushings
CT with Tap
CT
PILC (Paper Insulated Lead Covered) Cable
Reactor + Bushings
Reactor
Rotating Machine
Three Winding Transformer + Bushings
Three Winding Transformer
The form view of Object Template is pre-specified, however, it can be modified to satisfy different
requirements. For information about the commands in the IDAX C file see the Commands and
Variable Reference in the program help file or included pdf manual.
6.3.1 Overview
6.3.1.1 Introduction to MODS
General
The name MODS stands for MODelling Software.
Main purpose of this software is to facilitate the evaluation of insulation condition of power
transformers. The attention is focused on the estimation of moisture content in solid insulation
(pressboard) and oil conductivity, i.e. on the parameters reflecting the condition of the power
transformer.
When starting up MODS for the first time after installation you will be asked to enter your name and
serial code.
Use the serial name and number from the paper that came with your system.
Creates a new model curve tab with a user defined name or default name "Model-number".
If the file contains more than one measurement you will be asked which of them to open.
Note: For moisture assessment only CHL should be used (if available).
If you open a file that has a corresponding .mods file you will be asked if you want to load that file
instead. MODS will automatically load curve and modeling data from the .mods file if it is available.
Report...
Opens a Report Generator window where user can select reports created in
1) Microsoft Word (default choice)
Microsoft Word report generator allows user to create report from MODS and manipulate with
templates for report creation. The following functions are available:
Compile report - creates a report from modeling results
Add a template - allows for addition of a new template to the existing list of templates
Delete template - deletes the selected template
Modify template - modifies the selected template using Microsoft Word
New template - creates a new template
Export curve...
Export the currently selected measurement curve tab as either a .txt file or a .csv file. The text export
exports rows of three tab delimited values: frequency, e’ and e’’. It is possible to import the exported
txt file back into MODS. The CSV export makes a CSV export that contains the sweep name,
frequency, e', e", tan delta and PF values. If it has an associated model it also exports the model
settings.
Remove Curve
Displays a drop-out menu with all model and measurement curves loaded in MODS. From this menu
a curve can be selected and removed. Alternatively, it is possible to remove a curve by right-click on
the curve tab and select "Close".
Edit values
Displays a dialog where you can remove or edit measurement values.
Exit Alt+F4
Exits MODS.
The value of chosen entity will be shown in corresponding field in Material 2 panel.
Network Frequency
Power Factor and Tan Delta are frequency dependent, therefore the correct Network Frequency
must be chosen.
Available frequencies are:
50 Hz (Europe, Asia, etc)
60 Hz (USA, Canda, etc)
Activation Energy
This dialog is used to change the activation energy for the two insulation materials (default cellulose
(0.9) and oil (0.5)).
Reference temperature
This is used to set the reference temperature. This is used for the display of Power Factor and Tan
Delta in temperature curve mode.
Configure models
The configure models dialog makes it possible to add, remove and export model files.
At the top the currently installed models are listed with their name and file name.
Export saves a model file to a file.
Remove removed a model. Note: Unless you have exported it it will be lost unless you reinstall
MODS.
Import adds a model file to the list of models used. Note that you will have to restart MODS for the
change to take effect.
Close closes the dialog.
Coordinates
Displays position of the cursor in the graph window.
Focus Marker
When the user rich-clicks in the vicinity of a data point the frequency and the value for that point is
displayed.
Displays the location of power frequency on frequency axis and a window with Tan Delta or Power
Factor for both model and measurement.
Typical values of X and Y for a core form power transformer are X: 15-35% and Y: 15-25%
but in some designs X can go up to 55%. For Shell form power transformer, who normally
have a more solid insulation structure between windings X and Y is greater, sometimes up to
70-80%.
Auto (Best of R and K) - both algorithms are employed and the results from algorithm
giving best match is displayed
Comparison - both algorithms are employed and the results from both are displayed.
The user can select which results to use.
This setting allows you to set if a limit to oil conductivity should be used and if so to change the used
limits.
This setting allow you to change from using an automatic frequency range to set it manually. If set to
manual you can set the start and stop frequency.
C0 Settings
Geometrical capacitance, C0 settings can be used to enable or disable:
automatic C0 match when a measurement file is loaded
automatic readjustment of C0 when model parameters change
MODS Help
Invokes Help for MODS software.
The horizontal axis is frequency or temperature whereas from the vertical axis the values
permittivity, Power Factor and loss tangent can be read.
A standard pop-up color dialog is used for changing colors of model and measurement curves and it
is invoked by right-clicking on the colored rectangle next to the labels drop down menu.
To switch to temperature curves you can either use the option "Show Temperature Curves" in the
Diagram menu or use the "Temp." button in the lower right.
6.3.2.7 Zooming
Zoom feature is employed as follows.
Zooming in:
left click in a corner of the area to be magnified
hold down left mouse button
draw a rectangle around the area to be magnified
let go left mouse button
Zooming out:
right mouse click in the graph window restores the graph window
3. Information button
Using up and down arrows the geometrical capacitance can be adjusted to other values.
The button Show values invokes a window showing selected parameters of the
measurement and the test object at given power frequency.
The button Information brings up a window with the frequencies used for interpolation to obtain
values at power frequency.
6.3.2.8.4 Information
Information window displays the information about test object and measurement settings contained
in the IDAX data file.
When the Temperature correction window is closed, the Temperature correction panel is activated.
In order to change the temperature:
enter the desired value in the text field
use blue up- and down arrows to adjust temperature
Note: This temperature correction is only valid for a single material insulation with a certain
activation energy.
3. Geometry panel
4. Materials panel
The check boxes beside geometry parameters are used when geometry is unknown and it is
necessary to find these parameters.
Moisture content and oil conductivity are to be estimated. Automatic finding of these
parameters is enabled by checking corresponding check boxes.
Normally the low voltage winding is located closer to the core and high voltage winding encloses the
low voltage winding. In order to ensure required dielectric strength as well as provide sufficient
cooling the windings are kept apart. This is achieved by using a number of cylindrical pressboard
barriers between two windings. The barriers are mechanically separated by a certain amount of
spacers. Oil is allowed to freely flow in ducts between barriers and spacers. A simplified illustration of
two windings is given by a picture bellow.
Core
Oil
Pressboard barriers
Pressboard spacers
Describing the complex geometry of barriers and spacers/stick between the windings as one simple
capacitance is a simplification of the real conditions. This is one of the reasons why the modeled
geometry may not correspond exactly to the "real" geometry in the transformer. In most cases, best
matching/modelling is achieved by letting the SW automatically calculate the geometry.
The permittivities of oil, ε oil, spacers, ε spacers and barriers, ε barriers are complex functions of both
frequency and temperature resulting in the total or measured frequency response:
Y 1− Y
ε * (ω , T )duct = +
1− X X 1− X X
*
+ * *
+ *
ε spacers ε barriers ε oil ε barrier s
Oil can be represented by its relative permittivity 2,2 and conductivity, thus its frequency response
can be modelled as follows:
* σ
ε oil = 2, 2 − j
ε 0ω
Frequency response of pressboard is, however, much more complex function, for modelling of which
a number of reference frequency responses are used. The reference responses are obtained from
measurements on pressboard and impregnated paper samples having controlled parameters.
The influence of temperature must be also taken into account. Temperature changes result in
horizontal shift of the pressboard response, therefore it can be modelled using Arrhenius type
relationship:
For moisture assessment, the inter-winding capacitance, CHL, should always be used when available.
If CHL is not available, e.g. on an autotransformer, use CHT if available. If there is no tertiary winding
or for any other reason measurements between windings are not possible, select the winding closest
to the core for analysis. Most often this is the low voltage winding and CL should be used for
moisture assessment.
For best accuracy of the modelling it is important to perform the analysis over a sufficient frequency
range. Critical is the lowest frequency measured and this is pending insulation temperature. The
table below gives maximum stop frequency for various measurement temperatures.
If the data file contains the insulation temperature written in a corresponding field in the data file,
the dialog window asking to confirm this temperature will be displayed. (Note: this feature is
available only when loading IDAX D-files). If no temperature is given you will be asked to enter a
temperature.
After the measurement file is successfully loaded, the Model tab is active. To activate the
measurement tab, click on it (see pictures below)
There are three categories, Geometry, Material 1 and Material 2 containing six parameters in Model
tab. Default values and properties of these parameters are as follows:
Temperature can be varied only manually, therefore, if the temperature is not read from the
measurement file (i.e. not entered or erroneous) one should ensure that correct temperature value
is entered before proceeding. Since Material 2 is oil, having almost constant permittivity in given
frequency range it is set to 2,2 but can be changed if needed.
Automatic optimization of the parameters can take from a few seconds up till several tens of seconds
depending how much parameter adjustment is required to fit the model and which algorithms are
employed. When finished the final values of parameters are displayed in their respective boxes.
Starting point of manual match is the default values of parameters, which can then be either
increased or decreased. Values can be typed directly into respective fields or using coarse and fine up
and down arrows.
1. Launch MODS.
2. Load the desired file, for example "Dry.idf".
3. Default is that all checkboxes are selected.
4. Adjust temperature, Temp (°C), if not loaded from D-file. In "Dry.idf" the temperature
field is not available, and temperature is set at the default 20C therefore manual
adjustment to 18 degrees Celsius is done.
5. Leave the automatch boxes checked or enter a qualified guess qualified guess for
amount of barriers (X) and spacers (Y). A common estimate for barriers and spacers is
X = 20% and Y=20%.
6. Check the moisture content and oil conductivity checkboxes to enable automatch of
these parameters.
7. Selecting desired automatch algorithm or combination in Oil Conductivity Settings....
8. Press Automatch button.
9. Analyse the results. Obtained results show a match where measured and modeled
responses are very well matched. Moisture content is 1% by weight and oil
conductivity is 1.7 pS/m.
The report dialog shown below allows user to create a report by pressing Compile report. A report is
created using a selected template from the Report template list. This list can be changed by adding
new (Add template), deleting (Delete template) and creating new templates (New template). In
addition a selected template can be modified (Modify template) according to user needs. The
modification is done using MS Word as text editor. Note that the curve used in the report is the same
as is displayed in MODS so if you want a report with temperature curves just switch to temperature
mode before compiling the report.
Note that you must have MS Word installed to be able to generate a report.
7 Calibration
It is recommended to calibrate the IDAX at least once a year. There are two options regarding
calibration of IDAX:
The second option means that the IDAX instrument is available since it does not need to be sent
away for calibration. The calibration procedure is easy to perform using the optional Calibration Set,
which consists of:
Calibration Box CAL 300
Calibration Software
8 Appendix
8.1 Appendix A – IDAX 4.1 error messages
System messages
No Message Explanation
101 Filename missing File name not specified. Specify a file name.
103 Not string A string variable is expected.
104 Could not open file: Wrong file name, file is missing or path is incorrect.
105 Could not open or create Problem accessing Windows Registry.
key
106 Key could not be deleted Problem accessing Windows Registry.
107 Redeclaration of variable A variable is declared two or more times in the C–file.
108 Error writing to file The log file from a Calibration or a Self test could not be
created.
109 Incorrect data type The Data type specified does not exist or measurement
template not marked.
112 Cannot start measurement Displayed when IDAX is not properly set up in order to
perform a measurement.
113 Error reading Windows Problem accessing Windows Registry.
registry
No Message Explanation
typing error.
216 Wrong number of arguments Too few or too many arguments in a variable.
217 Syntax error Undefined error in the syntax. Usually accompanied by
additional information.
218 To few radio buttons in There must be at least two buttons in a radio group
group ("RB")
219 Not a radio buttons item The function is assigned to a value that does not match
("RB") any radio button. Most likely a typing error.
220 Not an array of integers Parameter values have to be of the same type as in the
parameter declaration.
221 Invalid bitmap "Image" Image is not a *.bmp file or the file is corrupted.
222 Variable not declared All new variables must be declared with a "New"
function.
223 "AmplitudeScale" not "AmplitudeScale" must be assigned. See User's Manual
assigned for options.
224 "VoltageSource" not "VoltageSource" must be assigned. See User's Manual
assigned for options.
225 "FileType" not assigned "FileType" must be assigned. See User's Manual for
options.
226 "SpecVersion" not assigned "SpecVersion" must be assigned. See User's Manual for
options
227 "DraftSpecVersion" not "DraftSpecVersion" must be assigned. See User's Manual
assigned for options.
228 "Configuration" not "Configuration" must be assigned if using a Termination
assigned Box with relays. See User's Manual for options.
229 Value not assigned If cat 'P' is assigned, the field can not be left blank.
No Message Explanation
313 Wrong number of arguments Too few or too many arguments in a command.
314 No "Amplitude" "Amplitude" must be assigned. See User's Manual
Appendix A.5 for options.
315 No "Frequency" "Frequency" must be assigned. See User's Manual
Appendix A.5 for options.
319 No measurement to comment Comments are possible during a measurement or
directly after the measurement.
320 No current measurement No measurement which has been started.
321 DSP variables wrong The DSP-variables (e.g. MinTime, MinCycles, etc.) are
wrong.
322 DSP code not found No or wrong DSP code found in the system.
323 Index does not exist Problems accessing Windows Registry.
in NT registry
324 Could not interpolate Could not interpolate transfer functions.
calibration data
325 No voltage source selected A voltage source must be selected before measuring.
327 Could not get info from Problems with the connection to the DSP instrument.
DSP
329 "MeasurementPerPoints" One measurement at each frequency must be
must be greater than zero. performed.
330 DSP reset failed Problems with the connection to the DSP instrument
331 "Else" without "if" Syntax error in C-file.
332 "End" without "if" Syntax error in C-file.
333 Only variables or Only variables or constants are allowed in an if–else
constants are allowed statement.
334 Types does not match Wrong type of data (e.g. string assigned to an integer).
339 Error updating "D-File" Measured data was not properly added to the D-file.
340 Invalid voltage The desired voltage is out of the specified limits.
Possible typing error in the C-file.
341 Invalid current The applied settings results in a current that is too high.
342 Invalid frequency The desired frequency is out of the specified limits.
Possible typing error in the C-file.
343 Cable parameter not found: The specified parameter was not found in the sample
cable.
344 Cable parameters not found No cable data was found.
345 "Temination box" parameter The specified parameter was not found in the
not found: "Termination Box".
346 "Termination box" No "Termination Box" data were found.
parameters not found
347 Output voltage is not The applied voltage differs from the desired.
within specified levels.
351 "InitMeasureZ" settings: Not possible to perform InitMeasureZ measurement.
This message is followed by one of 340-342.
352 No "Amplifier type" High Voltage amplifier not selected.
selected
356 Electrometer Overload Overload in Electrometer instrument during a
measurement.Measurement point is probably
inaccurate. Message also written in point property.
No Message Explanation
357 Error reading backplane Problem accessing instrument backplane.
serial
358 Cannot find the Name Syntax error in Calibration or Self test files.
option in file
359 Cannot find the Connection Syntax error in Calibration or Selftest files.
option in file
360 Error in file ... at line Syntax error in Calibration or Selftest files.
...
361 Overvoltage An overvoltage occurred.
364 Measured capacitances Disagreement between values of capacitance measured
don't match for different configurations
365 Specimen capacitance below Measured capacitance below specified limit.
limit.
366 Specimen capacitance above Measured capacitance above specified limit.
limit.
367 Measured DC Measured DC current exceeds limits set by
current>MaxDCCurrent MaxDCCurrent variable.
368 Measured hum Measured interference (hum) current exceeds limits set
current>MaxHumCurrent by MaxHumCurrent variable.
Example of red and green/finished and ready to measure, here the finished is chosen:
4) If you try to use an old version of IDAX SW, version 3.2 or earlier but the firmware in IDAX is for
IDAX SW 4.0 or newer, the IDAX SW do not understand the incapability and it usually result in error
347. Please check IDAX SW and if version 3.2 or earlier. Please upgrade to 4.0 or newer (this new SW
will automatically upgrade firmware if necessary).
1) When performing UST measurement, measuring electrode, Lo, is connected together with Ground
electrode or Lo is connected to ground:
check measurement set-up and make sure that measuring electrode, Lo, is connected
to a non-grounded terminal of the test object and that Ground electrode is connected to ground
For more information of actual measured capacitance, please see Message Window.
The far most common reason is too low resistance between measurement electrode and guard, e.g.
measuring a UST configuration between high and low voltage windings of a two winding transformer,
the low voltage winding has too low impedance to ground (inductive voltage transformer connected,
internal damage of transformer, neutral connected to ground via a Peterson coil. For a GST
measurement, same applies to guard electrode, i.e. a too low resistance Guard electrode to ground
may introduce DC currents
Make sure that floating electron has high resistance to ground, if not possible use
another set-up (e.g. measure to ground without use of guard)
It is possible increase limit level for DC current in Measurement Template, but only
when difference is very small and all other possibilities excluded
The impedance of a sample is measured by applying a voltage across the sample. This voltage will
generate a current through the sample. By accurately measuring the voltage and the current, the
impedance can be calculated, see illustration below.
The voltage (and the current) is generated by a voltage source. There are currently two internal
voltage sources available in the IDAX system, which can deliver a maximum peak output of 10 Vpeak
and 200 Vpeak, respectively. The voltage is measured by means of a voltmeter and the current is
measured by an ammeter or electrometer which acts as a current-to-voltage converter. The
analogue signals (voltages) are then converted to digital samples of the signals that are used in
subsequent calculations.
Polar:
Z = Abs{Z } ϕ = Arg {Z }
Rectangular:
Z Re = Re{Z } Z Im = Im{Z }
Two simple models which are usually used in circuit analysis, although more seldom in insulation
analysis, are capacitance, C, and resistance, R. The equivalent RC circuit models available are series
and parallel models (see Fig.1) calculated as follows:
a) b)
Parallel:
RP
Z=
1 + j ωRP C P
1
CP = Re
jω Z
1
RP =
1
Re
Z
Series:
1
Z = RS +
j ωC S
1
CS = −
ω Im{Z }
RS = Re {Z }
Another model, more often used in insulation diagnostics, is the complex capacitance model
describing the insulation impedance as a complex capacitance, where the imaginary part of the
capacitance represents the losses. The complex capacitance model is defined as follows:
1
Z=
j ωC C = C ′ − jC ′′
where
Complex C:
1
C′ = Re
j ω Z
∆C ′ = C ′ + k
1
C ′′ = − Im
j ω Z
The ∆C' is defined as the capacitance, C', with an arbitrary constant k (usually negative) added. The
aim of this parameter is to make it possible to distinguish between small changes in capacitance in
graphical presentation.
A model, very often used in insulation diagnostics, is a description of the insulation impedance as a
capacitance combined with a dissipation factor, tanδ, or a power factor (PF or cosϕ). The
capacitance, tanδ and cosϕ/PF are defined as follows:
C, PF, Tan-Delta:
1
C′ = Re
j ωZ
Re {Z }
PF = cos ϕ =
Z
Re {Z }
tan δ = −
Im{Z }
If tanδ and cosϕ (PF) are small, then tanδ » cosϕ = PF (E.g. tanδ=0.1 corresponds to PF=0.0995)
Insulation diagnostics is based on material characterization and therefore material models are often
used. To be able to define material parameters from measured impedance Z the geometry of the
sample, described in terms of the geometrical capacitance C0, has to be defined. In the illustration
below, a vacuum (or air-filled) capacitor of defined geometry is shown. Since no "material" is between
the electrodes, the capacitance of a) is the geometrical capacitance.
a) I b) I c) I
Material parameter models based on a geometrical capacitance C0 and material parameters.
In the above illustration b) and c) a material is inserted between the electrodes and it will influence
the current, I, flowing in the circuit. The influence of the material can be described by different
parameters using either a dielectric model or a conductive model. In the dielectric model the
"material capacitance", the permittivity, is a complex function describing both the capacitance and
the loss. Whereas in the conductive model the capacitance is described by a permittivity and the loss
by a conductivity (or resistivity). The dielectric and resistive models are derived as follows:
1
Z=
j ωC
C = C0 (ε ′ − jε ′′)
Dielectric:
1
ε ′ = Re
j ωC Z
0
∆ε ′ = ε ′ + k
1
ε ′′ = − Im
j ωC0Z
Resistive:
1
ε ′ = Re
j ωC0 Z
C0
ρ=
1
ε 0 Re
Z
1
σ=
ρ
If geometrical capacitance, C0, is unknown it can be set by the user by entering a permittivity
(dielectric constant), ε', for the material. Entering a permittivity, ε', will let the system calculate an
approximate C0 making use of the material models available. Yet, one must be aware that the
accuracy of the absolute values are limited by the accuracy of the entered permittivity. If the
geometrical capacitance is unknown and an approximate permittivity is given, the geometrical
capacitance is calculated as follows:
C′ 1 1
C0 = = ⋅ Re
εr εr jωZ
The impedance value, Z, used in the calculation is the first measurement point in the actual
measurement.
The IDAX 300 uses voltages up to 200 Vpeak (~140 V RMS). The voltage range can be increased by
adding an external high voltage unit (the high voltage unit is specified separately). The IDAX-300 is
operated using an external laptop/PC.
The IDAX 300 uses a three electrode set-up, which allows for measurements of non-grounded as well
as grounded objects with and without guard.
The IDAX 300 system applies sinusoidal shaped voltages at different frequencies across the test
object and simultaneously measures the current through it. The output voltage is either taken from a
10 Vpeak arbitrary waveform generator, from a 200 Vpeak arbitrary waveform generator or from a 2 -
30 kV peak external amplifier.
Output
Voltage/current range, 10V: 0 - 10 Vpeak, 0 - 50 mApeak
Voltage/current range, 200V: 0 - 200 Vpeak, 0 - 50 mApeak
Frequency range: 0.1 mHz - 10 kHz
Measurements
Inputs: Channel 1, channel 2, ground
Capacitance Range: 10 pF - 100 uF
Accuracy: 1 mHz < f < 1 kHz, C > 1 nF 0.5 % + 1 pF
f < 1 mHz, C < 1 nF 2% + 5 pF
f > 1 kHz, C > 1 nF 2% + 10 pF
Dissipation Factor Range: 0 - 10 (with retained accuracy of capacitance;
otherwise higher)
Accuracy: >1% +0.0003, 1 mHz - 100 Hz, C > 1 nF
>2% +0.0005, 100 Hz - 1 kHz, C > 1 nF
Noise Level: 1 mA or max 1:20 SNR
Test Modes, 2 Channels: UST-1, UST-2, UST-1+2, GST, GST-Guard-1, GST-Guard-
2, GST-Guard-1+2
With 2-ch measurement option additionally UST-
1+UST-2, UST-1+GST-Guard-2, UST-2+GST-Guard-1,
UST-1+2+GST
Calibration: Calibration box allows field calibration, recommended
interval 2 years
General
Mains Input (nominal): 100-240 V AC +/-10%, 47 to 63 Hz
Power Consumption (max): 60 VA, 80 VA (IDAX 350)
Communication Ports: USB 2.0 and Ethernet
Physical
Instrument Weight: 4.9 kg/11 lbs
Case and Instrument Weight: 9.9 kg/22 lbs
Accessories Weight: 8.5 kg/18 lbs (soft bag)
Dimensions: 335 x 300 x 99 mm
17.7 x 6.3 x 16.1 in
Dimensions with Case: 520 x 430 x 220 mm
20.5 x 17 x 8.7 in.
Environmental
Operating Ambient Temp: -20°C to +55°C /-4°F to +131°F
Storage Ambient Temp: -40°C to 70°C / -40°F to +158°F
Humidity: 20% - 95% RH, non-condensing
CE Standards: IEC61010 (LVD) EN61326 (EMC)
PC Requirements
Operating System: Windows XP / Vista / 7
CPU/RAM: Pentium 500 MHz/512 Mb or better
Physical
Instrument Weight: 13.4 kg
Dimensions: 520 x 430 x 220 mm
20.5 x 17 x 8.7 in.
PC
Screen: 12" touch screen, 1600 nits backlight
CPU: Intel Atom 1.8 GHz
Storage: 4 GB flash
Operating system: Windows Embedded Standard 7
VAX214
9 Index
347, 141 Dissipation factor, 144
361, 141 Edit, 89, 90, 91, 92
365, 142 Edit\:Change Sweep Color, 89, 91
366, 142 Edit\:Copy, 89, 90
367, 142 Edit\:Copy All, 89, 90
368, 143 Edit\:Cut, 89, 90
50/60 Hz Data Tab, 54 Edit\:Mark Sweep, 89, 91
About, 61 Edit\:Paste, 89, 90
Automatch panel, 127 Edit\:Read Only, 89, 92
Autoscale, 103 Edit\:Select All, 89, 91
Axis, 102, 103, 104, 105 Edit\:Toggle Sweep, 89, 91
Axis\:Autoscale, 103 Edit\:Undo, 89, 90
Axis\:Grid, 103 Edit\:Unmark Sweep, 89, 91
Axis\:Label, 104 Equivalent Circuit, 144
Axis\:Scale, 104 Equivalent Circuit\:Paralell, 144
Axis\:Settings, 102 Equivalent Circuit\:Series, 144
Axis\:Title, 105 Error Messages, 137
Axis\:Unit, 105 Export, 78, 79, 80, 82, 83
Calibration, 135 Export\:Data, 79
Calibration\:Calibration Procedure, 135 Export\:Measurement Templates, 82
Calibration\:EM0 Calibration, 135 Export\:Object Templates, 83
Calibration\:Full Calibration, 135 Export\:TXT format, 79
Capacitance, 144 Export\:With or without header, 80
Capacitance\:Geometrical, 144 Export\:WMF format, 79
Color, 119 File, 63, 64, 66, 67, 68, 69, 70, 84, 85, 87, 89
Conductivity, 144 File\:Add Templates, 68
Configuration, 100, 102, 106, 107 File\:Add/Remove Templates, 63
Configuration\:Graph, 100, 102 File\:Close Measurement, 63, 69
Configuration\:Information, 100 File\:Delete File/Folder, 63, 89
Configuration\:Models, 100, 107 File\:Exit, 63, 89
Configuration\:Power Data, 106 File\:Export, 63
Configuration\:Power Frequency Data, 100 File\:Import, 63
Configuration\:Send to..., 100 File\:Load D-file, 63, 70
Configuration\:Table, 100, 106 File\:New, 63, 64
Current, 148 File\:New Measurement Sequence, 67
Current\:Range, 148 File\:New Measurement Template, 66
Diagram, 92, 93 File\:New Object, 64
Diagram\:Zoom In, 92 File\:New Object Template, 66
Diagram\:Zoom Out, 92, 93 File\:Open Measurement, 63, 69
Directives, 148 File\:Print, 63, 84
Directives\:Emmision, 148 File\:Remake Measurement, 63, 69
Directives\:Immunity, 148 File\:Remove Templates, 68
Directives\:LVD, 148 File\:Rename File/Folder, 63, 89
Directives\:Mechanics, 148 File\:Report, 63, 85