Process validation in

medical devices
Fulfil requirements with expert
regulatory guidance

TÜV SÜD
1
Contents
INTRODUCTION 4

VALIDATION PLANNING 5

INSTALLATION QUALIFICATION 7

OPERATIONAL QUALIFICATION 9

PERFORMANCE QUALIFICATION 11

STATISTICAL SAMPLING PLANS 13

Guideline Contributors
Michael Schaefer, Co-Author
Quality Management and Regulatory Affairs in Medical Devices
michael@quality-on-site.com
www.quality-on-site.com

2 Process validation in medical devices | TÜV SÜD


Medical device technology
leadership from TÜV SÜD

End-to-end testing, certification and auditing solutions


With the highest number of medical device experts and the widest service offering of any notified body, TÜV SÜD
provides market access solutions and expert partnership for medical device manufacturers and suppliers worldwide.
Operating throughout a global network of world-class laboratories, our dedicated team of close to 600 medical experts,
engineers and doctors can support your business by providing assessments that cover the full medical device lifecycle.

Medical device lifecycle


Pre-Acceptance Testing &
Design & Production
Technical Tests Certification /
Development Monitoring
/ Clinical Tests Global Approvals

TÜV SÜD provides solutions for these stages of the medical device lifecycle.

Assuring safe and effective medical devices


Our team is ready to assess your innovative design with a suite of premium services that keep you in line with
regulations and ahead of your competition. We offer services for all categories of medical devices including:

Active medical devices


 Rehabilitative medical devices
 Medical device software and

Non-active medical devices
 In-vitro diagnostics
 applications
Implantable medical devices
 High-risk devices
 Innovative technologies

Our services

Testing & Product Auditing & System Market Approval Clinical Services
Certification Certification

Why choose TÜV SÜD?


Product safety Global expert Value-added High-risk device
assurance coverage services expertise
Independent and impartial Close to 600 medical Functional safety, Unique and specialised
certification that experts providing pre-assessments and services in high-risk
establishes confidence professional expertise with other services for early devices including clinical
among stakeholders and personal responsiveness. identification of issues to path analyses and
regulatory authorities. save time and money. expedited services.

TÜV SÜD | Process validation in medical devices 3


Introduction

Guideline on process validation activities


International regulations for medical When carrying out process
Standards and
devices [1, 2, 3, 4] stipulate validation validation activities, additional
regulations
of manufacturing processes for points to consider include
medical devices, where the resulting statistical sampling plans, process
This guideline is based on
process output cannot be or is not monitoring, validation change
the following standards and
verified by subsequent monitoring control and good documentation
regulations:
or measurement. This entails practice. Manufacturers of medical
[1] ISO 13485:2016
activities that are not only limited to devices should also note that the
[2] 21CFR820

the validation of the manufacturing requirements on process validations
[3] GHTF/SG3/N99-10:2004
process itself but may also include are not limited to their own premises
(Ed.2)
equipment qualification, computer but also apply to processes
[4] ZLG 3.9 B18
system validation and test method conducted offsite by suppliers. This
[5] F inal Guidance for
validation. guideline supports manufacturers
Industry and FDA Staff,
of medical devices in fulfilling these
General Principles of
regulatory requirements.
Software Validation

4 Process validation in medical devices | TÜV SÜD


Validation planning

Guarantee a predictable outcome


The Global Harmonization Task Force (GHTF) [3] defines process validation as ‘a term used in the medical device
industry to indicate that a process has been subject to such scrutiny that the result of the process … can be practically
guaranteed’. Prior to undertaking any validation activities, thorough planning is essential. Validation will include
Installation Qualification (IQ) for equipment and machines, as well as Operational Qualification (OQ) and Performance
Qualification (PQ) for the manufacturing process. It must be emphasised that additional process- and/or product-
specific requirements are to be considered. This includes but is not limited to sterilisation and sterile packaging
processes where PQ also includes challenging of process tolerances and limits.

Typical elements of process validation


Installation Required for equipment
Qualification (IQ) and machines

Operational Process runs at


Qualification (OQ) worst-case settings Required
if result of
the process
can/is NOT
Performance Process runs at fully verified
Qualification (PQ) normal settings

Documented procedures describing the approach for process validation are a requirement [1, 2]. Beyond listing the
definitions used, this documentation should describe responsibilities and delineate authorities. The different elements
of process validation and the sequence of execution are to be clarified, including requirements for re-validation.
Procedures for process validation should also contain further details regarding testing, training, documentation,
review and approval.

TÜV SÜD | Process validation in medical devices 5


Validation Master Plan
Manufacturers often choose to The elements typically found in a commitment to validation activities.
develop a Validation Master Plan VMP include listings and references They also support the planning and
(VMP) as a tool to control and monitor to the qualifications and validations allocation of human and financial
the status of validation activities. performed internally and externally resources, as requested by
As VMPs are not required by at suppliers, as well as schedules for regulations [1, 2, 4].
regulations, the content and structure re-validation activities. VMPs can be
can vary widely. used to demonstrate management

Individual validation plan


An individual validation plan is another validation plan may establish a (bracketing or grouping of products),
tool that is widely used to support the generic validation strategy and offer combined protocols and reports may
planning for more complex projects the potential to approach validation also be described and justified in an
such as new manufacturing lines or activities in a more structured individual validation plan.
production transfers. The individual manner. Product matrix approaches

6 Process validation in medical devices | TÜV SÜD


Installation Qualification

Ensure reliable operation of equipment


Installation Qualification (IQ) is or microbiological procedures. drawings or codes that are needed
defined by GHTF as ‘establishing Recommendations or instructions to operate, maintain or repair the
by objective evidence that all key from third-party equipment suppliers equipment should be compiled and
aspects of the process equipment are an essential source of information retained for future use.
and ancillary system installation and must be considered during
adhere to the manufacturer’s installation activities. IQ adds value for manufacturers by
approved specification and that the helping to prevent future issues.
recommendations of the supplier Applicable support systems and When done properly, it ensures
of the equipment are suitably utilities must also be verified and that equipment operates in a
considered’ [3]. Also referred to as completed during installation. reliable and predictable manner.
equipment qualification [1], IQ is This extends beyond the supply of Successful completion of the IQ is
regarded as an appropriate method to electricity and compressed air to usually a prerequisite to continue
fulfil regulatory requirements. encompass other topics such as process validation within the
material supply or data transfer Operational Qualification.
All equipment and machines used infrastructure. Functional testing
in the manufacturing process of the equipment, including alarm
of medical devices require IQ. and control systems, must also be IQ adds value for
It supports manufacturers in
properly establishing key areas of
carried out. Furthermore, operating
instructions for the equipment
manufacturers by
operation such as maintenance and training documentation needs helping to prevent
plans, calibration tolerances and to be prepared accordingly. Any future issues.
intervals, as well as cleaning documentation such as manuals,

TÜV SÜD | Process validation in medical devices 7


Computer system validation
Today, virtually all manufacturing these computerised systems and If a computerised system is used in
equipment is controlled by software tools are suitable for connection with the manufacturing
computerised systems such as their intended purpose [1, 2, 5]. IQ/ equipment, computer system
programmable logic controls or OQ/PQ terminology may be used validation must be successfully
computers. Additional software tools for computer system validation, completed before IQ can be
are also used to transfer or evaluate however it is acknowledged that approved.
process data. Validation is needed software professionals may apply
to provide objective evidence that different terms [5].

8 Process validation in medical devices | TÜV SÜD


Operational Qualification

Meet all predetermined requirements at process limits


GHTF defines Operational may choose to incorporate at the established process limits and
Qualification (OQ) as ‘establishing this essential process during adequate process documentation is
by objective evidence process developmental work in separate retained. Once OQ samples are
control limits and action levels documentation packages prior to tested, results are recorded and
which result in product that meets all OQ. Alternatively, DoE can feature evaluated thoroughly, including
predetermined requirements’ [3]. An directly in OQ protocols. Both statistical analyses. In most
OQ is usually accomplished in two pathways are possible as long as the cases, OQ should be successfully
main steps: goal of understanding the process completed as a prerequisite before
1. Identification of critical process and its parameters is achieved starting Performance Qualification.
parameters and demonstrated.
2. Provision of evidence that
production at worst-case settings The second step, providing OQ samples are
will still result in a product that
meets specifications
evidence that a product will meet
pre-determined requirements,
produced at the
necessitates statistically sound established
With regard to the first step, Design sampling plans and valid test process limits.
of Experiments (DoE) is a widely methods. Proper product acceptance
used tool for identifying critical criteria are key to successful
process parameters. Manufacturers validation. OQ samples are produced

TÜV SÜD | Process validation in medical devices 9


Basics of Operational Qualification
Scope of the OQ
 Identification of critical process parameters
 Evidence that product in specification will be produced at worst-case settings

ROBUST DESIGN Products must be


Robust in specification

Output

Sensitive
Min. spec
Sampling plans are needed
to provide evidence for
Input each worst-case situation
GHTF Study Group 3 – Quality Management Systems
Process Validation Guidance – January 2004

10 Process validation in medical devices | TÜV SÜD


Performance Qualification

Manufacture with consistency


Following the GHTF Guidance [3], influences that routine manufacturing be given to the number of different
Performance Qualification (PQ) may be exposed to. pieces of equipment used, variations
can be defined as ‘establishing by in machine set-up, different raw
objective evidence that the process, The PQ plan should clearly describe material or component batches,
under anticipated conditions, how many routine manufacturing working in shifts, environmental
consistently produces a product runs are required to show conditions or degradation of tooling.
that meets all predetermined repeatability and stability of the
requirements’. In order to determine processes under consideration. Successful completion of the PQ will
the scope, range and size of the Additionally, sampling plans need enable the manufacturer to initiate
PQ, it is vital to understand the statistically sound justifications. approval of the new or changed
natural process variability and other For example, consideration should manufacturing process.

Process monitoring
Once the manufacturing process has Automated alarm and control identified in the processes. This
been validated, it is a requirement systems (if qualified) may also be includes but is not limited to recalls,
to maintain the validated state. This used to control validated process complaints, internal and external
entails monitoring and controlling parameters. audit trends, failure investigation,
validated process parameters Corrective and Preventive Action
to ensure that the specified Detecting negative trends or (CAPA), equipment significant
requirements continue to be met [2, unexpected shifts of the process is maintenance or corrective
3]. Quality control charts, also known key, which necessitates thorough maintenance trends.
as Statistical Process Control (SPC), and timely data analysis. In the
are one of the tools commonly used event that these shifts or trends For specific processes such as
for this purpose. are detected, re-validation, sterilisation, re-validation is
investigations and/or corrective scheduled in defined intervals
This involves establishing action actions may be needed. in compliance with applicable
and warning limits then determining Re-validation should also be standards and regulations [1, 2, 4].
rules for atypical process behaviour. considered if deficiencies are

Validation change control


Documented procedures including determining the extent of production line transfers. Simpler
criteria for re-validation and approval re-validation activities needed. events including changes within
of changes to processes or qualified Complete re-validation will be validated limits may require less
equipment must be implemented triggered by significant changes extensive re-validation activities.
[1, 2, 3, 4]. The impact of changes such as the implementation of new
is evaluated and considered when manufacturing technologies or

TÜV SÜD | Process validation in medical devices 11


Good documentation practice
Qualification and validation records are maintained separately from the manufacturing of OQ or PQ samples
must be maintained to demonstrate qualification or validation package. are also documented. In addition, raw
conformance to specified data generated through testing is
requirements and the effective Validation records contain the recorded, including the identification
operation of the qualified equipment names and signatures of the of test persons and the equipment
and validated process. These individual(s) preparing, reviewing used as well as the calibration status.
records must contain appropriate and approving plans and reports.
references when supporting records Process parameters used during

12 Process validation in medical devices | TÜV SÜD


Statistical sampling plans

Select the optimal approach


The application of statistically sound Quality (LQ), as defined in Note that skip-lot sampling plans as
sampling plans is a requirement [1, ISO 3534-2:2006. per ISO 2859-3 cannot be used for
2, 3, 4] and was recently reinforced critical classes of nonconforming
[1]. For harmonised standards and/ AQL defines the level of quality a items or nonconformities.
or applicable regulations, medical sampling plan usually accepts with a
device specific requirements high level of probability. Applicable RQL defines the level of quality a
for statistical approaches may standards for AQL-based sampling sampling plan usually rejects with
need to be selected. In all cases, plans are: a high level of probability. Current
manufacturers are responsible  ISO 2859-1 Sampling inspection by versions of ISO 2859-1, ISO 2859-2
for determining and justifying attributes and ISO 3951-1 include tables for
statistically sound sampling plans.  ANSI/ASQ Z1.4 Sampling consumer’s risk (RQL, LQ) as well.
procedures and tables for
When it comes to determining inspection by attributes Both AQL and RQL can be found
such sampling plans, there are a  ISO 3951-1 Sampling inspection by on the Operating Characteristic
variety of statistical approaches variables Curve (OC-Curve) of the sampling
and methodologies to consider.  ANSI/ASQ Z1.9 Sampling plan. The OC-Curve shows the
Manufacturers often use the procedures and tables for relationship between the probability
principles of the Accepted Quality inspection by variables of a product’s acceptance and the
Limit (AQL) and the Rejected Quality incoming quality level for a given
Limit (RQL), also called Limiting sampling plan.

AQL versus RQL with an example of an OC-Curve

AQL defines the level LTPD (RQL, LQ, UQL)


of quality the sampling defines the level of quality
plan routinely accepts the sampling plan routinely
(producer’s risk). rejects (consumer’s risk).

Probability of 1
acceptance

0.5 OC Curve

0 0.5 1
Defectives [%]

TÜV SÜD | Process validation in medical devices 13


Accepted Quality Limit versus Rejected Quality Limit
The concept of AQL is based on a ISO 16269-6 is another applicable Test methods using attributive
well-known process and is not the standard that is often used for acceptance criteria may be easy
optimal measure for patient safety medical devices. This standard to implement and less expensive.
if the correlating customer risk is introduces the method of statistical However, attributive sampling plans
not known and justified. AQL-based tolerance intervals and is also based result in higher sample sizes than
sampling plans are usually applied in on consumer risk. Therefore, ISO variable sets of data.
routine manufacturing after process 16269-6 is suitable for statistical
reliability is established. sampling plans in process validation. Variable data is a set of continuous
A statistical tolerance interval measurements, ideally normally
For this reason, an RQL-based depends on a confidence level and distributed. By enabling statistical
sampling plan offers a more suitable a stated proportion of the sample evaluations, it delivers greater
approach for process validation, as it group. Once the confidence level and knowledge than attributive data.
more accurately predicts the amount the proportion targeted is known, the Done correctly, variable sampling
of defective products delivered standard will lead to sample sizes for plans result in lower sample sizes
to the customer. both variable and attribute data. than attributive sampling plans.
The advantage of having lower
The differentiation of attributive test sample sizes necessitates statistical
A RQL-based results versus variable test results calculations and evaluations as given

sampling plan offers is an additional consideration when


generating statistical sampling plans.
by the standards. Using variable
sampling plans without doing the
a more suitable Examples for attributive data sets statistical calculation is insufficient.
approach for include criteria such as go/no-go,
process validation. good/bad, leaking/tight, particle/
particle-free.

Statistical sampling plans in summary

AQL-based sampling plans are focussed on producer’s risk. Using


 Technical
these plans during process validation is not recommended unless
the referring RQL is justified.
understanding of
R
QL-based sampling plans are focussed on consumer’s risk. Using
 the manufacturing
these plans during process validation is recommended. Selection processes is key
of RQLs and confidence levels is to be justified.
ISO 16269-6 includes sampling plans for both variable and

to ensure that the
attribute data. process results
‘can be practically
guaranteed’ as
It must be emphasised that the application of statistical sampling plans is stated at the
required. However, these plans will not replace proper process development
or monitoring of critical process parameters. Technical understanding of the
beginning of this
manufacturing processes is key to ensure that the process results ‘can be guideline.
practically guaranteed’ as stated at the beginning of this guideline.

14 Process validation in medical devices | TÜV SÜD


GLOSSARY OF ACRONYMS
AQL – Accepted Quality Limit OC-CURVE – Operating Characteristic Curve
CAPA – Corrective and Preventive Action OQ – Operational Qualification
DoE – Design of Experiments PQ – Performance Qualification
GHTF – Global Harmonization Task Force RQL – Rejected Quality Limit
IQ – Installation Qualification SPC – Statistical Process Control
LQ – Limiting Quality UQL – Unacceptable Quality Level
LTPD – Lot Tolerance Percent Defective VMP – Validation Master Plan

IMPORTANT NOTICE
Process validation in medical devices
Revision: 0
Effective date: 15 September 2017

COPYRIGHT NOTICE
The information contained in this document represents the current view of TÜV SÜD on the issues discussed as of the date of publication. Because TÜV SÜD must respond to changing market
conditions, it should not be interpreted to be a commitment on the part of TÜV SÜD, and TÜV SÜD cannot guarantee the accuracy of any information presented after the date of publication.
This guideline is for informational purposes only. TÜV SÜD makes no warranties, express, implied or statutory, as to the information in this document. Complying with all applicable
copyright laws is the responsibility of the user. Without limiting the rights under copyright, no part of this document may be reproduced, stored in or introduced into a retrieval system, or
transmitted in any form or by any means (electronic, mechanical, photocopying, recording, or otherwise), or for any purpose, without the express written permission of TÜV SÜD.
TÜV SÜD may have patents, patent applications, trademarks, copyrights, or other intellectual property rights covering subject matter in this document. Except as expressly provided in
any written license agreement from TÜV SÜD, the furnishing of this document does not give you any license to these patents, trademarks, copyrights, or other intellectual property. ANY
REPRODUCTION, ADAPTATION OR TRANSLATION OF THIS DOCUMENT WITHOUT PRIOR WRITTEN PERMISSION IS PROHIBITED, EXCEPT AS ALLOWED UNDER THE COPYRIGHT
LAWS. © TÜV SÜD Group – 2017 – All rights reserved - TÜV SÜD is a registered trademark of TÜV SÜD Group.

DISCLAIMER
All reasonable measures have been taken to ensure the quality, reliability, and accuracy of the information in the content. However, TÜV SÜD is not responsible for the third-party content
contained in this guideline. TÜV SÜD makes no warranties or representations, expressed or implied, as to the accuracy or completeness of information contained in this guideline. This
guideline is intended to provide general information on a particular subject or subjects and is not an exhaustive treatment of such subject(s). Accordingly, the information in this guideline
is not intended to constitute consulting or professional advice or services. If you are seeking advice on any matters relating to information in this guideline, you should – where appropriate
– contact us directly with your specific query or seek advice from qualified professional people. TÜV SÜD ensures that the provision of its services meets independence, impartiality and
objective requirements. The information contained in this guideline may not be copied, quoted, or referred to in any other publication or materials without the prior written consent of
TÜV SÜD. All rights reserved © 2017 TÜV SÜD.

TÜV SÜD | Process validation in medical devices 15


Contact us for expert regulatory guidance on
manufacturing processes for medical devices
www.tuv-sud.com/medicaldevice
medicaldevice@tuv-sud.com

Choose certainty. Add value.


TÜV SÜD is a premium quality, safety and sustainability solutions provider that specialises in testing, inspection,
auditing, certification, training and knowledge services. Represented in over 1,000 locations worldwide, we hold
accreditations in Europe, the Americas, the Middle East, Asia and Africa. By delivering objective solutions to our
customers, we add tangible value to businesses, consumers and the environment.

TÜV SÜD AG
2017 © TÜV SÜD AG | V-M/MHS/20.1/en/SG

Westendstr. 199
80686 Munich
Germany
+49 89 5791 0
www.tuv-sud.com

16 Process validation in medical devices | TÜV SÜD

Das könnte Ihnen auch gefallen