SPECIFICATIONS

for
Diagnostic Ultrasound System

Model SSD-5000
A Platform for Pure Harmonic Detection

THE INNOVATOR IN ULTRASOUND


Scanning Method Advanced Technologies

・ Electronic Convex Sector ・ Harmonic Echo


・ Electronic Linear The Harmonic Echo is the technology to construct
・ Electronic Phased Array Sector the image by using the 2nd harmonic components.
・ Mechanical Sector/Radial The Harmonic Echo provides clearer edge
definition, reduced side lobe artifacts, and less
Image Display Modes
reverberation noise compared to the conventional
・ B: gray-scale imaging fundamental frequency imaging.
・ D: Spectral Doppler (PW, HPRF PW, and CW)
・ PureHD (Pure Harmonic Detection)
・ Dual B
The PureHD technology realizes distortion-free
・ B and M
ultrasound waveform transmission which enables
・ B and D
highest quality Harmonic Echo imaging.
・ B, M, and D
・ B(Flow) ・ HST (Hemispheric Sound Technology)
Flow: Color Doppler and PowerFlow imaging Forms ideal hemispheric sound source with less
・ Dual B(Flow) side lobe artifacts both in array direction and
・ M(Flow) elevation direction.
・ B(Flow) and M(Flow)
・ Impedance Matching Technology
・ B(Flow) and D (Triplex mode: B, Flow, and PW
Prevents signal loss by total impedance matching
Doppler simultaneous real-time display is
from transducer to the front-end.
possible.)
・ B(Flow), M(Flow) and D ・ New-generation Wide Dynamic Range Digital
・ Dual Dynamic Display: B and B(Flow) simultaneous Front-end

real-time display The 12-bit A/D converter provides superior


・ KI (Kinetic imaging)* — included in the A-SMA contrast resolution compared to conventional
function digital systems and analog systems.
・ TDI (Tissue Doppler Imaging)* ・ QFI (Quint Frequency Imaging)
Available on all imaging probes. You can select ultrasound frequencies from among
・ Volume Mode* up to 5 selections with one probe. The QFI can
* Option provide better resolution with a higher frequency
Request function: In multi-display mode, it is possible and better penetration with a lower frequency.
to select one mode for full screen display. (Selectable frequencies are probe dependent.)

CQ* (Cardiac Quantification)


Beamformer
Displays in real time the line graph of the left
12-bit Digital Beam Former
ventricular volume, EF (Ejection Fraction) and each
rate of change. Includes the A-SMA and KI functions
System Dynamic Range
described below.
172 dB A-SMA (Automated Segmental Motion Analysis):
・ On-line quantification of regional wall motion
Number of Processing Channels ・ Line graph display of FAC (Fractional Area Change)
256 channels and area.
・ Histogram display of FAC with KI (kinetic imaging)
Frame Rate KI (Kinetic Imaging)
Displays the movement of the endocardial walls
・ Max. 1000 frames/sec or more*
with color gradation with time.
*Probe and system setting dependent
* Option
(1,493 frames/sec in the case of UST-5531)

2
B-mode Spectral Doppler:
・ Imaging Gray Scale: 256 levels
・ Display: Power spectrum
・ Scanning area:
・ Doppler methods:
100% to 25%, continuously changeable
PW (Pulsed Wave)
・ Image Zoom
HPRF (High Pulse Repetition Frequency) PW
Write zoom (magnification of real-time image):
CW (Continuous Wave)
Max. 6 times
(Available methods are probe dependent.)
Read zoom (magnification of frozen image):
・ Real-time Doppler Auto Trace: Possible (option)
Up to 2 times
・ Reference frequency:
・ Depth range selections:
PW: 2.14, 2.5, 3, 3.75, 5, 6, and 7.5 MHz
2 to 24 cm*, changeable by minimum 1 cm
CW: 2.14 and 3 MHz
step (* Probe dependent)
・ Pulse repetition frequency on PW Doppler:
・ Image frequency select: 4 selections
PW: 0.5 to 20 kHz
Resolution, Standard 1 & 2, Penetration
・ Max. velocity range:
・ Scanning line density:
PW/HPRF PW: +5.6 or –5.6 m/sec *
3 selections on linear, convex sector and,
CW: +15.92 or –15.92 m/sec *
phased array sector probes
* 2.14 MHz reference freq., 0 degree, with
・ Gain: 30 to 90 dB, continuously variable *
base line shift)
* Possible to change also after freezing.
・ Base line shift: Available up to double velocity
・ STC (sensitivity time control) — gain versus depth
・ Steerable CW Doppler: Possible on phased array
curve control: 8 slide controls
sector probes
・ Angle gain (Lateral gain control)* — gain versus
・ Steered linear scanning: Possible on selected linear
angle curve control: 8 slide controls
probes
* Available on phased array sector image
Max. ±30 degrees *, 1 or 5 degrees step
・ Contrast: 16 steps
changeable
・ AGC—Suppression of brightness saturation:
* Probe dependent
16 steps
・ Spectrum inversion: Possible
・ Relief: 4 selections (Off + 3 levels)
・ Angle correction:
・ Frame correlation: Continuously variable
Available up to 80 degrees presetable
・ Horizontal smoothing: 3 selections (Off + 2 levels)
・ Sample volume size for PW Doppler:
・ Post Processing
0.5 - 10 mm, changeable in 0.5 mm step
Echo enhance curve: 3 kinds
・ Wall motion filter:
Rejection: 64 steps
Manual:
・ View gamma: 5 kinds
50, 100, 200, 400, 800 or 1600 Hz,
・ Image orientation
PW and CW independently selectable
Longitudinal and lateral inversion
Auto: varies in 12 steps for a velocity range
90 degrees rotation
10 Hz – 1500 Hz
M-mode: ・ Doppler gain: up to 50 dB, 1 dB step changeable*
・ Moving speed: 1, 1.5, 2, 3, 4, 6, or 8 s/screen * Possible to change also after freezing.
・ Gain: B-gain ±30 dB, continuously variable * ・ Contrast: 16 steps, selectable
* Possible to change also after freezing. ・ Black/white inversion: Possible
・ Contrast: 16 steps ・ Audio output: 2 channels stereo
・ AGC—Suppression of brightness saturation:
16 steps
・ Relief: 4 selections (Off + 3 levels)
・ Smoothing processing: On/off selection
・ Window function—zoom in on the region of interest
・ Free Angular M-mode*—M-mode cursor can be
freely set on real-time B-mode image (* Option)
3
Color Flow Imaging Volume Mode*
・ Represented parameters: ・ Scanning method:
Color Doppler mode: Mean Doppler frequency shift Electronic convex sector & mechanical sector
PowerFlow mode: Total power ・ Image processing
・ Color area size: Variable from 100% to 5% Pre-processing: Gain, Opacity, Contrast, Gamma
・ Steered linear scanning: Possible Post-processing: Brightness
Max. ±30 degrees *, 1 or 5 degrees step ・ Image magnification (possible on frozen images):
changeable From 0.5 to 2 times in 0.25 times increment

* Probe dependent ・ Image rotation (possible on frozen images):


・ Color enhance: 5 selections Possible by 45 degrees

・ Color line density: 3 selections


・ Image frequency select: 4 selections Volume Mode probes*

Resolution, Standard 1 & 2, Penetration ASU-1000C ASU-1001 ASU-1002


Scanning angle
・ Color smoothing: 7 selections Convex 60゚ 90゚ 180゚
・ Flow filter: 6 selections sector
Mechanical
・ Frame correlation: 4 selections sector
・ Motion Reject: 4 selections Manual: 60゚/45゚/30゚ 60゚/45゚/30゚ 90゚/60゚/45゚
Auto: 60゚ - 30゚ 60゚ - 30゚ 90゚ - 45゚
・ Sampling average rate: 3 steps selectable Scanning 4 - 24 cm 4 - 19 cm 3 - 17 cm
・ Color coding: depth range
(21 steps) (16 steps) (15 steps)
Abdomen: 5 selections
Maximum 2 frames/s 4 frames/s 2 frame/s
PV: 5 selections frame rate
Selectable 2.5 -6 MHz 2.5 - 6 MHz 3 - 7.5
Cardiology: 6 selections
frequency MHz
Variance: 1 kind range
PowerFlow: 5 selections * Option
PowerFlow
・ Gradation: 32 levels Cine memory
・ Color coding: 5 selections Memory capacity:
・ Background ON/OFF: Possible ・ Cine memory (B mode): Max. 1024 images
・ Color resolution: 4 selections Loop display: possible
Color Doppler ECG timed display: possible
・ Reference frequency: Quadruple (4-split screen) display: possible
2.14, 2.5, 3, 3.75, 5, 6 and 7.5 MHz (loop display is possible on 2 images.)
・ Pulse repetition frequency: ・ Store memory (B and/or M, or D mode):
0.5 to 10.0 kHz Max. 100 images (The upper limit varies
・ Max. velocity range: depending on ECG ON/OFF and other
+1.23 or –1.23 m/sec (2.14 MHz reference conditions.)
freq.) ・ Scrolling Memory (M or D mode):
・ Color base line shift: Max. 1024 seconds (at 8 seconds/plane)
Available up to double velocity (±31 steps) Note: The number of storable images depends on
・ Gradation: probe type, scanning angle and other conditions.
±32 levels for velocity (red and blue)
16 levels for variance (green) Physiological Signal Display*
・ Color polarity invert: Possible
・ 3 channels of ECG, PCG, and Pulse wave
・ ECG lead ― second lead system
・ ECG synchronized display:
Available for one phase
* Option

4
DMS (Data Management Subsystem):
Built-in digital interface option: DMS-5500B
<DICOM network communication>
・ Conformity to DICOM standard:
<Memory>
Service class user of storage, print , and basic work
● Cine Memory
list management. (For details, please refer to the
・ Display format: DICOM Conformance Statement issued by Aloka.)
Single (Full screen), Dual (2-split screen) and
・ Storage: Directly transmits image with patient
Quadruple (4-split screen) display
information to a DICOM file server
・ Loop image playback: Print: Directly printout images with DICOM
Possible independently on each split screen
compatible printer
Playback speed changeable
Work list management: Retrieval of patient and
・ Play-through function―simultaneous display of live reservation information from hospital information
image and cine memory image
system (HIS)
・ ECG synchronized image acquisition: possible NOTE: The HIS needs to be compatible with DICOM
・ Cine Memory capacity: standard supplement 10. The HIS network and the
NTSC
DICOM network need to be linked.
Max. 256 images (B/W) or 128 images (color)
・ Router setting
PAL
Max. 224 images (B/W) or 112 images (color)
<Image processing>
● Non-volatile Memory
●3-D image processing
・ Hard disk capacity: Up to 999 images ・ Volume Mode―photo-realistic display suitable for
・ Film Box function―thumbnail display of stored
imaging fetus
images and print format editing
・ Spatial Echo―pseudo 3 dimensional display of
・ Print Queue function―printout of stored images gray-scale image
directly with DICOM compatible printer
・ Spatial Flow―pseudo 3 dimensional display of
blood flow image
<Digital image transmission>
・ Spatial Color―pseudo 3 dimensional display of
・ Data compression: Lossless compression
gray-scale and blood flow images
・ Image degradation: 0%
・ 3D Slicing― display of planes viewed from the side
・ Output medium: and the bottom
MO disk (128 Mbyte/230 Mbyte/540 Mbyte)
●Kinetic Imaging
Ether net (Network interface: 10 BASE/T or 100
Displays the movement of the endocardial wall with
BASE/TX, automatically switched.)
color gradation with time
Max. data transfer rate: 10 Mbps or 100 Mbps,
according to the network interface)
<Stress Echo Package>
・ Output file format: BitMap and TIFF ・ Stress scoring*3
・ Data attendant to the image (Automatically
・ Quad-screen display
transferred from ultrasound main unit):
・ Exercise stress protocols:
Patient information
2 phases (pre-exercise/post-exercise)
Study information
4 phases (pre-exercise/peak
Display mode
exercise/post-exercise/recovery)
Scale factor
・ Pharmacological stress protocols:
Baseline position
BLPR (Baseline/low dose/ peak dose/ recovery)
Look-up Table for color
BLMP (Baseline/low dose/ mid dose/ peak dose)
・ User's protocol: The user can make a protocol in
<Measurements and calculations>
which 8 stages/phases can be set at maximum.
・ Those available on the main unit are also available
on the DMS.
・ BETA (Backscattered Energy Temporal Analysis) *1 Necessary optional kits:
・ OB/GYN report data output*2 *1 PM-DMS-10
*2 PM-DMS-15
*3 PM-DMS-16

The above specifications are applied to Ver. 3.5.

5
Measurement and Analysis: M mode
Mitral valve
● Basic measurements
Aortic valve
On B-mode image
Tricuspid valve
Distance
pulmonary valve
Area and Circumference by ellipse/by trace
Doppler mode
Angle
LV in-flow study
Volume
LV ejection-flow study 1, 2
Histogram
RV in-flow study
PSA density
RV ejection-flow study 1, 2
Hip joint angle
SV/CO
On M-mode image
Flow volume
Velocity
● Peripheral vessels analysis
Time interval
CCA (common carotid artery)
Heart rate
ICA (internal carotid artery)
On spectral Doppler
ECA (external carotid artery)
Velocity
Vertebral artery
Acceleration
% Stenosis area
Average flow velocity
% Stenosis distance
Time interval
S/D ratio
Resistance index
● Report Functions
Pulsatility index
・ Obstetrical report
Pressure gradient
・ Cardiac report
Pressure half time
・ PV report
Heart rate
● Post measurement on VCR playback image:
・ Doppler auto trace: Possible on frozen image as
Available with auto and manual calibration.
standard. Possible on real-time image with
optional kit.
Data Communication Function (ALK-3)*
It is possible to transmit patient data and OB/GYN
Others
Stenotic rate, ratio measurement data obtained by Aloka diagnostic
ultrasound system to a personal computer.
● Cardiac analysis * Optional Isolation Unit model SIU-5500-1 is
Left ventricle analysis: necessary. Please prepare personal computer,
M mode appropriate software, and connection cable at your
Pombo, Pombo (wall) side. As for necessary conditions of such items,
Teichholz, Teichholz (wall) please contact your local distributor or Aloka office.
Gibson, Gibson (wall)
B mode Direct Printout *
Pombo, Pombo (wall)
It is possible to printout report of OB/GYN and
Teichholz, Teichholz (wall)
urological measurement directly to an external
Gibson, Gibson (wall)
printer.
Single plane ellipsoid
* Optional Isolation Unit model SIU-5500-1 is
Bi-plane ellipsoid
necessary. Please prepare printer and connection
Bullet
Modified Simpson cable at your side.

Simpson Compatible printers: An HP printer of PCL (Printer


Measurements: Command Language) level 3 or higher
B mode Checked with HP Deskjet 890C
Long axis view
Long axis view with wall thickness
Short axis view
Apex view

6
Viewing Monitor
Acoustic Power ・ 15-inch diagonal
・ 0 to 100%, changeable Flicker-free non-interlaced scanning (option)
・ Tilt and swivel are possible.
Preset Function
・ 15 separate programs for specific clinical
applications and/or users Safety Regulation
・ User programmable and/or factory default settings ・ Complies with IEC 60601-1 Class Ⅰ, Type BF
・ Preset contents are storable in a floppy disk by
Environmental Requirements
using PC In Operation
・ Temperature: +10 to +40 degrees C
Characters and graphic displays
・ Relative Humidity: 30 to 85%
・ Character input area: (non condensing)
ID, name, age, sex, retained text ・ Atmospheric pressure: 700 to 1060 hPa
・ Automatic Annotation Labeling: In Storage
・ Temperature: -10 to +50 degrees C
Up to 120 words registerable to each application
・ Relative Humidity: 10 to 90%
・ Body mark: Default setting: 53 kinds (non condensing)
User edition: 24 kinds ・ Atmospheric pressure: 700 to 1060 hPa

Menu Control Power Requirement


・ Color LCD (liquid crystal display) touch control ・ 115/ 200 to 240V ±10%, 50 or 60 Hz,
panel Max. 950VA for main unit
Max. 350VA for optional units
Probe Connectors Total: Max. 1300VA
・ Electronic probes: 3 separate connectors
Dimensions
・ 54 cm (W) × 93 cm (D) × 137 cm (H)
Video Input/Output (for VCR)
・ Composite, RGB, and Y/C separated signals, 1 Weight

channel each ・ Approx. 180 kg (main unit only)

7
System Configuration

Standard Components
Main unit with 15-inch color viewing monitor Standard components and optional items depend on the country.
Probe * * Aloka recommend UST-9126 for general abdominal imaging and
Accessories UST-5297 for general cardiac examination.

Optional Functions
3-point footswitch MP-2614B
Tissue Harmonic Echo * : PM-5500-19

Ultrasound gel (300 g):


*2 Possible with UST-9126, UST-9119, UST-5297, UST-5545, and UST-5284-2.5. GEL-SCAN-PA
Contrast Harmonic Echo *3:
NOTE: Ultrasound gel cannot be
CHM 5500
*3
Possible with UST-9126, UST-9119, UST-5284-2.5, and UST-5297. sold in EU member countries.

CQ (Cardiac Quantification), A-SMA (Automated Segmental Motion Analysis) & KI (Kinetic Imaging):
EU-9078BQ
(Physiological signal display unit PEU-5500 is also necessary.)

Volume Mode :EU-9079CQ *2 NS: Non-sterilized, S: Sterilized


Volume Mode probe: ASU-1001 NOTE: Sterilized probe cover cannot be
sold in EU member countries.
Volume Mode probe: ASU-1002 Probe cover*2:
(Transvaginal type) RB-945BP-NS
RB-945BP-S

Probe holder for ASU-1000C-3.5: MP-PH5500-1 Volume Mode probe:


Cable hanger for ASU-1000C-3.5: MP-HA5500-2 ASU 1000C 3 5

TDI (Tissue Doppler Imaging): Flow Profile and TDI measurements (B TDI Profile,
PM 5500 16 MF Profile, MF Trace, M TDI Wall Thickness):
PM-5500-18
Free-angular M-mode: PM-5500-17

Real-time Doppler Auto Trace: SOP-5500-6

Non-interlaced scanning: EU-9080Q

Optional Module/Unit ECG lead cord for disposable electrodes:

Physiological signal display unit (ECG, PCG, and Pulse wave): PEU-5500Q Pulse wave transducer:TY-307A
(Includes ECG lead cord L-CABLE-674, clip type electrodes SEC141, and PCG
microphone TP-06 )

DMS (Data Management Subsystem):


DMS-5500B PM-DMS-10*4: BETA (Backscattered Energy Temporal Analysis)
*4 Some other modification kits may necessary depending on
the version of the unit. Please ask your local distributor for
d t il
PM-DMS-15: OB/GYN report data management

PM-DMS-16: Stress Echo scoring and report data


t

Optional Recorders *5 SVBK-110 enables remote control of the VCR from


MO Disk Drive the operation panel of SSD-5000.
SVRM-100 is a wired remote controller for the VCR.

Video Cassette Recorder (SONY®): Optional items for the SONY VCR*5
NTSC: SVO-9500MD 34-pin interface (SONY): SVBK-110
PAL: SVO-9500MDP Remote controller (SONY): SVRM-100

B/W printer, Worldwide (available in Japan): UP-895MD/SYN (SONY®) Print paper: UPP-110HG
B/W printer (available outside Japan)
NTSC (100-120V): UP-895MD (SONY®)
PAL (200-240V): UP-895CE (SONY®)

Color NTSC/PAL (not CE marked), Worldwide voltage:


RGB cable: L-CABLE-342-20 printer UP-21MD(UC) (SONY®) Color print pack:
UPC-21S (small size)
Slide adapter: UPA-21SA NTSC/PAL (CE marked), Worldwide voltage: UPC-21L (larege size)
UP-21MD(CED) (SONY®)

Optional Probes (See next page onward.)

8
Optional Probes
HST (Hemispheric Sound Technology) probes
*1
Convex sector probes B: B-mode
PW: Pulsed Wave Doppler
Application Ultrasound Scanning Radius of Model F: Flow (Color flow)
(Feature) frequency (MHz) *1 angle curvature BF: Flow in B-mode
CW: Continuous wave Doppler
General abdomen, B: 3.0/3.75/5.0/6.0 60° 60 mmR UST-9126 T.H.E.: Harmonic Echo
OB/GYN PW/F: 2.5/3.0/3.75/5.0 C.H.E.: Harmonic Echo
(Harmonic Echo) T.H.E.: 2.14
C.H.E.: B:2.14/2.5/1.88,
BF:1.88/2.14
General abdomen, B: 3.0/3.75/5.0/6.0 60° 60 mmR UST-9114-3.5
OB/GYN PW/F: 2.5/3.0/3.75/5.0
General abdomen, B: 3.75/5.0/6.0/7.5 60° 60 mmR UST-9115-5
OB/GYN (High resolution) PW/F:3.0/3.75/5.0/6.0 Puncture adapter:
MP-2473
General abdomen, B: 2.5/3.0/3.75/5.0 60° 60 mmR UST-9119
OB/GYN, PW/F: 2.14/2.5/3.0/3.75
Harmonic Echo T.H.E.: 2.14 Puncture
C.H.E.: B:2.14/2.5/1.88 adapter:
BF:1.88/2.14
Probe cover*2:
Endocavity B: 3.75/5.0/6.0/7.5 180° 9 mmR UST-9118 RB-945BP-NS
(Transvaginal) PW/F: 3.0/3.75/5.0/6.0 RB-945BP-S
Small part, Neonatal B: 5.0/6.0/7.5/10.0 70° 20 mmR UST-9120 Puncture adapter:
head PW/F:5.0/6.0/7.5 MP-2458
Endocavity B: 3.75/5.0/6.0/7.5 180° 9 mmR UST-675P Probe cover*2:
(Transrectal) PW/F: 3.0/3.75/5.0/6.0 RB-665P-NS
RB-665P-S
Linear probes*3 *3 Steered linear scanning is possible. Puncture adapter MP-2452
is attached as standard.
*2 NS: Non-sterilized,
Application Ultrasound Scanning Model S: Sterilized
(Feature) frequency (MHz) *1 width NOTE: Sterilized probe cover cannot
be sold in EU member countries.
Small parts, peripheral B: 5.0/6.0/7.5/10.0 34 mm UST-5539-7.5
vascular (Steered linear) PW/F: 5.0/6.0/7.5 Echo coupler (stand-off)
Flat type: ECM-12
Small parts, peripheral B: 5.0/7.5/10.0/13.0 34 mm UST-5545 Slant type: ECM-13
vascular (Steered linear) PW/F: 5.0/6.0/7.5
T.H.E.:3.75/5.0
Small parts, peripheral B: 3.75/5.0/6.0/7.5 42 mm UST-5548
vascular PW/F: 3.0/3.75/5.0/6.0
T.H.E.: 3.0
Puncture adapter: MP-2456
Small parts B: 5.0/6.0/7.5/10.0 60 mm UST-5712
PW/F: 5.0/6.0/7.5 Echo coupler (stand-off ):
MP-2463
Intra-operative B: 5.0/6.0/7.5/10.0 60 mm UST-5713T
PW/F: 5.0/6.0/7.5
Superficial tissue B: 7.5/10.0/13.0 34 mm UST-5543
PW/F: 5.0/6.0/7.5

Phased array sector probes


Application Ultrasound Scanning Model Number
(Feature) frequency (MHz) *1 angle

Cardiology B: 2.14/2.5/3.0/3.75 90° UST-5284-2.5


(Harmonic Echo) PW/F: 2.14/2.5/3.0/3.75, CW: 2.14
T.H.E.: 1.88
C.H.E.: B:1.88/2.14/2.5, BF: 1.88/2.5
Cardiology B: 2.14/2.5/3.0/3.75 90° UST-5297
(Harmonic Echo) PW/F: 2.14/2.5/3.0/3.75, CW: 2.14
T.H.E.: 1.88
C.H.E.: B:1.88/2.14/2.5, BF: 1.88/2.5
Pediatric heart B: 3.75/5.0/6.0/7.5 90°
PW/F: 3.0/3.75/5.0/6.0, CW: 3.75 UST-52108
T.H.E.:3.0
Cardiology B: 2.5/3.0/3.75/5.0 90° UST-5285-3.5
PW/F: 2.5/3.0/3.75/5.0, CW: 3.0
Transesophageal exam B: 3.75/5.0/6.0/7.5 90° UST-5293-5
PW/F: 3.0/3.75/5.0/6.0, CW: 3.75
Pediatric heart B: 3.75/5.0/6.0/7.5 90° UST-5294-5
PW/F: 3.0/3.75/5.0/6.0, CW: 3.75
Neonatal heart B: 5.0/6.0/7.5/10.0 90° UST-5296
PW/F:5.0/6.0/7.5, CW: 3.75

9
W-SHD (Wide-band Super High Density) probes and SHD probes
Convex sector probes
Application Ultrasound frequency Scanning Radius of Model
(MHz) angle curvature
Abdomen B: 2.14/2.5/3.0/3.75 60° 60 UST-988-2.5 Puncture adapter: MP-2486
(High PW/F: 2.14/2.5/3.0
penetration) mmR
Puncture adapter:
General B: 3.0/3.75/5.0/6.0 60° 60 UST-979-3.5 MP-2473
abdomen, PW/F: 2.5/3.0/3.75/5.0
OB/GYN mmR
General B: 3.75/5.0/6.0/7.5 60° 60 UST-990-5 Puncture adapter: MP-2414C
abdomen, PW/F: 3.0/3.75/5.0/6.0
OB/GYN mmR
Puncture adapter: MP-2482
Abdomen B: 3.0/3.75/5.0/6.0 90° 20 UST-9102-3.5
(intercostal scanning) PW/F: 2.5/3.0/3.75/5.0
mmR *1 This probe is not CE marked and
cannot be sold in EU member
Abdomen B: 5.0/6.0/7.5/10.0 60° 40 UST-9101-7.5
(High resolution) PW/F: 5.0/6.0/7.5
mmR
General B: 3.75/5.0/6.0/7.5 60° 40 UST-992-5*1 Puncture adapter: MP-2474
abdomen, PW/F: 3.0/3.75/5.0/6.0
OB/GYN mmR *2 Includes puncture adapter MP-2480.
Abdomen B: 3.0/3.75/5.0/6.0 120° 14 UST-9121 Puncture adapter set:
(intercostal PW/F:2.5/3.0/3.75/5.0
mmR MP-2445-SET
scanning)
Abdomen B: 3.0/3.75/5.0/6.0 Probe cover:
90° 14 UST-978-3.5 RB-945BP-S (Sterilized)
(intercostal) PW/F:2.5/3.0/3.75/5.0
mmR RB-945BP-NS (Nonsterilized)
Puncture & biopsy B: 3.0/3.75/5.0/6.0 60° 60 UST-9113P-3.5 *2 NOTE: Sterilized probe cover cannot
PW/F: 2.5/3.0/3.75/5.0 be sold in EU member countries.
mmR Puncture adapter:
Endo-cavity B:3.75/5.0/6.0/7.5 MP-2458
120° 14 UST-984-5
(Transvaginal) PW/F:3.0/3.75/5.0/6.0
mmR *3 Includes puncture adapter MP-2452.
Endo-cavity B:3.75/5.0/6.0/7.5 120° 14 UST-9112-5 Probe cover:
(Transvaginal) PW/F:3.0/3.75/5.0/6.0 RB-665P-S (Sterilized)
R RB-665P-NS (Nonsterilized)
Linear probes NOTE: Sterilized probe cover cannot
be sold in EU member countries.
Application Ultrasound frequency Scanning Model Number
(MHz) width
General abdomen, B: 3.0/3.75/5.0/6.0 96 mm UST-5046-3.5 Echo coupler (Disposable): ECM-14
OB/GYN PW/F: 2.5/3.0/3.75/5.0
Puncture & B: 5.0/6.0/7.5/10 28 mm UST-5540P-7.5 Guide tube:
biopsy*5 PW/F: 5.0/6.0/7.5 G18×45NS (Nonsterilized)
(superficial tissues)
*4 Includes puncture adapter MP-2416.
Puncture & biopsy B: 3.0/3.75/5.0/6.0 80 mm UST-5045P-3.5*
4
PW/F: 2.5/3.0/3.75/5.0
Small part B: 3.75/5.0/6.0/7.5 60 mm UST-5818-5 Water path: MP-2463
PW/F: 3.0/3.75/5.0/6.0
Puncture adapter
Small part B: 5.0/6.0/7.5/10.0 60 mm UST-5710-7.5 MP-2456
PW/F: 5.0/6.0/7.5
Echo coupler (Flat) : ECM-12
Peripheral B: 3.75/5.0/6.0/7.5 42 mm UST-5524-5
vessels*5 PW/F: 3.0/3.75/5.0/6.0
Echo coupler (Angled):
Peripheral B: 5.0/6.0/7.5/10.0 42 mm UST-5524-7.5
vessels*5 PW/F: 5.0/6.0/7.5
Intraoperative B: 3.75/5.0/6.0/7.5 61 mm UST-5819T-5
PW/F: 3.0/3.75/5.0/6.0 Puncture adapter MP-2448
Intraoperative B: 5.0/6.0/7.5/10.0 60 mm UST-579T-7.5
PW/F: 5.0/6.0/7.5
Intra-operative*5 B: 5.0/6.0/7.5/10.0 42 mm UST-5534T-7.5
PW/F: 5.0/6.0/7.5 6
Trocar—adaptable size: 10 mm inner dia.*
Laparoscopic probe B: 5.0/6.0/7.5/10.0 33 mm UST-5526L-7.5
(Rigid type) PW/F: 5.0/6.0/7.5 Trocar—adaptable size: 12 mm inner dia.*
6

Laparoscopic probe B: 5.0/6.0/7.5/10.0 33 mm UST-5536-7.5


(Flexible type) PW/F: 5.0/6.0/7.5
Handling tool T type: MP-2749
Microsurgery*5 B: 5.0/6.0/7.5/10.0 10 mm UST-5531 Handling tool I type: MP-2750
5 PW/F: 5 0/6 0/7 5
* Steered linear scanning is possible.
*6 Trocars of some manufacturers and models may not be physically adaptable to this probe. Trocars of some manufacturers
and models may damage the probe. For adaptability of the trocar, please contact your local distributor or Aloka office.

Puncture adapter MP-2451


Convex/Linear probe Probe cover:
Endo-cavity Convex sector: 120 deg./9 Linear: 60 mm BL-664-S (Sterilized)
(Transrectal) mmR B: 5.0/6.0/7.5/10.0 UST-672-5/7.5 BL-664-NS (Nonsterilized)
B: 3.75/5.0/6.0/7.5 PW/F: 5.0/6.0/7.5 Elastic band: FS5/16
Ultra-StepTM
NOTE: Sterilized probe cover cannot Stepping device and needle
be sold in EU member countries. guidance template (CIVCO)

10
SHD (Super High Density) probes

Phased Array probes


Application Ultrasound frequency (MHz) Scanning angle Model
Rotary-plane T.E.E. B: 3.75/5.0/6.0/7.5 90° UST-5280-5
PW/F: 3.0/3.75/5.0/6.0
CW: 3.75
Neonatal heart B: 3.75/5.0/6.0/7.5 90° UST-5281-5
PW/F: 3.0/3.75/5.0/6.0
CW: 3.75
Neuro surgery B: 3.75/5.0/6.0/7.5 90° UST-5268P-5*4 *4 Includes puncture adapter MP-2450.
PW/F: 3.0/3.75/5.0/6.0
Bi-plane T.E.E. for B: 3.75/5.0/6.0/7.5 90° UST-5271S-5
pediatrics PW/F: 3.0/3.75/5.0/6.0
Adult heart (High B: 2.14/2.5/3/3.75 90° UST-5283-2.5*1 Puncture adapter:
PW/F: 2.14/2.5/3/3.75 MP-2475
sensitivity)
CW: 2.14
*1 This probe is not CE marked and
cannot be sold in EU member

Independent probe connection kit:


EU-3040

Indepedent Probes (EU-3040 is necessary to connect the independent probes.)


Application Ultrasound frequency Model
CW Doppler (for heart) CW: 2.14 MHz UST-2265-2
CW Doppler (for peripheral vessels) CW: 5.0 MHz UST-2266-5

Mechanical probe connection unit:


SCU-5500Q

ANNULAR ARRAY MECHANICAL SECTOR PROBE Puncture adapter:


Small parts B: 10.0 MHz 40° ASU-36WL-10 MP-2493
MECHANICAL RADIAL PROBES
Transrectal (prostate) B: 7.5/10.0 MHz 360° ASU-67 Puncture adaptor MP-2438
Transrectal (prostate) B: 5.0/7.5 MHz 360° ASU-64B
Transurethral (bladder) B: 7.5 MHz 360° ASU-65B Outer sheath adapter
for STORZ #27040B MP-2421
for Olimpus A3128, A2101 MP-2422
for A.C.M.I. #8414 MP-2423
for Wolf #8654.021 MP-2424

Drive unit for


Microtip Transducers (ASU-101 and sheath are necessary)
Microtip transducer:
Application*3 Effective Ultrasound Scanning Model
ASU-101 length frequency angle
Intracavity 193 cm B: 10 MHz 360° MP-PN10-08L Sheath:
ST 08LV
Intracavity 137 cm B: 10 MHz 360° MP-PN10-08M Sheath:
ST 08MV
*3 The Microtip transducers are for gastroenterology use only.
Not for cardiovascular use (e.g. intravascular and heart cavity).

11
・ The specifications are subject to change without notice.

・ The standard components and optional items depend on the country.


Please contact your local ALOKA distributors for details.

・ ProSound is a registered trademark of ALOKA CO., LTD.

・ Harmonic Echo is a trademark of ALOKA CO., LTD.

We care, Ultrasound@Aloka

6-22-1 Mure, Mitaka-shi, Tokyo 181-8622 Japan


Telephone: +81 422 45 6049, Facsimile: +81 422 45 4058
URL: http://www.aloka.com/

Printed in Japan, 2002-11


SP-5000-V4.1-2.0

Das könnte Ihnen auch gefallen