Scale Terminal Calibration

Mettler-Toledo
IND560

Encompass Automation & Engineering Technologies LLC


Industrial Automation & Power Distribution | Engineering | Project Management | Electrical Contracting & Installation | Turn-Key Project Services
622 Eckel Road, Perrysburg, OH 43551 USA/ Phone (419) 873-0000 / Fax (419) 872-2612 www.EAETech.com
Copyright 2010 EAET-LLC. Confidential and Intellectual Property / Not For Disclosure Nor Abstraction / All Rights Reserved
How To : Traditional Calibration

Scale System Calibration


 Main area of proficiency:
- Mettler-Toledo IND560 Scale Terminal

2 Training Purposes Only


How To : Traditional Calibration

Entering Setup Mode


 Press the Down Arrow to view
the SETUP softkey.
 The configuration of the
IND560 terminal is accessed
through the SETUP softkey

 Softkeys

3 Training Purposes Only


How To : Traditional Calibration

 With the Down Arrow navigate to Scale Sub-


branch
 Use the UP and DOWN navigation keys to
move focus through the branches of the setup
menu tree.
 Press the RIGHT navigation key to expand a
branch and the LEFT navigation key to
collapse a branch. When the focus is on a sub-
branch, focus can quickly be moved back to
the main branch by pressing the LEFT
navigation key.
 When a branch such as Calibration is in
focus, press the ENTER key to display the
setup screen for that function.

4 Training Purposes Only


How To : Traditional Calibration

 Navigate to Calibration
- Press the Down Arrow twice.
 When at Calibration
- Press Enter

5 Training Purposes Only


How To : Traditional Calibration

 Navigate to Calibration
- Press the Down Arrow four (4) times.
 Geo Code
 Base Serial Number

 Calibration Units
 Linearity Adjust

6 Training Purposes Only


How To : Traditional Calibration

 Press Zero softkey


- Remove all weight from scale

- Press Start

7 Training Purposes Only


How To : Traditional Calibration

 Press Span softkey

- Last test weight value of 100 Lb.


is displayed

- Press Enter twice

- Add test weight either manually or


via Automated Test Weight
System
8 Training Purposes Only
How To : Traditional Calibration

 Span
- Press Start

9 Training Purposes Only


How To : Traditional Calibration

- Press Exit twice when done

- Unit will return to normal weighing


mode

- Weigh Hopper Discharge Feeder


needs to be placed back into
Automatic.

10 Training Purposes Only