Beruflich Dokumente
Kultur Dokumente
- REVISION RECORD 2
- INDEX 3
1.0 PURPOSE 4
2.0 SCOPE 4
3.0 DEFINITION 4
4.0 DIVERSION 4
5.0 RESPONSIBILITIES 4
6.0 CALIBRATION 5
9.0 DOCUMENTS 6
’
1. PURPOSE
2. SCOPE
This procedure details the requirement that is required to be implemented for the
welding machine calibration.
3. EXPLANATION / DEFINITIONS
4. DIVERSION
It is the difference in magnitude between the current rating of the welding machine
and the actual value measured in amperage volt ammeter clamp, taking as basis the
corresponding nominal value.
5. RESPONSIBILITIES
6. CALIBRATION
6.4.1. All welding machine such as inverter welding machine, rectifier welding
machine, A/C – transformer welding machine, an engine driven welding
machine should be Calibrated and certify before use, and thereafter every six
months.
6.4.2. All Master instruments should be Calibrated and certify before use, and
thereafter every six months.
Note: This period can be set by the contractor (CLIENT) but not higher than
that specified above.
6.4.3.1. Welding Electrode cable clamp connect to the positive terminal of the
welding machine to be test and calibrate.
6.4.3.2. Welding ground cable clamp connector to the negative terminal of the
welding machine to be test and calibrate.
6.4.3.3. Energize the welding machine. ((ON)Electrically or (Start)Engine
driven)
6.4.3.4. Test and record welding machine OCV (Open Circuit Voltage) and
verify with the welding machine name plate.
6.4.3.5. Set the welding parameter positioning liver in the desired
position.(Set point)
6.4.3.6. Engage with the volt – ammeter clamp the cable clamp electrode.
6.4.3.7. Carry out the reading of the clamp volt – ampere ammeter,
comparing it with the values shown on the recorder panel of the welding
machine.(Welding machine Dial)
Note; Evidence records calibration results of deviations in from. Calibrations should happen in
bands of every (It should indicate the calibration bands for gradual measurement are taken
e.g. every 40A, 60A, 80A
Numbering is the same numbering allocated as indicated in the procedure
)
7.1. The results of the calibration process must be registered in the Control of Calibration
of welding machine. It should be kept track of time calibration, so no equipment is
used with their deadlines expired.
7.2. The equipment may only be labelled when they are approved.
8. ACCEPTANCE CRITERIA
The maximum permissible deviations for calibration are: + / - 20% on the Set point.
Note: The maximum deviation is in accordance with the criteria established by Client.
Porum not exceeding the amount fixed by the Standard.
9. DOCUMENTS