Sie sind auf Seite 1von 6

Dysphagia

Dysphagia 9:63-68 (1994)

9 Springer-VerlagNewYorkInc, 1994

Stethoscope Acoustics and Cervical Auscultation of Swallowing


Sandra Hamlet, PhD,1 David G. Penney, PhD, 2'3 and John Formolo, MD 3
Departments of 1Otolaryngology and 2Physiology, School of Medicine, Wayne State University, Detroit, Michigan; and
Department of Internal Medicine, 3Division of Cardiology, St. John Hospital, Detroit, Michigan, USA

Abstract. Frequency response characteristics of six listening to recordings of swallowing sounds, that they
popular stethoscopes are reported for the higher fre- do not sound quite like what is heard with a stethoscope
quency range (to 3000 Hz) to supplement equivalent [15]. If the results of such research studies are to be
measurements for the lower frequencies (35-1000 Hz) translated into clinical practice, it is important to take into
published previously. Spectra of the sounds of swallow- account the characteristics of the different instruments.
ing from the throat, transduced with an accelerometer, The literature contains limited description of the
demonstrate important frequency composition in this spectrum of swallow sounds. Logan et al. [11] published
higher range. Two stethoscope models were found to spectrograms of liquid swallow sounds. Swallow sounds
have superior transmission characteristics for use in cer- are broad band, containing transient clicks, with energy
vical auscultation of swallowing sounds. up to 8,000 Hz. It can be appreciated from the illustra-
tions, however, that the greatest intensity is in the fre-
Key words: Sound of swallowing - - Stethoscope - -
quency range below 3,000 Hz, and that clicks with fre-
Auscultation - - Deglutition - - Deglutition disorders.
quency components higher than this contain energy in the
low frequencies also. Thus, there would be no loss of
temporal detail from filtering out the higher frequencies.
Cervical auscultation (listening at the throat with a Mackowiak et al. [10] described the spectrum of a "dry"
stethoscope) is being adopted by dysphagia clinicians as swallow as having peak energy at 1,000-1,100 Hz for the
a component of clinical evaluation of swallowing [1-9]. entire duration of the swallowing sound. Wet swallow
As in any observational method, the quality of informa- signals began with a low frequency component (400-600
tion obtained depends not only on the perceptual skills of Hz), then increased in intensity and frequency to a peak
the examiner but on utilization of appropriate instrumen- at 1,000 Hz at the end of the signal. They used a contact
tation and technique. Even a seemingly simple procedure microphone over the larynx lateral to the midline. Heinz
such as listening with a stethoscope contains a source of et al. [4] and Vice et al. [6] described swallow sounds
instrumental distortion. That is, the stethoscope is not recorded from infants, portions of which were periodic
completely faithful in picking up and transmitting to the with a high fundamental frequency. Hamlet et al. [ 12,13]
listener's ear the sound quality present at the body sur- published illustrations of a spectral change within a por-
face. tion of the sound of swallowing corresponding to fluid
Research studies on the acoustics of swallowing flow through the upper esophageal sphincter. Spectral
have generally recorded the sounds using a microphone information of this type is contained in the frequency
or accelerometer [5-7,9-14]. These transducers are de- region below 3,000 Hz.
signed to have a flat frequency response over a wide The acoustic spectrum of breath sounds following
range of frequencies. Clinicians have remarked, when swallowing were illustrated by Logan et al. [ 11]. Great-
est intensity is in the lower frequency range, below 1,500
Hz. Similarly, sounds of value in the diagnosis of heart
and lung disease by thoracic auscultation fall in the low
Address offprint requests to: David G. Penney, Ph.D., Professor, De- frequency range 20-1,000 Hz [16,17].
partment of Physiology and Occupational and Environmental Health,
Wayne State University School of Medicine, 540 E. Canfield, Detroit, A review of the acoustics of stethoscopes, and
M148201, USA frequency response data for several currently available
64 S. Hamlet et al.: Stethoscope Acoustics

Table 1. Six stethoscopes evaluated for frequency response characteristics in their various modes

Diaphragms Bells
Name Abbreviation Name Abbreviation
Littmann Classic II Classic II Littmann Classic II Classic II
Littmann Master Cardiology Mast HW Littmann Master Cardiology Mast LW
Littmann Cardiology II Cardiol II Littmann Cardiology II Cardiol II
Hewlett-Packard, Rappaport-Sprague HP-L Hewlett Packard, Rappaport- HP-L
large diaphragm~ Sprague, large bell
Hewlett-Packard, Rappaport-Sprague HP-S Hewlett-Packard, Rappaport- HP-M
small diaphragm b Sprague, medium bell
Tycos Harvey triple head, smooth Harv S Tycos Harvey triple head, bell Harv T
diaphragm
Tycos Harvey triple head, ribbed Harv R
diaphragm
Allen Medical Series 5-RPS, long tube Allen LT Alien Medical Series 5-RPS, long tube Allen LT
Allen Medical Series 5-RPS, short tube Allen ST
"Adult diaphragm, i3/8inch bell.
bPediatric diaphragm, 1 inch bell.

stethoscopes, have been published recently [18]. This miniature accelerometer (Vibro-meter model 501M601, frequency re-
w o r k only considered f r e q u e n c y response characteristics sponse 2~-0,000 Hz - 3 dB). The accelerometer was attached to one
side of the throat with double-sided tape, just above the larynx about 2
in the range 3 5 - 1 , 0 0 0 Hz, focusing on application to
cm below the angle of the mandible. A mandible rather than laryngeal
auscultation o f heart and lung sounds. All stethoscopes reference was used, because the larynx moves upward and forward
tested responded best in the v e r y low frequencies (be- during swallowing. The above location of the accelerometer was in-
low 100 Hz). T h e response then falls o f f with increas- tended to result in placement approximately over the hypopharynx and
ing f r e q u e n c y a b o v e 100 Hz, being d o w n 1 0 - 2 0 dB at posterior larynx when the larynx was elevated as the bolus was passing
through the pharynx.
1,000 Hz. Accelerometer recordings were made simultaneously with a
F r e q u e n c y response or acoustic transfer function videotape of a modified barium swallow on one audio channel of a
refers to the d e g r e e to which an instrument will process Panasonic AG-6300 video recorder (frequency response 50-12,000
all frequency c o m p o n e n t s o f a sound equally, without Hz). The simultaneous videotape of the modified barium swallow con-
e n h a n c e m e n t or suppression o f particular f r e q u e n c y re- firmed that the recorded event was a normal swallow.
The recording was played back from the line audio output of the
gions. Ideally, for faithful reproduction, a uniform or flat
same videorecorder, and bandpass filtered (80--4000 Hz). A/D conver-
frequency response should exist for the f r e q u e n c y range sion (10,000 Hz sampling rate) and spectral analysis was performed
o f importance in the sound to be studied. H o w e v e r , if a using Micro Speech Lab software on an IBM PC. A 100 ms time
particular f r e q u e n c y range in the sound is o f clinical window was located over three nonoverlapping portions of each swal-
interest, then a peak in the f r e q u e n c y response s p e c t r u m low signal (1) in the initial low amplitude portion, (2) centered on the
peak amplitude, (3) in the later low amplitude portion, and (4) from a
corresponding to that selected range will p r o v i d e s o m e
silent portion of the recording after the swallow for a comparison to the
amplification. noise floor present in the recording. An illustration of the time wave-
T h e r e are two issues addressed in this paper: (1) form of the swallow signal and selected spectra can be found in Hamlet
H o w well does the f r e q u e n c y response o f a stethoscope et al. [12,13]. Fast Fourier Transform analysis was performed with
e n c o m p a s s the frequencies present in the sound o f swal- "low" smoothing, which is a running average of four sampling inter-
vals. Smoothing reduces the variability in amplitude of spectral compo-
lowing? (2) A r e there significant differences in f r e q u e n c y nents between adjacent frequency locations, yielding a spectral display
response characteristics a m o n g currently available more representative of the overall contour. Using a cursor, the ampli-
stethoscopes, w h i c h w o u l d f a v o r particular types for use tude value corresponding to particular frequencies was read off the
in cervical auscultation of s w a l l o w i n g ? display. The frequencies chosen were every 200 Hz from 200-4000 Hz.
Composite average spectra across subjects were calculated from these
amplitude measurements.

Methods
Frequency Response of Stethoscopes
Spectrum of Swallow Sounds
The acoustic transfer function (frequency response) of six stethoscopes
High fidelity recordings were made of the sounds of swallowing from was measured by Western Electro-Acoustics Laboratory, Inc., Santa
the throat of 10 normal adults, 5 females and 5 males, between the ages Monica, CA (Table 1). The stethoscopes were tested in both their bell
of 34 and 67. One swallow of 10 ml thin barium suspension (40% w/w) and diaphragm modes. A white noise sound source was generated by an
was analyzed from each subject. The sounds were transduced by a electrodynamic head phone mounted in a 17 cubic cm coupler. The
S. Hamletet al.: StethoscopeAcoustics 65

Judgment of the suitability of particular stetho-


I ~ Max. Amplitude ]
[--C~ First Component / scope models for auscultation of lowfrequency swallow-
-10 I + Third Component /
ing sounds relied on the existence of amplification for
LU
~Z -20- very low frequencies, and flatness or uniformity of fre-
0
o. quency response in the lower and middle frequencies. In
LU -30" identifying particular stethoscope models as best suited
>UJ for faithful transmission of low frequency signals, the
-40 -
following criteria were chosen, to be met cumulatively in
sequence:

-6( r T
1. Amplification of sound below 100 Hz
1000 2000 3000 4000 2. Attenuation <10 dB in the range 100-700 Hz, with
Frequency (Hz)
variability confined between 0-10 dB
Fig. 1. Averagedspectralcharacteristicsof swallow signals recorded 3. Attenuation <15 dB in the range 700-1000 Hz, with
while swallowing 10 rnl of thin liquid barium. variability confined between 5 and 15 dB.

Evaluation of stethoscopes in the higher frequen-


stethoscope chest piece was placed on the coupler. Both earpieces of the cies rested on relative lack of attenuation in the high
stethoscope were placed at the entrance to artificial ears. Microphone frequencies and flatness or uniformity of frequency re-
systems, at the source coupler and the artificial ear, were used to detect sponse in the middle frequencies. The following criteria
the input and received sound for comparison. were chosen, to be met cumulatively in sequence:
Detailed frequency response data for these stethoscopes in the
low frequencies have been published [ 18]. We are providing summaries
of salient features of those results for low frequencies, and adding 4. Attenuation <30 dB in the range 1800-3000 Hz
information on frequency response characteristics in the higher range
5. Attenuation <20 dB in the range 1000-1800 Hz
up to 3000 Hz.
6. Attenuation <15 dB in the range 700-1000 Hz, with
variability confined between 5-15 dB.
Results

Bell Mode
Spectrum of Swallow Sounds Bell chestpieces are most suitable for transmitting very
low frequency sounds [18]. Criterion 1 was met by all of
The averaged spectra (Fig. 1) illustrate the frequencies
the stethoscope models tested in the bell mode. Criterion
prominent in the sound of swallowing when transduced
2 was met by the Littman Cardiology II, the Hewlett-
by an aecelerometer and recorded on a tape recorder.
Packard Rappaport-Sprague large and medium bell, and
Individual differences and detail in unsmoothed spectra
the Tycos-Harvey triple head bell. Criterion 3 was not
have been deemphasized so as to provide a general over-
met by the Tycos-Harvey triple head bell.
all indication of the characteristics of the signal. The
High frequency response in the bell mode was
highest amplitude components in the frequency spectrum
also evaluated. Criterion 4 was met by the Littman Cardi-
of the acoustic swallow signal were in the low frequen-
ology I! and Classic II, Hewlett-Packard Rappaport-
cies. Above 2200 Hz the averaged spectrum of the third
Sprague large + medium bell and Tycos-Harvey triple
component overlapped the noise floor, thus spectral com-
head bell. Criterion 5 was not met by the Tycos-Harvey
position for this component was limited to below 2200
triple head bell. Only the Littman Cardiology II and
Hz. Above 3000 Hz the spectra for the first two compo-
HP-M met criterion 6.
nents approximated the noise floor (about 5 dB above).
Figure 2 shows the bell mode frequency response
Thus, in general, significant spectral composition for
data in the range 900-3000 Hz. Detailed bell mode low
the recorded liquid bolus swallowing sounds was below
frequency response data for all the stethoscopes have been
3000 Hz.
published previously [18]. The stethoscopes not meeting
our performance criteria for bell mode failed because of
Frequency Response of Stethoscopes extreme variabifity in response (resonances), and response
curves exceeding criterion levels for attenuation.
Empirical criteria were chosen for evaluating frequency
response characteristics of stethoscopes. The criteria
were intended to separate the tested stethoscopes into Diaphragm Mode
"better" or "poorer" categories according to particular Most stethoscopes do not amplify the very lowest fre-
response characteristics. quencies when used in the diaphragm mode [18]. Low
66 S. Hamletet al.: StethoscopeAcoustics

B(~II~ ~aohragms

-10 -10 e, 9 ...... / \


LU
G9 LU
Z '7
O
e~ O
0,..
-20 ~n- -20

~ -30 .j -30 --0-- Cl~,~ie II ......9 ........... 9


m
w
I--~-_ Cardiol "I
l"""* ...... HP-L
l-" J
2'Oo ~8oo
...... " P - '
-4C -40
800 2o'oo 2,'00 23'o0 3200 800 1200 1600 2000 2400 2800 3200

0*

-I0 ILl -10"


LU
~0 03
Z
0 - i
O.
-20" w -20'
........= ............. i' ...... 9
....I ,

"1I.-- ""I ..........9.............


w ~::sF-
"--.o..:~.
~ff\l ~ o '~~ . "-,,"

~ -30 HP-M . '""" I ~ -30


W .........
9 ........

& HaW R ~ '';~''"" "''7',


........O""" Allen LT ~
I'--O-- Alien LT I -"o--~ o .......-Z~...-- Allen ST . ,
-4O -40
800 12'O0 16'O0 2000 24'00 28'O0 3200 800 1200 1800 2000 2400 2800 3200
F R E Q U E N C Y (Hz) F R E Q U E N C Y (Hz)

Fig. 2. Frequencyresponse of stethoscopebells over the range 900- Fig. 3. Frequencyresponseof stethoscopediaphragmsover the range
3000 Hz. Zero dB is the reference input signal level, which was the 900-3000 Hz. ZerodB is the referenceinputsignallevel, whichwas the
same for all stethoscopes. same for all stethoscopes.

frequency response in the diaphragm mode met criterion appeared to be superior: the Littman Cardiology II, and
1 by Littman Classic II, Cardiology II and Master Cardi- the Hewlett-Packard Rappaport-Sprague (medium bell,
ology. Criteria 2 and 3 were met only by Littman Cardi- small diaphragm). Figure 4 provides detailed low fre-
ology II. quency response data for these models, previously pub-
For higher frequency response, stethoscopes lished [18].
meeting Criterion 4 were the Littman Cardiology II and
Classic II, Hewlett-Packard Rappaport-Sprague small di-
aphragm, Allen LT, and the Tycos-Harvey triple head Discussion
smooth diaphragm. Criterion 5 was met by the Hewlett-
Packard Rappaport-Sprague small diaphragm, and the When clinicians listen to the sounds of swallowing with a
Littman Cardiology II. Criterion 6 was only met by the stethoscope, qualitative judgments are made about what
Littman Cardiology II. is heard. The characteristic swishing double-click as the
Figure 3 shows the diaphragm mode frequency bolus passes through the pharynx and into the esophagus
response data in the range 900-3000 Hz. Detailed dia- is judged for normalcy based on "crispness" of the sound
phragm mode low frequency response data for all the [15]. A definition of crispness has not been provided in
stethoscopes have been published previously [18]. physical terms, but in general, sharp, crackling, or bril-
Stethoscopes not meeting the criteria for diaphragm liant sound qualities have high frequency components.
mode performance failed mainly because of greater high In comparison to an accelerometer or micro-
frequency attenuation. phone, any stethoscope is poorly suited for transmitting
high frequencies. Even with the two stethoscopes we
Models Best Meeting Criteria considered most suitable for listening to swallow sounds,
For overall performance according to our criteria relative the higher frequencies were markedly attenuated. Stetho-
to auscultation of swallowing sounds, two stethoscopes scopes are best suited for transmitting extremely low
S. Hamlet et al.: Stethoscope Acoustics 67

listening conditions. It is worth being aware of the differ-


1 Bells Ill
ences in acoustic capabilities of a stethoscope in the bell
and diaphragm modes.
In summary, two stethoscopes were found to have
O~ -10 superior overall acoustical performance for use in the
-15 cervical auscultation of swallowing sounds based on a
200 400 600 800 1000 1200

~
LU
15" number of criteria: the Littman Cardiology II and the
~> 10-
~ol II
Hewlett-Packard Rappaport-Sprague (medium bell,
,; 5
small diaphragm). The latter stethoscope is intended for
er 0
-5
pediatric use, but could be utilized for adult cervical
-10 auscultation also.
-15
-20
-25
2;0 400 600 800 10'00 1200 Acknowledgments. This study was supported by a grant from the GME
F R E Q U E N C Y (Hz)
Biomedical Investigation Committee of St. John Hospital and Medical
Center, and in part by Grant CA-43838 from the National Institutes of
Fig. 4. Low frequency response characteristicsof two selected stetho-
Health. All stethoscopeswere purchased by the Research Committeeof
scopes in the bell and diaphragm modes. Zero dB is the reference input
St. John Hospital and Medical Center through standard medical supply
signal level, which was the same for all stethoscopes.
companies, and no input or monetary support was obtained from stetho-
scope manufacturers or suppliers.

frequency, dull pulsing sounds, with frequency composi-


References
tion below 1000 Hz.
In the very low frequency range (below 100 Hz), 1. Stott FD: The laryngeal microphone as an aid to treatment of
some stethoscopes actually provide a small amount of bulba poliomyelitis. Br Med 20:1416, 1953
amplification. The particular accelerometer used here has 2. RusselWR: Poliomyelitis, London: Arnold, 1956
3. BostonJF: Sensorimotorexaminationof the mouth and pharynx.
a fiat frequency response over an extremely wide r a n g e - -
In: Kawamura Y (ed.): Frontiers of Oral Physiology. Basel:
2 - 4 0 , 0 0 0 Hz -+ 3 dB. Certain types of microphones also Karger, 1976, pp 78-107
respond to very low frequencies. The limiting factor in 4. Heinz JM, Giuriati MA, Bosma JF, Ansel B, Hood M: Swallow
the very low frequencies is more likely to be the nature of sounds: cervical auscultation. Presented at the Academy of Ce-
the recorder used (audio tape recorder), or the character- rebral Palsy and Developmental Medicine, New Orleans, Oct
15-18, 1986
istics of input filtering prior to A/D conversion for a
5. Lebel D, Parel CL, Thouvenot J: Exploration de la deglutition a
computer. partir de son signal sonore. Arch lnt Physiol Biochem 98:75-86,
The human ear is sensitive in the 2000-3000 Hz 1990
region [19], and the auditory perceptual system is forgiv- 6. ViceFL, Heinz JM, Giuriati G, Hood M, Bosma JF: Cervical
ing of sound distortion. For these reasons it may not be auscultation of suckle feeding in newborn infants. Dev Med
Child Neuro132 :760-768, 1990
obvious to the listener when an instrument filters the
7. DempseyJE, Vice FL, Bosma JF: Combination of cervical
higher frequencies---especially when there is no standard auscultationand videoradiographyin evaluation of oral and pha-
against which to make a comparison. When listening for ryngeal dysphagia. Presented at the Symposium on Dysphagia,
the crispness of the sound of swallowing, it is important Johns Hopkins Hospital, Baltimore, Maryland, March 1990
to make sure that the stethoscope used is maximally capa- 8. Selley WG, Hon FDS, Flack FC, Ellis RE, Brooks WA: The
Exeter dysphagia assessment technique. Dysphagia 4:227-235,
ble o f transmitting the higher frequency range.
1990
Another auditory judgment made by dysphagia 9. Zenner PM: Clinical examination with cervical auscultation de-
clinicians relates to the breath sounds immediately after a tecting aspiration. Presented at the Cervical Auscultation Work-
swallow. If aspiration or laryngeal penetration o f the shop, Department of Pediatrics, University of Maryland, Balti-
bolus has occurred and there is residue, subtle gurgling or more, Maryland, April 22, 1992
10. MackowiakRC, Brenman HS, Friedman MHF: Acoustic profile
bubbling may be heard. W e did not analyze such sounds,
of deglutition. Proc Soc Exp Biol Med 125:1149-1152, 1967
not having examples available, so it is only by analogy to 11. LoganWJ, Kavanagh J: Deglutition. J Appl Physiol 23:279-
the low frequency range o f breath sounds [11] that we 284, 1967
assume these abnormal respiration sounds are low fre- 12. HamletAL, Nelson RJ, Patterson RL: Interpretingthe sounds of
quency also. Below 1000 Hz the human threshold of swallowing: fluid flow through the cricopharyngeus. Ann Otol
Rhinol Laryngo199:749-752, 1990
audibility begins to drop off with decreasing frequency
13. Hamlet SL, Patterson RL, Fleming SM, Jones LA: Sounds of
[19]. If low frequency sounds are of special interest, swallowing following total laryngectomy. Dysphagia 7:160-
using the bell mode of a stethoscope may provide the best 165, 1992
68 S. Hamlet et al.: Stethoscope Acoustics

14. Takahashi K, Groher ME, Michi K: Methodology for detecting 17. Hollins PJ: The stethoscope: some facts and fallacies. J Hosp
swallowing sounds. Dysphagia 9:53-61, 1994 Med 5 :509-516, 197l
15. Proceedings of the Cervical Auscultation Workshop, Department 18. Abella M, Formolo J, Penney DG: Comparison of the acoustic
of Pediatrics, University of Maryland, Baltimore, Maryland, April properties of six popular stethoscopes. J Acoust Soc Am
22, 1992 91:2224-2228, 1992
16. Dawson JB: Auscultation and the stethoscope. Practitioner 19. Yost WA, Nielsen DW: Fundamentals of Hearing. New York:
193:315-322, 1964 CBS College Publishing, 1985

Das könnte Ihnen auch gefallen