Sie sind auf Seite 1von 12

C H A P T E R 2 1

Special Instrumental Methods

Strictly speaking, there is no such thing as a instruments in this chapter.


non-instrumental method, since even ord- The 15 instruments whose usefulness for
inary volumetrie analysis depends ultimately partieular types of analytieal problems we
on use of an analytieal balance for standard- shall attempt to explain can be divided
ization. This chapter will be restricted to roughly into two categories: 1) those that are
explanations of the functions in analytical primarily used for qualitative and quantita-
chemistry of several instruments that are tive analysis of mixtures of compounds; and
"special" in that they are not apart of the 2) those that are primarily used for identi-
common equipment of analyticallaboratories. fying single pure compounds. It should be
These instruments differ in relative availability emphasized that this division is a very rough
in rough proportion to their cost. For in- one; for example, while the infrared spectro-
stance, flame photometers, polarimeters and photometer is almost uniquely valuable in its
less expensive types of fluorometers are found ability to identify an organic compound and
in quite a few food laboratories, while the groups that it contains, there are numerous
mass spectrometers and nuclear magnetic references to its use in quantitative analysis of
resonance devices are uncommon. In modern mixtures of several compounds (more often,
times, absorption spectrophotometers for the however, in the drug than in the food field).
visual-uItraviolet range such as the Beckman Nevertheless, the classification is a useful
DU have co me into almost universal use, so one, and we have employed it here.
we feel there is no need to discuss these

Instruments for Qualitative and


Quantitative Analysis of Mixtures

in principle with solution absorption spectro-


Atomic Absorption photometry except that the sampie to be
Spectrophotometers examined is in the vapor phase, produced
It has been pointed out! that atomie and maintained by a flame." Atomic absorp-
absorption spectrophotometry "is identieal tion instruments find their practical use,

459

F. L. Hart et al., Modern Food Analysis


© Springer-Verlag New York Inc. 1971
460] Special Instrumental Methods

however, as do flame photometers, primarily one-fifth as much or even less, is within the
as cheaper substitutes for the emission spect- means of more laboratories than the emission
trograph. spectrograph, and in addition is a more
Introduction of the atomic absorption sensitive means of determining certain ele-
spectrophotometer by Walsh in Australia in ments such as zinc; and 3) as noted above,
1957 2 was rapidly followed by the develop- the emission spectrograph can determine the
ment of commercial instruments, and this widest range of elements in the shortest time,
technique has become quite popular in the without prior knowledge of what elements are
years that have elapsed since that time. Like present. Under normal conditions only
the flame photometer, a solution of the metallic elements can be determined by any
sam pie is sprayed into a flame where it is of these instruments, and none determines
heated to incandescence. It differs from the how the element is combined in the sampie.
flame photometer in this respect: in the Method 6-24 is an example of an atomic
flame photometer, the emitted light of a absorption method.
particular wavelength, which comes only Manufacturers of atomic absorption spec-
from the activated (ionized) atoms of an trophotometers include the following:
element, is measured; in the atomic absorp-
tion spectrophotometer, the vaporized union- Aztec Instruments, Inc., 2 Railroad Place,
ized atoms absorb some of the light emitted Westport Conn.
from a hollow cathode lamp whose cathode Instrumentation Laboratory, Inc., 113
gives off light of the resonance frequency of a HartweIl Ave., Lexington, Mass.
particular element, and it is the amount of Jarrell-Ash., 590 Lincoln St., Waltham,
this absorption that is measured. One of the Mass.
claims for the atomic absorption instrument Perkin-Elmer Corp., 702G Main Ave.,
is that because only a minor proportion of Norwalk, Conn.
the vaporized atoms of an element is ionized,
atomic absorption spectrophotometry is more
sensitive than flame photometry. Claims for
Emission Spectographs
superiority over emission spectrography are The emission spectograph consists of a
also made on the basis that instrument read- power source to heat the sam pIe to incan-
ings can be made direct1y proportional to descence, a device (either a prism or a grating)
concentration. On the other hand, deter- to break up the spectrum of the emitted
minations of many elements can be made light, and a means of recording the spectrum
simultaneously with an emission spectro- (either a photographie plate or film or a
graph (and it is not even necessary to have bank of photocells). Grating spectrographs
prior knowledge of what elements are pres- have largely replaced prism instruments.
ent), while the atomic absorption spectro- Emission spectrographs find their chief use
photometer requires aseparate hollow cath- in food analysis either in the quantitative
ode lamp, and a separate determination, for analysis of plant materials for metallic ele-
each element (or at most each pair of ele- ments or in semi-quantitative tests for toxic
ments) estimated. elements such as arsenic, lead, beryllium and
As a practical matter: 1) the flame photo- thallium. For such purposes the sampie is
meter (which is the cheapest of the three usually ashed, taken up in dilute acid, and
instruments) finds its chief use, and is most applied to a crater in a carbon or graphite
efficient, in the determination of the alkali electrode. An are or spark is then formed
metals Na, K, and Li; 2) the atomic absorp- between this electrode and another of the
tion spectrophotometer, because its cost is same material, and the spectrum of the
Instruments for Qualitative and Quantitative Analysis of Mixtures [461
burning sam pIe is recorded and measured. note also particularly Selected Reference W.
In the metal industries, automatic recording The method of the late Waddy Mathis was
spectrographs employing photocells are used worked out in the laboratory of one of the
to measure the intensity of the chosen line in authors of this book, and is still in use there.
each element being determined, but such Both of these methods (with slight modifi-
devices are quite expensive, and in food cation) have been adopted by the AOAC
analysis the cheaper instruments in which the (AOAC Methods 41.001-41.016).
spectrum is recorded on a photographic Manufactures of emission spectrographs
plate or film are still in use. In photographic inc1ude:
recording, after development the film (or
plate) is placed in a densitometer which Applied Research Laboratories, Inc., 3717
first locates the lines corresponding to the Park Place, Glendale, Calif.
various elements present and then measures Baird-Atomic, 33 University Road, Cam-
their intensities. For semi-quantitative trace bridge, Mass. and Jarrell-Ash Co., 590
element detection, a d-c arc of 5-30 ampere Lincoln St., Waltham, Mass.
current is the preferred means of excitation
because it is the most sensitive, but in quanti- Spex Industries, Inc., 3880 Park Ave.,
tative analysis the a-c spark is employed Edison, N.J., is a source of spectrographic
because it yields more uniform spectra. supplies; graphite electrodes are available
For a more extended discussion of the from National Spectrographic Laboratories,
details of emission spectography the reader is Inc., 6300 Euclid Ave., Cleveland, Ohio, and
referred to the artic1e by Scribner and Union Carbide Corp., Carbon Products Div.,
Margoshes in Chapter 64 of Part I, Volume 6 270 Park Ave., New York, N.Y. Extremely
of Kolthoff, Elving and Sandell's Treatise on high purity elements and their salts for
Analytical Chemistry (New York-London- standardization, manufactured by Johnson-
Sydney: Interscience Publishers, 1965). Two Matthey Co., London, England, are available
things should be noted here, however: I) from Jarrell-Ash Co.
The densities of the spectrallines obtained in
emission spectography are not strictly pro-
portional to the concentrations of their
Flame Photometers
respective elements; and 2), because it is Flame photometry is essentially emission
nearly impossible to take two "shots" under spectrography using as excitation source a
completely identical conditions, it is the flame rather than the higher-temperature
common practice in quantitative analysis to eleetric are or spark. As a matter of fact,
correct for small variations in burning, etc., Lundegardh wrote two books describing
by adding a uniform amount of an element results obtained with an ordinary spectro-
such as cobalt (whieh is not presen~ in graph equipped with an acetylene-air flame
measurable amount in the sam pIe) to both source. 4 However, the modern-day flame
sampIe and standards. The ratio of the den- photometer, besides being a much less ex-
sity of the line of the element being deter- pensive device, in our opinion is most useful
mined to the density of a chosen cobalt line is just when the temperature of its flame, e.g.,
then what is calculated, and graphs are con- propane-air, is not high enough to excite most
structed relating these density ratios to con- elements, and it consequently can determine
centrations of the elements in question. the alkali metals with minimum interference.
Examples of emission spectrographic It is possible, by using higher-temperature
methods especially designed for plant and flames (hydrogen-oxygen at 2900 0 K or
food analysis are given in Text Reference 3; cyanogen-oxygen at 4800 0 K s), and separating
462] Special Instrumental Methods

the emission lines with a monochromator, to iIluminating ultraviolet lamp. For accurate
adapt the flame photometer to the deter- analysis it is necessary to ensure, by the use
mination of numerous other elements, but either of filters or of prisms or gratings, that
only at the price of loss in simplicity and both the incident and fluorescent light are
freedom from interference, and increase in monochromatic.
cost. Fluorometric methods are valuable where
An example of the determination of sodium they are applicable (i.e., when the compound
and potassium by flame photometry is being determined is fluorescent or can be
Method 1-17. converted to a fluorescent derivative) because
Relatively inexpensive flame photometers they are usually highly specific and about 10
that yield accurate results for Na and K times as sensitive as absorption methods; the
(particularly when using a lithium internal calibration curves are also usually linear
standard) are available from: over a 50-fold concentration range. There
are pitfalls, however; for instance, other
Advanced Instruments, Inc., 45 Kenneth compounds present in a solution of a sam pie
St., Newton Highlands, Mass. can quench the fluorescence of the fluores-
Coleman Instruments Corp., 42 West cing ingredient, and a practical method must
Madison St., Maywood, Ill. and be so designed as to avoid such interferences.
Jarrell-Ash Co. (See above). The better grade fluorometers use grat-
ings or prisms and a recorder of some sort to
separate and record the fluorescent spectrum
Fluorometers
of the substance being measured; when
Disregarding the cause of fluorescence, the this is done, a fluorescence spectrum is
visually observed phenomenon is that when obtained, frequently a mirror image of the
a substance is illuminated with light of one absorption spectrum, wh ich serves to charac-
wavelength, it emits light of another wave- terize the substance in addition to deter-
length; if the emitted light persists after the mining how much of it is present. 6
illumination ceases, the substance is said to Examples of fluorometric methods of food
phosphoresce. These phenomena have been analysis are those of Watkinson for selenium, 7
explained as being due to the release of energy McFariand for diethylstilbestrol in meat,8 and
in the form of light by an excited moleeule Roberts for cholesterol in egg noodles. 9
when it reverts to the ground state. A co m- Fluorometers of the quality needed for
plete explanation would require a knowledge such methods may be obtained from:
of quantum mechanics, but an understanding
of the theory of fluorescence is unnecessary American Instrument Co., Inc., 8030
for its use as an analytical too1. Georgia Ave., Silver Spring, Md.
In principle, a fluoro meter is a device in Baird-Atomic, 33 University Road, Ca m-
wh ich a solution of the substance being bridge, Mass. and
analyzed is illuminated by an ultraviolet Farrand Optical Co., Inc., 535-A South
light source of fixed wavelength, and the Fifth Ave., Mt. Vernon, N.Y.
intensity of the (usually visible) light emitted
by the irradiated solution is measured. The
Gas Chromatographs
incident radiation of the source must, of
course, be separated from the light emitted It can be questioned whether gas chroma-
by the sampie ; this can most readily be done tography is an "instrumental" method of
by measuring the light from the sam pie at a analysis in the same sense as are the deter-
right angle to the be am of light from the minations made by means of the other instru-
Instruments for Qualitative and Quantitative Analysis of Mixtures [463

ments described in this chapter. It is true that stages as the more volatile ingredients are
commercial gas chromatographs carry various removed. The more important developments
dials and controls on their faces and display in gas chromatography, however, have
their results on x-y recorders, but all that involved the discovery of new detectors that
these controls do is adjust temperatures and either were more sensitive than previous
gas pressures, maintain a detecting device, and ones, or had the desirable qllality of reacting
record when that device indicates that an toward only certain types of compounds. The
ingredient of a sam pie has come through a first and most general type of detector was
column and how much there was of it. In the thermal condllctivity gauge, wh ich was
contrast to the other instruments we will basically an electrically heated fine wire
describe, the gas chromatograph analyzes a mounted axially in a narrow tube; as the
sampie not by detecting its ingredients current of gas flowing through the gas
through their characteristic intramolecular chromatograph tube carried a sampIe ingred-
properties but by physically separating them. ient over this wire, its temperature (and
The heart of the gas chromatograph is a consequently its resistance) changed, and a
tube of glass or metal, of one of two types: Wheatstone bridge device converted this
either it is of appreciable diameter (about i" change to a small e.m.f. that activated the
o.d.) and fi11ed with an inert base coated recorder. A later device was the flame tem-
with an adsorbing medium, or it is a capillary perature detector; this consisted of a hydro-
whose inner surface bears a coating of adsor- gen flame impinging on a thermocouple. As
bent. A portion of the sam pie being analy- an organic compound emerged from the
zed-usually in the form of a solution con- column, its heat of combustion was added to
taining only milli grams or less of the ingred- that of the hydrogen, causing an increase in
ients being determined--is injected onto the flame temperature and change in e.mJ. of
heated column, and a stream of inert gas the thermocouple that was translated into a
(nitrogen, argon or helium) is passed through peak on the recorder.
the tube and onto the detecting device. The Nearly a11 of the later more sensitive detec-
gas carries the sam pie ingredients through the tors, as weil as those possessing specificity
tube at rates that vary inversely with the for particular types of compounds, have been
tenacity with wh ich they are held by the devices involving ionization. One exception
adsorbent the tube contains; in effect a sort is Coulson's microcoulometric detector, in
of distillation takes place, and the efficiency which the effluent compounds are burned
of a tube packing is sometimes caIculated in and then titrated for their chlorine or sulfur
terms of so many theoretical plates. As each contents. This device has been employed
compound reaches the detector at the end almost exclllsively for the detection and
of the tube, it causes the detector to activate a determination of organic chlorine or sulfur
recorder. This recorder plots a height vs time pesticide residues in foods; for this purpose
curve showing a peak for each time interval it has the advantage that non-pesticide
that a compound leaves the tube; the height residues carried through the column do not
of each peak (or more exactly, the area affect the recorded curve.
enc10sed by it) is proportional to the quantity Ionization detectors, in general, possess
of the compound present in the sam pIe higher sensitivity than thermal conductivity
aliquot. devices. They also differ in that they do not
The various gas chromatographs differ in respond to all compounds. There are several
control features such as their ability to make types, but one that was and still is being
rapid temperatllre changes to permit pro ces- extensively used in pesticide residue analysis
sing a sam pie in increasing temperature is the electron capture detector (Fig. 21-1):10
464] Special Instrumental Methods
Teflon tube sensitive to most organic phosphorus com-
pounds it is much less satisfactory for the
estimation of residues of organie phosphorus

-
pesticides that contain no chlorine atoms.
Inlet Outlet This deficiency was corrected with the
discovery by Giuffrida l l that a detector
composed of a hydrogen flame impinging on
a platinum coil coated with an alkali metal
salt had a sensitivity for phosphorus co m-
gauze
I cm
source pounds up to 10,000 times that of a regular
flame ionization detector. If the alkali salt
Fig. 21-1. The electron capture detector (after used were ahalide, the response to halogen
Lovelock). compounds was actually depressed. The
original coating material was sodium sulfate,
This deviee consists of achamber con- but potassium chloride was eventually adop-
taining two electrodes and a radioactive ted as the salt giving the best results.
ionizing source. A potential is applied across As was pointed out in Chapter 16 (Method
the electrodes just sufficient to collect all ions 16-11, Note 5), "gas chromatographic
and electrons. As the carrier gas passes methods identify aseparated compound only
through the chamber it is ionized by the beta to the extent that they show that, with a par-
partieies from the radioactive source, but ticular column and a particular form of detec-
when a sampie ingredient which is electron- tion, the compound gives a peak on achart at
capturing emerges from the chromatograph the same place as does the compound it is
tube and enters the ionization chamber, assumed to be." In order to make more
electron capture occurs to give negative ions certain identification of gas chromatographie
which are rapidly removed from the system by fractions, various authors have worked out
recombination. The ion current is thereby means of transferring these fractions to infra-
reduced, and this drop in current is translated red spectrophotometers or mass spectrometers,
into a peak on the recorder curve by suitable particularly the latter. This subject is discussed
devices. 10 The fact that this detector is below under "Mass Spectrometers".
exceedingly sensitive to halogen-containing For a more detailed yet concise discussion
compounds explains its appeal to the of gas chromatography the reader is referred
pesticide residue analyst. to Selected Reference U-l.
Until very recently the most common radio- Gas chromatographic methods described
active source was tritium adsorbed on in this book are Methods 2-13, 11-9A,
titanium. This source always had the dis- 13-17, 16-11, and 16-28.
advantage that as the temperature of the There is a rather large number of manu-
system rose above 200°, tritium began to facturers of gas chromatographs in the United
evaporate. To avoid this restrietion on States, inc1uding:
usable temperature and its associated hazard,
sources composed of nickel 63 have come on Barber-Colman Co., Industrial Instruments
the market,u Div., 1300 Rock St., Rockford, 111.;
Because of its high sensitivity and relative Beckman Instruments, luc., 2500 Harbor
specificity for halogen compounds, the Boulevard, Fullerton Calif.;
electron capture detector is nearly ideal for F & M Scientific Division of Hewlett-
the determination of residues of chlorinated Packard Co., P.O. Box 245, Avondale, Pa.;
organic pestieides, but because it is not very larrell-Ash Co.;
Instruments for Qualitative and Quantitative Analysis of Mixtures [465

Packard Instrument Co., Inc. 2200 sodium D light, are called saccharimeters.
Warrenville Road, Donners Grove, Ill.; While these instruments may be employed
Perkin-Elmer Corp.; and in the determination of other optically active
Varian Aerograph, 2700 Mitchell Drive, compounds than carbohydrates, they find
Walnut Creek, Calif. little use in food analysis other than for the
Manufacturers of gas chromatographie estimation of sugars, as in Method 16-6,
supplies inc1ude: "Determination" (A). Among the few ex-
Analabs, Inc., 80 Republic Drive, North ceptions are AOAC Method 20.062 for
Haven, Conn.; laevo-malic acid (which depends on enhance-
Applied Science Laboratories, State ment of rotation of the acid by uranium
College, Pa.; and salts) and Method 9-14 for lemon and
Pierce Chemical Co., P.O. Box 117, orange oils in extracts.
Rockford, 111. A discussion of the principles of optical
rotation may be found in Text Reference 13.
American manufacturers of better-grade
Neutron Activation Analysis
polarimeters are:
Apparatus
Neutron activation analysis depends in Cary Instruments, 2724 South Peck Road,
principle on the use of a neutron source to Monrovia, Calif.
convert a nonradioactive isotope to a O. C. Rudolph Sons, Inc., P.O. Box 446,
radioactive isotope, followed by determina- Caldwell, N.J.; and
tion of the concentration of that element by Rudolph Instruments Engineering Co.,
measurements of its radioactivity. An example Inc., 61 Stevens Ave., Little Falls, N.J.
is the determination of nitrogen (and there-
fore protein) by the N 14(n, 2n) N 13 reaction.
Polarographs: Amperometric
This method of analysis is often very
Titrations
sensitive; it can be both rapid and specific,
and requires little previous sam pIe prepa- The technique of polarography and the
ration. The chief obstac1e to its more frequent name itself were originated by Jaroslav
use has been the initial cost of the apparatus Heyrovsky at the Charles University in
and the shielding for the neutron source. Prague about 1929. 14 Polarography is a
Manufacturers supplying apparatus of this process of recording the variations in current
type include: that take place as a constantly increasing
voltage is applied to a system composed of
General Dynamics Corp., General Atomic a dropping mercury electrode and a reference
Division, P.O. Box 608, San Diego, Calif.; electrode immersed in a solution of the
Kaman Nuc1ear, 1700 Garden of the Gods sampIe being analyzed. The mercury electrode
Road, Colorado Springs, Colo. consists of a drop of the metal hanging for a
few seconds at the surface of a glass capillary
from which mercury regularly drops out.
Polarimeters Figure 21-2 is a schematic diagram of the
The polarimeter is of course an instrument polarograph; Figure 21-3 shows the type
for measuring the optical rotation of a solu- of curve obtained by plotting current vs
tion. Polarimeters calibrated to read directly voltage.
in terms of per cent sucrose when the rotation These polarograph curves are completely
of a solution of 26 grams of sam pIe in 100 ml reproducible, and if experimental conditions
is measured at 20° in a 200 mm tube, using are kept constant they depend only on the
466] Special Instrumental Methods

composition of the electrolyzed solution. In followed by a region of potentials in which


the presence of substances which undergo the current has a limiting value. The S-shaped
reduction or oxidation at the surface of the portion of the current-voltage curve is called
dropping electrode, or substances that a polarographic wave. The shapes and
catalytically affect the electrode process, positions of these waves provide information
or those that form stable compounds with on both the quantitative and qualitative
mercury, an increase in cathodic or anodic composition of the electrolyzed solution. The
current is observed. The current rises in a difference between the limiting current and
given potential range, and this increase is the current before the wave rise, which is

Current
meosuring
device

Potentiometer
slide
Mercury contoct
reservoir Storage
bottery

Potentiometer

Nitrogen
inlet

Electrolysis
vessel Reference electrode

Fig. 21-2. Schematic Diagram of Polarograph.


Instruments for Qualitative and Quantitative Analysis of Mixtures [467

an instrument in which the movements of a


galvanometer lamp were recorded photo-
graphieally. This was succeeded by various
instruments that plotted curves on achart
through an electrically-controlled pen. De-
rivative polarography has also been employed.
Limiting A method of analysis related to polaro-
...e: current
graphy is amperometry. Without going into
~::J
U
detail, in amperometric titrations the current
whieh passes through the titration cell
between an indicator electrode and a de-
polarized reference electrode, at a suitable
applied e.mJ., is measured as a function of
the volume of the titrating solution. The end
Voltoge, E - point of an amperometrie titration is the
Fig. 21-3. Current-Voltage Relation in Polaro- point of intersection of two lines giving the
graphy. change in current before and after the
equivalence pointP Amperometry is still
called the wave height, depends on the not a common technique, but it did find a
concentration of the electroactive substance certain use in testing for chlorinated organic
responsible for the wave, while the half-wave pesticide residues in the days before gas
potential, i.e., the potential at that point on chromatographie analyses became common:
the curve at which the current reaches half Extracts of the sam pIes were burned, and
its limiting value, is independent of the chloride was determined in solutions of the
concentration but is a function of the nature ash by amperometrie titration with silver
of the electroactive substance and conse- nitrate. t8
quently can be used to identify it quali-
tativelyY Manufacturers of polarographic apparatus
For a more detailed discussion of the inc1ude:
principle and practice of polarography, the Fisher Scientific Co., 711 Forbes Ave.,
reader is referred to one of the special texts Pittsburg, Pa.;
such as Selected Reference 1-2. This tech- Leeds & Northrup Co., 4901 Stenton Ave.,
nique has been used chiefly to analyze Philadelphia, Pa.;
sampies for different metallic elements that Melabs Scientific Instruments Operation,
can be made to yield specific polarographic 3300 Hillview Ave., Stamford Industrial
waves under the proper conditions; the Park, Palo Alto, Calif.;
titrations are carried out in a stream of E. H. Sargent & Co., 4647 West Foster Ave.,
nitrogen to eliminate the oxidation-reduction Chicago, IIl.
effects of atmospheric oxygen. The method is
however also used to determine organic
Raman Spectroscopes
compounds that yield half-wave potentials.
An example of the use of polarography in The phenomenon of the Raman shift
food analysis is the method of Christian interference effect was discovered inde-
et al. for traces of selenium. 16 pendently in 1928 by Raman and Krishnan
Originally polarographie curves were in India and Landsberg and Madelstam in
plotted manually, but because this was Russia. 19 When a beam of monochromatic
laborious, Heyrovsky and Shikata introduced light is passed through a tube containing a
468] Special Instrumental Methods

liquid sam pie, and some of the light that is vantages are: 1) analyses are practically
scattered in a direction perpendicular to the limited to c1ear, colorless (non-absorbing)
incident beam is passed into a spectrograph, liquids; 2) the sam pie must be completely
the resulting spectrogram shows (in addition non-fluorescent; and 3) the sam pie volume
to the expected strong line at a position required is larger than for infrared spectro-
corresponding to the wavelength of the scopy (5-15 ml is normally used). The
incident light) a pattern of relatively weak geatest disadvantage, however, has been the
lines or narrow bands on the red (long- difficulty in obtaining a light source that was
wavelength) side of the incident line, and sufficiently intense and monochromatic to
possibly a very few weak lines on the blue yield Raman lines that were bright and
side as weIl. These lines are the Raman sharp enough. Until recently almost all
spectrum of the sampie ; the lines on the instruments utilized the Hg4358A line, but it
long-wavelength side are called the Stokes is now generally expected that the illumina-
Raman fines, and those on the short- tion difficulty will be solved by the use of
wavelength side, the anti-Stokes fines. laser beams; in fact one instrument with a
If the wavelength of the incident light is laser source, the Cary, is already on the
changed to larger or sm aller values, the market.
pattern of Raman lines (i.e., their relative Manufacturers of Raman apparatus
positions to each other) is unaltered, but the inc1ude:
group as a whole is shifted to longer or
shorter wavelengths. The difference in fre- Cary Instruments;
quency between aRaman line and the incident Perkin-Elmer Corp.;
light is independent of the frequency of the Spex Industries, Inc., 3880 Park Ave.,
incident light, and the set of frequency Edison, N.J.
differences or shifts, which is the Raman
spectrum, is characteristic of the molecules
X-Ray Fluorescence
in the sampie. These shifts are commonly
Spectrographs
reported as wave-numbers.
Because the Raman spectrum of a com- There is a difference of opinion over the
pound is unique for that compound, and the name for the type of analysis to which we shall
spectrum of a mixture whose components refer here. Some authorities prefer the name
do not interact is a superposition of the X-ray emission spectrography on the grounds
Raman spectra of the components, Raman that this designation prevents confusion with
spectroscopy may be used for the identifi- ordinary fluorometry as discussed above, and
cation of both single compounds and "has a familiar ring to analytical chemists,
mixtures of as many as six different accustomed as they are to emission spectro-
substances. Since the intensity of each line graphy involving radiant energy in the visible
is proportional to the concentration of the and in the ultraviolet."20 Nevertheless, while
substance giving that line, quantitative such terminology may appear to prevent
analysis is also possible. confusion on the part of technicians knowing
The advantages of Raman spectroscopy little beyond, and interested only in, the
inc1ude: 1) the Raman spectrum of a com- practical application of instruments as
pound is relatively simple; 2) water may analytical tools, it (perhaps deliberately) has
be used as a solvent because the Raman the effect of completely confusing the
spectrum of water is diffuse and weak; and distinction between emission and fluorescence
3) the relationship between line intensity phenomena. We have therefore headed this
and concentration is linear. Some disad- section "X-Ray Fluorescence Spectrographs."
Instruments for Qualitative and Quantitative Analysis of Mixtures [469

When a be am of electrons impinges on a a bent mica crystal) that serves as a diffraction


metal target, X-rays characteristic of the grating. The diffracted fluorescent X-rays are
particular metal are expelled. This is X-ray detected by a Geiger counter attached to a
emission. To use it for analytical purposes goniometer, and their intensities are auto-
would require, however, construction either matically recorded as functions of the
of separate X-ray tubes for each analysis, or goniometer angles, which are proportional
of a single tube so designed that targets could to the wavelengths or frequencies of the
be changed. Because ofthe physical difficulties X-rays. Figure 21-4, which was obtained from
in this method of analysis, it has had no a chrome-nickel plating on a silver-copper
practical use, which probably explains in base, i1I ustrates the type of spectrum 0 bserved.
part why some authors in their description of Each element is identified by its position on
X-ray spectroscopy leave uncIear the dis- the abscissa, while the height of the ordinate
tinction between emission and fluoroscopic peak is proportional to the quantity present
effects. 21 in the sample. 21
The X-ray fluorescence spectrometer, on X-ray fluorescence analysis has an ad-
the other hand, is a practical device that has vantage over emission spectrography in that
extensive use and may be the instrument of X-ray spectra never consist of more than four
choice for so me analyses. lt works in principle lines, while the transition elements yield
as folIows: an intense beam from an X-ray many atomic emission lines. In trace metal
tube impinges on the sampie being analyzed, analysis particularly this plethora of lines can
producing fluorescent X-rays by secondary make it difficult to separate a line of the
emission. These rays are dispersed by a element sought from the lines of a more
crystal of known lattice constant (frequently abundant element such as iron. Another

...o
GI
0
~
Co

Z Ci
~

E
:I
'E
e
~
u
....GI0
111
~

E
:I

.
GI
. ~
u
.
'E
0

>
GI
Co
Co
o
u
...
I
30°
Angle in degrees

Fig. 21-4. X-ray fluorescence spectrum of chrome-nickel plating on silver~opper brass. [From O'Connor,
R. T.: J. Assoc. Ojfic. Anal. Chemists, 51 :244 (1968).]
470] Special Instrumental Methods

advantage is that X-ray fluorescence analysis include:


is not restricted to the determination of
metallic elements, as emission spectrography Baird-Automic; General Electric Co., X-
is in practice. Its disadvantages are that it is Ray Dept., 4855 Electric Ave., Milwaukee,
not capable of extreme sensitivity, and Wis.; Instrumentation Division, Engis Equip-
cannot be used to determine elements lighter me nt Co., 8035 Austin Ave., Morton Grove,
than sodium. 21 Tuchscheerer has applied Ill. and
X-ray f1uorescence analysis to the de- Norelco Philips Electronics Instrument
termination of K, P, Cu, Fe, Zn and Mn in Div., Philips Electronics and Pharmaceutical
foods. 22 Industries Corp., 750 South Fulton Ave.,
Manufacturers of this type of apparatus Mt. Vernon, N.Y.

Instruments Primarily for Identification of


Single Compounds

violet spectra. For our purposes, it is sufficient


Infrared Spectrophotometers
to note that the infrared curve of a compound
Despite the fact that they are expensive, is a fingerprint that uniquely identifies it,

°
infrared spectrophotometers are rapidlyon and further, that a specific group within a
their way to becoming part of the normal molecule, e.g., CH 3 , C = 0, P = or C = C,
equipment of all modern analytical labora- absorbs at nearly the same wavelength
tories. These instruments plot the absorption regardless of the compound in which it is
curves of sam pies in the infrared in essentially found. Therefore, if an analyst is confronted
the same manner that instruments such as with a compound of whose structure he is
the Beckman DU and the Cary do in the uncertain, he can run its infrared spectrum
visible and ultraviolet regions ofthe spectrum. and compare it with the spectrum of a
The source of light is different, being a reference substance he suspects it to be; ifthe
Globar (a bonded Sie rod) or a Nernst two prove to be the same, identification is
glower (a mixture of Zn, Th and Yoxides); certain. Ifthe curve for the unknown does not
the dispersing element of choice is rapidly match that ofthe postulated compound, it can
becoming a grating, but if it is a prism, it still be identified if a match can be found with
must be made of a salt such as NaCI or KBr any ofthe thousands ofspectra recorded in the
rather than of quartz; and the detector is literat ure. And if even this search fails, the
usually a thermocouple. These differences in unknown will usually show some charac-
the components of the infrared instrument teristic group absorption peaks that will help
from those of the visible-ultraviolet apparatus to point the way toward elucidation of its
are necessary because the lamps used as structure. It should be noted, however, that
sources of visible and ultraviolet light are infrared spectrometry does not distinguish
poor sources of infrared, and glass and even between optical isomers.
quartz are essentially opaque to the infra- Infrared spectrophotometry can also be
red. used for the quantitative determination of a
We shall not enter into a theoretical single compound or a mixture of several
discussion of why the infrared absorption compounds when each of the substances
curve of a substance teIls so much more about being determined possesses a sharp peak that
that substance than do the visible and ultra- is not shared with any other substance that is

Das könnte Ihnen auch gefallen