General US300PM
Specifications Ultrasonic Flowmeter, Portable type

GS 01G05B02-01E

■ General
US300PM is a portable type ultrasonic flowmeter that
measures the flowrate in a pipe fully filled with liquid.
It can be used not only for the maintenance and the
check of your facilities but also for the measurement at
laboratories or manufacturing lines. The housing and
transducers are made for rough industrial environment.
The measuring principle is based on the influence of the
flowing fluid to the travelling time of sound. The sound is
transmitted through the pipe and the transit time differ-
ence between the forward and backward beams is used
to determine the flow velocity (transit time method).
The measurement is non-intrusive. No cutting of the pipe
is necessary. The measurement is independent of fluid
pressure or conductivity. Owing to its clamp on transduc-
ers and a rechargeable battery the instrument is an ideal ■ Standard Specifications
tool for service work.
For the hygienic measurement, there is no risk of 䊉 General
contamination and suitable for ultra clean liquids as the Fluid: Liquid (Turbidity < 10,000 mg/L,
transducers do not touch the liquid. Also, there is no risk fluid sound speed 800 to 3500m/s)
of corrosion when used with aggressive media.
Measured quantities: Volume flow, mass flow (by
The instrument can work with up to two pairs of transduc- setting density), flow velocity,
ers. This enables you to measure the average flow for sound speed in the fluid
disturbed flow profile (two path measurement), the
Wall thickness of the pipe (when
differential flow of two independent flows, or simply two
optional wall thickness probe is
different flows independently.
available)
Measuring principal: Transit time method using
■ Features ultrasonic signal
• Two channel inputs as standard Pipe size: 25 to 6500 mm
• Two current outputs maximum (The flow data and the
Pipe and lining material: Carbon steel, Stainless steel,
sound speed in the liquid can be assigned at the same
Grey cast iron, Ductile iron,
time)
Copper, Glass, PVC, etc
• RS232 serial interface as standard
• Data storage memory 100,000 values maximum Flow velocity range: 0.01 to 25 m/s
• Wall thickness measurement function as standard Resolution: 0.025 cm/s
(probes are optional) Accuracy: 1 to 3% of reading depending on
• Battery operation possible (14 hours maximum) application
• Interactive setup menu for easy installation
Measuring cycle: 100 to 1000 Hz (when only one
channel input)
Straight pipe run in the upstream: 10 to 50 diameters,
depending on the kind of flow
disturbance

GS 01G05B02-01E
©Copyright Sep. 2001(YG)
1st Edition Sep. 17, 2001(YG)
<<Contents>> <<Index>> 2
䊉 Ultrasonic flowmeter (US300PM)
Construction: Parameter setting storage function:
Housing material: Aluminium (powder coated) Function: Storage of pipe and fluid parameters
Water and dust-proof: IP54 (EN60529) (Maximum 80 different settings)
IPx4 (JIS C 0920) Storage of all parameters (Maximum
14 different settings)
Input:
Calculation function:
Number of input channels: 2 (Cannel A, Channel B)
Flow value: Flow velocity
Both transducers and wall thickness
probe can be connected freely Volume flow or mass flow rate and
totalization (both positive and
Output:
negative flow totalization)
Current output: 0 to 2 outputs
Sound velocity: Sound velocity in the fluid
Range: 4 to 20 mA
Wall thickness: Wall thickness of the pipe available
Flow velocity, volume flow, or sound using wall thickness probe (optional).
speed in the liquid can be freely
Transfer function of the measured
assigned
wall thickness to the pipe parameter
Frequency output: 0 or 1 output available
Range: 0 to 1 kHz Calculation for the two flow inputs:
Contact type: Open-collector, Two values from average, sum, or
24 V/4 mA difference of the channel A and
The value indicates instantaneous channel B inputs are available freely
flow rate. Output assignment: Calculated values above except for
Binary output (pulse or alarm): the wall thickness can be freely
assigned to the actual outputs (two
0 to 2 outputs
channel-independent outputs
Contact type: Open-collector, available)
24 V/4 mA Output damping: 0 to 100 seconds
The output values are selectable for Alarm:
each output.
Alarm items: High limit, low limit, flow direction
The pulse outputs indicate the total change, quantity limit (for batch
volume flow (0.01 to 1000/unit) with operation), error (measurement
pulse width 80 to 1000 ms. impossible)
Terminal type: Banana plug terminals (+, –) Output hold type: Non-hold or Hold
Display and Setting: Output contact direction: Normal Open or Normal Close
LCD display: 216 characters LCD with back light Data logging function:
that can be stitched on or off.
Function: Store measured values in the internal
Two values can be displayed at the memory
same time.
Memory size: 27,000 values (standard)
LED lamp: SIGNAL lamp: Indicates the status of
100,000 values (optional)
the measuring signal from each input
channel (green or red light)
BATTERY lamp: Indicates the status Communication function:
of battery voltage, etc. Type: RS232
Connector: D-sub 9-pin connector, male
Keyboard: 15 keys (numeric and function keys) Function: On-line/Off-line output of the
Easy operation by the interacive measured values to personal
menu display computers or serial printers

Display language: Following languages selectable: Time-programmable measurement function:


Czech, Danish, Dutch, English, Function: Automatic start and stop of the
French, German, Norwegian, Polish, measurement using internal clock.
Turkey Can be used with data logging or
communication function

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 01G05B02-01E 1st Edition Sep.17,2001-00
<<Contents>> <<Index>> 3
Power supply: 䊉 Wall thickness probe
Power supply type: Internal rechargeable battery (6 V/4 Type of usage:
Ah) or power supply adapter (100 to Fluid temperature: General temp. type: 20 to 60 deg C
240 V AC input, used also as battery
(option /WTG or model USPA301)
charger)
High temp. type: 0 to 200 deg C
Battery operating time: Maximum 14 hours
(option /WTH or model USPA302)
Power consumption: less than 15W
Rating:
Safety and EMC standard:
Measuring range: 1.0 to 200 mm (This depends on the
General safety: EN61010 (CE marking)
material)
EMC regulation: EN61326 (CE marking)
Resolution: 0.01 mm
AS/NZS 2064 (C-Tick mark)
Operating conditions: 䊉 Accessories
Ambient temperature: 10 to 60 deg C Standard accessories for US300PM:
Transportation case
䊉 Transducers (US300PT)
User’s Manual
Type of usage:
Measuring tape
Dust and water-proof:
Battery set (built-in in the main unit)
General purpose:
Others (fixing hardware, couplant, etc):
IP65 (EN60529) , IPx5 (JIS C 0920)
Some are selectable in the model
Immersible: and suffix code (see next page) of the
IP67 (EN60529), IPx7 (JIS C 0920) main unit or transducers, or separate
Pipe size type: Medium size: 25 to 400 mm orders are also possible
Large size: 100 to 2500 mm
䊉 Data transfer software (Will be available soon)
Very large size: 2000 to 6500 mm
Function: Download and upload of the param-
Fluid temperature: General temp. type: eters or download of logging data via
30 to 130 deg C RS232 communication port
High temp. type: 30 to 200 deg C Standard accessories: RS232 cable and RS232 adapter 9/25
Construction: Operating system: Windows 95, 98, ME, NT, 2000
Case material: Stainless steel Language: English / German version
Contact surface material:
General temp. type: PEEK (Poly
Ether Ether Keton)
High temp. type:Polyimid
Cable protection material:
Stainless flexible tube

䊉 Optional extension cable (US300PC)


Length: 5 m, 10 m, 20 m
Use this when the length of the cable
part of the transducers is not enough.

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 01G05B02-01E 1st Edition Sep.17,2001-00
<<Contents>> <<Index>> 4
■ Model and Suffix Code
Ultrasonic flowmeter, portable type Transducers for portable type

Model Suffix code Specification Model Suffix code Specification


US300PM · · · · · · · · · · · · · · · Ultrasonic flowmeter, portable type US300PT · · · · · · · · · · · · · · · · Transducers for portable type

-A0 · · · · · · · · · · · · No current output Usage -G · · · · · · · · · · · · · General purpose (IP65)

Output -A1 · · · · · · · · · · · · One current output -W · · · · · · · · · · · · · Immersible (IP67)

-A2 · · · · · · · · · · · · Two current outputs Pipe Size / BG · · · · · · · · · · · Medium & General (with 3 m
Fluid cable) (Note)
Power Supply 1·········· Japan Temperature
BH · · · · · · · · · · · Medium & High (with 3 m cable)
Adapter and 2·········· USA (Note)
AC cable
3·········· Europe CG · · · · · · · · · · · Large & General (with 4.4 m
cable) (Note)
4·········· United Kingdom
CH · · · · · · · · · · · Large & High (with 4.4 m cable)
5·········· Australia
(Note)
6·········· South Africa DG · · · · · · · · · · · Very large & General (with 12 m
-2 · · · · · · · · Always 2 cable) (Note)

Always N (Note)
-N· · · · · ·
B: Medium size (25 to 400 mm)
Option /PU1 · · One binary (pulse or alarm) C: Large size (100 to 2500 mm)
output (open-collector) (Note) D: Very large (2000 to 6500 mm)
/PU2 · · Two binary (pulse or alarm)
outputs (open-collector) (Note) G: General temp. (–30 to +130
deg C)
/FQ1 · · Frequency output (open- H: High temp. (–30 to +200 deg C)
collector, 0 to 1kHz)
Mounting -S · · · · · · · · · Standard type (set of two
/DLX · · Data logging extension fixture blocks, including ruler marked
(100,000 values) length 330 mm)
/BGT · · Tag number on nameplate (in -M · · · · · · · · · Magnetic type for general temp.
the nameplate label, maximum (–30 to +100 deg C, set of two
16 characters) blocks, including ruler marked
/WTG · · Wall thickness probe for length 330 mm)
general temperature (–20 to -N · · · · · · · · · None
+60 deg C)
Fixing chain B · · · · · · · · For 25 to 1200mm
/WTH · · Wall thickness probe for high
Fixing chains (1 x 2)
temperature (0 to +200 deg C)
Extensional fixing chains (12)
T01.EPS
C · · · · · · · · For 1200 to 3000 mm
(Note) Option /PU1 and /PU2 are exclusive. Fixing chains (1 2)
Extensional fixing chains (42)

Optional extension cable for portable type D · · · · · · · · For 3000 to 6500 mm


Fixing chains (12)
Extensional fixing chains (102)
Model Suffix code Specification
N · · · · · · · · None
US300PC · · · · · · · · · · · · · · · Optional extension cable for
portable type Acoustic couplant G · · · · · · General temperature type (–30
to +130 deg C)
Length -A005 · · · · · · · · · · Cable length 5[m]
H · · · · · · High temperature type (–30 to
-A010 · · · · · · · · · · Cable length 10[m] +200 deg C)
-A020 · · · · · · · · · · Cable length 20[m] N · · · · · · None
T02.EPS
Option /TTP · · Transducer tag plate
(maximum 16 characters)
Data transfer software (Will be available soon) T04.EPS

Model Suffix code Specification


US300SA · · · · · · · · · · · · · · · Data transfer software
(Windows version) Including
connecting kit (RS232 cable for
connection with IBM-PC, RS
232 adapter 9/25)
Language -1 · · · · · · · · · · · · · · English / German version
00 · · · · · · · Always 00
T03.EPS

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 01G05B02-01E 1st Edition Sep.17,2001-00
<<Contents>> <<Index>> 5

Accessories (for ultrasonic flowmeter US300PM) Accessories (for transducers US300PT)

Model Description Model Description


USPA301 Wall thickness probe for general temperature USPA001 Fixing strap (10 m length)
(–20 to +60 deg C)
USPA002 Fixing strap (20 m length)
USPA302 Wall thickness probe for high temperature
(0 to +200 deg C) USPA011 Fixing clips
(medium type, for pipe size 40 to 100 mm, set of two
USPA311 Power supply adapter (100 to 240 V AC input) clips)
USPA321 AC cable (Japan) USPA012 Fixing clips
USPA322 AC cable (USA) (large type, for pipe size 100 to 6500 mm, set of two
clips)
USPA323 AC cable (Europe)
USPA021 Fixing band
USPA324 AC cable (United Kingdom)
(only for pipe size 25 to 50 mm)
USPA325 AC cable (Australia) USPA031 Fixing chain
USPA326 AC cable (South Africa) (for pipe size 25 to 600 mm)

USPA331 Battery set (6 V 4Ah) USPA032 Extensional fixing chain


(2 m length, equal to +600 mm diameter)
USPA341 Transportation case
USPA033 Repair set for fixing chain
T05.EPS

USPA054 Mounting fixture standard type


Accessories (others) (for transducers medium pipes size type, temperature
range –30 to +200 deg C, set of two blocks)
Model Description USPA055 Mounting fixture magnetic general temperature type
(for transducers medium pipe size type, temperature
USPA401 RS232 cable
range –30 to +100 deg C, set of two blocks)
USPA402 RS232 adapter 9/25
USPA057 Mounting fixture standard type
USPA411 Measuring tape (for transducers large or very large pipe size type,
temperature range –30 to +200 deg C, set of two blocks)
T06.EPS
USPA058 Mounting fixture magnetic general temperature type (for
transducers large or very large pipe size type,
temperature range –30 to +100 deg C, set of two blocks)
USPA073 Additional magnets for mounting fixture magnetic
general temperature type
(for transducers medium pipe size type, temperature
range –30 to +100 deg C, set of two magnets)
USPA075 Additional magnets for mounting fixture magnetic
general temperature type
(for transducers large or very large pipe size type,
temperature range –30 to +100 deg C, set of two
magnets)
USPA081 Ruler for the mounting fixture
(marked length 120 mm)
USPA082 Ruler for the mounting fixture
(marked length 330 mm)
USPA091 Acoustic couplant for general temperature
(100 g, –30 to +130 deg C)
USPA092 Acoustic couplant for high temperature
(100g, –30 to +200 deg C)
F07.EPS

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 01G05B02-01E 1st Edition Sep.17,2001-00
<<Contents>> <<Index>> 6
■ Dimensional Drawings
Unit: mm (inch)
Transducers US300PT-ⵧBⵧ
Ultrasonic flowmeter US300PM

276(10.9) 43(1.7)

(0.7)
18
115(4.53)

ø15

(0.9)
22
X Y
F02.EPS

Transducers US300PT-ⵧCⵧ, US300PT-ⵧDⵧ


60(2.4)
268(10.6)

315(12.4)

(1.3)
34
ø15

(1.2)
30
X Y
F01.EPS F03.EPS

Weight: 3.9 kg (8.60 lb) Length


(incl. battery set)
Transducers X m (inch) Y m (inch) X+Y m (inch) Weight kg (lb)
US300PT-ⵧBⵧ 2.0 ( 78.7) 1.0 ( 39.4) 3.0 (118.1) 0.6 (1.32)
US300PT-ⵧCⵧ 2.0 ( 78.7) 2.4 ( 94.5) 4.4 (173.2) 1.2 (2.65)
US300PT-ⵧDⵧ 5.0 (196.9) 7.0 (275.6) 12.0 (472.4) 2.2 (4.85)
T08.EPS

Optional extension cable US300PC-Aⵧⵧⵧ

ø15

L
F04.EPS

Length

Optional extension cable L m (inch) Weight kg (lb)


US300PC-A005 5.0 (196.9) 0.4 (0.88)
US300PC-A010 10.0 (393.7) 0.6 (1.32)
US300PC-A020 20.0 (787.4) 1.6 (3.53)
T09.EPS

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 01G05B02-01E 1st Edition Sep.17,2001-00
<<Contents>> <<Index>> 7

Unit: mm (inch)
Wall thickness probe for general temperature (20 to 60 deg C)
(option /WTG or model USPA301)

48(1.9)
ø15

F05.EPS

Wall thickness probe for high temperature (0 to 200 deg C)


(option /WTH or model USPA302)

ø22(0.9)
ø15

L
F06.EPS

Length

Wall thickness probe L m (inch) Weight kg (lb)


/WTG or USPA301 1.5 (59.1) 0.17 (0.37)
/WTH or USPA302 1.2 (47.2) 0.19 (0.42)
T10.EPS

Ordering Information
Specify the following when ordering.

1.Model, suffix and option code


(1) First, specify the main unit US300PM and the transducers US300PT (one pair of transducers when using only one
channel or two pairs when using two channels). The transducers can be added later.
(2) Specify the wall thickness probe if necessary in the option code for the main unit US300PM (code /WTG or /WTH).
They can be purchased later as model USPA301 or USPA302.
(3) Specify the optional extension cable if necessary.
(4) Specify the accessories when necessary or suitable. For example we may use “fixing strap” instead of “fixing chain” to
fix the transducers to the pipe. Please consult us for the details.

2. Tag number (when necessary)


Maximum sixteen (16) characters for the cases below.
(1) In the nameplate lable of the main unit US300PM (necessary to specify the option /BGT at the same time)
(2) In the tag plate for the transducers US300PT (necessary to specify the option /TTP at the same time)

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 01G05B02-01E 1st Edition Sep.17,2001-00