Sie sind auf Seite 1von 5

Basic Research—Technology

Effects of Electropolishing Surface Treatment on the Cyclic


Fatigue Resistance of BioRace Nickel-Titanium Rotary
Instruments
Hélio P. Lopes, LD,* Carlos N. Elias, PhD,† Victor T.L. Vieira, DDS,‡ Edson J.L. Moreira, PhD,‡
Raquel V.L. Marques, DDS,* Julio C. Machado de Oliveira, PhD,* Gilberto Debelian, PhD,§
and José F. Siqueira, Jr., PhD*

Abstract
Introduction: This study evaluated the influence of
electropolishing surface treatment on the number of
cycles to fracture of BioRace rotary nickel-titanium
C yclic fatigue fracture occurs when an endodontic instrument rotates in a curved
canal. In this condition, the instrument under elastic deformation is subjected to
a mechanical load represented by alternating tensile and compressive stresses. The
endodontic instruments. Methods: BioRace size BR5C cyclical repetition of the load leads to instrument fracture through low-cycle fatigue
instruments with or without electropolishing surface (1–4). The cyclic fatigue resistance comprises the number of cycles that an
treatment were used in an artificial curved canal under instrument can endure under a specific loading condition until fracture occurs.
rotational speed of 300 rpm until fracture. Fractured Because the number of cycles to fracture is cumulative, it can be calculated by
surfaces and the helical shafts of fractured instruments multiplying the rotation speed by the time elapsed until fatigue fracture occurs
were analyzed by scanning electron microscopy (SEM). (4–6).
Results: Polished instruments displayed a significantly Several factors can influence the cyclic fatigue resistance of an endodontic instru-
higher number of cycles to fracture when compared ment subjected to flexural fatigue (2, 4, 5, 7–10). One of these factors includes defects
with nonpolished instruments (P < .001). Actually, the in surface finishing resulting from the machining process used for manufacturing
number of cycles to fracture of a polished BR5C instru- nickel-titanium (NiTi) endodontic instruments. These defects are mostly grooves that
ment was 124% higher than that of a nonpolished work as stress concentration factors. Once fatigue cracks initiate on the instrument
instrument. SEM analysis showed that the fractured surface, stress concentration factors favor nucleation, growth, and propagation of
surface of both polished and nonpolished BR5C instru- cracks (3, 11).
ments had ductile morphologic characteristics. Evalua- Electropolishing is a method of surface finishing used by manufacturers of rotary
tion of the separated fragments after cyclic fatigue NiTi instruments to remove surface defects that might remain after the machining
testing showed the presence of microcracks near the process. Some previous studies showed that the surface finishing process increases
fracture surface. Polished instruments exhibited fine the fatigue resistance of rotary NiTi instruments (7, 10, 12–14). However, other
cracks that assumed an irregular path (zigzag crack studies reported that surface finishing defects do not influence the fatigue life of
pattern), whereas nonpolished instruments showed rotary NiTi instruments (15–18). The purpose of the present study was to analyze
cracks running along the machining grooves. Conclu- the influence of electropolishing surface treatment on the number of cycles to
sions: Electropolishing surface treatment of BioRace fracture of BioRace (FKG Dentaire, La Chaux-de-Fonds, Switzerland) rotary NiTi
endodontic instruments significantly increased the cyclic endodontic instruments.
fatigue resistance. (J Endod 2010;36:1653–1657)
Materials and Methods
Key Words
Instrument Dimensions
Chemomechanical preparation, cyclic fatigue, electropo-
lishing surface treatment, endodontic instruments
BioRace rotary NiTi endodontic instruments size BR5C, which have a nominal size
at D0 of 0.40 mm, a taper of 0.02 mm/mm, and total length of 25 mm, were used in this
experiment. Ten instruments were provided by the manufacturer with an electropol-
ished working part, whereas the other 10 instruments presented no surface treatment.
Specimens were from the same batch, with nonpolished instruments being removed
from the production line before the surface treatment approach.

From the *Department of Endodontics, Faculty of Dentistry, Estácio de Sá University, Rio de Janeiro, RJ, Brazil; †Department of Materials Science, Military Institute of
Engineering, Rio de Janeiro, RJ, Brazil; and ‡Department of Endodontics, Faculty of Dentistry, Grande Rio University, Rio de Janeiro, RJ, Brazil; and §Department of
Endodontics, University of Oslo, Oslo, Norway.
Address requests for reprints to José F. Siqueira Jr, DDS, MSc, PhD, Faculty of Dentistry, Estácio de Sá University, Av. Alfredo Baltazar da Silveira, 580/cobertura,
Recreio, Rio de Janeiro, RJ, Brazil 22790-710. E-mail address: jf_siqueira@yahoo.com.
0099-2399/$0 - see front matter
Copyright ª 2010 American Association of Endodontists.
doi:10.1016/j.joen.2010.06.026

JOE — Volume 36, Number 10, October 2010 Electropolishing Surface Treatment and Fatigue 1653
Basic Research—Technology

Figure 1. Schematic drawing of the artificial canal used in the cyclic fatigue
experiment.

For standardization of the instruments used, their diameters at D3


and D13 as well as their total lengths were determined by means of
a stereomicroscope (Pantec-Panambra, Cambuci, SP, Brazil). The taper
of the working part was obtained on the basis of D3 and D13 measure-
ments according to Stenman and Spangberg (19), following the equa-
tion: Taper = D13 – D3/10. The diameter at D0 was then calculated on
the basis of the values of D3 and taper (T) by using the following equa-
tion: D0 = D3 – T  3.

Surface Finishing
Surface finishing of the instrument’s working parts was assessed
under scanning electron microscopy (SEM) (Jeol JSM 5800, Tokyo,
Japan). Instruments were cleaned in acetone under ultrasonication,
and then photomicrographs were taken at magnifications of 150
and 500 for further analysis.
Figure 2. Nonpolished BR5C instrument showing machining grooves. (A)
Bending Resistance Tip (original magnification, 150); (B) helical shaft (original magnification,
The bending resistance of the test instruments was evaluated by the 500).
cantilever-bending test, as described by Serene et al (20). Briefly,
a universal testing machine (DL 10.000; Emic, São José dos Pinhais, A stainless steel apparatus was constructed with a square base and
PR, Brazil) was used. Load was applied by means of a flexible stainless a vertical axis. The vertical axis contained a structure that allowed for the
steel wire (with length of 50 cm and diameter of 0.34 mm), with one of fixture and movement of the micromotor/contra-angle. At the apparatus
the extremities fastened to the testing machine head and the other end 3 base, a bench vise was used to hold the stainless steel tube. A gap at the
mm away from the instrument tip. The bending test was conducted until base of the apparatus allowed for the movement of the bench vise in
the tip of each specimen underwent an elastic displacement of 45 a horizontal direction, allowing for a coincidence between the axis of
degrees. The test speed was 15 mm/min, and the load cell used was the instrument and the straight part of the stainless steel canal.
20 N. Data were statistically analyzed by using the Student t test, with During the test, the artificial canal was filled with glycerin to reduce
the significance level set at 5%. the friction of the instrument against the canal wall and to minimize the
release of heat. Each instrument was placed in a contra-angle at speed
Cyclic Fatigue reduction of 10:1 (TC–Motor 3000; Nouvag AG/AS/LTD, Goldach,
An artificial canal was made out of a cylindrical tube of stainless Switzerland) and introduced into the canal until the tip touched a shield
steel with inner diameter of 1.4 mm, total length of 19 mm, arc located positioned at the end of the canal (3-mm-long straight part). This shield
between the two straight segments of the canal, and curvature radius of was thereafter removed because its main purpose was to standardize the
6 mm. The arc measured 9 mm, the longest straight part 7 mm, and the distance the instrument penetrated in the artificial canal.
shortest straight part 3 mm. The curvature radius of the artificial canal Subsequently, BR5C instruments of each group (polished and
was measured, taking into account the concave surface of the interior of nonpolished) were worked in clockwise rotation at nominal speed of
the tube (Fig. 1). 300 rpm until fracture. The time to fracture was measured by the
same operator by using a digital stopwatch (Leroy) and was established
when there was visual observation of the instrument separation. The
TABLE 1. Means of Length, Diameter at D0, and Taper for the BR5C number of cycles was obtained by multiplying the rotational speed by
Instruments (in millimeters) Used in this Study the time (in seconds) until the fracture of each instrument occurred.
Data obtained on the number of cycles to fracture of the BioRace
Instrument Length D0 Taper
BR5C polished and nonpolished instruments were statistically analyzed
Polished BR5C 25.3 0.39 0.018 by the Student t test, with significance level set at 5%. The fractured
Nonpolished BR5C 25.1 0.41 0.018
surface and the helical shaft of fractured instruments were analyzed

1654 Lopes et al. JOE — Volume 36, Number 10, October 2010
Basic Research—Technology
TABLE 3. Mean ( standard deviation) of the Time and the Number of Cycles
to Fatigue Fracture (NCF) of Polished and Nonpolished BR5C Instruments
Instrument Time (sec) NCF
Polished BR5C 86.8 (9.4) 434 (47.0)
Nonpolished BR5C 38.8 (6.7) 194 (33.5)

in Table 2. No significant difference was observed between them (P


= .67).

Cyclic Fatigue
The average and standard deviation of the time and number of
cycles until the instrument fracture occurred are shown in Table 3. Pol-
ished instruments displayed a significantly higher number of cycles to
fracture when compared with nonpolished instruments (P < .001).
SEM analysis showed that the fractured surface of both polished
and nonpolished BR5C instruments had ductile morphologic charac-
teristics. No plastic deformation occurred in the helical shaft of the
instruments tested (Fig. 4). Evaluation of the separated fragment after
fatigue test showed the presence of microcracks near the fracture
surface. Fine cracks that assumed an irregular path (zigzag crack
pattern) were noted in polished instruments (Fig. 5), whereas nonpol-
ished instruments showed cracks running along the machining
grooves (Fig. 6).

Figure 3. Polished BR5C instrument exhibiting a smooth surface with limited


presence of manufacturing marks. (A) Tip (original magnification, 150);
(B) helical shaft (original magnification, 500).

by means of SEM (JEOL JSM 5800) to determine the type of fracture,


presence of fatigue cracks, and plastic deformation in the helical shaft.

Results
Instrument Dimensions
The average length, diameter at D0, and taper of the BR5C instru-
ments are shown in Table 1.

Surface Finishing
SEM analysis revealed that the surfaces of the working part of all
nonpolished instruments exhibited prominent transversal machining
marks (Fig. 2). Polished instruments in turn showed smooth surfaces
with limited presence of manufacturing marks (Fig. 3).

Bending Resistance
The average bending resistance as measured by the maximum
force (gf) to bend polished and nonpolished instruments is depicted

TABLE 2. Mean ( standard deviation) of Maximum Force (gf) to Bend


Polished and Nonpolished BR5C Instruments
Instrument Maximum force Figure 4. Fractured instruments. (A) Fracture surface of instruments
showing morphology characteristics of the ductile type (original magnification,
Polished BR5C 117.2 (8.1)
1500); (B) absence of plastic deformation is observed on the helical shaft
Nonpolished BR5C 118.6 (7.2)
(original magnification, 100).

JOE — Volume 36, Number 10, October 2010 Electropolishing Surface Treatment and Fatigue 1655
Basic Research—Technology

Figure 6. Nonpolished BR5C instrument. Cracks following machining grooves


are observed near the fracture surface. (A) Original magnification, 1000; (B)
Figure 5. Polished BR5C instrument. Cracks in a zigzag pattern are seen near original magnification, 1500; (C) original magnification, 2500.
the fracture surface. (A) Original magnification, 1000; (B) original magni-
fication, 1200; (C) original magnification, 1500.

for diameter and taper were within the tolerance limits permitted (ISO
Discussion 3630-1, 1992). These evaluations were done to guarantee that all tested
In this study, only 1 type of instrument was selected for testing to instruments had similar dimensions in spite of being of the same size as
eliminate other variables that could possibly interfere with the results. reported by the manufacturer. In this regard it is interesting to point out
Therefore, the only variable was the one under study, ie, instruments that the instruments to be polished are usually manufactured a little
with or without electropolishing surface treatment. All measurements overdimensioned, so that after the electropolishing process they fall
of instruments performed in this study revealed that the data obtained within the ISO tolerance limits.

1656 Lopes et al. JOE — Volume 36, Number 10, October 2010
Basic Research—Technology
As to these measurements, our results are in disagreement with In conclusion, the electropolishing surface treatment of BioRace
Bui et al (16), who stated that because electropolishing removes layers endodontic instruments significantly increased the number of cycles
of the surface of NiTi instruments, there are concerns that it reduces the to fracture under rotating-bending conditions within an artificial curved
dimensions of rotary files. However, reduction in the instrument dimen- canal.
sions caused by polishing appeared to be almost intangible, and the
variations observed in this study were within the tolerance limits for Acknowledgments
endodontic instruments (ISO 3630-1, 1992).
SEM analysis of nonpolished BR5C instruments before the fatigue The authors thank FKG Dentaire for providing the instruments
test revealed machining grooves, which ran almost perpendicularly to used in this study, including the nonpolished instruments that are
the long axis of the instrument. Polished instruments in turn showed not commercially available. This study was supported in part by
reduced surface irregularities, with remnants of the transverse- grants from Conselho Nacional de Desenvolvimento Cientı´fico e
running machining grooves vaguely discernible. These findings are in Tecnolo´gico (CNPq) and Fundaça˜o Carlos Chagas Filho de Amparo
consonance with other studies (7, 10, 12, 21). a` Pesquisa do Estado do Rio de Janeiro (FAPERJ), Brazilian Govern-
Bending resistance was tested by using the nondestructive mental Institutions.
cantilever-bending assay. The proposed displacement of the instrument
tip was 15 , so that the deformation induced by the applied load re- References
mained within the elastic limit of the NiTi alloy. Our findings showed 1. Zelada G, Varela P, Martin B, Bahillo JG, Magan F, Ahn S. The effect of rotational
no significant difference between polished and nonpolished instru- speed and the curvature of root canals on the breakage of rotary endodontic instru-
ments with regard to bending resistance. These findings conferred ments. J Endod 2002;28:540–2.
2. Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel-titanium
sample homogeneity to the next assay (cyclic fatigue), because the endodontic instruments. J Endod 1997;23:77–85.
instrument bending resistance can directly interfere with its perfor- 3. Callister WD Jr. Ciência e engenharia de materiais: uma introdução. 5a ed. Rio de
mance in the cyclic fatigue test (3). Janeiro: LTC; 2002.
The present findings demonstrated that electropolishing surface 4. Lopes HP, Ferreira AA, Elias CN, Moreira EJ, de Oliveira JC, Siqueira JF Jr. Influence
treatment significantly increased the instrument resistance to cyclic of rotational speed on the cyclic fatigue of rotary nickel-titanium endodontic instru-
ments. J Endod 2009;35:1013–6.
fatigue. Consequently, it can be inferred that surface finishing defects 5. Lopes HP, Moreira EJ, Elias CN, de Almeida RA, Neves MS. Cyclic fatigue of ProTaper
can influence the number of cycles to fracture. The number of cycles instruments. J Endod 2007;33:55–7.
to fracture of a polished BR5C instrument was 124% higher than that 6. Yao JH, Schwartz SA, Beeson TJ. Cyclic fatigue of three types of rotary nickel-titanium
of a nonpolished instrument. Overall, these findings are in line with files in a dynamic model. J Endod 2006;32:55–7.
7. Boessler C, Paque F, Peters OA. The effect of electropolishing on torque and force
several previous studies (7, 10, 12, 14, 21). However, they are in during simulated root canal preparation with ProTaper shaping files. J Endod 2009;
disagreement with others (15–18). 35:102–6.
Electropolishing surface treatment might have beneficial effects by 8. Yared GM, Bou Dagher FE, Machtou P. Cyclic fatigue of ProFile rotary instruments
prolonging fatigue life of rotary NiTi instruments (10). The benefits of after clinical use. Int Endod J 2000;33:204–7.
electropolishing are likely to be caused by reduction in surface irregu- 9. Martin B, Zelada G, Varela P, et al. Factors influencing the fracture of nickel-titanium
rotary instruments. Int Endod J 2003;36:262–6.
larities that serve as points for stress concentration and crack initiation. 10. Anderson ME, Price JW, Parashos P. Fracture resistance of electropolished rotary
On the other hand, one study (18) reported that electrochemical polish- nickel-titanium endodontic instruments. J Endod 2007;33:1212–6.
ing did not increase the mechanical resistance of NiTi K3 endodontic 11. Griffith AA. The phenomena of rupture and flow in solids. The Royal Society 1920;
files. The electrochemical treatment affected neither resistance to flex- 221:A163.
12. Kuhn G, Tavernier B, Jordan L. Influence of structure on nickel-titanium endodontic
ural fatigue nor torsional resistance of K3 instruments. According to their instruments failure. J Endod 2001;27:516–20.
SEM analysis, the surface of K3 files was not affected by electrochemical 13. Parashos P, Messer HH. Rotary NiTi instrument fracture and its consequences.
polishing. However, it is important to point out that since 1920 it has J Endod 2006;32:1031–43.
been widely recognized that surface finishing defects act as points of 14. Tripi TR, Bonaccorso A, Condorelli GG. Cyclic fatigue of different nickel-titanium
stress concentration, reducing the number of cycles to fracture (11). endodontic rotary instruments. Oral Surg Oral Med Oral Pathol Oral Radiol Endod
2006;102:e106–14.
According to Callister (3), instruments with finishing surface 15. Cheung GS, Shen Y, Darvell BW. Does electropolishing improve the low-cycle fatigue
defects present tensile resistance significantly lower than the theoretical behavior of a nickel-titanium rotary instrument in hypochlorite? J Endod 2007;33:
fracture resistance of the material. This behavior is related to the fact 1217–21.
that defects act as stress concentration factors and are sites highly 16. Bui TB, Mitchell JC, Baumgartner JC. Effect of electropolishing ProFile nickel-
titanium rotary instruments on cyclic fatigue resistance, torsional resistance, and
conducive to the formation and propagation of cracks. When no finish- cutting efficiency. J Endod 2008;34:190–3.
ing surface defects are present, the theoretical fracture resistance of the 17. Herold KS, Johnson BR, Wenckus CS. A scanning electron microscopy evaluation of
material is equal to the tensile stress of the material determined by microfractures, deformation and separation in EndoSequence and Profile
mechanical assays. nickel-titanium rotary files using an extracted molar tooth model. J Endod 2007;
SEM analysis showed that the fractured surface of both polished 33:712–4.
18. Barbosa FO, Gomes JA, de Araujo MC. Influence of electrochemical polishing on the
and nonpolished BR5C instruments had ductile morphologic character- mechanical properties of K3 nickel-titanium rotary instruments. J Endod 2008;34:
istics. Evaluation of the separated fragment after cyclic fatigue test re- 1533–6.
vealed microcracks near the fracture surface. Surface cracks 19. Stenman E, Spangberg LSW. Root canal instruments are poorly standardized.
followed machining grooves in nonpolished instruments, whereas pol- J Endod 1993;17:327–34.
20. Serene TP, Adams JD, Saxena A. Nickel-titanium instruments applications in
ished instruments exhibited cracks in a zigzag pattern. As for the latter, Endodontics. St Louis: Ishiyaku Euroamerica Inc; 1995.
its irregular path is typical of intergranular fracture, where propagation 21. Kim HC, Yum J, Hur B, Cheung GS. Cyclic fatigue and fracture characteristics of
of cracks occurs along the grain boundaries. ground and twisted nickel-titanium rotary files. J Endod 2010;36:147–52.

JOE — Volume 36, Number 10, October 2010 Electropolishing Surface Treatment and Fatigue 1657

Das könnte Ihnen auch gefallen